STANDARD OPERATING PROCEDURE IN
ARKRAY HGBA1C
HbA1c is a blood test that is used to help diagnose type 2 diabetes and monitor
blood glucose control in people who have diabetes. People with diabetes are advised to
have this test every 3 to 6 months.
2. ASSAY PRINCIPLE
The ADAMS A1c, HA-8180V is a fully automated Glycohemoglobin (HbA1c)
analyzer based on HPLC (High Performance Liquid Chromatography). Glucose in your
blood sticks to hemoglobin, a protein in your red blood cells. As your blood glucose
levels increase, more of your hemoglobin will be coated with glucose. An A1C test
measures the percentage of your red blood cells that have glucose-coated hemoglobin.
Specifically, the A1C test measures what percentage of hemoglobin proteins in your
blood are coated with sugar (glycated). Hemoglobin proteins in red blood cells transport
oxygen. The higher your A1C level is, the poorer your blood sugar control and the higher
your risk of diabetes complications.
3. CLINICAL SIGNIFICANCE
HbA1c values represents the average blood glucose level over the past 8-12 weeks
and is used for long-term glycemic control in diabetic individuals. Besides monitoring
JOBELLE C. AYALA, RMT KHRISTINE T. MORENO, RMT, MPH CARINA P. VILLAMAYOR, MD, FPSP Roy O. Arceo, MD, MHA, FPSO-HNS
Date: 06 September 2023 Date: 06 September 2023 Date Date:
Document No. Page: 2 of 2
Revision Date
Department: Pathology 06 September
2023
Section: Effective Date:
Revision No. 06September 2023
Policy Title:
STANDARD OPERATING PROCEDURE IN
ARKRAY HGBA1C
4. TEST INFORMATION
4.1 Running Day: Daily
4.2 Cut-off time: None
4.3 Result: After 2 hours
4.4 Relaying of Result: Results must be relayed once available.
4.5 Reference Range:
4.5.1 Normal: <5.5%
4.5.2 Pre-Diabetic: 5.6-6.4%
4.5.3 Diabetic: 6.5% (2 occassions)
4.6 Unit of Reporting: % percentage
6. SPECIMEN REQUIREMENTS
6.1 Sample Volume
6.1.1 Machine required volume per run: 3.5 uL EDTA whole blood
6.1.2 Volume required for submission: At least 1 ml EDTA whole
blood
JOBELLE C. AYALA, RMT KHRISTINE T. MORENO, RMT, MPH CARINA P. VILLAMAYOR, MD, FPSP Roy O. Arceo, MD, MHA, FPSO-HNS
Date: 06 September 2023 Date: 06 September 2023 Date Date:
Document No. Page: 3 of 2
Revision Date
Department: Pathology 06 September
2023
Section: Effective Date:
Revision No. 06September 2023
Policy Title:
STANDARD OPERATING PROCEDURE IN
ARKRAY HGBA1C
7. REAGENTS
JOBELLE C. AYALA, RMT KHRISTINE T. MORENO, RMT, MPH CARINA P. VILLAMAYOR, MD, FPSP Roy O. Arceo, MD, MHA, FPSO-HNS
Date: 06 September 2023 Date: 06 September 2023 Date Date:
Document No. Page: 4 of 2
Revision Date
Department: Pathology 06 September
2023
Section: Effective Date:
Revision No. 06September 2023
Policy Title:
STANDARD OPERATING PROCEDURE IN
ARKRAY HGBA1C
7.1.5 Label the reagent box with the date opened and user’s
initials.
7.1.6 If stored according to the recommended conditions, all
components are stable until the expiration date indicated on
the label.
8. SAMPLE RACKS
8.1 Normal rack - normal measurements, HbA1c reproducibility test (whole blood
sample) *normally use this rack to measure samples.
Object: whole blood sample
Adapter: adapters are not attached at the time of shipment. Attach
adapters if needed according to the type of sample containers to be used.
Sample container: sample tube
8.2 Start Rack - normal measurement, HbA1c control measurement, HbA1c control
calibration (if sample rack transportation is set to “loop transportation). Set this
rack as the first rack to be measured on the loading side of the sampler.
Measurement operation will stop automatically after all samples have been
measured and start rack has returned to its initial position. Example: For normal
measurements, set the start rack in the first position, followed by normal racks in
the 2nd to 10th positions.
Object: whole blood sample
Adapter: spinning unit adapter (transparent)
Sample container: sample tube
8.3 Whole blood pair rack - normal measurement. Use this rack to measure whole
blood sample cups. IDs are read from the barcode labels attached to sample
tubes.
