Oval Flowmeter

This document provides specifications for OVAL Corporation's ULTRA OVAL flow meters. It describes the meter sizes, features, general specifications for the meter body and register, and power supply information.

Uploaded by

Willard Custodio
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 12

ULTRA OVAL GENERAL SPECIFICATION

METER SIZES: 39, 41, 45, 50, 52, 53, 55, 56, 57 GS.No.GBU005-6-E
Register Type A and B

■ GENERAL
Our workhorse PD flowmeter series is redesigned in this
ULTRA OVAL with the most advanced multi-function
electronic register (ULTRA register) capable of
indicating the instantaneous flowrate and total flow on Meter sizes 55, 56
Meter sizes 39, 41, 45
an easy-to-read LCD and of providing the pulse and
analog output. Significantly improved performance
along with compact and lightweight design are among
the many benefits it offers.

■ FEATURES
1. Flow range expanded by 10 to 90% (over the
previously offered OVAL meters of the same sizes). Meter sizes 50, 52, 53 Meter size 57
2. Absence of mechanical reduction gear train combined
with special carbon bearings contributes to low
pressure loss and long life.
3. Thanks to pocketless design, the process fluid is 5. Output signal is available in two channels
virtually free from stagnation in the measuring simultaneously in the form of total flow (4/20mA DC
chamber-beneficial to a broad range of fluids, factored or unfactored current pulses) and instan-
particularly chemicals and foods. taneous flowrate signal (4~20mA DC analog).
4. Microprocessor-based ULTRA register indicates 6. A complete series of explosionproof models also
variables-total flow, both resettable and cumulative, available.
instantaneous flowrate, selectable with mode select 7. We also manufacture models approved for applicable
switch, plus alarm (low battery alarm) on the LCD. high-pressure gas control law.
■ GENERAL SPECIFICATIONS
● Meter Body
Item Description
Meter size 39 41 45 50 52 53 55 56 57
Nominal size 10mm 20mm 25mm 40mm 50mm
1 group JIS 10K RF, ANSI/JPI 150 RF
Flange rating
3 group JIS 16, 20, 30K RF, ANSI/JPI 300 RF, DIN 10, 16, 20, 25
Flow range See flow range tables (page 3).

Operating
Standard –10~+120˚C
Low ––––––––––– –60~+60˚C
temperature
High ––––––––––– 120~200˚C 120~350˚C
range
Jacketed ––––––––––– –––––––– 120~350˚C
Linearity ±0.35% or ±0.15%
Repeatability ±0.05% or ±0.02%
Body SCS14 or ✽SCS16
Materials

Rear cover SUS316 or ✽SUS316L


Rotors 39:Special carbon only, 41~57: SUS316 or ✽SUS316L
Bearings 39:Special carbon only, 41~57: Special carbon or ceramics
Shafts SUS316 or ✽SUS316L
Connection ––––––––––– Rc1/2 Rc3/4
Jacketed
Max. operating press. ––––––––––– 0.98MPa
Flow direction Right left (standard), left right, bottom top, top bottom
✽Special
●Flange Rating and Max. Operating Pressure (MPa)
Flange group Temp. JIS 10K RF ANSI/JPI 150 RF JIS 20K RF JIS 30K RF ANSI/JPI 300 RF
Max. 120˚C 1.18 1.51 –––– –––– ––––
1 Max. 200˚C 1.18 1.25 –––– –––– ––––
Max. 350˚C –––– –––– –––– –––– ––––
Max. 120˚C –––– –––– 2.45 2.94 2.94
3 Max. 200˚C –––– –––– 1.7 1.7 1.7
Max. 350˚C –––– –––– 1.5 1.5 1.5
Applies to body material code C. For body material code E, consult the factory.

OVAL Corporation http://www.oval.co.jp

Head Office: Tokyo; Tel. 81-3-3360-5121. Fax 81-3-3365-8605


Beijing Office: Tel., Fax. 86-10-6525-3601. Tel. 86-10-6513-3516
Overseas Branch Offices: Seoul. Singapore. Taipei
● Register Specifications
Item Description
Meter size 39 41 45 50, 52, 53 55, 56, 57
Local Cumulative totalizing counter 0.1 mL (std.) 0.001L (std.) 0.001L (std.) 0.01 (std.) 0.1 L (std.)
display Reset counter, 7digits C mode 1 mL, 0.01 L 0.01L, 0.1L 0.01L, 0.1L 0.1 L, 1L 1 L, 0.01 m3
(LCD) Instantaneous flow b1 mode 1 mL/h (std.) 0.01 L /h (std.) 0.1 L /h (std.) 1 L /h (std.) 1 L /h (std.)
✽1
rate (5 digits) b2 mode 0.01 mL/min (std.) 0.0001 L/min (std.) 0.001 L/min (std.) 0.01 L/min (std.)
—— None Local LCD only
Analog 4 ~ 20 mADC Refer to diagram, page 6.
Type Scaled or unscaled: 0/1=4/20 mADC
Current
Pulse Pulse width Scaled: 1ms (std.), 50 ms Unscaled: 2ms
Output signal

