0% found this document useful (0 votes)
514 views27 pages

Sentinel SVC Service Manual Manual

Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
514 views27 pages

Sentinel SVC Service Manual Manual

Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 27

ENTERAL FEEDING PUMP

SERVICE MANUAL

ALCOR Scientific Inc


20 Thurber Blvd Smithfield, RI
02917 USA
T 401 737 3774 / 800-495-7295

- May, 2017 -

Proprietary Notice

This document is the property of ALCOR Scientific. All information contained herein is considered
company proprietary. Its use is restricted solely for the purpose of assisting in the repair of this
product. Reproduction of all or any part of this document is prohibited without the prior written
consent of ALCOR Scientific.

CONTROLLED
TABLE OF CONTENTS

INTRODUCTION

Description 4

OPERATION

Displays & Switches 6

ROUTINE MAINTENANCE

Cleaning 8
Functional Tests 9

TEST PROCEDURES

Accuracy and Electrical 10

THEORY OF OPERATION

Software Description 14
Hardware Description 16

TROUBLESHOOTING 18

APPENDIX A

Specifications 22
Pump Block Diagram 23
Pump Hardware Diagram 24
Mechanical Parts List 25
Test Loop Assembly 27

CONTROLLED 2 of 27
DESCRIPTION

The purpose of this manual is to provide information necessary for the maintenance and repair of the
SENTINELplus Enteral Feeding Pump. Refer to the SENTINELplus Operator Manual for additional
description of pump operation. To aid in understanding, all references to keypad and display labels are
shown in boldface type.

The SENTINELplus pump is a rotary peristaltic design, and is capable of pumping all enteral feeding
formula. Key features of the pump are:

* Large displays can be read from a distance


* Total accumulated Dose (DOSE CHECK)
* Formula Free Flow detection and alarm
* Automatic maximum alarm volume during Free Flow alarm
* Rate and Volume stored in non-volatile memory
* Programmable Dose Limit from 5ml to 9999ml
* Low occlusion pressure
* Self-Test Mode
* Built-in pole clamp
* AC powered with Battery back-up operation
* Smooth enclosure surfaces for ease of cleaning
* Secure delivery set retention
* Delivery and Battery Operation service accessible history
* Display brightness control for night/battery use

HOLD MODE
In this non-pumping mode the Infusion Rate display flashes on and off, signaling inactivity. A flow rate is
programmed by pressing the UP or Down key, changing the Infusion Rate display. The right side display
normally shows Volume Delivered, the running volume total. Pressing CLEAR while viewing Volume
Delivered causes "cLr" to appear in the left display; after the second beep, the value clears. Pressing the SET
key causes the Dose Limit value to be shown in the right display, along with "doS/SEt" in the left display.
While being shown, Dose Limit can be changed by pressing UP or Down, or cleared after the second beep
while holding CLEAR. After three seconds with no key being pressed, the right display automatically returns to
displaying Volume Delivered.

Pressing DOSE CHECK shows total accumulated volume, along with "tOt" in left display. It can be
cleared with CLEAR while being viewed.

Volume Delivered, Dose Limit, and Infusion Rate may all be cleared at the same time by pressing and
holding CLEAR for about 6 seconds (after 7 beeps). Leaving the pump in the Hold Mode for more than 2.5
minutes without pressing a key will cause a “StOP” alarm.

RUN MODE
Pressing the RUN/STOP switch while in Hold Mode causes the pump to begin pumping fluid through the
tubing set, producing drops in the drip chamber. Drops are detected when they interrupt an IR beam passing
through the drip chamber. The motor rotation is duty-cycled to produce the programmed flow rate. While
pumping, Volume Delivered displays milliliters pumped since last cleared. When Volume Delivered is equal
to or greater than the

CONTROLLED 3 of 27
programmed Dose Limit value, the pump rotor will stop, the alarm will beep two times every 2½ minutes, and
the Dose Comp LED will continuously flash along with the Volume Delivered display.

