Technical Specification C86

The document provides specifications for the COMEN C86 patient monitor. It lists the standard and optional configurations, safety standards, physical characteristics, displays, interfaces, storage, alarms, and technical specifications for monitoring ECG, respiration, SpO2, NIBP, temperature and other parameters.

Uploaded by

Soporte Biosumma
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
157 views6 pages

Technical Specification C86

The document provides specifications for the COMEN C86 patient monitor. It lists the standard and optional configurations, safety standards, physical characteristics, displays, interfaces, storage, alarms, and technical specifications for monitoring ECG, respiration, SpO2, NIBP, temperature and other parameters.

Uploaded by

Soporte Biosumma
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 6

Digitally signed by Antonova Anna

Date: 2021.07.05 14:03:13 EEST


Reason: MoldSign Signature
Location: Moldova

Specification: C86

SHENZHEN COMEN MEDICAL INSTRUMENT CO.,LTD


No.2 of FIYATA Timepiece Building, Nanhuan Avenue, Gongming
sub-district, Guangming New District, Shenzhen, P.R. China
Tel: +86-755-26408879
Fax: +86-755-26431232
Email: infoatszcomen.com
Web: en.comen.com

2020 V 1.3
Patient Monitor
C86

Standard Configuration: QRS beep and alarm sound


5-lead ECG, RESP, Temp (Single Channel), COMEN SpO2, Operating key sound
NIBP, HR
Optional Configuration: Interface:
Dual-IBP, EtCO2, AG, ICG, C.O., BIS, Nellcor/Masimo SpO2, Parameter cable interface
3/12-lead ECG, Thermal Recorder, Dual-Temp, Suntech NIBP, AC power input socket
Trolley, Wall mount, Ground wire Two USB port
Safety Standards: RJ45 port
IEC 60601-1 IEC 60601-1-8 IEC 60601-2-27 EN 1060-3 IEC Optional Multi-functional interface
80601-2-30 IEC60601-2-34 IEC60601-2-49 ISO 80601-2-56 ISO VGA output
80601-2-61 Data storage
Physical Characteristics: Alarm Event Recall: 200 groups
Product Size: 344mm*291mm*165mm Wave Recall: 6 hours (8 waves)
Weight: 4.1kg NIBP Recall: 2000 groups
Display: 15” color TFT touch screen Trend Graph: 160 hours
Resolution: 1024*768 Trend Table: 160 hours
IP grade IPX1 Power-off Yes
Trace: 8 waveforms storage:
Operation Environment: Alarm: User-adjustable High and Low 3-level
Working Limits;
Temperature 5-40°C Prioritized audible and visual alarm
Humidity: ≤93% Network: Connected to Central Monitoring
Power Supply 100-240V~, 50/60Hz±1Hz System by hardwire/wireless
Battery Type: Rechargeable Lithium-ion battery Recorder:
Battery Capacity: 2200mAh (optional: 4400mAh) Type: Built-in; Thermal array
Battery Recharging Channel: 3 channel waveforms
Time: Maximum 5.5 hours for charging; Speed: 25mm/s,50mm/s
Battery backup: 2 hours for continuous working Record width: 50mm
Indicator: Real-time record
One alarm indicator time 8s, 16s, 32s or continual
Power indicator Alarm record Yes
Battery indicator Respiration:
Method: RA-LL Impedance Method ST SEGMENT
RR measurement detection: -2.0mV~+2.0mV (Automatic)
Range Adult: 0-120rpm Arrhythmia
Pediatric/Neonate:0-150rpm Analysis: 26 types
Accuracy: 7~150rpm: ±2rpm or 2%, whichever is Pacemaker
greater detection: Detectable
0-6rpm: unspecified Alarm: Yes, audible and visual alarm, alarm
Resolution: ±1 rpm events reviewable
RESP Apnea 10s-60s (Adu); 10s-40s (Ped/Neo) 12 lead ECG
Alarm: Audible and visual alarm; alarm events Analysis: Yes
reviewable NIBP:
Sweep Speed: 6.25,12.5,25mm/s Method Automatic oscillation
Gain Selection: X0.25, X0.5, X1, X2, X4 Work mode: Manual / Automatic/Continual (5min,
ECG: not applicable to neonates)
TM
Lead Type: CardioTec 5-leads ECG Analysis, 12- Measurement time: Adjustable(1-480min)