Unit of scaled pulse Same as of LCD counter ✽1

Scaled or unscaled: NPN transister output: Max. impressed voltage 30 VDC


Open Type allowable current: 50 mA,
transister on voltage: 1.5VDC and below
Pulse Scaled: 1ms (std.), 50 ms Unscaled: 2ms
collector Pulse width
Same as of LCD counter ✽1
Unit of scaled pulse
Type Scaled or unscaled: [0]1 VDC Max. [1]7 VDC Min.
Voltage Pulse Pulse width Scaled: 1ms (std.), 50 ms Unscaled: 2ms
Same as of LCD counter ✽1
Unit of scaled pulse
Without output signal Installed lithium battery Life: 8 years
Power External power source: 12 ~ 45 VDC (analog or current pulse)
With output signa 12 ~ 24 VDC (open collector pulse or voltage pulse)
supply 12 ~ 45 VDC (combination analog and current pulse)
Current consumption: Max. 30 mADC Refer to diagram in page 6. (✽2)
Captyre cable w/external shield
Transmission cable
(VCTF 1.25 mm2, finished O. D. 8.5 ~ 12mm) (✽3)
Transmission length Max.1 km
2-wire system Analog or current pulse
Transmission 3-wire system Open collector or voltage pulse
lines Analog + current pulse
4-wire system
Operating temperature range –10 ~ +60˚C
Select either one from following two
Explosionproof construction ① Non-explosionproof type
② Explosionproof type: ExdIIBT4 / ExiaIIBT4
Degree of Protection for enclosure IP66 (Dust-tight/Watertight Type)-IEC/EN60529, JIS C 0920
Material for housing Aluminum die casting
Finish Munsell No. 2.5PB5/8 (baked melamine)

Notes: (✽1) If factored pulse units other than above are required, consult the factory.
(✽2) Battery powered register features a local indicator alone; output signal is not available.
(✽3) For use of output type in hazardous locations, use cables heat resistant to 75˚C or higher.
Do not fail to use pressuretight packings furnished as standard accessories.

■ NOMINAL METER FACTORS


Meter size P/r Standard Low/high temp. and jacketed
39 2 0.09838mL/P –––––
41 2 0.4896mL/P –––––
45 2 1.2339mL/P –––––
50 4 4.968mL/P 9.936mL/P
52 4 9.664mL/P 9.664mL/P
53 4 17.513mL/P 17.513mL/P
55 6 23.07mL/P 34.60mL/P
56 6 37.33mL/P 74.66mL/P
57 8 98.04mL/P 196.08mL/P

2
■ FLOW RANGES
●Meter sizes: 39~45
Linearity: ±0.35% Linearity: ±0.15% (Option)
Operating temp. range: –10~+120˚C Unit in L/h Operating temp. range: –10~+120˚C Unit in L/h
Viscosity Less than 0.3mPa·s 0.8mPa·s 2mPa·s 5mPa·s ~ Viscosity Less than 0.3mPa·s 0.8mPa·s 2mPa·s 5mPa·s ~
Meter size 0.3mPa·s ~ 0.8mPa·s ~ 2mPa·s ~ 5mPa·s ✽200mPa·s Meter size 0.3mPa·s ~ 0.8mPa·s ~ 2mPa·s ~ 5mPa·s ✽200mPa·s
39 2~ 12 1.4~ 12 0.7~ 12 0.35~12 0.2~ 12 39 3~ 12 2~ 12 1~ 12 0.5~ 12 0.3~ 12
41 18~ 60 12 ~ 60 4~ 60 2.5 ~ 60 1 ~ 60 41 27~ 60 18~ 60 6~ 60 3.7~ 60 1.5~ 60
45 50~420 35 ~420 15~420 10 ~420 5 ~420 45 75~420 52~420 22~420 15~420 7.5~420
✽ Only model 45 can handle up to 1000mPa ·s
●The standard accuracy for model 39 is ±0.35%. Consult the factory
for ±0.2% accuracy model.

