Integrated Transducers
UFA-, UVA-, UTA-Ex-d
Operating Instructions
Contents
A1 Scope of Delivery
A1.1 Description, Type Plates
A2 Technical Specifications
A2.1 Operating Conditions
A2.2 Housing and Connections
A2.3 Electrical Data
A2.4 Measurement Uncertainty
A3 Installation
A3.1 Terminal Connections
A3.2 Wiring Diagrams
A3.2.1 Power supply
A3.2.2 Analog output v
A3.2.3 Digital output (relay contact)
A3.2.4 HART® interface
A3.2.5 optional LCD display
A4 Functional Description
A5 Settings
A6 Initial Operation
A7 Operation
A8 Shut-down, Dismantling
A9 Inspection
A10 Troubleshooting
A11 Replacement Parts
U377_UxAintExd_B_e_220803 1/11
Operating Instructions
Integrated Transducers
UFA-, UVA-, UTA-Ex-d
A1 Scope of Delivery
- Transducer UFA, UVA or UTA integrated in the connection housing of flow sensor FA, VA or TA
- Operating Instructions Flow Sensor FA, VA or TA
- Data Sheet flow sensor FA, VA or TA with integrated transducer UFA, UVA or UTA
- CD-ROM with PC configuration software UCOM (optional)
- HART® modem for PC connection to USB port (optional)
Please check that everything listed in the Delivery Note / Technical Data Sheet is included in the delivery.
One of the following type plates (or similar) can be found on the connection housing:
Tambient : ambient air temperature range connection housing -20 ... +60 °C
2/11 U377_UxAintExd_B_e_220803
Operating Instructions
Integrated Transducers
UFA-, UVA-, UTA-Ex-d
A2 Technical Specifications
Ambient temperature
of connection housing
when in use : -20 ... +60 °C
Supply voltage,
mains supply 24 V DC (20 ... 27 V DC), power < 5 W
supply current <150 mA
U377_UxAintExd_B_e_220803 3/11
Operating Instructions
Integrated Transducers
UFA-, UVA-, UTA-Ex-d
A3 Installation
The relevant national regulations for installing electrical equipment, the General Engineering Regulations
and these Operating Instructions apply.
Connecting terminals are accessible after removing the housing cover of the terminal compartment (hous-
ing cover on the cable gland side).
Electrical connection must be carried out according to the appropriate wiring diagram. Faulty connection
can cause damage to the electronics.
Do not install or wire up the transducer under mains voltage. Non-compliance can cause damage to
the electronics.
4/11 U377_UxAintExd_B_e_220803
Operating Instructions
Integrated Transducers
UFA-, UVA-, UTA-Ex-d
Before connecting please check that the power supply is within the specification. All relevant information
can be found on the type plate.
Protective ground
terminal on the
outside of the
housing
the exterior of
the housing
24 V DC
Equipotential
(20 … 27 V DC)
bonding
4 - 20 mA
4-20 mA
Resistance max. 500 Ohm or
Resistance max.
250 ... 500 Ohm500 Ohm
with HART
® interface
or with
HART interface 250 ... 500 Ohm
The analog output is configurable with the UCOM software via the HART® interface. The factory-set
values can be found in the accompanying documents.
U377_UxAintExd_B_e_220803 5/11
Operating Instructions
Integrated Transducers
UFA-, UVA-, UTA-Ex-d
The function of the digital output and the corresponding setting parameter are configurable using the
UCOM software via the HART® interface. The factory-programmed settings can be found in the parameter
printout included with delivery.
To connect the HART® interface an optional HART® modem is connected to the closed circuit of the analog
output 4 … 20 mA (connect to +4 ... 20mA terminal and -4 … 20mA terminal). The polarity of the HART®
modem connection is irrelevant. The resistance of the 4 ... 20 mA circuit must be between 250 Ohm and
500 Ohm.
HART modem ®
HART
modem
PC connection PC connection
6/11 U377_UxAintExd_B_e_220803
Operating Instructions
Integrated Transducers
UFA-, UVA-, UTA-Ex-d
(1)
(2)
(3)
The flat ribbon cable with 10-pin connector should not be plugged in or unplugged when live!