Lead and 3-leads electable Maximum
Lead selection 12-Lead I; II; III; aVR; aVL; aVF; V1-V6. measurement time Adu/Ped: 120s; Neo: 85s
5-lead: I; II; III; aVR; aVL; aVF; V Measurement Unit: mmHg / kpa selectable
3-lead: I; II; III Measurement
Waveform 5-lead: 2 -channel types: Systolic, Diastolic, Mean
3-lead: 1 -channel Range of systolic
Gain Selection: X0.125, X0.25, X0.5, X1, X2, X4, auto pressure: Adult Mode: 40-270mmHg
error<±5% Pediatric Mode: 40-200mmHg
Sweep Speed: 6.25,12.5, 25, 50mm/s, error≤±10% Neonate Mode: 40-135mmHg
Resp, lead Range of diastolic
disconnection pressure: Adult Mode: 10-215mmHg
detection and AC waveform: Pediatric Mode: 10-150mmHg
active noise Current :<0.1µA; Neonate Mode: 10-100mmHg
control: Frequency 64kHz, ±10% Range of mean
CMRR ≥105 dB pressure: Adult Mode:20-235mmHg
Heart Rate Pediatric Mode:20-165mmHg
measurement Neonate Mode 20-110mmHg
Range: Adult: 15~300bpm Static pressure
Pediatric/Neonate:15~350bpm range and accuracy:
Accuracy: ±1% or ±1bpm (whichever is greater) 0~300mmHg(0kPa~40.0kPa)
Protection: Withstand 4000VAC/50Hz voltage in ±3mmHg(±0.4kPa)
isolation, Again electrosurgical Over-pressure
interference and defibrillation protection: Adult Mode: 297mmHg
Accuracy: ±1% or ±1bpm (whichever is greater) Pediatric Mode: 240mmHg
Band width: Monitoring Mode: 0.5-40Hz Neonate Mode: 147mmHg
Diagnosis mode: 0.05-150Hz Accuracy: ±3mmHg
Surgery mode:1-20Hz Initial pressure
ST mode: 0.05-40Hz range(mmHg): Adult: 80~240;Pediatric: 80~200;
Neonate:60~120
Alarm: Systolic, Diastolic, Mean ±3% (70-100%, Neo, non-motion)
PR from NIBP: Measurement & alarm range: 40- Unspecified (1-69%)
240bpm Data averaging and
Resolution: 1bpm other signal
Accuracy: ±3bpm or ±3%, whichever is processing time: 2s
greater Data refresh rate: 8s
Nellcor SpO2: PR Measurement
Measurement Range: 20--254bpm
Range 0-100% Resolution: 1bpm
Alarm Range 20-100% Accuracy: ±2bpm
Resolution: 1% Alarm range: 20~254bpm
Accuracy: ±2% (70-100%, Adu/Ped, non-motion) Perfusion index: 0.05%~20%
±3% (70-100%, Neo, non-motion) Resolution: 0.01% (within
1-69% unspecified 0.05%~9.99% range) or 0.1% (within
Alarm Range: 20-100% 10.0%~20.0% range)
PR Measurement Temperature (Dual Channel)
Range: 20-300bpm Range: 0-50°C
Resolution: 1bpm TEMP sensor: Skin/rectal TEMP sensor
Accuracy: ±3bpm (20-250bpm); Resolution: 0.1°C
Unspecified (251-300bpm) Accuracy: ±0.1°C (exclusive of error of sensor)
Alarm Range: 20~300bpm Channel: T1, T2, TD (Temperature Difference)
Masimo SpO2: EtCO2
Measurement & Unit: mmHg, kPa
alarm Range 1~100% Measurement
Resolution: 1% range: 0mmHg~150mmHg
Accuracy: ±2% (70-100%, Adu/Ped, non-motion) Resolution 1mmHg or 0.1kPa or 0.1%
±3% (70-100%, Neo, non-motion) Accuracy 0mmHg ~40mmHg should be±2mmHg;
1-69% unspecified 41mmHg ~70mmHg should
Alarm range 1~100% be±5%×reading;
PR Measurement 71mmHg ~100mmHg should
Range 25~240bpm be±8%×reading;
Resolution: 1bpm 101mmHg~150mmHg should
Accuracy: ±3bpm (non-motion) be±10%×reading
±5bpm (motion); Oxygen
Alarm range: 25~240bpm compensation 0~100 mmHg
Perfusion index: 0.02~20% Equilibrium gas Helium, room air, nitrous oxide
Resolution: 0.01% (within IBP
0.02%~9.99% range) or 0.1% (within Channel: 2 Channels
10.0%~20.0% range) Measured Pressure: ART, PA, CVP, RAP, LAP, ICP, LV, AO,
COMEN SpO2: UAP, BAP, FAP, UVP, IAP, P1, P2, P3, P4
Measurement Measurement Unit: mmHg/ kPa selectable
& alarm Range 0~100% Measurement
Resolution: 1% Range: ART: 0~300mmHg
Accuracy: ±2% (70~100%, Adu/Ped, non-motion) PA: -6~120 mmHg
CVP: -10~40mmHg Data output: Fi and Et values
RAP: -10~40mmHg AG resolution: CO2: 1mmHg
LAP: -10~40mmHg awRR: 1rpm
ICP: -10~40mmHg Accuracy: For all measured values complies with