An insufficient drop rate or blocked delivery tubing will generate an Occlusion/Empty alarm, while an
excessive drip rate will cause a Free Flow alarm, with a distinctive audible alarm at maximum volume, and the
text "FLO FASt" appearing in the display.

Pump operation may be stopped at any time by pressing RUN/STOP, causing the pump to enter Hold
Mode. Attempting to start the pump with Volume Delivered equal to or greater then the Dose Limit will
cause an immediate alarm.

DOSE CHECK
DOSE CHECK allows monitoring of long term volume infusion, up to 10 liters. It is useful for long-
term dose monitoring, as it is cleared separately from Volume Delivered.

MEMORY
When the pump is first turned on, Infusion Rate, Volume Delivered, Dose Limit, and total accumulated
dose are recalled from EEPROM memory and are used as current values until changed or cleared. This
non-volatile memory aids in quick set-up for repeat usage on the same patient.

CONTROLLED 4 of 27
CONTROL PANEL

CONTROLLED 5 of 27
CONTROLS AND DISPLAYS

NUMERICAL DISPLAYS

Infusion Rate
Flow rate of formula in ml/hr.

Volume Delivered/Dose Limit


Volume Delivered is total quantity of formula delivered in ml. Normally displayed in Stop
and Run Modes.

When SET key is pressed, programmed Dose Limit set point is displayed. Value remains for three (3)
seconds unless key pressed. It can only be changed in Hold Mode.

LED INDICATORS

Bat
LED remains lit while running from battery power, extinguished when running on AC power.
LED flashes when battery reaches a low charge. After approximately 15 minutes the alarm beeps and
pumping stops. To maximize battery life, avoid frequent battery discharge. Always plug pump
into AC receptacle while in use.

AC
AC indicator remains lit whenever AC power is present, whether pump is off or operating.

Occl/Emp
LED flashes, audible alarm beeps when drop sensor fails to detect drops (tubing occlusion or
empty bag)

LED flashes, audible alarm beeps when pump left in Hold Mode for 2.5 minutes without pressing a
key (inactivity alarm).

Dose Comp
LED flashes, audible alarm beeps when programmed Dose Limit has been reached.

LED flashes, audible alarm beeps if attempt is made to enter Run Mode with a programmed Dose
Limit value less than or equal to the current Volume Delivered.

CONTROLLED
CONTROLS AND DISPLAYS

PUSHBUTTON SWITCHES

ON
Turns pump on

OFF
Turns pump off. Battery charging circuitry remain active

UP
Increases Infusion Rate or Dose Limit, whichever is displayed

DOWN
Decreases Infusion Rate or Dose Limit, whichever is displayed

RUN/STOP
Alternately starts and stops pumping. If alarming, press once to stop alarm then again to start pump
SET
Displays current Dose Limit and allows Dose Limit to be set or cleared. Can only be
changed in Hold Mode. Can be viewed in Run Mode.

CLEAR Resets whatever value being shown on right display to zero. Active only in Hold Mode, clears 1)
Volume Delivered ("cLr" in left display) immediately upon pressing, 2) Dose Limit ("doS" in left
display) after the second double beep, 3) total accumulated dose after the second double beep, and
4) Infusion Rate, Volume Delivered, and Dose Limit after single beep while "ALL" is in left display.

DOSE CHECK
Displays total accumulated volume since last cleared. "tot" shows on left display. Similar to Volume
Delivered, but cleared separately for long-term volume monitoring.

CONTROLLED
ROUTINE MAINTENANCE

CLEANING

Cleaning consists of removing any spillage from the pump housing, rotor, rollers, control panel, and drop sensor.

WARNING:

DISCONNECT THE PUMP FROM AC POWER SOURCE BEFORE


CLEANING TO PREVENT ELECTRICAL SHOCK!

PROLONGED EXPOSURE TO ALCOHOL, HOUSEHOLD DETERGENTS OR


STRONG CLEANERS COULD RESULT IN DAMAGE TO THE PUMP HOUSING!

