TH2817B Operation Manual
1.1 Foreword
TH2817B is a precision LCR meter with high accuracy, good stability, and wide
measurement range. Controlled by a 16 bits MPU, TH2817B can be used for evaluating LCR
components, materials and semiconductor devices over a wide range of frequencies (50 Hz
to 100 kHz) and test signal levels (0.1V, 0.3V, 1V). With its powerful function, excellent
performance, perspicuous LCD display and easy menu operation, TH2817B is suitable for
high speed measurement need on product line and high accuracy and stability
measurement need in laboratory. By using the Handler interface, IEEE488 interface
(optional), and RS232C interface, TH2817B can easily be used for automatic test system
and computer remote control.
The instrument provides variable test conditions, typical conditions are as follows:
z Test signal frequency
10 typical frequency points available from 50 Hz to 100 kHz: 50Hz, 60Hz, 100Hz,
120Hz, 1kHz, 10kHz, 20kHz, 40kHz, 50kHz, 100kHz.
z Test signal level
Programmable signal level from 0.01 V to 2.00 V in 0.01 V resolution.
z Measurement speed
Fast, Medium and Slow measurement speed can be selected. Using the averaging rate
(from 1 to 255 programmable) function, you can get more stable measurement
results.
z Constant selectable source resistance
30 Ω or 100Ω
z Correction function
The OPEN, SHORT correction for correcting the stray admittance, the residual
impedance, and the other errors can be performed. The correction function has two
kinds of correction methods. In one method the open and short correction can be
performed at all of the frequency points, and in the other method the open, short
correction can be performed at the frequency points you specify.
1
TH2817B Operation Manual
TH2817B provides several communication interfaces which make it easy to output the
measurement results to other equipments (for example computers), or build an automatic
test system.
z RS232C interface: RS232C interface makes it easy to communicate with other
quipments. You can set the measurement functions and input the parameters through
RS232C interface instead of keyboard on front panel.
z IEEE488 interface (optional): IEEE488 interface helps instrument to build an
automatic test system with computer and other measurement instruments.
Command of RS232C and IEEE488 interfaces are programmed in the format of
common SCPI.
z HANDLER interface: By using the handler interface, the instrument can easily be
combined with a component handler, and a system controller to fully automate
component testing, sorting, and quality control data processing.
TH2817B has the file function to store the set parameters, limit parameters and list sweep
parameters as a file in the internal nonvolatile memory, so that the setups can be easily
reloaded, and the defult file will be automatically reloaded when TH2817B is turned on.
2
TH2817B Operation Manual
1.2.4 Notices
z Please do not operate the instrument in the place that is vibrative, dusty, under direct
sunlight, or where there is corrosive air.
z Although the instrument has been specially designed for reducing the noise cased by
ac power, a place with low noise is still recommended. If this cannot be arranged,
please make sure to use power filter for the instrument.
z Please store the instrument in the place where temperature is between 5℃ and 40℃,
humidity is less then 85% RH. If the instrument will not be put in use for a time,
please have it properly packed with its original box or a similar box for storing.
z Don’t frequently turn on and off the instrument, doing so will lead to the loss of the
calibrated data and the data saved by users.
The following general safety precautions must be observed during all phases of operation,
service, and repair of this instrument.
3
TH2817B Operation Manual
This product has been designed and tested to the requirements of the Electromagnetic
Compatibility (EMC) Directive. To use a properly shielded cable or shielded coaxial cable to
connect each of the ports to their respective controllers, peripherals, equipments or
devices may ensure to meet the requirements.
