EndoScan-V User Manual
EndoScan-V User Manual
1 AND HIGHER
SOFTWARE MANUAL
contents
4.3.1 Import from CSV........................................................22
4.3.2 Add Sample................................................................23
4.3.3 Modify/Create From....................................................24
4.3.4 Delete Sample.................................................. ..........25
4.3.5 Print Sample Database............................................... 25
1 Introduction.................................................................1 4.3.6 Close Sample Database..............................................25
1.1 About..........................................................................1 4.4 Instruments................................................................25
1.2 Technical Support........................................................1 4.4.1 How to Add an Instrument:......................................... 26
1.3 License......................................................... ..............1 4.4.2 Adding a nexgen-PTS™..............................................26
2 Getting Started.............................................................2 4.4.3 Adding a nexgen-MCS™............................................ 28
2.1 PC System Requirements.............................................2 4.4.4 Discover MCS™/PTS™ (100 series)................... .......30
2.2 System Settings...........................................................3 4.4.5 Discover nexgen Instruments Function........................31
2.3 EndoScan-V™ Services Set-Up....................................3 4.4.6 Adding/Modifying BioTek ELx808™............................ 32
2.3.1 Installing EndoScan-V™ Services 4.4.7 BioTek Self-Test Report..............................................33
on a local PC or Central Server Location:......................3
4.5 Global Settings........................................................... 33
2.4 Installation Of EndoScan-V™........................................5
4.5.1 General Tab................................................................33
2.4.1 Installation Method:......................................................5
4.5.2 Security Tab...............................................................35
2.5 Licensing EndoScan-V™..............................................6
4.5.3 Report Tab.................................................................36
3 Navigating the Software ..............................................7
4.5.4 BioTek Layout Settings Tab.........................................36
3.1 Assay Icons ................................................................7
4.5.5 Cartridge Settings Tab................................................39
3.2 Tools/Administration Icons............................................8
4.6 Digital Signature Configuration................................... 39
3.3 Layout.........................................................................8
4.7 Manage Locks............................................................41
3.4 Information Icons.........................................................8
5 Results.......................................................................42
3.5 Account Icon............................................................... 9
5.1 BioTek Results............................................................42
4 Settings ......................................................................9
5.2 Cartridge Results........................................................44
4.1 User Management........................................................9
5.3 Results Options....................................................... ..44
4.1.1 Default User.................................................................9
6 Reports & Applying Digital Signatures........................45
4.1.2 Default Privilege Groups...............................................9
6.1 BioTek Reports...........................................................45
4.1.3 Default Sample Group.................................................15
6.2 Cartridge Reports.................................................... ..47
4.1.4 Users.........................................................................15
7 Import..................................................................... ..51
4.1.5 Privilege Groups......................................................... 16
8 Export..................................................................... ..52
4.1.6 Sample Groups.......................................................... 17
9 Copy from/Create Template(BioTek Mode Only)............53
4.2 Accessory Database.................................................. 18
9.1 Create Template.........................................................53
4.2.1 Manage Accessory Type............................................ 18
9.2 Copy From.................................................................54
4.2.2 Import from CSV........................................................19
10 Setting Up and Running a BioTek Assay......................55
4.2.3 Add Accessory...........................................................20
10.1 Adding Samples and Standard
4.2.4 Modify/Create From....................................................20 to the Template Grid....................................................55
4.2.5 Delete Accessory....................................................... 21 10.2 View List.................................................................. ..56
4.2.6 Print Accessory Database.......................................... 21
10.3 Print Grid....................................................................56
4.2.7 Close Accessory Database.........................................21
10.4 Read Number and Read Time.......................... ...........57
4.3 Sample Database...................................................... 22
P IES V 6 S F T-0 2
10.5 Well/Grid.....................................................................57
10.6 Standards and Samples Lists.................................... ..57
10.7 Properties....................................................................58
10.8 Starting an Assay...................................................... ..62
11 Cartridge MENU (CARTRIDGE MODE ONLY)................63
12 Setting Up and Running a Cartridge Assay...................65
12.1 nexgen-MCS™ Cartridge Panel...................................65
12.2 nexgen-MCS™ Assay Setup.......................................65
12.3 nexgen-MCS™ Cartridge Panel Test Sequence............66
12.4 nexgen-MCS™ Assay Result Screen...........................68
12.5 nexgen-PTS™ Cartridge Panel.................................. ..69
12.6 nexgen-PTS™ Assay Setup Section............................69
12.7 nexgen-PTS™ Cartridge Panel Test Sequence........... ..69
12.8 nexgen-PTS™ Assay Result Section......................... ..69
12.9 nexgen-PTS™ Multiple Bays Display...........................71
12.10 PTS™ Assay Setup................................................. ..71
12.11 MCS100 Assay Set Up.......................................... ....72
12.12 Running PTS™ and MCS™ Cartridge Assays.............72
13 Audit Trail....................................................................74
13.1 Date Range Fields.......................................................74
13.2 Print Button.............................................................. ..74
13.3 Search Criteria Field....................................................75
13.4 Search Keyword Field.................................................75
13.5 Show PC Name/PC Address Checkbox....................... 75
13.6 Show Windows User Name Checkbox........................75
13.7 Display Time in UTC Checkbox...................................75
13.8 Audit Entry Table...................................................... ..76
13.9 Audit Trail Entries - Examples......................................76
14 Back Up and Restore Database....................................78
14.1 Connect to Database...................................................79
14.2 Backup Database........................................................79
14.3 Auto-Backup............................................................ ..80
14.4 Backup Now...............................................................81
14.5 Restore Database.......................................................81
P IES V 6 S F T- 02
1 Introduction 1.3 LICENSE
This section provides a brief introduction to the software, the license Warning: This computer program is protected by copyright law and
agreement, as well as information about obtaining technical support. international treaty. Unauthorized reproduction or distribution of this
