Higrometro-Ficha Técnia Avanzada
Higrometro-Ficha Técnia Avanzada
Higrometro-Ficha Técnia Avanzada
SCH M I DT
c o n t r o l i n s t r u m e n t s Edition DHT-2.01-E
Model DHT-2
Operating Instructions
Valid as of: 01.12.2016 • Please keep the manual for future reference!
Contents
1 Warranty and Liability............................................................................................... 3
1.1 Notices within the Operating Instructions................................................................ 3
1.2 Responsibilities of the Operating Company............................................................ 3
1.3 Responsibilities of the Personnel............................................................................ 3
1.4 Informal Safety Measures....................................................................................... 4
1.5 Training of the Personnel........................................................................................ 4
1.6 Intended Use........................................................................................................... 4
1.7 Dangers in Handling the Device.............................................................................. 4
1.8 Copyright................................................................................................................. 4
1.9 EU Declaration of Conformity and WEEE Registration........................................... 4
2 Available Models....................................................................................................... 5
2.1 Specifications.......................................................................................................... 5
2.2 Delivery Includes . .................................................................................................. 5
2.3 Optional Accessories.............................................................................................. 5
2.4 Unpacking............................................................................................................... 6
3 Operation.................................................................................................................... 6
3.1 Notes Before Starting Measurement....................................................................... 6
3.2 Operating Elements................................................................................................ 6
3.3 Battery Insertion and Replacement......................................................................... 6
3.3.1 Switch-On.......................................................................................................... 7
3.3.2 Switch-Off.......................................................................................................... 7
3.3.3 Inserting the Measuring Electrode..................................................................... 7
3.4 Measuring Procedure . ........................................................................................... 8
3.4.1 Conversion Charts............................................................................................. 9
3.5 Verification of Calibration...................................................................................... 14
5 Cleaning................................................................................................................... 14
6 Verification Interval................................................................................................. 14
7 Correspondence...................................................................................................... 14
8 Repairs..................................................................................................................... 15
2
1 Warranty and Liability
In principle, the supply of the device is subject to our “General Conditions of Sale and
Delivery.” These have been provided to the operating company on conclusion of the
contract, at the latest.
Warranty:
- SCHMIDT moisture meters are warranted for 12 months.
Parts subject to wear, electronic components and measuring springs are not covered by
the warranty. No warranty or liability will be accepted for bodily injury or property damage
resulting from one or several of the following causes:
- Misuse or abuse of the device.
- Improper mounting, commissioning, operation and maintenance of the device
(e.g. verification interval).
- Operation of the device if any safeguards are defective or if any safety and protection
precautions are not properly installed or not operative.
- Failure to comply with the notices in the Operating Instructions regarding transport,
storage, mounting, commissioning, operation, maintenance and setup of the device.
- Any unauthorized structural alteration of the device.
- Insufficient inspection of device components that are subject to wear.
- Opening the device or improper repair work.
- Disasters caused by the effects of foreign objects or by force majeure.
3
1.4 Informal Safety Measures
The Operating Instructions must always be kept on hand where the device is operated.
Apart from the Operating Instructions, the generally and locally valid regulations on acci-
dent prevention and environmental protection must be provided and complied with.
The device must not be operated in potential explosive areas and must not
come into contact with aggressive substances.
1.8 Copyright
The copyright on these Operating Instructions remains with the company
HANS SCHMIDT & Co GmbH.
These Operating Instructions are intended for the operating company and its personnel
only. They contain instructions and notices that may only be reproduced on the prior writ-
ten permission of
HANS SCHMIDT & Co GmbH
and under indication of the complete reference data.
Violations will be prosecuted.
4
2 Available Models
DHT-2
2.1 Specifications
Application Area: Direct readings for:
(Relative Moisture Content) Wool, cotton, rayon, hemp, jute, paper, linen, nylon,
polyacrylic, polyester, 10 fabric blends
Comparison measurement for other materials:
1 - 100 scale for comparison with 20 conversion charts
(Room Temperature) -30° to 70° Celsius (optional)
(Air Humidity) 5% to 95% RH (optional)
Display: LCD, 0.1 resolution
Accuracy: 1% (of full scale)
Repeatability: 0.3%
Power Supply: 9V E block battery, low-battery indicator
Auto Power Off: After approx. 2 min.
Weight: 1650 g (complete set incl. carrying case)
Dimensions: 200 mm x 95 mm x 40 mm (LxWxH)
5
2.4 Unpacking
Unpack the moisture meter and inspect it for any shipping damage.
Notices of defect must be filed immediately, in writing, at the latest within 10 days on
receipt of the goods.
3 Operation
3.1 Notes Before Starting Measurement
Have you read and understood the Operating Instructions, in particular
Chapter 1 “Basic Safety Notices”? You are not permitted to operate the
moisture meter before doing so.
Before working with the instrument you must put on your personal protective
clothing, if necessary. For example, eye protectors, gloves, etc.
fig. 3.2
3.3.2 Switch-Off
Auto Power Off:
- The moisture meter switches off automatically after about 2 minutes of non-use.
- Plug in the CONNECTING CABLE at the BNC connectors of the ELECTRODE and
the DHT-2 moisture meter (fig. 3.3.3b).
CONNECTING CABLE
DHT-2
ELECTRODE
fig. 3.3.3b
7
3.4 Measuring Procedure
General information:
The measuring principle of the DHT-2 moisture meter is based on the electrical conductiv-
ity of the material which always bears a fixed relation to the moisture.
The moisture meter displays direct readings for the materials listed in the table below. For
other materials, note the reading on the upper 0 - 100 scale and then refer to the conver-
sion charts in Chapter 3.4.1 to convert the absolute reading into an accurate measurement
of percent moisture.
Requirements:
- Proper electrode inserted and connected to DHT-2 as described in Chapter 3.3.3.
To take measurements:
- Press the or key for about 2 sec. to select the next or previous
8 measuring range (as listed in the table above).
3.4 Measuring Procedure (Cont.)
- Insert the ends of the selected ELECTRODE into the TEST MATERIAL (fig. 3.4).
The DISPLAY shows the measured moisture content of the TEST MATERIAL.
CONNECTING CABLE
DHT-2
fig. 3.4
ELECTRODE
Relative
moisture [%]
Acetate Yarn (CA)
9
3.4.1 Conversion Charts (Cont.)
Relative
moisture [%]
Sisal Yarn
Relative
moisture [%]
55% Polyester (PES) - 45% Rayon (Zw)
Relative
moisture [%]
33% Rayon (Zw) - 67% Cotton (Bw)
Relative
moisture [%]
40% Cotton (Bw) - 60% Flax Yarn (FL / Li)
Relative
moisture [%]
40% Polyacrylic (PAC / PAN) - 60% Wool (Wo)
10
3.4.1 Conversion Charts (Cont.)
Relative
moisture [%]
50% Cotton (Bw) - 50% Polynosic (PON)
Relative
moisture [%]
50% Polyacrylic (PAC / PAN) - 50% Wool (Wo)
Relative
moisture [%]
55% Polyacrylic (PAC / PAN) - 45% Wool (Wo)
Relative
moisture [%]
55% Polyester (PES) - 45% Wool (Wo)
Relative
moisture [%]
56% Rayon (Zw) - 44% Cotton (Bw)
11
3.4.1 Conversion Charts (Cont.)
Relative
moisture [%]
60% Polyacrylic (PAC / PAN) - 40% Wool (Wo)
Relative
moisture [%]
65% Polyester (PES) - 35% Rayon (Zw)
Relative
moisture [%]
70% Polyacrylic (PAC / PAN) - 30% Cotton (Bw)
Relative
moisture [%]
70% Wool (Wo) - 20% Rayon (Zw) - 10% Polyamide (PA)
Relative
moisture [%]
70% Wool (Wo) - 30% Rayon (Zw)
12
3.4.1 Conversion Charts (Cont.)
Relative
moisture [%]
75% Polyacrylic (PAC / PAN) - 25% Wool (Wo)
Relative
moisture [%]
80% Polyacrylic (PAC / PAN) - 20% Wool (Wo)
Relative
moisture [%]
80% Polyester (PES) - 20% Flax Yarn (FL / Li)
Relative
moisture [%]
80% Rayon (Zw) - 20% Wool (Wo)
Relative
moisture [%]
84% Rayon (Zw) - 16% Polypropylene (PP)
13
3.5 Verification of Calibration
A set of calibration electrodes with different calibration values is optionally available to
allow verifying the calibration of the DHT-2.
Connect the electrodes to the moisture meter. The instrument should then display the
calibration value specified by the manufacturer. To adjust the calibration of the DHT-2,
please request our special calibration instructions.
5 Cleaning
For cleaning the unit, do not use any
AGGRESSIVE SOLVENTS
NO WARRANTY OR LIABILITY
6 Verification Interval
The question of finding the right frequency of calibration accuracy verification depends on
several different factors:
Operating time and load of the SCHMIDT moisture meter
Tolerance band defined by the customer
Changes to the tolerance band compared to previous calibrations
Therefore, the interval between verifications of calibration must be determined by the
user‘s Quality Assurance Department, based on the user‘s experience.
Assuming normal operating time and load as well as careful handling of the moisture
meter, we recommend a verification interval of one year.
7 Correspondence
Should you have any questions regarding the instrument or Operating Instructions, or their
use, please indicate above all the following details which are given on the ID plate:
1) Model
2) Serial number
14
8 Repairs
Shipping instructions:
We kindly ask for return free of charge for us, if possible by airmail parcel. All occurring
charges, if any (such as freight, customs clearance, duty etc.), will be billed to customer.
For return from foreign countries, we ask you to include a proforma invoice with a low value
for customs clearance only, e.g. 50 Euro, each and to advise the shipment in advance by
fax or eMail.
Notes: