Function - Host
Function - Host
Function Description
System
Host
MAGNETOM ESSENZA
2 Document Version / Disclaimer / Copyright
Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents cor‐
respond to the version at the time of system delivery and/or printout. Versions to hard‐
copy documentation are not automatically distributed.
Please contact your local Siemens office to order a current version or refer to our website
http://www.healthcare.siemens.com.
Disclaimer
Siemens provides this documentation “as is” without the assumption of any liability under
any theory of law.
The content described herein shall be used by qualified personnel who are employed by
Siemens or one of its affiliates or who are otherwise authorized by Siemens or its affiliates
to use such documents.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are not entitled to use this documen‐
tation without prior written authority.
Copyright
“© Siemens, 2006” refers to the copyright of a Siemens entity such as:
Siemens Healthcare GmbH - Germany
Siemens Aktiengesellschaft - Germany
Siemens Shenzhen Magnetic Resonance Ltd. - China
Siemens Shanghai Medical Equipment Ltd. - China
Siemens Healthcare Private Ltd. - India
Siemens Medical Solutions USA Inc. - USA
Siemens Healthcare Diagnostics Inc. - USA and/or
Siemens Healthcare Diagnostics Products GmbH - Germany
1 Introduction 4
1.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.3 Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.3.1 MR console (MRC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.3.2 MR satellite console (MRSC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2 Host 8
1.1 General
The Host computer concept for MAGNETOM ESSENZA systems is the same as used for all
syngo MR systems, i e. the Host is based on PC hardware.
This is true for both the main console (MRC) and the optional satellite console (MRSC).
Fig. 1: MRC components
What is new compared to older systems is that the measurement control (AMC) and the
image reconstruction system (MRIR) are combined in one single computer system called
MARS (Measurement And Reconstruction System).
Fig. 2: MARS
1.2 Overview
Before you continue, a few words on the designations used throughout the Host descrip‐
tion may be in order to avoid unnecessary confusion.
The term Host refers simply to a computing element that is used to control a system or
act as its "host". The "host" of our MR system is termed "MR Console", “MRAWP-MR”, or
MRC for short.
In addition to the MRC, an optional satellite MR console, or MRSC is available. Although
practically the same as the MRC, it is only intended for post-processing images. For this
reason, it may not contain some of the hardware components found in the MRC. Also, the
software configuration will be different, since the MRSC does not require all that system
control software. A description of the main differences is provided on the next page.
Fig. 3: Host and MARS
1.3 Function
The MARS - MRAWP serial cable must be properly connected to both units. Otherwise,
MARS boot problems will occur.
The USB keyboard and mouse must be connected to the correct ports, as specified on
the label on the host. Otherwise, problems may occur when software is being installed
on the host.
The host can switch the system into standby by sending a signal to the appropriate relay
in the Line Power Distribution section. In standby mode, all system components are
switched off with the exception of the MRC and MRSC hosts.
Switching those components back on (exiting standby mode) is possible only via power-
on button in the alarm box. A remote switch-on from standby is not possible.