Ba01728den 0117
Ba01728den 0117
Operating Instructions
Proline Promass F 300
EtherNet/IP
Coriolis flowmeter
Keep cover
ti
gh
tw
hile
n
ne
öff
ng
nnu
Sp a
Proline Promass F 300 EtherNet/IP
• Make sure the document is stored in a safe place such that it is always available when
working on or with the device.
• To avoid danger to individuals or the facility, read the "Basic safety instructions" section
carefully, as well as all other safety instructions in the document that are specific to
working procedures.
• The manufacturer reserves the right to modify technical data without prior notice. Your
Endress+Hauser Sales Center will supply you with current information and updates to
these instructions.
2 Endress+Hauser
Proline Promass F 300 EtherNet/IP Table of contents
Table of contents
Endress+Hauser 3
Table of contents Proline Promass F 300 EtherNet/IP
4 Endress+Hauser
Proline Promass F 300 EtherNet/IP Table of contents
13 Maintenance . . . . . . . . . . . . . . . . . . . . 203
13.1 Maintenance tasks . . . . . . . . . . . . . . . . . . . . 203
13.1.1 Exterior cleaning . . . . . . . . . . . . . . . 203
13.1.2 Interior cleaning . . . . . . . . . . . . . . . 203
13.2 Measuring and test equipment . . . . . . . . . . . 203
13.3 Endress+Hauser services . . . . . . . . . . . . . . . 203
14 Repairs . . . . . . . . . . . . . . . . . . . . . . . . . . 204
14.1 General notes . . . . . . . . . . . . . . . . . . . . . . . 204
14.1.1 Repair and conversion concept . . . . . 204
14.1.2 Notes for repair and conversion . . . . 204
14.2 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . 204
14.3 Endress+Hauser services . . . . . . . . . . . . . . . 204
14.4 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204
14.5 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . 205
14.5.1 Removing the measuring device . . . . 205
14.5.2 Disposing of the measuring device . . 205
15 Accessories . . . . . . . . . . . . . . . . . . . . . . 206
15.1 Device-specific accessories . . . . . . . . . . . . . . 206
15.1.1 For the transmitter . . . . . . . . . . . . . 206
15.1.2 For the sensor . . . . . . . . . . . . . . . . . 207
15.2 Communication-specific accessories . . . . . . . 207
15.3 Service-specific accessories . . . . . . . . . . . . . . 207
15.4 System components . . . . . . . . . . . . . . . . . . . 208
Endress+Hauser 5
About this document Proline Promass F 300 EtherNet/IP
Symbol Meaning
DANGER!
DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will
result in serious or fatal injury.
WARNING!
WARNING This symbol alerts you to a dangerous situation. Failure to avoid this situation can
result in serious or fatal injury.
CAUTION!
CAUTION This symbol alerts you to a dangerous situation. Failure to avoid this situation can
result in minor or medium injury.
NOTE!
NOTICE This symbol contains information on procedures and other facts which do not result in
personal injury.
Symbol Meaning
Direct current
Alternating current
Ground connection
A grounded terminal which, as far as the operator is concerned, is grounded via a
grounding system.
Symbol Meaning
LED
Light emitting diode is off.
6 Endress+Hauser
Proline Promass F 300 EtherNet/IP About this document
Symbol Meaning
LED
Light emitting diode is on.
LED
Light emitting diode is flashing.
Symbol Meaning
Allen key
Open-ended wrench
Symbol Meaning
Permitted
Procedures, processes or actions that are permitted.
Preferred
Procedures, processes or actions that are preferred.
Forbidden
Procedures, processes or actions that are forbidden.
Tip
Indicates additional information.
Reference to documentation.
A Reference to page.
Reference to graphic.
1. , 2. , 3. … Series of steps.
Result of a step.
Visual inspection.
Symbol Meaning
1. , 2. , 3. , … Series of steps
A, B, C, ... Views
Hazardous area
-
Endress+Hauser 7
About this document Proline Promass F 300 EtherNet/IP
Symbol Meaning
1.3 Documentation
For an overview of the scope of the associated Technical Documentation, refer to the
following:
• The W@M Device Viewer : Enter the serial number from the nameplate
(www.endress.com/deviceviewer)
• The Endress+Hauser Operations App: Enter the serial number from the nameplate
or scan the 2-D matrix code (QR code) on the nameplate.
For a detailed list of the individual documents along with the documentation code
→ 240
Sensor Brief Operating Instructions Guides you quickly to the 1st measured value - Part 1
The Sensor Brief Operating Instructions are aimed at specialists with
responsibility for installing the measuring device.
• Incoming acceptance and product identification
• Storage and transport
• Installation
Transmitter Brief Operating Guides you quickly to the 1st measured value - Part 2
Instructions The Transmitter Brief Operating Instructions are aimed at specialists with
responsibility for commissioning, configuring and parameterizing the
measuring device (until the first measured value).
• Product description
• Installation
• Electrical connection
• Operation options
• System integration
• Commissioning
• Diagnostic information
8 Endress+Hauser
Proline Promass F 300 EtherNet/IP About this document
TRI-CLAMP®
Registered trademark of Ladish & Co., Inc., Kenosha, USA
Endress+Hauser 9
Basic safety instructions Proline Promass F 300 EtherNet/IP
Depending on the version ordered, the measuring device can also measure potentially
explosive, flammable, poisonous and oxidizing media.
Measuring devices for use in hazardous areas, in hygienic applications or where there is an
increased risk due to process pressure, are labeled accordingly on the nameplate.
To ensure that the measuring device remains in proper condition for the operation time:
‣ Keep within the specified pressure and temperature range.
‣ Only use the measuring device in full compliance with the data on the nameplate and
the general conditions listed in the Operating Instructions and supplementary
documentation.
‣ Based on the nameplate, check whether the ordered device is permitted for the
intended use in the hazardous area (e.g. explosion protection, pressure vessel safety).
‣ Use the measuring device only for media to which the process-wetted materials are
sufficiently resistant.
‣ If the measuring device is not operated at atmospheric temperature, compliance with
the relevant basic conditions specified in the associated device documentation is
absolutely essential: "Documentation" section→ 8.
‣ Protect the measuring device permanently against corrosion from environmental
influences.
Incorrect use
Non-designated use can compromise safety. The manufacturer is not liable for damage
caused by improper or non-designated use.
LWARNING
Danger of breakage due to corrosive or abrasive fluids!
‣ Verify the compatibility of the process fluid with the sensor material.
‣ Ensure the resistance of all fluid-wetted materials in the process.
‣ Keep within the specified pressure and temperature range.
10 Endress+Hauser
Proline Promass F 300 EtherNet/IP Basic safety instructions
NOTICE
Verification for borderline cases:
‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance
in verifying the corrosion resistance of fluid-wetted materials, but does not accept any
warranty or liability as minute changes in the temperature, concentration or level of
contamination in the process can alter the corrosion resistance properties.
Residual risks
LWARNING
The electronics and the medium may cause the surfaces to heat up. This presents a
burn hazard!
‣ For elevated fluid temperatures, ensure protection against contact to prevent burns.
LWARNING
Danger of housing breaking due to measuring tube breakage!
‣ In the event of a measuring tube breakage for a device version without rupture disk it is
possible for the pressure loading capacity of the sensor housing to be exceeded. This
can lead to rupture or failure of the sensor housing.
Endress+Hauser 11
Basic safety instructions Proline Promass F 300 EtherNet/IP
It meets general safety standards and legal requirements. It also complies with the EU
directives listed in the device-specific EU Declaration of Conformity. Endress+Hauser
confirms this by affixing the CE mark to the device.
2.6 IT security
We only provide a warranty if the device is installed and used as described in the
Operating Instructions. The device is equipped with security mechanisms to protect it
against any inadvertent changes to the device settings.
IT security measures in line with operators' security standards and designed to provide
additional protection for the device and device data transfer must be implemented by the
operators themselves.
Write protection via hardware write Not enabled. Individually following risk assessment.
protection switch → 12
12 Endress+Hauser
Proline Promass F 300 EtherNet/IP Basic safety instructions
WLAN passphrase
A connection between an operating unit (e.g. notebook or tablet) and the device via the
WLAN interface (→ 72) which can be ordered as an option is protected by the
network key. The WLAN authentication of the network key complies with the IEEE 802.11
standard.
When the device is delivered, the network key is pre-defined depending on the device. It
can be changed via the WLAN settings submenu in the WLAN passphrase parameter
(→ 150).
Endress+Hauser 13
Basic safety instructions Proline Promass F 300 EtherNet/IP
14 Endress+Hauser
Proline Promass F 300 EtherNet/IP Product description
3 Product description
The device consists of a transmitter and a sensor.
The device is available as a compact version:
The transmitter and sensor form a mechanical unit.
1 2 3 4
Ni
Spcht Pow
an un
Do nu te er
en no ngr
öf
er t op öf
f ne
Ne gize fn
so pa den when
n
us s ou
tens vr en
ionir
ter
I/O
un
Nicht
la y
i sp
e
ar
ES
C
–
5
A0029586
Endress+Hauser 15
Incoming acceptance and product identification Proline Promass F 300 EtherNet/IP
A0028673
A0028673
note?
i i
Date:
• If one of the conditions is not satisfied, contact your Endress+Hauser Sales Center.
• Depending on the device version, the CD-ROM might not be part of the delivery!
The Technical Documentation is available via the Internet or via the Endress+Hauser
Operations App, see the "Product identification" section → 17.
16 Endress+Hauser
Proline Promass F 300 EtherNet/IP Incoming acceptance and product identification
For an overview of the scope of the associated Technical Documentation, refer to the
following:
• The chapters "Additional standard documentation on the device" → 8 and
"Supplementary device-dependent documentation" → 8
• The W@M Device Viewer: Enter the serial number from the nameplate
(www.endress.com/deviceviewer)
• The Endress+Hauser Operations App: Enter the serial number from the nameplate or
scan the 2-D matrix code (QR code) on the nameplate.
1 2 3 4 5
Order code:
Ser. no.:
Ext. ord. cd.:
20 6
19
7
18
17
16
15
14 8
i i
Date: 9
13 12 11 10
A0029192
Endress+Hauser 17
Incoming acceptance and product identification Proline Promass F 300 EtherNet/IP
2
Order code: 3
Ser. no.: 4
Ext. ord. cd.: 5
14
13
i i 8
Date:
9
10
12 11
A0029199
Order code
The measuring device is reordered using the order code.
Extended order code
• The device type (product root) and basic specifications (mandatory features) are
always listed.
• Of the optional specifications (optional features), only the safety and approval-
related specifications are listed (e.g. LA). If other optional specifications are also
ordered, these are indicated collectively using the # placeholder symbol (e.g. #LA#).
• If the ordered optional specifications do not include any safety and approval-related
specifications, they are indicated by the + placeholder symbol (e.g. XXXXXX-ABCDE
+).
18 Endress+Hauser
Proline Promass F 300 EtherNet/IP Incoming acceptance and product identification
Symbol Meaning
WARNING!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious
or fatal injury.
Reference to documentation
Refers to the corresponding device documentation.
Endress+Hauser 19
Storage and transport Proline Promass F 300 EtherNet/IP
A0029252
A0029214
20 Endress+Hauser
Proline Promass F 300 EtherNet/IP Storage and transport
Endress+Hauser 21
Installation Proline Promass F 300 EtherNet/IP
6 Installation
Mounting location
A0028772
To prevent measuring errors arising from accumulation of gas bubbles in the measuring
tube, avoid the following mounting locations in the pipe:
• Highest point of a pipeline.
• Directly upstream of a free pipe outlet in a down pipe.
A0028773
22 Endress+Hauser
Proline Promass F 300 EtherNet/IP Installation
8 ³⁄₈ 6 0.24
15 ½ 10 0.40
25 1 14 0.55
40 1½ 22 0.87
50 2 28 1.10
80 3 50 1.97
100 4 65 2.60
150 6 90 3.54
Orientation
The direction of the arrow on the sensor nameplate helps you to install the sensor
according to the flow direction (direction of medium flow through the piping).
Orientation Recommendation
A Vertical orientation
A0015591
1)
B Horizontal orientation, transmitter at
top Exceptions:
→ 5, 24
A0015589
2)
C Horizontal orientation, transmitter at
bottom Exceptions:
→ 5, 24
A0015590
1) Applications with low process temperatures may decrease the ambient temperature. To maintain the
minimum ambient temperature for the transmitter, this orientation is recommended.
2) Applications with high process temperatures may increase the ambient temperature. To maintain the
maximum ambient temperature for the transmitter, this orientation is recommended.
If a sensor is installed horizontally with a curved measuring tube, match the position of the
sensor to the fluid properties.
Endress+Hauser 23
Installation Proline Promass F 300 EtherNet/IP
1 2
A0028774
A0029322 A0029323
Installation dimensions
For the dimensions and installation lengths of the device, see the "Technical
Information" document, "Mechanical construction" section.
‣ If operating outdoors:
Avoid direct sunlight, particularly in warm climatic regions.
System pressure
It is important that cavitation does not occur, or that gases entrained in the liquids do not
outgas.
24 Endress+Hauser
Proline Promass F 300 EtherNet/IP Installation
A0028777
Thermal insulation
In the case of some fluids, it is important to keep the heat radiated from the sensor to the
transmitter to a low level. A wide range of materials can be used for the required
insulation.
The following device versions are recommended for versions with thermal insulation:
• Version with extended neck for insulation:
Order code for "Sensor option", option CG with an extended neck length of
105 mm (4.13 in).
• Extended temperature version:
Order code for "Measuring tube material", option SD, SE, SF or TH with an extended neck
length of 105 mm (4.13 in).
• High-temperature version:
Order code for "Measuring tube material", option TT or TU with an extended neck length
of 142 mm (5.59 in).
NOTICE
Electronics overheating on account of thermal insulation!
‣ Recommended orientation: horizontal orientation, transmitter housing pointing
downwards.
‣ Do not insulate the transmitter housing .
‣ Maximum permissible temperature at the lower end of the transmitter housing:
80 °C (176 °F)
‣ Thermal insulation with extended neck free: the insulation is omitted around the
extended neck. We recommend that you do not insulate the extended neck in order to
ensure optimum dissipation of heat.
A0034391
Endress+Hauser 25
Installation Proline Promass F 300 EtherNet/IP
Heating
NOTICE
Electronics can overheat due to elevated ambient temperature!
‣ Observe maximum permitted ambient temperature for the transmitter .
‣ Depending on the fluid temperature, take the device orientation requirements into
account .
NOTICE
Danger of overheating when heating
‣ Ensure that the temperature at the lower end of the transmitter housing does not
exceed 80 °C (176 °F).
‣ Ensure that convection takes place on a sufficiently large scale at the transmitter neck.
‣ Ensure that a sufficiently large area of the housing support remains exposed. The
uncovered part serves as a radiator and protects the electronics from overheating and
excessive cooling.
Heating options
If a fluid requires that no heat loss should occur at the sensor, users can avail of the
following heating options:
• Electrical heating, e.g. with electric band heaters
• Via pipes carrying hot water or steam
• Via heating jackets
Vibrations
The high oscillation frequency of the measuring tubes ensures that the correct operation of
the measuring system is not influenced by plant vibrations.
Rupture disk
Information that is relevant to the process: → 227.
LWARNING
Limited functional reliability of the rupture disk.
Danger to persons from escaping fluids!
‣ Do not remove the rupture disk.
‣ When using a rupture disk, do not use a heating jacket.
‣ Make sure that the function and operation of the rupture disk is not impeded through
the installation of the device.
‣ Take precautions to prevent damage and danger to persons if the rupture disk is
actuated.
‣ Observe information on the rupture disk sticker.
The position of the rupture disk is indicated on a sticker beside it.
26 Endress+Hauser
Proline Promass F 300 EtherNet/IP Installation
The existing connecting nozzles are not intended for the purpose of rinsing or pressure
monitoring, but instead serve as the mounting location for the rupture disk.
In the event of a failure of the rupture disk, a discharge device can be screwed onto the
internal thread of the rupture disk in order to drain off any escaping medium.
RUPTURE DISK
2 3 1 2 3
A0028903
Protective cover
A0029553
Endress+Hauser 27
Installation Proline Promass F 300 EtherNet/IP
Cover locking
NOTICE
Order code "Housing", option L "Cast, stainless": The covers of the transmitter housing
are provided with a borehole to lock the cover.
The cover can be locked using screws and a chain or cable provided by the customer.
‣ It is recommended to use stainless steel cables or chains.
‣ If a protective coating is applied, it is recommended to use a heat shrink tube to protect
the housing paint.
1
1
1
2 SW 2.5
5 (0.2) min. 15 (0.6)
M3
A0029800
For sensor
For flanges and other process connections: Corresponding mounting tools
28 Endress+Hauser
Proline Promass F 300 EtherNet/IP Installation
2. Install the measuring device or turn the transmitter housing so that the cable entries
do not point upwards.
A0029263
4.
öf
öf
f ne
f ne
n
n
te r
te r
un
un
Nicht
Nicht
ES
ES C
C
– –
+ +
E E
e
e
ar ar
3 mm 1. 4 mm 3.
2.
A0029993
1. Depending on the device version: Loosen the securing clamp of the connection
compartment cover.
2. Unscrew the connection compartment cover.
3. Release the fixing screw.
4. Turn the housing to the desired position.
5. Firmly tighten the securing screw.
6. Screw on the connection compartment cover
7. Depending on the device version: Attach the securing clamp of the connection
compartment cover.
Endress+Hauser 29
Installation Proline Promass F 300 EtherNet/IP
öf
öf
f ne
f ne
n
n
te r
te r
un
un
Nicht
Nicht
ES
ES C
C
– –
+ +
ES
C
E – E
e
e
ar ar
3 mm 1. 3.
2.
A0030035
1. Depending on the device version: Loosen the securing clamp of the connection
compartment cover.
2. Unscrew the connection compartment cover.
3. Turn the display module to the desired position: max. 8 × 45° in each direction.
4. Screw on the connection compartment cover.
5. Depending on the device version: Attach the securing clamp of the connection
compartment cover.
Does the arrow on the sensor nameplate match the direction of flow of the fluid through the
piping → 23?
Are the measuring point identification and labeling correct (visual inspection)?
30 Endress+Hauser
Proline Promass F 300 EtherNet/IP Electrical connection
7 Electrical connection
NOTICE
The measuring device does not have an internal circuit breaker.
‣ For this reason, assign the measuring device a switch or power-circuit breaker so that
the power supply line can be easily disconnected from the mains.
‣ Although the measuring device is equipped with a fuse, additional overcurrent
protection (maximum 10 A) should be integrated into the system installation.
Electrical safety
In accordance with applicable federal/national regulations.
Signal cable
EtherNet/IP
The standard ANSI/TIA/EIA-568-B.2 Annex specifies CAT 5 as the minimum category for
a cable used for EtherNet/IP. CAT 5e and CAT 6 are recommended.
For more information on planning and installing EtherNet/IP networks, please refer
to the "Media Planning and Installation Manual. EtherNet/IP" of ODVA Organization
Pulse/frequency/switch output
Standard installation cable is sufficient.
Endress+Hauser 31
Electrical connection Proline Promass F 300 EtherNet/IP
Relay output
Standard installation cable is sufficient.
Status input
Standard installation cable is sufficient.
Cable diameter
• Cable glands supplied:
M20 × 1.5 with cable ⌀ 6 to 12 mm (0.24 to 0.47 in)
• Spring-loaded terminals: Suitable for strands and strands with ferrules.
Conductor cross-section 0.2 to 2.5 mm2 (24 to 12 AWG).
Standard cable 2 × 2 × 0.34 mm2 (22 AWG) PVC cable with common shield (2 pairs, pair-
stranded)
Available cable length 5 m (15 ft)/10 m (35 ft)/20 m (65 ft)/30 m (100 ft)
Operating temperature When mounted in a fixed position: –50 to +105 °C (–58 to +221 °F); when cable
can move freely: –25 to +105 °C (–13 to +221 °F)
Cable length Maximum 300 m (1 000 ft), see the following table
32 Endress+Hauser
Proline Promass F 300 EtherNet/IP Electrical connection
Endress+Hauser 33
Electrical connection Proline Promass F 300 EtherNet/IP
"Electrical connection" 2 3
L, N, P, U Connector M12 × 1 –
1) 2) 1) 2) 1) 2) 1) 2)
R ,S ,T ,V Connector M12 × 1 Connector M12 × 1
1) Cannot be combined with an external WLAN antenna (order code for "Enclosed accessories", option P8) of
an RJ45 M12 adapter for the service interface (order code for "Accessories mounted", option NB) or of the
remote display and operating module DKX001
2) Suitable for integrating the device in a ring topology.
