Ecube 15 Troubleshooting Guide

The document provides troubleshooting solutions for various symptoms that can occur with an E-CUBE 15 ultrasound system, including replacing boards to address artifacts, replacing cables to resolve boot issues, and replacing boards and power supplies for other problems.

Uploaded by

Павел Ф
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
213 views6 pages

Ecube 15 Troubleshooting Guide

The document provides troubleshooting solutions for various symptoms that can occur with an E-CUBE 15 ultrasound system, including replacing boards to address artifacts, replacing cables to resolve boot issues, and replacing boards and power supplies for other problems.

Uploaded by

Павел Ф
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 6

Troubleshooting Guide

for E-CUBE 15

Customer Service Team

HIGHLY
All information in the presentation are highly confidential and no part CONFIDENTIAL
of contents may be informed or transmitted without permission from ALPINION Medical 18
Systems.
Service Trouble Shooting

Symptom Some artifacts that are being occurred consecutively on PW mode

Solution Replace BE board to new one

HIGHLY CONFIDENTIAL 19 /
Service Trouble Shooting

Symptom Not generated analog signals (Composite Video, S-Video, VGA)

Solution Replace IO sub board to new one

HIGHLY CONFIDENTIAL 20 /
Service Trouble Shooting

System locked up in the middle of the progress bar when being booted on.
Symptom
(Especially 2/3 of progress bar)

Solution Reinstall OS and SW. if it’s not effected, replace SATA cable for HDD to a new one.

HIGHLY CONFIDENTIAL 21 /
Service Trouble Shooting

Symptom Irregular signals and strange Doppler sound on CW mode

Solution Replace CW board to new one

HIGHLY CONFIDENTIAL 22 /
Service Trouble Shooting

Symptom All probes are recognized with same thing.

Replace AC PSU (Bottom board) to new one


Solution
(Bottom board of E-CUBE 15 is same with E-CUBE 9)

HIGHLY CONFIDENTIAL 23 /