SOP - Pipettes
SOP - Pipettes
rating
EN) Procedure
Operating for Pipettes
Procedure for Pipettes
Eppendorf SOP
Standard Operating Procedure for Pipettes
Copyright© 2014 Eppendorf AG, Hamburg. All rights reserved, including graphics and
images. No part of this publication may be reproduced without the prior permission of the
copyright owner.
Eppendorf® and the Eppendorf logo are registered trademarks of Eppendorf AG,
Hamburg, Germany.
Eppendorf ® and the Eppendorf logo are registered trademarks of Eppendorf AG,
Hamburg, Germany.
Biomaster®, Combitips®, Combitips advanced®, Eppendorf Reference®, Eppendorf
Research®, Eppendorf Xplorer®, Multipette®, PICASO®, Varipette® and Varispenser® are
registered trademarks of Eppendorf AG, Hamburg, Germany.
Registered trademarks and protected trademarks are not marked in all cases with ® or ™
in this manual.
Table of contents
2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.1 Scope of testing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.1.1 Adjusting adjustable volume pipettes . . . . . . . . . . . . . . . . . . . . . . . 8
2.1.2 Multi-channel pipettes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.1.3 Multipette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.1.4 Bottle-top dispenser and Top Buret . . . . . . . . . . . . . . . . . . . . . . . . 8
2.2 Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2.1 Work method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2.2 Removing the test volume from the reservoir . . . . . . . . . . . . . . . . . 9
2.2.3 Dispensing the test volume into the weighing vessel . . . . . . . . . . . 9
3 Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.1 Calculating the systematic error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.2 Calculating the random error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5 Leakage check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.1 Pipette leakage check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Table of contents
4 Eppendorf SOP
English (EN)
6 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
6.1 Potential error causes and corrections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
7 Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
7.1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
7.2 Adjusting adjustable-volume pipettes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
7.2.1 Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
7.2.2 Biomaster. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
7.2.3 Reference variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
7.2.4 Reference 2 adjustable-volume . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
7.2.5 Research variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
7.2.6 Research plusadjustable volume - adjustment to
environmental parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
7.2.7 Research plus adjustable volume - change to the
factory adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
7.2.8 Research Multi-channel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
7.2.9 Research plus Multi-channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
7.2.10 Varispenser plus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
7.2.11 Xplorer and Xplorer plus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
7.3 Adjusting fixed-volume pipettes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
7.3.1 Reference fixed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
7.3.2 Reference2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
7.3.3 Research fixed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
7.3.4 Research plus fixed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
7.4 Physical influences of liquids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
9 Technical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
9.1 Fixed-volume pipettes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
9.1.1 Reference fixed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
9.1.2 Reference 2 fixed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
9.1.3 Research fixed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
9.1.4 Research plus fixed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
9.2 Adjusting adjustable volume pipettes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
9.2.1 Reference variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
9.2.2 Reference 2 adjustable-volume . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
9.2.3 Research adjustable-volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
9.2.4 Research pro . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
9.2.5 Research plus adjustable-volume . . . . . . . . . . . . . . . . . . . . . . . . . 42
9.2.6 Biomaster. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
9.2.7 Varipette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
9.2.8 Xplorer and Xplorer plus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Table of contents
Eppendorf SOP 5
English (EN)
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Test and pipette conditions
6 Eppendorf SOP
English (EN)
1.1 Balances
1.1.1 Balance type
Use semi-microbalances and microbalances to calibrate pipettes.
Some companies offer balances that are specially designed to meet the requirements of
pipette calibration, e.g. Sartorius and Mettler.
1.1.2 Accuracy
When selecting the balances, ensure that they are suitable for the accuracy of the pipette.
This helps to ensure that deviations within a measuring series are recorded exactly for the
assessment of the systematic and random measurement deviations in accordance with EN
ISO 8655.
For a pipette volume of less than or equal to 10 μL, balances with a 6-digit display must be
used. With larger volumes, balances with a 5-digit display are sufficient.
* For practical reasons, the nominal volume can be used for the selection of the balances.
Test and pipette conditions
Eppendorf SOP 7
English (EN)
2 Calibration
2.1 Scope of testing
Note concerning the nominal volume:
The nominal volume of an adjustable-volume pipette is the largest volume to be set by the
user and specified by the manufacturer.
When using the Combitip with the Multipette, the nominal volume is the largest possible
dispensing volume of the Combitip. This means that for the Multipette plus and Multipette
M4, the nominal volume is 1/5 of the filling volume of the Combitip advanced. For the
Multipette stream / Xstream, the nominal volume is the filling volume of the Combitip
advanced.
2.1.3 Multipette
With the
• Multipette,
• Multipetteplus,
• Multipette stream,
• MultipetteXstream,
• MultipetteM4
check the nominal volume with 10 measured values with the Eppendorf Combitip used.
2.2 Measurement
2.2.1 Work method
1. Place the selected pipette tip on the tip cone of the pipette or on the corresponding
Combitip in the Multipette.