●Meter sizes: 50~57


1. Linearity: ±0.35% 4. Linearity: ±0.15% (Option)
Operating temp. range: –10~+120˚C (std.) Unit in m3/h Operating temp. range: –10~+120˚C Unit in m3/h
Viscosity Less than 0.3mPa·s 0.8mPa·s 2mPa·s 5mPa·s~ Viscosity Less than 0.3mPa·s 0.8mPa·s 2mPa·s 5mPa·s ~
Meter size 0.3mPa·s ~ 0.8mPa·s ~ 2mPa·s ~ 5mPa·s 1000mPa·s Meter size 0.3mPa·s ~ 0.8mPa·s ~ 2mPa·s ~ 5mPa·s 1000mPa·s
50 0.3~1.6 0.15~1.6 0.1 ~1.6 0.05~ 2 0.03 ~ 2 50 0.5~1.6 0.3~1.6 0.15~1.6 0.08~2 0.05~2
52 0.7~3 0.4 ~3 0.3 ~3 0.15~ 3.8 0.08 ~ 3.8 52 1 ~3 0.7~3 0.5 ~3 0.25~3.8 0.15~3.8
53 1.1~5 0.7 ~5 0.55~5 0.28~ 6.4 0.15 ~ 6.4 53 1.6~5 1.1~5 0.75~5 0.4 ~6.4 0.22~6.4
55 1.8~11 1.2 ~11 1 ~ 11 0.4 ~14 0.26~14
56 3.5~20 2.5 ~20 2 ~ 20 0.9 ~24 0.6 ~ 24 Operating temp. range: –10~+60˚C Unit in m3/h
Viscosity Less than 0.3mPa·s 0.8mPa·s 2mPa·s 5mPa·s ~
57 8 ~ 37 5 ~ 37 4 ~ 37 2 ~ 44 1.2 ~ 44 0.3mPa·s ~ 0.8mPa·s ~ 2mPa·s ~ 5mPa·s 1000mPa·s
Meter size
55 2.7~11 1.8~11 1.5~11 0.6 ~14 0.4~14
2. Linearity: ±0.35% 56 5.2~20 3.5~20 3 ~ 20 1.4 ~24 0.9~24
Operating temp. range: 120 ~ 200˚C Unit in m3/h
57 12 ~ 37 8 ~37 6 ~ 37 3 ~44 2 ~44
Viscosity Less than 0.3mPa·s 0.8mPa·s 2mPa·s 5mPa·s~
Meter size 0.3mPa·s ~ 0.8mPa·s ~ 2mPa·s ~ 5mPa·s 1000mPa·s
50 0.6~ 1.4 0.3~ 1.4 0.2~ 1.4 0.09~1.8 0.05~1.8 Operating temp. range: 60 ~ 120˚C Unit in m3/h
52 1 ~ 2.7 0.8~ 2.7 0.5~ 2.7 0.23~3.4 0.15~3.8 Viscosity Less than 0.3mPa·s 0.8mPa·s 2mPa·s 5mPa·s ~
Meter size 0.3mPa·s ~ 0.8mPa·s ~ 2mPa·s ~ 5mPa·s 1000mPa·s
53 2 ~ 4.5 1.4~ 4.5 0.9~ 4.5 0.35~5.7 0.28~6.4
55 4 ~ 11 2.7~11 2.2~ 11 0.9 ~14 0.6 ~ 14
55 3.6 ~ 9 2.4~ 9 1.5~ 9 0.6~12 0.4~14
56 8 ~ 20 5.2~20 4.5~ 20 2.1 ~24 1.3 ~ 24
56 7.5~18 5 ~ 18 3 ~ 18 1.4~21 0.9~24
57 18 ~ 37 12 ~37 9 ~ 37 4.5 ~44 3 ~ 44
57 15 ~ 33 10 ~ 33 6 ~ 33 3 ~ 39 2 ~ 44

3. Linearity: ±0.35%
Operating temp. range: 200 ~ 350˚C Unit in m3/h
5. Linearity: ±0.15% (Option)
Operating temp. range: 120 ~ 200˚C Unit in m3/h
Viscosity 0.3mPa·s 0.8mPa·s 2mPa·s 5mPa·s~ Viscosity 0.3mPa·s 0.8mPa·s 2mPa·s 5mPa·s ~
Meter size ~ 0.8mPa·s ~ 2mPa·s ~ 5mPa·s 1000mPa·s
Meter size ~ 0.8mPa·s ~ 2mPa·s ~ 5mPa·s 1000mPa·s
52 1 ~ 2.7 0.6 ~2.7 0.3 ~ 3.4 0.16 ~ 3.8
50 0.45~1.4 0.3~ 1.4 0.15~ 1.8 0.08 ~ 1.8
53 2 ~ 4.5 1.2 ~4.5 0.6 ~ 5.7 0.3 ~ 6.4
52 1.5 ~2.7 0.9~ 2.7 0.55~ 3.4 0.33 ~ 3.8
55 3.6 ~ 9 2~ 9 0.8~12 0.55~14
53 2.4 ~4.5 1.5~ 4.5 0.9 ~ 5.7 0.49 ~ 6.4
56 7.5~18 4 ~ 18 1.8~21 1.2 ~ 24
55 4 ~ 9 2.7 ~ 9 1.3 ~ 12 0.9 ~ 14
57 15 ~ 33 8 ~ 33 4 ~ 39 2.5 ~ 44
56 8 ~ 18 5.2~18 3.1 ~ 21 1.9 ~ 24
Note: 1. For measurement of high viscosity fluids (above 1000mPa·s),
57 18 ~ 33 12 ~ 33 6.7 ~ 39 7.5 ~ 44
consult the factory.
2. For flow range of meters for low temperature service
(–60~+60˚C) refer to Table 1 or 4. 6. Linearity: ±0.15% (Option)
Operating temp. range: 200 ~ 350˚C Unit in m3/h
3. For standard flowmeters (–10~+60˚C) refer to Table 1 or 4.
Viscosity 0.8mPa·s 2mPa·s 5mPa·s
4. For flow range of meters compatible with thermal shock
(–10~+120˚C), refer to Table 2 or 5. (Thermal shock means Meter size ~ 2mPa·s ~ 5mPa·s ~ 1000mPa·s
sharp fluid temperature changes at a rate in excess of 3˚C 52 1.5 ~ 2.7 0.9 ~ 3.4 0.49~3.8
/min. or staircase changes in excess of 30˚C between steps.
53 2.4 ~ 4.5 1.3 ~ 5.7 0.73~6.4
55 4 ~ 9 1.9~12 1.3~14
● For flow range with “water,” select by temperature
and viscosity brackets from the table below. 56 8 ~ 18 4.6~21 2.8~24
57 18 ~ 33 10 ~ 39 6.7~440
Temperature range Viscosity range
Max. 30˚C 0.8 ~ 2.0 mPa·s
30 ~ 80˚C 0.3 ~ 0.8 mPa·s
80 ~ 120˚C Less than 0.3 mPa·s

3
■ METER ERRORS AND PRESSURE LOSSES

Meter sizes: 39, 41, 45


Meter error (%)

39 41 45

3 10 30

Pressure loss (kPa)


7.5
2 20

5
1 10
2.5

0 25 50 75 100

Meter sizes: 50, 52, 53 Meter size: 55


Meter error (%)
Meter error (%)

Pressure loss (kPa)


Pressure loss (kPa)

40

30

20

10

Flow rate (%) Flow rate (%)

Meter size: 56 Meter size: 57


Meter error (%)
Meter error (%)

Pressure loss (kPa)


Pressure loss (kPa)

Flow rate (%) Flow rate (%)

NOTE: 100% flowrate shows the maximum flowrate at individual viscosity.

4
■ HOOKUP WITH RECEIVING INSTRUMENTS

Receiving instruments

ULTRA OVAL Totalizer


(EL0122)

Batch controller
Valve control
(EL18 series)
output

Open
MOS-FET
Pulse divider
(SU1308)
Totalizer
(EL0122)
Scaled pulse (current, open collector, voltage)

Unscaled pulse (current, open collector, voltage) Converter, Analog output


Compensator Pulse output
(EL084 )
Analog (4~20mADC)
OVAL
MODEL
EL4000
Flow computer
Pulse output
(EL4000 series)
Open
MOS-FET
Power supply unit Pulse divider
(SU1303) Indicator
(SU1308)
(EL1123)

Recorder

NOTE: Typical examples are shown here.


Many other receving instruments are acceptable.

■ WIRING CONNECTIONS
(1) Current pulse output (2-wire system) (2) Analog output (2-wire system)
Terminal block

V 24VDC 1
24VDC
Below
Terminal block 600 Terminal block
2
SIG V
1 1 12 ~ 45VDC
+
3
0V SIG
2 2 Power supply

4
G Below
3 3 +
0V 600

4 4 Analog indicator

In case of current input


Oval receiving instrument Oval receiving instrument
NOTE: OVAL offers two circuit configurations to accept current pulses as NOTE: To accept a voltage signal, couple an external
shown above. See the respective instruction manual of receiving load resistor (see acceptable load resistance
instrument for correct connections. range on page 6).

(3) Current pulse + Analog output (4-wire system)


NOTE 1:
Terminal block Tie point In an OVAL receiving instrument, an internal load
12 ~ 45VDC resistor plays the role of I/V conversion. But if you
+
1 Power supply build a system like the one shown here with a
Minimum commercially available totalizer, a combination
Load resistor 4/20mA x load resistance serves as a current
– electrical capacity 60mA
2 pulse/voltage pulse converter, make sure of the
+ input level of the totalizer before use.
3 ✽
G Totalizer NOTE 2:
– With a configuration like the one shown here
4 + where an OVAL receiving instrument is used,
make sure of the current carrying capacity of
Analog indicator
receiving instrument’s power supply before use.
– If found inadequate in current carrying capacity,
✽: Select Voltage signal input for the totalizer. prepare an additional power supply for the analog
Make sure of the trigger level of incoming voltage signal and determine indicator.
the supply voltage and load resistance value.

5
(4) Open collector pulse (3-wire system) (5) Voltage pulse (3-wire system)

Terminal block Terminal block


12 ~ 24VDC
1 V Constant 1 V
E ( 30VDC) voltage 12 ~ 24VDC
1K
2 SIG R Select R so that the 2 SIG
allowable current be
Max. 50mA 50mA max. Over 10K
3 0V 3 0V

4 4 1ms (Std.) or 50ms


Min.7V
1ms (Std.) or 50ms (Max.12V
V = 24V)
Max.1V

● Range of Load Resistance


(for current pulse and analog output) 1650

This instrument uses a two-wire transmission line for analog and Operating range
pulse signals, so the line serves for both power supply and signal.
A DC power supply is required for transmission loop.

Load resistance ( )
When connecting a meter to the loop, ensure that the meter and (600)
the load resistance of cable conductor is within the operating limits
shown in the figure at right. 300

Standard: Power supply voltage = 24VDC


0
Load resistance = 250Ω 0 12 (24) 45
Power supply voltage (V DC)

■ PRODUCT CODE EXPLANATION (Meter sizes 39, 41, 45)


Code No.
Item Description
1 2 3 4 5 6 7 8 – 9 10 11 12
Type L U S ULTRA OVAL, standard version (–10 ~ +120˚C)
3 9 Nominal size 10mm
Meter size 4 1 Nominal size 10mm
4 5 Nominal size 10mm
C SCS14
Body material
E SUSF316 [Special]
1 JIS 10K RF, ANSI/JPI 150 RF
Flange rating
3 JIS 16 ~ 30K RF, ANSI/JPI 300 RF, DIN10 ~ 25
1 – Standard bearings (Special carbon)
5 – Ceramic bearings
Bearing type ✽1
7 – Polymerization-inhibited ceramics bearings (Depends on liquid kind.)
8 – Polymerization-inhibited carbon bearings (Special)
A Basic ULTRA model
Register type
B Batch controller equipped ULTRA register (LW74E/LW76E) ✽2
3 Nonexplosionproof (w/battery unit)
Construction of register 4 Explosionproof (w/battery unit) TIIS
7 Explosionproof (w/battery unit) NEPSI ✽3
0 0 No output signal, Local LCD only (w/battery unit)
0 1 Unscaled pulse (Current pulse) 2 wires
0 2 Scaled pulse (Current pulse) 2 wires
0 5 Unscaled pulse (Open collector pulse) 3 wires
Types of output signal 0 6 Scaled pulse (Open collector pulse) 3 wires
(Register code A only.) 0 7 Unscaled pulse (Voltage pulse) 3 wires
0 8 Scaled pulse (Voltage pulse) 3 wires
1 0 Analog 2 wires
1 1 Analog + Unscaled pulse (Current pulse) 4 wires
1 2 Analog + Scaled pulse (Current pulse) 4 wires
Batch control function 7 4 Pneumatic batch controller (LW74E register) 1-step open/1-step close type
(Only register type B) 7 6 Pneumatic batch controller (LW76E register) 2-step open/2-step close type
Notes: ✽1. Meter sizes 41 and 45 use the same body material SUS316 or 316L for rotors.
Meter size 39 uses special carbon rotors only.
✽2. For register code B, see General Specification (No.GBC201-E).
✽3. Only Register type A.
6
■ PRODUCT CODE EXPLANATION (Meter sizes 50~57)
Code No.
Item Description
① ② ③ ④ ⑤ ⑥ ⑦ ⑧ – ⑨ ⑩ ⑪⑫
L U S ULTRA OVAL, Standard version (–10 ~ +120˚C)
L U H ULTRA OVAL, high temp. service (120 ~ 350˚C)
L U N ULTRA OVAL, low temp. service (–60 ~ +60˚C)
Type
L U J Jacketed ULTRA OVAL (Meter sizes 52 thru 57 only)
L U T ULTRA OVAL, compatible with sharp temp. changes (–10 ~ +120˚C)
K U S ULTRA OVAL, Standard Flowmeter (–10 ~ +60˚C)
5 0 Nominal size 20mm (3/4˜)
5 2 Nominal size 25mm (1˜)
5 3 Nominal size 25mm (1˜)
Meter size
5 5 Nominal size 40mm (1 1/2˜)
5 6 Nominal size 50mm (2˜)
5 7 Nominal size 50mm (2˜)
C SCS14
Body material
E SCS16 [Special]
1 JIS 10K RF, ANSI, JPI 150 RF
Flange rating
3 JIS16 ~ 30K RF, ANSI, JPI 300 RF, DIN10 ~ 25 (Without high/low temperature, jacket)
1 – Standard bearings (Special carbon)
5 – Ceramic bearings
Bearing type
7 – Polymerization-inhibited ceramic bearings
8 – Polymerization-inhibited carbon bearings
A Basic ULTRA model
Register type
B Batch controller equipped ULTRA register (LW74E/LW76E) ✽1
3 Nonexplosionproof (w/battery unit)
Construction of register 4 Explosionproof (w/battery unit) TIIS
7 Explosionproof (w/battery unit) NEPSI ✽2
0 0 No output signal, Local LCD only (w/battery unit)
0 1 Unscaled pulse (Current pulse) 2 wires
0 2 Scaled pulse (Current pulse) 2 wires
0 5 Unscaled pulse (Open collector pulse) 3 wires
Types of output signal 0 6 Scaled pulse (Open collector pulse) 3 wires
(Only in case register type A is used.) 0 7 Unscaled pulse (Voltage pulse) 3 wires
0 8 Scaled pulse (Voltage pulse) 3 wires
1 0 Analog 2 wires
1 1 Analog + Unscaled pulse (Current pulse) 4 wires
1 2 Analog + Scaled pulse (Current pulse) 4 wires
Batch control function 7 4 Pneumatic batch controller (LW74E) 1-step open/1-step close type
(Only register type B) 7 6 Pneumatic batch controller (LW76E) 2-step open/2-step close type
Note.: ✽ 1. For register code B, see General Specification Sheet (No.GBC201-E).
✽2. Only Register type A.

