Manual Fluke 1521
Manual Fluke 1521
1521
Handheld Thermometer Readout
User’s Guide
Each product from Fluke's Hart Scientific Division ("Hart") is warranted to be free from defects in mate-
rial and workmanship under normal use and service. The warranty period is one year for the Handheld
Thermometer Readout. The warranty period begins on the date of the shipment. Parts, product repairs,
and services are warranted for 90 days. The warranty extends only to the original buyer or end-user cus-
tomer of a Hart authorized reseller, and does not apply to fuses, disposable batteries or to any other prod-
uct which, in Hart's opinion, has been misused, altered, neglected, or damaged by accident or abnormal
conditions of operation or handling. Hart warrants that software will operate substantially in accordance
with its functional specifications for 90 days and that it has been properly recorded on non-defective me-
dia. Hart does not warrant that software will be error free or operate without interruption. Hart does not
warrant calibrations on Handheld Thermometer Readouts.
Hart authorized resellers shall extend this warranty on new and unused products to end-user customers
only but have no authority to extend a greater or different warranty on behalf of Hart. Warranty support is
available if product is purchased through a Hart authorized sales outlet or Buyer has paid the applicable
international price. Hart reserves the right to invoice Buyer for importation costs of repairs/replacement
parts when product purchased in one country is submitted for repair in another country.
Hart's warranty obligation is limited, at Hart's option, to refund of the purchase price, free of charge re-
pair, or replacement of a defective product which is returned to a Hart authorized service center within
the warranty period.
To obtain warranty service, contact your nearest Hart authorized service center or send the product, with
a description of the difficulty, postage, and insurance prepaid (FOB Destination), to the nearest Hart au-
thorized service center. Hart assumes no risk for damage in transit. Following warranty repair, the prod-
uct will be returned to Buyer, transportation prepaid (FOB Destination). If Hart determines that the
failure was caused by misuse, alteration, accident or abnormal condition or operation or handling, Hart
will provide an estimate or repair costs and obtain authorization before commencing the work. Following
repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for
the repair and return transportation charges (FOB Shipping Point).
THIS WARRANTY IS BUYER'S SOLE AND EXCULSIVE REMEDY AND IS IN LIEU OF ALL
OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IM-
PLIED WARRANTY OF MERCHANTABLILTY OR FITNESS FOR A PARTICULAR PURPOSE.
HART SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL. OR CONSE-
QUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHER ARISING FROM
BREACH OF WARRANTY OR BASED ON CONTRACT, TORT, RELIANCE OR ANY OTHER
THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or
limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not
apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court of com-
petent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
Rev. 571203
Table of Contents
2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
4 Quick Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.2 Use Proper Care with the 1521 and Accessories . . . . . . . . . . 11
4.3 Learn About the Features and Components . . . . . . . . . . . . . 11
4.4 Connect the Probe . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.5 Connect the Power Source . . . . . . . . . . . . . . . . . . . . . 12
4.6 Switch the Power On . . . . . . . . . . . . . . . . . . . . . . . . 12
4.7 Measure Temperature . . . . . . . . . . . . . . . . . . . . . . . . 12
6 General Operation . . . . . . . . . . . . . . . . . . . . . . . . 19
6.1 Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
6.2 DC Power Source . . . . . . . . . . . . . . . . . . . . . . . . . . 20
6.3 Power Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
6.4 Display and Backlight . . . . . . . . . . . . . . . . . . . . . . . . 21
6.5 Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
i
6.6 INFO-CON Connector . . . . . . . . . . . . . . . . . . . . . . . 23
7 Display Functions. . . . . . . . . . . . . . . . . . . . . . . . . 25
7.1 Blank Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
7.2 Min/Max Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
7.3 Hold Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
7.4 Delta(x) Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
7.5 Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
7.6 Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
7.7 Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
7.8 CAL MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
7.8.1 Mtr Due . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
7.8.2 Prb Due . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
7.8.3 Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
7.8.4 Passcode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
7.8.5 Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
7.8.6 Probe Lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
7.8.7 Prb #. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
7.8.8 Prb Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
7.8.9 ITS-90 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
7.8.10 IEC751 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
7.8.11 CVD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
7.8.12 YSI-400 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
7.8.13 THERM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
7.8.14 Probe Wires . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
7.8.15 PCal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
7.8.16 PDue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
7.8.17 Filter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
7.8.18 MCal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
7.8.19 MDue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
7.8.20 CAL1 and CAL2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
8 Communications Interface. . . . . . . . . . . . . . . . . . . . 37
8.1 RS-232 Connection . . . . . . . . . . . . . . . . . . . . . . . . . 37
8.2 Communication Command List . . . . . . . . . . . . . . . . . . . 38
8.2.1 Primary Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
8.2.2 Calibration Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
9 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
9.1 Required Equipment. . . . . . . . . . . . . . . . . . . . . . . . . 43
9.2 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . 43
10 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
ii
11 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . 47
11.1 An Error Message Is Displayed . . . . . . . . . . . . . . . . . . . 47
11.2 CE Comments . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
11.2.1 EMC Directive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
11.2.1.1 Immunity Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
11.2.1.2 Emission Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
11.2.2 Low Voltage Directive (Safety) . . . . . . . . . . . . . . . . . . . . . . . . . 49
iii
Figures and Tables
iv
1 Before You Start
Symbols Used
Symbol Description
AC (Alternating Current)
AC-DC
Battery
DC
Double Insulated
Electric Shock
Fuse
PE Ground
Off
On
1
1521 Handheld Thermometer Readout
User’s Guide
Symbol Description
Canadian Standards Association
1.2.1 Warnings
To avoid personal injury, follow these guidelines.
• DO NOT use this unit in environments other than those listed in the
User’s Guide.