DO NOT IMMERSE PUMP IN ANY LIQUID! DO

NOT AUTOCLAVE!
CHECK EFFECT OF CLEANING AGENTS BEFORE GENERAL USE!
DO NOT USE CLEANING AGENTS THAT ARE HARMFUL TO POLYCARBONATE PLASTIC.

1. Using a damp cloth or sponge, clean the pump housing and rollers with warm soapy water. Clean the
drop sensor parts with soft cotton swabs and isopropyl alcohol or warm soapy water.

2. Check that the rollers rotate freely. If they do not rotate easily, clean them as described in steps 3
to 8.

3. Loosen the set screw with a 3/32" Allen wrench and pull the pump rotor from its shaft.

4. Submerge the entire rotor in warm soapy water until rollers move freely.

5. After cleaning, align pump rotor with the flat on the motor shaft.

6. Push the rotor on the shaft, and use 0.125” Feeler Gauge between rotor and case.

7. Tighten the set screw. Warning: excess torque can strip rotor threads.

CONTROLLED
ROUTINE MAINTENANCE
FUNCTIONAL TESTS

These tests check operation of the motor, microcontroller, display, keys, audible alarm and drop sensor

NOTE: requires one disposable bag set or continuous loop test set.

1. Test Mode Check

1.1. Turn pump on


1.2. To enter the Test Mode, press UP and hold until 295ml/hr is displayed. Release and press UP and
hold for five seconds. The pump rotor turns continuously and the number 8 scrolls across the displays.
1.3. Press any switch except ON or OFF. Each time a switch is pressed and released, the motor stops,
the alarm sounds, and all LEDs light. Check each switch.
1.4. Insert test set with water, count the drops entering the drip chamber. As each drop of water enters
the drip chamber, the 3 LED’s should illuminate, the rotor stops, and alarm beeps. Count 25 drops.
NOTE! Each drop entering the drip chamber must be detected!
1.5. Press OFF to exit the Test Mode.

Any failure indicates further service must be performed to locate the malfunction.

2. No-Occlusion Check

2.1. Install test set with water (or test loop) in pump.
2.2. Press SET button; immediately press UP button until Dose Limit (right display) reaches 22.
2.3. Press Dose Check button; “tot” should appear on left display, and right display shows total
accumulated volume. Remember this value.
2.4. Set Infusion Rate to 295 ml/hr.
2.5. Press RUN/STOP button and let pump deliver 22 ml. When done, pump should alarm and
indicator LED “Dose Comp” flashes.
2.6. Press RUN/STOP to silence alarm.
2.7. Press Dose Check; “tot” should appear in left display.
The rate in the right display should have increased by 22 ml.
2.8. Press OFF button.

3. Motor Speed Check

Check motor speed by measuring the time required for the pump rotor to make ten revolutions while in the
Test Mode, as described above. To aid in counting, put a small piece of tape on one of the rotor lobes. Ten
revolutions take 38 seconds. Only one measurement is required to verify correct motor operation.

4. Occlusion Alarm Check

Without the drip chamber in place, set the Rate to 295 ml/hr and press RUN/STOP to start the
pump. An occlusion alarm should occur after approximately 5 revolutions of the rotor.

CONTROLLED
THEORY OF OPERATION
--------------------

A. PUMP ACCURACY TEST

Follow procedure OMI-239, latest revision

B. ELECTRICAL TESTS

This procedure checks system electrical specifications.

Equipment:

1. Digital multimeter
2. Assorted hand tools
3. Optional: Variable voltage power supply (0-12VDC, 0.500 amp)

WARNING

USE CAUTION WHEN CASE IS OPEN. 12VDC POWER PRESENT AT ALL TIMES.

