User Manual
v.030315 1
User Manual Table of Contents
The Zootec Double II vaccinator was developed 1. Safety 3
and manufactured by Zoo-Techniques located in
Bazainville, France (40 minutes southwest of Paris). 2. General Characteristics 3
15. N
ext Generation Twin Touch Plate
Troubleshooting 14
A SANOFI COMPANY
v.030315 2
1. Safety 6)
2. General Characteristics
The Zootec Double II is a pneumatic vaccinator that
ensures an automatic vaccination of day-of-age
poultry birds (chickens, ducks, turkeys, guinea fowls,
etc.) via subcutaneous injection in the neck. It is
4) Be sure the end user has been properly trained equipped with the Merial Next Generation Twin Touch
on all aspects of the vaccinator and they have Plate that improves vaccination accuracy and speed
received a copy of the user guide. while reducing the risks of injury to the operator and
chicks. Injection volumes range from 0.1 ml to 0.5 ml
5) Replace the defective parts or consumable spare depending on the syringe. The vaccinator is designed
parts with new parts that have been supplied by for one operator and can vaccinate up to 3000 chicks
Zoo-Techniques. per hour. The correct cadence for a good vaccina-
tion is 2,500 chicks per hour. The machine requires
external supply of compressed air with a pressure
comprised between 6 to 8 bars (85-120 psi). Pressure
will then be regulated to 5-6 bars (73-85 psi) inside
the machine.
A SANOFI COMPANY
v.030315 3
3. Presentation of the Machine
A SANOFI COMPANY
v.030315 4
Syringe 6. Injection Methods
The syringe is a precision device that is manufactured The single/double switch (ZT368) allows you to
to the most stringent of tolerances to deliver an choose from two different injection methods. The two
accurate and consistent dose. Unless otherwise injection methods are as follows:
specified, the Zootec is supplied with syringes that
deliver a fixed volume of 0.1 ml to 0.5 ml depending
on the piston size using water or oil-based vaccines.
Method I: Single injection
Half-dosage fixed volume syringes such as a 0.15 ml
dose are also available upon request. Any syringe
change does not require new machine settings.
4. General Recommendations
• I n order to avoid being injured by the needle,
all adjustments or repairs to the machine
must take place while the air pressure is
turned off (on/off test switch (ZT367), the
air inlet is disconnected, and the needle is
ZT626
removed from the syringe.
Never operate the machine without the syringe.
A SANOFI COMPANY
v.030315 5
7. Machine Startup
• ompressed air should be supplied by a
C • ush the on/off test switch (ZT367) located
P
standard air compressor. Although the on the front of the machine.
machine has its own filter, it is recommended
• et the working pressure of the machine to
S
to use an air compressor equipped with its
5-6 bars (73-85 psi)) on the pressure gauge
own filters. Remember to drain the residual
(ZT01NS) by adjusting the working
water in the compressed air tank of the air
pressure using the filter regulator (ZT17NS)
compressor once a week. The Zootec Double
if necessary.
II filter should be checked every month and
drained, if needed. • hoose the injection method that you want
C
with the single/double switch (ZT368).
• I nstalling an air dryer with the compressor
will help achieve the best air quality. • Mount the bottle holder on the machine.
• onnect the female air connector (ZT24NS)
C • djust the batch counter and reset the
A
to the male air connector on the left side of totalizer counter to position “0”.
the machine (ZT23NS).
The syringe connector (ZT21B) is preset on the primary air cylinder at 4 cm. This is
ZT50D
4.0 cm
ZT21B the distance from the middle dividing plate (ZT50D) and the front of the black piece
that holds the syringe piston end (ZT83M).
3.6 cm The secondary syringe spacer pusher (ZT212) is pre-set on the secondary air cylinder
at 3.6 cm. This is the distance from the middle dividing plate (ZT50D) and the front
of the black pusher button.
ZT212
These settings are important to ensure a good injection quality from the syringes
and should also be taken into account anytime an air cylinder or associated part is
replaced.