• Follow all safety guidelines listed in the User’s Guide.
• Calibration equipment should only be used by trained personnel.
• If this equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment may be impaired.
• Before initial use, or after transport, or after storage in humid or semi-hu-
mid environments, or anytime the instrument has not been energized for
more than 10 days, the instrument needs to be energized for a "dry-out"
period of 2 hours before it can be assumed to meet all of the safety re-
quirements of the IEC 1010-1. If the product is wet or has been in a wet
environment, take necessary measures to remove moisture prior to apply-
ing power such as storage in a low humidity temperature chamber
operating at 50°C for 4 hours or more.
• This instrument can measure extreme temperatures. Precautions must be
taken to prevent personal injury or damage to objects. Probes may be ex-
tremely hot or cold. Cautiously handle probes to prevent personal injury.
2
1 Before You Start
Safety Information
1.2.2 Cautions
To avoid possible damage to the instrument, follow these guidelines.
• If the instrument is dropped, struck, or handled in a way that causes inter-
nal or external physical damage, immediately unplug the AC adapter, re-
move the batteries, discontinue use, and contact the factory for repair. Do
not attempt to disassemble or repair the instrument, batteries, or AC
adapter. Refer repairs or replacement components to the manufacturer.
• The instrument and thermometer probes are sensitive and can be easily
damaged. Always handle these devices with care. DO NOT allow them to
be dropped, struck, stressed, or overheated.
• DO NOT leave the AC adapter plugged in for more than 24 consecutive
hours or the battery life could be degraded.
3
1521 Handheld Thermometer Readout
User’s Guide
Phone: +1.801.763.1600
Telefax: +1.801.763.1010
E-mail: [email protected]
Phone: +31-402-675300
Telefax: +31-402-675321
E-mail: [email protected]
4
1 Before You Start
Authorized Service Centers
Phone: +86-10-6-512-3436
Telefax: +86-10-6-512-3437
E-mail: [email protected]
5
2 Introduction
2 Introduction
The 1521 Handheld Thermometer Readout is a low-cost, high-accuracy digital
thermometer readout designed to be used with PRTs and thermistors. The
unique combination of features makes this instrument suitable for a wide vari-
ety of applications in industry. Features and capabilities of the 1521 include the
following:
• Measures with platinum resistance thermometers (PRTs) and thermistors
• Works with Hart’s special INFO-CON probe connector (which is partially
based on U.S. Patent 5,857,777) to automatically recognize the type of
sensor and its characteristics
• Automatically alerts the operator when the probe calibration or meter cal-
ibration expires
• Measures with an accuracy of ±0.025°C at 25°C with PRTs and ±0.005°C
at 25°C with thermistors
• Accepts three- or four-wire sensors to eliminate lead resistance errors
• Is immune to thermoelectric EMF
• Measures with a fast one-second measurement cycle
• Displays maximum, minimum, and delta temperatures
• Stores up to six measurements in memory
• Uses a unique factory-assigned pass-code to protect programmed settings
• Displays measurements and settings on a high-contrast LCD display
• Communicates with other equipment using an RS-232 serial interface
• Can be powered from its AC adapter or rechargeable battery
7
3 Specifications and Environmental Conditions
Specifications
3.1 Specifications
Resistance Range 0Ω to 500 kΩ
9
1521 Handheld Thermometer Readout
User’s Guide
10
4 Quick Start
Unpacking
4 Quick Start
This section briefly explains the basics of setting up and operating your 1521
Thermometer.
4.1 Unpacking
Carefully unpack the 1521 and accessories and inspect them to make sure all
components are present and in satisfactory condition. Verify that the following
items are present:
• 1521 Thermometer
• AC Adapter (clamp-on ferrite installed)
• Serial Cable
• User’s Guide
• Report of Calibration
• Calibration Label
• Pass-code Notice
• INFO-CON Connector
• Clamp-on ferrite with instructions for probe
• Probe (optional—must be purchased separately)
If all items are not present, call an Authorized Service Center (see Section 1.3).
CAUTION: The 1521 and any thermometer probes are sensitive instru-
ments that can be easily damaged. Always handle these devices with care.
DO NOT allow them to be dropped, struck, stressed, or overheated.
11
1521 Handheld Thermometer Readout
User’s Guide
top of the instrument with the ridged side down. It will fit snugly and lock into
place when it is fully inserted.
The INFO-CON probe connector includes a memory device that stores the
unique characteristics of the probe, allowing the 1521 to measure tempera-
ture accurately. If your probe was purchased from Hart for use with the 1521 it
should already have the connector attached and properly programmed. You can
use your own probe with the 1521, but you must attach it with an INFO-CON
connector. See Section 6.6 for more information on the INFO-CON connector.
12
5 Parts and Controls
Front Features
13
1521 Handheld Thermometer Readout
User’s Guide
14
5 Parts and Controls
Top and Side Features
Side View
The top and side of the 1521 feature the probe connector, DC power input, se-
rial port, and infrared window.
Probe Connector - At the top of the thermometer is the opening where the
probe connector is inserted. The probe must be connected using an INFO-CON
probe connector to measure temperature (see Section 6.6).
DC Power Input-The AC adapter plugs into the DC power input to recharge
the battery and to power the instrument while the battery is being charged (see
Section 6.2).
Serial Port-The RS-232 serial port provides a means of connecting the 1521 to
a computer or a printer using the included serial cable (see Section 8). The
baud rate is fixed at 2400 baud, the linefeed is fixed to ON (all carriage returns
are followed by a linefeed (ASCII decimal 10), and the duplex is fixed to Half
disabling the echo.
Infrared Window-The infrared window has no functionality.
15
1521 Handheld Thermometer Readout
User’s Guide
16
5 Parts and Controls
Accessories
5.5 Accessories
The 1521's accessories and their features are described here.