Risk of damage to equipment!
Visible are the readout potentiometer (3) for the display contrast, the reset button (4) for the counter, the
jumpers St1 m/s-m³/h (5) and St2 A-B (6) (see A4 Functional Description).
U377_UxAintExd_B_e_220803 7/11
Operating Instructions
Integrated Transducers
UFA-, UVA-, UTA-Ex-d
A4 Functional Description
UFA transducers for vane wheel probes FA and FAR (directional sensing) and vane wheel measuring
tubes FA Di and FAR Di (directional sensing) for measuring flow velocity or flow rate of air/gases and
water/liquids.
UVA transducers for vortex sensors VA for measuring flow velocity or flow rate of air/gases.
UTA transducers for thermal sensors TA for measuring standard flow velocity or standard flow rate of
air/gases.
The signal frequency or voltage generated from the flow sensor is converted to a linear analog output
signal 4-20 mA. The analog terminal value is configurable.
When logging directional sensing data, the zero point can be selected in the middle of the analog range, or
display of flow direction takes place with the aid of the digital output:
With transducers UFA and UVA the actual velocity/actual flow rate can be converted to standard velocity/
standard flow rate by entering the parameters 'temperature' and 'pressure'. Working temperature and
pressure should be constant.
For monitoring of power supply, data logging, sensor interface, parameter settings (see under A10:
Troubleshooting)
HART® interface
for changing calibration data and setting parameters.
Connect HART® modem (optional) to transducer UFA/UVA/UTA on the closed circuit of the analog output
(connect to both terminals of the analog output 4 ... 20 mA). Polarity of the connecting terminals of the
HART® modem is irrelevant, resistance between 250 Ohm and 500 Ohm to be strictly observed). Connect
USB plug to the USB PC connection.
Changes to the settings can now take place after starting the PC programme UCOM (optional)
(see under A5: Settings).
8/11 U377_UxAintExd_B_e_220803
Operating Instructions
Integrated Transducers
UFA-, UVA-, UTA-Ex-d
Configuration (see Fig. 2, under A3.2.5) via 2 jumper wrap connections St1 m/s-m³/h and St2 A-B
Display row 1:
m/s-m³/h = m/s and A-B = any: velocity in (N)m/s *
m/s-m³/h = m³/h and A-B = A: flow rate in (N)m³/h
m/s-m³/h = m³/h and A-B = B: flow rate in (N)l/h **
* standard values (N) UFA and UVA only, when parameter 'switching v/NV' =1 (see under A5)
** only when Di < 75.0 mm, otherwise display in (N)m³/h
Display row 2:
Quantity counter in m³ with 0 ... 3 decimal places
(see under A5: parameter 'switching pulse m³(cbm) / l (litre)' and
parameter 'm³ (cbm) / l (litre) per pulse' and
parameter 'decimal places quantity display')
with error : error 01 = parameter error
error 02 = sensor error
Reset button on the LCD display module behind the glass of the housing cover: see Fig. 2,
A3.2.5:
Reset the counter by pressing the reset button for more than 3 seconds. This can also be carried out via
the HART® interface.
A5 Settings
The following setting parameters can be read using the PC software UCOM and are also alterable.
The customer-specific settings are shown on the parameter print-out, which is included in the
documents.
A6 Initial Operation
U377_UxAintExd_B_e_220803 9/11
Operating Instructions
Integrated Transducers
UFA-, UVA-, UTA-Ex-d
A7 Operation
(Pay attention to A2.1 Operating Conditions)
(Pay attention to A2.3 Electrical Data)
A8 Shut-down, Dismantling
Please ensure that the supply voltage is switched off before disconnecting.
A9 Inspection
see under A4 Functional Description, self-diagnosis.
The screw threads of the housing cover have been treated with graphite to protect against corrosion.
10/11 U377_UxAintExd_B_e_220803
Operating Instructions
Integrated Transducers
UFA-, UVA-, UTA-Ex-d
A10 Troubleshooting
Measured value too Profile factor set too high Set profile factor to relevant nominal diameter
high and sensor type
EMC problem See reference to electromagnetic compatibility
(EMC)
®