software, or any portion of it, may result in severe civil or criminal
1.1 ABOUT penalties and will be prosecuted to the maximum extent under law.
EndoScan-V™ is an endotoxin detection and measuring software utilized to EndoScan-V™ is developed with the following third-party
generate and report quantitative LAL test data. Verified and validated to be software packages:
consistent with global regulatory agency requirements, the software
performs the requisite calculations and creates reports for product release • Bouncy Castle
or incoming testing. For added protection, it generates secure data files and • Entity Framework
audit trails on all actions involving files and raw test data. The program was
• Log4net
designed with flexible
configuration capabilities and performance options for gaining operational • Microsoft.Net
efficiencies. All aspects of the software form the pillars of data integrity • .Net Systems
compliance: fully searchable audit trail, flexible user control and manage- • NPGSQL
ment, and reliable data generation and backup. • SSH.NET
Instrumentation supported by EndoScan-V™ version 6.0 software are: Each of these licenses is listed within the software and can be
• BioTek ELx808IU Microplate Reader provided upon request.
• Endosafe®-MCS™ (Product Code MCS100) Warranty Disclaimer
• Endosafe®-PTS™ (Product Code PTS100) The SOFTWARE is supplied “AS IS”. Charles River
• Endosafe® nexgen-MCS™ (Product Code MCS150) disclaims (to the fullest extent permitted by applicable law) all other
• Endosafe® nexgen-PTS™ (Product Code PTS150) warranties and conditions, expressed or implied, including, without
limitation, warranties of merchantability, fitness for a particular purpose,
non-infringement of third-party intellectual property rights, title, and
those arising by statute or otherwise in law form a course of dealing
1.2 TECHNICAL SUPPORT or use of trade. The user must assume the entire risk of using the
EndoScan-V™ Software.
Charles River
Disclaimer of Damages
1023 Wappoo Rd
Suite 43B Charles River assumes no liability for damages, direct or
consequential, which may result from the use of EndoScan-V™
Charleston SC, 29407
Software even if Charles River has been advised of the possibility of
Phone: (843) 402-4900 such damages. Any liability of Charles River will be limited to refund of
the purchase price. Charles River will not be liable for special,
Email: [email protected]
occasional, non-pecuniary (moral), indirect or inconsequential damage
or loss of profit, business, revenue, goodwill, personnel costs or
anticipated (cost) savings. In purchasing the SOFTWARE, you
expressly release and discharge Charles River from any and all such
causes of action Acceptance of delivery of the SOFTWARE shall be
deemed agreement to these terms and conditions. No document
issued by you attempting to negate or otherwise modify the terms
hereof, including any purchase order or request for proposal, shall be
binding upon Charles River.
Please note – Server Core is not supported. Users may have an older version of EndoScan-V™ software
installed on the PC that has version 6.0 installed. This is
Minimum: acceptable, however they cannot run at the same time.
1.4 GHz 64-bit processor Only one instance of EndoScan-V™ can be run on a PC at a
time. You cannot have 2 different windows users trying to
Compatible with x64 instruction set
run EndoScan-V™ at the same time.
2 GB RAM
These settings are found in Control Panel, Power Option. After installing • The following screen will appear identifying the pre-required
on either of the stated Windows systems, make sure the power options software, with the status as pending, to the user.
are set as follows: • If any of the pre-requisites are already installed, EndoScan-V™
• Turn off monitor: Never will skip that installation and proceed to the other
pre-requisites that are not installed.
• Turn off hard disk: Never
• If the EndoScan-V™ Services installer finds all the
• System standby: Never pre-requisites already installed, you can then proceed to install
• Hibernate: Never the EndoScan-V™software.
Windows 7 and Windows 10:
2.3 ENDOSCAN-V™ SERVICES SETUP
Before beginning to install any of the required software, confirm
that you have logged into the operating system as a Windows
Administrator. EndoScan-V™ must be installed in a supported
language Operating System (English, German, French, Spanish, Italian,
and Portuguese).
The software pre-requisites included in the EndoScan-V™ Services
set up are:
• Net Redistributable Version Minimum of version 4.6.1 (If a higher
version than 4.6.1 is already available on the computer, this will
not appear in the list of required pre-requisites).
• Microbial Solutions Database (Custom PostgreSQL).
When the Install button is clicked, the first pre-requisite
• Microsoft Visual C++ 2017 Redistributable Package Minimum
Microsoft Visual C++ 2017 Redistributable Package will be installed
of version 14.14.26405 (If a version higher than 14.14.26405 is
already available on the computer, this will not appear in the list of and the installation status will be shown.
required pre-requisites).
2.3.1 Installing EndoScan-V™ Services on a local PC
or Central Server Location:
The following installation procedure will detail which Windows
operating system the screenshots shown are for. Some differences in
actual screens displayed on the PC during installation, compared to the
screenshots shown in this document, may be seen. This is normal and
should not be taken as a deviation.
Click the “Add” button and enter the Windows user name of a user who will
access Endoscan-V™. (This can be a group or an individual user).
Once the user name is typed in click “Check Names” and the user or
group name will be underlined.
Click “OK” and that user/group will be added to the EndoScan-V™
Users group.
Repeat the above steps for the local group named
“EndoscanAdmins”, adding the users who will require access to
perform database administrative tasks such as backup and restore.
Note: If installing EndoScan-V™ Services on a server, the
EndoScan-V™ application must be installed on the server as well.
Into the “User” box type an appropriate user name (if using the default
Software Administrator account, type “admin” (all lower case)). In the
“Password” box type the password for the entered user (if using the
default Software Administrator account, type “password” (again all
lower case)). Click “Login” (See section 4.1 for further information)
The Software Administrator can then proceed to create new users and
Privilege Groups so that the full functions of the software can be set up.