■ OUTLINE DIMENSIONS [Standard register type A provided] (Unit in mm)


● Meter sizes: 39, 41, 45 150 207.8

90 79.8

MODE RESET
MODE
130

X0.1 l
MIT
90

b1:X10 l/h b2:X0.1 l/min


138.4

Mass(Apporox.)
Conduit connection G1/2 8 127.8
7.5kg

Note: 1. For batch controller equipped ULTRA register, refer to General Specification (No. GBC201).
7
■ OUTLINE DIMENSIONS [Standard ULTRA register type A provided] (Unit in mm)
● Standard (Types: LUS, LUT, KUS)
L B
●Meter sizes : 50, 52, 53, 55, 56
90 C D

RESET

130
MODE

X10 ml
MIT

E
b1:X1 1/h b2:X10 s1/sis

137

8 F
Conduit connection G1/2 (female)
(Only for units with remote output signal)

1. Flange rating 1 group


Meter size Flange rating Mass
L B C D E F (Approx.)
JIS 10K RF 200
50 216.8 86.8 40 ø96 134.8 9kg
ANSI/JPI 150 RF 198
JIS 10K RF 200
52 207.8 74.8 43 106 122.8 10kg
ANSI/JPI 150 RF 200
JIS 10K RF 200
53 231.3 85.8 55.5 106 133.8 11kg
ANSI/JPI 150 RF 200
JIS 10K RF 230
55 244.8 92.8 62 130 140.8 16kg
ANSI/JPI 150 RF 233
JIS 10K RF 250
56 269.8 106.8 73 154 154.8 20kg
ANSI/JPI 150 RF 258

2. Flange rating 3 group


Meter size Flange rating Mass
L B C D E F (Approx.)
JIS 20K RF 204
50 JIS 30K RF 208 216.8 86.8 40 96 134.8 12kg
ANSI/JPI 300 RF 204
JIS 20K RF 204
52 JIS 30K RF 212 207.8 74.8 43 106 122.8 13kg
ANSI/JPI 300 RF 207
JIS 20K RF 204
53 JIS 30K RF 212 231.3 85.8 55.5 106 133.8 14kg
ANSI/JPI 300 RF 207
JIS 20K RF 234
55 JIS 30K RF 242 248.8 94.8 64 163 142.8 22kg
ANSI/JPI 300 RF 240
JIS 20K RF 254
56 JIS 30K RF 262 271.8 109.8 72 193 157.8 26kg
ANSI/JPI 300 RF 263