AC Adapter-The AC adapter recharges the internal battery pack and can also
be used to supply power to operate the 1521 while the battery is being charged
(see Section 6.2).
Serial Cable-The serial cable can be used to connect a computer or a printer to
the 1521 through its serial port (see Section 8).
INFO-CON Connector-Hart’s unique INFO-CON connector allows the probe
to be easily attached and detached from the 1521. It also contains a memory de-
vice that stores information about the probe and automatically transfers
this data to the 1521 when the probe is attached. This ensures that the set-
tings used to measure and calculate temperature always match the probe being
used (see Section 6.6).
IN
FO
-C
ON
Probe-The probe acts as the temperature sensor. Its resistance depends on its
temperature. The 1521 precisely measures the resistance of the probe and uses
the coefficients stored in the INFO-CON connector to calculate its temperature
(see Section 6.5). The following probe options are available.
17
1521 Handheld Thermometer Readout
User’s Guide
Ordering Information
Cal Uncertainty and Typical Drift
Model Type Range Size Repeatability (1 year)
5626 Pt-100 Ohm, 4-wire –200 to 660°C 1/4” x 12” or 15” ±0.015°C ±0.03°C
(6.35 x 305 or 381 mm)
5628 Pt-25 Ohm, 4-wire –200 to 660°C 1/4” x 12” or 15” ±0.015°C ±0.03°C
(6.35 x 305 or 381 mm)
5614 Pt-100 Ohm, 4-wire –200 to 450°C 1/4” x 12” (6.35 x 305 mm) ±0.025°C ±0.01°C
5613 Pt-100 Ohm, 4-wire –200 to 300°C 3/16” x 6” (4.76 x 152 mm) ±0.025°C ±0.01°C
5612 Pt-100 Ohm, 4-wire –200 to 450°C 3/16” x 9” (4.76 x 229 mm) ±0.025°C ±0.01°C
5627–6 Pt-100 Ohm, 4-wire –200 to 300°C 3/16” x 6” (4.76 x 152 mm) ±0.035°C ±0.13°C
5627–9 Pt-100 Ohm, 4-wire –200 to 450°C 3/16” x 9” (4.76 x 229 mm) ±0.035°C ±0.13°C
5627–12 Pt-100 Ohm, 4-wire –200 to 450°C 1/4” x 12” (6.35 x 305 mm) ±0.035°C ±0.13°C
5610-6 10K-ohm Thermistor 0 to 110°C 1/8” x 6” (3.2 x 152 mm) ±0.02°C ±0.01°C
5610-9 10K-ohm Thermistor 0 to 110°C 1/8” x 9” (3.2 x 229 mm) ±0.02°C ±0.01°C
5611 10K-ohm Thermistor 0 to 110°C 0.07” x 0.55” (1.8 x 14 mm) ±0.02°C ±0.01°C
5611T 10K-ohm Thermistor 0 to 110°C 0.11” x 1.1” (2.8 x 28 mm) ±0.02°C ±0.01°C
5640 4K-ohm Thermistor 0 to 60°C 1/4” x 9” (6.35 x 229 mm) ±0.0015°C ±0.005°C
LogWare Software - The Model 9934 LogWare software is a one channel data
acquisition software. This software can be used with your 1521 thermometer to
collect and graph data.
Carrying Cases - There are two types of cases available for your 1521 ther-
mometer. The Model 9318 case is a hard case for carrying the thermometer and
a 12" probe. The Model 9321 case is a soft case with a belt clip and a probe
slot.
18
6 General Operation
Battery
6 General Operation
This section explains the details of the operation of the 1521 with its compo-
nents and accessories.
6.1 Battery
The 1521 has a built-in nickel-metal-hydride battery pack that can power the
instrument for about 18 hours before needing to be recharged. The battery will
discharge much more quickly if the display backlight is used often. The battery
charge remaining or the message “External Power” is shown on the display
when the instrument is switched on. The battery percentages are approximate
and should be used as general guidelines when determining the length of
charge remaining.
The battery is recharged in situ (while in place) using the AC adapter that’s
provided. Plug the AC adapter into an AC socket of the appropriate voltage as
noted on the AC adapter (normally 115V, optionally 230V). Connect the DC
plug of the adapter into the DC input of the 1521 located on the right side. The
battery will be charged as necessary whether or not the instrument is switched
Battery cover
Battery
Battery plug
Battery connector
19
1521 Handheld Thermometer Readout
User’s Guide
on. The power control circuit inside the instrument manages battery charging
and will stop charging the battery automatically when it’s fully charged. It nor-
mally takes about four hours to fully charge the battery. The instrument can be
operated while the battery is being charged.
The battery pack can be used for a minimum of 500 charge-discharge cycles
before needing to be replaced. Replacement battery packs are available from
the manufacturer. The battery pack can be easily removed and replaced in the
field by following this procedure:
1. Power the 1521 off and unplug the AC adapter from the unit.
2. With the 1521 facing down, lift the stand to expose the battery cover (see
Figure 4). Remove the battery cover screw with a small Philip’s screw-
driver. Place the screw in a safe place so it won’t get lost. Remove the
battery cover to access the battery pack.
3. Pull the battery pack out. Note the orientation of the battery plug on
the connector. The new battery pack will be plugged in the same direc-
tion. Gently pull the battery plug off the connector.
4. Attach the polarized plug of the new battery pack onto the connector in
the same direction as before (see Figure 4 on page 19). Place the battery
pack in the battery compartment.
5. Close the battery cover and replace the screw.
6. Plug in the AC adapter and charge for a minimum of four hours for its
initial charge.
20
6 General Operation
Power Button
adapter supplied with the instrument. The DC power source plugs into the DC
power input on the right side of the 1521.