2 Pin Assignment
1 + Tx
2 + Rx
1 3
3 - Tx
4 - Rx
4
A0016812
Coding Plug/socket
D Socket
34 Endress+Hauser
Proline Promass F 300 EtherNet/IP Electrical connection
1
2
3
4
A0026781
In addition to connecting the device via EtherNet/IP and the available inputs/outputs,
additional connection options are also available:
• Integrate into a network via the service interface (CDI-RJ45) → 39.
• Integrate the device into a ring topology → 40.
3.
öf
öf
f ne
f ne
n
n
te r
ter
un
un
Nicht
Nicht
ES
ES C
C
– –
+ +
E E
ES
C
–
e
e
ar + ar
4. E
3 mm 1. 3.
2.
A0029813
Endress+Hauser 35
Electrical connection Proline Promass F 300 EtherNet/IP
Nich Po Nich Po
Sp t Sp t
an un w an un w
Do nu te er Do nu te er
en no ngr en no ngr
öf
öf
er t
f ne
er t
f ne
öf öf
Ne gi op fn Ne gi op fn
so pa zeden when so pa zeden when
n
us s ou er us s ou
tens vr en tens vr en
ionir Powr en ionir
te fn n
un g öf whe
te r
te r
I/O ichtnnun en I/O
un
un
Npa t op
6. S noze d vrir
Doergis ouion
Nicht
Nicht
en patens
Neus
so
I/O
e
ar ar
ES ES
C C
– –
+ +
E E
5.
A0029814
11.
22 mm
24 mm 9.
7.
N
Spicht Pow
a u
8.
D nn nte er
eno no ungr
öf
f ne
e t ö
N rgizop ffn
soe pa eden w en
n
us s o he
ten uv n
siorir
n
te r
un
t
Nich
≤4
e
2
ar
(1
.6
)
10.
A0033722
7. Push the cable through the cable entry . To ensure tight sealing, do not remove the
sealing ring from the cable entry.
8. Strip the cable and cable ends and connect to the RJ45 connector.
9. Connect the protective ground.
10. Plug in the RJ45 connector.
11. Firmly tighten the cable glands.
This concludes the EtherNet/IP connection process.
36 Endress+Hauser
Proline Promass F 300 EtherNet/IP Electrical connection
1.
2.
10 (0.4)
öf
f ne
n
ter
un
Nicht
mm (in)
e
ar
ES
C
–
3.
+
A0033983
1. Push the cable through the cable entry . To ensure tight sealing, do not remove the
sealing ring from the cable entry.
2. Strip the cable and cable ends. In the case of stranded cables, also fit ferrules.
3. Connect the protective ground.
5.
öf
f ne
n
22 mm
te r
un
Nicht
24 mm
e
ar
ES
C
–
4.
A0033984
Endress+Hauser 37
Electrical connection Proline Promass F 300 EtherNet/IP
Removing a cable
3 (0.12)
1.
2.
A0029598
1. To remove a cable from the terminal, use a flat-blade screwdriver to push the slot
between the two terminal holes
2. while simultaneously pulling the cable end out of the terminal.
38 Endress+Hauser
Proline Promass F 300 EtherNet/IP Electrical connection
N
Spicht Pow
a u
Do nnunte er
en no ngr
öf
f ne
e t ö
Ne rgizope ffne
so pa ed n w n
n
us s o he
ten uv n
siorir
n
te r
I/O
un
t
Nich
e
ar
≤4
2
(1
.6
)
1
A0033703
Endress+Hauser 39
Electrical connection Proline Promass F 300 EtherNet/IP
N
Spicht Pow
a u
Do nnunte er
en no ngr
öf
f ne
e t ö
Ne rgizope ffne
so pa ed n w n
n
us s o he
ten uv n
siorir
n
te r
un
t
Nich
e
ar
≤4
2
(1
.6
)
1
A0033717
1 EtherNet/IP connection
2 Service interface (CDI-RJ45)
40 Endress+Hauser
Proline Promass F 300 EtherNet/IP Electrical connection
Gnd
Vcc
A B
81 82 83 84
1
2
3
81 82 83 84
4
Gnd
A B
Vcc
5
A0027518
7.3.1 Requirements
No special measures for potential equalization are required.
Endress+Hauser 41
Electrical connection Proline Promass F 300 EtherNet/IP
EtherNet/IP
1 2 3 4
A0028767
1 2 3 4
A0027544
1 2
3
4...20 mA
A0028758
42 Endress+Hauser
Proline Promass F 300 EtherNet/IP Electrical connection
1 2 3
4
4...20 mA
A0028759
Pulse/frequency output
1 2
12345
A0028761
Switch output
1 2
A0028760
Endress+Hauser 43
Electrical connection Proline Promass F 300 EtherNet/IP
Relay output
1 2
A0028760
Current input
1 2 3
A0028915
Status input
1 2
A0028764
44 Endress+Hauser
Proline Promass F 300 EtherNet/IP Electrical connection
Addressing data
Addressing mode ex Software addressing; all DIP switches for hardware addressing are set to OFF.
works
1 128
IP Address setting
2 64
3 32
(last octet)
4 16
5 8
6 4
7 2
8 1
A0029635
1. Depending on the housing version, loosen the securing clamp or fixing screw of the
housing cover.
2. Depending on the housing version, unscrew or open the housing cover and
disconnect the local display from the main electronics module where necessary.
3. Set the desired IP address using the corresponding DIP switches on the I/O
electronics module.
4. Reverse the removal procedure to reassemble the transmitter.
5. Reconnect the device to the power supply.
The configured device address is used once the device is restarted.
Endress+Hauser 45
Electrical connection Proline Promass F 300 EtherNet/IP
Off On
1
2
3
4
A0034499
1. Depending on the housing version, loosen the securing clamp or fixing screw of the
housing cover.
2. Depending on the housing version, unscrew or open the housing cover and
disconnect the local display from the main electronics module where necessary .
3. Set DIP switch No. 2 on the I/O electronics module from OFF → ON.
4. Reverse the removal procedure to reassemble the transmitter.
5. Reconnect the device to the power supply.
The default IP address is used once the device is restarted.
To guarantee IP66/67 degree of protection, Type 4X enclosure, carry out the following
steps after the electrical connection:
1. Check that the housing seals are clean and fitted correctly.
2. Dry, clean or replace the seals if necessary.
3. Tighten all housing screws and screw covers.
4. Firmly tighten the cable glands.
5. To ensure that moisture does not enter the cable entry:
Route the cable so that it loops down before the cable entry ("water trap").
A0029278
46 Endress+Hauser
Proline Promass F 300 EtherNet/IP Electrical connection
Are all the cable glands installed, firmly tightened and leak-tight? Cable run with "water trap"
→ 46 ?
Endress+Hauser 47
Operation options Proline Promass F 300 EtherNet/IP
8 Operation options
ESC
- + E
1 2 3 4 5
A0034513
48 Endress+Hauser
Proline Promass F 300 EtherNet/IP Operation options
Parameter 1
Operator
Parameter n
Submenu 1
Submenu n
Wizard 1 / Parameter 1
Maintenance
Task-oriented
Wizard n / Parameter n
Parameter 1
Parameter n
Submenu 1
Submenu n
Diagnostics Parameter 1
Parameter n
Submenu 1
Submenu n
Parameter n
System
Function-oriented
Sensor
Expert
Input
Output
Communication
Application
Diagnostics
A0018237-EN
Endress+Hauser 49
Operation options Proline Promass F 300 EtherNet/IP
Diagnostics "Maintenance" role Contains all parameters for error detection and analyzing process and device
Fault elimination: errors:
• Diagnostics and elimination of • Diagnostic list
process and device errors Contains up to 5 currently pending diagnostic messages.
• Measured value simulation • Event logbook
Contains event messages that have occurred.
• Device information
Contains information for identifying the device.
• Measured values
Contains all current measured values.
• Data logging submenu with "Extended HistoROM" order option
Storage and visualization of measured values
• Heartbeat
The functionality of the device is checked on demand and the verification
results are documented.
• Simulation
Is used to simulate measured values or output values.
50 Endress+Hauser
Proline Promass F 300 EtherNet/IP Operation options
Expert function-oriented Tasks that require detailed Contains all the parameters of the device and makes it possible to access
knowledge of the function of the these parameters directly using an access code. The structure of this menu is
device: based on the function blocks of the device:
• Commissioning measurements • System
under difficult conditions Contains all higher-order device parameters which do not concern the
• Optimal adaptation of the measurement or the communication interface.
measurement to difficult • Sensor
conditions Configuration of the measurement.
• Detailed configuration of the • Input
communication interface Configuring the status input.
• Error diagnostics in difficult cases • Output
Configuring of the analog current outputs as well as the pulse/frequency
and switch output.
• Communication
Configuration of the digital communication interface and the Web server.
• Application
Configure the functions that go beyond the actual measurement (e.g.
totalizer).
• Diagnostics
Error detection and analysis of process and device errors and for device
simulation and Heartbeat Technology.
2 XXXXXXXXX F 3
4 1120.50
kg/h
A0029348
1 Operational display
2 Device tag
3 Status area
4 Display area for measured values (4-line)
5 Operating elements→ 57
Status area
The following symbols appear in the status area of the operational display at the top right:
• Status signals→ 177
– F: Failure
– C: Function check
– S: Out of specification
– M: Maintenance required
• Diagnostic behavior→ 178
– : Alarm
– : Warning
• : Locking (the device is locked via the hardware )
• : Communication (communication via remote operation is active)
Endress+Hauser 51
Operation options Proline Promass F 300 EtherNet/IP
Display area
In the display area, each measured value is prefaced by certain symbol types for further
description:
↓ ↓ ↓
Example
Measured values
Symbol Meaning
Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
Temperature
Totalizer
displayed.
The measurement channel number indicates which of the three totalizers is
Output
Measurement channel 1 to 4
The measurement channel number is displayed only if more than one channel is present for the same measured
variable type (e.g. Totalizer 1 to 3).
Diagnostic behavior
The diagnostic behavior pertains to a diagnostic event that is relevant to the displayed measured variable.
For information on the symbols → 178
The number and display format of the measured values can be configured via the
Format display parameter (→ 122).
52 Endress+Hauser
Proline Promass F 300 EtherNet/IP Operation options
1 1
5 5
A0013993-EN A0013995-EN
1 Navigation view
2 Navigation path to current position
3 Status area
4 Display area for navigation
5 Operating elements → 57
Navigation path
The navigation path - displayed at the top left in the navigation view - consists of the
following elements:
↓ ↓ ↓
/ ../ Display
For more information about the icons in the menu, refer to the "Display area" section
→ 54
Status area
The following appears in the status area of the navigation view in the top right corner:
• In the submenu
– The direct access code for the parameter you are navigating to (e.g. 0022-1)
– If a diagnostic event is present, the diagnostic behavior and status signal
• In the wizard
If a diagnostic event is present, the diagnostic behavior and status signal
• For information on the diagnostic behavior and status signal → 177
• For information on the function and entry of the direct access code → 59
Endress+Hauser 53
Operation options Proline Promass F 300 EtherNet/IP
Display area
Menus
Symbol Meaning
Operation
Appears:
• In the menu next to the "Operation" selection
• At the left in the navigation path in the Operation menu
Setup
Appears:
• In the menu next to the "Setup" selection
• At the left in the navigation path in the Setup menu
Diagnostics
Appears:
• In the menu next to the "Diagnostics" selection
• At the left in the navigation path in the Diagnostics menu
Expert
Appears:
• In the menu next to the "Expert" selection
• At the left in the navigation path in the Expert menu
Submenu
Wizard
Parameter locked
When displayed in front of a parameter name, indicates that the parameter is locked.
• By a user-specific access code
• By the hardware write protection switch
Wizard operation
Symbol Meaning
54 Endress+Hauser
Proline Promass F 300 EtherNet/IP Operation options
Numeric editor
1 +0.000 Xx
0 1 2 3 4
2 5 6 7 8 9
+ –
3
A0034250
Text editor
1 Xxxxxxx
A B C D E F G H I J K
2 L M N O P Q R S T U V W
X Y Z 6
3 7
5
4
A0034114
Minus key
Move the entry position to the left.
Plus key
Move the entry position to the right.
Endress+Hauser 55
Operation options Proline Promass F 300 EtherNet/IP
Enter key
• Press the key briefly: confirm your selection.
• Press the key for 2 s: confirm the entry.
Input screens
Symbol Meaning
Upper case
Lower case
Numbers
Symbol Meaning
Reject entry
Confirm entry
56 Endress+Hauser
Proline Promass F 300 EtherNet/IP Operation options
Minus key
In a menu, submenu
Moves the selection bar upwards in a choose list.
With a Wizard
Confirms the parameter value and goes to the previous parameter.
With a text and numeric editor
Move the entry position to the left.
Plus key
In a menu, submenu
Moves the selection bar downwards in a choose list.
With a Wizard
Confirms the parameter value and goes to the next parameter.
With a text and numeric editor
Move the entry position to the right.
Enter key
For operational display
• Pressing the key briefly opens the operating menu.
• Pressing the key for 2 s opens the context menu including the option for activating the
keypad lock.
In a menu, submenu
• Pressing the key briefly:
– Opens the selected menu, submenu or parameter.
– Starts the wizard.
– If help text is open, closes the help text of the parameter.
• Pressing the key for 2 s for parameter:
If present, opens the help text for the function of the parameter.
With a Wizard
Opens the editing view of the parameter.
With a text and numeric editor
• Press the key briefly: confirm your selection.
• Press the key for 2 s: confirm the entry.
Endress+Hauser 57
Operation options Proline Promass F 300 EtherNet/IP
XXXXXXXXXX
Setup
Conf.backup
Simulation
Keylock on
l/h
A0034608-EN
2. Press + simultaneously.
The context menu is closed and the operational display appears.
58 Endress+Hauser
Proline Promass F 300 EtherNet/IP Operation options
XXXXXXXXX
20.50
Main menu 0104-1
Language
1. English
Display/operat.
Setup
Main menu
Language
2. Display/operat.
Setup
/ ../Display/operat. 0091-1
Access stat.disp
3. Operator
Locking status
Display
/ ../Display/operat.
Locking status
4. Display
/ ../Display 0098-1
Format display
5. 1 value, max.
Contrast display
Display intervall
/ ../Format display 0098-1
1 value, max.
6. Bargr. + 1 value
2 values
Val. large+2val.
/ ../Format display 0098-1
1 value, max.
7. Bargr. + 1 value
2 values
Val. large+2val.
XXXXXXXXX
10.50
8. 2s mA
2800
Hz
A0029562-EN
Endress+Hauser 59
Operation options Proline Promass F 300 EtherNet/IP
The direct access code consists of a 5-digit number (at maximum) and the channel
number, which identifies the channel of a process variable: e.g. 00914-2. In the navigation
view, this appears on the right-hand side in the header of the selected parameter.
00914-2 1
A0029414
A0014002-EN
2. Press + simultaneously.
The help text is closed.
60 Endress+Hauser
Proline Promass F 300 EtherNet/IP Operation options
For a description of the editing view - consisting of the text editor and numeric editor
- with symbols → 55, for a description of the operating elements → 57
1) The user only has write access after entering the access code.
1) Despite the defined access code, certain parameters can always be modified and thus are excepted from
the write protection, as they do not affect the measurement. Refer to the "Write protection via access code"
section
The user role with which the user is currently logged on is indicated by the Access
status parameter. Navigation path: Operation → Access status
Parameter write protection via local operation can be disabled by entering the user-specific
access code in the Enter access code parameter (→ 135) via the respective access
option.
1. After you press , the input prompt for the access code appears.
Endress+Hauser 61
Operation options Proline Promass F 300 EtherNet/IP
If the user attempts to access the operating menu while the keypad lock is active, the
Keylock on message appears.
62 Endress+Hauser
Proline Promass F 300 EtherNet/IP Operation options
8.4.2 Prerequisites
Computer hardware
Hardware Interface
CDI-RJ45 WLAN
Interface The computer must have an RJ45 The operating unit must have a
interface. WLAN interface.
Connection Standard Ethernet cable with RJ45 Connection via Wireless LAN.
connector.
Computer software
Software Interface
CDI-RJ45 WLAN
Computer settings
Settings Interface
CDI-RJ45 WLAN
User rights Appropriate user rights (e.g. administrator rights) for TCP/IP and proxy server
settings are necessary (for adjusting the IP address, subnet mask etc.).
Proxy server settings of the The Web browser setting Use a Proxy Server for Your LAN must be
Web browser deselected .
Ifenter
JavaScript cannot be enabled:
http://192.168.1.212/basic.html in the address line of the Web
browser. A fully functional but simplified version of the operating menu
structure starts in the Web browser.
Network connections Only the active network connections to the measuring device should be used.
Switch off all other network Switch off all other network
connections such as WLAN. connections.
Endress+Hauser 63
Operation options Proline Promass F 300 EtherNet/IP
Web server Web server and WLAN must be enabled; factory setting: ON
64 Endress+Hauser
Proline Promass F 300 EtherNet/IP Operation options
IP address 192.168.1.XXX; for XXX all numerical sequences except: 0, 212 and 255 → e.g.
192.168.1.213
Endress+Hauser 65
Operation options Proline Promass F 300 EtherNet/IP
To ensure the safe and swift assignment of the WLAN network to the measuring
point, it is advisable to change the SSID name. It should be possible to clearly assign
the new SSID name to the measuring point (e.g. tag name) because it is displayed as
the WLAN network.
Disconnecting
‣ After configuring the device:
Terminate the WLAN connection between the operating unit and measuring device.
66 Endress+Hauser
Proline Promass F 300 EtherNet/IP Operation options
2. Enter the IP address of the Web server in the address line of the Web browser:
192.168.1.212
The login page appears.
1 2 3 4 5
7
8
10
A0029417
1 Picture of device
2 Device name
3 Device tag
4 Status signal
5 Current measured values
6 Operating language
7 User role
8 Access code
9 Login
10 Reset access code (→ 153)
8.4.4 Logging on
1. Select the preferred operating language for the Web browser.
2. Enter the user-specific access code.
3. Press OK to confirm your entry.
Endress+Hauser 67
Operation options Proline Promass F 300 EtherNet/IP
1
2
A0029418
1 Function row
2 Local display language
3 Navigation area
Header
The following information appears in the header:
• Device name
• Device tag
• Device status with status signal → 180
• Current measured values
Function row
Functions Meaning
Device status Displays the diagnostic messages currently pending, listed in order of priority
Configuration and checking of all the parameters required for establishing the connection to
Network the measuring device:
configuration • Network settings (e.g. IP address, MAC address)
• Device information (e.g. serial number, firmware version)
68 Endress+Hauser
Proline Promass F 300 EtherNet/IP Operation options
Navigation area
If a function is selected in the function bar, the submenus of the function open in the
navigation area. The user can now navigate through the menu structure.
Working area
Depending on the selected function and the related submenus, various actions can be
performed in this area:
• Configuring parameters
• Reading measured values
• Calling up help text
• Starting an upload/download
Navigation
"Expert" menu → Communication → Web server
Web server functionality Switch the Web server on and off. • Off On
• HTML Off
• On
Option Description
Endress+Hauser 69
Operation options Proline Promass F 300 EtherNet/IP
3. If no longer needed:
Reset modified properties of the Internet protocol (TCP/IP) → 65.
If communication with the Web server was established via the default IP address
192.168.1.212, DIP switch No. 10 must be reset (from ON → OFF). Afterwards, the IP
address of the device is active again for network communication.
Star topology
1 2 3
5 5 5
A0032078
Ring topology
The device is integrated via the terminal connection for signal transmission (output 1) and
the service interface (CDI-RJ45).
70 Endress+Hauser
Proline Promass F 300 EtherNet/IP Operation options
1 2 3
5 5 5
A0033725
Service interface
Via service interface (CDI-RJ45)
A point-to-point connection can be established to configure the device onsite. With the
housing open, the connection is established directly via the service interface (CDI-RJ45) of
the device.
