PATROL UPLC Process Analyzer System Guide Rev B
PATROL UPLC Process Analyzer System Guide Rev B
PATROL UPLC Process Analyzer System Guide Rev B
Analyzer
System Guide
Revision B
Trademarks
ACQUITY, ACQUITY UPLC, NuGenesis, PATROL UPLC, UPLC, and Waters
are registered trademarks of Waters Corporation, and Empower and “THE
SCIENCE OF WHAT’S POSSIBLE.” are trademarks of Waters Corporation.
Phillips is a registered trademark of Phillips Screw Co.
TORX is a registered trademark of Textron, Inc.
Windows is a registered trademark of the Microsoft Corporation.
Other registered trademarks or trademarks are the sole property of their
owners.
ii
Customer comments
Waters’ Technical Communications department invites you to tell us of any
errors you encounter in this document or to suggest ideas for otherwise
improving it. Please help us better understand what you expect from our
documentation so that we can continuously improve its accuracy and
usability.
We seriously consider every customer comment we receive. You can reach us
at [email protected].
Contacting Waters
®
Contact Waters with enhancement requests or technical questions regarding
the use, transportation, removal, or disposal of any Waters product. You can
reach us via the Internet, telephone, or conventional mail.
Safety considerations
Some reagents and samples used with Waters instruments and devices can
pose chemical, biological, and radiological hazards. You must know the
potentially hazardous effects of all substances you work with. Always follow
iii
Good Laboratory Practice, and consult your organization’s safety
representative for guidance.
When you develop methods, follow the “Protocol for the Adoption of Analytical
Methods in the Clinical Chemistry Laboratory,” American Journal of Medical
Technology, 44, 1, pages 30–37 (1978). This protocol addresses good operating
procedures and the techniques necessary to validate system and method
performance.
iv
Warning: To avoid igniting solvents, do not use an electrical stirrer
inside the enclosure.
Requirements:
• Ensure that a master AC disconnect is located near the system and is
easily accessible to system users.
v
• Ensure that use of the PATROL system, including any emergency
procedures, is consistent with the safety practices followed by your
organization. If a serious malfunction occurs that requires you to
®
immediately remove AC power from the PATROL UPLC Process
Analyzer system, press the emergency-stop button located on the front of
the options compartment.
• Locking the PATROL system enclosure restricts access to system
components and reduces the safety risks of users who have not received
specific training from Waters on performing tasks within the enclosure.
Users who have been trained by Waters to perform tasks inside the
enclosure, must observe all safety warnings and cautions associated
with the PATROL UPLC Process Analyzer system and enclosure.
• For installations outside of the United States, a grounded AC plug must
be provided by the customer.
Safety advisories
Consult Appendix A for a comprehensive list of warning and caution
advisories.
Disposal
Do not dispose of instrument components in municipal waste.
Dispose of the PATROL UPLC Process Analyzer system according to local,
state, provincial, federal, and international regulations.
Contact Waters with questions regarding the transportation, removal, or
disposal (consistent with the WEEE Directive) of any Waters product.
vi
Applicable symbols
Symbol Definition
Authorized representative of the European
Community
vii
Calibrating
To calibrate LC systems, follow acceptable calibration methods using at least
five standards to generate a standard curve. Include the entire range of QC
samples, typical specimens, and atypical specimens in the concentration range
for standards.
When calibrating mass spectrometers, consult the calibration section of the
operator’s guide for the instrument you are calibrating. In cases where an
overview and maintenance guide, not operator’s guide, accompanies the
instrument, consult the instrument’s online Help system for calibration
instructions.
Quality control
Routinely run three QC samples that represent subnormal, normal, and
above-normal levels of a compound. Ensure that QC sample results fall within
an acceptable range, and evaluate precision from day to day and run to run.
Sometimes, data collected when QC samples are out of range is invalid. Do not
report these data until you are certain that the instrument performs
satisfactorily.
ISM classification
viii
EC Authorized Representative
Telephone: +44-161-946-2400
Fax: +44-161-946-2480
Contact: Quality manager
ix
x
Table of Contents
Copyright notice ................................................................................................... ii
Trademarks ............................................................................................................ ii
Customer comments ............................................................................................ iii
Table of Contents xi
Accessing the system through the software interface ........................... 1-11
Topic Page
System technology 1-2
System components 1-2
Accessing the system through the software interface 1-11
Determining tasks allowed for a user level 1-11
1-1
System technology
The PATROL system is a process analytical technology (PAT) system. Its
UPLC® technology produces real-time, quantitative results. The PATROL
®
system is part of the same technology platform as Waters’ ACQUITY UPLC
System. The system performs process sampling, sample handling,
chromatographic analysis, and data management tasks.