–
+
WARNING: The AC adapter has circuits with high voltages inside that
could present danger of electric shock or fire if exposed. If the AC adapter
is damaged in any way or becomes hot, discontinue its use immediately,
disconnect it from any AC supply, and have it replaced. Do not attempt to
open, repair, or continue using a damaged or defective AC adapter.
Function Action
Power on Press the power button for at least two seconds when the power is off
Power off Press the power button momentarily (1/2 second or less) when the power is on
Backlight on Press the power button for three seconds or more when the power is on and
the backlight is off
Backlight off Press the power button for three seconds or more when the backlight is on
21
1521 Handheld Thermometer Readout
User’s Guide
smaller type to the right of the measurement. The possible units are degrees
Celsius (C), resistance in ohms (Ω), degrees Fahrenheit (F), Kelvin (K), and
degrees Rankine (R). The units can be easily changed by placing the display in
the units setting mode using the MODE button (see Section 7.5).
The lower part of the display has various functions depending on the selected
mode. It has smaller alphanumeric characters. It can be used to display mini-
mum and maximum measurements, delta(x) measurements, or measurements
stored in memory. It is also used to view and set various operating parameters.
The mode is easily changed using the MODE button. (See details on the vari-
ous modes beginning with Section 7.)
The display has a backlight that can be switched on for better viewing in dim
light. The backlight is switched on and off by pressing and holding the power
button for three seconds (see Section 6.3 above). Note: The battery dis-
charges more quickly when the backlight is used.
6.5 Probe
The probe is used to sense temperature. The probe attaches to the 1521 using a
Hart INFO-CON probe connector that plugs into the top of the instrument. The
probe connector must be properly programmed with the correct charac-
teristics of the probe for measurements to be accurate (see Section 6.6).
The 1521 can be used with various types of PRT and thermistor probes:
• ITS-90 calibrated 25Ω or 100Ω PRT
• IEC-751 or DIN-43760 PRT (RTD)
• Callendar-Van Dusen calibrated 100Ω PRT
• YSI-400 series or equivalent 2252Ω thermistor
• Steinhart-Hart thermistor polynomial; nominal R(25°C) 2kΩ to 100kΩ
See Section 7.8.8 for details on the various probe types and their programming.
The 1521 cannot be used with thermocouples.
The 1521 can be used with probes having two, three, or four wires. Three wires
allow partial compensation for wire resistance. The 1521 measures three-wire
probes using the following algorithm. It measures the sensor resistance, includ-
ing the lead resistance in C2. The 1521 then directly measures the resistance in
the C1 lead of the sensor. The 1521 then subtracts the measured C1 lead resis-
tance from the measured sensor resistance (including the C2 lead resistance) to
obtain the sensor resistance used to calculate temperature. Note: This algorithm
assumes that the lead resistance of C1 and C2 are equal. If C1 and C2 have dif-
ferent resistance there will be an error in the calculated sensor resistance equal
to this difference. Four wires allow complete rejection of wire resistance and
should be used when the best accuracy is desired. The 1521 wire setting must
be set to match the actual number of wires of the probe (see Section 7.8.14).
22
6 General Operation
INFO-CON Connector
Temperature is generally sensed at the tip of the probe. To get an accurate tem-
perature measurement the probe sheath should be well immersed, with ade-
quate depth and fit, into the medium to be measured.
CAUTION: Probes are fragile devices that can be easily damaged by me-
chanical shock, overheating, and absorption of moisture or fluids in the
wires or hub. Damage may not be visibly apparent but nevertheless can
cause drift, instability, and loss of accuracy. Observe the following pre-
cautions:
23
1521 Handheld Thermometer Readout
User’s Guide
Short
Short
J1 J2 J1 J2 J1 J2
24
7 Display Functions
7 Display Functions
The 1521 operates in any of several modes which determine what information
is visible in the lower alphanumeric portion of the display. The MODE button
changes the mode. This button can be pressed repeatedly until a desired mode
is set. The various modes are listed below in the order they appear. (See Figure
7).
Blank - No data shown
Min/Max - Minimum and maximum measurements (CLR to reset)
Hold - Data stored in the hold registers
Delta(x) - Delta from reference measurement (ENT to set reference)
Units: C/Ω/F/K/R - Set the units of measurement
Rate: 00000 - Set the period for serial transmission
Resol: XXX.XXX - Set the display resolution
CAL MODE - Set probe and instrument calibration parameters (MODE
button must be held for three seconds) See Section 7.8.
Note: Units, Rate, and Resol are temporary modes. The instrument will return
to Delta(x) mode if left in one of these modes when the power is cycled.
25
1521 Handheld Thermometer Readout
User’s Guide
Blank Mode
MODE button
Delta(x) Mode
Min/Max Mode
Set Rate
MODE button
Set Resolution
MODE button
any mode
Cal Mode
26
7 Display Functions
Min/Max Mode
In this mode nothing appears on the alphanumeric portion of the display. Use
this mode if you are only interested in the latest measurements and want to sim-
plify the appearance of the display.
The message “Prb cal expired”, “Mtr cal expired”, “Recharge Needed”, or
“Probe is locked” may be displayed. See Section 11.1.
27
1521 Handheld Thermometer Readout
User’s Guide
7.5 Units
This mode is identified by “Units: C/Ω/F/K/R” that appears on the display.
This mode allows you to select the unit of measurement: C for degrees Celsius,
Ω for resistance in ohms, F for degrees Fahrenheit, K for Kelvin, and R for
Rankine. The Up and Down buttons are used to select one of the displayed
units. The underline cursor indicates which of the units is selected. The ENT
button stores the selection. If the MODE button or CLR button is pressed with-
out pressing ENT the unit remains unchanged. Once the unit is selected, subse-
quent measurements appear in that unit. The Min/Max and Delta(x) values
are automatically converted when changing to C, K, F, or R units.
