4-20ma Callibrator Manual

This document provides an instruction manual for a 4-20mA loop calibrator and thermometer. It details the features, applications, panel descriptions, operation instructions, scanning functions, specifications and maintenance of the device.

Uploaded by

jamie77mj
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
84 views40 pages

4-20ma Callibrator Manual

This document provides an instruction manual for a 4-20mA loop calibrator and thermometer. It details the features, applications, panel descriptions, operation instructions, scanning functions, specifications and maintenance of the device.

Uploaded by

jamie77mj
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 40

Instruction Manual

RS-135
4-20mA Loop Calibrator & Thermometer
4-20mA Loop Calibrator & Thermometer / EN

This unit passes the following tests:

EN 61326-1 : 2013

Safety Symbols

Please read the statement thoroughly to prevent injury or loss

of life, and prevent damage to this product.

Earth (ground)

DC (Direct Current)

Conforms to relevant European Union


directives.
4-20mA Loop Calibrator & Thermometer / EN

Do not dispose of this instrument as unsorted


municipal waste. Contact a qualified
recycler for disposal.

Please remove all the test leads


before performing maintenance,
cleaning and battery replacement,
etc.
4-20mA Loop Calibrator & Thermometer / EN
4-20mA Loop Calibrator & Thermometer / EN

Table of Contents

FEATURES: .............................................................................................................1

APPLICATIONS: .....................................................................................................1

I. PANEL DESCRIPTION.........................................................................................2

II. OPERATION ........................................................................................................7


1. VOLTAGE SOURCE .................................................................................................... 7
1a. 0~24V ................................................................................................................... 7
2. CURRENT SOURCE ................................................................................................... 9

2a. SOURCE (4~20mA, 0~24mA) .......................................................................... 9

2b. LED Brightness Testing (0~24mA) ................................................................ 10

3. FREQUENCY SOURCE (HZ)..................................................................................... 12


3a. 3Vpp, 1~20KHz, Square Wave ..................................................................... 12
3b. DUTY .................................................................................................................. 13
4. VOLTAGE MEASURE ............................................................................................... 14
4a. DC 0~24V .......................................................................................................... 14
5. CURRENT MEASURE .............................................................................................. 16
5a. DC 0.005~24mA ............................................................................................... 16
5b. LOOP+ MEASURE and Supply 24V Loop................................................. 18
6. TEMPERATURE MEASURE ...................................................................................... 19
6a. Measure TC Thermocouple Signals ........................................................... 19
6b. Select TYPE and Temperature units .......................................................... 20

III. SCANNING FOR SOURCE ..............................................................................21


1. VOLTAGE SCANNING FOR SOURCE ........................................................................ 21
1a. Functions of Voltage Scanning for SOURCE .......................................... 21
1b. Connections and LCD displays for Auto-Ramp and Auto-Step ......... 22
4-20mA Loop Calibrator & Thermometer / EN

2. CURRENT SCANNING FOR SOURCE ....................................................................... 24


2a. Functions of Current Scanning for SOURCE .......................................... 24
2b. Connections and LCD displays for Auto-Ramp and Auto-Step ......... 25

IV. ELECTRICAL SPECIFICATIONS ....................................................................27

V. GENERAL SPECIFICATIONS ..........................................................................30

VI. MAINTENANCE & CLEANING........................................................................31

LIMITED WARRANTY ...........................................................................................32


4-20mA Loop Calibrator & Thermometer / EN

FEATURES:

1. Source 4~20mA, 0~24mA with 1μA resolution (1KΩ load, 24V loop supply),
basic accuracy 0.025%.
2. Source 0~24.000V with 1mV resolution.
3. Source frequency 1~20000Hz square wave with programmable duty cycle
(1~99%).
4. Auto step and auto ramp for sourcing mA, V.
5. Easy fixed 25% up and down, or programmable percentage (1~99%) increase
and decrease (mA, V).
6. Measure DC mA with loop power (24V output).
7. Measure DC V.
8. Measure Temperature (°C, °F) of 11 types of thermocouples (K, J, E, T, R, S,
N, L, U, B and C).
9. Calibrator grade precision temperature measurement with 0.1C & 0.1F
resolution.
10. Detection of thermocouples disconnection.
11. Source and Measure at the same time.
12. Warning for overload, output open (mA) or short (V).
13. Short circuit protection for output.
14. Battery power indication (%) when turn on the calibrator.