LV: 0~300mmHg EN ISO 21647:2004 and EN 864:1996
AO: 0~300mmHg Alarm: EtCO2: 0mmHg~190mmHg
UAP: 0~300mmHg Fi CO2: 0mmHg~190mmHg
BAP: 0~300mmHg AwRR: 2mmHg~150mmHg
FAP: 0~300mmHg EtO2: 18% ~ 100%
UVP: -10~ 40mmHg FiO2: 18% ~ 100%
IAP: -10~40mmHg EtN2O: 0% ~ 100%
P1, P2, P3, P4: -50~300mmHg FiN2O: 0% ~ 82%
Accuracy: ±2% or ±1mmHg, whichever is greater EtHal/EtEnf/EtIso/EtSev/EtDes: 0% ~
Resolution: 0.1kPa or 1mmHg 25%
(-50mmHg~300mmHg) FiHal/FiEnf/FiIso/FiSev/FiDes: 0% ~
Alarm Range: -50mmHg~300mmHg 25%
Pressure sensor: sensitivity: 5 V/V/mmHg Others: Up to 4 waveforms displayed
Impedance range: 300~3000Ω MAC value displayed
PR from IBP: Measurement & alarm range: ISATM (AG) Sidestream Gas Analyzer
20bpm~350bpm Method: Infrared gas measurement
Resolution: 1bpm No Breaths
Accuracy: ±1bpm or ±1%, whichever is Timeout range Adult: 10s, 15s, 20s, 25s, 30s, 35s,
greater 40s ,45s, 50s, 55s or 1min;
Pediatric and neonate: 20s, 25s, 30s,
AG 35s or 40s
AG (complies with ISO 80601-2-55) Accuracy: ±5s
Method: Infrared Radiation Absorption No Breath Alarm
Characteristics Delay: 10s, 15s, 20s, 25s, 30s, 35s, 40s, 45s,
AG preheating time 50s, 55s, 1min or Off
<20s Working
Gas sorts: CO2, N2O, DES, ISO, ENF, SEV, HAL, O2 conditions: ISA AX+: 0~50°C (32~122°F); ISA OR+:
(optional paramagnetic sensor) 5~50°C (41~122°F)
Measurement Storage conditions -40~70°C (-40~158°F)
range: CO2: 0~15%: ±(0.2kPa+reading×2%), RH <4kPa H2O (non-condensing) 95% RH,
15~25%: unspecified 30°C
N2O: 0~100 %: ±(2kPa+reading×2%) Barometric
HAL, ISO, ENF: 0~8%: ± pressure 52.5~120kPa (4572m)
(0.15%+reading×5%); 8~25vol%: Water treatment Sampling tube: patented dehydration
unspecified tube
SEV: 0~10%: ± (0.15%+reading×5%); Data output: Fi and Et values
10~25vol %: unspecified Waveform: Display up to 4 gas concentration
DES: 0~22%: ± (0.15%+reading×5%); waveforms at a time
22~25%: unspecified Diagnostic
O2: 0-100%: ± (1%+reading×2%) parameter: Barometric pressure
ISA sensor: 2~9-channel NDIR gas analyzer Step: 0.1°C
(measurement range: 4~10µm) Impedance Cardiography (ICG)
Compensation: CO2 broadening effect
Calibration No calibration is required. The Monitor Method: Indirect measurement by the
will auto perform zeroing when impedance cardiogram
powered on and perform auto zeroing Measurement
every 24h (ISACO2) or 8h (ISA range: SV: 5~250 ml/ beat
AX+/OR+) subsequently. HR: 40~250 bpm
Preheating time ISA CO2: <10s; ISA OR+/AX+: < 20s C.O.: 1.4~15 L/min
Rise time CO2: ≤250ms; N2O: ≤ 350ms; Accuracy: HR: ±2bpm
anesthetic gases: ≤ 350ms; O2: ≤ C.O.: Unspecified
450ms SV: unspecified
Overall system Alarm range: C.I.:0~15.0L/min/m2
response time <3s (2m sampling tube) TFC:10~150KΩ
Respiration
Alarm Deviation C.I.: ±0.1L/min/m2
detection Self-adaptive threshold (minimum CO2
TFC: ±1k Ω
concentration change: 1 vol%)
RR 0~150 breaths/min BIS
Anesthetic gas Measurement
threshold Threshold of main anesthetic gases range: BIS: 0~100; Accuracy: 1%
(ISA OR+/AX+): 0.15 vol%. The SQI: 0~100%; Accuracy: 1%
concentration of any identified EMG: 0~100dB; Accuracy: 1%
anesthetic gas will be reported, even if ESR: 0~100%; Accuracy: 1%
it is lower than 0.15 vol% Resolution: 1
Cardiac Output (C.O.) Alarm range: BIS: 0~99
Method: Thermodilution Wall Mounting
Range: C.O.: 0.1~20L/min Net weight: 3.1kg
BT: 25~43°C Bracket: 275*150*165mm
IT: 0~25°C Length 330mm
Resolution: C.O.: 0.01L/min (Support Arm)
BT, IT: 0.1°C Max. Slip 140mm
Accuracy: C.O.: ±5% or ±0.1 L/min,whichever is Rotation 180°
greater
BT, IT: ±0.1°C (no sensor) *Notice: Specifications subject to changes without
Alarm Range: BT Hi limit: (LO limit +0.4)-43°C prior notice. All rights reserved by COMEN
BT Lo limit:25.0~(Hi limit-0.4) °C