An adapter for RJ45 and the M12 connector is optionally available:
Order code for "Accessories", option NB: "Adapter RJ45 M12 (service interface)"
The adapter connects the service interface (CDI-RJ45) to an M12 connector mounted
in the cable entry. Therefore the connection to the service interface can be established
via an M12 connector without opening the device.
A0027563
Endress+Hauser 71
Operation options Proline Promass F 300 EtherNet/IP
3 4 1 2 3 4
5 6 7
A0034570
72 Endress+Hauser
Proline Promass F 300 EtherNet/IP Operation options
NOTICE
In principle, avoid simultaneous access to the measuring device via the service
interface (CDI-RJ45) and the WLAN interface from the same mobile terminal. This
could cause a network conflict.
‣ Only activate one service interface (CDI-RJ45 service interface or WLAN interface).
‣ If simultaneous communication is necessary: configure different IP address ranges, e.g.
192.168.0.1 (WLAN interface) and 192.168.1.212 (CDI-RJ45 service interface).
To ensure the safe and swift assignment of the WLAN network to the measuring
point, it is advisable to change the SSID name. It should be possible to clearly assign
the new SSID name to the measuring point (e.g. tag name) because it is displayed as
the WLAN network.
Disconnecting
‣ After configuring the device:
Terminate the WLAN connection between the operating unit and measuring device.
8.5.2 FieldCare
Function scope
FDT-based plant asset management tool from Endress+Hauser. It can configure all smart
field devices in a system and helps you manage them. By using the status information, it is
also a simple but effective way of checking their status and condition.
Access is via:
• CDI-RJ45 service interface → 71
• WLAN interface→ 72
Typical functions:
• Configuring parameters of transmitters
• Loading and saving device data (upload/download)
• Documentation of the measuring point
• Visualization of the measured value memory (line recorder) and event logbook
For additional information about FieldCare, see Operating Instructions BA00027S
and BA00059S
Establishing a connection
1. Start FieldCare and launch the project.
Endress+Hauser 73
Operation options Proline Promass F 300 EtherNet/IP
User interface
2 3 4 5 6 7
Xxxxxx/…/…/
Device name: Xxxxxxx Mass flow: 12.34 kg/h
Device tag: Xxxxxxx
1 Volume flow: 12.34 m³/h
Status: Good
10 11
A0021051-EN
1 Header
2 Picture of device
3 Device name
4 Tag name
5 Status area with status signal→ 180
6 Display area for current measured values
7 Edit toolbar with additional functions such as save/restore, event list and create documentation
8 Navigation area with operating menu structure
9 Working area
10 Range of action
11 Status area
8.5.3 DeviceCare
Function scope
Tool to connect and configure Endress+Hauser field devices.
74 Endress+Hauser
Proline Promass F 300 EtherNet/IP Operation options
The fastest way to configure Endress+Hauser field devices is with the dedicated
"DeviceCare" tool. Together with the device type managers (DTMs) it presents a convenient,
comprehensive solution.
For details, see Innovation Brochure IN01047S
Endress+Hauser 75
System integration Proline Promass F 300 EtherNet/IP
9 System integration
For an overview of the different firmware versions for the device → 202
Add-on Profile • Major System file for "Studio 5000" www.endress.com → Download-
Level 3 revision 1 software (Rockwell Automation) Area
• Minor
revision 1
76 Endress+Hauser
Proline Promass F 300 EtherNet/IP System integration
Endress+Hauser 77
System integration Proline Promass F 300 EtherNet/IP
Possible configurations
78 Endress+Hauser
Proline Promass F 300 EtherNet/IP System integration
Possible connections
No. #1 #2 #3 #4 #5 #6 #7 #8 #9
Number of connections 1 1 1 1 1 1 1 1 1
6. Temperature 21-24
7. Density 25-28
9. Totalizer 1 33-36
Endress+Hauser 79
System integration Proline Promass F 300 EtherNet/IP
Description Byte
4. Density 13-16
5. Temperature 17-20
6. Totalizer 1 21-24
5. Totalizer 2 45-48
6. Totalizer 3 49-52
80 Endress+Hauser
Proline Promass F 300 EtherNet/IP System integration
Description Byte
8. Concentration 57-60
Endress+Hauser 81
System integration Proline Promass F 300 EtherNet/IP
Description Byte
82 Endress+Hauser
Proline Promass F 300 EtherNet/IP System integration
1. Totalizer 1 1 0
2. Totalizer 2 1
3. Totalizer 3 2
4. Verification 3 • 0: Enable
5. Concentration of medium type 4 • 1: Disable
6. Compensation, pressure 5
8. Compensation, temperature 7
Endress+Hauser 83
System integration Proline Promass F 300 EtherNet/IP
22. Select concentration of medium type 13-14 16 • 3062 (0) : Aqueous Fructose
• 3063 (0) : Aqueous Glucose
• 3068 (0) : Aqueous Hydrochloric Acid
• 3077 (0) : Aqueous Hydrogen
Peroxide
• 3065 (0) : Aqueous Sucrose
• 3064 (0) : Aqueous Invert Sugar
• 3069 (0) : Aqueous Nitric Acid
• 3070 (0) : Aqueous Phosphoric Acid
• 3075 (0) : Aqueous Potassium
Hydroxide
• 3071 (0) : Aqueous Sodium Hydroxide
• 3060 (0) : Ethanol Water
• 3061 (0) : Methanol Water
• 3066 (0) : Ammonium Nitrate In
Water
• 3067 (0) : Ferric Chloride In Water
• 3073 (0) : High Fructose Corn Syrup
42
• 3074 (0) : High Fructose Corn Syrup
55
• 3072 (0) : High Fructose Corn Syrup
90
• 3092 (0) : Percent Volume / Percent
Mass
• 3081 (0) : Wort
• 3082 (0) : Coef Set 1
• 3083 (0) : Coef Set 2
• 3084 (0) : Coef Set 3
28. External reference density unit (integer) 29-30 16 • 32840 (240): kg/Nm3
• 32841 (240): kg/Nl
• 32842 (240): g/Scm3
• 32843 (240): kg/Scm3
• 32844 (240): lb/Sft3
84 Endress+Hauser
Proline Promass F 300 EtherNet/IP System integration
1. None 32 4 0
3. None 8 1 5
13. None 16 2 24
21. None 32 4 40
22. None 16 2 44
Endress+Hauser 85
System integration Proline Promass F 300 EtherNet/IP
64. Parameter 37 Low flow cut off Assign process variable 16 2 134
66. Parameter 41 Corrected volume flow calculation Corrected volume flow calculation 16 2 138
67. Parameter 188 Low flow cut off On value low flow cut off 32 4 140
68. Parameter 187 Low flow cut off Off value low flow cutoff 32 4 144
69. Parameter 209 Low flow cut off Pressure shock suppression 32 4 148
70. Parameter 191 Empty pipe detection Low value partial filled pipe detection 32 4 152
71. Parameter 189 Partially filled pipe detection High value partial filled pipe detection 32 4 156
72. Parameter 190 Empty pipe detection Response time part. filled pipe detect. 32 4 160
86 Endress+Hauser
Proline Promass F 300 EtherNet/IP System integration
73. Parameter 182 Corrected volume flow calculation Fixed reference density 32 4 164
74. Parameter 186 Corrected volume flow calculation Linear expansion coefficient 32 4 168
75. Parameter 211 Corrected volume flow calculation Square expansion coefficient 32 4 172
76. Parameter 210 Corrected volume flow calculation Reference temperature 32 4 176
85. Parameter 133 Medium selection Temperature coefficient sound velocity 32 4 204
87. Parameter 115 Empty pipe detection Max. damping empty pipe detection 32 4 212
Endress+Hauser 87
System integration Proline Promass F 300 EtherNet/IP
000 – 0
88 Endress+Hauser
Proline Promass F 300 EtherNet/IP System integration
Endress+Hauser 89
System integration Proline Promass F 300 EtherNet/IP
90 Endress+Hauser
Proline Promass F 300 EtherNet/IP System integration
Endress+Hauser 91
System integration Proline Promass F 300 EtherNet/IP
92 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
10 Commissioning
XXXXXXXXX
20.50
XX XX
Main menu 0104-1
1. Display language
English
Operation
Setup
Display language 0104-1
2. Ã English
Deutsch
Español
Français
Display language 0104-1
3. Ã English
Deutsch
Español
Français
Hauptmenü 0104-1
4. Sprache
Deutsch
Betrieb
Setup
A0029420
Endress+Hauser 93
Commissioning Proline Promass F 300 EtherNet/IP
XXXXXXXXX
20.50
mA
Main menu 0104-1
Display language
1. English
Display/operat.
Setup
Main menu
Display/operat.
2. Setup
Diagnostic
/ ../Setup
Medium selection
3.
XXXXXXXXX
XXXXXXXXX
A0032222-EN
Depending on the device version, not all submenus and parameters are available in
every device. The selection can vary depending on the order code.
Setup
Device tag → 95
‣ System units → 95
‣ Communication → 97
‣ Medium selection → 99
94 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
‣ Display → 121
1 XXXXXXXXX
A0029422
Navigation
"Setup" menu → Device tag
Device tag Enter the name for the measuring point. Max. 32 characters, such as Promass
letters, numbers or special
characters (e.g. @, %, /).
Navigation
"Setup" menu → System units
‣ System units
Mass unit → 96
Endress+Hauser 95
Commissioning Proline Promass F 300 EtherNet/IP
Volume unit → 96
Density unit → 96
Temperature unit → 97
Pressure unit → 97
Mass flow unit Select mass flow unit. Unit choose list Country-specific:
• kg/h
Result
• lb/min
The selected unit applies for:
• Output
• Low flow cut off
• Simulation process variable
Volume flow unit Select volume flow unit. Unit choose list Country-specific:
• l/h
Result
• gal/min (us)
The selected unit applies for:
• Output
• Low flow cut off
• Simulation process variable
Corrected volume flow unit Select corrected volume flow unit. Unit choose list Country-specific:
• Nl/h
Result
• Sft³/min
The selected unit applies for:
Corrected volume flow parameter
(→ 161)
Corrected volume unit Select corrected volume unit. Unit choose list Country-specific:
• Nl
• Sft³
96 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Reference density unit Select reference density unit. Unit choose list Country-dependent
• kg/Nl
• lb/Sft³
Pressure unit Select process pressure unit. Unit choose list Country-specific:
• bar a
Result
• psi a
The unit is taken from:
• Pressure value parameter (→ 100)
• External pressure parameter
(→ 100)
• Pressure value
Navigation
"Setup" menu → Communication
‣ Communication
MAC address → 98
DHCP client → 98
IP address → 98
Subnet mask → 98
Default gateway → 98
Endress+Hauser 97
Commissioning Proline Promass F 300 EtherNet/IP
MAC address Displays the MAC address of the measuring Unique 12-digit character Each measuring device is given
device. string comprising letters and an individual address.
numbers, e.g.:
MAC = Media Access Control 00:07:05:10:01:5F
Default network settings Select whether to restore network settings. • Off Off
• On
IP address Displays the IP address of the Web server of 4 octet: 0 to 255 (in the 192.168.1.212
the measuring device. particular octet)
Subnet mask Displays the subnet mask. 4 octet: 0 to 255 (in the 255.255.255.0
particular octet)
Default gateway Displays the default gateway. 4 octet: 0 to 255 (in the 0.0.0.0
particular octet)
98 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Navigation
"Setup" menu → Select medium
‣ Medium selection
Endress+Hauser 99
Commissioning Proline Promass F 300 EtherNet/IP
Select gas type The Gas option is selected in Select measured gas type. • Air Methane CH4
the Select medium parameter. • Ammonia NH3
• Argon Ar
• Sulfur hexafluoride
SF6
• Oxygen O2
• Ozone O3
• Nitrogen oxide
NOx
• Nitrogen N2
• Nitrous oxide N2O
• Methane CH4
• Hydrogen H2
• Helium He
• Hydrogen chloride
HCl
• Hydrogen sulfide
H2S
• Ethylene C2H4
• Carbon dioxide
CO2
• Carbon monoxide
CO
• Chlorine Cl2
• Butane C4H10
• Propane C3H8
• Propylene C3H6
• Ethane C2H6
• Others
Reference sound velocity In the Select gas type Enter sound velocity of gas at 1 to 99 999.9999 m/ 415.0 m/s
parameter, the Others option 0 °C (32 °F). s
is selected.
Temperature coefficient sound The Others option is selected Enter temperature coefficient Positive floating- 0 (m/s)/K
velocity in the Select gas type for the gas sound velocity. point number
parameter.
Pressure value The Fixed value option is Enter process pressure to be Positive floating- 0 bar
selected in the Pressure used for pressure correction. point number
compensation parameter.
External pressure The External value option is Shows the external process Positive floating- 0 bar
selected in the Pressure pressure value. point number
compensation parameter.
100 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Navigation
"Setup" menu → I/O configuration
‣ I/O configuration
I/O module 1 to n terminal numbers Shows the terminal numbers used by the I/O • Not used –
module. • 26-27 (I/O 1)
• 24-25 (I/O 2)
• 22-23 (I/O 3)
I/O module 1 to n information Shows information of the plugged I/O • Not plugged –
module. • Invalid
• Not configurable
• Configurable
• EtherNet/IP
I/O module 1 to n type Shows the I/O module type. • Off Off
• Current output *
• Current input *
• Status input *
• Pulse/frequency/switch
output *
Alteration code Enter the code in order to change the I/O Positive integer 0
configuration.
Navigation
"Setup" menu → Current input
‣ Current input 1 to n
Endress+Hauser 101
Commissioning Proline Promass F 300 EtherNet/IP
20 mA value → 102
Signal mode The measuring device is not Select the signal mode for the • Passive Active
approved for use in the current input. • Active
hazardous area with type of
protection Ex-i.
Failure value In the Failure mode Enter value to be used by the Signed floating-point 0
parameter, the Defined value device if input value from number
option is selected. external device is missing.
Navigation
"Setup" menu → Status input
‣ Status input 1 to n
102 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Terminal number Shows the terminal numbers used by the • Not used –
status input module. • 24-25 (I/O 2)
• 22-23 (I/O 3)
Assign status input Select function for the status input. • Off Off
• Reset totalizer 1
• Reset totalizer 2
• Reset totalizer 3
• Reset all totalizers
• Flow override
Active level Define input signal level at which the • High High
assigned function is triggered. • Low
Response time status input Define the minimum amount of time the 5 to 200 ms 50 ms
input signal level must be present before the
selected function is triggered.
Navigation
"Setup" menu → Current output
‣ Current output 1 to n
20 mA value → 106
Endress+Hauser 103
Commissioning Proline Promass F 300 EtherNet/IP
Signal mode – Select the signal mode for the • Passive Active
current output. • Active
104 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Assign current output 1 to n – Select process variable for • Off Mass flow
current output. • Mass flow
• Volume flow
• Corrected volume
flow
• Target mass flow *
• Carrier mass flow *
• Target volume
flow
• Carrier volume
flow
• Target corrected
volume flow
• Carrier corrected
volume flow
• Density
• Reference density
• Reference density
alternative
• GSV flow
• GSV flow
alternative
• NSV flow
• NSV flow
alternative
• S&W volume flow
• Water cut
• Oil density
• Water density
• Oil mass flow
• Water mass flow
• Oil volume flow
• Water volume flow
• Oil corrected
volume flow
• Water corrected
volume flow
• Concentration *
• Temperature
• Carrier pipe
temperature *
• Electronic
temperature
• Oscillation
frequency 0
• Oscillation
amplitude 0 *
• Frequency
fluctuation 0
• Oscillation
damping 0
• Oscillation
damping
fluctuation 0
• Signal asymmetry
• Exciter current 0
• Pressure
Endress+Hauser 105
Commissioning Proline Promass F 300 EtherNet/IP
0/4 mA value One of the following options is Enter 4 mA value. Signed floating-point Country-specific:
selected in the Current span number • 0 kg/h
parameter (→ 105): • 0 lb/min
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
• 0...20 mA
20 mA value One of the following options is Enter 20 mA value. Signed floating-point Depends on country
selected in the Current span number and nominal
parameter (→ 105): diameter
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
• 0...20 mA
Fixed current The Fixed current option is Defines the fixed output 0 to 22.5 mA 22.5 mA
selected in the Current span current.
parameter (→ 105).
Damping output 1 to n One of the following options is Set reaction time for output 0.0 to 999.9 s 1.0 s
selected in the Assign current signal to fluctuations in the
output parameter (→ 105): measured value.
• Mass flow
• Volume flow
• Corrected volume flow
• Target mass flow *
• Carrier mass flow *
• Density
• Reference density
• Concentration *
• Temperature
• Carrier pipe temperature *
• Electronic temperature
• Oscillation frequency 0
• Oscillation amplitude 0 *
• Frequency fluctuation 0
• Oscillation damping
fluctuation 0
• Tube damping fluctuation 0
• Signal asymmetry
• Exciter current 0
Detailed description of
the options Oscillation
frequency, Oscillation
amplitude, Oscillation
damping and Signal
asymmetry: Value 1
display parameter
One of the following options is
selected in the Current span
parameter (→ 105):
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
• 0...20 mA
106 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Failure mode One of the following options is Define output behavior in • Min. Max.
selected in the Assign current alarm condition. • Max.
output parameter (→ 105): • Last valid value
• Mass flow • Actual value
• Volume flow • Defined value
• Corrected volume flow
• Target mass flow *
• Carrier mass flow *
• Density
• Reference density
• Concentration *
• Temperature
• Carrier pipe temperature *
• Electronic temperature
• Oscillation frequency 0
• Oscillation amplitude 0 *
• Frequency fluctuation 0
• Oscillation damping 0
• Oscillation damping
fluctuation 0
• Signal asymmetry
• Exciter current 0
One of the following options is
selected in the Current span
parameter (→ 105):
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
• 0...20 mA
Failure current The Defined value option is Enter current output value in 0 to 22.5 mA 22.5 mA
selected in the Failure mode alarm condition.
parameter.