A system enclosure facilitating walk-up sample submission houses the system
components.
To use the system, you introduce individual, bar-coded vials to the process
sample manager, which then reads the bar codes, accepts or rejects the vials,
and runs the specified test procedures. You can also configure the process
sample manager to aspirate samples directly from a manufacturing process
stream.
The modular design of the PATROL system’s components lends itself to fast
and efficient maintenance and support.
System components
The PATROL UPLC Process Analyzer system comprises the following
elements:
• ACQUITY binary solvent manager
• ACQUITY column manager (CM-A)
• ACQUITY auxiliary column manager (CM-Aux), optional
• ACQUITY detector (TUV, PDA, or eλPDA)
• ACQUITY process sample manager (PSM)
• Leak sensors
• Computer, with touch screen and keyboard
™
• PATROL and Empower software
• Solvent and waste storage
• Solvent recycling valve module, optional
• Options compartment
System enclosure:
Light tower
Vent
Emergency-stop
button (e-stop)
Instrument
compartment
Options
compartment
Computer touch
screen
Sample access
door (to access
process sample
manager)
AC power
See also: For more information on connections to the enclosure, see the
PATROL UPLC Process Analyzer Site Preparation Guide.
System components:
Light tower
E-stop
Detector Circuit
breakers
Auxiliary DC supply
column
manager System PC
Process
sample
manager Solvent
reservoirs
Binary
solvent
manager
Waste
reservoirs
(not shown)
Recycle valve
(not shown)
System enclosure
The system enclosure houses three adjoining compartments: instrument,
solvent and waste, and options. Each compartment is fitted with a separate
front-panel door, which you can secure using a locking mechanism, if desired.
Requirement: The front-panel doors interlock such that the solvent and waste
compartment door must be opened before opening either the instrument or
options compartment doors.
Tip: System administrators use the separate doors and locking mechanisms to
restrict access to specific user types.
Requirement: Locking the PATROL system enclosure restricts access to
system components and reduces the safety risks of users who have not
received specific training from Waters on performing tasks within the
enclosure. Users who have been trained by Waters to perform tasks inside the
enclosure, must observe all safety warnings and cautions associated with the
PATROL UPLC Process Analyzer system and enclosure.
The enclosure includes a light tower on the top surface to indicate system
status:
Instrument compartment
Inside the instrument compartment (left-hand side of the enclosure), each of
the system components rests on a sliding drawer so that, when the
compartment door is fully open, you can slide the components forward,
enabling access to panels and connections.
Note: An optional solvent recycling valve module is also located in the
instrument compartment if it is included as part of the system.
When the main instrument compartment door is closed, you can open a
smaller entry door to directly access the process sample manager from the
front of the enclosure.
Note: From the solvent and waste compartment, you can access a regulated
instrument air supply (758 kPa/110 psig/7.58 bar maximum pressure). You
can use the instrument air to supply an ancillary laboratory device (a
pneumatic stir plate for chromatographic solvents, for example), if desired.
See the PATROL UPLC Process Analyzer Site Preparation Guide for
additional details.
Options compartment
Authorized users can access the options compartment to perform installation
and maintenance functions.
ACQUITY components
The system uses four ACQUITY UPLC components (the binary solvent
manager, column manager, auxiliary column manager, and detector) along
with a process sample manager specific to the PATROL system. See the
Power button
and LEDs
Access panel
Fuse holder
Power connection
Ethernet connection
Sample door
(for vials)
Access panel
You can submit samples by using the process sample manager. When you
insert a bar-coded vial through the process sample manager’s door and into
the sample carousel, the process sample manager reads the bar code to verify
sample identity. It then performs any required dilutions and injects the
sample mixture onto the appropriate analysis column (located in the column
manager or auxiliary column manager). You can also configure the process
sample manager to aspirate samples directly from a manufacturing process
stream.
