Final Measurement Lab. Manual
Final Measurement Lab. Manual
Final Measurement Lab. Manual
Course Outcomes:
At the end of the course a student will be able to –
1. Identify different analogue & digital instruments both AC and DC, source and sink devices,
their specifications, constructions using basic knowledge of electrical measurement.
2. Perform the experiments, interpret measured data and compare the measured value with
the true value of a quantity, calculate error in measurement, draw calibration & error curve
using appropriate techniques.
3. Develop the concept of calibration and understand the limitations of the different measuring
instruments.
5. Learn the necessity of safety measures of using different instruments and handling of high
voltage AC.
6. Work as a member in a team, communicate with each other, and share their independent
thinking to perform the experiment successfully
TECHNO MAIN SALTLAKE
Name of Department: Electronics andInstrumentation Name of Subject with code: Electrical &
Engineering Electronic Measurement Lab [PC-EI491]
Objectives:
Objective/Criteria
Lab Participation
Lab Report
Interaction with
Group
Safety
Cleaning / Rearranging
after
experiment
Total Marks with grade
20
Mark Evaluation : Between 20-16 O (5)
Between 15-11 E (4)
Between 10-6 A (3)
Between 5-0 B (2)
Comments:
Experiment No.1
As the potentiometer measures voltage, it can be measure current by measuring the voltagedrop
produced by the unknown current passing through a known standard resistance. To begin a
measurement with potentiometer it is to be standardizing with a reference voltage. This is done
by adjusting the series rheostat in between potentiometer supply and potentiometer terminals.
When a null
point is obtained, with the jockey placed at the positive end of potentiometer, the voltage across
the potentiometer = reference voltage.
When a voltmeter is to be calibrated, say, a null point is obtained at 𝐿 cm from the ground terminaland
voltmeter is connected to 𝑃 volt terminal of voltage ratio box.
1.2×𝐿×𝑃
Then the voltage across voltmeter =
1000×1.5
Reference voltage being 1.2 volts. When an ammeter is to be calibrated. Say a null point is obtained
at 1 cm from the ground terminaland ammeter is connected in series with a standard shunt of say one
ohm. The current through
1.2×𝐿
ammeter = (Resistance of shunts)
1000×1
Specifications:
Potentiometer resistance = 20 ohms
Potentiometer power supply = 2 volts
Series resistance with potentiometer circuit = 25 ohm
Reference voltage = 1.21volts, shunt = 1 ohm. Current capacity 10 amp.
Procedure
1. Ensure power is not switched on.
2. Make connections as per the circuit diagram.
3. Put both current set and voltage set to zero position.
4. Check connections once more, ensure it is correct.
5. Switch on the power.
6. Connect ‘slide wire input’ to potentiometer supply positive(red) terminal.
7. Press ‘Press to calibrate button’ adjust calibrate pot to obtain null point.
Potentiometer calibration is over. Now the potentiometer supply is 1.21v v DC.
8. Connect ammeter as per the circuit.
9. Set current to say 1 amp. Set sensitivity to minimum. Adjust Jockey position to obtain
nullpoint. Put Sensitivity maximum make adjustment for null point (Fine adjustment).
Techno Main Salt Lake
Department of Electronics & Instrumentation Engineering
APPARATUS:
Precautions:
1. Ensure connections are perfect before switching on, wrong connections may damage
theequipment, and/or may cause accident to the attendants.
2. Turn both set pots. To zero position, before starting and end of the experiments.
3. Do not exceed the specified voltage and current limit. Exceeding the limit may damage
the meters. Recommended maximum limits are,
Voltage = 120v DC
maxCurrent = 1A DC max.
4. Set sensitivity to minimum, while searching for null point to avoid damage to
galvanometer.
5. Do not keep pressed the jockey key if the galvanometer pointer exceeds the end limits.
Techno Main Salt Lake
Department of Electronics & Instrumentation Engineering
RESULT:
Experiment No.2
Theory: The Kelvin double bridge is a modification of the Wheatstone bridge and provides
greatly increased accuracy in measurement of low value resistances. It incorporates the idea of
a second set of ratio arms-hence the name double bridge – and the use of four terminal resistors
for the lowresistance arms. The first ratio arms are P and Q. The second set of ratio arms, p and
q is used to connect the galvanometer to a point c at the appropriate potential between points
m and n to eliminate the effect of connecting lead of resistance r between the known resistance
R and the standard resistance S.