APPLICATIONS:

1. Calibration of 4~20mA panel meters.


2. LED brightness testing (0~24mA).
3. Temperature calibration of panel meters or instruments for 11 types of
thermocouples.

06/04/2018 Version No. 1 1


4-20mA Loop Calibrator & Thermometer / EN

I. PANEL DESCRIPTION

1. LCD display 5. Output terminal (+)


2. Function buttons 6. Output terminal (-)
3. Temperature input socket 7. V, mA input terminal (+)
4. Output function sliding switch 8. Input terminal (-)

2 06/04/2018 Version No. 1


4-20mA Loop Calibrator & Thermometer / EN

1. “MEASURE” symbol appears when under MEASURE mode.

2. HOLD symbol appears when MEASURE data is held.

3. Symbol indicates Auto-power-off.

4. This symbol appears when sub-functions of MEASURE are being used.

5. Low battery symbol.

6. Negative symbol for MEASURE value.

7. Symbol indicates the MEASURE is for DC.

8. Temperature units for MEASURE.

9. Status of MEASURE function.

10. MEASURE value display.

06/04/2018 Version No. 1 3


4-20mA Loop Calibrator & Thermometer / EN

11. “SOURCE” symbol appears when under SOURCE mode.

12. This symbol appears when sub-functions of SOURCE are being used.

13. Symbol indicates % step function output.

14. Symbol indicates 0~100% ramp scanning of output.

15. Warning symbol for Output overload (or there is no connection on SOURCE
terminals).

16. Symbol indicates the type of thermocouples.

17. mA Current output range.

18. SOURCE value display.

19. Symbol indicates “duty cycle” of frequency output.

20. % for step output or duty cycle % of frequency output (SOURCE).

21. Value display for %, or Battery power indication, or Sourcing percentage.

22. Status of SOURCE function.

4 06/04/2018 Version No. 1


4-20mA Loop Calibrator & Thermometer / EN

Buttons Functions

Power on/off

SHIFT (black):
Select a range/function under MEASURE mode.
Hold the present value under MEASURE mode.
Keep pressing it and turn on the power at the same
time to disable Auto-power-off.

Switch button for selecting MEASURE or SOURCE or


MEASURE+SOURCE.
time to disable Auto-power-off.
SHIFT (blue): Select a function under SOURCE mode.

1. MEASURE mode: press SHIFT (black) and then


this button to enter DCV measure.
2. SOURCE mode: press it to enter number 1.

1. MEASURE mode: press SHIFT (black) and then


this button to enter DC mA measure.
2. SOURCE mode: press it to enter number 2.

1. MEASURE mode: press SHIFT (black) and then


this button to enter Temperature measure.
2. SOURCE mode: press it to enter number 3.

1. SOURCE mode (V, mA) : press SHIFT (blue) and


then this button to increase 25% (0%  25% 
50%  75%  100%).
2. SOURCE mode: press it to enter number 4.
1. SOURCE mode (V, mA) : press SHIFT (blue) and
then this button to decrease 25% (100%  75% 
50%  25%  0%).
2. SOURCE mode: press it to enter number 5.

06/04/2018 Version No. 1 5


4-20mA Loop Calibrator & Thermometer / EN

1. MEASURE temperature range: press SHIFT (black)


and then this button to select the type of
thermocouples (K, J, E, T, R, S, N, L, U, B, C).
2. SOURCE mode: press it to enter number 6.
1. SOURCE mode: press SHIFT (blue) and then this
button to perform an Auto-ramp sourcing of whole
range. Press any button to stop Auto-ramp
sourcing.
2. SOURCE mode: press it to enter number 7.
1. SOURCE mode: press SHIFT (blue) and then this
button to increase % of the range.
2. SOURCE mode: press it to enter number 8.
1. MEASURE mode: press SHIFT (black) and then
this button to select the temperature unit (℃ or
℉)
2. SOURCE mode: press SHIFT (blue) + this button to
enter DUTY cycle of output frequency.
3. SOURCE mode: press it to enter number 9.
1. SOURCE mode: press SHIFT (blue) and then this
button to decrease % of the range.
2. SOURCE mode: press it to enter number 0.
1. SOURCE mode: press SHIFT (blue) and then this
button to set up the increase/decrease % of the
range.
2. SOURCE mode: press it to enter decimal point.
1. SOURCE mode: press ENTER button after key in a
value.
2. SOURCE mode: press SHIFT (blue) and then this
button to enter Auto step sourcing; press them
again to enter Pause step sourcing.

