KF Quick Reference Guide Method Parameters

This document provides definitions and default parameters for Karl Fischer titration methods. It explains key terms like polarization voltage, drift correction, stabilization time and stop criteria. It also gives default parameters for volumetric and coulometric KF titration and the calculations for determining water content from titration data.

Uploaded by

aldi_dudul
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
74 views2 pages

KF Quick Reference Guide Method Parameters

This document provides definitions and default parameters for Karl Fischer titration methods. It explains key terms like polarization voltage, drift correction, stabilization time and stop criteria. It also gives default parameters for volumetric and coulometric KF titration and the calculations for determining water content from titration data.

Uploaded by

aldi_dudul
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

KARL FISCHER QUICK REFERENCE

KF Method Parameters Glossary

I(pol) – Polarization voltage applied to the platinum double-pin electrode during


measurement.

Start drift – Consumption of KF reagent during conditioning. Since the cell is not
absolutely impervious, we need to establish a baseline water content of the cell. This value
sets the acceptable drift value for the start of the determination (i.e. cond OK). Drift
accounts for water entering the system and is labeled as µL of KF reagent needed per
minute.

Drift Correction – A value obtained by taking the time of the titration multiplied by the drift
value. That amount is subtracted from the added volume. Default is off. Parameter might
be helpful if water content is very low or if a long titration time is necessary.

Stabilizing time – Time requirement for the conditioned cell to be stable before “cond OK”

Dynamics – The control range before the specified endpoint. Within the control range,
dosing is slowed until the minimum volume increment has been reached. Outside the
control range, dosing is continuously carried out at the maximum rate.

Stop Criterion – Required as drift because EP alone is not sufficient for determining
endpoint since some fluctuation in water will always be detected.
Drift – Considered an absolute drift value. The titration will stop once the EP and
this drift value have been reached.
Rel. Drift – The stop drift is calculated as the sum of the drift value at the titration
start and the value entered here.

Extraction time – The titration will not stop until the extraction time has elapsed, even if
the EP has already been reached. This parameter is helpful when analyzing samples with
solubility challenges or also when working with the KF oven. Setting an extraction time will
keep the titration from ending prematurely.

Default KF Titration Parameters

Volumetric KF Coulometric KF
EP at U 250 mV 50 mV
Dynamics 100 mV 70 mV
Start drift 20 µL/min 20 µg/min
Stop criteria drift / time drift / relative drift
Stop drift 20 µL/min 5 µg/min
I(pol) 50 µA 10 µA
KARL FISCHER QUICK REFERENCE

Volumetric KF Calculations

titer = sample size / EP x standard concentration

% water = EP x titer x 0.1 / sample size

ppm water = EP x titer x 1000 / sample size

titer: mg/mL
sample size: g
EP: mL
Standard concentration: mg/g

Coulometric KF Calculations

% water = EP / sample size / 10000

ppm water = EP / sample size

EP: ug
sample size: g

Conversion Chart

1% = 10 mg/g = 10,000 ppm

0.1% = 1.0 mg/g = 1,000 ppm

0.01% = 0.1mg/g = 100 ppm