Endress+Hauser 107
Commissioning Proline Promass F 300 EtherNet/IP
Navigation
"Setup" menu → Advanced setup → Pulse/frequency/switch output
‣ Pulse/frequency/switch output
1 to n
Navigation
"Setup" menu → Pulse/frequency/switch output
‣ Pulse/frequency/switch output
1 to n
108 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Signal mode – Select the signal mode for the • Passive Passive
PFS output. • Active
Assign pulse output 1 to n The Pulse option is selected in Select process variable for • Off Off
the Operating mode pulse output. • Mass flow
parameter. • Volume flow
• Corrected volume
flow
• Target mass flow *
• Carrier mass flow *
• Target volume
flow
• Carrier volume
flow
• Target corrected
volume flow
• Carrier corrected
volume flow
• GSV flow
• GSV flow
alternative
• NSV flow
• NSV flow
alternative
• S&W volume flow
• Oil mass flow
• Water mass flow
• Oil volume flow
• Water volume flow
• Oil corrected
volume flow
• Water corrected
volume flow
Value per pulse In the Operating mode Enter measured value at which Signed floating-point Depends on country
parameter, the Pulse option is a pulse is output. number and nominal
selected, and one of the diameter
following options is selected in
the Assign pulse output
parameter (→ 109):
• Mass flow
• Volume flow
• Corrected volume flow
• Target mass flow *
• Carrier mass flow *
Pulse width In the Operating mode Define time width of the 0.05 to 2 000 ms 100 ms
parameter, the Pulse option is output pulse.
selected, and one of the
following options is selected in
the Assign pulse output
parameter (→ 109):
• Mass flow
• Volume flow
• Corrected volume flow
• Target mass flow *
• Carrier mass flow *
Endress+Hauser 109
Commissioning Proline Promass F 300 EtherNet/IP
Failure mode In the Operating mode Define output behavior in • Actual value No pulses
parameter, the Pulse option is alarm condition. • No pulses
selected, and one of the
following options is selected in
the Assign pulse output
parameter (→ 109):
• Mass flow
• Volume flow
• Corrected volume flow
• Target mass flow *
• Carrier mass flow *
Navigation
"Setup" menu → Pulse/frequency/switch output
‣ Pulse/frequency/switch output
1 to n
110 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Signal mode – Select the signal mode for the • Passive Passive
PFS output. • Active
Endress+Hauser 111
Commissioning Proline Promass F 300 EtherNet/IP
Assign frequency output The Frequency option is Select process variable for • Off Off
selected in the Operating frequency output. • Mass flow
mode parameter (→ 108). • Volume flow
• Corrected volume
flow
• Target mass flow *
• Carrier mass flow *
• Target volume
flow
• Carrier volume
flow
• Target corrected
volume flow
• Carrier corrected
volume flow
• Density
• Reference density
• Reference density
alternative
• GSV flow
• GSV flow
alternative
• NSV flow
• NSV flow
alternative
• S&W volume flow
• Water cut
• Oil density
• Water density
• Oil mass flow
• Water mass flow
• Oil volume flow
• Water volume flow
• Oil corrected
volume flow
• Water corrected
volume flow
• Concentration *
• Temperature
• Carrier pipe
temperature *
• Electronic
temperature
• Oscillation
frequency 0
• Oscillation
amplitude 0 *
• Frequency
fluctuation 0
• Oscillation
damping 0
• Oscillation
damping
fluctuation 0
• Signal asymmetry
• Exciter current 0
• HBSI
• Pressure
112 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Minimum frequency value In the Operating mode Enter minimum frequency. 0.0 to 10 000.0 Hz 0.0 Hz
parameter, the Frequency
option is selected, and one of
the following options is
selected in the Assign
frequency output parameter
(→ 112):
• Mass flow
• Volume flow
• Corrected volume flow
• Target mass flow *
• Carrier mass flow *
• Density
• Reference density
• Concentration *
• Temperature
• Carrier pipe temperature *
• Electronic temperature
• Oscillation frequency 0
• Frequency fluctuation 0
• Oscillation amplitude 0 *
• Oscillation damping 0
• Oscillation damping
fluctuation 0
• Signal asymmetry
• Exciter current 0
Maximum frequency value In the Operating mode Enter maximum frequency. 0.0 to 10 000.0 Hz 10 000.0 Hz
parameter, the Frequency
option is selected, and one of
the following options is
selected in the Assign
frequency output parameter
(→ 112):
• Mass flow
• Volume flow
• Corrected volume flow
• Target mass flow *
• Carrier mass flow *
• Density
• Reference density
• Concentration *
• Temperature
• Carrier pipe temperature *
• Electronic temperature
• Oscillation frequency 0
• Frequency fluctuation 0
• Oscillation amplitude 0 *
• Oscillation damping 0
• Oscillation damping
fluctuation 0
• Signal asymmetry
• Exciter current 0
Endress+Hauser 113
Commissioning Proline Promass F 300 EtherNet/IP
Measuring value at minimum In the Operating mode Enter measured value for Signed floating-point Depends on country
frequency parameter, the Frequency minmum frequency. number and nominal
option is selected, and one of diameter
the following options is
selected in the Assign
frequency output parameter
(→ 112):
• Mass flow
• Volume flow
• Corrected volume flow
• Target mass flow *
• Carrier mass flow *
• Density
• Reference density
• Concentration *
• Temperature
• Carrier pipe temperature *
• Electronic temperature
• Oscillation frequency 0
• Frequency fluctuation 0
• Oscillation amplitude 0 *
• Oscillation damping 0
• Oscillation damping
fluctuation 0
• Signal asymmetry
• Exciter current 0
Measuring value at maximum In the Operating mode Enter measured value for Signed floating-point Depends on country
frequency parameter, the Frequency maximum frequency. number and nominal
option is selected, and one of diameter
the following options is
selected in the Assign
frequency output parameter
(→ 112):
• Mass flow
• Volume flow
• Corrected volume flow
• Target mass flow *
• Carrier mass flow *
• Density
• Reference density
• Concentration *
• Temperature
• Carrier pipe temperature *
• Electronic temperature
• Oscillation frequency 0
• Frequency fluctuation 0
• Oscillation amplitude 0 *
• Oscillation damping 0
• Oscillation damping
fluctuation 0
• Signal asymmetry
• Exciter current 0
114 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Failure mode In the Operating mode Define output behavior in • Actual value 0 Hz
parameter, the Frequency alarm condition. • Defined value
option is selected, and one of • 0 Hz
the following options is
selected in the Assign
frequency output parameter
(→ 112):
• Mass flow
• Volume flow
• Corrected volume flow
• Target mass flow *
• Carrier mass flow *
• Density
• Reference density
• Concentration *
• Temperature
• Carrier pipe temperature *
• Electronic temperature
• Oscillation frequency 0
• Frequency fluctuation 0
• Oscillation amplitude 0 *
• Oscillation damping 0
• Oscillation damping
fluctuation 0
• Signal asymmetry
• Exciter current 0
Failure frequency In the Operating mode Enter frequency output value 0.0 to 12 500.0 Hz 0.0 Hz
parameter, the Frequency in alarm condition.
option is selected, and one of
the following options is
selected in the Assign
frequency output parameter
(→ 112):
• Mass flow
• Volume flow
• Corrected volume flow
• Target mass flow *
• Carrier mass flow *
• Density
• Reference density
• Concentration *
• Temperature
• Carrier pipe temperature *
• Electronic temperature
• Oscillation frequency 0
• Frequency fluctuation 0
• Oscillation amplitude 0 *
• Oscillation damping 0
• Oscillation damping
fluctuation 0
• Signal asymmetry
• Exciter current 0
Endress+Hauser 115
Commissioning Proline Promass F 300 EtherNet/IP
Navigation
"Setup" menu → Pulse/frequency/switch output
‣ Pulse/frequency/switch output
1 to n
Signal mode – Select the signal mode for the • Passive Passive
PFS output. • Active
116 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Switch output function The Switch option is selected Select function for switch • Off Off
in the Operating mode output. • On
parameter. • Diagnostic
behavior
• Limit
• Flow direction
check
• Status
Assign diagnostic behavior • In the Operating mode Select diagnostic behavior for • Alarm Alarm
parameter, the Switch switch output. • Alarm or warning
option is selected. • Warning
• In the Switch output
function parameter, the
Diagnostic behavior option
is selected.
Assign limit • The Switch option is Select process variable for limit • Mass flow Mass flow
selected in the Operating function. • Volume flow
mode parameter. • Corrected volume
• The Limit option is selected flow
in the Switch output • Target mass flow *
function parameter. • Carrier mass flow *
• Target volume
flow
• Carrier volume
flow
• Target corrected
volume flow
• Carrier corrected
volume flow
• Density
• Reference density
• Reference density
alternative
• GSV flow
• GSV flow
alternative
• NSV flow
• NSV flow
alternative
• S&W volume flow
• Water cut
• Oil density
• Water density
• Oil mass flow
• Water mass flow
• Oil volume flow
• Water volume flow
• Oil corrected
volume flow
• Water corrected
volume flow
• Concentration *
• Temperature
• Totalizer 1
• Totalizer 2
• Totalizer 3
• Oscillation
damping
• Pressure
Endress+Hauser 117
Commissioning Proline Promass F 300 EtherNet/IP
Assign flow direction check • The Switch option is Select process variable for flow • Off Mass flow
selected in the Operating direction monitoring. • Volume flow
mode parameter. • Mass flow
• The Flow direction check • Corrected volume
option is selected in the flow
Switch output function
parameter.
Assign status • The Switch option is Select device status for switch • Partially filled pipe Partially filled pipe
selected in the Operating output. detection detection
mode parameter. • Low flow cut off
• The Status option is
selected in the Switch
output function parameter.
Switch-on value • In the Operating mode Enter measured value for the Signed floating-point Country-specific:
parameter, the Switch switch-on point. number • 0 kg/h
option is selected. • 0 lb/min
• In the Switch output
function parameter, the
Limit option is selected.
Switch-off value • In the Operating mode Enter measured value for the Signed floating-point Country-specific:
parameter, the Switch switch-off point. number • 0 kg/h
option is selected. • 0 lb/min
• In the Switch output
function parameter, the
Limit option is selected.
Switch-on delay • The Switch option is Define delay for the switch-on 0.0 to 100.0 s 0.0 s
selected in the Operating of status output.
mode parameter.
• The Limit option is selected
in the Switch output
function parameter.
Switch-off delay • The Switch option is Define delay for the switch-off 0.0 to 100.0 s 0.0 s
selected in the Operating of status output.
mode parameter.
• The Limit option is selected
in the Switch output
function parameter.
Navigation
"Setup" menu → Relay output 1 to n
‣ RelaisOutput 1 to n
118 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Relay output function – Select the function for the • Closed Closed
relay output. • Open
• Diagnostic
behavior
• Limit
• Flow direction
check
• Digital Output
Assign flow direction check In the Relay output function Select process variable for flow • Off Mass flow
parameter, the Flow direction direction monitoring. • Volume flow
check option is selected. • Mass flow
• Corrected volume
flow
Endress+Hauser 119
Commissioning Proline Promass F 300 EtherNet/IP
Assign limit In the Relay output function Select process variable for limit • Mass flow Mass flow
parameter, the Limit option is function. • Volume flow
selected. • Corrected volume
flow
• Target mass flow *
• Carrier mass flow *
• Target volume
flow
• Carrier volume
flow
• Target corrected
volume flow
• Carrier corrected
volume flow
• Density
• Reference density
• Reference density
alternative
• GSV flow
• GSV flow
alternative
• NSV flow
• NSV flow
alternative
• S&W volume flow
• Water cut
• Oil density
• Water density
• Oil mass flow
• Water mass flow
• Oil volume flow
• Water volume flow
• Oil corrected
volume flow
• Water corrected
volume flow
• Concentration *
• Temperature
• Totalizer 1
• Totalizer 2
• Totalizer 3
• Oscillation
damping
• Pressure
Assign diagnostic behavior In the Relay output function Select diagnostic behavior for • Alarm Alarm
parameter, the Diagnostic switch output. • Alarm or warning
behavior option is selected. • Warning
Assign status In the Relay output function Select device status for switch • Partially filled pipe Partially filled pipe
parameter, the Digital Output output. detection detection
option is selected. • Low flow cut off
Switch-off value In the Relay output function Enter measured value for the Signed floating-point Country-specific:
parameter, the Limit option is switch-off point. number • 0 kg/h
selected. • 0 lb/min
Switch-off delay In the Relay output function Define delay for the switch-off 0.0 to 100.0 s 0.0 s
parameter, the Limit option is of status output.
selected.
Switch-on value In the Relay output function Enter measured value for the Signed floating-point Country-specific:
parameter, the Limit option is switch-on point. number • 0 kg/h
selected. • 0 lb/min
120 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Switch-on delay In the Relay output function Define delay for the switch-on 0.0 to 100.0 s 0.0 s
parameter, the Limit option is of status output.
selected.
Navigation
"Setup" menu → Display
‣ Display
Endress+Hauser 121
Commissioning Proline Promass F 300 EtherNet/IP
Format display A local display is provided. Select how measured values • 1 value, max. size 1 value, max. size
are shown on the display. • 1 bargraph + 1
value
• 2 values
• 1 value large + 2
values
• 4 values
122 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Value 1 display A local display is provided. Select the measured value that • Mass flow Mass flow
is shown on the local display. • Volume flow
• Corrected volume
flow
• Target mass flow *
• Carrier mass flow *
• Target volume
flow
• Carrier volume
flow
• Target corrected
volume flow
• Carrier corrected
volume flow
• Density
• Reference density
• Reference density
alternative
• GSV flow
• GSV flow
alternative
• NSV flow
• NSV flow
alternative
• S&W volume flow
• Water cut
• Oil density
• Water density
• Oil mass flow
• Water mass flow
• Oil volume flow
• Water volume flow
• Oil corrected
volume flow
• Water corrected
volume flow
• Weighted density
average
• Weighted
temperature
average
• Concentration *
• Temperature
• Carrier pipe
temperature *
• Electronic
temperature
• Oscillation
frequency 0
• Oscillation
amplitude 0 *
• Frequency
fluctuation 0
• Oscillation
damping 0
• Oscillation
damping
fluctuation 0
• Signal asymmetry
• Exciter current 0
• Totalizer 1
• Totalizer 2
• Totalizer 3
• Current output 1
• Current output 2 *
• Current output 3 *
• Pressure
Endress+Hauser 123
Commissioning Proline Promass F 300 EtherNet/IP
0% bargraph value 1 A local display is provided. Enter 0% value for bar graph Signed floating-point Country-specific:
display. number • 0 kg/h
• 0 lb/min
100% bargraph value 1 A local display is provided. Enter 100% value for bar Signed floating-point Depends on country
graph display. number and nominal
diameter
124 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Value 2 display A local display is provided. Select the measured value that • None None
is shown on the local display. • Mass flow
• Volume flow
• Corrected volume
flow
• Target mass flow
• Carrier mass flow
• Target volume
flow
• Carrier volume
flow
• Target corrected
volume flow
• Carrier corrected
volume flow
• Density
• Reference density
• Reference density
alternative
• GSV flow
• GSV flow
alternative
• NSV flow
• NSV flow
alternative
• S&W volume flow
• Water cut
• Oil density
• Water density
• Oil mass flow
• Water mass flow
• Oil volume flow
• Water corrected
volume flow
• Oil corrected
volume flow
• Water volume flow
• Weighted density
average
• Weighted
temperature
average
• Concentration
• Dynamic viscosity
• Kinematic viscosity
• Temp.
compensated
dynamic viscosity
• Temp.
compensated
kinematic viscosity
• Temperature
• Carrier pipe
temperature
• Electronic
temperature
• Oscillation
frequency 0
• Oscillation
frequency 1
• Oscillation
amplitude 0
• Oscillation
amplitude 1
• Frequency
fluctuation 0
• Frequency
fluctuation 1
Endress+Hauser 125
Commissioning Proline Promass F 300 EtherNet/IP
126 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Value 3 display A local display is provided. Select the measured value that • None None
is shown on the local display. • Mass flow
• Volume flow
• Corrected volume
flow
• Target mass flow
• Carrier mass flow
• Target volume
flow
• Carrier volume
flow
• Target corrected
volume flow
• Carrier corrected
volume flow
• Density
• Reference density
• Reference density
alternative
• GSV flow
• GSV flow
alternative
• NSV flow
• NSV flow
alternative
• S&W volume flow
• Water cut
• Oil density
• Water density
• Oil mass flow
• Water mass flow
• Oil volume flow
• Water corrected
volume flow
• Oil corrected
volume flow
• Water volume flow
• Weighted density
average
• Weighted
temperature
average
• Concentration
• Dynamic viscosity
• Kinematic viscosity
• Temp.
compensated
dynamic viscosity
• Temp.
compensated
kinematic viscosity
• Temperature
• Carrier pipe
temperature
• Electronic
temperature
• Oscillation
frequency 0
• Oscillation
frequency 1
• Oscillation
amplitude 0
• Oscillation
amplitude 1
• Frequency
fluctuation 0
• Frequency
fluctuation 1
Endress+Hauser 127
Commissioning Proline Promass F 300 EtherNet/IP
0% bargraph value 3 A selection was made in the Enter 0% value for bar graph Signed floating-point Country-specific:
Value 3 display parameter. display. number • 0 kg/h
• 0 lb/min
128 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
100% bargraph value 3 A selection was made in the Enter 100% value for bar Signed floating-point 0
Value 3 display parameter. graph display. number
Endress+Hauser 129
Commissioning Proline Promass F 300 EtherNet/IP
Value 4 display A local display is provided. Select the measured value that • None None
is shown on the local display. • Mass flow
• Volume flow
• Corrected volume
flow
• Target mass flow
• Carrier mass flow
• Target volume
flow
• Carrier volume
flow
• Target corrected
volume flow
• Carrier corrected
volume flow
• Density
• Reference density
• Reference density
alternative
• GSV flow
• GSV flow
alternative
• NSV flow
• NSV flow
alternative
• S&W volume flow
• Water cut
• Oil density
• Water density
• Oil mass flow
• Water mass flow
• Oil volume flow
• Water corrected
volume flow
• Oil corrected
volume flow
• Water volume flow
• Weighted density
average
• Weighted
temperature
average
• Concentration
• Dynamic viscosity
• Kinematic viscosity
• Temp.
compensated
dynamic viscosity
• Temp.
compensated
kinematic viscosity
• Temperature
• Carrier pipe
temperature
• Electronic
temperature
• Oscillation
frequency 0
• Oscillation
frequency 1
• Oscillation
amplitude 0
• Oscillation
amplitude 1
• Frequency
fluctuation 0
• Frequency
fluctuation 1
130 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Endress+Hauser 131
Commissioning Proline Promass F 300 EtherNet/IP
Navigation
"Setup" menu → Low flow cut off
Assign process variable – Select process variable for low • Off Mass flow
flow cut off. • Mass flow
• Volume flow
• Corrected volume
flow
On value low flow cutoff One of the following options is Enter on value for low flow cut Positive floating- Depends on country
selected in the Assign process off. point number and nominal
variable parameter diameter
(→ 132):
• Mass flow
• Volume flow
• Corrected volume flow
Off value low flow cutoff One of the following options is Enter off value for low flow cut 0 to 100.0 % 50 %
selected in the Assign process off.
variable parameter
(→ 132):
• Mass flow
• Volume flow
• Corrected volume flow
Pressure shock suppression One of the following options is Enter time frame for signal 0 to 100 s 0s
selected in the Assign process suppression (= active pressure
variable parameter shock suppression).
(→ 132):
• Mass flow
• Volume flow
• Corrected volume flow
132 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Navigation
"Setup" menu → Partially filled pipe detection
Low value partial filled pipe One of the following options is Enter lower limit value for Signed floating-point 200
detection selected in the Assign process deactivating partialy filled pipe number
variable parameter detection.
(→ 133):
• Density
• Reference density
High value partial filled pipe One of the following options is Enter upper limit value for Signed floating-point 6 000
detection selected in the Assign process deactivating partialy filled pipe number
variable parameter detection.
(→ 133):
• Density
• Reference density
Response time part. filled pipe One of the following options is Enter time before diagnostic 0 to 100 s 1s
detect. selected in the Assign process message is displayed for
variable parameter partially filled pipe detection.
(→ 133):
• Density
• Reference density
Endress+Hauser 133
Commissioning Proline Promass F 300 EtherNet/IP
XXXXXXXXX
20.50
mA
Main menu 0104-1
1. Display language
English
Display/operat.
Setup
Main menu
2. Display/operat.
Setup
Diagnostic
/ ../Setup
3. Medium selection
XXXXXXXXX
XXXXXXXXX
/ ../Setup
4. XXXXXXXXX
XXXXXXXXX
Advanced setup
Device tag
****
Def. access code
A0032223-EN
The number of submenus can vary depending on the device version. Some submenus
are not dealt with in the Operating Instructions. These submenus and the parameters
they contain are explained in the Special Documentation for the device.
Navigation
"Setup" menu → Advanced setup
‣ Advanced setup
‣ Totalizer 1 to n → 137
‣ Display → 139
134 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
‣ Concentration
‣ Heartbeat setup
‣ Administration → 152
Enter access code Enter access code to disable write protection of parameters. 0 to 9 999
Navigation
"Setup" menu → Advanced setup → Calculated values
‣ Calculated values
Endress+Hauser 135
Commissioning Proline Promass F 300 EtherNet/IP
Corrected volume flow calculation – Select reference density for • Fixed reference Calculated reference
calculating the corrected density density
volume flow. • Calculated
reference density
• External reference
density
• Current input 1 *
• Current input 2 *
External reference density In the Corrected volume flow Shows external reference Floating point –
calculation parameter, the density. number with sign
External reference density
option is selected.
Fixed reference density The Fixed reference density Enter fixed value for reference Positive floating- 1 kg/Nl
option is selected in the density. point number
Corrected volume flow
calculation parameter
parameter.
Reference temperature The Calculated reference Enter reference temperature –273.15 to 99 999 °C Country-specific:
density option is selected in for calculating the reference • +20 °C
the Corrected volume flow density. • +68 °F
calculation parameter.
Linear expansion coefficient The Calculated reference Enter linear, medium-specific Signed floating-point 0.0 1/K
density option is selected in expansion coefficient for number
the Corrected volume flow calculating the reference
calculation parameter density.
parameter.
Square expansion coefficient The Calculated reference For media with a non-linear Signed floating-point 0.0 1/K²
density option is selected in expansion pattern: enter the number
the Corrected volume flow quadratic, medium-specific
calculation parameter expansion coefficient for
parameter. calculating the reference
density.