6 06/04/2018 Version No. 1


4-20mA Loop Calibrator & Thermometer / EN

II. OPERATION

1. Voltage SOURCE

1a. 0~24V

1. Turn on the power. Turn the sliding switch to V.


2. Press M/S button to select SOURCE (output) mode.
3. Type in a voltage value (including decimal point); then press ENTER.
4. Connect the banana plugs of Test leads to SOURCE terminals (red to +,
black to -).
5. Then connect the other ends (prods or alligator clips) of Test leads to the
object for calibration.
6. Perform voltage SOURCE function (refer to the scanning operation).

06/04/2018 Version No. 1 7


4-20mA Loop Calibrator & Thermometer / EN

Remark:
1. The max. output is 24.000V.
2. Type in a voltage value (including decimal point), press ENTER, then
the calibrator will output this voltage value.
3. When the output value is <1 and >0, please type in “0.” first.

Warning:
1. Do not make an input with voltage potential or connect any charged
circuitry to SOURCE (terminals) to prevent from damaging the calibrator.
2. When there is a short circuit or overload at the output terminals, the
calibrator can not output the correct voltage. Please remove connecting
leads and check when there is a symbol of OL.

8 06/04/2018 Version No. 1


4-20mA Loop Calibrator & Thermometer / EN

2. Current SOURCE

2a. SOURCE (4~20mA, 0~24mA)

1. Turn on power. Turn sliding switch to mA range (4~20mA) or mA (24).


2. Press M/S button to select SOURCE (output) mode.
3. Type in a current value (including decimal point); then press ENTER.
4. Connect the banana plugs of Test leads to SOURCE terminals (red to +,
black to -).
5. Then connect the other ends (prods or alligator clips) of Test leads to the
object for calibration.
6. Perform current SOURCE function (refer to the scanning operation).

06/04/2018 Version No. 1 9


4-20mA Loop Calibrator & Thermometer / EN

2b. LED Brightness Testing (0~24mA)

1. Turn on power. Turn sliding switch to mA (24).


2. Press M/S button to select SOURCE (output) mode.
3. Type in a current value (including decimal point); then press ENTER.
4. Connect the banana plugs of Test leads to SOURCE terminals (red to +,
black to -).
5. Then connect the other ends (prods of alligator clips) of Test leads to
LED.
6. Perform current SOURCE function (refer to the scanning operation).

10 06/04/2018 Version No. 1


4-20mA Loop Calibrator & Thermometer / EN

Remark:
1. Users are allowed to type in max. 5 digits.
2. Type in a current value (including decimal point), press ENTER, then
the calibrator will output this current value.
3. When the output value is <1 and >0, please type in “0.” first.

Warning:
1. Do not make an input or connect any charged device to SOURCE
(terminals) to prevent from damaging the calibrator.
2. When there is an output open circuit or overload, the calibrator can not
output the correct current. Please remove connecting leads and check
when there is a symbol of OL.

06/04/2018 Version No. 1 11


4-20mA Loop Calibrator & Thermometer / EN

3. Frequency SOURCE (Hz)

3a. 3Vpp, 1~20KHz, Square Wave

1. Turn on power. Turn sliding switch to Hz range.


2. Press M/S button to select SOURCE (output) mode.

3. Press SHIFT (blue) and then to select DUTY% or HZ.


4. Type in a value (including decimal point) then press ENTER.
(If an unusual frequency value is typed in, the LCD will temporarily
display -- % which means the calculation inside takes a longer time.)
5. Connect the banana plugs of Test leads to SOURCE terminals (red to +,
black to -).
6. Then connect the other ends of Test leads to the object for calibration.

12 06/04/2018 Version No. 1


4-20mA Loop Calibrator & Thermometer / EN

Remark:
1. Users are allowed to type in max. 5 digits.
2. Type in each parameter (including decimal point), press ENTER, then
the calibrator will output the parameter values.
3. When the output value is <1 and >0, please type in “0.” first.

Warning:
1. Do not make an input or connect any charged device to SOURCE
(terminals) to prevent from damaging the calibrator.
2. When there is an output short circuit or overload, the calibrator can not
output the correct frequency.

3b. DUTY

DUTY: for a square wave, users can decide the band width of the positive
wave.