A SANOFI COMPANY
v.030315 6
The batch counter (ZT02) permits you to count the quantity of chicks per box. Press
the upper button (1) and simultaneously preselect the number on the counter using
1
the lower buttons (2). When the number is reached, the machine will emit an air
signal inside the machine and will stop functioning. Press the reset counter push
button (ZT14) on the top cover to restart the machine. Pressing the batch counter
(1) can also restart the machine.
2
The totalizer counter (ZT03) is used for totalization of daily vaccinations. The totalizer
counter can be reset to zero at the beginning of each vaccination period by pushing
the reset button (1).
A) Loosen the syringe holder nut then gently lift the lid
clamp of the connector syringe (ZT21B). Hold the
A B primary syringe by the syringe end piece (ZT83M)
and create a gap between the syringe piston end
and the connector syringe (ZT21B).
To reposition the syringes, the steps mentioned above will be repeated in the reverse order, using both hands
to reinsert the syringes.
A SANOFI COMPANY
v.030315 7
Priming the syringe: Option 1
Caution: Be sure the O-rings are clean and lubricate them with food-grade silicone lubricant spray or
oil spray vaseline (ZT101) before starting the priming process, if necessary.
2) R
emove the syringe piston from the syringe 7) O
nce the piston is completely inserted,
body and lay it beside the machine. Do not loosely hand-tighten the piston hex connector
touch the piston with your fingers. Un-clamp to the body.
the vaccine line and let the vaccine flow
8) M
anually pump the piston several times until
through the line via gravity. Pinch the end of
the system is completely primed. Fully tighten
the vaccine line to stop the flow of vaccine
the hex connector using the key provided.
once the line is completely filled. Hold the sy-
Avoid over tightening the hex connector.
ringe body upright and fill it with the vaccine.
9) V
erify the dose for the primary syringe by
3) O
nce the syringe is full, connect the vaccine
placing it into the top section of the syringe
line to the syringe inlet (ZT64M). (see page 14)
holder, connecting the air line, and activating
4) C
heck for any vaccine leaks around the inlet the machine 10 times into a graduated test vial
and outlet ports. (ZT84). For the secondary syringe, manually
pump the piston 10 times into a graduated
5) W
ithout touching the piston with your fingers,
test vial.
prepare to insert it into the syringe body.
10) Position both syringes in the syringe holder
6) B
e sure the piston and syringe body are lined
and check the combined dose of both
up perfectly before inserting the piston. Slowly
syringes by pressing the touch button trigger.
The dose can be verified by measuring the volume collected in a graduated test vial
(ZT84) after 10 injections. This process can be performed with or without a needle.
Very important: Make sure the O-rings, kit springs, tac valves and fiber washers are changed after
50,000 injections and the syringe is kept clean.
A SANOFI COMPANY
v.030315 8
Priming the syringe: Option 2
The syringe should remain intact during the priming process.
1) ang a bag or bottle of vaccine on the vaccine holder. The diluent bag may contain dye. Clamp the
H
vaccine line.
2) U
n-clamp the vaccine line and let the vaccine flow through the line via gravity. Pinch the end of the
vaccine line to stop the flow of vaccine once the line is completely filled.
3) With the syringe fully assembled, connect the vaccine line to the syringe inlet (ZT64M).
4) U
sing the vaccine line clamp, push the vaccine through the syringe until reach the syringe tip (ZT62M).
The syringe inlet may need to be slightly loosened during this process.
ZT64M
5) M
anually pump the piston several times until the system is completely primed. Fully tighten the syringe
inlet.
A SANOFI COMPANY
v.030315 9
9. Adjusting the Needle Position (and the Injection Level
and Depth)
Place the needle on the syringe.
Close the machine lid and supply the machine with compressed air. Turn the on/off test switch (ZT367) to the on
position and adjust the pressure to 5-6 bars (73-85 psi). The operator should be twice as careful to avoid any
injury from needle movements.
The syringe will thrust forward and lock into the injection position (see picture).
Extended position: the needle should always be at the center of the needle hole in the cover. If it is not
centered, adjust by loosening the hinge screws on the cover on the front of the machine.
Injection position:
Needle position adjustments must be made
with the needle in the injection position:
Injection point:
The needle length To adjust the point of injection, move the twin touch
is 14-18 mm plate by loosening the two screws (ZT410) that
holds the plate to the lid. Raising the twin touch
plate on the lid will lower the point of injection
in the bird and lowering it will raise the point of
injection. The position of the twin touch plate may
vary slightly depending on the size of the chicks and
operator skill level. There should be a 2-3 mm gap
between the needle and the side button.
2-3 mm
A SANOFI COMPANY
v.030315 10
10. Vaccinating Day-Old Chicks by Subcutaneous Injection
in the Neck
Seize the chick at the base of its neck with your left hand. With your right hand, hold the chick’s body and
stretch the neck gently. Position the bird on the machine. The head of the chick must be positioned so that it
pushes against both of the buttons in the twin touch plate. (see picture)
The quality of vaccination can be controlled by examining the necks of the birds vaccinated with a dye for water-
based vaccines. Using a dye is unnecessary for oil-based vaccines, which are generally of a pearly white color
and clearly visible under the skin.
The dye enables visualization of the injection site in the chick’s neck.
Before starting each vaccination, always inject 20 chicks and verify if the vaccination quality is correct. Observe
the chicks for 10 minutes to confirm that there is no discomfort or injury to the chick, or leakage of the vaccine.
Important: If the injection site is too close to the head, then move the twin touch plate up towards
the top of the machine. Recheck injection.
A SANOFI COMPANY
v.030315 11
11. Daily Maintenance
For any maintenance activity or replacement Weekly Maintenance:
of needle, always switch off the machine Lubricate one time per week the following
completely by turning the “on/off test switch” spare parts:
(ZT367) to the off position and disconnecting
the air to the machine. The pressure gauge • Air cylinder rod (ZT19)
should read 0 psi (0 bars).
A SANOFI COMPANY
v.030315 12
13. General Troubleshooting
With time and normal wear, the syringe is liable to deliver a higher dose at each vaccination. Do not hesitate to
replace it.
After a series of vaccinations, the Vaccination counter has reached Reset the counter to the num-
machine suddenly stops function- zero. ber of vaccinations desired by
ing and a continuous air flow can be pressing the reset button located
heard. on top of the machine or the top
button on the batch counter.
Syringe makes too many injections a) Insufficient displacement range Check function by manually
for a set volume; this means doses activating air circulation system.
b) Syringe out of order
are insufficient
c) Poor suction of the inlet tac valve Check syringe manually and
replace it if necessary. Replace
d) Needle diameter too small
valves and springs.
Syringe does not make enough Excessive displacement range Check manually (very important)
injections, doses are too large that syringe rear nut is correctly
tightened.
ZT80M Indicates
ZT81M Volumes used* ZT625NS
NS
ZT82M ZT627
ZT90 NS
ZT79M ZT63M
ZT622M
ZT77M
ZT83M ZT72M (0.1) ZT62M
ZT73M (0.2) ZT76M NS
ZT74M (0.3, 0.4, 0.5)
ZT626
ZT64M
A SANOFI COMPANY
v.030315 13
15. Next Generation Twin Touch Plate Troubleshooting
The machine activates Tubing going to the twin touch plate. Check for a crimp in the tubing going to the
when only one button is All parts twin touch plate when the lid is closed.
depressed.
Remove twin touch plate and check for pinched
or clogged airline. Remove buttons and sensors.
Inspect all parts for wear or damage clean and/
or replace if necessary.
The machine makes Inspect all components This issue is not typically associated with the
multiple injections when twin touch plate and is most likely a cause of
bird is placed in position dirt, water, oil etc. getting inside the pneumatic
for injection. controls. Inspect all components starting
with the impulse valve (ZT369) and working
backwards through the air logic system.
A SANOFI COMPANY
v.030315 14
16. Spare Parts
ZT57FM to ZT61FM
Primary Syringe
Complete
ZT57SM to ZT61SM
Secondary Syringe
Complete
Primary Syringe
Note: The primary syringe is shown without the connectors to attach the secondary syringe.
A SANOFI COMPANY
v.030315 15
Syringe Connectors for Primary and Secondary Syringes
Page 14 diagram
ZT626
ZT63M
ZT623NS
ZT621NS
Note: The exploded view of the completed primary syringe is shown on page 12.
Secondary Syringe
Secondary Syringe
ZT626
ZT621NS
A SANOFI COMPANY
v.030315 16
Case
A SANOFI COMPANY
v.030315 17
Pneumatic
Pneumatic
A SANOFI COMPANY
v.030315 18
Air Cylinder
A SANOFI COMPANY
v.030315 19
Accessories
ZT101
ZT84 ZT87
ZT88
ZT56
ZT85
ZT75
ZT66M
ZT68M
ZT70M
ZT77M ZT76M
ZT72M
ZT73M
ZT74M
A SANOFI COMPANY
v.030315 20
17. Spare Parts List
Reference Description
Machine Complete
Pneumatic
A SANOFI COMPANY
v.030315 21
ZT30 Mounting plate for pressure gauge with screws
ZT364 Label
A SANOFI COMPANY
v.030315 22
ZT408 Tubing kit for TTP
Case
ZT481NS Case
Syringe
ZT57CM Syringe optimized for single injection when using Zootec double
A SANOFI COMPANY
v.030315 23
ZT621NS Tip syringe NS
ZT627 Fiber washer repair kit for the inlet of the second syringe (x40)
ZT66M Repair kit for 50,000 inj. for 0.1 ml for water or oil based vaccine
ZT640 Syringe metal inlet for water & oil based vaccines
ZT66 Repair kit 50,000 injections for 0.1ml water or oil based
ZT661M Repair kit for 1 million inj for 0.1ml for water or oil based vaccine
ZT68M Repair kit for 50,000 inj for 0.2 ml for water or oil based vaccine
ZT68M-B Repair kit for 50,000 inj. for 0.15 ml for water or oil based vaccine
ZT681M Repair kit for 1 million inj for 0.2 ml for water or oil based vaccine
ZT681M-B Repair kit for 1 million inj. for 0.15 ml for water or oil based vaccine
ZT70M Repair kit for 50,000 inj for 0.3 ml, 0.4 ml, 0.5 ml for water or oil based vac
ZT701M Repair kit for 1 million inj for 0.3 ml, 0.4 ml, 0.5 ml for water or oil based vac
ZT74M Syringe piston O-ring for 0.3, 0.4, 0.5 ml syringe (x10)
A SANOFI COMPANY
v.030315 24
ZT84 Test vial
Needles
Screws
Accessories
A SANOFI COMPANY
v.030315 25
18. F
or Commercial orders and technical service, contact your
local Merial representative.
ONE WAY
VALVE
YELLOW TWIN
RESET OFF/ON MODE TOUCH
SWITCH SENSOR
SWITCH RELAY PLATE
Y-
FITTINGS
Y-
FITTINGS
IMPULSE
VALVE
AIR SUPPLY
FILTER
REGULATOR 4
1
FREE FLOW
VALVE 2
MANIFOLD
Y-FITTINGS MAC VALVE
GAUGE
Y
Z
Z PRIMARY
P A
Y-FITTINGS
TOTAL Y-FITTINGS SECONDARY
COUNTER
BATCH COUNTER
SIGNAL
A SANOFI COMPANY
v.030315 26
Notes
A SANOFI COMPANY
v.030315 27
Notes
A SANOFI COMPANY
v.030315 28
Notes
A SANOFI COMPANY
v.030315 29
Notes