3.3 LAYOUT
The About screen allows to you view the System
The layout section provides the user with options as to what will be information. This can help Charles River staff when
displayed on the screen in either mode. By checking the boxes those troubleshooting something for the user.
items will display; by unchecking they will be removed from the main
view. Standard Sets and Sample Info display in the left-hand panel and Save EndoScan-V™ Logs – If the user needs to contact
the Accessory DB and Sample DB appear in the bottom window. The Charles River for system issues, Charles River may
ask you to select this button so that you can send your
area they occupy can be adjusted by moving the window edge. system logs for analysis.
BioTek Mode:
Cartridge Mode:
4 Settings
When a user with appropriate privileges selects the Global 4.1.1 Default User
Settings icon, a drop down menu appears with the following
options: The software will ship with one user
• Accessory Database
Login Name = “admin”
• Digital Signature Configuration
Full Name = “EndoScanV Installer”
• Global Settings
Password = “password”
• Instruments
• Sample Database The software does not allow any changes to the Full Name for the
default user “admin”. This user cannot be deleted from the software.
• User Management
• Manage Locks 4.1.2 Default Privilege Groups
This section of the software will walk through each of these The following are the default privilege groups:
functions in detail.
“Software Administrator”, “Administrator”, “Manager”, and “User”.
Allows the user to make post assay chang- Allows the user to select the Configure
M/A Post Assay Changes es to settings that can alter the reported Inline spiking checkbox (BioTek reader
M/A Inline Spiking
results. (BioTek Reader only). only). This checkbox is in Global Settings
-> BioTek tab.
Global Settings - BioTek BioTek Calculation Privileges in
Calculations Global Settings Allows the user to select the Show CV prefix
M/A Show CV Prefix (BioTek reader only). This checkbox is in
Allows the user to configure the regression Global Settings -> BioTek tab.
M/A Regression settings (BioTek reader only). This field is in
Global Settings -> BioTek tab. Spike Recovery for Allows the user to select the Spike Recovery
Unreacted Samples - for Unreacted Samples – Use Zero. (BioTek
Allows the user to configure the default M/A
reader only). This checkbox is in Global
M/A Default Collection Type collection type (BioTek reader only). This Use Zero Settings -> BioTek tab.
field is in Global Settings - BioTek tab.
Allows the user to configure the order Global Settings - Cartridge Cartridge Setting Privileges
when using Polynomial Regression (BioTek
M/A Polynomial Order Allows the user to Configure CV limit for
reader only). This field is in Global Settings
-> BioTek tab. M/A CV Limit - Cartridges Cartridges (Cartridge Reader only). This field
is in Global Settings -> Cartridge tab.
Allows the user to configure the number of
# of Points to Use for
A points to use for MaxV used in the PQ. This Allows the user to select Sample Value by
MaxV
field is in Global Settings -> BioTek tab. Display Sample Value Channel (Cartridge Reader Only). This
M/A
Allows the user to configure the Onset OD by Channel checkbox is in Global Settings ->
M/A Onset OD (BioTek reader only). This field is in Global Cartridge tab.
Settings -> BioTek tab.
Cartridge Menu Settings Privileges
Allows the user to configure the num- Cartridge Menu Settings
# of Decimals to and Cog Privileges
Display for ber of decimals reported for Endotoxin Allows the user the ability to download logs
M/A
Value. This field is in Global Settings M/A Download Logs from a cartridge instruments or Save Test
Endotoxin Value.
-> BioTek tab. Reports (Cog) or Save Audit Log (Cog).
Allows the user to configure the Sample CV Allows the user the ability to download
M/A CV Limit - Samples settings (BioTek reader only). This field is in M/A Download Results results from the nexgen cartridge
Global Settings -> BioTek tab. instrument (Cartridge reader only).
Allows the user to configure the Standards
M/A CV Limit - Standards CV settings (BioTek reader only). This field
is in Global Settings -> BioTek tab.
P IE S V 6S F T-0 2 ENDOSAFE ® ENDOSCAN-V™ VERSION 6.0.1 AND HIGHER SOFTWARE MANUAL
Default Default
Privilege Privilege Name Description Privilege Privilege Name Description
Group Group
General Assay Settings General Assay Settings
Global Settings - General Global Settings - General
Privileges in Global Settings Privileges in Global Settings
nexgen Instrument Allows the user the ability to update the Sample Database Sample Database Privileges
M/A
Software Update nexgen instrument software.
Allows the user the ability to reconnect a Add/Create From/ Allows the user to add and modify a
U/M/A Reconnect M/A
cartridge instrument (Cartridge reader only). Modify Sample sample in the Sample Database.
Allows the user the ability to shut down or Allows a user to delete samples from
U/M/A Shutdown M/A Delete Sample
Restart the connected cartridge instrument. the Sample Database.
Accessory Database Accessory Database Privileges M/A Import Allows a user to import .pltx files.
M/A Export Allows a user to export data.
Manage Accessory Allows the user to create new
M/A
Types Accessory Types. Instrument Instrument Privileges
Add/Create From/ Allows the user the ability to Add and Modify
M/A M/A Add/Modify Instrument Allows a user add and modify instruments.
Modify Accessory Accessories in the Accessory Database.
M/A Delete Instrument Allows a user to delete an instrument.
Allows the user to Delete Accessories from
M/A Delete Accessory
the Accessory Database. Assay Properties Assay Properties privileges (BioTek only)
Print Accessory Allows the user to print Accessory
M/A
Database Database List.
• The Sample Group “All Samples” will be associated with all users by default.
Button Functionality
Manage Accessory Allows you to add an Accessory Type for future use.
Types
Import from CSV Allows you to import Accessories from a .CSV file.
Add Add a new record to the Accessory Editor.
Modify Allows you to modify a record in the Accessory Editor.
Create From Copies a current Accessory(s) from the Database so
that you can use it to create a new Accessory.
Delete Deletes the Accessory(s) that is highlighted in the
Accessory Editor or in the Accessory Database.
Save Saves the current records in the Accessory Editor into
the Accessory Database.
Print Print a report of all the Accessories in the Accessory
database.
Close Closes the Accessory Database window. These accessories are those most frequently used in Bacterial Endotoxin
Testing. The Added By user and Added Date will reflect the installation date.
These accessories can be deleted if not used by the customer and new
ones can be added. The only exception is that Cartridges cannot be
deleted or recreated.
• To Add: Type a new accessory type name in the field and
click Add.
• To Delete: Highlight the required Accessory using the rows
(a small arrow on the left indicates which row is selected)
and press delete. A warning box will appear as follows:
To ensure consistency of naming conventions, the Manage Accessory Type • Click OK to confirm the deletion.
feature allows those users with the appropriate privileges to name all
accessories used in the facility. Other users just creating accessory entries
A user can import an Accessory file but the .csv file must abide by the data
rules outlined below.
If a spreadsheet in CSV format is already used to list details of accessories
used in testing which is external to the software, the rules below can be
followed to ensure this file can be imported as a CSV into EndoScan-V™
Version 6.
To successfully import to the Accessory Database within EndoScan-V™, the
CSV file needs to be in the CSV UTF-8 format to ensure all special characters
are imported. In addition, the separation character for a CSV needs to be When you import an Accessory file, EndoScan-V™ creates an Excel
a comma, not a semi-colon. (Note that CSV stands for Comma Separated spreadsheet status file that opens upon completion of the import.
Variable). The Status file will list any files that were not imported.
See example below:
If importing to the Accessory Database via CSV you cannot import from a file
that is open the software will state that 0 of 0 records were imported.
Comments 1 to 255 characters To add an accessory, click Add. If multiple accessories are to be creat-
ed a user can click on add multiple times to create multiple rows in the
Added By n/a Editor.
Added Date n/a
Modified By n/a
Modified Date n/a
When importing, the user will receive a message reporting the number of
To print a list of all accessories listed in the database, select Print. The user
will be directed to a Print Preview page where they can choose to print or
save as a .pdf file.
If a user has searched for a selection of accessories, then when the print
button is selected the Print Preview will display only those
selected accessories.
4.3 SAMPLE DATABASE
4.2.7 Close Accessory Database
The sample database can be used to add all samples routinely tested in
To close the Accessory Database select Close. If items are still in the the laboratory, allowing for easy addition to assays.
Accessory Editor, the following warning will appear:
The Sample Database window is divided into 2 sections. The top
section is the Editor and the bottom section is the Sample Database.
You can add and modify samples in the Editor. Both sections have the
titles (Sample Editor and Sample Database).
Comments 60 characters
Complete the Sample details in each row as follows:
Added By N/A
Added Date N/A Product Family: Maximum 40 characters. User can choose to use a
Product Family name for the sample; this would be a grouping of
Modified By N/A similar samples.
Modified Date N/A
Sample Name: Maximum120 characters. User should enter a
Sample Name for the sample (this is a unique identifier field and must
be completed).
When you import a Sample file, EndoScan-V™ creates an Excel spreadsheet
status file that opens upon completion of the import. The Status file will list any
files that were not imported.
See example below: When importing, the user will receive a message If using Charles River Cortex™ software you must restrict
reporting the number of successfully imported Samples. character length for sample name to a maximum of 100 characters.
To delete an existing sample, the user must highlight the row to be deleted
either in the editor or in the database. This is denoted by an arrow to the
left of the sample: Choose Cancel if you wish to return to the Sample Database to
complete edits. Or choose OK if you wish to exit. If you choose OK, any
items in the Editor will not be saved.
Sample Database Notes:
The database can be searched by utilizing the search fields at the top of
each column:
4.4 INSTRUMENTS
The Instruments function allows users to select which instruments they
would like to use. There are 2 modes in EndoScan-V™: Cartridge and
BioTek. When viewing the Results window, you will see assays depending
Select OK to delete the Sample. the mode you are in. EndoScan-V™ version 6.0 supports the following
instrument types:
4.3.5 Print Sample Database
• BioTek ELx808IU Microplate Reader
To print a list of all samples listed in the database, select print. The user will be
directed to a Print Preview page where they can choose to print or save as a • Endosafe®-MCS™ (Product Code MCS100)
.pdf File. • Endosafe®-PTS™ (Product Code PTS100)
If a user has searched for a selection of samples, then when the print button • Endosafe® nexgen-MCS™ (Product Code MCS150)
is selected the Print Preview will display only those selected accessories. • Endosafe® nexgen-PTS™ (Product Code PTS150)
4.3.6 Close Sample Database All users will have access to the Instruments menu in Settings. Only those
users with the privilege to Add/Modify or Delete will be able to carry out
To Close the Sample Database, select Close. If items are still in the Sample these functions. Those without the privilege will only be able to activate and
Editor, the following warning will appear: deactivate instruments to allow them to switch between different
instruments/modes.
Select Add, then nexgen-PTS™, and the following window will appear:
Once in the Instrument window, you need to select the nexgen-PTS™ that
you just added to be your Active instrument.
The software will then show the new nexgen-PTS™ in the Instruments main
window. To activate the nexgen-PTS™, check the
Active Instrument box as shown below.
EndoScan-V™ is now ready to run a cartridge using the nexgen-PTS™. The initial setup of a nexgen-MCS™ assumes you would like to run the
instrument in your first tab and that each brick of the MCS™ should be
Up to 5 PTS™ systems can appear in one tab at a time. assigned to panels 1-5 in order. You can change the tab number (to 3)
and the panel numbers.