06/04/2018 Version No. 1 13


4-20mA Loop Calibrator & Thermometer / EN

4. Voltage MEASURE

4a. DC 0~24V
1. Turn on power.
2. Press M/S button to select MEASURE (input) mode.
3. Press SHIFT (black) and then to select MEASURE DCV mode.
4. Connect the banana plugs of Test leads to MEASURE terminals (red to
V, black to COM). Then connect the other ends (prods or alligator
clips) of Test leads to the object for measurement.
5. LCD will display the DCV measurement value.

14 06/04/2018 Version No. 1


4-20mA Loop Calibrator & Thermometer / EN

Remark:
The measurement result is 5-digit (including decimal point and negative
sign).

Warning:
Do not measure over DC 30V for MEASURE (terminals) to prevent from
damaging the calibrator.

06/04/2018 Version No. 1 15


4-20mA Loop Calibrator & Thermometer / EN

5. Current MEASURE

5a. DC 0.005~24mA

1. Turn on power.
2. Press M/S button to select MEASURE (input) mode.

3. Press SHIFT (black) and then to select MEASURE DCmA mode.


4. Connect the banana plugs of Test leads to MEASURE terminals (red to
mA, black to COM). Then connect the other ends (prods of alligator
clips) of Test leads to the object for measurement.
5. LCD will display the DCmA measurement value.

16 06/04/2018 Version No. 1


4-20mA Loop Calibrator & Thermometer / EN

Remark:
The measurement result is 5-digit (including decimal point and negative
sign).

Warning:
Do not make connections to MEASURE and TC at the same time. Always
connect to only one (MEASURE or TC) to prevent from damaging the
calibrator.

06/04/2018 Version No. 1 17


4-20mA Loop Calibrator & Thermometer / EN

5b. LOOP+ MEASURE and Supply 24V Loop

1. Turn on power. Turn sliding switch to mA (24) range.


2. Press M/S button to select MEASURE (input) + SOURCE (output)
mode.

3. Press SHIFT (black) and then to select MEASURE DCmA mode


and SOURCE mode (setup 24mA output) – Dual mode function.
4. On the top of LCD will display LOOP+ PWR.
5. Connect the banana plugs of Test leads to below terminals: + (positive)
to SOURCE “red” terminal; - (negative) to MEASURE “red” terminal.
6. Then connect the other ends (prods or alligator clips) of Test leads to
the transmitter.
7. LCD will display the DCmA value of the transmitter.

18 06/04/2018 Version No. 1


4-20mA Loop Calibrator & Thermometer / EN

6. Temperature MEASURE

6a. Measure TC Thermocouple Signals


(for Types K, J, E, T, R, S, N, L, U, B, C)

1. Turn on power.
2. Press M/S button to select MEASURE (input) mode.

3. Press SHIFT (black) and then to select MEASURE TC mode.


4. Connect the mini plug of Thermocouples to MEASURE TC sockets.
And then connect the other end to the object for measurement.
5. LCD will display the temperature measurement value.
6. Please refer to the next page for “Select TYPE and temperature units”.

06/04/2018 Version No. 1 19


4-20mA Loop Calibrator & Thermometer / EN

Remark:
The measurement result is 5-digit (including decimal point and negative
sign).

Warning:
1. The Input TC terminals are for measuring mV, do not measure over
30V for MEASURE (terminals) to prevent from damaging the
calibrator.
2. Do not make connections to MEASURE and TC at the same time.
Always connect to only one (MEASURE or TC) to prevent from
damaging the calibrator.

6b. Select TYPE and Temperature units

1. Press SHIFT (black) and then to select TC TYPE.

2. Press SHIFT (blue) and then to select ℃ or ℉.

20 06/04/2018 Version No. 1


4-20mA Loop Calibrator & Thermometer / EN

III. SCANNING FOR SOURCE

1. Voltage Scanning for SOURCE

1a. Functions of Voltage Scanning for SOURCE

Press SHIFT (blue) then the lower left of LCD will display various functions
(see the descriptions on below).

SHIFT
Functions
(blue) +
Auto-Ramp scanning
1% 2%.... 100% 99%.... 2% 1%.
Press again to HOLD/CONTINUE scanning.
Auto-Step scanning
0% 25% 50% 75% 100% ....(increase and
decrease progressively).
Press again to HOLD/CONTINUE scanning.
Press SHIFT (blue) and then this button once to
increase 25%.
Press SHIFT (blue) and then this button once to
decrease 25%.

Set up the increase/decrease percentage.