When measuring in units of ohms, make sure that the appropriate probe type
has been selected in the Cal Mode menu. For example, to measure a sensor be-
tween 0 and 400 ohms, set the units to ohms and select the probe type "CVD".
To measure a sensor between 400 and 500K ohms, set the units to ohms and se-
lect the probe type "THERM".
7.6 Rate
This mode is identified by “Rate: 00000" that appears on the display.
This mode allows you to set the samples that are transmitted through the serial
port. The sample for a 2-wire and a 4-wire probe is approximately one second
and for a 3-wire probe is approximately two seconds. For example, if the rate is
set to 2, every other sample is sent through the serial interface. Note: If the
probe wire is set to 3, the time between samples is twice as long. A value of 0
disables transmission. The Up and Down buttons are used to increase or de-
crease the value for the underlined digit. The ENT button moves to the next
digit. If the ENT button is pushed when the last digit is underlined, the setting
is stored. If the MODE button or CLR button is pressed, the value remains un-
changed. Once the value is set, measurements are transmitted through the serial
port at the given interval.
7.7 Resolution
This mode is identified by “Resol: XXX.XXX” that appears on the display.
This mode allows you to set the resolution of measurements to one, two, or
three digits after the decimal point. The Up and Down buttons are used to in-
crease or decrease the resolution setting. The ENT button stores the setting. If
the MODE button or CLR button is pressed, the setting remains unchanged.
Once the resolution is set, measurements are displayed and transmitted from
the serial port at the given resolution if possible. The resolution may be reduced
if necessary for large values of temperature or resistance.
28
7 Display Functions
CAL MODE
29
1521 Handheld Thermometer Readout
User’s Guide
7.8.3 Time
This function displays and sets the time of day. The time is always set, dis-
played, or printed in 24-hour format. To set each time segment use the Up and
Down buttons. Use the ENT button to move between time segments and to
save the changed values. The CLR button moves to the next function.
7.8.4 Passcode
All functions below this function are protected by a pass-code so that only au-
thorized users can access them. The passcode must be entered correctly to con-
tinue. If the passcode is entered incorrectly or the MODE or CLR buttons are
pressed, the message “Invalid Passcode” is displayed and the instrument is re-
turned to the next display mode after the display mode that was active before
the CAL Mode was entered.
Each individual instrument has its own unique passcode. The passcode number
is given on a notice provided with this instrument. The equipment manager is
responsible for keeping the passcode number stored in a proper location and
30
7 Display Functions
CAL MODE
ensuring that it is not lost or forgotten. If necessary the passcode can be re-
stored by contacting the manufacturer. To enter the passcode use the Up and
Down buttons to set each digit of the passcode and the ENT button to move to
the next digit on the right. Press ENT twice when all digits are entered to move
to the next function. If the passcode is entered correctly the next function in the
CAL Mode will appear.
7.8.5 Date
This function displays the present date and allows it to be changed. This func-
tion is placed in the passcode protected part of the CAL Mode since the calibra-
tion due warning messages depend on the present date as well as the stored due
dates. Use the Up and Down buttons to set each date segment and the ENT
button to move between the date segments. Press ENT when all digits are en-
tered to save the changed values. The CLR button moves to the next function.
7.8.7 Prb #
This function displays the serial number of the probe and allows it to be set.
The probe serial number is stored in the INFO-CON probe connector. Use
the Up and Down buttons to set each digit and the ENT button to move to the
next digit on the right. Press ENT when all digits are entered.
31
1521 Handheld Thermometer Readout
User’s Guide
7.8.9 ITS-90
The ITS-90 type is for PRTs calibrated and characterized using the Interna-
tional Temperature Scale of 1990 equations. This type of probe generally has
excellent accuracy (0.005 to 0.025°C), a wide operating range (–180°C to
420°C or higher), but may be rather expensive. The coefficients that appear
when ITS-90 is selected are “R(.01)”, “a”, “b”, “c”, “d”, “a4", and ”b4".These
should be set to the coefficient values that appear on the PRT calibration certifi-
cate according to the following table.
Unused coefficients must be set to 0.
R(.01) Rtpw
a a5, a6, a7, a8, a9, a10, or a11
b b5, b6, b7, b8 or b9
c c6 or c7
d d
a4 a4 or a5
b4 b4 or b5
32
A PRT was calibrated to ITS-90 and its calibration certificate states values for
coefficients R(273.16K), a4, b4, a8, and b8. Set the 1521 coefficients to the
certificate values as follows.
b b8
c 0.0
d 0.0
a4 a4
b4 b4
Example 2:
A PRT was calibrated to ITS-90 and its calibration certificate states values for
coefficients R(273.16K), a5, and b5. Set the 1521 coefficients to the certificate
values as follows.
a a5
b b5
c 0.0
d 0.0
a4 a5
b4 b5
7.8.10 IEC751
The IEC-751 probe type is for standard PRTs (RTDs) conforming to the
IEC-751 or DIN-43760 standards. This type of probe is generally low cost and
has a wide operating range (–180°C to 420°C or higher), but has limited accu-
racy (0.15°C to 4°C). There are no user programmable coefficients associated
with this probe type.