If the nexgen-MCS™ is the first and only instrument you are connecting to,
you can use the defaults where the nexgen-MCS™ is in tab 1 and the bays
are displayed in panels 1-5 in that tab.
Select Add, then select “nexgen-MCS™”, and the following window will appear:
Select Save, and the instrument configuration will be saved into the
database. A notification message will confirm the Instrument was saved.
The software will then present the Instruments main window with the new
nexgen-MCS™ added to the table.
Example:
CRL – 17360016-320190506-001.SUF
When EndoScan-V™ is connected to up to 3 nexgen-MCS™ units, each
instrument can display in its own tab (see below); the tab will be named
with the serial number of the instrument. Multiple nexgen-MCS™ units can
also be displayed in one tab, however, the tab name will be that of the first
instrument connected. Panel numbers should be numbered sequentially,
1-5 for the first nexgen-MCS™ and 6-10 for the second nexgen-MCS™.
Select Add, then Discover MCS100/PTS100, and the following window will
appear.
When the Add button is selected, the software will connect to all COM ports
on the computer, then establish if any 100 series instruments are connected.
The software will determine the instrument serial number(s) and bay serial
number(s), determine if the unit attached is a PTS™ or MCS™, and will
update the information on the screen accordingly.
The top part of the PTS™ cartridge panel contains the following:
• Tab number and instrument type in the tab.
• PTS™ Icon: icon that portrays the PTS™.
Check the box marked Select, then finish adding the instrument by clicking • Tab number, Panel number and Brick serial number.
Add. Once the instrument is added, the Instruments window will display the • Status indicator: (green/red/blue circle).
new instrument. • Assay Name: the assay name is a concatenated combination of:
Prefix
Suffix
designated Sequence
designated
in Global PTS Date in number (3
in Global
settings - serial - the format - digits) which is -
settings (up
(up to 10 number YYYYMMDD reset every day
to 10
charac- at 00:00:01 )
characters)
ters)
Select Add, then Discover nexgen instruments, and the following window
will appear.
Select the bays for use, then fill in the tab and panel number. If you
would like all 5 bays displayed in a single tab, the tab number for each
bay would be 1 and the panel number would be 1 for bay 1, 2 for bay 2,
3 for bay 3, 4 for bay 4 and 5 for bay 5. Each bay can also be displayed
as 5 different tabs. Once all information is completed see below:
Enter the Base IP and a Range for the software to discover the instrument.
(Example: Base IP: 192.168.1.1 and Range 255, means that EndoScan-V™ will
search IP addresses that are in the IP range of 192.168.1.1 to 192.168.1.255)
For the example above, the Discover function will attempt to find MCS™
bays at 192.168.1.101 to 192.168.1.106.
In the Instrument window, activate the instrument you have configured
Select Add, and the software will complete a series of steps to connect to by checking the Activate checkbox.
the instrument.
Select Close, and the software will display the cartridge panels for the
activated instrument. You need to log out and log back in when you
activate an instrument.
4.4.6 Adding/Modifying BioTek ELx808™ If you are unsure of the BioTek COM port, you can use the COM Port field
pull-down menu and choose from the COM ports on your PC.
From the Instruments tab in Global Settings, select Add, then BioTek You can test the connection by clicking on the ‘Test Connection’ button.
ELx808, and the following window will appear:
Once the BioTek reader is added, you will get the following confirmation:
Select the COM port that the BioTek is connected to, then enter the BioTek
serial number. This information can be found under the Windows Device
Manager. It is recommended that one BioTek is added at a time. The serial
number is located on the right-hand side of the instrument.
If you are connecting the BioTek for the first time, please enter 1 in the Tab
Order field. Completing the description field is optional.
Up to 3 BioTek instruments can be active and run at the same time.
Each BioTek instrument connected must have its own tab number, i.e
two BioTek instruments cannot be added to the same tab number.
The software is now ready to run an assay.
The report is generated as .pdf file and stored in the directory specified.
C:/ProgramData/Charles River/EndoScan-V/Documents/Directory with Data
stamp/file name.pdf
Where the file name is BioTek ELx808 Self-Test Report-timestamp.pdf
The Self-Test runs diagnostics of the instrument and diagnoses any
failures. Refer to the BioTek Instrument Manual for details on the
System Self-Test.
4.5.2 Security Tab Password Pull 14 0-14 How many days 0 will turn off
Warning-Days down before a password security features
menu expiration warning
message appears.
Backup File path c:\ Any available Allows the user to choose
Destination ProgramData\ directory on the backup destination file
CharlesRiver\ PC path.
EndoScan-V\
Database\
Backup\
Allows the user to establish settings for the reports. Allows a user to establish certain parameters for a BioTek assay.
Field Label Field Type Default Valid Range Information BioTek Layout Settings Section:
Value
Report Logo File display CRL logo See The Report Logo function can accept Field Field Type Default Value Valid Range Information
Information .JPEG, .JPG, and .PNG images. The Label
column report header section for the logo is
213 by 55 pixels. Any logo with this Number of Pull down menu 2 1-96 Allows the user to
approximate ratio will load into the Replicates choose the number
software correctly and with no of replicates to
distortion. The default CRL logo is be displayed on
300 DPI. Any logo with this DPI or the assay. Can be
higher will render accurately. edited at the assay
level. See further
Logo Posi- Pull down Left Left, Center, Allows the user to choose the information below.
tion menu Right position of the logo on printed
reports. SPL color Color swatch - Blue Any color Allows the user
brings up color to choose the text
Lab Name Text input Blank Text up to Allows the user to choose the palette color for Samples
40 characters Lab Name to be displayed on that will appear in
the reports. the assay grid. See
below for further
For Invalid Checkbox Unchecked Unchecked/ Allows the user to choose if information.