Navigation
"Setup" menu → Advanced setup → Sensor adjustment
‣ Sensor adjustment
136 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Installation direction Set sign of flow direction to match the • Flow in arrow direction Flow in arrow direction
direction of the arrow on the sensor. • Flow against arrow direction
Navigation
"Setup" menu → Advanced setup → Sensor adjustment → Zero point adjustment
Progress → 137
Zero point adjustment control Start zero point adjustment. • Cancel Cancel
• Busy
• Zero point adjust failure
• Start
Navigation
"Setup" menu → Advanced setup → Totalizer 1 to n
‣ Totalizer 1 to n
Endress+Hauser 137
Commissioning Proline Promass F 300 EtherNet/IP
Assign process variable – Select process variable for • Off Mass flow
totalizer. • Mass flow
• Volume flow
• Corrected volume
flow
• Target mass flow *
• Carrier mass flow *
• Target volume
flow
• Carrier volume
flow
• Target corrected
volume flow
• Carrier corrected
volume flow
• GSV flow
• GSV flow
alternative
• NSV flow
• NSV flow
alternative
• S&W volume flow
• Oil mass flow
• Water mass flow
• Oil volume flow
• Water volume flow
• Oil corrected
volume flow
• Water corrected
volume flow
Unit totalizer 1 to n One of the following options is Select process variable totalizer Unit choose list Country-specific:
selected in the Assign process unit. • kg
variable parameter • lb
(→ 138) of the Totalizer
1 to n submenu:
• Volume flow
• Mass flow
• Corrected volume flow
• Target mass flow *
• Carrier mass flow *
Totalizer operation mode One of the following options is Select totalizer calculation • Net flow total Net flow total
selected in the Assign process mode. • Forward flow total
variable parameter • Reverse flow total
(→ 138) of the Totalizer
1 to n submenu:
• Volume flow
• Mass flow
• Corrected volume flow
• Target mass flow *
• Carrier mass flow *
Failure mode One of the following options is Define totalizer behavior in • Stop Stop
selected in the Assign process alarm condition. • Actual value
variable parameter • Last valid value
(→ 138) of the Totalizer
1 to n submenu:
• Volume flow
• Mass flow
• Corrected volume flow
• Target mass flow *
• Carrier mass flow *
138 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Navigation
"Setup" menu → Advanced setup → Display
‣ Display
Header → 148
Separator → 149
Backlight → 149
Endress+Hauser 139
Commissioning Proline Promass F 300 EtherNet/IP
Format display A local display is provided. Select how measured values • 1 value, max. size 1 value, max. size
are shown on the display. • 1 bargraph + 1
value
• 2 values
• 1 value large + 2
values
• 4 values
140 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Value 1 display A local display is provided. Select the measured value that • Mass flow Mass flow
is shown on the local display. • Volume flow
• Corrected volume
flow
• Target mass flow *
• Carrier mass flow *
• Target volume
flow
• Carrier volume
flow
• Target corrected
volume flow
• Carrier corrected
volume flow
• Density
• Reference density
• Reference density
alternative
• GSV flow
• GSV flow
alternative
• NSV flow
• NSV flow
alternative
• S&W volume flow
• Water cut
• Oil density
• Water density
• Oil mass flow
• Water mass flow
• Oil volume flow
• Water volume flow
• Oil corrected
volume flow
• Water corrected
volume flow
• Weighted density
average
• Weighted
temperature
average
• Concentration *
• Temperature
• Carrier pipe
temperature *
• Electronic
temperature
• Oscillation
frequency 0
• Oscillation
amplitude 0 *
• Frequency
fluctuation 0
• Oscillation
damping 0
• Oscillation
damping
fluctuation 0
• Signal asymmetry
• Exciter current 0
• Totalizer 1
• Totalizer 2
• Totalizer 3
• Current output 1
• Current output 2 *
• Current output 3 *
• Pressure
Endress+Hauser 141
Commissioning Proline Promass F 300 EtherNet/IP
0% bargraph value 1 A local display is provided. Enter 0% value for bar graph Signed floating-point Country-specific:
display. number • 0 kg/h
• 0 lb/min
100% bargraph value 1 A local display is provided. Enter 100% value for bar Signed floating-point Depends on country
graph display. number and nominal
diameter
Decimal places 1 A measured value is specified Select the number of decimal • x x.xx
in the Value 1 display places for the display value. • x.x
parameter. • x.xx
• x.xxx
• x.xxxx
142 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Value 2 display A local display is provided. Select the measured value that • None None
is shown on the local display. • Mass flow
• Volume flow
• Corrected volume
flow
• Target mass flow
• Carrier mass flow
• Target volume
flow
• Carrier volume
flow
• Target corrected
volume flow
• Carrier corrected
volume flow
• Density
• Reference density
• Reference density
alternative
• GSV flow
• GSV flow
alternative
• NSV flow
• NSV flow
alternative
• S&W volume flow
• Water cut
• Oil density
• Water density
• Oil mass flow
• Water mass flow
• Oil volume flow
• Water corrected
volume flow
• Oil corrected
volume flow
• Water volume flow
• Weighted density
average
• Weighted
temperature
average
• Concentration
• Dynamic viscosity
• Kinematic viscosity
• Temp.
compensated
dynamic viscosity
• Temp.
compensated
kinematic viscosity
• Temperature
• Carrier pipe
temperature
• Electronic
temperature
• Oscillation
frequency 0
• Oscillation
frequency 1
• Oscillation
amplitude 0
• Oscillation
amplitude 1
• Frequency
fluctuation 0
• Frequency
fluctuation 1
Endress+Hauser 143
Commissioning Proline Promass F 300 EtherNet/IP
Decimal places 2 A measured value is specified Select the number of decimal • x x.xx
in the Value 2 display places for the display value. • x.x
parameter. • x.xx
• x.xxx
• x.xxxx
144 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Value 3 display A local display is provided. Select the measured value that • None None
is shown on the local display. • Mass flow
• Volume flow
• Corrected volume
flow
• Target mass flow
• Carrier mass flow
• Target volume
flow
• Carrier volume
flow
• Target corrected
volume flow
• Carrier corrected
volume flow
• Density
• Reference density
• Reference density
alternative
• GSV flow
• GSV flow
alternative
• NSV flow
• NSV flow
alternative
• S&W volume flow
• Water cut
• Oil density
• Water density
• Oil mass flow
• Water mass flow
• Oil volume flow
• Water corrected
volume flow
• Oil corrected
volume flow
• Water volume flow
• Weighted density
average
• Weighted
temperature
average
• Concentration
• Dynamic viscosity
• Kinematic viscosity
• Temp.
compensated
dynamic viscosity
• Temp.
compensated
kinematic viscosity
• Temperature
• Carrier pipe
temperature
• Electronic
temperature
• Oscillation
frequency 0
• Oscillation
frequency 1
• Oscillation
amplitude 0
• Oscillation
amplitude 1
• Frequency
fluctuation 0
• Frequency
fluctuation 1
Endress+Hauser 145
Commissioning Proline Promass F 300 EtherNet/IP
0% bargraph value 3 A selection was made in the Enter 0% value for bar graph Signed floating-point Country-specific:
Value 3 display parameter. display. number • 0 kg/h
• 0 lb/min
100% bargraph value 3 A selection was made in the Enter 100% value for bar Signed floating-point 0
Value 3 display parameter. graph display. number
Decimal places 3 A measured value is specified Select the number of decimal • x x.xx
in the Value 3 display places for the display value. • x.x
parameter. • x.xx
• x.xxx
• x.xxxx
146 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Value 4 display A local display is provided. Select the measured value that • None None
is shown on the local display. • Mass flow
• Volume flow
• Corrected volume
flow
• Target mass flow
• Carrier mass flow
• Target volume
flow
• Carrier volume
flow
• Target corrected
volume flow
• Carrier corrected
volume flow
• Density
• Reference density
• Reference density
alternative
• GSV flow
• GSV flow
alternative
• NSV flow
• NSV flow
alternative
• S&W volume flow
• Water cut
• Oil density
• Water density
• Oil mass flow
• Water mass flow
• Oil volume flow
• Water corrected
volume flow
• Oil corrected
volume flow
• Water volume flow
• Weighted density
average
• Weighted
temperature
average
• Concentration
• Dynamic viscosity
• Kinematic viscosity
• Temp.
compensated
dynamic viscosity
• Temp.
compensated
kinematic viscosity
• Temperature
• Carrier pipe
temperature
• Electronic
temperature
• Oscillation
frequency 0
• Oscillation
frequency 1
• Oscillation
amplitude 0
• Oscillation
amplitude 1
• Frequency
fluctuation 0
• Frequency
fluctuation 1
Endress+Hauser 147
Commissioning Proline Promass F 300 EtherNet/IP
Decimal places 4 A measured value is specified Select the number of decimal • x x.xx
in the Value 4 display places for the display value. • x.x
parameter. • x.xx
• x.xxx
• x.xxxx
Display language A local display is provided. Set display language. • English English
• Deutsch * (alternatively, the
• Français * ordered language is
• Español * preset in the device)
• Italiano *
• Nederlands *
• Portuguesa *
• Polski *
• русский язык
(Russian) *
• Svenska *
• Türkçe *
• 中文 (Chinese) *
• 日本語
(Japanese) *
• 한국어 (Korean) *
• Bahasa Indonesia *
• tiếng Việt
(Vietnamese) *
• čeština (Czech) *
Display interval A local display is provided. Set time measured values are 1 to 10 s 5s
shown on display if display
alternates between values.
Display damping A local display is provided. Set display reaction time to 0.0 to 999.9 s 0.0 s
fluctuations in the measured
value.
Header A local display is provided. Select header contents on local • Device tag Device tag
display. • Free text
Header text In the Header parameter, the Enter display header text. Max. 12 characters ------------
Free text option is selected. such as letters,
numbers or special
characters (e.g. @,
%, /)
148 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Separator A local display is provided. Select decimal separator for • . (point) . (point)
displaying numerical values. • , (comma)
Backlight One of the following Switch the local display • Disable Enable
conditions is met: backlight on and off. • Enable
• Order code for "Display;
operation", option F "4-line,
illum.; touch control"
• Order code for "Display;
operation", option G "4-line,
illum.; touch control
+WLAN"
• Order code for "Display;
operation", option O
"Separate 4-line display,
illum.; 10m/30ft cable;
touch control"
Navigation
"Setup" menu → Advanced setup → WLAN settings
‣ WLAN settings
WLAN → 150
Endress+Hauser 149
Commissioning Proline Promass F 300 EtherNet/IP
WLAN mode – Select WLAN mode. • WLAN access WLAN access point
point
• WLAN Client
WLAN IP address – Enter IP address of the device 4 octet: 0 to 255 (in 192.168.1.212
WLAN interface. the particular octet)
WLAN passphrase The WPA2-PSK option is Enter the network key (8 to 32 8 to 32-digit Serial number of the
selected in the Security type characters). character string measuring device
parameter. comprising numbers, (e.g. L100A802000)
The network key
supplied with the device letters and special
characters
should be changed
during commissioning
for security reasons.
Assign SSID name – Select which name will be used • Device tag User-defined
for SSID: device tag or user- • User-defined
defined name.
SSID name • The User-defined option is Enter the user-defined SSID Max. 32-digit EH_device
selected in the Assign SSID name (max. 32 characters). character string designation_last 7
name parameter. comprising numbers, digits of the serial
• The WLAN access point The user-defined SSID
name may only be letters and special number (e.g.
option is selected in the assigned once. If the SSID characters EH_Promass_300_A
WLAN mode parameter. 802000)
name is assigned more
than once, the devices
can interfere with one
another.
150 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Navigation
"Setup" menu → Advanced setup → Configuration backup
‣ Configuration backup
Operating time Indicates how long the device has been in Days (d), hours (h), minutes –
operation. (m) and seconds (s)
Last backup Shows when the last data backup was saved Days (d), hours (h), minutes –
to HistoROM backup. (m) and seconds (s)
Configuration management Select action for managing the device data in • Cancel Cancel
the HistoROM backup. • Execute backup
• Restore
• Compare
• Clear backup data
Backup state Shows the current status of data saving or • None None
restoring. • Backup in progress
• Restoring in progress
• Delete in progress
• Compare in progress
• Restoring failed
• Backup failed
Comparison result Comparison of current device data with • Settings identical Check not done
HistoROM backup. • Settings not identical
• No backup available
• Backup settings corrupt
• Check not done
• Dataset incompatible
Endress+Hauser 151
Commissioning Proline Promass F 300 EtherNet/IP
Options Description
Execute backup A backup copy of the current device configuration is saved from the HistoROM
backup to the memory of the device. The backup copy includes the transmitter data
of the device.
Restore The last backup copy of the device configuration is restored from the device
memory to the device's HistoROM backup. The backup copy includes the
transmitter data of the device.
Compare The device configuration saved in the device memory is compared with the current
device configuration of the HistoROM backup.
Clear backup data The backup copy of the device configuration is deleted from the memory of the
device.
HistoROM backup
A HistoROM is a "non-volatile" device memory in the form of an EEPROM.
While this action is in progress, the configuration cannot be edited via the local
display and a message on the processing status appears on the display.
Navigation
"Setup" menu → Advanced setup → Administration
‣ Administration
Navigation
"Setup" menu → Advanced setup → Administration → Define access code
152 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Define access code Restrict write-access to parameters to protect the configuration Max. 16-digit character string comprising
of the device against unintentional changes. numbers, letters and special characters
Confirm access code Confirm the entered access code. Max. 16-digit character string comprising
numbers, letters and special characters
Navigation
"Setup" menu → Advanced setup → Administration → Reset access code
Operating time Indicates how long the device has been in Days (d), hours (h), minutes –
operation. (m) and seconds (s)
Reset access code Reset access code to factory settings. Character string comprising 0x00
numbers, letters and special
For a reset code, contact your
Endress+Hauser service organization. characters
Navigation
"Setup" menu → Advanced setup → Administration
10.7 Simulation
The Simulation submenu enables you to simulate, without a real flow situation, various
process variables in the process and the device alarm mode and to verify downstream
signal chains (switching valves or closed-control loops).
Endress+Hauser 153
Commissioning Proline Promass F 300 EtherNet/IP
Navigation
"Diagnostics" menu → Simulation
‣ Simulation
154 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
Assign simulation process variable – Select a process variable for • Off Off
the simulation process that is • Mass flow
activated. • Volume flow
• Corrected volume
flow
• Target volume
flow
• Carrier volume
flow
• Target corrected
volume flow
• Carrier corrected
volume flow
• Density
• Reference density
• Reference density
alternative
• GSV flow
• GSV flow
alternative
• NSV flow
• NSV flow
alternative
• S&W volume flow
• Water cut
• Oil density
• Water density
• Oil mass flow
• Water mass flow
• Oil volume flow
• Water volume flow
• Oil corrected
volume flow
• Water corrected
volume flow
• Weighted density
average
• Weighted
temperature
average
• Temperature
• Concentration *
• Target mass flow *
• Carrier mass flow *
Process variable value One of the following options is Enter the simulation value for Depends on the 0
selected in the Assign the selected process variable. process variable
simulation process variable selected
parameter (→ 155):
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
• Concentration *
• Target mass flow *
• Carrier mass flow *
Input signal level In the Status input simulation Select the signal level for the • High High
parameter, the On option is simulation of the status input. • Low
selected.
Endress+Hauser 155
Commissioning Proline Promass F 300 EtherNet/IP
Value current input 1 to n In the Current input 1 to n Enter the current value for 0 to 22.5 mA 0 mA
simulation parameter, the On simulation.
option is selected.
Value current output 1 to n In the Current output 1 to n Enter the current value for 3.59 to 22.5 mA 3.59 mA
simulation parameter, the On simulation.
option is selected.
Frequency output simulation 1 to n In the Operating mode Switch the simulation of the • Off Off
parameter, the Frequency frequency output on and off. • On
option is selected.
Frequency value 1 to n In the Frequency output Enter the frequency value for 0.0 to 12 500.0 Hz 0.0 Hz
simulation 1 to n parameter, the simulation.
the On option is selected.
Pulse output simulation 1 to n In the Operating mode Set and switch off the pulse • Off Off
parameter, the Pulse option is output simulation. • Fixed value
selected. • Down-counting
Pulse
For Fixed value option:
width parameter value
(→ 109) defines the
pulse width of the pulses
output.
Pulse value 1 to n In the Pulse output Enter the number of pulses for 0 to 65 535 0
simulation 1 to n parameter, simulation.
the Down-counting value
option is selected.
Switch output simulation 1 to n In the Operating mode Switch the simulation of the • Off Off
parameter, the Switch option switch output on and off. • On
is selected.
Switch status 1 to n The On option is selected in Select status of the relay • Open Open
the Switch output simulation output for the simulation. • Closed
1 to n parameter parameter.
Pulse output simulation – Set and switch off the pulse • Off Off
output simulation. • Fixed value
• Down-counting
For Fixed value option:
Pulse width parameter value
defines the pulse width
of the pulses output.
Pulse value In the Pulse output Set and switch off the pulse 0 to 65 535 0
simulation parameter, the output simulation.
Down-counting value option
is selected.
Device alarm simulation – Switch the device alarm on and • Off Off
off. • On
156 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
The device automatically locks the write-protected parameters again if a key is not pressed
for 10 minutes in the navigation and editing view. The device locks the write-protected
parameters automatically after 60 s if the user skips back to the operational display mode
from the navigation and editing view.
• If parameter write protection is activated via an access code, it can also only be
deactivated via this access code → 61.
• The user role with which the user is currently logged on via the local display is
indicated by the → 61 Access status parameter. Navigation path: Operation
→ Access status
Endress+Hauser 157
Commissioning Proline Promass F 300 EtherNet/IP
↓ ↓
Via Web browser, FieldCare, DeviceCare (via CDI-RJ45 service interface), fieldbus
For a reset code, contact your Endress+Hauser service organization.
158 Endress+Hauser
Proline Promass F 300 EtherNet/IP Commissioning
1.
OFF ON
1
2
3
4
A0029630
Setting the write protection (WP) switch on the main electronics module to the ON
position enables hardware write protection.
In the Locking status parameter the Hardware locked option is displayed
→ 160. In addition, on the local display the -symbol appears in front of the
parameters in the header of the operational display and in the navigation view.
XXXXXXXXX
20.50
XX XX
A0029425
2. Setting the write protection (WP) switch on the main electronics module to the OFF
position (factory setting) disables hardware write protection.
No option is displayed in the Locking status parameter → 160. On the local
display, the -symbol disappears from in front of the parameters in the header
of the operational display and in the navigation view.
Endress+Hauser 159
Operation Proline Promass F 300 EtherNet/IP
11 Operation
None The access status displayed in the Access status parameter applies → 61. Only
appears on local display.
Hardware locked The DIP switch for hardware locking is activated on the PCB board. This locks write
access to the parameters (e.g. via local display or operating tool) .
Temporarily locked Write access to the parameters is temporarily locked on account of internal
processes running in the device (e.g. data upload/download, reset etc.). Once the
internal processing has been completed, the parameters can be changed once
again.
Navigation
"Diagnostics" menu → Measured values
‣ Measured values
‣ Totalizer → 162
160 Endress+Hauser
Proline Promass F 300 EtherNet/IP Operation
Navigation
"Diagnostics" menu → Measured values → Measured variables
‣ Measured variables
Density → 161
Temperature → 162
Concentration → 162
Corrected volume flow – Displays the corrected volume flow Signed floating-point
currently calculated. number
Dependency
The unit is taken from the Corrected
volume flow unit parameter
(→ 96).
Endress+Hauser 161
Operation Proline Promass F 300 EtherNet/IP
Concentration For the following order code: Displays the concentration currently Signed floating-point
"Application package", option ED calculated. number
"Concentration" Dependency
Target mass flow With the following conditions: Displays the target fluid mass flow Signed floating-point
• Order code for "Application package", currently measured. number
option ED "Concentration" Dependency
• The WT-% option or the User conc. The unit is taken from the Mass flow
option is selected in the unit parameter (→ 96).
Concentration unit parameter.
Carrier mass flow With the following conditions: Displays the carrier fluid mass flow Signed floating-point
• Order code for "Application package", currently measured. number
option ED "Concentration" Dependency
• The WT-% option or the User conc. The unit is taken from the Mass flow
option is selected in the unit parameter (→ 96).