Procedure

1. Test Mode

1.1. Plug the AC power supply into a 100-240vac mains receptacle.


1.2. Install and fill a pump set with water
1.3. Enter the Test Mode as described above.

2. Check Numerical Displays

2.1. While in Test Mode verify all display segments are lit uniformly as the number “8” scrolls across
the displays.

3. Check drop sensor and discrete LED indicators

3.1. While in Test Mode, verify that Occ/Emp, Low Bat, Dose Comp, and Bat LEDs blink and the
alarm sounds for every drop passing sensors.

4. Check Keypad Switches

4.1. In Test Mode, press each key (except ON or OFF) several times and check for a beep with each
press and release.

5. Check Audible Alarm

5.1. Exit Test Mode by turning pump off then on again.


5.2. Set Rate at 295 ml/hr and start pump by pressing RUN/STOP.
5.3. At 5.3 revolutions of the rotor, the audible alarm should begin beeping and the Volume
Delivered/Dose Limit display should read "2" ml.
5.4. Press RUN/STOP to silence alarm.

CONTROLLED
THEORY OF OPERATION
--------------------

6. Check AC/Battery Sensing

6.1. With the pump in Hold (or Run) Mode and the AC Power Supply plugged into an outlet, verify
that the Bat LED is extinguished.
6.2. Remove the AC Power Supply from the outlet and verify that the Bat LED lights within a few
seconds (NOTE: in the case of a deeply discharged battery, it may take up to a few minutes to
charge the battery to the threshold voltage point, at which time the Bat LED should
extinguish).

7. Electrical Measurements

The following measurements on the internal PCB may be performed to check for proper electrical
operation. Test points are noted below.

WARNING

USE CAUTION WHEN CASE IS OPEN. 12VDC POWER PRESENT AT ALL TIMES.

TEST POINT voltage

TP1 flow / no flow 5v / 0v


U1 pin18 & pin19 ON >2 and <3 vdc
U5 pin1 12v / press ON 0v
U5 pin6 12v / press OFF 0v
TP2 AC operation 15 to 18v
TP3 13.8v
TP4 ground
TP5 13.2v
TP6 13v
TP7 13 nominal
TP8 4.9 to 5.0vdc

CONTROLLED
THEORY OF OPERATION
--------------------

CONTROLLED
THEORY OF OPERATION
--------------------

CONTROLLED
THEORY OF OPERATION
--------------------

GENERAL

The basic operating Modes are:

1. Hold Mode (all keys functional)


2. Run Mode (only RUN/STOP and SET keys functional)
3. Alarm Mode (only RUN/STOP functional)
4. Test Mode (entered from Hold Mode, must turn pump off to exit)

The OFF key is functional in all modes.

SOFTWARE DESCRIPTION

At power-on, the microcontroller recalls the previous values of Infusion Rate, Dose Limit, Volume
Delivered, and DOSE CHECK. This allows quick set- up for repeated operation and prevents loss of values if
OFF is inadvertently pressed. After displaying the initial values, a single beep of the audible alarm is
performed as a safety check to insure it is functional.

After entering the Hold Mode, an alarm will occur after 2½ minutes of inactivity (no key presses). This is
used as a reminder to the operator that the pump has not been started. The 2½ minute timer is reset after
each key press while in Hold Mode.

Clearing of Volume Delivered is achieved by pressing CLEAR (“cLr” displayed) while viewing the value.
The pump will beep once, wait 1½ seconds, then, if still pressed, a double beep occurs and the value is
cleared to 0. Additionally, Volume Delivered, Dose Limit, and Infusion Rate will all clear if CLEAR is held
for four more beeps (approximately 6 seconds). The word "ALL" appears in the left display. After a final 1½
sec, another dual beep and all parameters EXCEPT DOSE CHECK will be cleared to their lowest allowed
value.

From Hold mode, pressing RUN/STOP once starts pump. The rotor will turn. The intermittent rotation of the
pump rotor has a constant period of 1.9sec, regardless of the selected Rate. Only the duty cycle is varied.
The speed of the rotor during an "on" cycle is always the same and corresponds to a theoretical rate of
354ml/hr. The duty cycle (percentage of "on" to "off" time) is changed to vary the actual delivery rate. At
295ml/hr the motor operates at 100% duty cycle, at 147ml/hr the duty cycle is 50%, etc. Accuracy of the
motor speed, and hence the theoretical fluid flow rate, is determined by the crystal-controlled microprocessor.