Assay show checked the Sample Value should be
SPK color Color swatch - Black Any color Allows the user
N/A for displayed on a report if other to choose the text
brings up color
Sample parameters are invalid. color for Spikes
palette
Value (Positive Product
Controls) that will
Add Checkbox Unchecked Unchecked/ Allows the user to choose if a appear in the assay
Disclaimer to checked Disclaimer should be added at the grid. See below for
All Summary bottom of all summary reports. further information.
Reports
STD color Color swatch - Black Any color Allows the user
Show Checkbox Checked Unchecked/ Allows the user to choose if the list brings up color to choose the text
Accessories checked of Accessories used in an assay palette color for Standards
in Report should be printed on the report. that will appear in
the assay grid. See
Disclaimer text input N/A 0-300 Allows the user to type a below for further
characters Disclaimer to print on the summary information.
reports if the ‘Add Disclaimer to All
Summary Reports’ is checked.
Onset OD Option
The Onset OD option is the default optical density (OD)
reaction-time threshold.
CV Limit – Samples
EndoScan-V™ calculates the coefficient of variation (CV) of the well results for CV Limit – Cartridges
each sample and spike. The limit entered into this field, in percent, is the default The cartridge reader calculates the coefficient of variation (CV) of the
acceptable limit for sample and spike CV for the limit set. channel results for the sample and spike. The limit entered into this field,
in percent, is the acceptable limit for sample and spike CV. The default is
CV Limit – Standards set to 25% as per the product insert for cartridges.
EndoScan-V™ calculates the coefficient of variation (CV) of the well results for
Display Sample Value by Channel
each standard. The limit entered into this field, in percent, is the default acceptable
limit for each standard CV for the limit set. For cartridge reader assays, sample endotoxin results can be displayed
and reported for each sample channel or for the average reaction time of
Average Replicates both sample channels. Default is off.
Average Replicates determines how the data is used in regression and endotoxin 4.6 DIGITAL SIGNATURE CONFIGURATION
calculations. When this option is selected, the program averages the data from
all the replicates prior to performing the calculations. When this option is not This feature allows the user to configure the Digital Signature Roles.
selected the individual optical density (OD) data values are used in the regression These roles must be configured in order for the Digital Signature
function to work.
and endotoxin calculations.
To delete a digital signature meaning, highlight in the main text box and
click delete.
For further information on applying Digital Signature please refer to
Section 6 – Reports.
Lock The object that has been locked. It could be an Assay name,
template name, or Global Settings.
Lock Type There are 4 different lock types - Cartridge Assay, BioTek
Assay, Template (BioTek), or Settings.
PC Name The PC name of user that owns the lock.
Login Name The Login of the user that owns the lock.
Table Column
Definition Information
Label
Assay Name Assay Name Unique identifier
Created By Who the assay was
Details the user’s full name
created by
Test Date/Time Assay Date/Time
Collection Type Kinetic Turbidimetric, Kinetic
Collection Type
Chromogenic, End Point.
Instrument Serial Instrument Serial
Number Number
Signature Status Digital Signature
Signed or unsigned
Status
Signature Level If signed, at what
Blank, Analyst, Reviewer, or QC
level?
Auto Export Was the assay
Displays Yes or No.
exported?
A user can search the table by date, Assay Name, keyword or by hovering over
the column header they would like to search on. When the mouse hovers over
the column header, the search arrow appears to the right of the column header if
the field is searchable - see below:
A search column window appears. A user can select the values needed by clicking
the check boxes. The window also allows the user to type
in a specific value (Text Filter). To apply the search criteria select Apply.
Blank, Analyst,
Fields available in Results Window: Signature Level If signed, at what level?
Reviewer, or QC
Field Label Field Type Default Value Valid Range Definition Auto Export Was the assay exported? Yes or No
You will be required to type the reason for cancellation before clicking OK. If
you do not type comments and attempt to click OK, you will not be allowed
to proceed with the signature cancellation.
Once OK is selected the following screen will appear:
Click Sign Assay to apply the signature. Digital Signature roles and
meanings are set up under User Management and Digital Signature
Configuration.
Once the digital signature is applied this will be displayed at the bottom of the
report as shown below:
Analyst B
Analyst A Analyst D Analyst E
cannot Analyst
cannot can cancel can cancel Analyst F
cancel the C can
Analyst A cancel their the the can cancel
signature cancel the
only has signature signature signature the signature
even if signature if
signed unless they if they if they if they have
they have they have
the assay. have the have the have the the privilege
the the privilege
privilege privilege privilege assigned.
privilege assigned.
assigned. assigned. assigned.
assigned.
Analyst Analyst
Analyst Analyst D Analyst E
A cannot B cannot Analyst F
Assay is C cannot can cancel can cancel
cancel the cancel the can cancel
signed by cancel the the signa- the signa-
signatures signatures the signature
Analyst B signature if ture if they ture if they
even if they even if they if they have
and Analyst they have have the have the
have the have the the privilege
D only. the privilege privilege privilege
privilege privilege assigned.
assigned. assigned. assigned.
assigned. assigned.
*NOTE: This file will be locked and signatures cannot be cancelled if the Level 3
signatory does not have the cancel signature privilege. The Cancel Digital Signa-
ture Privileges is a suggested privilege in the Admin and Manage Privilege groups.
Charles River recommends that all L3 authorized signatories are provided with the
Cancel Digital Signature Privilege. When cancel signature is chosen, the following pop up window will appear:
When the user chooses to Remove signature while in the Report preview window,
they will first get warning:
7 Import
Once the user has selected the files to import, a status bar should appear
listing the files being imported. The user will be given a chance to cancel
the import.
If the user imports 1200 files and 2 are corrupt and do not import, the
user will be notified which ones did not import.
8 Export
Select the location of the files/folders to be imported. Once the user selects a
folder to import from, the software will check all levels of sub folders for all .pltx
files and automatically import them.
