DIGITAL MULTIMETER
OPERATOR’S
INSTRUCTION MANUAL
GENERAL INSTRUCTIONS
This instrument complies with IEC 1010-1 (61010-1@IEC: 2001), CAT. II 1000V and CAT. III 600V
overvoltage standards. See Specifications.
To get the best service from this instrument, read carefully this user's manual and respect the detailed
safety precautions.
International symbols used on the Meter and in this manual are explained in chapter 1.1.3
1.1.1 Preliminary
* Measurement category III is for measurements performed in the building installation.
NOTE: Examples are measurements on distribution boards, circuit-breakers, wiring, including cables,
bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and equipment for industrial
use and some other equipment, for example, stationary motors with permanent connection to the fixed
installation.
* Measurement category II is for measurements performed on circuits directly connected to the low
voltage installation.
NOTE: Examples are measurements on household appliances, portable tools and similar equipment.
* Measurement category I is for measurements performed on circuits not directly connected to MAINS.
NOTE: Examples are measurements on circuits not derived from MAINS, and specially protected
(internal) MAINS derived circuits. In the latter case, transient stresses
are variable; for that reason, requires that the transient withstand capability of the equipment is made
known to the user.
* When using this Multimeter, the user must observe all normal safety rules concerning:
― protection against the dangers of electric current.
― protection of the Multimeter against misuse.
* For your own safety, only use the test probes supplied with the instrument. Before use, check that they
are in good condition.
1.1.3 Symbols:
Symbols used in this manual and on the instrument:
2.3 Keypad
See Table 2 indicated for information about the keypad operations.
Table 2. Keypad
Key Function Operation performed
ON/O Any switch turn the meter on or
FF position off
HOLD Any switch Press HOLD to enter
position and exit the Data
Hold mode.
LIGH Any switch Press LIGHT to turn
T position the backlight on.
After about 5
seconds, the
backlight is auto-off.
2.4 Terminals
See Table 4 indicated for information about the terminals.
Table 4. Terminals
Termi Description
nal
Return terminal for all measurements.
( Receiving the black test lead or the
COM
“com” plug of the special multi-function
socket)
Input for voltage, resistance, diode and
V( continuity measurements. (Receiving
the red test lead)
Input for capacitance, hFE and
mA 0.001mA to 200mA current
hF measurements. ( Receiving the red
E test lead or the “+” plug of the special
multi-function socket)
Input for 200mA to 10A current
10A measurements. (Receiving the red test
lead)
2.5 Accessories
Delivered with the multimeter:
( User's manual One piece
( Test leads One piece
( Carry case One piece
( Special Multi-function One piece
socket
. FUNCTION DESCRIPTION
3.1 General Functions
3.1.1 DATA HOLD mode
Data Hold mode makes the meter stop updating the display. Data Hold function can be cancelled by
changing the measurement mode, or push HOLD key again.
To enter and exit the Data Hold mode:
1. Press HOLD key. Fixes the display on the current value, is displayed.
2. A second short press returns the meter to normal mode.
3.1.2 Battery Saver
Turn on the meter. And then The Meter will be turned off automatic after approx. 30 minutes.
NOTE:
( Unstable display may occur especially at DC200mV and AC2V ranges, even though you do not put test
leads into input terminals, in this case, if an erroneous reading is suspected, short the V terminal and the
COM terminal, and make sure the zero display.
To measure resistance:
1. Set the rotary switch to proper range.
2. Connect the black and red test leads to the COM and ( terminals respectively.
3. Connect the test leads to the circuit being measured and read the displayed value.
Some tips for measuring resistance:
( The measured value of a resistor in a circuit is often different from the resistor's rated value. This is
because the Meter's test current flows through all possible paths between the probe tips.
( In order to ensure the best accuracy in measurement of low resistance, short the test leads before
measurement and memory the test probe resistance in mind. This necessary to subtract for the
resistance of the test leads.
( The resistance function can produce enough voltage to forward-bias silicon diode or transistor junctions,
causing them to conduct. To avoid this, do not use the 40M( range for in-circuit resistance
measurements.
( On 20M( and 200M( ranges, the meter may take a few seconds to stabilize reading. This is normal for
high resistance measuring.
( On 200M( range, the display is approx. 10 digits when test leads are shorted. These 10 digits have to be
subtracted from measuring results. For example, when measuring 100 M( resistance, the reading will be
101.0 and the correct measuring result should be 101.0-1.0=100.0 M(.
( When the input is not connected, i.e. at open circuit, the figure "1" will be displayed for the overrange
condition.
Use the diode test to check diodes, and other semi- conductor devices. The diode test sends a current
through the semiconductor junction, and then measures the voltage drop across the junction; a good
silicon junction drops between 0.5V and 0.8V.
To test a diode out of a circuit:
1. Set the rotary switch to range.
2. Connect the black and red test leads to the COM and terminals respectively.
3. For forward-bias readings on any semiconductor component, place the red test lead on the
component's anode and place the black test lead on the component's cathode.
4. The meter will show the approx. forward voltage of the diode. If the test lead connection is reversed,
only figure "1" displayed.
In a circuit, a good diode should still produce a forward bias reading of 0.5V to 0.8V; however, the
reverse-bias reading can vary depending on the resistance of other pathways between the probe tips.
To measure capacitance:
1. Set the rotary switch to proper range.
2. Connect the black and red test leads to the COM and terminals respectively (or you can measure
the capacitance by using the special Multi-Function Socket).
3. Connect the test leads to the capacitor being measured and read the displayed value.
Some tips for measuring capacitance:
( The meter may take a few seconds to stabilize reading. This is normal for high capacitance measuring.
( To improve the accuracy of measurements less than 20nF, subtract the residual capacitance of the
Meter and leads.
3.2.6 Transistor measurement
To avoid electrical shock and/or damage
to the instrument, do not apply more than
250Vdc or 250Vac rms between the hFE
terminal and the COM terminal.
4.2.2 AC Voltage
Range Resoluti Accuracy
on
2V 1mV
20V 10mV ±(0.8% of rdg +3 digits)
200V 100mV
750V 1V ±(1.2% of rdg +3 digits)
Input impedance: 10M(
Max. input voltage: 250Vdc or ac rms for 200mV range and 1000Vdc or 750V ac rms for other ranges,
Frequency Range: 40Hz-200Hz for 750V range, 40Hz-400Hz for other ranges.
Response: Average, calibrated in rms of sine wave
4.2.3 Resistance
Rang Resolut Accuracy
e ion
200( 0.1( ±(0.8% of rdg+3 digits)
2k( 1(
20k( 10(
±(0.8% of rdg+1 digit)
200k( 100(
2M( 1k(
20M( 10k( ±(1.0% of rdg+2 digits)
200M( 0.1M( ±[5.0% of (rdg-10digits) +10
digits]
Overload protection: 380V dc or 380Vac rms.
Open Circuit Voltage: Less than 700mV.
4.2.4 Diode
Rang Resoluti Function
e on
Display read approx.
1mV
forward voltage of diode
Forward DC Current: approx. 1mA
Reversed DC Voltage: approx. 2.8V
Overload protection: 380Vdc or 380Vac rms.
4.2.7 Capacitance
Range Resolutio Accuracy
n
2nF 1pF
20nF 10pF
200nF 0.1nF ±(4.0% of rdg+3 digits)
2(F 1nF
20(F 10nF
Overload protection: Resettable fuse (F200mA/250V).
4.2.8 DC Current
Resolutio
Range Accuracy
n
2mA 1(A ±(0.8% of rdg+1 digit)
20mA 10(A
±(1.5% of rdg+1 digit)
200mA 0.1mA
10A 10mA ±(2.0% of rdg+5 digits)
4.2.9 AC Current
Resolutio
Range Accuracy
n
2mA 1(A ±(1.0% of rdg+3 digits)
200mA 0.1mA ±(1.8% of rdg+3 digits)
10A 10mA ±(3.0% of rdg+7 digits)
Overload protection: Resettable fuse (F200mA/250V). 10A range unfused.
Max. input current: 200mA dc or 200mA ac rms for mA range, 10A dc or 10A ac rms for 10A ranges.
For measurements>5A, 4 minutes maximum ON to measure 10 minutes OFF; Above 10A unspecified.
Frequency Range: 40Hz-400Hz
Response: Average, calibrated in rms of sine wave
5. MAINTENANCE
Do not attempt to repair or service your Meter unless you are qualified to do so and have the relevant
calibration, performance test, and service information.
Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents.
Dirt or moisture in the terminals can affect readings.
To clean the terminals:
( Turn the meter off and remove all test leads.
( Shake out any dirt that may be in the terminals.
( Soak a new swab with a cleaning and oiling agent (such as WD-40).
( Work the swab around in each terminal. The oiling agent insulates the terminals from moisture-related
contamination.
9V Battery
CAUTION
Using this appliance in an environment with a strong radiated radio-frequency electromagnetic field
(approx. 3V/m), may influence its measuring accuracy. The measuring result can be strongly deviating
from the actual value.