ABS absolute
ADDR address
ALCZ auto LCZ
ASCII American Standard Code for Information Interchange
AUX auxailiary
BIASM bias mode
BINn bin number
COMP compare
Cs equivalent serial capacitance
Cp equivalent parallel capacitance
D dissipation factor
DCL device clear
DCHR discharge
DCR DC resistance
DELIM delimiter
DIR direct
DUT the device under test
EXT external
EOM end of measurement
EOI end or identify
ESR equivalent serial resistance
4
TH2817B Operation Manual
FREQ frequency
GTL go to local
GET group execution trigger
GPIB general-purpose interface Bus
HDL Handler interface
IDX index
IFC interface clear
INT internal
KBEEP kep beep
KLOCK key lock
L2A primary inductance by 2-wire measurement
L2B secondary inductance by 2-wire measurement
LCD liquid crystal display
LCR inductance, capacitance, resistance
LCZ inductance, capacitance, impedance
Lk leakage inductance
Lp equivalent parallel inductance
Ls equivalent serial inductance
M mutual inductance
MAN manual
N turns ratio
NOM nominal value
PARA A parameter A
PARA B parameter B
PER percentage
PHI primary high
PLO primary low
Q quality factor
R resistance, real part of impedance
R2 DC resistance by 2-wire measurement
RMT remote control
Rs equivalent serial resistance
RS recommandation standard
Rp equivalent parallel resistance
SCPI standard command of programmable instruments
SDC selected device clear
SEC secondary parameter
SMD surface mount devices
5
TH2817B Operation Manual
6
TH2817B Operation Manual
Ls=Lp/(1+D2)
D=2πFLp/Rp=1/Q
Rs=RpD2/(1+D2)
L
Lp=(1+D2)Ls
D=Rs/2πFLs=1/Q
Rp=(1+D2)Rs/D2
Cs=(1+D2)Cp
D=1/2πFCpRp=1/Q
Rs=RpD2/(1+D2)
C
Cp=Cs/(1+D2)
D=2πFCsRs=1/Q
Rp=Rs(1+D2)/D2
7
TH2817B Operation Manual
The real-world capacitor, resistor and inductor are not ideal, but component of complex
impedance in the form of serial connection or parallel connection. The instrument
calculates necessary values according to equivalent circuit connected in series or parallel.
And different result will be obtained in different circuits, differing in different components.
Two equivalent modes can be conversed through some formulas, and Q and D don’t
change in different modes, as shown in the above figure.
Generally speaking, component with low impedance (such as large capacitor and small
inductor) is usually measured in series mode, and component with high impedance (such
as small capacitor and large inductor) is usually measured in parallel mode. And the
equivalent mode can be decided according to the application of component. If a capacitor
is used for power filtering, series mode is the best choice, and if it’s used for LC oscillat
circuit, parallel mode should be selected.
2.1.4 Range
Auto, Manual
9 ranges for LCR measurement
Internal: When the trigger mode is set to INTernal trigger mode, the instrument
continuously repeats measurements.
External: When the trigger mode is set to EXTernal, the instrument performs a single
measurement every time a low-to-high transition TTL level signal is applied to
the Handler interface on the rear panel.
BUS: When the trigger mode is set to BUS trigger mode, the instrument performs a
single measurement every time the *TRG common command is sent to
TH2817B via GPIB. The BUS trigger mode cannot be set on the front panel.
Send the TRIGger:SOURce BUS command via GPIB or RS232C to set the
trigger mode to the BUS trigger mode.
Manual: When the trigger mode is set to MANual trigger mode, the instrument performs
a single measurement every time when TRIGGER key on the front panel is
pressed.
Waiting time after measurement completes. If HANDLER interface clear has been set,
signal on HANDLER interface will be cancelled after delay.
8
TH2817B Operation Manual
Note: Make sure shielding terminals of Hcur and Lcur are reliably short with low impedance
when the test fixtures or cables used don’t come standard with the instrument.
At the frequency of being lower than 1kHz, measurement speed will slow down. And
transformer parameter measurement at the same conditions, especially resistance
measurement, will have lower speed.
9
TH2817B Operation Manual
2.3 Function
2.3.1 Correction
Open correction: to eliminate measurement errors due to parasitic stray admittance (C,
G) of the test fixture.