2. Implement the following settings:
• Adjustable-volume piston-stroke pipettes: the smallest volume to be tested
• Multipettes: the nominal volume
• Bottle-top dispenser and Top Buret: the nominal volume
3. For all Multipettes with fully drawn up Combitip, always discard the first dispensing
step.
4. Fill test liquid up to a height of min. 3 mm in the weighing vessel.
5. For piston-stroke pipettes fill the pipette tip 5x with test liquid and empty it (pre-wet),
in order to create a moisture balance in the dead air volume.
6. Replace the single-use tip.
7. Pre-wet the tip 1x.
3 Evaluation
3.1 Calculating the systematic error
Mean value of the dispensed volume:
The specifications of the tested pipette can be found in the relevant operating instructions
or technical specifications (see Technical specifications on p. 34).
Sterilization and cleaning
Eppendorf SOP 11
English (EN)
4.1 Sterilization
Parameter for autoclaving
• 121 °C
• 20 minutes
• 1 bar overpressure
4.1.6 Varispenser
The Varispenser and Varispenser plus are only autoclavable when fully assembled.
With theVarispenser plus
1. Set the discharge valve toggle to the dispensing position (–>).
2. Release the volume setting knob, move to the middle position and leave released.
3. Place the dispenser on a cloth and autoclave, avoiding contact with hot surfaces.
4. Only use the dispenser again when it has cooled down to room temperature.
4.2 Cleaning
All areas of the underside of the pipette can be cleaned in a soap solution or
isopropanol 60 % provided that no differing instructions are given in the
operating manual.
1. Clean the parts in soap solution or isopropanol.
2. Rinse the parts in distilled water.
3. Allow the parts to dry completely and then assemble.
4. Lightly lubricate the piston of the pipette (Eppendorf special grease
Leakage check
Eppendorf SOP 13
English (EN)
5 Leakage check
5.1 Pipette leakage check
In order to carry out a leakage check on a pipette, follow the procedure outlined below:
1. For adjustable-volume pipettes: set the nominal volume.
2. For volumes < 20 μL pre-wet the tip several times.
3. Hold the pipette vertically with a full tip for approx. 15 seconds. Do not touch the
pipette tip.
4. Observe the meniscus of the liquid on the tip opening. If there is a leak in the pipette, a
drip will form on the tip opening.
Troubleshooting
14 Eppendorf SOP
English (EN)
6 Troubleshooting
6.1 Potential error causes and corrections
7 Adjustment
7.1 General information
Prior to delivery, all pipettes are adjusted with distilled or de-ionized, degassed water
under the conditions described in Chapters 1 to 3 and according to ISO 3696
(see Calibration on p. 8).
If you have doubts regarding the accuracy of the volume entered, please check the
following points first:
• Is there a leak in the device (see Leakage check on p. 13)?
Exception: Biomaster
• Does the temperature of the pipetted liquid correspond to:
– the temperature of the device?
– the ambient air temperature?
• Is the set volume correct?
• Has the liquid density and air pressure been taken into consideration?
• Does the density of the pipetted liquid deviate from the double-distilled, degassed
water?
• Was the work completed correctly, as described in the "Calibration" chapter (see
p. 8)and "Evaluation" chapter (see p. 10)?
• Were original Eppendorf pipette tips used?
Volume errors can also occur when pipetting liquids with high vapor pressure, where the
density or viscosity deviates considerably from the values of water.
Adjustments should only be carried out on the device after these conditions have been
taken into consideration.
7.2.1 Procedure
Requirements
Device, original Eppendorf tip, test liquid and ambient air must have the same
temperature (15 °C – 30 °C) at a constant temperature of ±0.5 °C during the test
(according to EN ISO 8655, Part 6).
7.2.2 Biomaster
Auxiliary equipment
• Supplied pipette wrench (order no. 4910 092.001)
Reference 2
Reference 2
J
AD
1. Insert the pin in the middle of the gray adjustment seal (ADJ).
2. Remove the adjustment seal.
3. Insert the adjustment tool.
4. Turn the adjustment tool until the adjustment display shows the desired value.
5. Read the set value distortion-free using the
aligning aid in the viewing window.
6. Carry out weighings to check the accuracy and precision.
1000 µL
ADJ
1
- 1
+
The random and systematic errors recorded on delivery can be found in the Eppendorf
Certificate. This certificate is included in delivery. The certificate is no longer valid if the
default settings have been changed.
If the dispensing accuracy is only be to changed temporarily, the correct method is to
change the user adjustment. Before changing the user or factory adjustment, observe the
general notes and the associated gravimetric tests.
For the factory setting, the volume display will be set to match the piston stroke and the
detected actual volume. If a gravimetric test indicates that an error needs to be corrected
and you will be required to change to factory adjustment, proceed as follows:
AD
J
Re
AD
J
Re
AD
J
Re
5. Turn the adjustment tool to set the volume display. Set the volume display (from the
volume setting used during the inspection to the calculated volume of the gravimetric
test).
The volume change applies to the entire volume range. For the volume change, you
should initially set the value for 10% of the nominal volume of the value calculated
during the gravimetric test. Then carry out gravimetric tests to check 50% and 100%
of the nominal volume with this setting. If necessary, change the selected setting again
to achieve the best possible correction for all volumes. Use the error limits in
Adjustment
Eppendorf SOP 21
English (EN)
accordance with ISO 8655-2 and the technical data provided by Eppendorf AG to
decide whether the data obtained meets your requirements.
6. Carry out a gravimetric check of the changes that have been completed.
7. Insert the red safety plug into the
adjustment opening.
The red safety plug on the pipette
indicates that the pipette has been
adjusted and calibrated by the user.
AD
J
Re
+.
This adjusts the piston stoke of the
Re
se
C
h
0
20 change in the process.
Adjustment
22 Eppendorf SOP
English (EN)
You can recognize a change to the factory adjustment that was made by the user of the
Research plus from the red safety plug behind the ejector. If the Research plus has been
adjusted and calibrated by Eppendorf AG, this is indicated by a gray safety plug.
Adjustment
24 Eppendorf SOP
English (EN)
multi-channel
Nominal +½ revolution +¼ revolution -¼ revolution -½ revolution
volume
Color code
10 μL 0.53 μL 0.27 μL -0.27 μL -0.53 μL
Medium gray
100 μL 5.4 μL 2.7 μL -2.7 μL -5.4 μL
Yellow
300 μL 10.7 μL 5.4 μL -5.4 μL -10.7 μL
Orange
The values mentioned are theoretical values and are for orientation purposes only. The
volume changes mentioned apply to each volume setting.
Adjustment
Eppendorf SOP 27
English (EN)
9. Document the changes made and the measurements conducted. Remove the mark on
the volume setting ring and on the ejector. The red safety plug on the pipette indicates
that the Research plus has been adjusted and calibrated by the user.
Adjustment
28 Eppendorf SOP
English (EN)
One revolution of the control button corresponds to the following (based on water):
7.3.2 Reference2
Adjustment is carried out as described for the Reference 2 adjustable-volume(see p. 18).
With reference to water, one revolution of the volume setting ring corresponds to:
With adjustable-volume pipettes, all other values are then readjusted, as the
adjustable-volume pipette then becomes a fix-volume pipette.
The adjustment process is comparable to the described procedure (see p. 16). The
difference is that the mean value of the weighings is converted to microliters according to
the formula:
9 Technical specifications
The following specifications are valid for the following conditions:
9.2.6 Biomaster
9.2.7 Varipette
9.3.4 Reference2
9.4 Multipette
The following specifications for the Multipette plus, Multipette stream / Multipette
Xstream, Multipette M4 apply to the following conditions:
• Using the Combitips advanced
• Liquid: Distilled or deionized water
• Reference temperature: 20 °C to 25 °C, ±0.5 °C
• Number of determinations: 10 according to EN ISO 8655, with the original Eppendorf
Combitip advanced
Multipette stream / Multipette Xstream:
• Volume test in "DIS" mode
• Set speed level: 7
Technical specifications
Eppendorf SOP 49
English (EN)
9.4.2 Multipette M4
For Multi-channel pipettes, the error values are two times the values specified
for single-channel pipettes.
Technical specifications
54 Eppendorf SOP
English (EN)
A J
Accuracy ................................................... 6 Justierung
Adjustable-volume pipettes Adjustable-volume pipettes...............15
Adjust ............................................... 16
Adjusting adjustable volume pipettes L
Technical specifications ................... 37 Leakage check.........................................13
Adjustment ............................................. 15
Fixed-volume pipettes ...................... 30
General information ......................... 15 M
Measurement
Work method.......................................9
B
Measuring station .....................................7
Balance
Multi-channel pipettes
Type .................................................... 6
Technical specifications ....................45
Balances ................................................... 6
Multipette
Accuracy ............................................. 6
Technical specifications ....................48
minimum requirements ...................... 6
P
C
Physical influences of liquids..................32
Calibration software ................................. 7
Pipette conditions .....................................6
Cleaning ................................................. 12
Climatic conditions................................... 7
R
Random error ..........................................10
D
Dispensing the
test volume ..................................... 9, 9 S
Scope of testing
Adjusting adjustable volume pipettes.8
E
Bottle-top dispenser, Top Buret ..........8
Error causes ........................................... 14 Multi-channel pipettes ........................8
Error correction ...................................... 14 Multipette............................................8
Error limits ............................................. 53 Sterilization .............................................11
Evaporation .............................................. 7 Systematic error......................................10
F T
Factor Z Technical specifications
Table ................................................. 33 Adjusting adjustable volume pipettes37
Fixed-volume pipettes Conditions .........................................34
Adjust ............................................... 30 Fixed-volume pipettes.......................34
Index
58 Eppendorf SOP
English (EN)
V
Varispenser
Technical specifications ................... 52
Volume errors......................................... 15
W
Work method ............................................ 9
Evaluate your manual