Process line #2
Process line #3
Process line #6
Process line #1
Process line #5
Sampling
valve
Priming
valve Injection
valve
Plugged
Waste
To column
Transducer
Transducer
Sensor manager
Waste Mixer
10 μL To binary
Diluent Process solvent
pump manager
Process
valve
Wash
Vial
Injection
port Wash pump
Topic Page
Configuring PATROL software 2-2
Logging in to Empower software from the touch screen 2-4
2-1
Configuring PATROL software
After the instrument control software for the PATROL system is installed, a
system administrator must configure a PATROL system in the Empower
software at system installation. Consult the Empower online Help or the
Empower 2 System Administrator’s Guide for details on how to configure a
new chromatographic system.
Configuring timeouts
The process sample manager requires you to complete certain tasks within a
specified time interval. An authorized PATROL user can configure intervals
(timeouts) that allow these operations:
• Insertion or removal of a sample or standard
• Removal of a rejected sample or standard
To configure timeouts:
1. In the process sample manager information window, click Configure >
Timeouts.
2. Specify appropriate time intervals for each operation.
Topic Page
Introducing a bar-coded sample vial into the system 3-2
Loading standards 3-4
Unloading an accepted, analyzed, or rejected vial 3-5
Configuring methods 3-5
3-1
Warning: To avoid spills and inadequate solvent supply, verify daily
that fluid reservoirs are at an appropriate level. Mobile-phase and
liquid-waste counters are not currently active.
To run a sample, initiate the vial loading process via the touch screen.
Loading standards
You must log in to the PATROL system to load standard vials.
To load standards:
1. Touch the Tech button in the lower right-hand corner of the PATROL
touch screen.
2. Enter a valid password, and then touch OK.
3. In the window that appears, touch Insert Standards.
4. Specify the vial position, method set, volume, and threshold for the
standard.
Configuring methods
You can configure instrument methods in Empower by using the process
sample manager’s instrument method editor.
Additional channels are available to use for diagnosing problems with the
system.
Topic Page
Maintenance schedule 4-2
Configuring maintenance warnings 4-3
Spare parts 4-4
Maintenance considerations 4-6
Accessing or removing instrument components 4-8
Adding or replacing solvents 4-11
Emptying waste containers 4-11
Switching columns 4-12
Relocating the system 4-12
Maintaining the process sample manager 4-13
Cleaning the system enclosure’s exterior 4-47
4-1
Maintenance schedule
For details about the routine maintenance of the binary solvent manager,
column manager, auxiliary column manager, TUV detector, PDA detector, or
eλPDA detector, consult the ACQUITY UPLC System Operator’s Guide and
the appropriate column compartments or detector information (on the
PATROL documentation CD).
Waters recommends that you perform the following routine maintenance on
the process sample manager to ensure reliable operation and accurate results.
For
Maintenance procedure Frequency
information...
Resolving leak sensor errors As needed See page 4-17
Replacing the leak sensor As needed See page 4-23
Replacing a sample loop When the See page 4-26
injection valve is
replaced during
scheduled
preventative
maintenance, or
as needed
Replacing the injection valve cartridge During scheduled See page 4-31
preventative
maintenance
Replacing the priming valve cartridge During scheduled See page 4-31
preventative
maintenance
For
Maintenance procedure Frequency
information...
Replacing the sampling valve During scheduled See page 4-31
cartridge preventative
maintenance
Replacing the process valve cartridge During scheduled See page 4-31
preventative
maintenance
Replacing the sample pump During scheduled See page 4-36
preventative
maintenance, or
as needed
Replacing the diluent pump During scheduled See page 4-39
preventative
maintenance
Replacing the process pump and As needed See page 4-41
installing the tubing
Replacing the mixer As needed (see See page 4-43
tip, below)
Calibrating the needle seal As needed See page 4-16
Cleaning the air filters During scheduled See page 4-44
preventative
maintenance, or
as needed
Replacing the fuses As needed See page 4-45
Spare parts
Replace only the component parts mentioned in this document and in the
ACQUITY UPLC System Operator’s Guide (on the PATROL documentation
CD). See the Waters Quality Parts Locator on the Waters Web site Services &
Support page.
Front of
drawer slide
TP03345
Fully extended
drawer slide
4. Press the release mechanism for the desired drawer, and slowly pull the
platform forward until it stops.
5. Remove the power and ethernet cables from the rear of the instrument.
Fuse holder
Ethernet connection
Power connection
1. Open the solvent/waste compartment door, and pull out the waste
drawer.