Manual Multi-step scanning:


increase progressively per the setup %.
Manual Multi-step scanning:
decrease progressively per the setup %.

06/04/2018 Version No. 1 21


4-20mA Loop Calibrator & Thermometer / EN

1b. Connections and LCD displays for Auto-Ramp and Auto-Step

Auto-Ramp Scanning

22 06/04/2018 Version No. 1


4-20mA Loop Calibrator & Thermometer / EN

Auto-Step Scanning

06/04/2018 Version No. 1 23


4-20mA Loop Calibrator & Thermometer / EN

2. Current Scanning for SOURCE

2a. Functions of Current Scanning for SOURCE

Press SHIFT (blue) then the lower left of LCD will display various
functions (see the descriptions on below).

SHIFT
Functions
(blue) +
Auto-Ramp scanning
1% 2%.... 100% 99%.... 2% 1%.
Press again to HOLD/CONTINUE scanning.
Auto-Step scanning
0% 25% 50% 75% 100% ....(increase and
decrease progressively).
Press again to HOLD/CONTINUE scanning.
Press SHIFT (blue) and then this button once to
increase 25%.
Press SHIFT (blue) and then this button once to
decrease 25%.

Set up the increase/decrease percentage.

Manual Multi-step scanning:


increase progressively per the setup %.
Manual Multi-step scanning:
decrease progressively per the setup %.

24 06/04/2018 Version No. 1


4-20mA Loop Calibrator & Thermometer / EN

2b. Connections and LCD displays for Auto-Ramp and Auto-Step

Auto-Ramp Scanning

06/04/2018 Version No. 1 25


4-20mA Loop Calibrator & Thermometer / EN

Auto-Step Scanning

26 06/04/2018 Version No. 1


4-20mA Loop Calibrator & Thermometer / EN

IV. ELECTRICAL SPECIFICATIONS


(23+/- 5℃, 10 minutes after turning on the power)

DCmA SOURCE
(Max. load 1K, Max. voltage supply 24V)
Range Resolution Accuracy
0~4mA 1A ± 0.025% ±10A
4~20mA 1A ± 0.025% ±5A
20~24mA 1A ± 0.025% ±5A
When output open, the LCD displays “OL”.

DCV SOURCE
(Max. load 24mA, Short circuit protection<25mA)
Range Resolution Accuracy
0~4V 1mV ± 0.05% ± 10mV
4~20V 1mV ± 0.05% ±5mV
20~24V 1mV ± 0.05% ±5mV
When output short circuit, the LCD displays “OL”.

Frequency SOURCE
(TTL, Square wave, Duty cycle = 1~99%)
Range (Hz) Resolution Accuracy
1.0~1000.0Hz 0.1Hz 0.1Hz
1000~10000Hz 1Hz 1Hz
10000~20000Hz 1Hz 10Hz

DCmA MEASURE
Range Resolution Accuracy
0.005~4mA 1uA ±0.025%±5dgts
4~20mA 1uA ±0.025%±5dgts
20~24mA 1uA ±0.025%±5dgts
If reading of mA (measure) is less than 5 digits, it is displayed as 0.

06/04/2018 Version No. 1 27


4-20mA Loop Calibrator & Thermometer / EN

DCV MEASURE
Range Resolution Accuracy
0~24V 0.001V ±0.025%±5dgts
If reading of V (measure) is less than 5 digits, it is displayed as 0.

28 06/04/2018 Version No. 1


4-20mA Loop Calibrator & Thermometer / EN

Temperature of Thermocouples
(MEASURE only, 0.1C & 0.1F Resolution, Internal Cold Junction Compensation,
thermocouples accuracy not included, 3 minutes after plugging in thermocouples.)

