Lab 2 PDF
Lab 2 PDF
Lab 2 PDF
Glassware Calibration
1
Introduction:
Two basic types of equipment in an analytical laboratory are volume measuring devices and
balances.
Volumetric glassware used in analysis should be clean.
42
2
Very often Low-priced glassware does not meet these tolerances. For this
reason, it is often advisable for the student to calibrate his volumetric
glassware. Volumetric glassware can be calibrated to measure the volume that
is actually contained in or delivered by a particular piece of equipment.
Calibration is done by measuring the mass of water contained or delivered and
using Table 2-5 to convert mass into volume:
True volume (actual volume) = (mass of water) x (correction factor in table 4)
Objectives:
• This lab will introduce the basic laboratory equipment and skills you will need
for accurate and precise quantitative analysis.
• To determine through the calibration process the actual volume delivered by
volumetric tools, as opposed to the volume they say they deliver; this
information will be contained in correction factors and a burette correction
graph specific for your personal stash of glassware and you will use the data
throughout semester.
Laboratory Procedure:
A. Volumetric flask calibration
Figure 1 shows different kinds of
volumetric flasks.
1. Select a clean dry volumetric flask to
calibrate.
2. Record the mass of the empty, dry
volumetric flask.
3. Fill the flask to the mark with deionized
water and record the mass.
Because of the time constraints, we will not repeat this procedure-the flask would need to be
dried between measurements.
3
B. Pipette Calibration
Figure 2 shows different kinds of volumetric pipettes. Pipettes are made to deliver specific
volumes.
The following procedure will be used to calibrate your 10 mL pipette.
1. Determine the weight of an empty large weighing bottle or small beaker.
2. Fill the pipette to the mark with distilled water.
3. Empty the pipette into the weighing bottle or beaker. Be sure to touch off the pipette
on the side of the weighing bottle.
4. Weigh the container and its contents.
5. Repeat this three times to obtain total of three measurements. (Report the mass of
each individual addition- the net mass on the data sheet)
Figure 2
C. Calibrating a 50-mL Burette
This procedure tells how to construct a graph such as the figure 3 below to convert the
measured volume delivered by a burette to the true volume delivered at 20°C.
4
1. Fill the burette with distilled water and force any air bubbles out the tip (Figure 4).
o See whether the burette drains without leaving drops on its walls.
o If drops are left, clean the burette with soap and water soak it with cleaning
solution.
• Adjust the meniscus (Figure 4) to be at or slightly below 0.00 mL, a touch the burette
tip to a beaker to remove the suspended drop of water.
• Allow the burette to stand for 5 min while weigh a 100 mL flask (hold the flask with a
paper towel to prevent fingerprints for changing its mass).
• If the level of the liquid in the burette has changed, tighten the stopcock and rep the
procedure. Record the level of the liquid.
2. Drain approximately 10 mL of water into weighed flask, and cap it tightly to prevent
evaporation.
• Allow 30 s for the film of liquid on the walls to descend before you read the burette.
Estimate all readings to the nearest 0.01 mL.
• Weigh the flask again to determine the mass of water delivered.
3. Drain the burette from 10 to 20 mL, and measure the mass of water delivered.
5
aBurette Volume= (burette reading for this aliquot)-(Initial burette reading)
bMass of Water= (mass of this aliquot)-(initial mass)
cActual Volume = (mass of water) × correction factor-see Table 4 in book)
d Correction = (Actual volume) - (burette volume)
6
7