During each first milliliter delivered, the drop sensor is disabled to help avoid false alarms by allowing any
final priming of the feeding set. After the first 1ml, drop monitoring is activated via the photo IC / infrared LED.
After the initial 1ml allowance for priming, if the motor is driven 1925 steps without detecting a drop, an
Occ/Emp alarm is generated. The motor is stopped and the audible alarm is activated. This will occur if no

CONTROLLED
THEORY OF OPERATION
--------------------
drops fall to break the drop sensor light path. This is caused by either an empty feeding bag or a tubing
occlusion.

Free Flow alarms are generated by three different conditions: drop sensor blockage for 9 seconds (e.g.
fluid streaming), excessive drop rate, and falling drop detection during a periodic rotor pause. "FLO FASt"
appears in the displays, and a triple note/dual tone beep is generated always at maximum volume.

If a Dose Limit has been set while in Hold Mode, then the microprocessor continuously compares Volume
Delivered to the set limit. When Volume Delivered equals the Dose Limit, a Dose Complete Alarm
occurs, where the alarm beeps two times every 2½ minutes. The pump remains idle, flashing the Dose
Comp LED. Entering Run Mode with Dose Limit set to a value less than the current Volume Delivered will
cause an immediate Dose Complete Alarm.

While operating on battery, voltage is monitored during all pump modes and when the voltage drops below a
nominal 11.2VDC, the Bat LED flashes.
Because the battery voltage drops slightly when the motor is on, the voltage level is valid only during motor
"off" cycles in the pumping mode. The Bat LED begins flashing almost immediately after detecting low charge
voltage, and remains on for approximately 15 minutes before generating a single-tone, beeping alarm.
Pumping stops during a Low Battery Alarm. Note that the 15-minute timer re-starts if the pump is turned off
then back on, which could cause battery charge depletion before alarming.

If SET is pressed during Run, the current Dose Limit value is displayed for approximately 2
seconds on the Volume Delivered/Dose Limit display. If RUN/STOP is pressed, pumping stops and the
pump enters Hold Mode. If any other key is pressed during Run a beep will occur signaling an invalid key
press.

In any Alarm Mode the microprocessor stops the motor, flashes both displays, and beeps the alarm at a rate
and frequency dependent upon alarm type. When RUN/STOP is pressed, the audible alarm stops, and the
pump enters the Hold Mode. Any indicator LEDs that were lit as a result of the alarm condition remain lit as a
reminder of what caused the alarm.
Pressing RUN/STOP (re-starting the motor) will cause any of the alarm LEDs to be extinguished.

To control the LEDs and numerical displays, microcontroller U1 sends commands to the display controller
U12. The display controller also modulates the display brightness. The brightness level can be reduced by
pressing and holding UP and SET simultaneously until display dims. The default level is high when power
is cycled off then on.

CONTROLLED
THEORY OF OPERATION
--------------------

HARDWARE DESCRIPTION

Primary Power for the pump is provided by either AC mains voltage or the internal sealed lead-
acid(SLA)battery. For AC power mode, the mains voltage is reduced to a nominal 18 volts DC. This regulated
DC voltage is used both to power the pump and charge the battery. When the wall power supply is energized
by plugging in the AC mains, regulated DC voltage is divided by R6/R7 and presents a HI to the
microcontroller U1, indicating that AC voltage is present. This signal is used by U1 to control the Bat LED. V+
is logically OR'ed by D4 and D5 which permits operation from the line voltage, if present, or the battery. If line
voltage is present D4 is forward biased allowing a charge current to flow into the battery. When the input line
voltage is removed, D4 is reverse biased and all current from the battery flows through D5. Thus there is
always uninterrupted DC power for pump operation. V+ is always present at the supply switching latch, U5.

