Dpi 104 Is
Dpi 104 Is
Dpi 104 Is
Sensing
Features Applications
• Intrinsically safe • Process monitoring and control
• 0.05% full scale (FS) accuracy The DPI 104-IS is an intrinsically safe version of the
microprocessor-controlled DPI 104-digital pressure
• Pressure ranges to 20,000 psi (1400 bar) gauge. The DPI 104-IS matches advanced silicon sensor
technology with several convenient design features,
• Eleven selectable pressure units resulting in an accurate, versatile yet affordable digital
test gauge. Supplied as a stand-alone process indicator
or in a kit with the widely proven Druck hand pumps, the
• Large, easy-to-read display with five-digit resolution
DPI 104-IS provides a reliable and economic solution for
a wide range of pressure sensing applications.
• % pressure indication and additional bar graph for
quick visual reference
DPI 104-IS
Druck Digital
Test Gauge
DPI 104-IS is a Druck product.
Druck has joined other GE
high-technology sensing
businesses under a new name_
GE Sensing.
g
GE
Sensing
DPI 104-IS
Specifications
Accuracy
0.05% FS including non-linearity, hysteresis, repeatability Display Update
and temperature effects from 14°F to 122°F Two times per second
(-10°C to 50°C)
Minimum/Maximum Values
Resolution The minimum and maximum pressure values can
Maximum five digits be displayed on the DPI 104-IS. This function can be
enabled/disabled and reset by the user.
Pressure Ranges
Range Resolution Maximum Working Switch Test
Pressure The DPI 104-IS features a switch test function that will
psi bar type psi mbar psi bar
capture and display the open and closed values from an
0-10** 0-0.7 G* 0.001 0.01 11.2 0.77 external pressure switch. Maximum switch impedance
0-30 0-2 G* or A 0.001 0.1 32 2.2 200 Ω.
0-100 0-7 G* or A 0.01 0.1 111.7 7.7
Alarm Output
0-300 0-20 G* or A 0.01 1 319 22 The alarm output consists of an open drain field effect
0-1000 0-70 G* or A 0.1 1 1117 77 transistor (maximum current 22 mA, maximum voltage
16.9 V).
0-3000 0-200 SG 0.1 10 3190 220
Mechanical Vibration
Specifications To Def Stan 66-31, 8.4 Cat III
Mechanical Shock
EN 60079-0:2006
Battery
The DPI 104-IS is supplied with a 9V alkaline battery, Electrical Environments
type MN1604. Battery life is approximately four • EMC: EN61326:1997 + A1:1998 + A2:2001 + A3:2003
months when used daily for one hour per day. • Electrical Safety: EN61010-1:2001
• Mechanical Safety Pressure Equipment
Pressure Port Directive_Class: Sound Engineering Practice (SEP)
• 1/4 NPT or BSP male for units to 10,000 psi (700 bar)
• 9/16 x 18 UNF male cone connection 15,000 psi Options
(1000 bar) and 20,000 psi (1400 bar) units
Storage Temperature
-4°F to 158°F (-20°C to 70°C)
Dimensions
Diameter: 3.7 in (95 mm) excluding pressure fittings
Depth: 2.2 in (55 mm)
GE
Sensing
DPI 104-IS
Specifications
Pneumatic and Hydraulic Test Kits
The DPI 104-IS is included as a standard component in
these test and calibration kits:
Low pressure pneumatic test kit Pneumatic test kit
Low Pressure Pneumatic Test Kit
Includes DPI 104-IS; ranges to 30 psi (2 bar), PV 210
low pressure pneumatic test pump, hose, adaptors,
seal kit and case
g
©2007 GE. All rights reserved.
920-406C_E
All specifications are subject to change for product improvement without notice. GE® is
a registered trademark of General Electric Co. Intel® and Pentium® are registered
trademarks of Intel Corporation. Windows® is a registered trademark of Microsoft
Corporation. Intel and Microsoft are not affiliated with GE, in the U.S. and other www.gesensing.com
countries. Other company or product names mentioned in this document may be
trademarks or registered trademarks of their respective companies, which are not
affiliated with GE.