Instalación de ProX
Instalación de ProX
EN
installation manual
10029964
The manufacturer, assembler and importer are responsible for the safety, reliability and
performance of the unit only if:
- installation, calibration, modification and repairs are carried out by qualified
authorised personnel
- electrical installations are carried out according to the appropriate requirements such
as IEC 60364
- equipment is used according to the operating instructions.
Planmeca pursues a policy of continual product development. Although every effort
is made to produce up-to-date product documentation this publication should not be
regarded as an infallible guide to current specifications. We reserve the right to make
changes without prior notice.
COPYRIGHT PLANMECA
Publication number 10029964 Revision 17
Released 19 December 2022
Table of contents
1 Introduction................................................................................................................................................. 1
2 Symbols on product labels..........................................................................................................................3
3 Pre-installation requirements...................................................................................................................... 4
3.1 Environmental requirements...........................................................................................................4
3.1.1 Wall / ceiling construction...............................................................................................4
3.1.2 Floor construction...........................................................................................................4
3.1.3 Temperature & humidity.................................................................................................4
3.2 Electrical requirements................................................................................................................... 4
3.2.1 Mains voltage................................................................................................................. 4
3.2.2 Mains frequency.............................................................................................................4
3.2.3 Internal mains fuse ratings............................................................................................. 5
3.2.4 External mains fuse recommendation............................................................................5
3.2.5 Grounding...................................................................................................................... 5
3.2.6 Power consumption........................................................................................................5
4 Associated documentation..........................................................................................................................6
5 Installation alternatives............................................................................................................................... 7
5.1 Mounting alternatives..................................................................................................................... 7
5.1.1 Standard wall mounting in treatment room.................................................................... 8
5.1.2 Wall mounting with remote control in treatment room....................................................9
5.1.3 Ceiling mounting in treatment room............................................................................. 10
5.1.4 X-ray room with generator box/control panel outside the room................................... 11
5.1.5 Other installation alternatives.......................................................................................12
5.2 Control panel assembling alternatives..........................................................................................13
5.2.1 Standard assembly...................................................................................................... 13
5.2.2 Remote assembly........................................................................................................ 14
5.2.3 Fixed control panel (allows remote exposure switch assembly).................................. 14
6 Control panel holder................................................................................................................................. 15
6.1 Attaching control panel holder to generator housing.................................................................... 15
6.2 Attaching control panel holder to side of generator adapter plate................................................ 16
7 Wall installation.........................................................................................................................................18
7.1 Attaching adapter plate to wall..................................................................................................... 18
7.1.1 Concealed wiring with metallic armor...........................................................................22
7.2 Installing extension arm................................................................................................................23
7.3 Attaching bracket arm...................................................................................................................29
7.4 Installing generator assembly.......................................................................................................35
7.5 Connecting cables........................................................................................................................ 37
7.5.1 Arm cable..................................................................................................................... 38
7.5.2 Mains cable (concealed wiring)....................................................................................39
7.5.3 Mounting generator housing........................................................................................ 40
7.5.4 Connecting control panel cable....................................................................................41
8 Ceiling installation.....................................................................................................................................42
8.1 Concrete ceiling............................................................................................................................42
8.2 Wooden ceiling............................................................................................................................. 44
8.3 Adjustable ceiling adapter............................................................................................................ 53
1 Introduction
The Planmeca ProX X-ray unit produces intraoral X-ray images for the
diagnosis of teeth and adjacent structures. The unit is allowed to be used
only under supervision of a dental/ health care professional.
This manual contains all the information required to install and set up
the Planmeca ProX X-ray unit. Please read this manual carefully before
installing the X-ray unit.
NOTE
This manual is valid for software revisions 4.00 or later.
NOTE
The Planmeca ProX intraoral X-ray unit is allowed to be used only under
supervision of a dental/health care professional.
The Planmeca ProX intraoral X-ray unit fulfils the requirements of Directive
93/42/EEC.
WARNING
Do not connect a multiple portable socket outlet (MPSO) or extension cord
to the system.
WARNING
To avoid the risk of electric shock, the x-ray unit must be connected to a
protectively grounded power line.
WARNING
Failure to install the Planmeca ProX X-ray unit in an approved location may
be dangerous to both patient and operator.
WARNING
No modification of this equipment is allowed.
Do not modify this equipment without authorisation of the manufacturer.
If this equipment is modified, appropriate inspection and testing must be
conducted to ensure continued safe use of equipment.
CAUTION
While servicing and installing Planmeca ProX X-ray unit there must be
means to isolate its circuits electrically from the supply mains on all poles
simultaneously. Ensure that the mains plug is disconnected from the supply
mains or shall be capable of being locked in the off position.
CAUTION
The unit contains hazardous voltages. While servicing internal parts, always
turn off externally the power to the unit, and wait for 2 minutes before
touching any electrical parts.
Always replace the fuses with ones of the same type and rating. Otherwise
patient, operator or equipment safety cannot be guaranteed.
The circuit boards can be damaged due to static discharges and require
careful handling.
CAUTION
It is very important that the room in which the Planmeca ProX X-ray unit is
installed and the position from which the equipment is operated are correctly
shielded against radiation.
Since radiation safety requirements vary from country to country and state
to state, it is the responsibility of the installer to ensure that all local and
national safety regulations are met.
NOTE
In case the Planmeca ProSensor digital intraoral sensor system is going to
be installed to the Planmeca ProX X-ray unit, use the ProSensor digital intra
X-ray system installation manual in conjunction with this manual.
Medical Device
3 Pre-installation requirements
NOTE
If the X-ray unit has been stored at temperatures below +10°C for more
than a few hours, time must be allowed for the X-ray unit to reach the room
temperature before connecting it to the mains voltage.
NOTE
Note that the ratings of the internal (as well as external) mains fuses are
depending on the mains voltage setting.
CAUTION
Never connect the X-ray unit to the mains without first checking the voltage
setting. Incorrect voltage setting can cause damage to the X-ray electronics.
3.2.5 Grounding
The X-ray unit must always be connected to a grounded outlet to fulfill the
safety directives stated.
4 Associated documentation
The Planmeca ProX X-ray unit is supplied with the following manuals:
• Planmeca ProX user’s manual
• Planmeca ProX installation manual
• Planmeca ProX technical manual
These manuals are intended to be used in conjunction with the
documentation for the Planmeca Romexis imaging software. The imaging
software package contains the following manuals:
• Planmeca Romexis user’s manual
• Planmeca Romexis technical manual
5 Installation alternatives
5.1 Mounting alternatives
NOTE
The X-ray unit can be installed in several ways. Read these installation
instructions carefully, because some installation accessories are optional,
and are not included in the standard delivery. Make sure that required
accessories are available before starting the installation.
NOTE
The following cables are used in some installations described in this section:
EXTENSION CABLE:
• 1 pcs. protective earth min. 1.0 mm2 AWG 12
• 4 pcs. signal wires min. 1.0 mm2 AWG 18
• 4 pcs. signal wires min. 0.5 mm2 AWG 20 rated 300 V, flexible
TELEPHONE CABLE:
• Ordinary 6-pole telephone cable AWG 24
1 Control panel
2 Dental unit
3 Generator box
NOTE
Only 391 mm and 518 mm extension arms can be used for the ceiling
installation.
1 Generator box
2 Control panel
3 Extension cable, max. 12 m
4 Dental unit
5.1.4 X-ray room with generator box/control panel outside the room
1 Generator box
2 Extension cable, max. 12 m
1 Generator box
2 Control panel
3 Telephone cable
Attach the control panel holder to the generator housing with the two PT
3x22 rst WN1451 screws.
Attach the control panel holder to its assembly plate with two PT 3x22 rst
WN1451 screws.
Attach the holder assembly to the adapter plate with two M4x8 DIN 912
screws using a 3 mm Allen key.
7 Wall installation
7.1 Attaching adapter plate to wall
Attach the adapter plate to the wall according to the following instructions.
NOTE
The minimum clearance widths for the installation vary according to the
length of the X-ray unit's extension arm.
• If the wall is made of wood or plaster, use the ø 8x80 DIN 571 lag
screws
Do not use the expansion anchors with wooden or plaster wall.
Use the adapter plate as a template and mark the positions where the holes
for the attaching screws will be drilled. Use a spirit level to ensure that the
adapter plate will be level.
To correctly prepare a concrete or brick wall for the adapter plate installation:
1. Drill a ø10 mm (0.4 in.) hole, 32 - 35 mm (1.25 in.) in depth.
Remove dust from the drill hole with a vacuum cleaner or compressed
air.
3. Hammer the wedge into the hole with a ø6 mm drift and hammer.
The X-ray unit must be positioned in accordance with the information given
in the installation pattern supplied with the unit.
The X-ray unit should be positioned within the reach of the power supply
cable 3 m (118 in.). In the case that concealed wiring is used, the cable
outlet should be positioned to the upper part of the adapter plate (grey area
on figure below).
If the wall is made of concrete or brick, use the four side attachment
openings (white arrows in the figure below).
If the wall is made of wood or plaster, attach the plate to the wall with as
many screws as possible to make the attachment as strong as possible.
You can also use the three middle attachment openings (black arrows in the
figure below) if you attach the plate to the wall supporting rail.
The metallic armor is secured with the attachment plate to the adapter plate
(see figure above). Attach the plate to its position with two M4x8 DIN 912
screws using a 3 mm Allen key. Attach the metallic armor to the plate with
a RACO connector: RACO 3/8” Straight Squeeze 2191 for 3/8” armor and
RACO 1/2” Straight Squeeze 2192 for 1/2” armor.
Strip the mains cable outer jacket and the jackets of cable leads as shown
on the figure below.
NOTE
Before attaching the extension arm, add a small amount of grease to the
extension arm bearings to enable smooth arm movement.
2. Route the grounding cable (2) through the wall adapter bearing.
3. Mount the arm into position by pushing the arm shaft (3) into the wall
adapter bearing.
4. Use a spirit level to check that the arm is horizontal when set to the left
and the right. Adjust the arm position, if needed.
1. Unlock the arm shaft by loosening the three arm shaft screws (black
arrows, figure abowe).
2. Adjust the arm position with the three adjustment screws (white
arrows).
3. Lock the arm shaft by tightening the arm shaft screws.
5. Check the arm alignment using a spirit level, moving the arm from
left position to right as in the image below. Take careful note of the
alignment.
In the following image, the balance is incorrect even though the right-
hand alignment is horizontal.
6. Adjust the arm so that the alignment on the left offsets the alignment
on the right. If the spirit level shows that the arm is off-horizontal in
one direction, balance the arm so that it is off-horizontal in the opposite
direction in its second position. See the image below.
7. Adjust the arm in its respective positions by using the screws shown in
the image below.
8. Lock the arm shaft with the three attachment screws as described
above.
10. The arm position is adjusted at the factory. In case the position needs to
be adjusted, turn the adjustment screws located on the both sides of the
bearing.
CAUTION
Care must be taken when assembling the arm for not to damage the arm
cable.
Assemble the bracket arm to the extension arm by pushing the bracket arm
shaft into the housing.
Attach the arm grounding lead to the bottom of the extension arm.
NOTE
When attaching the grounding cable, position the lock washer between the
frame and the cable connector.
NOTE
It is recommended to use cable installation tool (part number 10033043) to
sheath the cable heads and better enable their routing through the extension
arm
1 Cable tie
2 Cable installation tool
Attach the cover plate grounding lead to the plate and attach the cover plate
to the extension arm.
Adjust the arm rotational friction with the adjustment screw (1). Tightening
the screw increases the friction.
NOTE
After the bracket arm is installed at the end of the extension arm, its weight
can pull the extension arm downwards. After the bracket arm is installed,
check the horizontal balance of the extension arm again.
There are four M4x8 ULS screws attached to the adapter plate at the factory.
Loosen these screws a couple of turns with a 2.5 mm Allen key. Slip the
generator assembly under the heads of the screws.
Secure the generator assembly into position by tightening the four screws
through the openings on the generator PCB.
Connect the grounding lead coming from the extension arm to the grounding
point located on the left side of the generator assembly.
Secure the grounding lead to the upper edge of the generator assembly with
a cable tie.
WARNING
Ensure that the power supply is switched off before connecting the cables.
NOTE
The figure above shows a concealed wiring setup with a metallic armour
attachment plate (1). The mains cable without metallic armor is routed and
connected in the same way.
NOTE
If the extension arm is not attached to the generator box, attach the
generator cover plug to the generator housing with one attachment screw
using a T10 Torx key.
8 Ceiling installation
Two versions of the ceiling columns are available. A version that can be
used in rooms up to 2460 mm (97”) in height, and a version that can be used
in rooms up to 3080mm (121") in height. An adjustable ceiling adapter can
be used, if needed. Refer to section "Adjustable ceiling adapter" on page
53.
Attach the plate (1) to the ceiling with four M8x35 DIN 912 screws and
ø8.4/17 DIN 125 washers. Use the adjustment nuts and the adjustment
plates with the screws at each corner of the adapter plate.
NOTE
The extension cable and Planmeca ProSensor data cable must be lead so
that the cables reach the lower edge of the ceiling arm opening.
Route the cables through the arm adapter. Attach the arm adapter to the
adapter plate with four attachment screws. Note, that the grounding lead
attachment hole (1) must be in the same side as the cable opening. Even if
concealed wiring is used, the grounding lead is routed through this opening
and attached to the arm adapter.
Shorten the ceiling column if needed. You have to drill ø10 mm hole to the
shortened arm. Drill the opening to the opposite side of the column than the
cable opening (3 in the figure below) and 131 mm distance from the new
upper edge.
Attach the mounting rings (1) to the arm (2). Do not tighten the screws yet.
NOTE
If the optional adapter cover (1) is used, attach it to its position before
attaching the ceiling column (2).
Use the adapter cover if the ceiling does not cover the adapter plate. Set the
opening away from the patient chair (3).
Move the lower mounting ring to the position shown on the figure below
(measures in mm). Attach the ceiling column to the adapter with the lower
mounting ring as follows. Attach the M8 x 16 attachment screw to the hole
located on the arm. Tighten three mounting ring securing screws firmly.
Move the upper mounting ring to the position shown on the figure above.
Tighten the upper mounting rings’ screws firmly.
With a spirit level make sure that the ceiling column is vertical. Adjust its
position with the adjustment nuts (1), if needed.
Make sure that you have a extension arm to which the bracket arm is
attached upwards (i.e. joint upwards). Place the extension arm to its position
and attach it with the outer, upper attachment screw M5x10 DIN 7991 (1 in
the figure below). Do not tighten the screw firmly yet.
Attach the inner, lower securing screw M4x8 DIN 914 to its position (2 in the
figure below). You can reach the inner screw hole through the outer screw
hole. Tighten the screw firmly.
Attach the outer, lower attachment screw M5x10 DIN 7991 (3 in the figure
below). Do not tighten the screw totally. Tighten the outer, upper attachment
screw firmly (4 in the figure below). Tighten the outer, lower attachment
screw.
The extension arm position is adjusted slightly upwards with the inner, lower
and outer, upper attachment screws.
Install the X-ray unit to the extension arm as described in section "Attaching
bracket arm" on page 29. Tape the arm cable and ProSensor data cables
so that the connectors are consecutively. Attach a draw cord to the cables.
Route the cables through extension arm and ceiling column to the ceiling
column opening.
Attach the grounding lead coming from the opening cover assembly to the
grounding point located inside the ceiling column using a screw, washer and
serrated washer.
Connect the extension and arm cables to the extension cable connector
PCB according to the figure below.
1 Generator PCB
2 Extension cable max. 12 m
3 Extension cable connectors
4 Arm cable (std.)
5 Ferrite
Attach the grounding leads to the cover assembly using screws, washers
and serrated washers.
Connect the ProSensor data cable (1 in the figure below) to the connector
located on the cover assembly.
Bundle the wires as shown on the figure below (white arrows).
The X-ray unit ceiling installation wiring is shown on the figure below.
1 Generator
2 Ethernet cable for ProSensor (for more information, see the
Planmeca ProSensor installation manual)
Attach the cover to its position.
Adjust the extension arm rotating friction with an adjustment screw located in
the ceiling column, if needed.
Attach the standard adapter plate to the adjustable ceiling adapter with four
M8x30 DIN 912 screws and ø8.4/17 DIN 125 washers. Use the adjustment
nuts and the adjustment plates with the screws at each corner of the adapter
plate. Continue the installation according to the instructions given in section
"Concrete ceiling" on page 42 or section "Wooden ceiling" on page 44.
1. The height of the adjustable ceiling adapter is 2. Upper part of the adjustable adapter
470 - 815 mm (18.5 - 32.1 in.)
3. Lower plate of the adjustable adapter 4. Adapter plate
NOTE
The attachment surface must fulfill requirements given in section
"Environmental requirements" on page 4.
Steps
1. Select the installation height according to the instructions given in Single
stud installation template (publication number 10035371).
1.a. Use the installation template or the single stud frame (1) as a
template and use a spirit level to make sure that the frame is
vertical (2).
1.b. Mark the positions (3) where the holes for the attaching screws will
be drilled.
2. Attach the single stud frame to the wall.
• If the wall is made of concrete or brick, drill ø10mm (0.4 in.), 32 - 35
mm (1.25 in.) in depth, holes and place the expansion anchors into
them. Tighten the anchors with hammer and wedge.
• If the wall is made of wood, drill ø5 mm (0.2 in.), 75 mm (3 in.) in
depth, holes for the attachment screws.
2.a. Route the cables through the opening on the frame (2).
2.b. The single stud frame (1) must be attached to the wall with six
screws.
• If the wall is made of concrete or brick, use the M8x18 DIN
912 screws (3), ø8.4 DIN 125 washers (4) and the expansion
anchors (5).
• If the wall is made of wood or plaster, use the ø8x80 DIN 571
lag screws (6) and ø8.4 DIN 125 washers (3).
Do not use the expansion anchors with wooden wall.
• Make sure that the extension cable extends 280 mm (11 in.) out from
the wall (1).
• If the interconnection cable is lead from the wall, make sure that it
extends 250 mm (10 in.) out from the wall (2).
4.a. Attach the shield (1) to the frame with two M4x5 DIN 7984 screws
(2) using a 2.5 mm Allen key.
4.b. Route the cable(s) (3) through the shield opening (4).
5. Secure the cable(s) coming from the wall with cable ties (1) to the frame.
Turn the cable tie lockings sideways.
6. Attach the extension PCB assembly (1) to the frame with three M4x5
DIN 7984 screws (2) using a 2.5 mm Allen key.
7.a. Secure the extension cable to the extension PCB assembly with a
cable tie (1).
7.b. Connect the extension cable according to the figure below (2).
7.c. Attach the extension cable grounding lead to the extension PCB
attachment plate with M4x5 DIN 7984 screw and serrated washer
(3).
7.d. Bundle the primary and secondary wires with cable ties (4).
NOTE
There are four grounding points on the extension PCB attachment plate,
one for each grounding lead: cover plate, extension cable, arm and arm
cable.
8.a. Attach the cover plate (1) to the frame with two M4x5 DIN 7984
screws (2) using a 2.5 mm Allen key.
8.b. Route the grounding lead (3) down to the extension PCB
attachment plate and attach the grounding lead to the plate with
one M4x5 DIN 7984 screw and serrated washer (4).
9. Attach three set screws to the arm adapter using a 4 mm Allen key (1, 2
and 3).
10. Attach the arm adapter to the single stud frame with two M8x70 DIN 912
screws (1) and one M8x25 DIN 912 screw (2) using a 4 mm Allen key.
11.a. Route the arm and ProSensor cables through the arm adapter.
11.b. Mount the extension arm into position and use a spirit level (1) to
check that the arm is horizontal in three directions (2, 3 and 4).
11.c. Adjust the arm adapter to vertical position with three set screws
(5), if needed.
For more information on balancing the extension arm, see section
"Installing extension arm" on page 23.
12. Attach the rotation limit spacer and M8x10 DIN 912 screw to its position
using a 4 mm Allen key (1).
You can change the bracket arm movement area with two securing
screws (2 and 3). Loosening the screws will narrow the movement area.
Adjust the arm rotational friction with the adjustment screw (4), tightening
the screw increases the friction.
13.a. Connect the arm cable to the extension PCB (1 and 2).
13.b. Attach the arm cable grounding lead with one M4x5 DIN 7984
screw and serrated washer (3).
13.c. Attach the arm grounding lead (4) to the extension PCB plate with
one M4x5 DIN 7984 screw and serrated washer.
13.d. The different ProSensor data cable (5) and interconnection cable
connection possibilities (6, 7 and 8), are shown on below.
Interconnection cable is routed directly to the computer
NOTE
You can also attach the control panel to the wall, refer to section
"Remote control panel installation" on page 103.
15. Push the excess cables into the extension arm and attach the end cover
to its position.
10 Floor installation
NOTE
The concrete or stone floor must be at least 100 mm (4 in.) thick.
1. Use the floor stand attachment plate as a template and mark the
positions of the attachment holes gently using a 10 mm size drill.
2. Use the M8x60 DIN 912 screws (1 in the figure below), washers (2 in the
figure below) and PFG ES 8-50 anchors (3 in the figure below) supplied
with the floor stand. Drill 14 mm (0.6 in.) holes until the required depth
of 55 mm (2.2 in.) (4 in the figure below). Clean the dust caused by the
drilling from the holes with a vacuum cleaner or with compressed air.
3. Attach the anchors (4 in the figure below) to the floor (refer also to
the anchor manufacturer’s instructions). Make sure that the distance
between the anchor and floor edge is at least 75 mm (2 in.).
4. Attach the plate to the floor with the screws (1 in the figure below) and
DIN 125 8.4 washers (2 in the figure below). Tighten the screws with
25 Nm installation torque. Note, that it is really important to tighten the
screws, this will ensure that the anchors are tightened firmly against the
concrete.
5. Insert the plugs to the floor plate (3 in the figure below) and to the pillar
(4 in the figure below).
6. Attach the generator frame (1) to the back plate with four M8x35 DIN 912
screws (2) and DIN 125 A 8.4 (3) washers using a 6 mm Allen key.
If it is not possible to attach the Planmeca ProX X-ray unit to the floor in the
standard way, a single-stud installation setup can be substituted.
As well as the main Planmeca ProX installation kit, this optional installation
variant requires the following additional parts (available upon request):
NOTE
The three screws required for the single stud adapter plate attachment (2 in
the figure above) are the same as are used in the attachment of the regular
adapter plate. This installation variant requires three attachment screws
rather than the original four.
The following parts from the standard Planmeca ProX installation are not
required in this variant, and can be removed:
• Single-stud assembly shield
12 Mobile installation
In the following figure, the mobile base dimensions are given.
Follow the steps below to install the Planmeca ProX with the mobile base.
1. Attach the wheel assemblies to the base plate with three M10x40 DIN
912 screws using a 8mm Allen key.
2. Rotate the column holder (2, located in the column) so that the base
plate can be attached to the column (1, base plate’s and column’s screw
holes in line). Make sure that the gap between the column edge and
column holder is approx. 3mm (0.12 in.) so that the column is firmly
attached to the base plate.
3. Attach the base plate to the column with four M10x40 DIN 912 screws
using a 8mm Allen key. Check that the column is vertical with e.g. a spirit
level.
4. Route the extension arm grounding lead out from the column. Attach the
extension arm to the column with two M8x20 DIN 912 screws using a 6
mm Allen key.
6. Attach the generator back plate to the column with six M8x10 DIN 912
screws using a 6 mm Allen key.
7. Attach the handle to the generator back plate with two M10x20 DIN 912
screws using a 8 mm Allen key.
8. Attach the generator frame to the back plate with four M8x35 DIN 912
screws and DIN 125 A 8.4 washers using a 6 mm Allen key.
12. Attach the column plug to its position with two attachment screws using a
3 mm Allen key. Insert the plugs to the column openings (black arrows).
13. If the optional column tray is installed, attach the grounding lead to the
bottom of the tray axle.
14. Attach the tray to the column with two locking screws using a 3 mm Allen
key. Make sure that the locking screws go into holes on the attachment
block.
16. Attach “No pushing” labels (included in to the X-ray unit delivery
package, figure below) to the both sides of the bracket arm.
NOTE
Only 391 mm and 518 mm extension arms can be used with the KaVo
treatment unit installation.
NOTE
The attachment surface must fulfill requirements given in section
"Environmental requirements" on page 4.
Steps
1. Install the KaVo treatment unit mounting bar adapter.
For more information, refer to the Kit adapter for accessory device for
light mounting post 3.002.7708 (KaVo document ID 1.012.8655).
2. Attach the single stud frame to the KaVo treatment unit mounting bar
adapter.
2.a. Route the cables through the opening on the adapter plate.
2.b. The single stud frame must be attached to the adapter with all six
screws.
• The extension cable (1) should extend 280 mm (11 in.) out from the
adapter.
• The interconnection cable (2) should extend 250 mm (10 in.) out
from the adapter.
4.a. Attach the shield (1) to the frame with two M4x5 DIN 7984 screws
(2) using a 2.5 mm Allen key.
4.b. Route the cable(s) (3) through the shield opening (4).
5. Attach the extension PCB assembly (1) to the frame with three M4x5
DIN 7984 screws (2) using a 2.5 mm Allen key.
6.a. Secure the extension cable to the extension PCB assembly with a
cable tie (1).
6.b. Connect the extension cable according to the diagram (2).
6.c. Attach the extension cable grounding lead to the extension PCB
attachment plate with M4x5 DIN 7984 screw and serrated washer
(3).
6.d. Bundle the primary and secondary wires with cable ties (4).
NOTE
There are four grounding points on the extension PCB attachment plate,
one for each grounding lead: cover plate, extension cable, arm and arm
cable.
8.a. Attach the cover plate (1) to the frame with two M4x5 DIN 7984
screws (2) using a 2.5 mm Allen key.
8.b. Route the grounding lead (3) down to the extension PCB
attachment plate and attach the grounding lead to the plate with
one M4x5 DIN 7984 screw and serrated washer (4).
9. Attach three set screws to the arm adapter using a 4 mm Allen key (1, 2
and 3).
10. Attach the arm adapter to the single stud frame with two M8x70 DIN 912
screws (1) and one M8x25 DIN 912 screw (2) using a 4 mm Allen key.
11.a. Route the arm and ProSensor cables through the arm adapter.
11.b. Mount the extension arm into position and use a spirit level (1) to
check that the arm is horizontal in three directions (2, 3 and 4).
11.c. Adjust the arm adapter to vertical position with three set screws
(5), if needed.
For more information on balancing the extension arm, see section
"Installing extension arm" on page 23.
12. Attach the rotation limit spacer and M8x10 DIN 912 screw to its position
(1) using a 4 mm Allen key.
You can change the bracket arm movement area with two securing
screws (2 and 3). Loosening the screws will narrow the movement area.
Adjust the arm rotational friction with the adjustment screw (4), tightening
the screw increases the friction.
13.a. Connect the arm cable to the extension PCB (1 and 2).
13.b. Attach the arm cable grounding lead with one M4x5 DIN 7984
screw and serrated washer (3).
13.c. Attach the arm grounding lead (4) to the extension PCB plate with
one M4x5 DIN 7984 screw and serrated washer.
13.d. The different ProSensor data cable (5) and interconnection cable
connection possibilities (6, 7 and 8), are shown on below.
Interconnection cable is routed directly to the computer
NOTE
You can also attach the control panel to a wall, refer to section "Remote
control panel installation" on page 103.
15. Push the excess cables into the extension arm and attach the end cover
to its position.
WARNING
Ensure that the power supply is switched off before connecting the cables.
NOTE
Use the telephone cable (max. 12 m) between the generator box and the
data wall socket.
Attach the frame of the data wall socket to the wall with two ø4x30 DIN 7981
screws. If the wall is made of concrete or brick, drill ø6 mm (0.23 in.), 32 -
35 mm (1.25 in.) in depth, holes and place the expansion anchors into them.
If the wall is made of wood or plaster, attach the frame to the wall with two
ø4x30 DIN 7981 screws to the wall without drilled holes. Do not use the
expansion anchors with wooden or plaster wall.
The data wall socket frame has a selection of possible through-holes (1) for
the cable.
Route the telephone cable through the hole on the side or bottom of the
frame. Assemble the data wall socket according to the figure above.
P13-2 to socket 2
P13-3 to socket 3
P13-4 to socket 4
P13-5 to socket 5
P13-2 to socket 5
P13-3 to socket 4
P13-4 to socket 3
P13-5 to socket 2
Install the holder attachment hook to the wall with two ø4x30 DIN 7981
screws. If the wall is made of concrete or brick, drill ø6 mm (0.23 in.), 32 - 35
mm (1.25 in.) in depth, holes and place the expansion anchors into them. If
the wall is made of wood, attach the screws to the wall without drilled holes.
A Wooden wall
B Concrete wall
1 Expansion anchor
2 ø4x30 DIN 7981 screw
3 Holder attachment hook
Attach the plate to the holder with two PT 3x8 WN 1451 screws. Press the
control panel holder to the attachment hook. Attach the control panel to the
holder and connect the control panel cable to the control panel and to the
wall socket.
NOTE
The control panel holder can be attached with a sticker to a surface which
can not be drilled (e.g. glass).
NOTE
If you are using a wall socket, attach the bottom plate according to
instructions given in section "Attaching bottom plate to wall socket" on page
110.
Use the bottom plate (1) as a template and mark the positions of the holes
for the four attaching screws to the wall.
NOTE
Do not use expansion anchors with wooden/plaster wall.
3 Telephone cable
1 Wall socket
2 Telephone cable
3 Bottom plate
NOTE
You can connect the telephone cable to the Extension cable PCB before
attaching the bottom plate to the wall socket. Refer to section "Connecting
cables" on page 111.
1 Telephone cable
2 Double exposure button cable
3 Generator PCB
4 Telephone cable
5 Remote exposure switch
6 Extension cable PCB
Connect the cross connection (part number 10007597) cable between the
Extension cable PCB and fixed control panel.
Press the fixed control panel to its position so that the mounting pins located
on the bottom plate hit the mounting springs on the control panel. Secure the
control panel to its position with one attachment screw.
Press and hold down the Mode key for more than 2 seconds, until the four
uppermost preprogrammed setting indicator lights come on.
17 Exposure lamp
The exposure lamp cable can be connected to the fixed control panel as
shown on the figure below.
19 Arm adjustments
19.1 Adjusting balance of the bracket arm
Adjust the balance of the arm by turning the adjustment nuts with a slot head
screwdriver. Tightening the nut will increase the tension.
Note that there are two arms, adjust them both.
If needed, increase the arm friction after adjusting the balance with the
adjustment screw using a slot head screwdriver. Tightening the screw will
increase the friction.
Place the arm cover to its position. Tighten three attachment screws using
a 3 mm Allen key and insert the cover plugs to the screw openings. Attach
also the cover of the other arm.
Adjust, if needed.
Refer to section "Arm adjustments" on page 119.
Documentation
Check that you can open the documents on the USB sticks.