Concentration unit parameter.
Navigation
"Diagnostics" menu → Measured values → Totalizer
‣ Totalizer
162 Endress+Hauser
Proline Promass F 300 EtherNet/IP Operation
Totalizer value 1 to n A process variable is selected in the Displays the current totalizer counter Signed floating-point
Assign process variable parameter value. number
(→ 138) of the Totalizer 1 to n
submenu.
Totalizer overflow 1 to n A process variable is selected in the Displays the current totalizer overflow. Integer with sign
Assign process variable parameter
(→ 138) of the Totalizer 1 to n
submenu.
Navigation
"Diagnostics" menu → Measured values → Input values
‣ Input values
Navigation
"Diagnostics" menu → Measured values → Input values → Current input 1 to n
‣ Current input 1 to n
Measured values 1 to n Displays the current input value. Signed floating-point number
Measured current 1 to n Displays the current value of the current input. 0 to 22.5 mA
Endress+Hauser 163
Operation Proline Promass F 300 EtherNet/IP
Navigation
"Diagnostics" menu → Measured values → Input values → Status input 1 to n
‣ Status input 1 to n
Value status input Shows the current input signal level. • High
• Low
Navigation
"Diagnostics" menu → Measured values → Output values
‣ Output values
Navigation
"Diagnostics" menu → Measured values → Output values → Value current output 1 to n
‣ Current output 1 to n
164 Endress+Hauser
Proline Promass F 300 EtherNet/IP Operation
Output current 1 Displays the current value currently calculated for the current 3.59 to 22.5 mA
output.
Measured current Displays the current value currently measured for the current 0 to 30 mA
output.
Navigation
"Diagnostics" menu → Measured values → Output values → Pulse/frequency/switch output
1 to n
‣ Pulse/frequency/switch output
1 to n
Output frequency 1 to n In the Operating mode parameter, the Displays the value currently measured 0.0 to 12 500.0 Hz
Frequency option is selected. for the frequency output.
Pulse output 1 to n In the Operating mode parameter, the Displays the pulse frequency currently Positive floating-point
Pulse option is selected. output. number
Switch status 1 to n The Switch option is selected in the Displays the current switch output • Open
Operating mode parameter. status. • Closed
Navigation
"Diagnostics" menu → Measured values → Output values → Relay output 1 to n
‣ Relay output 1 to n
Endress+Hauser 165
Operation Proline Promass F 300 EtherNet/IP
Switch cycles Shows number of all performed switch cycles. Positive integer
Max. switch cycles number Shows the maximal number of guaranteed switch cycles. Positive integer
Navigation
"Operation" menu → Totalizer handling
‣ Totalizer handling
166 Endress+Hauser
Proline Promass F 300 EtherNet/IP Operation
Control Totalizer 1 to n One of the following options is Control totalizer value. • Totalize Totalize
selected in the Assign process • Reset + hold
variable parameter • Preset + hold
(→ 138) of the Totalizer • Reset + totalize
1 to n submenu: • Preset + totalize
• Volume flow • Hold
• Mass flow
• Corrected volume flow
• Target mass flow *
• Carrier mass flow *
Preset value 1 to n One of the following options is Specify start value for totalizer. Signed floating-point Country-specific:
selected in the Assign process number • 0 kg
Dependency
variable parameter • 0 lb
(→ 138) of the Totalizer
1 to n submenu:
The unit of the selected
process variable is
• Volume flow specified for the totalizer
• Mass flow in the Unit totalizer
• Corrected volume flow parameter (→ 138).
• Target mass flow *
• Carrier mass flow *
Options Description
Reset + hold The totaling process is stopped and the totalizer is reset to 0.
Preset + hold The totaling process is stopped and the totalizer is set to its defined start value
from the Preset value parameter.
Reset + totalize The totalizer is reset to 0 and the totaling process is restarted.
Preset + totalize The totalizer is set to the defined start value from the Preset value parameter and
the totaling process is restarted.
Options Description
Reset + totalize Resets all totalizers to 0 and restarts the totaling process. This deletes all the flow
values previously totalized.
Endress+Hauser 167
Operation Proline Promass F 300 EtherNet/IP
Function range
• A total of 1000 measured values can be stored
• 4 logging channels
• Adjustable logging interval for data logging
• Display of the measured value trend for each logging channel in the form of a chart
/ ../XXXXXXXX
175.77
40.69 kg/h
-100s 0
A0016357
• x-axis: depending on the number of channels selected displays 250 to 1000 measured
values of a process variable.
• y-axis: displays the approximate measured value span and constantly adapts this to the
ongoing measurement.
If the length of the logging interval or the assignment of the process variables to the
channels is changed, the content of the data logging is deleted.
Navigation
"Diagnostics" menu → Data logging
‣ Data logging
‣ Display channel 1
‣ Display channel 2
168 Endress+Hauser
Proline Promass F 300 EtherNet/IP Operation
‣ Display channel 3
‣ Display channel 4
Endress+Hauser 169
Operation Proline Promass F 300 EtherNet/IP
Assign channel 1 The Extended HistoROM Assign a process variable to • Off Off
application package is logging channel. • Mass flow
available. • Volume flow
• Corrected volume
The software options
currently enabled are flow
• Target mass flow *
displayed in the
• Carrier mass flow *
Software option
• Target volume
overview parameter.
flow
• Carrier volume
flow
• Target corrected
volume flow
• Carrier corrected
volume flow
• Density
• Reference density
• Reference density
alternative
• GSV flow
• GSV flow
alternative
• NSV flow
• NSV flow
alternative
• S&W volume flow
• Water cut
• Oil density
• Water density
• Oil mass flow
• Water mass flow
• Oil volume flow
• Water volume flow
• Oil corrected
volume flow
• Water corrected
volume flow
• Concentration *
• Dynamic viscosity
• Kinematic viscosity
• Temp.
compensated
dynamic viscosity
• Temp.
compensated
kinematic viscosity
• Temperature
• Carrier pipe
temperature *
• Electronic
temperature
• Oscillation
frequency 0
• Oscillation
frequency 1
• Oscillation
amplitude *
• Oscillation
amplitude 1
• Frequency
fluctuation 0
• Frequency
fluctuation 1
• Oscillation
damping 0
170 Endress+Hauser
Proline Promass F 300 EtherNet/IP Operation
Assign channel 2 The Extended HistoROM Assign process variable to Picklist, see Assign Off
application package is logging channel. channel 1 parameter
available. (→ 170)
Assign channel 3 The Extended HistoROM Assign process variable to Picklist, see Assign Off
application package is logging channel. channel 1 parameter
available. (→ 170)
Assign channel 4 The Extended HistoROM Assign process variable to Picklist, see Assign Off
application package is logging channel. channel 1 parameter
available. (→ 170)
Logging interval The Extended HistoROM Define the logging interval for 0.1 to 3 600.0 s 1.0 s
application package is data logging. This value
available. defines the time interval
between the individual data
points in the memory.
Clear logging data The Extended HistoROM Clear the entire logging data. • Cancel Cancel
application package is • Clear data
available.
Logging delay In the Data logging Enter the time delay for 0 to 999 h 0h
parameter, the Not measured value logging.
overwriting option is selected.
Data logging control In the Data logging Start and stop measured value • None None
parameter, the Not logging. • Delete + start
overwriting option is selected. • Stop
Endress+Hauser 171
Operation Proline Promass F 300 EtherNet/IP
Data logging status In the Data logging Displays the measured value • Done Done
parameter, the Not logging status. • Delay active
overwriting option is selected. • Active
• Stopped
Entire logging duration In the Data logging Displays the total logging Positive floating- 0s
parameter, the Not duration. point number
overwriting option is selected.
172 Endress+Hauser
Proline Promass F 300 EtherNet/IP Diagnostics and troubleshooting
Local display dark and no output Supply voltage does not match the Apply the correct supply voltage .
signals value indicated on the nameplate.
Local display dark and no output The polarity of the supply voltage is Correct the polarity.
signals wrong.
Local display dark and no output No contact between connecting Check the connection of the cables
signals cables and terminals. and correct if necessary.
Local display dark and no output Terminals are not plugged into the Check terminals.
signals I/O electronics module correctly.
Terminals are not plugged into the
main electronics module correctly.
Local display dark and no output I/O electronics module is defective. Order spare part → 204.
signals Main electronics module is
defective.
Local display is dark, but signal Display is set too bright or too dark. • Set the display brighter by
output is within the valid range simultaneously pressing + .
• Set the display darker by
simultaneously pressing + .
Local display is dark, but signal The cable of the display module is Insert the plug correctly into the
output is within the valid range not plugged in correctly. main electronics module and
display module.
Local display is dark, but signal Display module is defective. Order spare part → 204.
output is within the valid range
Backlighting of local display is red Diagnostic event with "Alarm" Take remedial measures
diagnostic behavior has occurred.
Text on local display appears in a Incorrect operating language is 1. Press + for 2 s ("home
foreign language and cannot be configured. position").
understood. 2. Press .
3. Set the desired language in the
Display language parameter
(→ 148).
Message on local display: Communication between the • Check the cable and the
"Communication Error" display module and the electronics connector between the main
"Check Electronics" is interrupted. electronics module and display
module.
• Order spare part → 204.
Signal output outside the valid Main electronics module is Order spare part → 204.
range defective.
Device shows correct value on local Configuration error Check and correct the parameter
display, but signal output is configuration.
incorrect, though in the valid range.
Device measures incorrectly. Configuration error or device is 1. Check and correct parameter
operated outside the application. configuration.
2. Observe limit values specified in
the "Technical Data".
Endress+Hauser 173
Diagnostics and troubleshooting Proline Promass F 300 EtherNet/IP
For access
Error Possible causes Solution
No write access to parameters Hardware write protection enabled Set the write protection switch on
main electronics module to the OFF
position → 158.
No write access to parameters Current user role has limited access 1. Check user role → 61.
authorization 2. Enter correct customer-specific
access code → 61.
No connection via EtherNet/IP Device plug connected incorrectly Check the pin assignment of the
connector .
Not connecting to Web server Web server disabled Using the "FieldCare" or "DeviceCare"
operating tool, check whether the
Web server of the measuring device
is enabled, and enable it if
necessary→ 69.
Not connecting to Web server • Incorrect IP address 1. If addressing via hardware: open
• IP address is not known the transmitter and check the IP
address configured (last octet).
2. Check the IP address of the
measuring device with the network
manager.
3. If the IP address is not known,
set DIP switch no. 10 to ON, restart
the device and enter the factory IP
address 192.168.1.212.
EtherNet/IP communication
is interrupted by enabling the
DIP switch.
Web browser setting "Use a Proxy Disable the use of the proxy server
Server for Your LAN" is enabled in the Web browser settings of the
computer.
Using the example of MS Internet
Explorer:
1. Under Control Panel open
Internet options.
2. Select the Connections tab and
then double-click LAN settings.
3. In the LAN settings disable the
use of the proxy server and select
OK to confirm.
Apart from the active network • Make sure that no other network
connection to the measuring device, connections are established by
other network connections are also the computer (also no WLAN)
being used. and close other programs with
network access to the computer.
• If using a docking station for
notebooks, make sure that a
network connection to another
network is not active.
Not connecting to Web server Incorrect WLAN access data • Check WLAN network status.
• Log on to the device again using
WLAN access data.
• Verify that WLAN is enabled on
the measuring device and
operating device .
174 Endress+Hauser
Proline Promass F 300 EtherNet/IP Diagnostics and troubleshooting
Not connecting to Web server, No WLAN network available • Check if WLAN reception is
FieldCare or DeviceCare present: LED on display module
is lit blue
• Check if WLAN connection is
enabled: LED on display module
flashes blue
• Switch on instrument function.
Network connection not present or WLAN network is weak. • Operating device is outside of
unstable reception range: Check network
status on operating device.
• To improve network
performance, use an external
WLAN antenna.
Web browser frozen and operation Data transfer active Wait until data transfer or current
no longer possible action is finished.
Content of Web browser Not using optimum version of Web 1. Use the correct Web browser
incomplete or difficult to read server. version .
2. Clear the Web browser cache and
restart the Web browser.
Operation with FieldCare or Firewall of computer or network is Depending on the settings of the
DeviceCare via CDI-RJ45 service preventing communication firewall used on the computer or in
interface (port 8000) the network, the firewall must be
adapted or disabled to allow
FieldCare/DeviceCare access.
Flashing of firmware with FieldCare Firewall of computer or network is Depending on the settings of the
or DeviceCare via CDI-RJ45 service preventing communication firewall used on the computer or in
interface (via port 8000 or TFTP the network, the firewall must be
ports) adapted or disabled to allow
FieldCare/DeviceCare access.
12.2.1 Transmitter
Different LEDs in the transmitter provide information on the device status.
Endress+Hauser 175
Diagnostics and troubleshooting Proline Promass F 300 EtherNet/IP
1 2 3 4 5
A0029629
1 Supply voltage
2 Device status
3 network status
4 Port 1 active: EtherNet/IP
5 Port 2 active: EtherNet/IP and service interface (CDI)
3 Network status Off The device does not have an EtherNet/IP address.
176 Endress+Hauser
Proline Promass F 300 EtherNet/IP Diagnostics and troubleshooting
2 1
XXXXXXXXX S XXXXXXXXX S
20.50 S801
Supply voltage
3
4
X XX i Menu
5
A0029426-EN
1 Status signal
2 Diagnostic behavior
3 Diagnostic behavior with diagnostic code
4 Short text
5 Operating elements
If two or more diagnostic events are pending simultaneously, only the message of the
diagnostic event with the highest priority is shown.
Other diagnostic events that have occurred can be displayed in the Diagnostics menu:
• Via parameter
• Via submenus → 197
Status signals
The status signals provide information on the state and reliability of the device by
categorizing the cause of the diagnostic information (diagnostic event).
The status signals are categorized according to VDI/VDE 2650 and NAMUR
Recommendation NE 107: F = Failure, C = Function Check, S = Out of Specification, M
= Maintenance Required
Symbol Meaning
Failure
A device error has occurred. The measured value is no longer valid.
Function check
The device is in service mode (e.g. during a simulation).
Out of specification
The device is operated:
Outside its technical specification limits (e.g. outside the process temperature range)
Maintenance required
Maintenance is required. The measured value remains valid.
Endress+Hauser 177
Diagnostics and troubleshooting Proline Promass F 300 EtherNet/IP
Diagnostic behavior
Symbol Meaning
Alarm
• Measurement is interrupted.
• Signal outputs and totalizers assume the defined alarm condition.
• A diagnostic message is generated.
Warning
Measurement is resumed. The signal outputs and totalizers are not affected. A diagnostic
message is generated.
Diagnostic information
The fault can be identified using the diagnostic information. The short text helps you by
providing information about the fault. In addition, the corresponding symbol for the
diagnostic behavior is displayed in front of the diagnostic information on the local display.
Diagnostic information
Diagnostic code
Diagnostic Diagnostic
Status signal Short text
behavior number
↓ ↓ ↓
Operating elements
Key Meaning
Plus key
In a menu, submenu
Opens the message about remedy information.
Enter key
In a menu, submenu
Opens the operating menu.
178 Endress+Hauser
Proline Promass F 300 EtherNet/IP Diagnostics and troubleshooting
XXXXXXXXX S XXXXXXXXX S
20.50 S801
Supply voltage
X XX i Menu
1.
Diagnostic list S
Diagnostics 1
1 S801 Supply voltage
Diagnostics 2
Diagnostics 3
2.
3.
A0029431-EN
The user is in the Diagnostics menu at an entry for a diagnostics event, e.g. in the
Diagnostic list submenu or Previous diagnostics parameter.
1. Press .
The message for the remedial measures for the selected diagnostic event opens.
2. Press + simultaneously.
The message for the remedial measures closes.
Endress+Hauser 179
Diagnostics and troubleshooting Proline Promass F 300 EtherNet/IP
2 3
A0031056
In addition, diagnostic events which have occurred can be shown in the Diagnostics
menu:
• Via parameter
• Via submenu → 197
Status signals
The status signals provide information on the state and reliability of the device by
categorizing the cause of the diagnostic information (diagnostic event).
Symbol Meaning
Failure
A device error has occurred. The measured value is no longer valid.
Function check
The device is in service mode (e.g. during a simulation).
Out of specification
The device is operated:
Outside its technical specification limits (e.g. outside the process temperature range)
Maintenance required
Maintenance is required. The measured value is still valid.
The status signals are categorized in accordance with VDI/VDE 2650 and NAMUR
Recommendation NE 107.
180 Endress+Hauser
Proline Promass F 300 EtherNet/IP Diagnostics and troubleshooting
Xxxxxx/…/…/
Device name: Xxxxxxx Mass flow: 12.34 kg/h
Device tag: Xxxxxxx Volume flow: 12.34 m³/h
Status signal: Function check (C)
Xxxxxx
P Diagnostics 1: C485 Simu...
P Remedy information: Deactivate...
P Access status tooling: Mainenance Failure (F)
A0021799-EN
In addition, diagnostic events which have occurred can be shown in the Diagnostics
menu:
• Via parameter
• Via submenu → 197
Diagnostic information
The fault can be identified using the diagnostic information. The short text helps you by
providing information about the fault. In addition, the corresponding symbol for the
diagnostic behavior is displayed in front of the diagnostic information on the local display.
Diagnostic information
Diagnostic code
Diagnostic Diagnostic
Status signal Short text
behavior number
↓ ↓ ↓
Endress+Hauser 181
Diagnostics and troubleshooting Proline Promass F 300 EtherNet/IP
Bytes 0 1 2 3 4 5 6 7
↓ ↓ ↓
Diagnostic number
of the diagnostic
event that is
"Empty" or "Padding displayed in the
Content "Empty" or "Reserved"
bytes" Actual
diagnostics
parameter , e.g.
242
Options Description
Alarm The device stops measurement. The totalizers assume the defined alarm condition. A
diagnostic message is generated.
The background lighting changes to red.
Warning The device continues to measure. The totalizers are not affected. A diagnostic message is
generated.
Logbook entry only The device continues to measure. The diagnostic message is displayed only in the Event
logbook submenu (Event list submenu) and is not displayed in alternation with the
operational display.
Off The diagnostic event is ignored, and no diagnostic message is generated or entered.
182 Endress+Hauser
Proline Promass F 300 EtherNet/IP Diagnostics and troubleshooting
Endress+Hauser 183
Diagnostics and troubleshooting Proline Promass F 300 EtherNet/IP
184 Endress+Hauser
Proline Promass F 300 EtherNet/IP Diagnostics and troubleshooting
Endress+Hauser 185
Diagnostics and troubleshooting Proline Promass F 300 EtherNet/IP
Status signal F
332 Writing in HistoROM backup failed Replace user interface board 0x10002C7
Ex d/XP: replace transmitter
Status signal F
186 Endress+Hauser
Proline Promass F 300 EtherNet/IP Diagnostics and troubleshooting
373 Sensor electronic (ISEM) faulty 1. Transfer data or reset device 0x10002D1
2. Contact service
Status signal F
Endress+Hauser 187
Diagnostics and troubleshooting Proline Promass F 300 EtherNet/IP
Status signal F
188 Endress+Hauser
Proline Promass F 300 EtherNet/IP Diagnostics and troubleshooting
Status signal C
Status signal C
Endress+Hauser 189
Diagnostics and troubleshooting Proline Promass F 300 EtherNet/IP
Status signal C
Status signal C
Status signal C
190 Endress+Hauser
Proline Promass F 300 EtherNet/IP Diagnostics and troubleshooting
Status signal C
Status signal C
Status signal C
Status signal C
Status signal C
Endress+Hauser 191
Diagnostics and troubleshooting Proline Promass F 300 EtherNet/IP
Status signal C
Status signal C
192 Endress+Hauser
Proline Promass F 300 EtherNet/IP Diagnostics and troubleshooting
830 Sensor temperature too high Reduce ambient temp. around the 0x80000C0
sensor housing
Status signal S
831 Sensor temperature too low Increase ambient temp. around the 0x80000C2
sensor housing
Status signal S
Endress+Hauser 193
Diagnostics and troubleshooting Proline Promass F 300 EtherNet/IP
Status signal S
Status signal S
194 Endress+Hauser
Proline Promass F 300 EtherNet/IP Diagnostics and troubleshooting
941 API temperature out of specification 1. Check process temperature with 0x8000380
selected API commodity group
Status signal S 2. Check API related parameters
942 API density out of specification 1. Check process density with 0x800033B
selected API commodity group
Status signal S 2. Check API related parameters
943 API pressure out of specification 1. Check process pressure with 0x800037F
selected API commodity group
Status signal S 2. Check API related parameters
Endress+Hauser 195
Diagnostics and troubleshooting Proline Promass F 300 EtherNet/IP
Status signal S
Navigation
"Diagnostics" menu
Diagnostics
Actual diagnostics A diagnostic event has occurred. Shows the current occured diagnostic Symbol for diagnostic
event along with its diagnostic behavior, diagnostic code
information. and short message.