Tip: See “Adding or replacing solvents” on page 4-11 for information
about how to access the compartment and drawers.
2. Identify which waste container you are emptying.
3. Secure the end of the waste tubing to prevent spills.
4. Remove the waste container, and transfer the contents to an appropriate
disposal location.
5. Replace the waste container and tubing line.
6. Slide the drawer back into place, and close the front-panel door.
Switching columns
Access to the column manager to add/remove analysis columns can be
restricted to specific user types. Authorized users can find information on
performing these tasks in the column compartments user documentation,
provided on the PATROL documentation CD.
When you access the priming window, the sample needle moves to the
injection port position.
From the priming window, you can use the process pump to prime the process
valve and each of the six process lines.
You can select which section of the flow path to test for leaks:
• Sample – tests for leaks along the flow path from the sample pump to
the priming valve (the priming valve is blocked).
• Diluent – tests for leaks along the flow path from the diluent pump to
the priming valve (the priming valve is blocked).
• System – tests for leaks along the flow paths from both pumps.through
the injection port. The priming valve and the injection valve are
blocked.and the needle is in the injection port during the test. The leak
test includes the needle seal.
When you access the seal calibration window and start a calibration,
each step is highlighted as it is executed.
You can upgrade the process sample manager’s firmware from this window.
Required materials
• Clean, chemical-resistant, powder-free gloves
• Cotton swabs
• Nonabrasive, lint-free wipes
®
• T10 TORX screwdriver
• 4-mm hex wrench
2. Using the hex wrench, remove the two screws securing the bottom-front
panel to the process sample manager.
3. Remove the bottom panel from the front of the process sample manager
by gently pulling both the right-hand and left-hand edges toward you.
Top panel
Bottom panel
4. Locate the source of the leak, and make the repairs necessary to stop the
leak.
6. Lower the reservoir until the top of the leak sensor is accessible.
Leak sensor
Leak sensor reservoir
8. Tip: If you cannot easily manipulate the leak sensor after removing it
from its reservoir, detach the connector from the front of the instrument.
Connector
Ribbon cable
Prism
Lint-free wipe
TP02891
10. Roll up a nonabrasive, lint-free wipe, and use it to absorb the liquid from
the leak sensor reservoir and its surrounding area.
11. With a cotton swab, absorb any remaining liquid from the corners of the
leak sensor reservoir and its surrounding area.
12. Align the leak sensor’s T-bar with the slot in the side of the leak sensor
reservoir, and slide the leak sensor into place.
TP02892
T-bar TP03343
Slot in leak
sensor reservoir
13. If you detached the connector from the front of the instrument, reattach
it.
14. In the ACQUITY UPLC Console, select Process Sample Manager from
the system tree.
15. In the process sample manager information window, click Control >
Reset PSM, to reset the process sample manager.
Required materials
• Clean, chemical-resistant, powder-free gloves
• New leak sensor
• T10 TORX screwdriver
• 4-mm hex wrench
4. Lower the reservoir until the top of the leak sensor is accessible.
5. Press the tab to detach the leak sensor connector from the front of the
instrument.
Connector
Ribbon cable
Leak sensor
Leak sensor reservoir
TP02892
T-bar TP03343
Slot in leak
sensor reservoir
9. Plug the leak sensor connector into the front of the instrument.
10. Reattach the leak sensor reservoir to the process sample manager.
11. In the ACQUITY UPLC Console, select Process Sample Manager from
the system tree.
12. In the process sample manager information window, click Control >
Reset PSM, to reset the process sample manager.
Caution:
• To prevent bandspreading or excessive carryover, be sure to
reinstall a sample loop in exactly the same way that it was
previously installed, so that the same end goes into the same port.
• To avoid leaks, remove the O-rings prior to installing the sample
loop. The O-rings on the new sample loop hold the fittings in place
during shipping.
Replace the sample loop when it clogs or when its capacity is inappropriate for
your chromatographic need.
Required materials
• 1/4-inch open-end wrench
• ACQUITY UPLC sample loop
TP03023
4. Remove the sample loop and its fittings from the valve.
5. If you are going to reuse the sample loop, label each end with the port it
was connected to.
Tip: If you replaced the injection valve cartridge, do not install
previously used sample loops on the new cartridge.