7.8.11 CVD
The CVD (Callendar-Van-Dusen) type is used with PRTs that have unique cal-
ibration coefficients or do not match the IEC-751 specification. This type of
1521 Handheld Thermometer Readout
User’s Guide
higher), and has moderate accuracy (0.05°C to 1°C). This type uses the
Callendar-Van Dusen equation to calculate temperature from resistance:
⎧ ⎧ ⎡ 1 ⎛ t ⎞ ⎤⎫
⎪R0 ⎨1 + α ⎢t − δ ⎜ − 1⎟ ⎥⎬ t≥0
⎪ ⎩ ⎣ 100 100 ⎠ ⎦⎭
⎝
r( t[ °C ]) = ⎨
⎧ ⎡ 3
⎤⎫
⎪R ⎨1 + α ⎢t − δ t ⎛⎜ t − 1⎞⎟ − β ⎛⎜ t − 1⎞⎟ ⎛⎜ t ⎞⎟ ⎥⎬ t < 0
⎪ ⎪
0
⎢⎣ ⎝
100 100 ⎠ ⎝ 100 ⎠ ⎝ ⎠
100 ⎥⎦⎭
⎩ ⎩
The coefficients R0, α, δ, and β can be set by the user using the functions
“R0", ”Alpha", “Delta”, and “Beta” respectively. When editing “Alpha”, there
is an assumed exponent of 10–3 (i.e. 3.8500 is actually 0.0038500).
7.8.12 YSI-400
The YSI-400 type is for 2252Ω thermistors matching the characteristics of the
YSI-400 series of thermistors. This type of probe generally is low cost, has a
limited operating range (–40°C to 150°C), good accuracy (0.1°C to 0.4°C), and
excellent resolution (0.002°C). There are no user programmable coefficients as-
sociated with this probe type. Temperature is converted from resistance based
upon a standard YSI-400 table resident in the instrument.
7.8.13 THERM
The THERM thermistor type is for thermistors that have unique calibration co-
efficients or do not match the YSI-400 specification. This type of probe gener-
ally has moderate cost, a limited operating range (–40°C to 100°C), excellent
accuracy (0.005°C to 0.02°C) and excellent resolution (0.002°C). This type
uses the Steinhart-Hart equation in the following form:
R( T[ K ])[ Ω ] = exp[ b0 + b1 T −1 + b2 T −2 + b3 T −3 ]
The coefficients b0 through b3 are usually found on the calibration certificate
of the probe. These coefficients are set with the functions “b0", ”b1", “b2", and
”b3". In some cases the equation is used without the b2 term. For this case, the
“b2" coefficient should be set to 0. Consider the following examples.
Example 1:
A thermistor has coefficients for the equation ln(R) as a function of T given as
a, b, c, and d. Set the 1521 coefficients to the certificate values as follows:
34
7 Display Functions
CAL MODE
Example 2:
A thermistor has coefficients for the equation ln(R) as a function of T given as
a, b, and c. Set the 1521 coefficients to the certificate values as follows:
b2 0.0
b3 c
7.8.15 PCal
This function shows the date the probe was calibrated and allows it to be set.
The date is stored in the INFO-CON connector and is used simply for refer-
ence. Use the Up and Down buttons to set each date segment and the ENT but-
ton to move to the next date segment to the right. Press ENT when the date is
correctly entered. Press CLR to move to the next function without changing the
date.
7.8.16 PDue
This function shows the date the probe is due for calibration and allows it to be
set. The date is stored in the INFO-CON connector. The probe due date is regu-
larly checked against the present date and if the calibration has expired the user
is alerted with the message “Prb cal expired”. This message only disappears
when the probe is recalibrated and the probe calibration due date is set to a fu-
ture date. To set the date, use the Up and Down buttons to set each date seg-
ment and the ENT button to move to the next date segment to the right. Press
ENT when the date is correctly entered. Press CLR to move to the next func-
tion without changing the date.
35
1521 Handheld Thermometer Readout
User’s Guide
Note: Probe parameters are uploaded to the INFO-CON connector at this point.
7.8.17 Filter
The digital filter helps to smooth variations in the measurements and improve
resolution. A negative effect of filtering is that it tends to slow the response to
changes in temperature. You can increase the filter time constant to further im-
prove accuracy and resolution or decrease the time constant to reduce the re-
sponse time. Valid values are any number between 0.0 and 60.0 seconds. A
value of 0.0 disables the filter. The default is 0.0. Use the Up and Down but-
tons to set the value and press the ENT button to store and continue. Press
CLR to move to the next function without changing the filter.
7.8.18 MCal
This function shows the date the 1521 was calibrated and allows it to be set.
The date is used simply for reference. Use the Up and Down buttons to set
each date segment and the ENT button to move to the next date segment on the
right. Press ENT when the date is correctly entered. Press CLR to move to the
next function without changing the date.
7.8.19 MDue
This function shows the date the 1521 is due for calibration and allows it to be
set. The meter calibration due date is regularly checked against the present date
and if the calibration has expired the user is alerted with the message “Mtr cal
expired”. The message will only disappear when the 1521 is recalibrated and
the meter calibration due date is set to a future date. To set the meter calibration
due date use the Up and Down buttons to set each date segment and the ENT
button to move to the next date segment to the right. Press ENT when the date
is correctly entered. Press CLR to move to the next function without changing
the date.
36
8 Communications Interface
RS-232 Connection
8 Communications Interface
Remote communications allows an external device, such as a computer, to com-
municate with the 1521 to obtain measurement data and control its operation.
Communication is accomplished with various commands issued to the 1521
through the RS-232 serial port.
GND 5
9
GND 4
8
RxD TxD 3
7
RxD 2
TxD 6
1
37
ON (all carriage returns are followed by a linefeed (ASCII decimal 10)), and
the duplex to HALF disabling echo.
The serial port can be used to transmit measurements to a computer or printer.
It can also be used to change settings of the 1521 from a computer or interface
with software. A full list of commands follows in Section 8.2.
Commands sent to the 1521 must end with an EOS character which is a car-
riage return (ASCII decimal 13) or linefeed character (ASCII decimal 10).