Short correction: to eliminate measurement errors due to parasitic residual impedances
(L, R) of the test fixture.
Cable length calibration:
0m: to measure with terminal fixture;
1m: to measure with 1m cable of TH26009, TH26011
CUST: to calibrate with cable of required length. And the shipped instrument hasn’t been
performed calibration.
2.3.2 Comparator
The instrument can perform multi-bin sorting and support signal output of HANDLER
interface.
BIN1 to BIN3: Both primary and secondary measurement parameters are within limits.
OUT BIN: Primary parameter is not within limits. Or primary parameter is within limits,
secondary parameter is not within limits, and AUX bin is off.
10
TH2817B Operation Manual
AUX BIN: Primary parameter is within limits, but whose secondary parameter
measurement result is not within limits.
Sorting mode:
ABS (absolute deviation): absolute deviation between measured result and nominal
value is compared with bin limit.
PER (percentage deviation): percentage deviation between measured result and
nominal value is compared with bin limit.
At AUTO mode, primary and secondary parameters are selected according to impedance
angle:
Impedance angle Primary and secondary parameters
o
θ>5 L-Q
θ<-5o C-D
others Z-θ
Keep the original parameter at similar status of short circuit or open circuit. If primary and
secondary parameters are modified, automatic LCZ function will be cancelled.
11
TH2817B Operation Manual
2.3.6 Memory
The instrument’s measurement parameter, such as frequency and level, measurement
setups, comparator’s limit data can be stored in nonvolatile storage in the form of files.
When they’re needed, they can be loaded directly from file menu to save time. And the
latest saved and loaded file will be automatically loaded when the instrument is turned on
next time.
2.3.7 Interface
z IEEE488 interface: also called GPIB interface and equal to IEC625 interface, but
having different linker. Communication command is in the format of SCPI and all
commands and data on BUS are sent in ASCII code. The interface supports automatic
FETCH, and has the functions of SH1, AH1, T5, L4, RL1, DC1, DT1, C0, and E1. It has
the compatibility of IEEE488.1 and IEEE488.2.
z RS232C interface: It uses standard minimum subclass of RS232C and doesn’t
support hardware connection. The interface has transfer baud rate: 4800bps, 9600bps,
11520bps, 12800bps, 14400bps, 19200bps, 28800bps, and 38400bps, signal logic
level of ±8V and maximum transfer distance of 15m. The communication command
is in the format of SCPI, and all commands and data on BUS are sent in ASCII code.
The interface supports automatic FETCH, and ending symbol supports LF, CR, CR+LF.
z HANDLER interface: It receives trigger signal and output sorting comparison result
signal and outputs synchronous signal of measurement (IDX, EOM). Low level is valid
for logic level. It has internal pull-up resistance and external power is defautly used.
Trigger edge has two options of rising or falling. And measurement output signal is
optional to be cleared after measurement delay.
12
TH2817B Operation Manual
短路 比较 系统 极限 偏置
副参数 电平 量程 ▲ 触发
▼
m
▼
ENT
测试端
LCUR LPOT HPOT HCUR
NG P1 P2 P3 AUX
2) LCD
240×64 dot-matrix Liquid Crystal Display (LCD) displays measurement results, test
conditions, etc.
3) Sorting indicators
It indicates sorting result: NG—not good
P1—pass BIN1
P2—pass BIN2
P3—pass BIN3
AUX—auxiliray bin
4) Power on/off
Power on/off switch. In the “I” position all operating voltages are applied to the instrument.
13
TH2817B Operation Manual
5) Softkeys
They have functions of three levels:
a) Functions by directly pressing softkeys
PARA A: to select primary parameters
PARA B: to select secondary parameters
FREQ: to set frequency
LEVEL: to select level
SPEED: to select measurement speed
RANGE: to set range HOLD/AUTO
SHIFT: to shift to second function
TRIGGER: to trigger single measurement
▲ ▼ is used to select select PAR, PAR B, FREQ, LEVEL, SPEED or RANGE,the object
is determined by the formal press. If pressPARA A firstly, then ▲ ▼ corresponds
with parameter selection;pressFREQ key,then ▲ ▼ corresponds with frequency
selection.