If the user selects a folder without any .pltx files, it will prompt the user to select
a folder that does.
An authorized user can export any record or records in the database to a
.CRL, .CSV or .XML file(s). The user can choose a record in the database,
the file format and then choose the file directory where the exported file will
be stored.
The user can choose to export one file or a group of files. They can select a
group of files by holding down the control and shift keys as they select each
file.
This icon in BioTek Mode will present the user with a drop down box as
follows:
If you choose to Export more than 1 assay, the assay records are concatenated into
one file. The user can choose what folder to place the file into.
The row can be highlighted and Delete can be selected to delete the
template. To modify, double click the template; this will then open the
template in the main interface. Edits can be made and Save selected again.
A user can then add samples and standards to the plate as required and set plate
properties for the template. Once completed, click the Save icon at the top of the
template:
• If the desired sample or standard replicate will not fit in the desired
pasted location, the software will notify the user with a message. Undo
• When adding standards to a plate, if the replicates will not fit in the
desired wells, the software will not allow the standards to be added in
that location. The software will notify the user with a message. Redo
• When adding samples to the template in outer wells, the samples and
spikes will not wrap to the other side of the plate. Print Grid
• When cutting samples/standards, the user will be allowed to cut and
paste to open wells, provided there are enough wells to accommodate
the number of replicates needed to paste. Save (for use
• When deleting standards from a template prior to an assay being run, when saving
the standards will be deleted from the plate and the standards list. The Templates)
software will notify the user with a message.
• When deleting samples from a plate prior to an assay being run, all
highlighted fields will be deleted, including the spike wells.
The software will notify the user with a message.
• Polynomial Order
Polynomial Order value is established by selecting the value from
the pull-down list. The possible values are 2, 3, and 4.
• Onset OD
The onset optical density (OD) defined for a assay. Enter the
desired onset OD.
• CV Limit, Standards
EndoScan-V™ calculates the coefficient of variation (CV) of the
well results for each standard and control. The limit entered into
this field, in percent, is the acceptable limit for standard and
control CV and will be included on reports.
• Enable Validation
Selecting this option directs the program to evaluate the well
criteria (found under the well/group tab) for the assay.
• Regression
• Average Replicates
The EndoScan-V™ software is capable of performing either Linear
or Polynomial Regression to determine the standard curve. Select When Average Replicates is selected, EndoScan-V™ will
the type of regression desired for the assay. If Polynomial average the reaction times of each well in a sample, spike,
Regression is selected, the Polynomial Order option will be enabled. control, or standard group before calculating the grou
endotoxin value. Otherwise, the group endotoxin value is the
For Linear Regression to be performed there must be at least three mean of the well endotoxin values.
data points defined. For Polynomial Regression to be performed
there must be one more data point than the defined Polynomial • Spike Recovery Range
Order. EndoScan-V™ calculates spike recovery as a percentage for
spiked sample groups.
• Sample Name
The Sample Name field identifies the sample name of the
sample or spike selected in the Group/Well list. This will be
pre-populated if added from the sample database.
• Sample Lot Number
A user can add the sample lot number.
• Sample ID
A user can enter a sample ID for the sample. This will be
pre-populated if added from the sample database.
• Dil./Conc.
Well/Group
Enter the Concentration or Dilution for the sample or spike. If
The Well/Group tab allows the user to set the properties for the well and the dosage units are milliliters, this value can be expressed
samples, spikes, controls, and standards (groups of wells) contained in as a Dilution, i.e., a ratio (1:10). If dosage units are other than
the current assay. milliliters, this value must be expressed as a Concentration, i.e.,
a decimal number.
Changing the Concentration/Dilution for a sample changes the
Concentration/Dilution displayed for the sample’s spike, and
vice-versa. This will be pre-populated if added from the sample
database.
• Spike/Standard Concentration
For a sample, enter the amount of endotoxin spike for the sample. The assay will begin collecting data.
For a spike or standard, enter the endotoxin concentration of the The assay can be stopped before it completes via time assigned or auto
spike or standard. This will be pre-populated if added from the stop. This can be done by clicking the Stop Collection icon:
sample database.
• Group Criteria
The Group Criteria are the equations to be evaluated as true or
false for the assay by EndoScan-V™. Criteria entered into the
Group Criteria field will be included on reports for the assay. Enter
all applicable criteria for the plate, connecting criteria
expressions with logical “ AND ” operators. If the instrument does not connect, click the Reconnect Instrument
button.If connection issues continue, refer to section 4.4 on Instruments
in this manual.
11 Cartridge Menu
(Cartridge Mode Only)
When in the Cartridge Reader Mode a user can select the Cartridge menu
under the cartridge icon:
The screen below shows when the instrument is Running and Collecting
Assay Data:
Sample Information can be filled by dragging and dropping a row from the
Sample Database.
The Cog menu allows you to run cartridge functions on an individual bay.
The functions here are:
• Download Logs
• Save Audit Log
The top part of the nexgen-PTS™ cartridge panel contains the following:
• Restart (see section 11.0 Cartridge Menu for more info)
• nexgen-PTS™ Icon - icon that portrays the nexgen-PTS™ • Shutdown (see section 11.0 Cartridge Menu for more info)
• nexgen-PTS™ serial number • Save Test Reports
• Status indicator (green/red circle) • Download Results (see section 11.0 Cartridge Menu for more info)
• Assay Name - the assay name is a concatenated combination of: • Reconnect (see section 11.0 Cartridge Menu for more info)
Sequence Suffix • Assay Status
Prefix
number (3 designated
designated in PTS Date in Download Logs
digits) which in Global
Global - serial - the format - -
is reset settings
settings (up to number YYYYMMDD This function takes all the logs off the nexgen instrument and saves them in
every day at (up to 10
10 chars) the default directory
00:00:01 ) chars)
The cartridge panel will display “Save Logs” once the download
is complete.