°C °F
Range Accuracy Range Accuracy
K -200 to -150 2.0 -382 to -238 3.6
-150 to 0 1.2 -238 to 32 2.1
0 to 1000 0.8 32 to 1832 1.4
1000 to 1370 1.2 1832 to 2498 2.1
J -200 to -150 2.0 -382 to -238 3.6
-150 to 0 1.0 -238 to 32 1.8
0 to 1050 0.7 32 to 1922 1.2
E -200 to -150 1.5 -382 to -238 2.7
-150 to 0 0.9 -238 to 32 1.6
0 to 850 0.7 32 to 1562 1.2
T -200 to -150 1.5 -382 to -238 2.7
-150 to 0 1.2 -238 to 32 2.1
0 to 400 0.8 32 to 752 1.4
R 0 to 500 1.8 32 to 932 3.2
500 to 1760 1.5 932 to 3200 2.7
S 0 to 500 1.8 32 to 932 3.2
500 to 1760 1.5 932 to 3200 2.7
N -200 to 0 1.5 -328 to 32 2.7
0 to 1300 0.9 32 to 2372 1.6
L -200 to 0 0.9 -328 to 32 1.6
0 to 900 0.7 32 to 1652 1.2
U -200 to 0 1.1 -328 to 32 1.9
0 to 600 0.7 32 to 1112 1.2
B 600 to 800 2.2 1112 to 1472 3.9
800 to 1000 1.8 1472 to 1832 3.2
1000 to 1820 1.4 1832 to 3308 2.5
C 0 to 1800 1.0 32 to 3272 1.8
1800 to 2310 1.5 3272 to 4190 2.7

06/04/2018 Version No. 1 29


4-20mA Loop Calibrator & Thermometer / EN

V. GENERAL SPECIFICATIONS

Dimension 214.0 (L) x 98.7(W) x 56.0(H) mm

8.4" (L) x 3.9" (W) x 2.2" (H)

Weight 650g / 22.9oz

(batteries included)

Operation Environment 0℃ ~ 50℃ , < 85% RH

Storage Environment -20℃ ~ 60℃ , < 75% RH

Accessories Carrying case x 1

User manual x 1

1.5V SUM-3 Battery x 5


K-type thermocouples (1M, single plug) x 1

Alligator clips x 2 (black and red)

Test leads x 2 (black and red)

30 06/04/2018 Version No. 1


4-20mA Loop Calibrator & Thermometer / EN

VI. MAINTENANCE & CLEANING

1. Servicing not covered in this manual should only be performed by qualified


personnel. Repairs should only be performed by qualified personnel.

2. Periodically wipe the case and cable with a damp cloth and detergent; do not
use abrasives or solvents.

3. Please remove the batteries if users won’t use the meter for a long time.

Address of Agent, Distributor, Importer, or Manufacturer

06/04/2018 Version No. 1 31


4-20mA Loop Calibrator & Thermometer / EN

Limited Warranty

This meter is warranted to the original purchaser against defects in


material and workmanship for 3 years from the date of purchase. During
this warranty period, RS Components will, at its option, replace or repair
the defective unit, subject to verification of the defect or malfunction. This
warranty does not cover fuses, disposable batteries, or damage from
abuse, neglect, accident, unauthorized repair, alteration, contamination, or
abnormal conditions of operation or handling.
Any implied warranties arising out of the sale of this product, including but
not limited to implied warranties of merchantability and fitness for a
particular purpose, are limited to the above. RS Components shall not be
liable for loss of use of the instrument or other incidental or consequential
damages, expenses, or economic loss, or for any claim or claims for such
damage, expense or economic loss. Some states or countries laws vary,
so the above limitations or exclusions may not apply to you. For full terms
and conditions, refer to the RS website.

32 06/04/2018 Version No. 1


4-20mA Loop Calibrator & Thermometer / EN

06/04/2018 Version No. 1 33


4-20mA Loop Calibrator & Thermometer / EN

Africa
RS Components SA
P.O. Box 12182,
Vorna Valley, 1686
20 Indianapolis Street,
Kyalami Business Park,
Kyalami, Midrand
South Africa
www.rs-components.com
Asia
RS Components Pte Ltd.
31 Tech Park Crescent
Singapore 638040
www.rs-components.com
China
RS Components Ltd.
Suite 23 A-C
East Sea Business Centre
Phase 2
No. 618 Yan'an Eastern Road
Shanghai, 200001
China
www.rs-components.com
Europe
RS Components Ltd.
PO Box 99, Corby,
Northants.
NN17 9RS
United Kingdom
www.rs-components.com
Japan
RS Components Ltd.
West Tower (12th Floor),
Yokohama Business Park,
134 Godocho, Hodogaya,
Yokohama, Kanagawa 240-0005
Japan
www.rs-components.com
U.S.A
Allied Electronics
7151 Jack Newell Blvd. S.
Fort Worth,
Texas 76118
U.S.A.
www.alliedelec.com
South America
RS Componentes Limitada
Av. Pdte. Eduardo Frei M. 6001-71
Centro Empresas El Cortijo
Conchali, Santiago, Chile
www.rs-components.com

34 06/04/2018 Version No. 1

You might also like