Perillaje de Ecocardiografía
Perillaje de Ecocardiografía
Perillaje de Ecocardiografía
197]
Review
Article Transesophageal echocardiography:
Instrumentation and system controls
Mahesh Prabhu, Dinesh Raju, Henning Pauli
Department of Cardiothoracic Anesthesia and Intensive Care, Freeman Hospital, Newcastle Upon Tyne, NE7 7DN, UK
ABSTRACT Transesophageal echocardiography (TEE) is a semi-invasive, monitoring and diagnostic tool, which is used
in the perioperative management of cardiac surgical and hemodynamically unstable patients. The low degree
of invasiveness and the capacity to visualize and assimilate dynamic information that can change the course
of the patient management is an important advantage of TEE. Although TEE is reliable, comprehensive,
credible, and cost-effective, it must be performed by a trained echocardiographer who understands the
indications and the potential complications of the procedure, and has the ability to achieve proper acquisition
and interpretation of the echocardiographic data. Adequate knowledge of the physics of ultrasound and the
TEE machine controls is imperative to optimize image quality, reduce artifacts, and prevent misinterpretation of
diagnosis. Two-dimensional (2D) and Motion (M) mode imaging are used for obtaining anatomical information,
while Doppler and Color Flow imaging are used for information on blood flow. 3D technology enables us to
view the cardiac structures from different perspectives. Despite the recent advances of 3D TEE, a sharp,
optimized 2D image is pivotal for the reconstruction. This article describes the relevant underlying physical
principles of ultrasound and focuses on a systematic approach to instrumentation and use of controls in the
practical use of transesophageal echocardiography.
Received: 08-01-12
Accepted: 03-03-12 Key words: Cardiac ultrasound, instrumentation, system controls, transesophageal echocardiography
Address for correspondence: Dr. Mahesh Prabhu, Department of Cardiothoracic Anesthesia and Intensive Care, Freeman Hospital, Newcastle Upon Tyne,
NE7 7DN, UK. E-mail: [email protected]
CARDIAC ULTRASOUND
Field of view
The field of view (FOV) is the plane or area scanned by
the ultrasound transducer.
Sector
The sector is a fan-shaped image with the narrow end
displaying objects closer to the transducer and the wider
Figure 5: Diagram of a sector showing progression of scan lines section showing the more distant structures. The region
of interest (ROI) should be kept as close as possible to
The small wheel allows for lateral flexion or side-to- the TEE probe. 2-D imaging is displayed on a gray scale,
side movement of the tip. In addition, the probe can with the high amplitude signals being ascribed a white
be advanced or withdrawn in a vertical fashion into color and low amplitude a black color. The shades of
the esophagus and stomach or manually turned to the gray form a picture of the heart in a sector.
left or right with respect to the esophagus[11] [Figure 3].
The array rotation buttons turn the mutiplane angle Scan lines
from 0 to 180°. The piezoelectric elements can be Each piezoelectric element of the transducer transmits
electronically rotated either clockwise or anticlockwise a US pulse along a line. All echoes from the structures
using the multiplane angle control, to enable viewing along this line are received, thus producing a scan line
of the heart in any desired plane. The newer probes whose length is the image depth.[12] The ultrasound
are enabled to suppress the electrocautery artifacts. beam then moves on to the next scan line position where
Paediatric probes are of smaller diameter (5-7mm) and the process is repeated [Figure 5]. Increased scan lines
higher frequency (5-10MHz) with greater flexibility. improve the resolution of the image.
The echocardiography unit combines two component by twice the original and a decrease of 3 dB will halve
technologies: Imaging (Two-dimensional imaging and the power- the 3 dB rule.[23] Minimum power which
M-mode) and Doppler (continuous wave, pulse wave, is consistent with good image quality should be used.
and color-flow Doppler mapping). The individual
components that execute the different modes of imaging Thermal Index (TI) is an estimate of risk from thermal
are a video display unit and a power source.[19] effects of US and is defined as the ratio of total acoustic
power to that required to raize the temperature by 1°C
System controls under defined assumptions.[9]
Power is the rate of energy delivered in a sound wave.
Intensity is the concentration of power within the Mechanical Index (MI) is an estimate of risk from the
cross-sectional area of the US beam.[20] There are several non-thermal effects and is the amount of negative
system controls that can be used to obtain a good image acoustic pressure within an ultrasonic field. It is defined
by using minimal acoustic intensity. These controls can as the peak negative pressure divided by the square root
be divided into three categories: of the ultrasound frequency. Values of indices less than
1 are generally considered safe and though a high index
Direct controls reading is not proportional to a bioeffect, every effort
The Power control has a direct impact on acoustic must be made to reduce the chances of a high index.[5]
intensity and is used to select the minimal intensity
levels required to produce an optimal image.
Indirect controls
The controls of imaging mode, pulse repetition
frequency (PRF), focus depth and frequency have
an indirect effect on acoustic intensity. 2D imaging
mode disperses energy over the entire scanned area
but Doppler concentrates energy in a particular area.
Increasing the rate and time of the ultrasound signals
increases the time-averaged intensity value.
Receiver controls
Receiver controls such as gain, dynamic range and image
Figure 6: Mid-esophageal AV short axis view. LA - Left atrium; NCC - Non-
processing influence ultrasound without affecting the coronary cusp; LCC - Left coronary cusp; RCC - Right coronary cusp;
intensity output. Hence they should be optimized first TV - Tricuspid valve; PV - Pulmonary valve
before increasing the power.
Table 1: Commonly used echocardiography controls
Another classification of these controls can be 2D Pulse wave Continuous Color Doppler
based on the time of modification of the signal. imaging Doppler wave Doppler
Power Sample volume Beam axis Color box size
Preprocessing controls modify the analog and digital size and placement placement and placement
signal prior to storage in the computer memory whereas Frequency Trackball Trackball Trackball
postprocessing controls manipulate the image after its Depth Scale Scale Color scale
entry into memory.[21] Focus Baseline Baseline Color map
Gain Gain Gain Color gain
Two-dimensional imaging TGC Sweep speed Sweep speed Baseline
2D imaging helps to image the heart motion in real- LGC Depth Depth Filter
time enabling detailed morphological and functional Sector Frequency Frequency Measurement
size and trace
assessment [Figure 6]. It also helps quantitative
Zoom Measurement Measurement
assessment of cardiac dimensions and provides the and trace and trace
framework for M-mode and Doppler imaging.[22] The Compression Reject Reject
various control buttons used to get an optimal image Reject Invert Invert
are described in Table 1. Persistence Filter Filter
Freeze Angle correction Angle correction
Transmit power Measurement
Most machines describe power in terms of decibels and trace
Annotation
(dB). An increase of 3 dB increases power or intensity
Frequency
The frequency of most TEE probes is between 3.5-7.5
MHz. A higher transducer frequency improves
resolution but lacks depth penetration whilst a
lower frequency diminishes resolution and improves
penetration.[17] The highest possible frequency that
allows adequate penetration upto the ROI should be
used in two-dimensional (2-D) imaging.[14]
Focus
Focusing the US beam by mechanical or electrical means
helps to optimize resolution. The area of interest should
be in the “near field” between the transducer and the
focal length as the resolution is highest in this region.
Therefore repositioning of the focal zone is particularly
important when performing 2D measurements and
examination of specific regions of interest.[14]
Gain
Gain is the degree of amplification of the returning
US signal.[20] The gain button alters the “brightness”
of the whole image by adjusting the amplification of
Figure 7: Mid-esophageal AV short axis view. In the image the a arrow highlights
all returning echo signals. Minimal gain should be a low gain setting of 20; and the coaptation line between NCC and LCC is not
used to provide an optimal image with good quality visualized. In image b, the gain is set too high at 100. LCC - Left coronary cusp;
NCC - Non coronary cusp; RCC - Right coronary cusp
without dropout or blooming of signals[14] [Figure 7]. As
a rule of thumb, fluid and blood should appear black,
myocardium a medium grey and the pericardium and
calcification, a bright white color on the grey scale.
Figure 9: Transgastric mid-papillary short axis view. In panel A the septal and Figure 10: The arrows highlight the compression setting. With low compression,
lateral wall areas are poorly defined. In panel B both wall regions are adequately Panel A displays a grainy image with stark contrast. With high compression Panel
imaged by increasing the lateral gain. PPM - Posteromedial papillary muscle and B provides a smoother image with more shades of gray, but increased unwanted
APM - Anterolateral papillary muscle signals. A mid-level compression is usually appropriate
M-mode imaging
Motion mode imaging is one of the basic forms of
imaging. One scan line provides a single dimensional
“ice-pick” view of the heart and the signal is plotted
as brightness mode against time on the X-axis. The
amplitude of movement of an object and its distance from
the transducer is displayed on the Y-axis[26] [Figure 11].