6. Place the sample loop and its fittings in a contamination-free bag.
Caution:
• To prevent dead volume in the system, ensure that the ends of
the loop are fully inserted into the ports of the injection valve
before swaging the ferrules.
• Use only Waters approved ACQUITY UPLC loops with
ACQUITY UPLC fittings. Each loop must be fitted to its
injector.
9. Ensure the 2-piece ferrule is in place on one end of the sample loop. If
not, slide a 2-piece ferrule and compression screw onto one end of the
sample loop, and then seat the end into injection valve port 1.
10. Finger-tighten the fitting and then add 3/4-turn with the wrench.
11. Remove the fitting and inspect the ferrule to ensure it has not moved.
TP02506 TP02507
Correct Incorrect
12. Repeat step 8 through step 10 for the other end of the sample loop and
for injection valve port 4.
13. Reinstall the sample loop fitting into valve ports 1 and 4.
Process pump
Sampling
valve
Sample
Injection
pump
valve
Diluent
pump Process
valve
Priming
valve
Mixer
Required materials
• 2-mm hex wrench
• 1/4-inch open-end wrench
• Pump or valve cartridge
• Sample loop, if necessary
Requirement: When you replace the injection valve cartridge, you must
also replace the sample loop.
Set screw
TP03023
Set screw
Process line #1
Process line #2
To process
pump
Process line #4 Process line #5
Set screw
Plugged Plugged
Connected to To waste
port 1 of priming
valve
Needle Sensor
(from
process
pump)
TP03023
Set screw
Connected to port 3
Diluent of sampling valve
Transducer Transducer
(diluent pump) (sample pump)
5. Use the 2-mm hex wrench to remove the set screw at the 10 o’clock
position on the valve cartridge.
6. Remove the valve cartridge from the valve assembly by pulling it toward
you.
7. Carefully unpack the replacement valve cartridge.
8. Ensure that the groove in the cartridge housing and the groove on the
drive clamp are aligned. If they are not, turn the drive clamp until the
grooves line up.
Aligned grooves
Drive clamp
Required materials
• 3/32-inch hex wrench
®
• Phillips screwdriver
• Replacement sample pump
Sample pump
Transducer
6. Remove the three electrical connectors from the top of the sample pump.
Important: Use the Phillips screwdriver to remove the two screws
securing the sensor connector to the pump.
Motor
connection Sensor connection
Encoder
connection Fitting
Sample pump
Diluent pump
Mounting bracket
8. Grasp the front of the pump and remove it from the instrument.
9. Unpack the new pump, if necessary.
10. Position the pump on the mounting bracket
11. Using the hex wrench and mounting screws, secure the replacement
pump to the bracket.
12. Attach the sensor, motor, and encoder connectors to the appropriate
locations.
13. Replace the process sample manager’s top and front panels.
Required materials
• 3/32-inch hex wrench
• Phillips screwdriver
• T20 TORX screwdriver
• Replacement diluent pump
Diluent pump
Transducer
Sample pump
Mounting bracket
Diluent pump
Mounting bracket
11. Grasp the front of the pump and remove it from the instrument.
12. Unpack the new pump, if necessary.
Required materials
• T20 TORX screwdriver
6. Remove the two electrical connectors from the back of the process pump.
Electrical connections
Mounting screws
7. Using the TORX wrench, remove the two mounting screws securing the
process pump assembly to the instrument’s support platform.
8. Remove the pump assembly from the instrument.
Tip: The process pump tubing is connected to the flow path by two
finger-tight fittings. The fittings are pre-installed on new tubing
assemblies.
Finger-tight fittings
Screw
Front of process
sample manager
Filter panel
Screw
Metal support
Metal tab
Required materials
• Fuses (see Appendix C for type)
• Flat-blade screwdriver
6. Insert the new fuses into the holders and the holders into the power
entry module.
7. Reconnect the power cord to the power entry module.
Contents:
Topic Page
Warning symbols A-2
Caution symbol A-4
Warnings that apply to all Waters instruments A-4
Electrical and handling symbols A-11
A-1
Warning symbols
Warning symbols alert you to the risk of death, injury, or seriously adverse
physiological reactions associated with an instrument’s use or misuse. Heed
all warnings when you install, repair, and operate Waters instruments.
Waters assumes no liability for the failure of those who install, repair, or
operate its instruments to comply with any safety precaution.
Specific warnings
The following warnings can appear in the user manuals of particular
instruments and on labels affixed to them or their component parts.
Burst warning
This warning applies to Waters instruments fitted with nonmetallic tubing.
Important: Toute modification sur cette unité n’ayant pas été expressément
approuvée par l’autorité responsable de la conformité à la réglementation peut
annuler le droit de l’utilisateur à exploiter l’équipement.
注意:未經有關法規認證部門允許對本設備進行的改變或修改,可能會使使用者喪失操作該設
備的權利。
注意:未经有关法规认证部门明确允许对本设备进行的改变或改装,可能会使使用者丧失操
作该设备的合法性。
注意:規制機関から明確な承認を受けずに本装置の変更や改造を行うと、本装置のユー
ザーとしての承認が無効になる可能性があります。
警告:當在有壓力的情況下使用聚合物管線時,小心注意以下幾點。
• 當接近有壓力的聚合物管線時一定要戴防護眼鏡。
• 熄滅附近所有的火焰。
• 不要使用已經被壓癟或嚴重彎曲管線。
• 不要在非金屬管線中使用四氫呋喃或濃硝酸或濃硫酸。
• 要了解使用二氯甲烷及二甲基亞楓會導致非金屬管線膨脹,大大降低管線的耐壓能力。
警告:圧力のかかったポリマーチューブを扱うときは、注意してください。
• 加圧されたポリマーチューブの付近では、必ず保護メガネを着用してください。
• 近くにある火を消してください。
• 著しく変形した、または折れ曲がったチューブは使用しないでください。
• 非金属チューブには、テトラヒドロフラン(THF)や高濃度の硝酸または硫酸などを流
さないでください。
• 塩化メチレンやジメチルスルホキシドは、非金属チューブの膨張を引き起こす場合が
あり、その場合、チューブは極めて低い圧力で破裂します。
Attention: L’utilisateur doit être informé que si le matériel est utilisé d’une
façon non spécifiée par le fabricant, la protection assurée par le matériel risque
d’être défectueuses.
警告:使用者必須非常清楚如果設備不是按照製造廠商指定的方式使用,那麼該設備所提供
的保護將被消弱。
警告:使用者必须非常清楚如果设备不是按照制造厂商指定的方式使用,那么该设备所提供
的保护将被削弱。
警告: ユーザーは、製造元により指定されていない方法で機器を使用すると、機器が提供
している保証が無効になる可能性があることに注意して下さい。
警告 : 為了避免火災,更換保險絲時,請使用與儀器保險絲蓋旁面板上所印刷之相同類
型與規格的保險絲。
警告 : 为了避免火灾,应更换与仪器保险丝盖旁边面板上印刷的类型和规格相同的
保险丝。
警告 : 火災予防のために、ヒューズ交換では機器ヒューズカバー脇のパネルに記
載されているタイプおよび定格のヒューズをご使用ください。
Electrical symbols
These can appear in instrument user manuals and on the instrument’s front
or rear panels.
Electrical power on
Standby
Direct current
Alternating current
Fuse
Keep upright!
Keep dry!
Fragile!
Use no hooks!
Contents
Topic Page
Preventing contamination B-2
Items exposed to solvent B-2
Solvents used to prepare mobile phases B-2
B-1
Preventing contamination
For information on preventing contamination, visit www.waters.com and refer
to Controlling Contamination in LC/MS Systems (part number 715001307).
Materials of construction:
Item Material
Wetted materials 316 SST, UHMWPE, synthetic
sapphire, synthetic ruby, FEP,
PTFE, ETFE, diamond-like coating,
PEEK, PEEK alloys, titanium alloys
Process sample manager, chassis 5051-H32 Aluminum
Process sample manager, external 5051-H32 Aluminum
panels
Process sample manager, front 5051-H32 Aluminum
panel
Enclosure panels 304 SST, 11 ga and 13 ga
Enclosure finish Number 4 finish
Drip trays Polyethylene
Caution: To avoid damage to the system, do not use the solvents listed
below on the PATROL system.
• Toluene
• Methylene chloride
• Trichlorobenzene
• Ethyl acetate
• Chloroform
• Chlorinated solvents
• Strong acids (greater than 5% v/v)
• Hexane
• Tetrahydrofuran
Important: Tetrahydrofuran and hexane are not recommended for use on the
PATROL system. This restriction differs from solvent recommendations of
some other ACQUITY UPLC systems.
