Datacolor 650 Users Guide
Datacolor 650 Users Guide
Datacolor 600™
Datacolor 400™
User’s Guide
Datacolor 650/600/400 User’s Guide
Part No. 4230-0395M Rev 1, January, 2007
All efforts have been made to ensure the accuracy of the information presented in this format. However,
should any errors be detected, Datacolor appreciates your efforts to notify us of these oversights.
©2007 Datacolor. Datacolor, SPECTRUM and other Datacolor product trademarks are the property of
Datacolor. Microsoft and Windows are either registered trademarks of Microsoft Corporation in the United
States and/or other countries.
To obtain information on local agents, contact either of the offices listed below, or visit our website at
www.datacolor.com.
Support Questions?
If you need help with a Datacolor product, please contact one of our top-rated technical support teams
located around the world for your convenience. You can find contact information below for the Datacolor
office in your area.
Americas
+1.609.895.7465
+1.800.982.6496 (toll-free)
+1.609.895.7404 (fax)
[email protected]
Europe
+41.44.835.3740
+41.44.835.3749 (fax)
[email protected]
Asia Pacific
+852.2420.8606
+852.2420.8320 (fax)
[email protected]
Or Contact Your Local Representative
Datacolor has representatives in over 60 countries. For a complete list, visit
www.datacolor.com/locations.
Manufactured by Datacolor
288 Shengpu Road, Suzhou Industrial Park, Suzhou, China
+86 512 625 80338
Maintenance..................................................................................................... 39
About Instrument Maintenance ..........................................................................................39
Sphere Cleaning.........................................................................................................39
Operating the Minivac Portable Cleaning System ...............................................39
Tile Handling/Cleaning ...............................................................................................40
Handling Tiles .....................................................................................................40
Cleaning Tiles .....................................................................................................40
Storage................................................................................................................41
Cleaning the Black Trap ......................................................................................41
Cleaning the Spectralon® Plaque .......................................................................41
Appendix .......................................................................................................... 43
Optical Block Diagram .......................................................................................................43
Datacolor 650/600/400 Instrument Specifications .............................................................44
Features by Model .............................................................................................................45
Dimension Comparison - Horizontal vs. Vertical Mount .............................................46
Optional Accessories .........................................................................................................46
Miscellaneous Technical Information .................................................................................47
System Requirements for USB Connection................................................................47
RS-232C Connector Pin Assignments .......................................................................47
Compliance Statements ....................................................................................................47
FCC Compliance Statement .......................................................................................47
Canadian Compliance Statement ...............................................................................48
Index ................................................................................................................. 49
Datacolor 600
Below is a summary of the standard and optional features included with instruments.
Please refer to Appendix, Features by Model to determine the standard and optional
features available for a specific model.
The configuration of the adjustable UV filter is handled through the software. See also
Datacolor TOOLS user Guide for detailed instructions to configure the option and calibrate
the instrument.
Datacolor 650V
• The Datacolor 600V and 400V are typically used by the paper industry.
• The Datacolor 650V offers the option to make transmission measurements in a
vertical configuration.
Electrical/Environmental Requirements
Power 85 to 264 VAC, 47 to 63 Hz, 80 VA peak, 35 VA typical
Absolute Operating 5º to 40º C, 20-85% RH, non-condensing
Environment
Interfaces Serial: RS232 9600/115200 baud (shipped at 19200 baud)
USB: 1.1 or higher
CAUTION
There are no user-serviceable parts for this equipment.
Light Source
Do not stare directly into the open port located in the front door panel when the
measurement is in progress. Staring directly into the light source can result in eye
discomfort similar to that of staring at a camera flash.
Power
CAUTION
Disconnect power before servicing.
Feature Summary
These instruments employ state-of-the-art features including the spectrometer, integrating
sphere, light source, and optics. Below is a summary of those features:
FEATURE DESCRIPTION PURPOSE/BENEFIT
Integrating sphere Barium coated sphere. Industry standard.
Measurement D/8º geometry, Specular Provides for more uniform
Geometry Component Included (SCI) or measurement of samples with
Excluded (SCE). irregular surfaces.
