Basic Principles of Gamma Camera Imaging & QC
Basic Principles of Gamma Camera Imaging & QC
No financial disclosures.
Gamma camera images and photographs of
equipment are for illustrating concepts and not
intended to advertise or endorse any particular
manufacturer or vendor.
Page 2
Learning Objectives
4
Gamma Camera Operation
Array of Photomultiplier Tubes (PMTs):
Localizes the position where the gamma ray
interacts in the crystal
Collimator:
Forms a projection image by
allowing only gamma rays traveling
g in certain directions to reach crystal (for
g a parallel hole collimator, gamma rays
approximately perpendicular to crystal
pass through).
g
Gamma rays emitted from patient
5
Conventional Gamma Cameras
Page 6
Position Determination
Page 7
Types of Imaging
Static Planar
Dynamic Planar
Whole body
Tomographic (SPECT)
Not all gamma cameras do all types of imaging
– some do only planar, or only SPECT.
Page 8
SPECT Operation
Static Planar
Spatial Resolution
Efficiency/Sensitivity
Energy resolution
Page 11
Spatial Resolution
• Intrinsic resolution (Rint) refers to how well the
crystal and PMT system localize an interaction
in the crystal. Affected by crystal thickness,
gamma ray energy, scatter in crystal.
• Collimator resolution (Rcoll) refers to how well
the collimator localizes the gamma ray source
in the patient, affected by hole diameter and
length, distance from collimator to patient.
• System resolution (Rsys) is a combination of
intrinsic and collimator resolution:
12
Intrinsic Spatial Resolution
Page 13
Intrinsic Spatial Resolution
Page 14
Bar pattern using Thallium, one
peak at a time
Lower energy peak only, 69 keV Upper energy peak only, 167 keV –
Better resolution at higher energy15
Collimators
Page 16
Collimators and Scatter
Page 17
Scatter in Patients
d
d x Rcoll ( L x)
L
19
At collimator surface 5 cm from surface
Page 22
Efficiency or Sensitivity
Page 23
Efficiency or Sensitivity
Page 24
Efficiency or Sensitivity
• Parallel hole collimator efficiency
proportional to:
ædö
2
d 2
ç ÷ ´
è L ø (d + t) 2
d=hole diameter
L=hole length
t=septal thickness 25
System Sensitivity
Page 26
Energy Resolution
Page 27
Energy Resolution
Page 28
Image Acquisition Options
Page 29
Image Acquisition Options
Page 30
SPECT Phantom
• Jaszczak Phantom
for SPECT quality
control.
• Approved by ACR
for SPECT ACR
accreditation images
• Standing on end,
used for evaluation
of planar spatial
resolution with
scatter – rod sizes:
12.7, 11.1, 9.5, 7.9,
31
6.4 and 4.8 mm
Image Acquisition Options
Page 32
64 X 64 128 X 128
33
64X64 128 X 128
Pixel 8.8 mm Pixel 4.4 mm
34
Bar spacings 2.0, 2.5, 3.0, 3.5 mm
Image Acquisition Options
Page 35
512 X 512 matrix,
1 M counts
36
Image Acquisition Options
Page 37
500K 1M
38
1.25M 5M
39
Gamma Camera Calibrations
Page 40
Correction Tables
Page 41
Effects of Correction Tables
No Energy
corrections And
Linearity
Energy,
Energy Linearity,
only Uniformity
(all corrections)
42
99mTc Intrinsic Flood Images
Center of Rotation
Pixel number Offsets between physical center of
recording signal
rotation and center of image matrix must
128x128 matrix
be corrected for.
1 63
COR error
Source
65 1 Image
Image acquisition Backprojection Matrix
43
What is the primary function of a
collimator in a gamma camera?
A. Protect the crystal 96%
B. Define direction of
incoming gamma rays
entering crystal
C. Substantially eliminate
scatter
D. Reduce count rate to
prevent dead time 0%
4%
0% 0%
44
What is the primary function of a
collimator in a gamma camera
A. Protect the crystal
B. Define direction of incoming gamma rays
entering crystal
C. Substantially eliminate scatter
D. Reduce count rate to prevent dead time
E. Shield the electronics
thicker crystal
B. Use a 64X64 matrix
rather than 256X256
C. Image with lowest energy
gamma rays available
D. Position patient as close 8%
4%
as possible to collimator 0%
face A. B. C. D.
46
Which would improve spatial resolution in
gamma camera images?