See also: For the specifications of the binary solvent manager, column
manager, and auxiliary column manager, and detectors, consult the
appropriate user documentation on the PATROL documentation CD.
Contents
Topic Page
Specifications for the PATROL UPLC Process Analyzer C-2
Specifications for the process sample manager C-3
C-1
Specifications for the PATROL UPLC Process Analyzer
Attribute Specification
Height 72.0 inches (77.0 inches with light tower installed)
Width 30.0 inches
Depth 33.0 inches
Weight 1,019 pounds
Rear clearance 0.5 inch, minimum clearance required
Side clearance 0.5 inch, minimum clearance required
Attribute Specification
Moisture protection
a Normal (IPX0)
Voltage range 100 to 240 VAC
Frequency 50/60 Hz
Power consumption 1000 VA nominally
a. Moisture Protection – Normal (IPX0) – IPX0 means that no Ingress Protection against any
type of dripping or sprayed water exists. The X is a placeholder that identifies protection
against dust, if applicable.
Attribute Specification
Operating temperature 4 to 35 °C (39.2 to 95 °F)
Operating humidity 20% to 80%, noncondensing
Shipping and storage temperature -40 to 60 °C (-22 to 140 °F)
Shipping and storage humidity 10% to 90%, noncondensing
See also: For the performance specifications of the PATROL UPLC Process
Analyzer System, consult the information for the individual instrument
C-2 Specifications
components, available in the user documentation on the PATROL
documentation CD.
Attribute Specification
Height 21.5 inches
Width 13.5 inches
Depth 26 inches
Weight 110 pounds
Rear clearance 2 inches, required minimum clearance
Side clearance 1 inch, required minimum clearance
Attribute Specification
Pollution degree
a 2
Moisture protection
b Normal (IPX0)
Attribute Specification
Operating temperature 4 to 40 °C (39.2 to 104 °F)
Operating humidity 20% to 80%, noncondensing
Shipping and storage temperature -30 to 60 °C (-22 to 140 °F)
Shipping and storage humidity 10% to 90%, noncondensing
Attribute Specification
Maximum sample capacity 32 available sites for process and
control samples
Number of sample injections Up to 60 injections per sample vial
Sample Access Total random access
Injection modes Full-loop injections only. Injected
material can be from a vial or process
source.
Injection loop volumes 1-μL, 2-μL, or 5-μL sample loops
Dilution range Up to 1:50 in 0.1 increments
Injection precision (without dilution) Less than 0.5% RSD for caffeine, full
loop injection
Injection precision with dilution Less than 1.0% RSD for caffeine, full
loop injection
Injection linearity Greater than 0.999 coefficient of
deviation
1 to 50 dilution factor
Wash solvent consumption Variable (configured by user)
Sample carryover Less than 0.01% of the previous
injection for caffeine
C-4 Specifications
Process sample manager performance specifications: (Continued)
Attribute Specification
Diluent solvent consumption Variable (configured by user).
Minimum of 500 μL per injection for
wash.
Minimum injection cycle time 120 seconds (vial), or 180 seconds
(process line)
Minimum sample volume required 1,200 µL must be dispensed into the
vial in order to perform an injection.