Before Assay Status, “Save Audit Log” is displayed over the current Assay Status.
After Assay Status, the assay status is now displayed on the cartridge panel.
2 Enter User
Assay inserts
Setup a new N/A
cartridge
Insert
Cartridge User
enters
cartridge
and
12.12 RUNNING PTS™ AND MCS™ CARTRIDGE ASSAYS sample
parame-
• When the cartridge instrument is ready, it will display Insert Cartridge (as ters
displayed in sections above). This indicates that the bay is at the correct
temperature.
• Input cartridge lot number, cartridge calibration code, product name, product lot
number, Conc./Dil (Concentration/Dilution), and the
appropriate units, and finally the Endotoxin Limit (if applicable). Drag and Drop
can be used from the Sample and Accessory Database.
EndoScan-V™ will send this information for each cartridge to the respective
PTS™ or MCS™ bay. Once all information has been entered, select the Send
Setup button.
EndoScan-V™ will display Add sample and click Start when the sample and
cartridge information has been saved. Dispense 25µL into each of the four wells
for each cartridge and select the Start Button.
Each panel has a status bar that displays the current activity. For a typical assay on
the PTS™ or MCS™, the assay will move through the following stages:
Please
wait
5 Add User
sample adds
and click sample N/A
8 Assay View Beep
Start and complete / Print
selects results
Start
Remove
cartridge Remove
if inserted cartridge
Return to
the
beginning
of this
table
This checkbox allows the user to add the PC Name and PC Address to the
audit table. The PC Address is the actual network address of the computer
that generated the audit trail entry. This checkbox is unchecked by default.
The Search Criteria field has the following entries: This checkbox allows the user to add the Windows User Name to the audit
table. This information can be useful when troubleshooting database
• Blank – No search criteria selected. connection issues. This checkbox is unchecked by default.
• Add/Modify/Delete/Archive Users - Will search all audit trail entries and
detail when a user has been added/modified, deleted, or archived. 13.7 DISPLAY TIME IN UTC CHECKBOX
• Add/Modify/Delete/Privilege Group - Will search all audit trail entries
and detail when a privilege group has been added, edited, or deleted.
• Add/Modify/Delete Sample - Will search all audit trail entries and detail
when a sample has been added/modified or deleted. This checkbox allows the user to change the Date/Time time zone from
• Add/Modify/Delete Sample Group - Will search all audit trail entries and local time to UTC time. (Universal Time Coordinated). This checkbox is
detail when a sample group has been added/modified or deleted. unchecked by default.
• Changes to Global Settings - Will search all audit trail entries and detail
when global settings have been changed.
• Changes to Global Settings Comments - Will search for all audit trail
entries and detail when comments have been listed when making a
change to a global setting.
• Post Assay Comments - Will search for all audit trail entries and detail when
comments have been listed when making post assay changes.
• Signature Applied - Will search all audit trail entries that show a digital
signature has been applied.
• Signature Cancelled - Will search all audit trail entries that detail a
cancelled signature.
If the connection is unsuccessful, the Database Address and Database Port will
14.1 CONNECT TO DATABASE not be stored, and the fields will be filled in with the last successful connection
The following fields are required to be filled in correctly in order to successfully Database Address and Database Port or blank if there has not been a successful
connect. connection yet.
Database Address is the IP address of the Database. If the database is on the 14.2 BACKUP DATABASE
local PC, the Database Address will be localhost.
Backups of all the data in the database should be regularly created. This feature
Database Port is used to communicate to the database. By default, it is set to can be set up as a recurring task by using the Auto-backup feature or can also be
the CRL standard of 18083. created in the moment by simply clicking on the Backup Now button.
When these fields are populated, click the Connect to Database button. The Authorized users are granted access to the Global Settings/Security tab.
user should then receive a message saying ‘Successfully connected to
Database’, as below.
Then, select the Security tab on the left-hand side, where the Server Backup
Details section is found.
The Prefix field can be used to create an identifier to change the default,
which is discussed in the Naming Convention section below.
The time of the Next Backup can be set by clicking the calendar icon in the
drop down, as below. The date can be selected from the calendar, and once
the date is selected, the time of day can be manually entered.
The backup process will be run using the Computer Account of the system where
EndoScan-V™ Services is installed. If a network share is being used to save the
backup data this Computer Account must be given access to that network share.
If a database error occurs with EndoScan-V™ the software will create offline files
located in “C:\ProgramData\Charles River\EndoScan-V\Offline\Audit Trail and C:\
ProgramData\Charles River\EndoScan-V\Offline\Plate”. Once a database connection
is reestablished the files will then be imported into the database with the correct
timestamps and actions or assays that were performed during the disconnected
period. These files are locked down to the SYSTEM user of the PC and will not be
able to be deleted before or after they are successfully imported. Finally, Backup Destination can be established by using the Browse button
and navigating to a folder. We recommend using a network location and a
14.3 AUTO-BACKUP secure folder where the data is protected.
To enable automatic backups, select Yes from the Backups Enabled drop down
menu.
From the next screen, select Browse and the software will open a window
where the user can navigate to the folder where backups are stored.
Select the .tar (backup) file to restore, and click Open.
A message will appear in the lower right-hand corner of the screen which states After Restore is clicked, the software will create a backup of the database
Starting Database Backup. Once completed, a message will appear that states and then populate it with the file that was selected for restore.
Backup Complete.
Charles River recommends performing a Backup of the current database before Click Yes to restore the database, the following confirmation screen will
performing the Restore function. Once a backup is done, proceed with the display.
Restore as below.
The Restore function is found from the login screen of the software.
Click Settings, then select Connect/Restore Database.
Any Backup or Restore functions will be noted in the audit trail, as below.