The ratio p/q is made equal to P/Q. Under balance conditions there is no current through the
galvanometer which means that the voltage drop between a and d, Eab is equal to voltage drop
Eamc between a and c.
Equation (2) is the usual working equation for the Kelvin bridge. It indicates that the resistance of
connecting lead, r, has no effect on the measurement, provided that the two sets of ratio arms
haveequal ratios. Equation (1) is useful, however, as it shows the error that is introduced in case
the ratios are not exactly equal. It indicates that it is desirable to keep r as small as possible in
order to minimize the errors in case there is difference between the ratios P/Q and p/q.
APPARATUS:
PROCEDURE:
RESULT:
Experiment No.3
Theory: Wien’s Bridge is primarily used for determination of an unknown frequency. However, it can
be used for various other applications including capacitance measurement, in harmonic distortion
analyzers, where it is used as notch filter, and also in audio and HF oscillators. It has a series RC
combination in one and a parallel RC combination in the adjoining arm.
Under balanced condition, since no current flows through the detector, nodes B and C are at the same
potential. i.e, V1 = V2 and V3 = V4
𝑅 𝐽
(1+𝑗𝜔𝐶1 𝑅 ) ( 𝑅2 − 𝜔𝐶 )
1 1 2
At balance, =
𝑅3 𝑅4
𝑅1 𝑅4 𝜔 𝐶2 𝑅2 𝑅3−𝐽 𝑅3
Or, =
1+𝑗𝜔𝐶1 𝑅1 𝜔𝐶2
Techno Main Salt Lake
Department of Electronics & Instrumentation Engineering
𝐶2 𝑅1 𝑅4 = 𝐶2 𝑅2 𝑅3 + 𝐶1 𝑅1 𝑅3
𝑅4 𝑅2 𝐶1
= + ----------------------------- (1)
𝑅3 𝑅1 𝐶2
𝜔2 𝐶1 𝐶2 𝑅1 𝑅2 𝑅3 = 𝑅3
1
𝜔=√
𝐶1 𝐶2 𝑅1 𝑅2
1
𝑓= ---------------------------- (2)
2𝛱√ 𝐶1 𝐶2 𝑅1 𝑅2
R1 = R2 = R and C1 = C2 = C
1
Then we have, 𝑓= 2𝛱𝑅𝐶
Sliders for resistors R1 and R2 are mechanically coupled to satisfy the criteria R 1= R2
Wien’s bridge is frequency sensitive. Thus, unless the supply voltage is purely sinusoidal, achieving
balance may be troublesome, since harmonics may disturb balance condition. Use of filters with the
null detector in such cases may solve the problem.
Techno Main Salt Lake
Department of Electronics & Instrumentation Engineering
APPARATUS:
PROCEDURE:
RESULT:
Experiment No.4
Theory: AC bridges are most popular, convenient and accurate instruments for measurement of
unknown inductance, capacitance and some other related quantities. In its simplest form, ac
bridges can be thought of to be derived from the conventional dc Wheatstone bridge. An AC
bridge, in its basic form, consists of four arms, with the four arms being represented by four
unspecified impedancesZ₁, Z₂, Z₃, Z₄, an alternating power supply, and a balance detector.
I3
Let
𝐿1 = self-inductance to be measured
𝑅1 = resistance of self-inductor
𝑟1 = resistance connected in series with inductance to be measured
𝑟1, 𝑅2, 𝑅3, 𝑅4 = known non-inductive resistances
𝐶 = fixed standard capacitor
At balance, 𝐼1 = 𝐼3 and 𝐼2 = 𝐼c + 𝐼4
Techno Main Salt Lake
Department of Electronics & Instrumentation Engineering
Ic = I₁jωCR₃……. (1)
Writing the other balance equations
VCD = V CE + V ED
Or
Then putting the value of Ic from equation (1) in equation (3), we have
An examination of balance equations reveals that to obtain easy convergence of balance, alternate
adjustments of r₁ and r should be done as they appear in only one of the two balance equations.