●Meter size : 57 B
L 90 C D
G

G
130

MODE RESET
RESET
MODE
MODE

X0.1 l
E

MIT

b1:X10 l/h b2:X0.1 l/min


137

Conduit connection G1/2 (female) 8 F


(Only for units with remote output
signal)

1. Flange rating 1 group

Meter size Flange rating Mass


L B C D E F G (Approx.)
JIS 10K RF 350
57 299.8 124.8 85 ø240 172.8 171.5 36kg
ANSI/JPI 150 RF 357
2. Flange rating 3 group

Meter size Flange rating Mass


L B C D E F G (Approx.)
JIS 20K RF 354
57 JIS 30K RF 362 308.8 131.8 87 ø260 179.8 171.5 47kg
ANSI/JPI 300 RF 363
In case the other Flange rating dimensions are requested consult factory.
Note: For batch controller equipped ULTRA register, refer to General Specification (No. GBC201).

8
■ OUTLINE DIMENSIONS [Standard ULTRA register type A provided] (Unit in mm)
● With cooling or heating tube (Types: LUH, LUN)

● Metersize : 52, 53
L B B

90 157.3 C D 90 157.3 C D
Cooling Heating
tube tube

130

130
RESET
MODE

X10 ml
MIT

E
E
b1:X1 I/h b2:X10 sI/sis
137

8 F F

Conduit connection 2 (female) G1/


(Only for units with remote output signal)

Flange rating 1 group


Meter size Flange rating B Mass (Approx.)
L (Overall length) C D E F
JIS 10K RF 200 W/Cooling tube 17kg
52 353.3 59 47 106 264.3 W/Heating tube 16kg
ANSI/JPI 150 RF 200
JIS 10K RF 200 W/Cooling tube 18kg
53 376.8 70 59.5 106 275.3 W/Heating tube 17kg
ANSI/JPI 150 RF 200

● Metersize : 55, 56

L B B
90 167.3 C D 90 167.3 C D
Cooling Heating
tube tube
130
130

RESET
MODE MODE

X0.1 l

E
E

MIT

b1:X10 I/h b2:X0.1 l/min


137

8 F F

Conduit connection G1/2 (female)


(Only for units with remote output signal)

Flange rating 1 group


Meter size Flange rating L B C D E F Mass (Approx.)
JIS 10K RF 230 W/Cooling tube 23kg
55 371.3 43 71 ø163 258.3 W/Heating tube 22kg
ANSI/JPI 150 RF 233
JIS 10K RF 250 W/Cooling tube 27kg
56 391.3 63 76 ø193 273.3 W/Heating tube 26kg
ANSI/JPI 150 RF 258

● Metersize : 57

L B B
90 C D 90 C D

Cooling Heating
tube tube
G

G
130

130

RESET
MODE

X10 ml
MIT

b1:X1 1/h b1:X10 s1/sis


E
E
137

8 F F
Conduit connection G1/2 (female)
(Only for units with remote output signal)

Flange rating 1 group


Meter size Flange rating L B C D E F G Mass (Approx.)
JIS 10K RF 350 W/Cooling tube 43kg
57 424.3 90 87 ø260 295.3 171.5 W/Heating tube 42kg
ANSI/JPI 150 RF 357

Note: For batch controller equipped ULTRA register, refer to General Specification (No. GBC201).
9
■ OUTLINE DIMENSIONS [Standard ULTRA register type A provided] (Unit in mm)
● Jacketed type (Type: LUJ)

● Meter sizes : 52, 53 B


L 90 157.3 C D
12

RESET

130
MODE MODE

190
X0.1 l

E
MIT

b1:X10 I/h b2:X0.1 I/min


74

Heating fluid connection


F
8 Rc 1/2
Conduit connection G1/2 (female)
(Only for units with remote output signal)

Mass
Meter size Flange rating L B C D E F (Approx.)
JIS 10K RF 200
52 352.8 58.5 47 106 263.8 13kg
ANSI/JPI 150 RF 200
JIS 10K RF 200
53 376.3 69.5 59.5 106 274.8 13.7kg
ANSI/JPI 150 RF 200

● Meter sizes : 55, 56


B
L 90 167.3 C D
14
130

MODE RESET
MODE

×0.1 l
C 213
E
MIT

b1:×10 l/h b2:×0.1 l/min


137

8 F
Heating fluid connection
Conduit connection G1/2 (female) Rc 3/4
(Only for units with remote output signal)

Meter size Flange rating Mass


L B C D E F (Approx.)
JIS 10K RF 250
55 396.3 69 70 ø163 283.3 25kg
ANSI JPI 150 RF 253
JIS 10K RF 300
56 423.3 89 77 ø193 299.3 31kg
ANSI JPI 150 RF 306

