SOP For Calibration of Viscometer
SOP For Calibration of Viscometer
-->
4.0 ACCOUNTABILITY
Head of the Department
5.0 PROCEDURE
Frequency of calibration: Once in three month
5.1 Operate the instrument as per SOP
5.2 Calibrate the instrument with standard viscosity solution of 10, 100, 500, 5000 & 12,500 cps using respective spindles and record the results in
Annexure-I
5.3 The performance of the instrument is satisfactory, if the measured viscosity is within the given tolerance limit, otherwise follow the S.O.P.
6.0 ABBREVIATION
6.1 0C = Centigrade
6.2 S.O.P. = Standard operating procedure
6.3 cps = Centipoise
Annexure-I
CALIBRATION REPORT OF BROOK FIELD VISCOMETER
INSTRUMENT DETAILS
INSTRUMENT NAME INSTRUMENT MAKE INSTRUMENT
IDENTIFICATION NO.
OBSERVATION
Sr. No. Fluid of Certified Spindle Speed factor Reading (B) Viscocity in Acceptance criteria %
Labeled Value in cps at 250C (A) cps AXB Tolerance from certified
viscosity in viscosity
cps
01 10 +/- 10%
02 100 +/- 7.5%
03 500 +/- 7.5%
04 5000 +/- 5.0%
05 12,500 +/- 5.0%