Ifsimultaneously,
two or more messages occur
the message with
the highest priority is shown on
the display.
Previous diagnostics Two diagnostic events have already Shows the diagnostic event that Symbol for diagnostic
occurred. occurred prior to the current diagnostic behavior, diagnostic code
event along with its diagnostic and short message.
information.
Operating time from restart – Shows the time the device has been in Days (d), hours (h),
operation since the last device restart. minutes (m) and seconds
(s)
Operating time – Indicates how long the device has been Days (d), hours (h),
in operation. minutes (m) and seconds
(s)
196 Endress+Hauser
Proline Promass F 300 EtherNet/IP Diagnostics and troubleshooting
/ ../Diagnose list
Diagnostics
F273 Main electronic
Diagnostics 2
Diagnostics 3
A0014006-EN
/ ../Eventlist F
I1091 Config. change
I1157 Mem.err. ev.list
0d01h19m10s
F311 Electr. failure
A0014008-EN
Endress+Hauser 197
Diagnostics and troubleshooting Proline Promass F 300 EtherNet/IP
In addition to the operation time of its occurrence, each event is also assigned a symbol
that indicates whether the event has occurred or is ended:
• Diagnostic event
– : Occurrence of the event
– : End of the event
• Information event
: Occurrence of the event
To call up the measures to rectify a diagnostic event:
• Via local display → 179
• Via Web browser → 180
• Via "FieldCare" operating tool → 181
• Via "DeviceCare" operating tool → 181
For filtering the displayed event messages → 198
I1089 Power on
198 Endress+Hauser
Proline Promass F 300 EtherNet/IP Diagnostics and troubleshooting
I1451 Monitoring on
Endress+Hauser 199
Diagnostics and troubleshooting Proline Promass F 300 EtherNet/IP
Options Description
To delivery settings Every parameter for which a customer-specific default setting was ordered is reset
to this customer-specific value. All other parameters are reset to the factory
setting.
Restart device The restart resets every parameter whose data are in the volatile memory (RAM)
to the factory setting (e.g. measured value data). The device configuration remains
unchanged.
Restore S-DAT backup Restore the data that are saved on the S-DAT. The data record is restored from the
electronics memory to the S-DAT.
Navigation
"Diagnostics" menu → Device information
‣ Device information
200 Endress+Hauser
Proline Promass F 300 EtherNet/IP Diagnostics and troubleshooting
Device tag Shows name of measuring point. Max. 32 characters, such as Promass
letters, numbers or special
characters (e.g. @, %, /).
Serial number Shows the serial number of the measuring A maximum of 11-digit –
device. character string comprising
letters and numbers.
Firmware version Shows the device firmware version installed. Character string in the format –
xx.yy.zz
Order code Shows the device order code. Character string composed of –
letters, numbers and certain
The order code can be found on the
nameplate of the sensor and punctuation marks (e.g. /).
transmitter in the "Order code" field.
Extended order code 1 Shows the 1st part of the extended order Character string –
code.
Extended order code 2 Shows the 2nd part of the extended order Character string –
code.
Extended order code 3 Shows the 3rd part of the extended order Character string –
code.
ENP version Shows the version of the electronic Character string 2.02.00
nameplate (ENP).
Endress+Hauser 201
Diagnostics and troubleshooting Proline Promass F 300 EtherNet/IP
It is possible to flash the firmware to the current version or the previous version using
the service interface.
For the compatibility of the firmware version with the previous version, the installed
device description files and operating tools, observe the information about the device
in the "Manufacturer's information" document.
The manufacturer's information is available:
• In the Download Area of the Endress+Hauser web site: www.endress.com →
Downloads
• Specify the following details:
– Product root: e.g. 8F 3B
The product root is the first part of the order code: see the nameplate on the
device.
– Text search: Manufacturer's information
– Media type: Documentation – Technical Documentation
202 Endress+Hauser
Proline Promass F 300 EtherNet/IP Maintenance
13 Maintenance
Endress+Hauser 203
Repairs Proline Promass F 300 EtherNet/IP
14 Repairs
14.4 Return
The measuring device must be returned if it is need of repair or a factory calibration, or if
the wrong measuring device has been delivered or ordered. Legal specifications require
Endress+Hauser, as an ISO-certified company, to follow certain procedures when handling
products that are in contact with the medium.
To ensure safe, swift and professional device returns, please refer to the procedure and
conditions for returning devices provided on the Endress+Hauser website at
http://www.endress.com/support/return-material
204 Endress+Hauser
Proline Promass F 300 EtherNet/IP Repairs
14.5 Disposal
LWARNING
Danger to persons from process conditions.
‣ Beware of hazardous process conditions such as pressure in the measuring device, high
temperatures or aggressive fluids.
2. Carry out the mounting and connection steps from the "Mounting the measuring
device" and "Connecting the measuring device" sections in reverse order. Observe the
safety instructions.
Endress+Hauser 205
Accessories Proline Promass F 300 EtherNet/IP
15 Accessories
Various accessories, which can be ordered with the device or subsequently from Endress
+Hauser, are available for the device. Detailed information on the order code in question is
available from your local Endress+Hauser sales center or on the product page of the
Endress+Hauser website: www.endress.com.
Accessories Description
Transmitter Promass300 Transmitter for replacement or storage. Use the order code to define the following
specifications:
• Approvals
• Output
• Input
• Display / operation
• Housing
• Software
206 Endress+Hauser
Proline Promass F 300 EtherNet/IP Accessories
Accessories Description
Heating jacket Is used to stabilize the temperature of the fluids in the sensor.
Water, water vapor and other non-corrosive liquids are permitted for use as fluids.
If using oil as a heating medium, please consult with Endress+Hauser.
Heating jackets cannot be used with sensors fitted with a rupture disk.
Field Xpert SFX350 Field Xpert SFX350 is a mobile computer for commissioning and maintenance. It
enables efficient device configuration and diagnostics for and can be used in non-
hazardous areas.
Endress+Hauser 207
Accessories Proline Promass F 300 EtherNet/IP
Memograph M graphic The Memograph M graphic data manager provides information on all the relevant
data manager measured variables. Measured values are recorded correctly, limit values are
monitored and measuring points analyzed. The data are stored in the 256 MB
internal memory and also on a SD card or USB stick.
BA00247R
For details, see "Technical Information" TI00133R and Operating Instructions
Cerabar M The pressure transmitter for measuring the absolute and gauge pressure of gases,
steam and liquids. It can be used to read in the operating pressure value.
Cerabar S The pressure transmitter for measuring the absolute and gauge pressure of gases,
steam and liquids. It can be used to read in the operating pressure value.
BA00271P
For details, see "Technical Information" TI00383P and Operating Instructions
iTEMP The temperature transmitters can be used in all applications and are suitable for
the measurement of gases, steam and liquids. They can be used to read in the
medium temperature.
208 Endress+Hauser
Proline Promass F 300 EtherNet/IP Technical data
16 Technical data
16.1 Application
The measuring device is suitable for flow measurement of liquids and gases only.
Depending on the version ordered, the measuring device can also measure potentially
explosive, flammable, poisonous and oxidizing media.
To ensure that the device remains in proper operating condition for its service life, use the
measuring device only for media against which the process-wetted materials are
sufficiently resistant.
Measuring principle Mass flow measurement based on the Coriolis measuring principle
Endress+Hauser 209
Technical data Proline Promass F 300 EtherNet/IP
16.3 Input
15 ½ 0 to 6 500 0 to 238.9
25 1 0 to 18 000 0 to 661.5
40 1½ 0 to 45 000 0 to 1 654
50 2 0 to 70 000 0 to 2 573
DN x
8 ³⁄₈ 60
15 ½ 80
25 1 90
40 1½ 90
50 2 90
80 3 110
210 Endress+Hauser
Proline Promass F 300 EtherNet/IP Technical data
DN x
100 4 130
150 6 200
250 10 200
Current input
The measured values are written from the automation system to the measuring device via
the current input → 211.
Digital communication
The measured values are written from the automation system to the measuring device via
EtherNet/IP.
Resolution 1 µA
Endress+Hauser 211
Technical data Proline Promass F 300 EtherNet/IP
Status input
212 Endress+Hauser
Proline Promass F 300 EtherNet/IP Technical data
16.4 Output
Ex-i, passive
Open-circuit voltage DC 28.8 V (active)
Load 0 to 700 Ω
Resolution 0.38 µA
The range of options increases if the measuring device has one or more
application packages.
Pulse/frequency/switch output
Ex-i, passive
Maximum input values DC 30 V, 250 mA (passive)
Pulse output
Endress+Hauser 213
Technical data Proline Promass F 300 EtherNet/IP
Frequency output
Output frequency Adjustable: end value frequency 2 to 10 000 Hz (f max = 12 500 Hz)
The range of options increases if the measuring device has one or more
application packages.
Switch output
The range of options increases if the measuring device has one or more
application packages.
Relay output
214 Endress+Hauser
Proline Promass F 300 EtherNet/IP Technical data
The range of options increases if the measuring device has one or more
application packages.
EtherNet/IP
4 to 20 mA
Failure mode Choose from:
• 4 to 20 mA in accordance with NAMUR recommendation NE 43
• 4 to 20 mA in accordance with US
• Min. value: 3.59 mA
• Max. value: 22.5 mA
• Freely definable value between: 3.59 to 22.5 mA
• Actual value
• Last valid value
0 to 20 mA
Failure mode Choose from:
• Maximum alarm: 22 mA
• Freely definable value between: 0 to 20.5 mA
Endress+Hauser 215
Technical data Proline Promass F 300 EtherNet/IP
Pulse/frequency/switch output
Pulse output
Frequency output
Switch output
Relay output
Local display
Interface/protocol
• Via digital communication:
EtherNet/IP
• Via service interface
– CDI-RJ45 service interface
– WLAN interface
Web server
216 Endress+Hauser
Proline Promass F 300 EtherNet/IP Technical data
Low flow cut off The switch points for low flow cut off are user-selectable.
Galvanic isolation The outputs are galvanically isolated from one another and from earth (PE).
Protocol-specific data Protocol • The CIP Networks Library Volume 1: Common Industrial Protocol
• The CIP Networks Library Volume 2: EtherNet/IP Adaptation of CIP
Manufacturer ID 0x11
Baud rates Automatic ¹⁰⁄₁₀₀ Mbit with half-duplex and full-duplex detection
Polarity Auto-polarity for automatic correction of crossed TxD and RxD pairs
Configuration options for • DIP switches on the electronics module for IP addressing
measuring device • Manufacturer-specific software (FieldCare)
• Add-on Profile Level 3 for Rockwell Automation control systems
• Web browser
• Electronic Data Sheet (EDS) integrated in the measuring device
Configuration of the EtherNet • Speed: 10 MBit, 100 MBit, auto (factory setting)
interface • Duplex: half-duplex, full-duplex, auto (factory setting)
Configuration of the device • DIP switches on the electronics module for IP addressing (last octet)
address • DHCP
• Manufacturer-specific software (FieldCare)
• Add-on Profile Level 3 for Rockwell Automation control systems
• Web browser
• EtherNet/IP tools, e.g. RSLinx (Rockwell Automation)
Terminal assignment → 34
Endress+Hauser 217
Technical data Proline Promass F 300 EtherNet/IP
DC24 V ±20% –
Option I
AC100 to 240 V –15…+10% 50/60 Hz
Electrical connection → 35
Potential equalization → 41
terminals Spring-loaded terminals: Suitable for strands and strands with ferrules.
Conductor cross-section 0.2 to 2.5 mm2 (24 to 12 AWG).
Cable entries • Cable gland: M20 × 1.5 with cable ⌀ 6 to 12 mm (0.24 to 0.47 in)
• Thread for cable entry:
– NPT ½"
– G ½"
– M20
• Device plug for digital communication: M12
Cable specification → 31
218 Endress+Hauser
Proline Promass F 300 EtherNet/IP Technical data
Base accuracy
Design fundamentals → 223
Density (liquids)
Temperature
±0.5 °C ± 0.005 · T °C (±0.9 °F ± 0.003 · (T – 32) °F)
15 ½ 0.200 0.007
25 1 0.540 0.019
40 1½ 2.25 0.083
50 2 3.50 0.129
80 3 9.0 0.330
Endress+Hauser 219
Technical data Proline Promass F 300 EtherNet/IP
High-temperature version: order code for "Measuring tube material", option TT, TU
DN Zero point stability
25 1 1.80 0.0661
50 2 7.00 0.2572
80 3 18.0 0.6610
For devices with low-temperature version, order code for "Measuring tube mat., wetted
surface", option LA, please note the following:
NOTICE
Zero point confirmation and zero point adjustment are difficult to carry out in the
field due to the vaporization of the cryogenic liquid.
‣ As a general rule, the factory-set zero point should not be changed. Please ensure that
the medium is in the liquid phase if a zero point adjustment is to be carried out.
Flow values
Flow values as turndown parameter depending on nominal diameter.
SI units
250 2 200 000 220 000 110 000 44 000 22 000 4 400
US units
220 Endress+Hauser
Proline Promass F 300 EtherNet/IP Technical data
Accuracy of outputs
The outputs have the following base accuracy specifications.
Current output
Accuracy ±5 µA
Pulse/frequency output
o.r. = of reading
Accuracy Max. ±50 ppm o.r. (over the entire ambient temperature range)
Base repeatability
Design fundamentals → 223
Density (liquids)
±0.00025 g/cm3
Temperature
±0.25 °C ± 0.0025 · T °C (±0.45 °F ± 0.0015 · (T–32) °F)
Pulse/frequency output
Endress+Hauser 221
Technical data Proline Promass F 300 EtherNet/IP
Density
When there is a difference between the density calibration temperature and the process
temperature, the typical measured error of the sensor is
±0.00005 g/cm3 /°C (±0.000025 g/cm3 /°F). Field density calibration is possible.
Wide-range density specification (special density calibration)
If the process temperature is outside the valid range (→ 219) the measured error is
±0.00005 g/cm3 /°C (±0.000025 g/cm3 /°F)
3
[kg/m ]
18
16
14
12 1
10
8
2
6
4
2
0
-50 0 100 200 300 [°C]
Temperature
±0.005 · T °C (± 0.005 · (T – 32) °F)
Influence of medium The table below shows the effect on accuracy of mass flow due to a difference between
pressure calibration pressure and process pressure.
o.r. = of reading
It is possible to compensate for the effect by:
• Reading in the current pressure measured value via the current input.
• Specifying a fixed value for the pressure in the device parameters.
Operating Instructions .
DN [% o.r./bar] [% o.r./psi]
[mm] [in]
8 3/8 no influence
15 ½ no influence
25 1 no influence
40 1½ –0.003 –0.0002
222 Endress+Hauser
Proline Promass F 300 EtherNet/IP Technical data
DN [% o.r./bar] [% o.r./psi]
[mm] [in]
50 2 –0.008 –0.0006
80 3 –0.009 –0.0006
ZeroPoint
³ BaseAccu ⋅ 100 ± BaseAccu
A0021339
A0021332
ZeroPoint ZeroPoint
< BaseAccu ⋅ 100 MeasValue ⋅ 100
±
A0021333 A0021334
½ ⋅ ZeroPoint
³ BaseRepeat ⋅ 100
A0021340
A0021335
½ ⋅ ZeroPoint ZeroPoint
< BaseRepeat ⋅ 100 ± ½ ⋅ MeasValue ⋅ 100
A0021336 A0021337
E [%]
2.5
2.0
1.5
1.0
0.5
0
0 10 20 30 40 50 60 70 80 90 100 Q [%]
A0028808
16.7 Installation
"Mounting requirements"
Endress+Hauser 223
Technical data Proline Promass F 300 EtherNet/IP
16.8 Environment
Ambient temperature → 24
range
Temperature tables
Observe the interdependencies between the permitted ambient and fluid
temperatures when operating the device in hazardous areas.
For detailed information on the temperature tables, see the separate document
entitled "Safety Instructions" (XA) for the device.
Mechanical load Never use the transmitter housing as a ladder or climbing aid.
16.9 Process
224 Endress+Hauser
Proline Promass F 300 EtherNet/IP Technical data
Standard version –50 to +150 °C (–58 to +302 °F) Order code for "Measuring tube
mat., wetted surface", option HA,
SA, SB, SC
Extended temperature version –50 to +240 °C (–58 to +464 °F) Order code for "Measuring tube
mat., wetted surface", option SD,
SE, SF, TH
High-temperature version –50 to +350 °C (–58 to +662 °F) For nominal diameters DN 25 (1"),
DN 50 (2") and DN 80 (3")
Order code for "Measuring tube
mat., wetted surface", option TT,
TU
Low-temperature version –196 to +150 °C (–320 to +302 °F) Order code for "Measuring tube
mat., wetted surface", option LA
NOTICE
Material fatigue due to excessive
temperature differential.
‣ The maximum temperature
differential of the fluid used
must not exceed 300 K.
Ta
A
B
Tm
A0031121
Endress+Hauser 225
Technical data Proline Promass F 300 EtherNet/IP
A B A B
Version Ta Tm Ta Tm Ta Tm Ta Tm
Standard version 60 °C (140 °F) 150 °C (302 °F) – – 60 °C (140 °F) 110 °C (230 °F) 55 °C (131 °F) 150 °C
(302 °F)
Extended 60 °C (140 °F) 170 °C (338 °F) 55 °C (131 °F) 240 °C 60 °C (140 °F) 110 °C (230 °F) 50 °C (122 °F) 240 °C
temperature (464 °F) (464 °F)
version
High-temperature 60 °C (140 °F) 350 °C (662 °F) – – 60 °C (140 °F) 350 °C (662 °F) – –
version
Secondary containment For the Standard version with the temperature range –50 to +150 °C (–58 to +302 °F), the
sensor housing is filled with dry nitrogen gas and protects the electronics and mechanics
inside.
For all other temperature versions the sensor housing is filled with dry inert gas.
The following secondary containment pressure ratings/burst pressures are only valid for
standard devices and/or devices equipped with closed purge connections (never opened/as
delivered).
If a device fitted with purge connections (order code for "Sensor option", option CH "Purge
connection") is connected to the purge system, the maximum nominal pressure is
determined by the purge system itself or by the device, depending on which component
has the lower nominal pressure classification.
If the device is fitted with a rupture disk (order code for "Sensor option", option CA "Rupture
disk"), the rupture disk trigger pressure is decisive for the maximum nominal pressure
→ 227.
The secondary containment burst pressure refers to a typical internal pressure achieved
prior to mechanical failure of the secondary containment as determined during type
testing. The corresponding type test declaration can be ordered with the device (order code
for "Additional approval", option LN "Type test containment").
226 Endress+Hauser
Proline Promass F 300 EtherNet/IP Technical data
If a measuring tube fails (e.g. due to process characteristics like corrosive or abrasive
fluids), the fluid will be contained by the secondary containment.
If there is a need to drain the leaking medium into a discharge device, the sensor should be
fitted with a rupture disk. Connect the discharge to the additional threaded connection .
If the sensor is to be purged with gas (gas detection), it should be equipped with purge
connections.
Do not open the purge connections unless the containment can be filled immediately
with a dry, inert gas. Use only low pressure to purge. Maximum pressure:
5 bar (72.5 psi).
In case of a tube failure, the pressure level inside the secondary containment will rise
according to the operating process pressure. If the user judges that the secondary
containment pressure rating/burst pressure does not provide an adequate safety margin,
the device can be fitted with a rupture disk. This prevents excessively high pressure from
forming inside the secondary containment. Therefore, the use of a rupture disk is strongly
recommended in applications involving high gas pressures, and particularly in applications
in which the process pressure is greater than 2/3 of the secondary containment burst
pressure.