The rechargeable lead-acid battery is a nominal 12 volt/1.2 to 1.3 amp- hour unit and is fused at 2 amps
for board protection by F1, which is solder-mounted to the PC board. The battery is charged by precision
regulator U10 through D5. As the voltage level on the battery increases with increasing charge, U10
charging current diminishes. At full battery charge only 20-30 milliamps to flow to the battery. The battery
voltage can range from near 0VDC (loaded, fully discharged) to approximately 13.5VDC (fully charged).
U10 charging current is self-limiting to a maximum of approximately 250 mA. The 1% resistors, R15 and
R17, accurately set U10 to the nominal 13.5VDC charge voltage.

To monitor a low battery condition, the U14-switched supply voltage is scaled down by R20/R21 and applied
to voltage detector IC, U13. The result is U13 output goes HI when its input drops below 4V. This
corresponds to a battery voltage of 11.2V. The HI level is detected by microcontroller U1 which flashes the
Bat LED and beeps alarm.

The power control R-S latch U5 is always powered, either by AC line or directly from the battery. U5
presents an ultra-low current drain to the battery, and therefore is not a factor in determining battery
operating time. When the pump is in the off state, the output U5-A of the power control latch is HI which
turns off transistor Q4. When the ON key is pressed, U5-A goes LO, turning on Q1 and allows current to
flow into the
+5VDC regulator U14. The +5V powers the pump circuit. Power-on also generates a relatively slowly rising
voltage to the microcontroller RESET line via C3. The forces a normal start sequence for the pump
software.

The microcontroller U1 controls all actions of the pump. It monitors inputs from the keypad, drop
sensor, battery voltage, and ac on/off. It controls outputs including driving the stepper motor, generating
the audible alarms, illuminating LEDs and displaying alpha and numerical information. It executes
instructions fetched from EPROM memory, U3.
Addresses are provided via multiplexed data/address lines DB0-7, and latched address lines A8-A11.
The low-order 8 bits of all addresses are latched by U2. Instructions are then read in on DB0-7.

CONTROLLED
THEORY OF OPERATION
--------------------
Unipolar stepper motor drive is delivered to the four motor phases by current-sink drivers U8 and U9. Vs
(+12v)00 is always applied to one side of each motor winding. Windings are then activated two at a time in a
four step sequence by active LO's from drivers, controlled in turn by active HI's from U1.

Non-volatile memory data storage is accomplished by EEPROM. On shut-down (power off) microcontroller
U1 serially transmits all memory parameters to EEPROM, U6. At power on, microcontroller fetches
parameters from EEPROM, U6.

The drop sensor consists of an infrared emitter LED and light modulating photo detector IC. The light
modulation photo IC consists of an oscillator, a timing generator, an LED driver, a photodiode, a
preamplifier, a comparator, a signal processing circuit and an output circuit. All functions are integrated on
a monolithic chip. Optical synchronous detection is achieved by connecting the infrared LED to the photo
IC. In optical synchronous detection, signal light is modulated in pulses and detected in synchronization
with the modulation timing. Noise generated by background light is eliminated. When the light path is
obstructed by falling liquid drops, P3-3 connected to U1-13 goes to a logic HI. It remains LO when the path
is clear.

The audio alarm consists a piezo buzzer driven by the combination of U11, L1, C16. The frequency of the
audio tone is generated by the microcontroller. A logic level signal to U11-2 sets the alarm volume to either
high or low level. The default volume level for annunciation is low volume, and all alarms conditions are at
high level.

Keyboard functions are buffered by U4 which sends information along databus DB0-DB7, with timing
controlled by Microcontroller U1. U4 also detects active LO signal KSCAN to "read" the status of the
keyboard. When one of the front panel keys is pressed, the KSCAN signal will appear at one of U4's
keyboard inputs via shorting action of the key, thus allowing U1 to decode which key was pressed.

