Biosafety Levels Specifications
Biosafety Levels Specifications
All penetrations are sealed “gas tight”. Supply air is HEPA-filtered and exhaust air passes through
two HEPA filters. Air flow is maintained by a dedicated exhaust system exterior to the cabinet,
which keeps the cabinet interior under negative pressure (about 124.5 Pa). Access to the work
surface is by means of heavy duty rubber gloves, which are attached to ports in the cabinet. The
Class III BSC should have an attached pass-through box that is sterilizable and is equipped with
HEPA-filtered exhaust. The Class III cabinet may be connected to a double-door autoclave used
to decontaminate all materials entering or exiting the cabinet. Several glove boxes can be joined
together to extend the work surface. Class III BSCs are suitable for work in Biosafety Level 3 and
4 laboratories.
Materials are passed into the cabinet through a dunk tank, that is accessible through the cabinet
floor, or double-door pass-through box (e.g., autoclave) that can be decontaminated between
uses. The same process is reversed to take items out of the cabinet. Both supply and exhaust air
are HEPA filtered on a Class III cabinet and the exhaust air must pass through two HEPA filters, or
a HEPA filter and an air incinerator, before discharge to the outside.
Certification of hard-duct connected BSCs is more time-consuming than that for BSCs that
recirculate air to the room or which are thimble-connected.
Figure 1: Schematic representation of a class III biosafety cabinet (glove box) without dunk tank.
Figure 2: Schematic representation of a class III biosafety cabinet (glove box) with a dunk tank.
A- Glove ports with O-ring for attaching arm-length gloves to cabinet
B- Sash
C- Exhaust HEPA filter
D- Supply HEPA filter
E- Double-ended autoclave or pass-through box
F- chemical dunk tank
Note: A chemical dunk tank may be installed which would be located beneath the work surface of
the BSC with access from above. The cabinet exhaust needs to be connected to the building exhaust
system. Connection of cabinet exhaust to an independent building exhaust air system is required.
International Standards
Biological safety cabinets are generally required to comply with that country or region's standards.
International Standards are mainly issued by two organizations. International Standards Association
(ISO), based in Geneva, Switzerland, and the Committee Europeen de Normalisation (CEN) based in
Brussels, Belgium.
These two standards replace all other local standards on the same subject. The aim of the standards
is to construct a mechanism to provide manufacturer and user of a common site to specify technical
specifications and features.
In general terms, a regular service maintenance schedule may include the following tasks:
Airflow and filter capacities are verified. The filters have a limited lifespan, determined by the air
quality within the laboratory space, the amount of particles & aerosols generated inside the BSC
work zone, and the volume of air passing through the filters. As these filters load, the internal fan is
required to do more work to push/pull the same volume of air through them. Newer cabinets
measure the airflow constantly and self-compensate fan performance to ensure constant volumes of
air moving through the filters and the cabinet. However, self-adjusting cabinets should be cross-
verified by calibrated instruments to ensure that performance is maintained to the correct levels. If
the flow drops below desired performance, an audio and visual alarm will alert the operator.
Changing the filter should be limited to trained persons as the filter is potentially contaminated. This
can be done either after the cabinet has been decontaminated using a gaseous procedure (using
formaldehyde, chlorine dioxide, or vaporized hydrogen peroxide) or a "bag-in/bag-out" procedure.
UV lights are checked and changed. UV lights decrease in power over time, resulting in diminished
disinfection of the working area.
Summary of requirements
Biosafety Level (BSL) III:
Applicable where work is done with indigenous or exotic agents which may cause serious or
potentially lethal disease as a result of exposure by inhalation.
Figure 3: A CDC laboratory personnel dons an older-model positive-pressure suit before entering one
of the CDC's earlier BSL-4 labs.
In Zimbabwe, Ideal Medical Solutions, F6, First Floor, Travel Plaza. Cnr Mazoe & Chinamano Avenue.
Harare, Harare, Zimbabwe. And LabLad at
Online and probable suppliers