SP2000 Spectral Proprietary dual-channel Dual channel design provides
Analyzer holographic grating. 256- continuous monitoring of sample
photodiode linear arrays used illumination and compensates for
for both the reference and changes. 256-photodiode array
sample channels. enhances the precision of the
measurement.
Light source Pulsed xenon flash lamp Filtered to provide D65 illumination
including UV component.
Effective 5nm or 10nm Based on reporting interval.
Bandwidth
Calibration Tiles
A black trap, white tile and green tile are provided with all
instruments:
• The black trap and white tile are used each time the
instrument is calibrated.
• The green tile is used to perform an optional diagnostic
test.
• A CD and diskette containing the calibration values for
the white tile are also provided.
Aperture Plates
Multiple aperture plates with openings of different sizes
are included as standard accessories with all models.
The identification is engraved on each plate.
NOTE
Some of these apertures are standard accessories, depending on instrument model.
NOTE
The area illuminated is always 4mm greater than the diameter measured to reduce
translucency errors. The exception to this is the XUSAV plate.
Optional Accessories
If you have purchased a Datacolor 650 or Datacolor 650V, you will receive the following
accessories to make transmission measurements:
• Transmission Sample Holder Base. Datacolor 650 users must install this
holder into the transmission chamber. For Datacolor 650V (vertical mount) units,
the sample holder base is installed at the factory.
• Transmission Sample Holders. Used to secure the transmission sample in the
chamber. One is used for liquids and the other for solid samples. Each has two
dowels on the bottom that fit into the holes on the transmission sample holder
base.
• Transmission Cell. Used as a container for liquid materials. It is placed into the
liquid sample holder.
• Medium Area View, Barium Coated Aperture Plate. The inside of this plate
has a barium coating around the opening. This aperture plate must be used for
all transmission measurements.
• Black Calibration Card. This is used instead of the black trap for the zero
calibration.
• Spectralon® (White) Plaque. This white standard must be used for
transmission calibration. In addition, it is placed at the port for most transmission
measurements. Consult the standard procedure (ASTM, ISO, etc.) you are
following to determine if this sample should be placed at the port.
Overview
You must install a power cable, and either a USB cable or a serial cable to connect the
instrument to a computer.
NOTE
Do not connect both cables to the unit. You have a choice of using either a serial cable
or a USB cable, and both are provided with the instrument.
The connections for the cables are found on the back of the instrument.
WARNING
Read the "Electrical and Environmental Requirements” section BEFORE connecting your
instrument.
Power Cable
A six-foot power cable is provided with the instrument.
NOTE
The power cord supplied with the unit must be used.
Power is supplied to the back of the unit via a 3-prong male connector.
Power cable
Plug the female end of the power cord into the power receptacle on the rear panel of the
instrument. Insert the plug into a standard AC outlet.
NOTE
Do not connect both cables to the unit.
1. Connect the 9-pin male connector on the serial cable to the RS-232C female
connector on the rear panel of the instrument.
2. Connect the female connector on the serial cable to a serial communications port
on the back of the computer.
3. If there is more than one serial port on the computer, make a note of the port
being used. You may need to enter this information into the program.
4. Tighten each connection securely to ensure proper signal.
See the Appendix for a description of the RS-232C connector pin assignments.
You can connect the instrument to the computer using either a serial cable or a USB
cable. Select one of them. This section provides instructions for using the USB cable.
NOTES
Do not connect both cables to the unit.
If you use the USB cable, you must install additional software to activate the USB port.
See also Driver Installation below for instructions to add this software.
Three steps must be completed to enable the USB port for use with the instrument:
Cable Installation
The USB cable provided with the system is shown below:
To Instrument
To Computer
One end of the cable connects to a USB port on the instrument. The other connects to a
USB port on the computer.
• Connect the USB cable to the instrument and to the USB port on the back of the
computer.
1. Place the USB Drivers and Documentation CD into the CD-Rom drive.
2. Connect the USB cable to both the instrument and the computer. The window
below will automatically display:
3. Click on the Next button at the bottom of the window. The wizard will step you
through the installation of the USB software files. For most installations, you can
accept the defaults selected by the Wizard.
CAUTION
Do not remove the USB Driver CD from the drive. The Wizard will close
and then immediately reopen and then you will have to repeat the procedure
a second time to successfully copy the file.