Figure 11: M mode applied to a transgastric long axis view to quantify chamber
size. LVIDd - Left ventricular diameter in diastole; LVIDs - Left ventricular diameter
in systole; PM - Papillary muscles; PERI - Pericardium The short sampling time ensures a higher sampling
frequency and frame rate, which allows for extremely
ultrasound and electronic “noize”. Care must be taken not accurate measurement and timing of events. M-mode is
to eliminate low-intensity echoes from fresh intracardiac typically used for measurements of the aortic and mitral
thrombi by setting the reject threshold too high.[24] valve, ventricle cavity size and thickness during systole
and diastole and timing of flow patterns in combination
Persistence with color flow mapping.[17,27]
Persistence was the phenomenon of superimposition of
new images on the old fading images in the cathode ray DOPPLER IMAGING
tubes of earlier ultrasound machines.[25] Nowadays it is
a postprocessing control that averages and merges the The Doppler shift is defined as the change in frequency
frames thereby producing a smooth moving image of the of sound reflected by a moving object and is determined
heart. Persistence should be set low to retain temporal by the Doppler equation[28] [Figure 12]. Doppler shift
resolution and a real-time appearance.[24] caused by red blood cells reflecting ultrasound can be
Figure 12: The Doppler frequency shift (fd) depends on the transmitted frequency
(ft) and the velocity (V) of the moving blood and the angle (θ) between the
ultrasound beam and the direction of blood flow. c = speed of sound in the tissue
Figure 14: The transgastric long axis view and the CW Doppler applied through
the stenosed calcific aortic valve displays a typical ‘filled-in’ velocity envelope. Figure 15: A mid esophageal, four-chamber view superimposed with color
By tracing the velocity signal the VTI, peak, and mean gradient are calculated flow doppler showing mitral regurgitation. The vena contracta (VC) depicts
the regurgitant orifice. LA - Left atrium; RA - Right atrium; RV - Right ventricle;
LV - Left ventricle
the different blood velocities along the beam axis,
continuously. This makes it difficult to accurately know error. This does not actually change the direction of the
the depth of the reflected signal. The main disadvantage Doppler beam nor does it alter the quality of the spectral
of CWD is its lack of depth discrimination (Range recording. It is preferable to adjust the transducer to
ambiguity). position the beam axis parallel to the blood flow rather
than count on angle correction.
Display
The CWD is displayed as a waveform with a filled- Compression, and the Reject and Invert buttons can be
in spectrum caused by the different velocities being used as for the pulse wave Doppler.
measured along the scan line [Figure 14]. A turbulent
flow, caused by obstruction, results in spectral Color flow Doppler
broadening and increased velocities. The Color Flow Doppler (CFD) provides a striking
display of both blood flow and the cardiac anatomy
Scale factor
by superimposing the PWD flow data on 2D images[29]
The scale control adjusts the range of velocity that can
[Figure 15]. It is useful in diagnosing the abnormal flow
be displayed, and should be set so that the highest
as well as confirming the normal structures.
velocity spectral trace can be displayed without cutting
off the peak of the trace.
Color maps
Sweep speed The PWD used for color mapping, records the mean
This control indicates how fast the spectral waveform velocity data from multiple sample volumes – color
can sweep across the screen. Increasing the sweep packets along each scan line. This velocity data is
speed decreases the number of cycles and increases then color-coded; blue indicates flow away from the
the width of the waveform. A low sweep speed transducer and red is the flow toward the transducer.
shows more waveforms on the screen and is useful to The color bar on the screen provides a reference frame
demonstrate waveform variations caused by respiration for interpreting the colors. The color flow maps are
and arrhythmias. then integrated with 2-D imaging to provide a real-time
display.
Wall filters
All Doppler systems have a variable wall filter control Sampling and depth ― Color sector size and placement
that sets a threshold beyond which frequency signals The size, position, and trackball controls are used to
can be removed from the display. change the width, length, and position of the color
box. The color box, while enclosing the ROI, should
Angle correction be narrow and as close to the transducer as possible.
This control changes the angle calculation in the This prevents aliasing, lower frame rates, and ‘swimmy
Doppler equation by adjusting the beam axis to the images’ caused by the longer time taken to interrogate
direction of assumed flow, thereby minimizing the pulses that are far away from the transducer.[19]
ACKNOWLEDGMENT
REFERENCES
and artifacts. In: Sidebotham D, Merry A, Legget M, editors. Practical Silvestre FE, Wiegers SE, editors. Intracardiac Echocardiography. Oxon:
Perioperative Transoesophageal Echocardiography. 2nd ed. Philadelphia: Taylor and Francis; 2006. p. 7-17.
Elsevier Saunders; 2011. p. 2-21. 27. Chambers J. Echocardiographic Techniques. In: Chambers J, editor.
16. Oh JK, Seward JB, Tajik AJ. How to obtain a good Echocardiographic Echocardiography in clinical practice. London: The Parthenon
Examination. In: Oh JK, Seward JB, Tajik AJ, editors. Ultrasound Publishing Group; 2002. p. 9-20.
physics, techniques and medical knowledge, The Echo Manual. 3rd ed. 28. Thomas JD, Licina MG, Savage RM, Aronson S. Physics of
Philadelphia: Lippincott Williams and Wilkins; 2006. p. 1-6. Echocardiography. In: Savage RM, Aronson S, editors. Comphrensive
17. Walsh CA, Wilde P. Introduction to Echocardigraphic Modalities. Textbook of Intraoperative Transoesophageal Echocardiography. 1st ed.
In: Walsh CA, Wilde P, editors. Practical Echocardiography. London: Philadelphia: Lippincott Williams and Wilkins; 2005. p. 3-23.
Greenwich Medical Media Limited; 1999. p. 21-36. 29. Callaway M, Wilde P. Instrumentation. In: Izzat MB, Sanderson JE,
18. Thomas JD, Adams DB, Devries S, Ehler D, Greenberg N, Garcia M, et al. Sutton, editors. Echocardiography in adult cardiac surgery. Oxford:
Guidelines and recommendations for digital echocardiography. J Am MG, ISIS Medical Media; 1999. p. 3-13.
Soc Echocardiogr 2005;18:287-97. 30. Sutherland GR, Stewart MJ, Groundstroem KW, Moran CM, Fleming A,
19. Lawrence JP. Physics and Instrumentation of Ultrasound. Crit Care Med Guell-Peris FJ, et al. Color Doppler myocardial imaging: a new technique
2007;35 (8 Suppl): S314-22. for the assessment of myocardial function. J Am Soc Echocardiogr
1994;7:441-58.
20. Feigenbaums H, Armstrong WF, Ryan T. Physics and Instrumentation.
In: Feigenbaums H, Armstrong WF,Ryan T, ed. Text book of 31. Sugeng L, Shernan SK, Salgo IS, Weinert L, Shook D, Raman J, et al. Live
Echocardiography. 6th ed. Philadelphia: Lippincot Williams and Wilkins; 3-dimensional transesophageal echocardiography initial experience
2005. p. 11-45. using the fully-sampled matrix array probe. J Am Coll Cardiol
2008;52:446-9.
21. Kahn RA, Salgo IS. Principles of Ultrasound. In: Konstadt SN, Shernan
32. Hung J, Lang R, Flachskampf F, Shernan SK, McCulloch ML, Adams DB,
SK, Oka, editors. Clinical Transoesophageal Echocardiography. 2nd ed.
et al. 3D echocardiography: A review of the current status and future
Philadelphia: Lippincott Williams and Wilkins; 2003. p. 3-9.
directions. J Am Soc Echocardiogr 2007;20:213-33.
22. Lee D, Solomon SD. Essential Echocardiography. In: Solomon SD, editor.
33. Perk G, Lang RM, Garcia-Fernandez MA, Lodato J, Sugeng L, Lopez J, et al.
A practical Handbook. New Jersey: Humana Press; 2007. p. 9-33.
Use of real time three-dimensional transesophageal echocardiography
23. Forsberg E, Adams D. Understanding of Ultrasound System Controls. in intracardiac catheter based interventions. J Am Soc Echocardiogr
In: Mathew JP, Ayoub C, editors. Clinical manual and review of 2009;22:865-82.
transesophageal echocardiography. USA: Mcgraw-Hill Companies Inc; 34. Vegas A, Meineri M. Core review: three-dimensional transesophageal
2005. p. 16-30. echocardiography is a major advance for intraoperative clinical
24. Dyal HW, Frith MD, Reeves ST. Techniques and Tricks for Optimizing management of patients undergoing cardiac surgery: a core review.
Transesophageal Images. In: Perrino AC, Reeves ST, editors. A Practical Anesth Analg 2010;110:1548-73.
Approach to Transesophageal Echocardiography. 2nd ed. Philadelphia:
Lippincott Williams and Wilkins; 2003. p. 435-45.
Cite this article as: Prabhu M, Raju D, Pauli H. Transesophageal
25. Weyman AE. Physical Principles of Ultrasound. In: Weyman AE, editor.
echocardiography: Instrumentation and system controls. Ann Card Anaesth
Cross-sectional echocardiography. Philadelphia: Lea and Febiger; 1982.
2012;15:144-55.
p. 3-61.
Source of Support: Nil, Conflict of Interest: None declared.
26. Wiegers SE, Scott CH. Physics and Instrumentation of Ultrasound. In:
Announcement
iPhone App
A free application to browse and search the journal’s content is now available for iPhone/iPad. The application
provides “Table of Contents” of the latest issues, which are stored on the device for future offline browsing.
Internet connection is required to access the back issues and search facility. The application is Compatible
with iPhone, iPod touch, and iPad and Requires iOS 3.1 or later. The application can be downloaded from http://
itunes.apple.com/us/app/medknow-journals/id458064375?ls=1&mt=8. For suggestions and comments do
write back to us.