FM Operation Manual
FM Operation Manual
Table of Contents
a = Capacity: between 100kg and 10T; 200lb and 20klb; 1kN and 100kN.
b = Capacity: between 5kg and 500kg; 10lb and 1klb; 50N and 5kN.
d = Capacity: between 5 klb and 500 klb; 2 ton and 300 ton; 20 kN and 2.5 MN.
e = Capacity: any between 0.5 kg and 2500 kg; 1 lb and 5 klb; 5 N and 25 kN.
f = Capacity: between 200lb and 20klb; 100kg and 10000kg; 1kN and 100kN.
g = Capacity: between 20kg and 1000kg; 50lb and 2000lb; 200N and 10kN.
h = Capacity between 100kg and 100T; 200lb and 200klb; 1kN and 1MN.
i = Capacity between 100lbs and 10klbs.
x = Customer: F, M, N or W.
Preamble
This manual covers only the “Ex” relevant aspects.
Equipment Function
Flintec load cells are designed to be used in various kinds of industrial scales and meet the most stringent
accuracy requirements. Certifications have been obtained from Weights & Measures Authorities worldwide.
These load cells are available with different maximum capacities and include accuracy classifications
according to OIML R 60 and / or NTEP.
They offer stainless steel or aluminum construction sealed by welding or improved potting.
This makes them suitable for use in tough industrial environments.
The load cells can be used in all hazardous areas. The basic structure is always the same.
All standard equipment is provided with a 4-wire shielded conductor cable; equipment with the coding
extension –6w is provided with a 6-wire shielded conductor cable.
Entity Parameters:
VMax (Ui) = 29.9V, IMax (Ii) = 0.450A, Ci = 0, Li = 0.
The intrinsically safe circuit including the load cells must be built up with approved safety barriers
or switch amplifiers, matching the connected weighing indicator.
Control drawing
Warning: The specialist who installs the equipment must take responsibility
for proper operation in combination with various measuring equipment .
The intrinsically safe circuit including the load cells must be built up with approved safety barriers or switch
amplifiers, matching the connected weighing indicator.
Designation
All Flintec load cells follow the same electrical design and meet the requirements for category1 equipment.
Usage
WARNING : Misuse will cause the loss of warranty and manufacturer’s responsibility.
The load cells are only allowed for professional applications in accordance with the load cell data sheet and
Flintec application parts.
a) If the load cells are not powered from an intrinsically-safe circuit the connection cables must either be lead
out of the hazardous area to terminate them or terminated in suitable junction boxes.
b) If used in hazardous dust environment, the dust layer on the load cell body must not exceed
5 mm in thickness.
Maintenance
Maintenance interventions on the load cells have to be carried out by Flintec personnel only.
Repair
This equipment is certified for use in hazardous locations, therefore no modifications are allowed. Repairs
must only be performed by Flintec personnel.
Waste Disposal
The waste disposal of package and shipped parts must be done in accordance with the regulations of the
country in which the equipment is installed.