LOCAL: to enter local operation at remote control mode (RMT status)
c) Only the following menu keys are valid after entering menu:
【ESC】
【ENTER】
【▲】: to select option or add 1
【▼】: to select option or minus 1
【 】: to select option
▼ ▲
【 】: to select option
【m】、 【μ】、 【n】、 【M】、 【p】 、【k】 in limit set menu,when modifying limit data, press
unit rate directly to replace ENT key.
Unit rate:【p】means×10-12,【n】means×10-9,【μ】means×10-6,【m】means×
10-3,【k】means×103, 【M】means×106.
6) UNKNOWN terminals !
These are the UNKNOWN Terminals used to connect a four-terminal pair test fixture or test
14
TH2817B Operation Manual
Note: Make sure shielding terminals of Hcur and Lcur are reliably short with low impedance
when the test fixtures or cables used don’t come standard with the instrument.
7) Frame terminal !
This is the FRAME Terminal which is tied to the instrument's chassis and which can be used
for measurements that require guarding.
GPIB HANDLER
RS-232C
! WARNING
DO NOT CONNECT ANY DC/AC CURRENT/VOLTAGE SOURCE
TO TEST OR MONITOR TERMINAL.
NO USER SERVICEABLE PARTS CONTAINED, LINE VOLTAGE FUSE
DO NOT REMOVE COVERS.HAZARDOUS VOLTAGES PRESENT. ~ 220V/50Hz 250V T1AL
4) Name plate
Name plate is used to provide the information of date, model, lot number and
manufacturer etc.
5) Grounding
15
TH2817B Operation Manual
If the power cord doesn’t ground or isn’t reliably grounding, please connect grounding line!
7) Fuse holder
Fuse holder for TH2817B line fuse, 220Vac, 1A.
1) Alarm display
“ ”on: to turn on alarm
“ ” off: to turn off alarm
16
TH2817B Operation Manual
8) Range indicator
“AUTO”: range auto
“HOLD”: range hold
Digit 0-9 indicates current range number.
TH2817B has the function of measuring DC resistance. DC resistance doesn’t have
common range with AC parameters, so current display and operatable range can be
switched.
12) RMT
RMT means the instrument is at the status of remote communication when only LOCAL key
can be operated on panel.
17
TH2817B Operation Manual
TH2817B also has the function of measuring DC resistance and transformer parametersw.
Secondary parameters differ with parimary parameters:
Primary parameters Secondary parameters
CS, CP, RS, RP, Z θ° (angle), θr (radian), R, X, Q, D
LS, LP θ° (angle), θr (radian), R, X, Q, D, DCR
Press PARA B key, and the secondary parameter changes. Or pressPARA B, then▲、▼ to
switch parameter.
18
TH2817B Operation Manual
Press FREQ key, and the frequency changes among the foregoing frequencies. Or
pressPREQ, then▲、▼ to switch parameter.
And the current frequency is displayed in frequency area on LCD.
Press LEVEL key, and the level changes. Or pressLEVEL, then▲、▼ to switch level.
And the current voltage is displayed in voltage area on LCD.
Press SPEED key, and the speed changes. Or pressSPEED, then▲、▼ to switch speed.
And the current measurement speed is displayed in speeda area on LCD in the form of F, M
or S.
4.2.6 Trigger
When the trigger mode is set manual, the instrument performs one measurement by
pressing TRIGGER key once.
TH2817B totally has 9 ranges: 10Ω, 30Ω, 100Ω, 300Ω, 1kΩ, 3kΩ, 10kΩ, 30kΩ, 100k.
Please select range accoding to the impedance of the device under test.
Please refer to the following table for range and its corresponding measurement range.