The displays consist of seven LED segment digits and four (4) discrete LEDs. The eight digit display driver,
U12, decodes serial data from U1 and causes appropriate segments or LEDs to illuminate. The discrete LEDs
are treated as the eighth digit with respect to the driver U12. Display drive current is multiplexed to reduce
overall drive current and pin count. The maximum segment current is controlled by R22. U12 also allows
programming of drive currents that are less than the maximum. This feature has been utilized as a brightness
control. Pressing and holding the UP and SET simultaneously toggles the display brightness.

CONTROLLED
TROUBLESHOOTING

The following table is provided to aid in correcting potential operational problems.

Additional troubleshooting information is provided in the FunctionalTest section.

WARNING

USE CAUTION WHEN CASE IS OPEN. 12VDC POWER PRESENT AT ALL TIMES.

Symptoms Possible Cause Solution

Occ/Emp alarm Pump is on hold for Press RUN/STOP


more than 2-1/2 minutes. switch to return to hold
mode

Feeding container is empty Replace or refill feeding


container

Occlusion is restricting flow Identify point of occlusion

Pump set:
Closed tubing clamp Kinked
tubing Blocked tubing
FLO FASt alarm Drip chamber walls Invert to clear drip chamber coated with
formula Replace pump set

Tubing set not Assemble properly


properly placed in pump

Too much fluid in Invert container and drip


chamber clear fluid from drip
chamber

Drop sensor parts Clean with cotton


need cleaning swab dampened with alcohol or warm
soapy water

Drop sensor parts Parts must be snapped into not in


place place on their respective
sides accessible from inside the
pump case, so
that optical path is aligned

CONTROLLED
TROUBLESHOOTING

Bat LED flashing Battery discharged Charge battery


by connecting to AC mains

Poor harness connections Check connection at battery

Battery duration below Batteries that have been expected


performance left in discharged state
will require 2-3 charge / discharge cycles
to resume operation at rated capacity.

Batteries are normally good for at least

1) 200 full discharges, or


2) 300+partial discharges,
whichever comes first

Bat LED always on.

Poor wire harness Check and repair P4-1


connections or P4-8

Rate change alarm UP, Down, or CLEAR Press RUN/STOP once


activated pressed with pump in to stop pump, before
hold mode pressing other switches.

Dose Complete Alarm Pump has delivered Press RUN/STOP, Press CLEAR
preset volume Press RUN/STOP Pump
will operate until dose
is completed again.

AC operation only, Battery end of replace battery


charging no effect life

F1 (on Pcboard) blown Replace F1 with 2amp fuse Poor

wire harness Check and repair


connections. P4 pins 6,8 or battery
connectors

No power No AC Check AC outlet (no


displays)

External power supply Check that External power not


attached to pump supply is fully seated in
the DC jack.

External power supply Check that 18VDC is present Open or


sorted. at power supply DC jack when
connected to AC.

CONTROLLED
TROUBLESHOOTING

Keyboard connector not Check P1 connection


attached to PCB. (inside case)

Bad on/off latch U5 Check for LO/HI toggling when


ON/OFF keys pressed

Poor harness connections Check and repair P4,


pins 1
or 8

F1 blown on pc board Replace F1 with 2amp fuse

No audible alarm Poor harness connections Check and repair P4,


pins 8 and 9.

Motor hesitates or fails to Bent shaft, rotor rubbing Replace motor


turn against case

Poor harness Check and repair P3,


connections pins 5-12.

Bad driver IC Check and replace U8 or U9

Under delivery Rotor rollers sticking or loose Clean rotor and rollers and tighten
set screw set screw

Rotor misaligned Align rotor with a 0.125” Feeler


Gauge between rotor and case

CONTROLLED
APPENDIX A

Specifications Pump Block


Diagram
Mechanical Assembly
Spare Parts List Test Loop
Assembly

ATTACHMENTS

Service Procedures (latest revision)


113-24-002 QC Work Instruction – SENTINELplus
113-28-020 Service Operating Procedure
113-28-021 SENTINELplus Quick Check

OMI-239 Delivery Test


113-28-001 Servicing – Case Replacement, Front
113-28-002 Servicing – Case Replacement, Rear
113-28-014 Troubleshooting Power Problems
113-28-003 Servicing – Battery, S-3000-IU
113-28-004 Servicing – Pole Clamp

113-28-005 Servicing – Rotor


113-28-006 Servicing – EPROM

113-28-007 Servicing – Drip Sensor


113-28-008 Servicing – PC Board

113-28-009 Servicing – Motor


113-28-011 Servicing – Labels
113-28-010 Servicing – Keypad
OMI-277 Troubleshooting Power Problems

Service Forms (latest revision)

OM-4014 Pump Work Sheet


OM-4112 Pump Delivery Test Form

CONTROLLED
APPENDIX A
SPECIFICATIONS

Type of Device............Volumetric enteral feeding pump Mechanism.................Rotary peristaltic

Operating Rates...........5-295 ml/hr in 1 ml/hr increments

Selectable Dose Limit.....1-1999 in 1ml increments, 2000-9999 ml in 5 ml increments accuracy..................±10% of flow

rate/hr [with ALCOR sets]

Occlusion Pressure........15 psi, maximum

Battery...................Sealed lead-acid, rechargeable.


Automatically recharges when unit is plugged in. Recharge time from complete discharge to full
charge is 12 hours with pump turned off.

Self-discharge............Battery should be recharged every six months when not in use.

Minimum Operating Time....5 hours @ 295ml/hr new


battery, full charge 6 hours @ 125ml/hr
8 hours @ 25ml/hr
When the pump is first entered into service or allowed to self-discharge, a charge/discharge/charge cycle is
necessary to test minimum operating time.

Dimensions................15cm H x 20cm W x 9cm D

Weight....................1.42kg

Case Material.............plastic alloy Power

Requirement.........18VDC, 1.33 amp

ALCOR Power Supply .......100 to 240VAC, 400-200ma, 50-60Hz

Fuse .....................Battery: 2A fast-acting PCB mount. Power Supply Plug.........all common

international plug adapters

CONTROLLED
Pump Block Diagram

CONTROLLED 23 of 27
SENTINELplus mechanical assembly

CONTROLLED 24 of 27
Service Spare Parts
SENTINELplus

P/N Description Reference


000197 #8-32 x 5/16"L Phil PH SS used on Case Body and Battery Bracket
000196 #4 x 3/8"L Phil PH Thread Cutting SS used on PC Board Case Body
000194 Retaining E-Ring, 3/8" used on Pole Clamp
113-01-009 Washer, fit .415"ID, .812"OD, .0625thk SS used on Pole Clamp
113-08-004 Knob/Shaft Assembly, Long used on Pole Clamp
113-08-007 Pole Clamp Assembly used on Pole Clamp
000207 Ferrite, Keypad Flex Cable Used on front case
113-02-001 Keypad SENTINELplus Used on front case
113-02-002 PC Board, S-3000 IU main circuit board / with EPROM
000193 EPROM programmed used on PC Board
113-02-004 Motor / Optical Harness Used on front case
113-02-005 Power Supply Used with pump
113-02-003 DC Power Jack – Harness with Piezo Used on rear case
113-02-006 PC Board, Optical Sensor Used on front case
000195 Battery, 12V 12V - 1.2/1.3 AH, sealed lead-acid
113-01-003 Plate Battery Retainer holds battery in rear case
113-01-001 Case, Front shell only
113-01-002 Case, Rear shell only
113-08-006 Rotor Sub Assembly Used on front case
000198 Set Screw holds rotor to motor shaft

CONTROLLED 25 of 27
TEST LOOP ASSEMBLY

1 2

McMaster 5463K38

CONTROLLED 26 of 27
REV DESCRIPTION DATE
A 113-28-001 SENTINELplus Service Manual May, 2017

CONTROLLED 27 of 27

You might also like