Maximum sample consumed per 100 μL
analysis (vial)
Sample compartment temperature 4 to 40 ºC programmable in 0.1 ºC
control increments
Sample compartment temperature ± 0.5 ºC from set point measured at
stability the control sensor
Sample compartment temp. ± 1.0 ºC from set point measured at
accuracy the control sensor
Sample compartment temp. ± 3 ºC from set point as measured in
distribution air, not in a vial
Sample compartment cooling time Less than 60 minutes from 25 ºC to
4 ºC
Audible noise Less than 65 dBA
Index-1
sliding drawer 1-7 login
equipment guidelines vi, A-4 advanced access 1-11, 2-1
Empower 2-4
F Tech 2-4, 3-4, 3-5
fan speed 1-6, 1-7 Windows 1-11, 2-4
ferrule installation 4-29
filters, cleaning 4-44 M
firmware, upgrade 4-17 maintaining the system 4-2, 4-3
flow path manufacturing processes 1-2, 1-9, 3-1,
process sample manager 1-10 3-4
front-panel buttons, process sample master AC disconnect v
manager 1-11 materials of construction B-1
full-loop injections 1-11 method editor 3-5
fuses, type 4-45, C-3 method set 2-3
method sets, selecting 2-2
H methods 3-5
handling symbols A-12 mixer 4-43
hexane and tetrahydrofuran B-3 mobile phases, solvents used in
high priority sample 3-4 preparation of B-2
I moving the system 4-12
injection cycles, process sample N
manager 4-4 needle seal
injection valve 4-27, 4-32, 4-35 calibrating 4-16
injections, full-loop 1-11 NuGenesis Scientific Data
inserting a vial 2-4 Management System 1-11
instrument air 1-8
instrument compartment 1-7 O
instrument methods 3-5 online decision-maker file 2-4
intended use vii online prime 4-14
interface, basic software 1-12 online valve control 4-14
ISM classification viii options compartment 1-8
L P
laser, bar-code reader v, C-3 PATROL system
leak sensor 4-18, 4-23 administration 2-1
leak test 4-15 compatible solvents B-1
light tower 1-3, 1-6, 2-2 components 1-2
loading a sample or standards 3-3, 3-4 maintenance and support 1-2
locking the enclosure vi materials of construction B-1
Index-2
software 1-11 replacing
user types 1-11 diluent pump 4-39
PDA detector 4-2 fuses 4-45, 4-46
preventing contamination B-2 mixer 4-43 I
priming 4-35 process pump 4-41
online 4-14 sample loop 4-26
priming valve 4-34 sample pump 4-36
priority sample 3-4 valve cartridge 4-31
process analytical technology (PAT) 1-2
process pump 4-41 S
replacing 4-41 safety advisories A-1
process sample manager (PSM) 1-2 sample carousel 1-11
diluent pump, replacing 4-39 sample door 1-11
disconnecting 4-9 sample loop 4-31
flow path 1-10 replacing 4-26
front-panel buttons 1-11 sample pump 4-36
fuses, replacing 4-46 replacing 4-36
injection cycles 4-4 samples
needle seal, calibrating 4-16 high priority 3-4
process pump, replacing 4-41 process stream 1-2, 1-9, 3-1, 3-4
sample door 1-11 status 2-3
sample loop, replacing 4-26 submitting 1-9, 3-1
sample pump, replacing 4-36 sampling valve 4-33, 4-35
software control 1-11 scientific data management 1-11
submitting samples 1-9 screen saver interface 1-12
valve cartridge, replacing 4-31 SDMS 1-11
process sampling 1-2, 1-9, 3-4 secondary containment 1-7
process valve 4-33 selecting a project 2-2
project, selecting 2-2 selecting method sets 2-2
pump control 4-13 sliding drawer 1-7, 4-8, 4-9
pumps 4-30, 4-36, 4-39, 4-41 software 1-11, 2-2
purpose and audience vii solvent recycling valve 1-7
solvents
R compatible B-1
release lever 4-8 exposure of PATROL system
relocating the system 4-12 components to B-2
remote connection 1-11 managing supply 4-11
removing manifold 1-7
instrument components 4-8 reservoir configurations 1-8
vials 2-4, 3-5
Index-3
supply 1-7 inserting or removing 2-4, 3-3, 3-4,
use in mobile phases B-2 3-5
spare parts 4-4 size 3-2
standards 3-4 viewing the carousel 3-5
status, sample 2-3
submitting samples 3-1 W
switching columns 4-12 walk-up sample submission 3-1
symbols warning symbols A-2, A-4
caution A-4 warnings and cautions, PATROL
electrical A-11 specific 4-6
handling A-12 waste containers 1-7
warning A-2 emptying 4-11
system waste outlet 1-7
administration 2-1 Windows login 1-11, 2-4
components 1-2 Windows Remote Desktop Connection
enclosure 1-3, 4-1 software 1-11, 2-4
relocating 4-12
software 1-11
system enclosure 1-6
T
Tech login 2-4, 3-4, 3-5
tetrahydrofuran and hexane B-3
thermal management 1-6, 1-7
timeouts 2-4, 3-5
touch screen, computer 1-11, 2-4, 3-3
training 1-12
training, for PATROL users vi
TUV detector 4-2, 4-8
types of users, PATROL system 1-11
U
upgrade firmware 4-17
user access 1-12
user type, Empower 1-12
V
valve control 4-13
valves 4-30, 4-32, 4-33, 4-34, 4-35
vials
Index-4