When the instrument is working at automatic range, it can look for the most suitable range
to take measurement, therefore, it’s recommended to select range auto mode mostly, to
avoid incorrect measurement caused by wrong selection of range.
But in some special cases, range should be held, such as measurement of the same type in
batch and sorting test, in order to improve efficiency, speed and stability. Inductance’s bias
test should be performed at held range.
The way to hold range is to make instrument automatically measure, and hold range after
19
TH2817B Operation Manual
measuring out correct and stable value, or select range according to the impedance of the
device under test, referring to the next table.
LCR range:
Press RANGE key, and the range changes between AUTO and HOLD. Press UP or DOWN
key to manually select range. And the current range No. is displayed in range area on LCD.
TH2817B’s open correction function can eliminate the effect of test cable, test fixture and
stray admittance in parallel with the device under test, such as stray capacitance.
20
TH2817B Operation Manual
▼
)Note:Press【 】【 】 to enter short correction menu.
4. Press ESC key to cancel correction and exit to test status.
5. Press ENTER key to start correction.
TH2817B’s open correction can eliminate the effect of test cable, test fixture and stray
impedance in series with the device under test, such as lead resistance or lead inductance.
Display mode only works for primary parameter, not secondary parameter.
Shift of display mode is related to tolerance mode of primary parameters:
When tolerance mode is absolute value (ABS) → Display mode shifts between DIR and ⊿
When tolerance mode is percentage (PER) → Display mode shifts between DIR and %.
21
TH2817B Operation Manual
For example:
1. The current display mode is DIR, and “DIR” is displayed on LCD, as shown in Figure4-5.
4.2.11 Comparator
Press SHIFT + COMP to turn on or off comparator.
After comparator is turned off, comparator indicator on panel is off, alarm doesn’t beep,
comparison result doesn’t output through HANDLER interface, EOM of measurement’s
ending signal doesn’t output, but AD shift signal IDX and trigger input TRIG are still valid.
After comparator is turned on, “COMP” is displayed on LCD.
22
TH2817B Operation Manual
At measurement setup page, press 【 】or 【 】 key to select item, press【▲】or 【▼】
key to modify value, then press 【 ESC 】 to exit form measurement setup or data
modification. When data input is valid, press【ENTER】 key to enter data modification, and
when data input isn’t valid, press【ENTER】to exit, as【ESC】 key works.
1) TRIG
The instrument starts one measurement only after receiving one trigger signal.
There are four trigger modes: INT (internal, EXT (external), MAN (manual), BUS. The
instrument only receives trigger signal at current trigger mode.
Internal: When the trigger mode is set to INTernal trigger mode, the instrument
continuously repeats measurements.
External: When the trigger mode is set to EXTernal, the instrument performs a single
measurement every time a low-to-high transition TTL level signal is applied to
the Handler interface on the rear panel.
BUS: When the trigger mode is set to BUS trigger mode, the instrument performs a
single measurement every time the *TRG common command is sent to
TH2817B via GPIB. The BUS trigger mode cannot be set on the front panel.
Send the TRIGger:SOURce BUS command via GPIB or RS232C to set the
trigger mode to the BUS trigger mode.
Manual: When the trigger mode is set to MANual trigger mode, the instrument performs
a single measurement every time when TRIGGER key on the front panel is
pressed.
23
TH2817B Operation Manual
2) DELAY
Programmable delay from the trigger command to the start of the measurement, 0 to
60.000 s in 1 ms steps.
▼
a) Select “DELAY” through 【 】or【 】key, as shown in Figure4-9.
b) Press 【▲】or【▼】 to add or minus delay time.
b) Press 【ENTER】key to enter data modification, and the data which is going to be
modified twinkles, press 【▲】or【▼】key to change value. Then press 【ENTER】
key to ensure data input, and 【ESC】key to cancel.
c) Press 【ESC】key to exit.