For information on the dimensions: see the "Mechanical construction" section of the
"Technical Information" document
Rupture disk To increase the level of safety, a device version with a rupture disk with a trigger pressure
of 10 to 15 bar (145 to 217.5 psi)can be used (order code for "Sensor option", option CA
"rupture disk").
Rupture disks cannot be combined with the separately available heating jacket .
Special mounting instructions: → 26
For information on the dimensions: see the "Mechanical construction" section
Flow limit Select the nominal diameter by optimizing between the required flow range and
permissible pressure loss.
For an overview of the full scale values for the measuring range, see the "Measuring
range" section → 210
• The minimum recommended full scale value is approx. 1/20 of the maximum full scale
value
• In most applications, 20 to 50 % of the maximum full scale value can be considered ideal
• A low full scale value must be selected for abrasive media (such as liquids with entrained
solids): flow velocity < 1 m/s (< 3 ft/s).
• For gas measurement the following rules apply:
– The flow velocity in the measuring tubes should not exceed half the sound velocity
(0.5 Mach).
– The maximum mass flow depends on the density of the gas: formula → 210
To calculate the flow limit, use the Applicator sizing tool → 207
Endress+Hauser 227
Technical data Proline Promass F 300 EtherNet/IP
Pressure loss To calculate the pressure loss, use the Applicator sizing tool → 207
Promass F with reduced pressure loss: order code for "Sensor option", option CE "reduced
pressure loss"
System pressure → 24
Design, dimensions For the dimensions and installation lengths of the device, see the "Technical
Information" document, "Mechanical construction" section.
Weight All values (weight exclusive of packaging material) refer to devices with EN/DIN
PN 40 flanges. Weight specifications including transmitter as per order code for "Housing",
option A "Aluminum, coated".
Different values due to different transmitter versions:
• Transmitter version for the hazardous area
(Order code for "Housing", option A "Aluminum, coated"; Ex d): +2 kg (+4.4 lbs)
• Cast transmitter version, stainless
(Order code for "Housing", option L "Cast, stainless"): +6 kg (+13 lbs)
• Transmitter version for hygienic area
(Order code for "Housing", option B "Stainless, hygienic"): +0.2 kg (+0.44 lbs)
Weight in SI units
8 11
15 12
25 14
40 19
50 30
80 55
100 96
150 154
250 400
Weight in US units
3/8 24
½ 26
1 31
1½ 42
2 66
3 121
4 212
228 Endress+Hauser
Proline Promass F 300 EtherNet/IP Technical data
6 340
10 882
Window material
Order code for "Housing":
• Option A "Aluminum, coated": glass
• Option B "Stainless, hygienic": polycarbonate
For order code for "Approval", option BS, CZ, GS, MS and NS: glass
• Option L "Cast, stainless": glass
Seals
Order code for "Housing":
Option B "Stainless, hygienic": EPDM
Adapter for cable entry with internal thread G ½" Nickel-plated brass
Adapter for cable entry with internal thread G ½" Nickel-plated brass
Endress+Hauser 229
Technical data Proline Promass F 300 EtherNet/IP
Sensor housing
• Acid and alkali-resistant outer surface
• DN 08 to DN 150: stainless steel, 1.4301 (304)
Optional: order code for "Sensor option", option CC: stainless steel, 1.4404 (316L)
• DN 250:
– Order code for "Measuring tube material", option SA, SD (stainless steel, 1.4404
(316L)):
Stainless steel 1.4301 (304)
Optional: order code for "Sensor option", option CC: stainless steel, 1.4404 (316L)
– Order code for "Measuring tube material", option LA (stainless steel, 1.4404 (316L)):
Stainless steel 1.4301 (304)
Optional: order code for "Sensor option", option CC: stainless steel, 1.4404 (316L)
– Order code for "Measuring tube material", option HA, TH (Alloy C22):
Stainless steel, 1.4404 (316L)
Measuring tubes
• DN 8 to 100 (3/8…4"): stainless steel, 1.4539 (904L);
Manifold: stainless steel, 1.4404 (316/316L)
• DN 150 (6"), DN 250 (10"): stainless steel, 1.4404 (316/316L);
Manifold: stainless steel, 1.4404 (316/316L)
• DN 8 to 250 (3/8 to 10"): Alloy C22, 2.4602 (UNS N06022);
Manifold: Alloy C22, 2.4602 (UNS N06022)
High-temperature version
DN 25, DN 50, DN 80 (DN 1", DN 2", DN 3"): Alloy C22, 2.4602 (UNS N06022)
Process connections
• Flanges according to EN 1092-1 (DIN2501) / according to ASME B 16.5 /
as per JIS B2220:
– Stainless steel, 1.4404 (F316/F316L)
– Alloy C22, 2.4602 (UNS N06022)
– Lap joint flanges: stainless steel, 1.4301 (F304); wetted parts Alloy C22
• All other process connections:
Stainless steel, 1.4404 (316/316L)
High-temperature version
Flanges according to EN 1092-1 (DIN2501) / according to ASME B 16.5 /
as per JIS B2220:
– Stainless steel, 1.4404 (F316/F316L)
– Alloy C22, 2.4602 (UNS N06022)
List of all available process connections → 231
230 Endress+Hauser
Proline Promass F 300 EtherNet/IP Technical data
Seals
Welded process connections without internal seals
Accessories
Protective cover
Stainless steel, 1.4404 (316L)
Surface roughness All data relate to parts in contact with fluid. The following surface roughness quality can
be ordered.
• Not polished
• Ramax = 0.8 µm (32 µin)
• Ramax = 0.4 µm (16 µin)
16.11 Operability
Endress+Hauser 231
Technical data Proline Promass F 300 EtherNet/IP
A0026785
Display elements
• 4-line, illuminated, graphic display
• White background lighting; switches to red in event of device errors
• Format for displaying measured variables and status variables can be individually
configured
• Permitted ambient temperature for the display: –20 to +60 °C (–4 to +140 °F)
The readability of the display may be impaired at temperatures outside the temperature
range.
Operating elements
• External operation via touch control (3 optical keys) without opening the housing: ,
,
• Operating elements also accessible in the various zones of the hazardous area
232 Endress+Hauser
Proline Promass F 300 EtherNet/IP Technical data
A0026786
Material
The housing material of the display and operating module DKX001 depends on the choice
of transmitter housing material.
Cable entry
Corresponds to the choice of transmitter housing, order code for "Electrical connection".
Connecting cable
Dimensions
Information on the dimensions:
"Mechanical construction" section of the "Technical Information" document.
Remote operation → 70
Service interface → 71
Supported operating tools Different operating tools can be used for local or remote access to the measuring device.
Depending on the operating tool used, access is possible with different operating units and
via a variety of interfaces.
Endress+Hauser 233
Technical data Proline Promass F 300 EtherNet/IP
Other operating tools based on FDT technology with a device driver such as DTM/
iDTM or DD/EDD can be used for device operation. These operating tools are available
from the individual manufacturers. Integration into the following operating tools,
among others, is supported:
• Field Device Manager (FDM) by Honeywell → www.honeywellprocess.com
• FieldMate by Yokogawa → www.yokogawa.com
• PACTWare → www.pactware.com
The associated device description files are available at: www.endress.com →
Downloads
Web server
Thanks to the integrated Web server, the device can be operated and configured via a Web
browser and via a service interface (CDI-RJ45) or via a WLAN interface. The structure of
the operating menu is the same as for the local display. In addition to the measured values,
status information on the device is also displayed and allows the user to monitor the status
of the device. Furthermore the device data can be managed and the network parameters
can be configured.
A device that has a WLAN interface (can be ordered as an option) is required for the
WLAN connection: order code for "Display; operation", option G "4-line, illuminated; touch
control + WLAN". The device acts as an Access Point and enables communication by
computer or a mobile handheld terminal.
Supported functions
Data exchange between the operating unit (such as a notebook for example) and the
measuring device:
– Upload the configuration from the measuring device (XML format, configuration
backup)
– Save the configuration to the measuring device (XML format, restore configuration)
– Export event list (.csv file)
– Export parameter settings (.csv file or PDF file, document the measuring point
configuration)
– Export the Heartbeat verification log (PDF file, only available with the "Heartbeat
Verification" application package)
– Flash firmware version for device firmware upgrade, for instance
– Download driver for system integration
– Visualize up to 1000 saved measured values (only available with the Extended
HistoROM application package → 239)
Webserver special documentation → 241
234 Endress+Hauser
Proline Promass F 300 EtherNet/IP Technical data
HistoROM The measuring device features HistoROM data management. HistoROM data management
data management comprises both the storage and import/export of key device and process data, making
operation and servicing far more reliable, secure and efficient.
When the device is delivered, the factory settings of the configuration data are stored
as a backup in the device memory. This memory can be overwritten with an updated
data record, for example after commissioning.
There are different types of data storage units in which device data are stored and used by the device:
Device memory T-DAT S-DAT
Available data • Event logbook such as diagnostic events • Measured value logging ("Extended • Sensor data: nominal diameter etc.
for example HistoROM" order option) • Serial number
• Parameter data record backup • Current parameter data record (used • Calibration data
• Device firmware package by firmware at run time) • Device configuration (e.g. SW options,
• Driver for system integration for exporting • Maximum indicators (min/max fixed I/O or multi I/O)
via Web server, e.g.: values)
EDS for EtherNet/IP • Totalizer values
Storage location Fixed on the user interface board in the Attachable to the user interface board in In the sensor plug in the transmitter
connection compartment the connection compartment neck part
Data backup
Automatic
• The most important device data (sensor and transmitter) are automatically saved in the
DAT modules
• If the transmitter or measuring device is replaced: once the T-DAT containing the
previous device data has been exchanged, the new measuring device is ready for
operation again immediately without any errors
• If the sensor is replaced: once the sensor has been replaced, new sensor data are
transferred from the S-DAT in the measuring device and the measuring device is ready
for operation again immediately without any errors
• If exchanging the electronics module (e.g. I/O electronics module): Once the electronics
module has been replaced, the software of the module is compared against the current
device firmware. The module software is upgraded or downgraded where necessary. The
electronics module is available for use immediately afterwards and no compatibility
problems occur.
Manual
Additional parameter data record (complete parameter settings) in the integrated device
memory HistoROM backup for:
• Data backup function
Backup and subsequent restoration of a device configuration in the device memory
HistoROM backup
• Data comparison function
Comparison of the current device configuration with the device configuration saved in
the device memory HistoROM backup
Data transfer
Manual
• Transfer of a device configuration to another device using the export function of the
specific operating tool, e.g. with FieldCare, DeviceCare or Web server: to duplicate the
configuration or to store in an archive (e.g. for backup purposes)
• Transmission of the drivers for system integration via Web server, e.g.:
EDS for EtherNet/IP
Endress+Hauser 235
Technical data Proline Promass F 300 EtherNet/IP
Event list
Automatic
• Chronological display of up to 20 event messages in the events list
• If the Extended HistoROM application package (order option) is enabled: up to 100
event messages are displayed in the events list along with a time stamp, plain text
description and remedial measures
• The events list can be exported and displayed via a variety of interfaces and operating
tools e.g. DeviceCare, FieldCare or Web server
Data logging
Manual
If the Extended HistoROM application package (order option) is enabled:
• Record up to 1 000 measured values via 1 to 4 channels
• User configurable recording interval
• Record up to 250 measured values via each of the 4 memory channels
• Export the measured value log via a variety of interfaces and operating tools e.g.
FieldCare, DeviceCare or web server
CE mark The measuring system is in conformity with the statutory requirements of the applicable
EU Directives. These are listed in the corresponding EU Declaration of Conformity along
with the standards applied.
Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.
C-Tick symbol The measuring system meets the EMC requirements of the "Australian Communications
and Media Authority (ACMA)".
Ex approval The devices are certified for use in hazardous areas and the relevant safety instructions are
provided in the separate "Safety Instructions" (XA) document. Reference is made to this
document on the nameplate.
Pharmaceutical • FDA
compatibility • USP Class VI
• TSE/BSE certificate of suitability
236 Endress+Hauser
Proline Promass F 300 EtherNet/IP Technical data
EtherNet/IP certification The measuring device is certified and registered by the ODVA (Open Device Vendor
Association). The measuring system meets all the requirements of the following
specifications:
• Certified in accordance with the ODVA Conformance Test
• EtherNet/IP Performance Test
• EtherNet/IP PlugFest compliance
• The device can also be operated with certified devices of other manufacturers
(interoperability)
Pressure Equipment • With the identification PED/G1/x (x = category) on the sensor nameplate,
Directive Endress+Hauser confirms conformity with the "Essential Safety Requirements" specified
in Appendix I of the Pressure Equipment Directive 2014/68/EC.
• Devices not bearing this marking (PED) are designed and manufactured according to
good engineering practice. They meet the requirements of Art. 4, Par. 3 of the Pressure
Equipment Directive 2014/68/EU. The range of application is indicated in tables 6 to 9
in Annex II of the Pressure Equipment Directive 2014/68/EC.
CRN approval
Some device versions have CRN approval. A CRN-approved process connection with a CSA
approval must be ordered for a CRN-approved device.
ISO 23277 AL2x (PT) ASME ASME NORSOK Measuring tube Process connection
ISO 10675-1 AL1 (RT, DR) B31.3 VIII Div.1 M-601
NFS Appx. 4+8
CF x PT RT
KK x PT RT
KP x PT RT
KR x VT, PT VT, RT
K1 x PT DR
K2 x PT DR
K3 x PT DR
Endress+Hauser 237
Technical data Proline Promass F 300 EtherNet/IP
ISO 23277 AL2x (PT) ASME ASME NORSOK Measuring tube Process connection
ISO 10675-1 AL1 (RT, DR) B31.3 VIII Div.1 M-601
NFS Appx. 4+8
K4 x VT, PT VT, DR
238 Endress+Hauser
Proline Promass F 300 EtherNet/IP Technical data
The application packages can be ordered with the device or subsequently from
Endress+Hauser. Detailed information on the order code in question is available from your
local Endress+Hauser sales center or on the product page of the Endress+Hauser website:
www.endress.com.
Detailed information on the application packages:
Special Documentation for the device → 240
Extended HistoROM Comprises extended functions concerning the event log and the activation of the
measured value memory.
Event log:
Memory volume is extended from 20 message entries (standard version) to up to
100 entries.
Data logging (line recorder):
• Memory capacity for up to 1000 measured values is activated.
• 250 measured values can be output via each of the 4 memory channels. The
recording interval can be defined and configured by the user.
• Measured value logs can be accessed via the local display or operating tool e.g.
FieldCare, DeviceCare or Web server.
Endress+Hauser 239
Technical data Proline Promass F 300 EtherNet/IP
Petroleum The most important parameters for the Oil & Gas Industry can be calculated and
displayed with this application package.
• Corrected volume flow and calculated reference density in accordance with the
"API Manual of Petroleum Measurement Standards, Chapter 11.1"
• Water content, based on density measurement
• Weighted mean of the density and temperature
16.14 Accessories
Overview of accessories available for order → 206
Technical Information
ATEX/IECEx Ex ec XA01439D
cCSAus XP XA01373D
240 Endress+Hauser
Proline Promass F 300 EtherNet/IP Technical data
cCSAus Ex d/ Ex de XA01372D
cCSAus Ex nA XA01507D
INMETRO Ex ec XA01470D
NEPSI Ex nA XA01471D
ATEX/IECEx Ex i XA01494D
ATEX/IECEx Ex ec XA01498D
cCSAus IS XA01499D
cCSAus Ex nA XA01513D
INMETRO Ex i XA01500D
INMETRO Ex ec XA01501D
NEPSI Ex i XA01502D
NEPSI Ex nA XA01503D
Special documentation
Radio approvals for WLAN interface for A309/A310 display module SD01793D
Petroleum SD02096D
Installation Instructions
Contents Comment
Installation instructions for spare part sets and accessories • Access the overview of all the available spare part sets via
W@M Device Viewer → 204
• Accessories available for order with Installation Instructions → 206
Endress+Hauser 241
Index Proline Promass F 300 EtherNet/IP
Index
0…9 D
3-A approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 236 Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . 11
Define access code . . . . . . . . . . . . . . . . . . . . . . 157, 158
A Degree of protection . . . . . . . . . . . . . . . . . . . . . 46, 224
Access authorization to parameters Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 226
Read access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61 Design fundamentals
Write access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61 Maximum measured error . . . . . . . . . . . . . . . . . 223
Access code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61 Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . 223
Incorrect input . . . . . . . . . . . . . . . . . . . . . . . . . . . 61 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 218 Device components . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Adapting the diagnostic behavior . . . . . . . . . . . . . . . 182 Device description files . . . . . . . . . . . . . . . . . . . . . . . 76
Additional certification . . . . . . . . . . . . . . . . . . . . . . 237 Device documentation
Ambient temperature Supplementary documentation . . . . . . . . . . . . . . . . 8
Influence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 221 Device locking, status . . . . . . . . . . . . . . . . . . . . . . . 160
Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 209 Device name
Application packages . . . . . . . . . . . . . . . . . . . . . . . . 238 Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Applicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 210 Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 236 Device repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204
Device revision . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
C Device type ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
C-Tick symbol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 236 DeviceCare . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Cable entries Device description file . . . . . . . . . . . . . . . . . . . . . . 76
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . 218 Diagnostic behavior
Cable entry Explanation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . 46 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178
CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11, 236 Diagnostic information
Certificates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 236 Communication interface . . . . . . . . . . . . . . . . . . 182
Check Design, description . . . . . . . . . . . . . . . . . . . 178, 181
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 DeviceCare . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180
Checklist FieldCare . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180
Post-connection check . . . . . . . . . . . . . . . . . . . . . 47 Light emitting diodes . . . . . . . . . . . . . . . . . . . . . 175
Post-installation check . . . . . . . . . . . . . . . . . . . . . 30 Onsite display . . . . . . . . . . . . . . . . . . . . . . . . . . 177
Cleaning Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183
Cleaning in place (CIP) . . . . . . . . . . . . . . . . . . . . 203 Remedial measures . . . . . . . . . . . . . . . . . . . . . . 183
Exterior cleaning . . . . . . . . . . . . . . . . . . . . . . . . 203 Web browser . . . . . . . . . . . . . . . . . . . . . . . . . . . 179
Interior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . 203 Diagnostic list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197
Sterilization in place (SIP) . . . . . . . . . . . . . . . . . . 203 Diagnostic message . . . . . . . . . . . . . . . . . . . . . . . . . 177
Climate class . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 224 Diagnostics
Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177
Advanced settings . . . . . . . . . . . . . . . . . . . . . . . 134 DIP switches
Configuring the measuring device . . . . . . . . . . . . . 94 see Write protection switch
Connecting cable . . . . . . . . . . . . . . . . . . . . . . . . . 31, 32 Direct access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Connecting the measuring device . . . . . . . . . . . . . . . . 35 Direct access code . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Connecting the signal cables . . . . . . . . . . . . . . . . . . . 35 Disabling write protection . . . . . . . . . . . . . . . . . . . . 157
Connecting the supply voltage cables . . . . . . . . . . . . . 35 Display
Connection see Onsite display
see Electrical connection Display and operating module DKX001 . . . . . . . . . . 232
Connection preparations . . . . . . . . . . . . . . . . . . . . . . 34 Display area
Connection tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 For operational display . . . . . . . . . . . . . . . . . . . . . 52
Context menu In the navigation view . . . . . . . . . . . . . . . . . . . . . 54
Calling up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 Display values
Closing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 For locking status . . . . . . . . . . . . . . . . . . . . . . . . 160
Explanation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 205
Current consumption . . . . . . . . . . . . . . . . . . . . . . . . 218 Document
Cyclic data transmission . . . . . . . . . . . . . . . . . . . . . . 77 Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
242 Endress+Hauser
Proline Promass F 300 EtherNet/IP Index
Endress+Hauser 243
Index Proline Promass F 300 EtherNet/IP
244 Endress+Hauser
Proline Promass F 300 EtherNet/IP Index
Endress+Hauser 245
Index Proline Promass F 300 EtherNet/IP
246 Endress+Hauser
Proline Promass F 300 EtherNet/IP Index
Write protection
Via access code . . . . . . . . . . . . . . . . . . . . . . . . . . 157
Via write protection switch . . . . . . . . . . . . . . . . . 158
Write protection switch . . . . . . . . . . . . . . . . . . . . . . 158
Endress+Hauser 247
www.addresses.endress.com