Object: whole blood samples
Adapter: odd-numbered ports - spinning unit adapter (transparent)
JOBELLE C. AYALA, RMT KHRISTINE T. MORENO, RMT, MPH CARINA P. VILLAMAYOR, MD, FPSP Roy O. Arceo, MD, MHA, FPSO-HNS
Date: 06 September 2023 Date: 06 September 2023 Date Date:
Document No. Page: 5 of 2
Revision Date
Department: Pathology 06 September
2023
Section: Effective Date:
Revision No. 06September 2023
Policy Title:
STANDARD OPERATING PROCEDURE IN
ARKRAY HGBA1C
JOBELLE C. AYALA, RMT KHRISTINE T. MORENO, RMT, MPH CARINA P. VILLAMAYOR, MD, FPSP Roy O. Arceo, MD, MHA, FPSO-HNS
Date: 06 September 2023 Date: 06 September 2023 Date Date:
Document No. Page: 6 of 2
Revision Date
Department: Pathology 06 September
2023
Section: Effective Date:
Revision No. 06September 2023
Policy Title:
STANDARD OPERATING PROCEDURE IN
ARKRAY HGBA1C
JOBELLE C. AYALA, RMT KHRISTINE T. MORENO, RMT, MPH CARINA P. VILLAMAYOR, MD, FPSP Roy O. Arceo, MD, MHA, FPSO-HNS
Date: 06 September 2023 Date: 06 September 2023 Date Date:
Document No. Page: 7 of 2
Revision Date
Department: Pathology 06 September
2023
Section: Effective Date:
Revision No. 06September 2023
Policy Title:
STANDARD OPERATING PROCEDURE IN
ARKRAY HGBA1C
JOBELLE C. AYALA, RMT KHRISTINE T. MORENO, RMT, MPH CARINA P. VILLAMAYOR, MD, FPSP Roy O. Arceo, MD, MHA, FPSO-HNS
Date: 06 September 2023 Date: 06 September 2023 Date Date:
Document No. Page: 8 of 2
Revision Date
Department: Pathology 06 September
2023
Section: Effective Date:
Revision No. 06September 2023
Policy Title:
STANDARD OPERATING PROCEDURE IN
ARKRAY HGBA1C
10.3.1 Attach the calibration information barcode labels to each of two empty
sample tubes
10.3.2 Set the sample tubes in port 1 to 3 of the calibration rack
-use any of the ports 1 to 3
-set barcodes in any order
10.4 Load the calibration rack onto the sampler
10.4.1 Load the calibration rack onto the sampler
10.5 Check the measurement mode (Variant or Fast)
10.5.1 On the standby screen, check the [measurement mode] setting.
10.6 Set the calibration information. [For the numeric button operation]
10.6.1 On the standby screen, press Cal
10.6.2 Set the HbA1c calibration conditions.
10.6.3 Press OK
10.6.4 Below <CODE>, check the current reagent code.
10.6.5 To change the reagent code, enter the 10-digit code that is written on
the standard value list.
JOBELLE C. AYALA, RMT KHRISTINE T. MORENO, RMT, MPH CARINA P. VILLAMAYOR, MD, FPSP Roy O. Arceo, MD, MHA, FPSO-HNS
Date: 06 September 2023 Date: 06 September 2023 Date Date:
Document No. Page: 9 of 2
Revision Date
Department: Pathology 06 September
2023
Section: Effective Date:
Revision No. 06September 2023
Policy Title:
STANDARD OPERATING PROCEDURE IN
ARKRAY HGBA1C
JOBELLE C. AYALA, RMT KHRISTINE T. MORENO, RMT, MPH CARINA P. VILLAMAYOR, MD, FPSP Roy O. Arceo, MD, MHA, FPSO-HNS
Date: 06 September 2023 Date: 06 September 2023 Date Date:
Document No. Page: 10 of 2
Revision Date
Department: Pathology 06 September
2023
Section: Effective Date:
Revision No. 06September 2023
Policy Title:
STANDARD OPERATING PROCEDURE IN
ARKRAY HGBA1C
JOBELLE C. AYALA, RMT KHRISTINE T. MORENO, RMT, MPH CARINA P. VILLAMAYOR, MD, FPSP Roy O. Arceo, MD, MHA, FPSO-HNS
Date: 06 September 2023 Date: 06 September 2023 Date Date:
Document No. Page: 11 of 2
Revision Date
Department: Pathology 06 September
2023
Section: Effective Date:
Revision No. 06September 2023
Policy Title:
STANDARD OPERATING PROCEDURE IN
ARKRAY HGBA1C
JOBELLE C. AYALA, RMT KHRISTINE T. MORENO, RMT, MPH CARINA P. VILLAMAYOR, MD, FPSP Roy O. Arceo, MD, MHA, FPSO-HNS
Date: 06 September 2023 Date: 06 September 2023 Date Date: