Agilent 1146A
Agilent 1146A
User’s Guide
Contents
Receiving your Shipment. . . . . . . . . . . . . . . . . . 2
Packaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Warning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Control and Connector Identification . . . . . . . . 3
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Electrical Specifications . . . . . . . . . . . . . . . . 4
Environmental Specifications . . . . . . . . . . . 5
Mechanical Specifications . . . . . . . . . . . . . . 5
Safety Specification . . . . . . . . . . . . . . . . . . . 5
Electromagnetic compatibility . . . . . . . . . . . 5
Typical Response Curves . . . . . . . . . . . . . . . 6
Linearity for DC signal Range 10 mV/A . . . . 7
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Zero Adjustment . . . . . . . . . . . . . . . . . . . . . . 7
Current Measurement . . . . . . . . . . . . . . . . . 7
Battery Indication . . . . . . . . . . . . . . . . . . . . . 8
Overload Indication . . . . . . . . . . . . . . . . . . . 8
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Battery Replacement . . . . . . . . . . . . . . . . . . 8
Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Technical Assistance . . . . . . . . . . . . . . . . . . . . . 8
Packaging
The 1146A AC/DC Current Oscilloscope Probe is shipped with a separate battery
(not installed) and a User’s Guide.
WARNING • Connect the probe to the oscilloscope or voltage measuring instrument before
clamping the probe around a conductor.
• Never use the probe on circuits rated higher than 600 V or with float voltage
greater than 600 V.
• Never leave the probe clamped around a conductor while it is not connected
to an oscilloscope or voltage measuring instrument.
• Carefully center the conductor inside the probe jaws and ascertain that the
probe is perpendicular to the conductor before closing the jaws.
• Avoid, if possible, the proximity of other conductors which may create noise.
• Check the magnetic mating surfaces of the probe jaws; these should be free
of dirt, rust, or other foreign matter
• Do not use a probe which is cracked, damaged or has defective leads.
This symbol signifies that the 1146A AC/DC Oscilloscope Current Probe is
protected by double or reinforced insulation. Only use specified replacement
parts when servicing the instrument.
This symbol signifies CAUTION! and requests that the user refer to the user
manual before using the instrument.
2
Agilent 1146A AC/DC Oscilloscope Current Probe
Description
Description
The 1146A AC/DC Current Oscilloscope Probe expands oscilloscope applications
in industrial, automotive or power environments, and is ideal for analysis and
measurement of distorted current waveforms and harmonics. The probe permits
accurate display and measurement of currents from 100 mA to 100A rms, DC to
100 kHz without breaking into the circuit. The probe uses Hall effect technology
to measure AC and DC signals. The probe connects directly to an oscilloscope
through a 2 meter coaxial cable with an insulated BNC.
Compatibility
The 1146A AC/DC Current Oscilloscope Probe is compatible with any analog or
digital oscilloscope or other voltage measuring instrument which has the
following features:
• BNC input connector.
• Range capable of displaying 0.2 to 0.5 V per division.
• Minimum input impedance of 1 MΩ.
φ = 0.46”
(11.8 mm)
3
Agilent 1146A AC/DC Oscilloscope Current Probe
Specifications
Specifications
Electrical Specifications
Current Range 100 mV/A:100 mA to 10 A peak
10 mV/A:1 to 100 A peak
Output Signal 1000 mV peak max
AC current accuracy (After calibration and for one year)
(zero probe before making measurement)
Range Accuracy
100 mV/A (50 mA to 10 A peak) 3% of reading ±50 mA
10 mV/A (500 mA to 40 A peak) 4% of reading ±50 mA
10 mV/A (40 A to 100 A peak) 15% max at ±100A
Phase Shift1 < 1º from dc to 65 Hz on 10 mV/A
< 1.5º from dc to 65 Hz on 100 mV/A
Frequency Range DC to 100 kHz (-3dB with current derating)
Noise Range 10 mV/A: 480 µV
Range 100 mV/A: 3 mV
Slew Rate Range 10 mV/A: 20 mV/µs
Range 100 mV/A: 0.3 V/µs
Load Impedance >1 MΩ / 100 pF
Insertion Impedance (50/60 Hz) 100 mV/A: 0.01Ω
10 mV/A: 0.01Ω
Rise or Fall time Range 100 mV/A: 3 µs
Range 10 mV/A: 4 µs
Working Voltage 600 Vrms max.
Common mode voltage 600 Vrms max.
Influence of adjacent conductor <0.2 mA/A AC
Influence of conductor position in jaw 0.5% of reading at 1 kHz
Battery 9 V alkaline (NEDA 1604A, IEC 6LR61)
Low battery green LED when ≥6.5 V
Overload indication red LED indicates input greater than the selected
range.
Typical Consumption 8.6 mA
Battery Life 55 hours typical
1 Reference conditions 23 °C ±5 °C, 20 to 75% relative humidity, DC to 1kHz, probe zeroed,
1 minute warmup, batteries at 9 V ±0.1 V, external magnetic field <40 A/m, no DC component, no
external current carrying conductor, 1 MΩ / 100 pF load, conductor centered in jaw.
4
Agilent 1146A AC/DC Oscilloscope Current Probe
Specifications
Environmental Specifications
Operating Temperature 0º to 50 ºC
Storage Temperature -30º to 80 ºC
Temperature Influence <0.2% per ºC
Operating Relative Humidity
10º to 30 ºC: 85 5% RH (without condensation)
40º to 50 ºC: 45 5% RH (without condensation)
Altitude Operating:0 to 2000 m
Non operating:0 to 12,000 m
Mechanical Specifications
Zero Adjustment 20 turn potentiometer
Maximum cable diameter 11.8 mm
Case Protection IP20 per IEC 529
Drop Test 1.0 m on 38 mm of oak on concrete; test according to IEC 1010
Mechanical Shock 100 G; test per IEC 68-2-27
Vibration Test per IEC 68-2-6, frequency range 10 Hz to 55 Hz, amplitude
0.15 mm
Handle Lexan¨ 920A, UL 94 V2
Dimensions 231 x 36 x 67mm
Weight 330 g (11.6 oz) with battery
Color light gray
Output lead insulated coaxial cable with insulated BNC connector
Cable length 2 m
Safety Specifications
• Double insulation or reinforced insulation between primary or secondary
and outer case of the handle, per IEC 1010.
• 600 V category III, Pollution: 2.
• 300 V category IV, Pollution: 2.
• 5550 V 50/60 Hz between primary or secondary and the outer case of the
handle.
• 3250V 50/60 Hz between primary and secondary.
Electromagnetic compatibility
EN 50081-1 Class B
EN 50082-2 Electrostatic discharge IEC 1000-4-2
Radiated Field IEC 1000-4-3
Fast transients IEC 1000-4-4
Magnetic Field at 50/60 Hz IEC 1000-4-8
5
Agilent 1146A AC/DC Oscilloscope Current Probe
Specifications
Frequency
Phase Shift
6
Agilent 1146A AC/DC Oscilloscope Current Probe
Operation
-2%
-4%
-6%
Uncertainty
-8%
-10%
-12%
-14%
-16%
0A 20A 40A 60A 80A 100A
Accuracy
Operation
Zero Adjustment
The probe has a zero adjustment which should be adjusted before measurement.
Alternatively, you may "zero" with the oscilloscope instead.
Current Measurement
Connect the current probe to the proper input channel on the oscilloscope. Begin
with the least sensitive range on the current probe (10 mV/A). Select the
0.5 V/Division range on your oscilloscope. Clamp the probe on the conductor to
be measured and read the current flowing directly on your oscilloscope.
You may also use your oscilloscope to amplify the signal while using the 100 mV/A
probe range (which offers the best accuracy and least phase shift).
Important
It is possible to change the range on the current probe without removing the probe
from the current carrying conductor, but it is important to remember not to exceed
the permissible peak ratings of 1000 mV peak or 2000 mV peak to peak maximum. The
peak ratings by range are: 10 A peak on the 100 mV/A range, 100 A peak on the 10 mV/A
range.
7
Agilent 1146A AC/DC Oscilloscope Current Probe
Maintenance
Maintenance
Be sure that mating surfaces of the jaw are free of dirt or foreign matter. If they
are rusted, gently clean with a soft, lightly oiled cloth. Do not leave excessive oil
residue.
Battery Replacement
When the probe is turned on, the green battery indication LED should light up.
If not, replace the 9 V battery (see Figure 1). To replace the battery, disconnect
the probe from the circuit and the oscilloscope. Turn the probe "OFF". Unscrew
the battery compartment screw and pull out the battery compartment cover.
Replace the battery and put the cover back on. Do not replace the battery while
probe is in use.
8
Agilent 1146A AC/DC Oscilloscope Current Probe
Calibration
Calibration
Repair and Calibration
To guarantee that your instrument complies with the factory specifications, we
recommend that the 1146A AC/DC Oscilloscope Current Probe be submitted to
our factory service center at one year intervals for recalibration, or as required
by other standards.
Equipment Required
To calibrate the 1146A, carefully pop out the plastic dot located on the left side
of the probe, to access the potentiometer.
Turn the calibrator on and allow 15-30 minutes for it to warm up.
Figure 2
9
Agilent 1146A AC/DC Oscilloscope Current Probe
Calibration
Positive Negative
Measure Minimum Maximum Measure Minimum Maximum
10 A 98 mV 102 mV -10 A -98 mV -102 mV
30 A 294 mV 306 mV -30 A -294 mV -306 mV
60 A 576 mV 624 mV -60 A -576 mV -624 mV
90 A 837 mV 963 mV -90 A -837 mV -963 mV
100 A 910 mV 1.09 V -100 A -910 mV -1.09 V
Positive Negative
Measure Minimum Maximum Measure Minimum Maximum
1A 98 mV 102 mV -1 A -98 mV -102 mV
3A 294 mV 306 mV -3 A -294 mV -306 mV
10
Agilent 1146A AC/DC Oscilloscope Current Probe
Calibration
Positive Negative
6A 588 mV 614 mV -6 A -588 mV -612 mV
10 A 980 mV 1.02 V -10 A -980 mV -1.02 V
Measure Range
+10 A ±2% 100 mV/A
+120 A 10 mV/A
-120 A 10 mV/A
11
Agilent 1146A AC/DC Oscilloscope Current Probe
Calibration
12
DECLARATION OF CONFORMITY
According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014
Date: 06/28/2000
Name
For further information, please contact your local Agilent Technologies sales office, agent, or distributor.
Product Regulations
EMC IEC 61326-1:1997+A1:1998 / EN 61326-1:1997+A1:1998
CISPR 11:1990 / EN 55011:1991
IEC 61000-4-2:1995+A1:1998 / EN61000-4-2:1995 A
IEC 61000-4-3:1995 / EN 61000-4-3:1995 B
IEC 61000-4-4:1995 / EN 61000-4-4:1995 A
IEC 61000-4-5:1995 / EN 61000-4-5:1995 A
IEC 61000-4-6:1996 / EN 61000-4-6:1996 A
IEC 61000-4-11:1994 / EN 61000-4-11:1994 A
Canada: ICES-001:1998
Australia/New Zealand: AS/NZS 2064.1
Performance Codes:
A PASS - Normal operation, no effect.
B PASS - Temporary degradation, self recoverable.
C PASS - Temporary degradation, operator intervention required.
D FAIL - Not recoverable, component damage.
Notes: (none)
ICES/NMB-001
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est confomre à la norme NMB-001 du Canada.
Agilent Technologies
P.O. Box 2197
1900 Garden of the Gods Road
Colorado Springs, CO 80901
Notices with regard to this man-
ual and any information
WARNING
contained herein, includ- A WARNING notice
© Agilent Technologies, Inc. ing but not limited to the denotes a hazard. It
1996-2002 implied warranties of calls attention to an
No part of this manual may be merchantability and fit-
reproduced in any form or by ness for a particular pur- operating procedure,
any means (including elec- pose. Agilent shall not be practice, or the like
tronic storage and retrieval or liable for errors or for that, if not correctly
translation into a foreign lan- performed or adhered
guage) without prior agree- incidental or consequen-
ment and written consent tial damages in connec- to, could result in
from Agilent Technologies, tion with the furnishing, personal injury or
Inc. as governed by United use, or performance of death. Do not proceed
States and international copy- this document or of any
right laws. information contained
beyond a WARNING
herein. Should Agilent notice until the
Manual Part Number
and the user have a sepa- indicated conditions
01146-92002, September 2002 are fully understood
rate written agreement
Print History with warranty terms cov- and met.
01146-92000, November 1996 ering the material in this
document that conflict
01146-92001, August 2000
with these terms, the war-
01146-92002, July 2002 ranty terms in the sepa- CAUTION
Agilent Technologies, Inc. rate agreement shall
1900 Garden of the Gods Road control. A CAUTION notice
Colorado Springs, CO 80907 denotes a hazard. It
USA Technology Licenses calls attention to an
The hardware and/or soft-
Restricted Rights Legend ware described in this docu- operating procedure,
If software is for use in the ment are furnished under a practice, or the like
performance of a U.S. Govern- license and may be used or that, if not correctly
ment prime contract or sub- copied only in accordance
contract, Software is with the terms of such license.
performed or adhered
delivered and licensed as to, could result in
“Commercial computer soft- damage to the product
ware” as defined in DFAR or loss of important
252.227-7014 (June 1995), or as
a “commercial item” as data. Do not proceed
defined in FAR 2.101(a) or as beyond a CAUTION
“Restricted computer soft- notice until the
ware” as defined in FAR
52.227-19 (June 1987) or any indicated conditions
equivalent agency regulation are fully understood
or contract clause. Use, dupli- and met.
cation or disclosure of Soft-
ware is subject to Agilent
Technologies’ standard com-
mercial license terms, and
non-DOD Departments and
Agencies of the U.S. Govern-
ment will receive no greater
than Restricted Rights as
defined in FAR 52.227-19(c)(1-
2) (June 1987). U.S. Govern-
ment users will receive no
greater than Limited Rights as
defined in FAR 52.227-14 (June
1987) or DFAR 252.227-7015
(b)(2) (November 1995), as
applicable in any technical
data.
Document Warranty
The material contained in
this document is provided
“as is,” and is subject to
being changed, without
notice, in future editions.
Further, to the maximum
extent permitted by appli-
cable law, Agilent dis-
claims all warranties,
either express or implied,
sA
Agilent Technologies
Printed in the Malaysia
*01146-92002*