5. When you once again see the “Welcome” screen, click the Next button.
6. Repeat the steps mentioned above as you move through the Wizard again.
When you again the last wizard screen, click Finish. At this time, the Wizard
closes and the installation is complete.
NOTES
Some Datacolor programs automatically configure the USB port assignment. Others
require that you manually assign the port number. Refer to the program User’s Guide for
information to assign/edit the instrument port number.
If the computer has more than one USB port, the program will default to the first one
available. If you want to connect the instrument to a different USB port, you must assign
the port manually in the applications software.
1. Click the Start button at the bottom left of the screen. The Start menu displays.
2. Place the cursor on Settings so that the program submenu displays, and click
Control Panel.
3. When the Control Panel window displays, double-click on System.
6. Go to the Ports heading. Click on the plus (+) to display the port selections.
7. Click USB Serial Port. The USB Serial Port (COM3) Properties dialog box
displays.
9. Click the Advanced button. The Advanced Settings for COM3 dialog box
displays.
The field labeled COM Port Number displays the current assignment for the USB
port. To change the port setting, click the drop-down list arrow. The list of COM
ports is displayed.
10. Click on the port assignment to be used for the USB port.
11. When you have completed the changes, click OK.
Datacolor 600/650
The status panel is the same for the Datacolor 650 and 600. It is shown below:
OPTION DESCRIPTION
420 The 420nm UV cutoff filter is selected. All wavelengths below 420nm are
excluded from the light source in the instrument.
460 The 460nm UV cutoff filter is selected. All wavelengths below 460nm are
excluded from the light source in the instrument.
UV Cal The UV filter option has been properly/successfully calibrated.
Datacolor 400
Below is the status panel for a 400:
OPTION DESCRIPTION
Spec Inc. The specular port is closed, and all measurements include the specular
reflection from the sample.
Ready Instrument is ready for use.
Spec Exc The specular port is open and all measurements exclude the specular
reflection from the sample.
Powering Up
NOTE
The power cord supplied with the unit must be used.
1. Verify all cable connections. The AC power cord should be plugged into an
outlet. The communications cable (serial or USB) should be attached to the
instrument and host computer.
2. Turn on the computer.
NOTE
If the instrument power is not turned on before a Datacolor program is
launched, you may receive an error message.
4. Depending on the model you are using, the LED lights on the status panel may
include the status of the following mechanisms:
− Aperture plate/zoom lens position
− UV filter status
− Specular port status
When the Ready light remains on, your instrument is ready to use.
Overview
The instrument must be calibrated regularly to ensure that the measurements are accurate.
This section provides instructions for performing both reflectance and transmittance
calibrations.
NOTE
We recommend that you calibrate the instrument every 8 hours. Please refer to your
software documentation for specific calibration instructions.
NOTE
Depending on the program you are running, the screens shown below may not match
those in your program. See also the User’s Guide for the program you are running to
obtain detailed information on the instrument configuration options you will use to set up
the instrument.
1. You must access the instrument setup. This is typically done from an Instrument
Setup option on a menu. You will see a screen similar to the one below:
3. Two files should be displayed. If they are not, browse to the correct drive/folder
location.
4. You must load both files. Highlight both file names, and click Open. The files
will be transferred to the system. When the transfer is completed, the program
will prompt you to calibrate the instrument. A calibration screen similar to the one
below will appear:
5. Click Calibrate to start the instrument calibration or Cancel to close the window.
NOTE
Instrument calibration can also be accessed through a separate menu option
or function button in the program window. The User’s Guide for the program
you are running contains detailed instructions to access the calibration
options.
The black trap and white tile are used each time the instrument is calibrated. The green tile
is used to perform an optional diagnostic test.
Before you begin, verify that the aperture plate at the port matches the aperture selection in
the program you are running.
The Datacolor Users Guide for the program you are running (e.g., Datacolor Tools, etc.)
provides step-by-step instructions regarding the (instrument) software setup and calibration
procedure.
NOTES
If sequential total transmittance and reflectance measurements are going to be made using
the 650, the instrument should be calibrated for both transmittance and reflectance
conditions, and measurements are made using the barium-coated MAV aperture plate.
NOTES
This calibration procedure is critical to the accurate measurement, either reflectance or
transmittance, of fluorescent samples.
