3.9 Microscope
3.9 Microscope
A second lens near the eye called the eyepiece (ocular) enlarges the
primary image converting it into one that can enter the pupil of the eye.
The higher the resolving power of an objective, the closer can be the fine
lines or small dots in the specimen which the objective can separate in
the image.
lens (n) and the sine of half the angle of the cone of light absorbed by the
front lens of the objective (u) or
Is a number that expresses the ability of a lens to resolve fine detail in an
When a beam of light passes from air into glass it is bent and when
it passes back from glass to air it is bent back again to its original
direction.
This has effect on oil immersion objective and affects the NA of the
objective and consequently it’s resolving power.
Fluorescence microscope
is a light compound
microscope, which is routinely
used in medical labs of
hospitals and/or health centers
the colored specimens will have a refractive index that will differentiate
it from the surrounding, presenting a combination of absorption and
refractive contrast
needs a strong light source for magnification and sometimes the light
source may produce a lot of heat which may damage or kill the specimen
the dark field microscope removes the dispersed light so that only the
scattered beams hit the sample
the only light entering the eye comes from the micro-organisms
themselves, no light entering the eye directly from light source, the mo
are seen brightly illuminated against a black background, like stars in a
night sky.
is well suited for uses involving live and unstained biological samples,
such as
Treponema palladium; Borreliae in blood; Microfilariae in blood
the techniques are almost entirely free of artifacts, due to the nature of
the process.
The main limitation is the low light levels seen in the final image
The sample must be very strongly illuminated, which can cause damage
to the sample.
In a phase-contrast microscope,
these phase shifts are converted
into changes in amplitude, which
can be observed as differences
in image contrast
Excitation filter
Dichroic mirror
Dichroic mirror
Emission filter
Objective lens
Ocular lens
Sample
Detector
The emitted light is then focused to the detector by the ocular lens
29 Introduction to MLT compiled by Waqtola C
Advantages and limitations
Advantages Limitations
high energy accelerating voltage (about 20kv) is used to focus the electrons
into a straight line passing through the sample (ultra thin, 200 times thinner
than those used in optical microscope) in air tight vacuum chamber.
Heavy atoms in the sample absorb more electrons but light atoms absorb
less and thus transmit more; ie the result of this varying absorbance result
in image formation in black and white
Magnetic field helps to maintain the beam of electron in air tighter vacuum
chamber
33 Introduction to MLT compiled by Waqtola C
34 Introduction to MLT compiled by Waqtola C
Scanning EM (SEM)
Help to visualize the sample in 3D shape and details of the cellular structures
with greater magnification and resolution
Little difference from TEM in that anode and scan coil exists. Anode is situated
between the source and the condenser whereas scan coil is b/n the condenser
and objective lens.
Once the beam hits the sample then scattered electron based on the difference
of depth is diffracted and then detected by CMOS detector to form 3D image
more electrons escape from the higher surface of the sample versus low electron
escape from lower side of the surface of the sample creates differential image which is
then captured by the CMOS sensor and projected as a 3D image, i.e., presence of the
different surface components of the sample
35 Introduction to MLT compiled by Waqtola C
Applications of EM
used to investigate the ultrastructure of a wide range of biological and
inorganic specimens including microorganisms, cells, large molecules,
biopsy samples, metals, and crystals.
Condenser adjustments: -
focused usually by rotating a knob to one side of it to move the
condenser up or down.
Objectives: -
components that magnify the image of the
specimen to form the primary image.
For most routine laboratory work 10x, 40x
and 100x (oil immersion) objectives are
adequate.
Eyepiece:-
the upper optical component that further
magnifies the primary image and brings the
light rays to a focus at the eye point.
Multi head microscope
It consists of two lenses mounted at the
correct distance.
It is available in a range of magnifications
usually of 10x, 15x and sometimes as high as
20x.
N.B: Based on their number of eyepiece,
microscopes can be classified as monocular,
binocular, trinocular, multi head microscopes
Electric light
An ordinary 60-watt pearl electric bulb placed about 18 inches from the
microscope is sufficient for most routine work.
Quartz halogen (quartz iodine) are very good light sources because they
give excellent white illumination and do not blacken like ordinary tungsten
lamps.
Many microscopes are now provided with correctly aligned built-in sources
of illumination, which use tungsten or quartz halogen lamps operating on
6,8 or 12 volts through variable transforms.
42 Introduction to MLT compiled by Waqtola C
Filters : Light filters are used in the microscope to:
Reduce the intensity of light
Adjust the color balance of the light to give the best visual effect
Absorb light
The user must be seated at the correct height for the convenient use of
the microscope.
3. Place the specimen on the stage, making sure that the underside of
the slide is completely dry.
10. For a higher magnification, swing the 40x objective into place.
Focus the 40x objective, using the fine adjustment.
If for any reason the image is not visible, lower the objective until it is
nearly but not quite touching the specimen.
Then looking through the eyepiece, focus up wards with the fine adjustment
until the image comes into view.
11. For the highest magnification, add a drop of immersion oil to the specimen
and swing the 100x oil immersion objective into place, then open the iris fully
to fill the objective with light. Example. Stained blood smear, acid-fast stain,
etc
49 Introduction to MLT compiled by Waqtola C
Care, Cleaning and Repair of microscope
The microscope is one of the most expensive and delicate instruments.
Good microscopy practice includes:
a) Using a clean cloth, wipe any dust from stage and other surfaces of
microscope
2. Check stage and under side of the specimen re DRY and CLEAN.
6. Always lift and carry the microscope well supported with hands.