APPARATUS:
Observation table:
𝑋L = 2𝜋𝑓𝐿
Experiment No.5
Theory: This is the simplest method of finding out the value of an unknown capacitor in terms of a
known standard capacitor. The unknown capacitor C 1 in the branch AB is compared with the standard
known capacitor C2 on arm AC. The bridge can be balanced by varying either of the non-inductive
resistors R3 or R4. Under balanced condition, since no current flows through the detector, nodes B and
C are at the same potential, i.e V1 = V2 and V3 = V4.
At balance, V1 = V2
The advantage of De Sauty’s Bridge is its simplicity. However, this advantage may be nullified by
impurities creeping in the measurement if the capacitors are not free from dielectric losses. This
method is thus best suited for use under balanced condition.
APPARATUS:
PROCEDURE: -
The connections are made as shown in the figure. The resistance R3 and a capacitor C1 are in
seriesas current I1 is flowing through both the branches of the bridge and resistance R 4 and
another capacitor C2 are in series as current I2 flows through the other two branches. A resistance
is unplugged in R3 and the resistance R4 is adjusted till the sound in the headphone gets reduced
to zero level. The value of R4 is measured with a multi-meter and is noted down. Measurement of
the resistances should be done in open circuit condition. The above process is repeated for different
values of R3 and the values are noted in the table. When the hum in the headphone is at zero level,
then the time constants of the upper and the lower branches of bridge are equal i.e.
Techno Main Salt Lake
Department of Electronics & Instrumentation Engineering
RESULT:
𝐶2(µF)
(µF)
Experiment No.6
TITLE: - STUDY OF STATIC CHARECTERISTICS (ACCURACY, PRECISION,
HYSTERISIS, REPEATABILITY, LINEARITY) OF A MEASUREING INSTRUMENT.
OBJECT: - To knows about the Static Characteristics.
THEORY: - The treatment of instrument & measurement system characteristics can be divided
into two distinct categories i) Static, & ii) Dynamic. Important factors of static characteristics are i)
Accuracy, ii) Precision, iii) Hysteresis, iv) Repeatability, v) Linearity.
ACCURACY: - It is the degree of closeness or conformity to the true value of the quantity under
measurement. It is the closeness with when the instrument reading approached the value of the
quantity being measured. Accuracy of a measurement means conformity to truth. Accuracy can
be specified in terms of inaccuracy or limit of error and can be expressed in the following ways.
i) Point accuracy, ii) Accuracy as percentage of scale range, iii) Accuracy as percentage of true
value.
APPARATUS:
Result:
PRECISION: - The term ‘precise’ means carely or sharply defined. Precision is a measure of
productivity of measurements that is given a fixed value of a quantity precision is the measure of
the degree of agreement within a group of measurements or instruments. Precision is used in
measurement to describe the consistency or the reproductivity of results. Accuracy can be
improved upon but not the precision of the instrument calibration.
Result:
Result:
REPEATABILITY: - It is a measure of closeness with which a given input may be measured over
and over again. It is defined as the variation of scale reading and is random in nature (more
repeatability indicates less hysteresis and hence the goods is better).
Result:
Graph: -
Experiment No.7
The measurement system goes through the transient state before it finally settles to its steady
state. Total time Response [C(t)] of a system is given by,
1) Fidelity
2) Speed of Response
3) Measuring Lag
4) Dynamic Error
1) FIDELITY: It is defined as the degree to which a measurement system indicates
changes in the measured quantity without any dynamic error.
APPARATUS:
FIDELITY:
CALCULATION:
SPEED OF RESPONSE:
Sl. no. True Value (A) Time Taken (t) Speed A/t
1 12
2 15
3 18
Graph: -
Experiment No.8
TITLE: - ACQUAINTANCE WITH BASIC STRUCTURE OF DMM AND
MEASUREMENT OF DIFFERENT ELECTRICAL PARAMETERS.