If you are sharing either reflectance or transmittance data for fluorescent samples, the
adjustable UV filter in each of the instruments must be calibrated regularly.
• Haze evaluations, in compliance with ASTM D1003 Standard Test Method for
Haze and Luminous Transmittance of Transparent Plastics, for materials that
exhibit slight light scattering.
• APHA evaluation, which measure the yellowness of liquids.
The transmission compartment of the Datacolor 650 is large enough to allow the colorist to
easily insert and remove samples. It has been fitted with a removable spill tray in the event
of minor spillage of liquid samples.
NOTES
The transmission option is available on both the Datacolor 650 and Datacolor 650V. See
also Datacolor 650 Vertical Mount Options in this guide for instructions to use the
transmission option in the vertical mount unit (Datacolor 650V).
To calibrate the instrument for transmittance and to make transmittance measurements, the
sample holder base and transmission sample holder base must be installed.
The transmission sample holder base is installed at the factory for Datacolor 650V units.
Button
Front Back
Depress these buttons, and slide the case toward the back of the instrument, to open the
transmission compartment. The case will lock in the Open position:
NOTE
The cuvette holder is designed to hold a cuvette 50mm X 10mm (Datacolor Part No. 2950-
0004).
Spill Tray
The base attaches to a metal bar at the bottom of the transmission compartment. It is
placed directly over the spill tray, and is secured through magnets.
Installed sample
holder base
Spill tray
NOTES
Exercise care when installing the base into the transmission compartment. The design is
very tight.
Install the base in the transmission compartment before attaching either of the sample
holders. No special tool is required for installation.
Two thumb screws provide the ability to adapt This holder is designed for use with a 50mm X
the holder to samples of varying thicknesses. 10mm cuvette.
This holder is used for transmission
measurements using the 650V.
Dowel pins
The base includes three positions to locate the sample between the sphere and the lens.
The location of the sample holder along the base depends on the type of material being
measured.
NOTES
Loosen the thumbscrews on the solid sample holder to aid in the installation of the sample
holder.
Make sure the sphere opening/lens opening is completely covered by the sample you are
going to measure.
Calibration Procedure
Before you begin, you should verify the current aperture selection. The setting must be
Medium Area View (MAV) and the barium-coated MAV plate should be placed at the
instrument port.
NOTE
Transmission calibration and measurements must be made using the barium-coated
Medium Area View (MAV) aperture plate.
NOTES
The white plaque remains at the port for “Total” Transmittance measurements, and is
replaced with the black trap for “Diffuse” Transmittance measurements. See also
Sample Presentation and Measurement, Transmittance Measurements.
Reflectance Measurements
1. Grasp the sample holder and pull forward.
2. Position the sample, then carefully bring arm back up to normal operating
position.
WARNING
Do not allow sample arm to spring back. For the Datacolor 600V (vertical
mount), the cylindrical, spring-loaded sample holder can be depressed to
position your sample. You can return the cylinder to its normal position with a
quarter turn of the cylinder.
1. Place the sample at the port. Grasp the black tab above the aperture plate.
3. The backside of this door reveals the area of the sample covering the port
opening. If necessary, adjust the placement of the sample to target the portion of
the sample to be measured.
Transmittance Measurements
Three techniques are used to measure the transmittance of samples: “regular”
transmittance, “total” transmittance and “diffuse” transmittance.
When working with samples that are completely transparent (such as dyes in solution or
transparent plastics) you are measuring “regular” transmittance. These samples need to
be positioned close to the lens opening at the back of the transmission compartment:
Lens Opening
Front Back
Front Back
“Total” transmittance measurements are made with the white Spectralon® plaque in place.
“Diffuse” transmittance measurements are made by replacing the Spectralon® plaque at
the port with a light trap.
Consult with your Applications Specialist regarding the measurement technique you should
use for specific evaluations.
NOTE
The remote measure button is not available on all models. For those models equipped with
it, it is active only when the software supports remote measurement.
CAUTION
There are no user-serviceable parts for this equipment.
Sphere Cleaning
The sphere should be examined visually for the presence of dust, sample particles, fibers,
and excessive yellowing due to environmental influences. This inspection should be
performed frequently, especially if you measure loose materials.