● Meter size : 57 B
L 90 157.3 C D
25
35ß

130

MODE RESET
307

MODE

X0.1 l
E

MIT

b1:X10 I/h b2:X0.1 I/min


157

8 F Heating fluid
Conduit connection G1/2 (female) connection
(Only for units with remote output signal) Rc 3/4

Meter size Flange rating Mass


L B C D E F (Approx.)
JIS 10K RF 350
57 ANSI/JPI 150 RF 357 451.8 117.5 87 ø260 322.8 52kg

10
■ STRAINERS
It is essential that a strainer be provided immediately
upstream of, or as close as to, the flowmeter to prevent
solids suspended in the process fluid from entering the
meter, possibly leading to costly downtime.

H
H

H
L L
1 3 3
SR012 3(CE)0323 C
SR0 45 1(E) 031 C
SR0 45 8 (E)032

Model number Nominal size Flange rating L (mm) H (mm) Body material Screen material std. screen mesh Applicable meters
SR013C031 JIS 10K RF 180
ANSI/JPI 150 RF 178 100 39
SR013C032
10mm JIS 20K RF 184 SCS14A SUS316 200 41
JIS 30K RF 188 45
SR013C032 126
ANSI/JPI 300 RF 185
C
SR023(E)031 JIS 10K RF 180
C SCS14A SUS316
SR023(E)032 20mm ANSI/JPI 150 RF 177 100 200 50
C (SCS16A) (SUS316L)
SR023(E)032 JIS 20K RF 184
C
SR031(E)031 JIS 10K RF 230 165
C SCS14A SUS316 52
SR038(E)032 25mm ANSI/JPI 150 RF 231 100
C 209 (SCS16A) (SUS316L) 53
SR038(E)032 JIS 20K RF 234
C
SR041(E)031 JIS 10K RF 230 165
C SCS14A SUS316
SR048(E)032 40mm ANSI/JPI 150 RF 233 60 55
C 209 (SCS16A) (SUS316L)
SR048(E)032 JIS 20K RF 234
C
SR051(E)031 JIS 10K RF 290 190
C SCS14A SCS316 56
SR058(E)032 50mm ANSI/JPI 150 RF 296 60
C
242 (SCS16A) (SCS316L) 57
SR058(E)032 JIS 20K RF 294

11
■ REQUIRED SPACE AROUND THE METER
Unit in mm

Min 300

Min 300
Min 500

Min 300
Min300

Min 500

■ OPERATING PRECAUTIONS
1. Every OVAL flowmeter is carefully assembled and 5. It is esseritial that a strainer (supplied by OVAL)
precisely adjusted to measure flows down to minute exclusively designed for OVAL flowmeters be used.
flows before is leaves the factory. Take every 6. This flowmeter is not provided with subtract
precaution in uncrating, installation in the piping function. If pulsation in the flow (where the fluid
assembly, and testing. moves back and forth in the pipeline under the
2. Never allow foreign solids to enter the measuring influence of pressure) or reversal of flow exists, the
chamber. total counter may show erratic reading, accumulating
3. Flush the piping assembly thoroughly. all incoming pulses irrespective of flow direction.
4. Avoid allowing the meter rotors to spin uncontrolled
by directing a stream of air, etc. or allowing the fluid
to flow excessively - even momentarily.

■ ORDERING INFORMATION
Please complete the following form when making inquiries.
1. Model L_______________________ Standard High temp. Low temp. Jacketed
2. Fluid to be measured Name_________________ Viscosity________mPa·s Specific gravity _________
3. Flowrate (L/h, m3/h) Maximum______________ Normal______________ Minimum______________
4. Fluid temperature (˚C) Maximum______________ Normal______________ Minimum______________
5. Ambient temperature (˚C) Maximum______________ Normal______________ Minimum______________
6. Pressure (MPa) Maximum______________ Normal______________ Minimum______________
7. Flow direction Right Left, Bottom Top
8. Flange connection Nominal size______________mm, Flange rating______________
9. Required Linearity ±______________%
10. Explosionproof construction Required class______________ Not required
11. Accessories Strainer, Air eliminator, Companion flange
12. Quantity Including accessories______________
_____________(dosing, sampling, blending process, etc.)
13. Application Flow integration, Flow indication, Record, Flow control,
Batch control, CPU interface, Others
14. Receiving instrument Type, manufacturer, model, specifications (input, output, power supply, etc.)
15. Distance between flow meter
and receiving instrument ______________m

The specification as of July, 2006 is stated in this GS Sheet. Specifications and design are subject to change without notice.