A. Choose a camera with a thicker crystal
B. Use 64 X 64 matrix rather than 256 X 256
C. Image with lowest energy gamma ray available
D. Position patient as close to collimator face
as possible
Reference: The Essential Physics of Medical
Imaging, JT Bushberg, JA Seibert, EM Leidholdt
Jr, JM Boone, 3rd edition, 2012, 686-691.
Physics in Nuclear Medicine, SR Cherry, JA
Sorenson, ME Phelps, 4th edition, 2012,365-366.
47
Quality Assurance
Page 48
Joint Commission Requirements
Page 49
American College of Radiology Accreditation
At least annually:
Intrinsic uniformity
System uniformity
Intrinsic or System spatial resolution
Relative sensitivity
Energy Resolution
Count Rate Parameters
System performance for SPECT: tomographic
uniformity, contrast and spatial resolution
Page 50
Intersocietal Accreditation Commission
(IAC) Recommendations
Page 51
Routine QC - Uniformity
Page 52
System Uniformity
With collimator on, use planar sheet source:
Page 54
Uniformity- Annual Testing
Page 55
Good uniformity images
56
Poor uniformity
57
Uniformity - Quantification
Page 61
Spatial Resolution – Four Quadrant
Bar Pattern
62
Intrinsic Bars – Linearity Correction
Off
63
Extrinsic Resolution (FWHM) with Line
Source
Page 64
Line source profile and curve
FWHM= 7 pixels =
7.7 mm
65
System Resolution with Scatter
Tc-99m Tl-201
Static images of SPECT phantom standing on end on top of
collimator. Provides a measure of planar system resolution with
66
scatter.
Measuring Sensitivity
– Place ~1-2 mCi 99mTc,
and small volume of
water in plastic flat-
bottomed vial on top of
Styrofoam cup 10 cm
from collimator face.
– Record exact activity
and time
– Count for 1 min, also
count and subtract
background
– Use total counts in
image, not an ROI
drawn around image 67
Measuring Sensitivity
Page 68
Energy Resolution Measurement
69
Estimating Energy Resolution
Page 71
Low to High Count Rate Intrinsic
Floods
19 kcps 79 kcps
109 kcps
(too high
without high
count rate
mode
corrections)
72
SPECT Image Quality
Page 73
SPECT Phantom Imaging
74
SPECT phantom imaging procedure
Page 75
SPECT phantom imaging procedure
Page 76
SPECT phantom imaging procedure
Page 77
SPECT phantom reconstructed
slices
78
SPECT phantom reconstructed slices –
no attenuation correction
79
SPECT phantom image quality
Page 80
Ring Artifacts
Ring artifacts
visible
81
Severe Ring Artifacts
82
SPECT Ring Artifacts
Page 83
Phantom filled with 99mTc Sestamibi
rather than Sodium Pertechnetate
Artifact
84
Which of these is not specifically listed as an
annual physics test required by TJC or
ACR?
13% A. Sensitivity
74% B. Center of Rotation
0% C. Uniformity
13% D. Energy Resolution
85
Which of these is not specifically listed as an
annual physics test required by TJC or
ACR?
A. Sensitivity
B. Center of Rotation
C. Uniformity
D. Energy Resolution
References: The Joint Commission Revised Requirements
for Diagnostic Imaging Services,
http://www.jointcommission.org/assets/1/6/HAP-
CAH_DiagImag_Prepub_July2015release_20150105.pdf
ACR Nuclear Medicine Accreditation Program
Requirements, http://www.acr.org/Quality-
Safety/Accreditation/Nuclear-Med-PET
86
What is the primary cause of ring
artifacts in SPECT phantom images?
86% A. Non-uniformities
14% B. Center of Rotation error
0% C. Phantom off center in field of view
0% D. Using the wrong matrix size
87
What is the primary cause of ring artifacts in
SPECT phantom images?
A. Non-uniformities
B. Center of Rotation error
C. Phantom off center in field of view
D. Using the wrong matrix size
88
Imaging and recording counts for a known
amount of activity in a small flask for 1 min is
a method of measuring _____?
13% A. Uniformity
4% B. Spatial Resolution
79% C. Sensitivity
4% D. Energy Resolution
89
Imaging and recording counts for a known
amount of activity in a small flask for 1 min is
a method of measuring _____?
A. Uniformity
B. Spatial Resolution
C. Sensitivity
D. Energy Resolution
Page 91