If you discover loose materials in the sphere, they should be removed. An optional
accessory, the Minivac portable cleaning system is a useful tool for cleaning the sphere.
WARNING
Do not touch the inside of the sphere. The coating is very fragile.
Tile Handling/Cleaning
Handling Tiles
• Handle calibration tiles with extreme care. Do not drop them, or scratch the
glazed surface.
• Always grasp the tile using its edges.
Cleaning Tiles
• The calibration tiles should be cleaned before each use. Wiping it clean with a
soft, lint-free cloth can do this.
WARNING
Do NOT use Freon to air brush this plaque.
Cleaning Instructions
If the tile is lightly soiled, it may be air brushed with a jet of clean dry air or nitrogen. For
heavier soil, the material can be cleaned using the following procedure:
1
1. Sand the surface under running water with a 220 – 240 grit waterproof emery
cloth, until the surface is totally hydrophobic (water beads and runs off
immediately).
2. Blow dry with clean air or nitrogen, or allow the material to air dry.
1
Low reflectance Spectralon (<10% reflectance) should be dry sanded.
Below is a comparison of the dimension specifications for the horizontal and vertical mount
instruments:
Datacolor 600™ Datacolor 600 V
Optional Accessories
Compatible Models Datacolor Part No. Accessory
Datacolor 600, 650 3775-0260 MAV Aperture Plate
Datacolor 400, 600, 650 3775-0259 XUSAV Aperture Plate
Datacolor 400, 600, 650 1035-0004 Textile Skein Holder
Datacolor 400 1035-0186 Glass Aperture Plate - LAV
Datacolor 400 1035-0189 Textile Sample Holder - SAV
Datacolor 400 1035-0190 Textile Sample Holder - USAV
Datacolor 400 1035-0185 Carpet Tube Holder - LAV
Datacolor 400 1035-0188 Textile Sample Holder - LAV
Datacolor 400 1035-0191 Compression Sample Holder
Datacolor 650 2950-0004 10mm Transmission Cell
Datacolor 650, 600, 400 5400-0036 Minivac Portable Cleaning System
CAUTION
(1) DO NOT WIRE to pins other than 2, 3, and 7.
(2) For all unused inputs, the default is active high.
Compliance Statements
FCC Compliance Statement
WARNING
Changes or modifications to this unit not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
This equipment has been tested and found to comply with the limits for a Class A digital
device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference when the equipment is operated in a
commercial environment. This equipment generates, uses and can radiate radio frequency
energy and, if not installed and used in accordance with this User’s manual, may cause
harmful interference to radio communications.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following
conditions: (1) this device may not cause harmful interference, and (2) this device must
accept any interference received, including interference that may cause undesired
operation.
Francais
El present appareil numerique n’emet pas de bruits radioelectriques depassant les limites
applicables aux appareils numeriques de la class A prescrites dans le Reglement sur le
brouillage radioelectrique edicte par le ministere des Communications du Canada.
2 G
20 Read Repeatability on White Tile Using Dual Green Tile, 4
Flash (CIELAB), 45
H
A
Horizontal Mount, 46
Absolute Operating Environment, 44
Accessories, 4
Adjustable UV Filter Option, 2
I
Aperture Illumination Source, 44
configuration, 44 Indicators, 17
detection, 45 Instrument
opening, 33 calibration, 21
plates, 4, 45 calibration data, 21
Maintenance, 39
B specifications, 44
type, 44
Black Trap, 4 Integrating sphere, 3
cleaning, 41 Interface, 44
Interinstrument Agreement, 45
Inter-instrument Agreement for Diffuse
C Transmittance, 45
Cable Inter-instrument Agreement for Regular
installation, 7 Transmittance, 45
power, 7 Inter-instrument Agreement for Transmission
USB, 9 Haze Measurements, 45
COM Port Assignment, 15
Compliance Statements, 47 L
Controls, 17
Lens, 45
Light source, 3
D
Dimension Comparison, 46 M
Dimensions, 44
Measurement Geometry, 3
Measuring Geometry, 44
E Minivac, 39
Effective Bandwidth, 3 Miscellaneous Technical Information, 47
F O
Feature Summary, 3 Optical Block Diagram, 43
Features by Model, 45 Optional Accessories, 46