Commands can be sent with upper or lower case letters. Data returned from the
1521 ends with a carriage return and a linefeed.
Response Response
Description Command Example Format Example Range
Read measurement, extended t[emperature] t t: 999.999 t: 125.745 C
format {C, O, F, R, or K} 03-01-1999
mm-dd-yyyy 15:49:15
hh:mm:ss
Read measurement, simple format f[etch?] fetch? t: 999.999 t: 25.587 C
rea[d?] read? {C, O, F, R, or K}
me[asure?] meas?
Read minimum m[inimum] min min: 999.999 min: -14.653 C
{C, O, F, R, or K}
Read maximum ma[ximum] max max: 999.999 max: 416.781 C
{C, O, F, R, or K}
Clear minimum and maximum cl[ear] clear
Read a hold register h[ld]<n> hld6 hld<n>: 999.999 {C, O, F, hld6: 26.553 C <n> can be 1 to
R, or K} 6
8 Communications Interface
Communication Command List
Response Response
Description Command Example Format Example Range
Set resolution digits re[solution]=n res=1 1 to 3
Read time ti[me] time ti: hh:mm:ss ti: 16:23:45
Set time ti[me]=hh:mm:s ti=08:15:0 00:00:00 to
Note: You must set hours, min- s 0 23:59:59
utes, and seconds for the time to
be set correctly.
Read identification information *i[dn?] *idn? <manufacturer>, 2.11
Hart,
<model>, 1521,
<serial number>, 95001,
<firmware version>
Read model and firmware version *v[ersion] *ver ver.<model>, ver.1521,
<firmware version> 2.11
Legend: [] Optional Command data
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
Response Response
Description Command Example Format Example Range
Unprotect calibration parame- *pas[scode]=nnnnn *pas=12345
ters with the pass-code
Read date da[te] date da: yyyy-mm-dd da: 1999-03-28
Read passcode protection *pas *pas ps:9 ps:0 0 or 1
state
Set date da[te]=yyyy-mm-dd da=1999-03-28 1999-01-01 to
2100-12-31
Read probe serial number se[rial] ser ser: xxxxxxxxxx ser: 100231A
Set probe serial number se[rial]=xxxxxxxxxx ser=001 0 to 9, A to Z,
., -, /
Read probe type p[robe] pr pr: {ITS-90, IEC751, pr: ITS-90
CVD, YSI400, THERM}
39
1521 Handheld Thermometer Readout
User’s Guide
Response Response
Description Command Example Format Example Range
Set probe type p[robe]=IT[S-90]/ pr ITS-90,
IE[C751]/ C[VD]/ IEC751,
Y[SI400]/ T[HR] CVD, YSI400,
THERM
Read ITS-90 R(.01) or CVD R0 r[0] r0 r0: 9999.999 r0: 100.035
Set ITS-90 R(.01) or CVD R0 r[0]=n r0=100.106 1 to 900
Read ITS-90 coefficient a/b/c/d/a4/b4 a {a/b/c/d/a4/b4}: a: -2.8644101E-5
-9.999999E-99
Set ITS-90 coefficient a/b/c/d/a4/b4=n a=-2.8644E-5 1.0 to 9.9e-18
Read CVD coefficient al[pha]/de[lta]/be[ta] al {al/de/be}: 9.99999999 al: 0.0038123
Set CVD coefficient al[pha]/de[lta]/be[ta]=n al=0.0038512 E+/-18
Read THERM thermistor b0/b1/b2/b3 b3 {b0/b1/b2/b3}: b3: -9.944070e-12
coefficient 9.999999e-99
Set THERM thermistor b0/b1/b2/b3=n b3=-9.944E-12 E+/-18
coefficient
Read probe wires wi[res] wi wi: 9 wi: 4
Set probe wires wi[res]=2/3/4 wi=4 2,3,4
Read probe calibration date pc[al] pcal pc: yyyy-mm-dd pc: 1999-03-28
Set probe calibration date pc[al]=yyyy-mm-dd pc=1999-03-28 1999-01-01 to
2100-12-31
Read probe due date pd[ue] pdue pd: yyyy-mm-dd pd: 1999-03-28
Set probe due date pd[ue]=yyyy-mm-dd pd=1999-03-28 1999-01-01 to
2100-12-31
Read filter fi[lter] fi fi: 99.99999 fi: 0.5
Set filter fi[lter]=n fi=0 0.0 to 60.0
Read low range calibration *c1 *c1 c1: 99.999 c1: -0.0121
Set low range calibration *c1=n *c1=-0.012 -1.999 to 1.999
Read high range calibration *c2 *c2 c2: 9999.9 c2: -0.9
Set high range calibration *c2=n *c2=-0.9 -1980.0 to
1980.0
Read meter calibration date mc[al] mcal mc: yyyy-mm-dd mc: 1999-03-28
Set meter calibration date mc[al]=yyyy-mm-dd mc=1999-03-28 1999-01-01 to
2100-12-31
Read meter due date md[ue] mdue md: yyyy-mm-dd : 1999-03-28
Set meter due date md[ue]=yyyy-mm-dd md=1999-03-28 1999-01-01 to
2100-12-31
Legend: [] Optional Command data
{} Returns either information
40
8 Communications Interface
Communication Command List
Response Response
Description Command Example Format Example Range
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
41
9 Calibration
Required Equipment
9 Calibration
The 1521 should be calibrated at regular intervals to ensure that it continues to
measure with proper accuracy. Calibration should only be done by qualified,
authorized personnel.