OBJECTIVE: - To Measure the Electrical Parameter.
THEORY: In the early age Analog Multimeter was used for measurement. But now, due to the
more accuracy, DMM is being used for,
i) High Accuracy
ii) High input impedance
iii) Smaller in size
iv) Better resolution
v) Capability to eliminate observational, parallax & approximation errors.
vi) The measured output of DMM can be directly fed to a computer for further
analysis and use.
In maximum DMM the, A/D converter is used. Often the converter is of dual slop integrating type.
Techno Main Salt Lake
Department of Electronics & Instrumentation Engineering
DMM is made up of
1) A/D converter
2) Circuitry for counting
3) An attenuation circuit
APPLICATION:
1) To measure Voltage
2) To measure Current
3) To measure Resistance
4) To measure Frequency
5) To measure Temperature
6) To measure Capacitance
7) To test Diode
8) To test Transistor
PROCEDURE:
APPARATUS:
OBSERVATION:
Experiment No.9
THEORY:
Q-meter: The Q meter is an instrument designed to measure some of the electrical properties
of coils and capacitor. The operation of this useful laboratory instrument is based on the familiar
characteristics of a series-resonant circuit, namely, that the voltage across the coil or the capacitor
is equal to the applied voltage times the Q of the circuit. If a fixed voltage is applied to the circuit,
a voltmeter across the capacitor can be calibrated to read Q directly.
The voltage and current relationships of a series-resonant circuit are shown in Fig. 1
XC = XL
EC = IX C = IX L
E = IR
50 Hz to 50 KHz delivers currents to a low-value shunt resistance RSH . The value of this shunt is
very low, typically on the order of 0.02 Ω. Its introduce almost no resistance into the oscillatory
circuit and its therefore represents a voltage source of magnitude E with a very small (in most
cases negligible) internal resistance. The voltage E across the shunt, corresponding to E in Fig.2
, is measured with a thermocouple meter, marked “Multiply Q by ”. The voltage across the
variable capacitor, corresponding to E C in Fig. , is measured with an electronic voltmeter whose
scale is calibrated directly in Q values.
To make a measurement, the unknown coil is connected to the test terminals of the instrument,
and the circuit is turned to resonance either by setting the oscillator to a given frequency and
varying the terminal resonating capacitor or by presetting the capacitor to a desired value and
adjusting the frequency of the oscillator. The Q reading on the output meter must be multiplied by
the index setting of the “Multiply Q by”, meter to obtain the actual Q value.
The indicated Q (which is the resonant reading on the “circuit Q” meter) is called the circuit Q
because the losses of the resonating capacitor, voltmeter and insertion resistor are all included in
the measuring circuit. The effective Q of the measured coil will be somewhat greater than the
indicated Q. This difference can generally be neglected, except in certain cases where the
resistance of the coil is relatively small in comparison with the value of the insertion resistor.
The inductance of the coil can be calculated from the known values of frequency ( f ) and
resonating capacitance ( C ), since
1
X L = X C and L = henry
(2f ) 2 C
So frequency (f) =
Now to measure frequency spectrum of unknown mixed signal, Q-meter connect can be used to
find out the frequency contents.
In this trainer kits we have used Q-meter concept to find out 2 different signal from a complex
mixed signal. 12 types of complex signal can be generated in our trainer kit. This signal feed into
the I/P of matching module to match with the Q-meter network.
Select some value for “C” & adjust “L” box until we obtain max. reading in digital high sensitive
meter. Again try to adjust next max. reading for another component of this signal.
This signal can be checked in our supplied software in time domain as ‘DSO’ or at frequency
domain as ‘SPECTRUM ANALYZER / FFT’.
Techno Main Salt Lake
Department of Electronics & Instrumentation Engineering
APPARATUS:
PROCEDURE:
Some calculated value for frequency wise different inductance value for fixed capacitance. In
practical there may be some value difference due to source impedance.
Techno Main Salt Lake
Department of Electronics & Instrumentation Engineering
1209 1 17.3
1336 1 14.2
1477 1 11.6
GROUP “B”
697 1 52.1
770 1 42.7
852 1 39.9
941 1 28.6