43
1521 Handheld Thermometer Readout
User’s Guide
4. Verify the accuracy at 0Ω, 4kΩ, 10 kΩ, 40 kΩ, 100 kΩ, and 500 kΩ.
5. Set the meter calibration date to the present date (see Section 7.8.18).
6. Set the meter due date to one year from the present date (see Section 7.8.19).
44
10 Maintenance
10 Maintenance
• This instrument has been designed with the utmost care. Ease of operation
and simplicity of maintenance have been a central theme in the product
development. Therefore, with proper care the instrument should require
very little maintenance. Avoid operating the instrument in oily, wet, dirty,
or dusty environments.
• If the outside of the instrument becomes soiled, it may be wiped clean
with a damp cloth and mild detergent. Do not use harsh chemicals on the
surface which may damage the LCD or the plastic outside shell.
• If a hazardous material is spilt on or inside the equipment, the user is re-
sponsible for taking the appropriate decontamination steps as outlined by
the national safety council with respect to the material.
• Before using any cleaning or decontamination method except those rec-
ommended by Hart, users should check with an Authorized Service
Center to be sure that the proposed method will not damage the equip-
ment.
• If the AC adapter becomes damaged, have it replaced immediately. Never
disassemble the AC adapter or attempt to repair it.
• If the instrument is used in a manner not in accordance with the equip-
ment design, the operation of the thermometer may be impaired or safety
hazards may arise.
45
11 Troubleshooting
An Error Message Is Displayed
11 Troubleshooting
In case you run into difficulty while operating the 1521, this section provides
some suggestions that may help you solve the problem. Below are several situa-
tions that may arise followed by possible causes of the problem and suggested
actions you might take.
Problem Solution
“Recharge Needed” The battery is low and needs to be recharged. When this message first
appears the user has approximately 30 minutes of charge left. If the
battery is not recharged, the 1521 continues to operate until the bat-
tery drops below 0% and then automatically shuts itself off after dis-
playing the error message.
“No Probe” The display shows ‘ - - - - - - - ‘. The instrument is unable to detect
that a probe is connected. Check that a probe with a properly pro-
grammed INFO-CON connector is attached.
“Probe is locked” The incorrect probe is attached. The meter is set to accept only the
probe with the given serial number. Attach the correct probe or have
the instrument properly programmed for the new probe.
“Prb cal expired” The probe calibration has expired. Have the probe calibrated and
the calibration due date reset.
“Mtr cal expired” The meter calibration has expired. Have the instrument calibrated
and the calibration due date reset.
“Invalid Passcode” The passcode entered was not correct. Check the passcode on the
Pass-code Notice sent with the instrument.
The instrument display is blank Make sure the instrument has power, either from the batteries or
when normally it should show the AC adapter. Verify that the initialization information appears on
measurements. the display when the power is switched on. Check to make sure that a
probe is properly connected. Check to make sure that the probe is
programmed properly.
47
1521 Handheld Thermometer Readout
User’s Guide
Problem Solution
While attempting to measure re- Poor or incorrect connection of the probe. A common mistake is to
sistance the display shows an in- connect the wires of the probe to the wrong terminals. Check the wir-
correct value. ing carefully (see Figure 6).
Open, shorted, or damaged sensor or lead wires. Check the resis-
tance across the sensor using a handheld DMM. Also check the resis-
tance between common pairs of leads. Check to make sure there is
no conductivity between any of the leads and the probe sheath. Use a
good-quality sensor to avoid errors caused by drift, hysteresis, or in-
sulation leakage.
Stem conduction error. Make sure the stem of the probe is im-
mersed in the medium with adequate depth and a tight fit.
Electrical interference. Intense radio-frequency radiation near the
1521 or the probe can induce noise into the measurement circuits re-
sulting in erratic readings. Try eliminating the source of interference or
moving the 1521 to a different location. A well-grounded, shielded ca-
ble should be used for the probe.
Memory Check Error Memory Error.
A non-critical parameter is corrupt in memory. The non-critical
praameters are Units, Rate, Resolution, Filter, and serial port settings.
Press any key and the display reads “Check Settings”. Press any key
again to resume normal operation. Check your current settings for the
specific parameters that were set.
Memory Recover X.
A critical parameter is corrupt and the 1521 can recover the correct
value for that parameter. The critical parameters are CAL1, CAL2,
Passcode, Meter Due Date, Meter Cal Date, and Serial Number.
Press any key and the display reads “Check Settings”. Press any key
again to resume normal operation. Also, check non-critical settings as
in the Memory Error above. The number following the message is
used by authorized technicians.
Memory Invalid!
A critical parameter is corrupt and the 1521 cannot recover the cor-
rect value for the parameter. The critical parameters are CAL1, CAL2,
Passcode, Meter Due Date, Meter Cal Date, and Serial Number.
Press any key and the display reads “Contact Factory”. Contact an
Authorized Service Center (see Section 1.3) for assistance in guiding
you through the process of reentering critical parameters.
Note: The parameters stored in the INFO-CON, Probe Wires, Probe
Type, Probe Parameters, are not checked by the meter during mem-
ory testing.
11.2 CE Comments
11.2.1 EMC Directive
Hart Scientific’s equipment has been tested to meet the European Electromag-
netic Compatibility Directive (EMC Directive, 89/336/EEC). The Declaration
of Conformity for your instrument lists the specific standards to which the unit
was tested.
48
11 Troubleshooting
CE Comments
The instrument was designed specifically as a test and measuring device. Com-
pliance to the EMC directive is through IEC 61326-1 Electrical equipment for
measurement, control and laboratory use – EMC requirements (1998).
As noted in the IEC 61326-1, the instrument can have varying configurations.
The instrument was tested in a typical configuration with shielded, grounded
probe and RS-232 cables. Emissions may, in non-typical applications, exceed
the levels required by the standard. It is not practical to test all configurations,
as the manufacturer has no control over the probes the user may connect to the
instrument.
49