3) FILT
Perform the following steps to set filtering depth:
▲
4) SRES
Perform the following steps to set source resistance:
▲
24
TH2817B Operation Manual
5) DCHR
Perform the following steps to turn on or off discharging test:
▼
a) Select “DCHR” through 【 】or【 】key, as shown in Figure4-12.
b) Press 【▲】or【▼】key to select “ON” or “OFF”.
c) Press 【ENTER】key or 【ESC】key to exit.
6) ALCZ
Perform the following steps to turn on or off auto LCZ:
▲
10) CABLE
Perform the following steps to set cable length:
▲
11) ZERO
Perform the following steps to turn on or off correction function:
▲
25
TH2817B Operation Manual
The result can also be output from alarm, referring to system setup.
Comparator’s comparison and sorting:
26
TH2817B Operation Manual
Measurement completes
Please first set limit list before using comparator function. If data in limit table aren’t set, it
will be 0.
27
TH2817B Operation Manual
▼
At limit setup page, press 【 】or 【 】 key to select item, press【▲】or 【▼】 key
to modify value. When data input is valid, press 【ENTER】 key to enter data modification,
and when data input isn’t valid, press【ENTER】to exit, as【ESC】 key works.
1) AUX
Perform the following steps to turn on or off auxiliary item.
▲
▼
a) Select “AUX” item through【 】or【 】key at limit setup page, as shown in
Figure4-21.
b) Press【▲】or【▼】key to select “ON” or “OFF”.
c) Press 【ENTER】or【ESC】key to exit.
2) MODE
Primary parameter’s high and low limits can be set in the form of absolute tolerance and
percentage tolerance.
a) Select “MODE” item through【 】or【 】key at limit setup page, as shown in
Figure4-18.
b) Press【▲】or【▼】key to select “ABS” or “PER”.
c) Press 【ENTER】or【ESC】key to exit.
3) NOM
Perform the following steps to set nominal value:
▲
a) Select “MODE” item through【 】or【 】key at limit setup page, as shown in
Figure4-19. The current nominal value is displayed at the second line with unit
corresponding with current primary parameter’s type and multiple at unit column.
b) Press 【ENTER】key to enter data modification status, and the data which is going to
be modified twinkles.
▲
e) If decimal point needs repositioning, press 【 】or【 】key to select the decimal
28
TH2817B Operation Manual
▼
f) If multiple needs modifying, press 【 】or【 】key to select the unit and multiple
to twinkle, then press【▲】or【▼】key to modify multiple.
g) After setting, press 【ENTER】key to ensure the modification, or【ESC】key to cancel
it. Or press 【m】、【μ】 、【p】、【k】 directly, select rate and confirm data
、【n】、【M】
modification.
h) At the status of non-data modification, press【ESC】key to exit.
At absolute tolerance mode, high and low limits of primary parameters are displayed in the
similar way of nominal value with unit and multiple; at percentage tolerance mode,
corresponding data are displayed in the form of percentage without unit and multiple. For
example, at percentage tolerance mode, low limit of BIN1 is -5.000, then it will be
displayed as -5%, as shown in Figure 4-20 and 4-21.
At limit setup page, select SEC L or SEC H through【 】or【 】key, then set low limit
and high limit of secondary parameter, as shown in Figure 4-22.
29
TH2817B Operation Manual
1) SAVE
This operation saves the current setups to appointed file No. Perform the following steps to
save file:
▲
a) Select “SAVE” at file operation page through【 】or【 】key, as shown in Figure4-25.
b) Press【▲】or【▼】key to select appointed file No. from 0-9. Displayed “Y” means the
file has existed, and it will be covered at the time of saving, and “N” means the file doesn’t
exist.
c) Press【ENTER】key to perform the saving operation, and “WAIT” displays.
d) Press【ESC】to exit.
2) LOAD
The operation loads setups of appointed file. Perform the following steps to load file.
▲
a) Select “LOAD” at file operation page through【 】or【 】key, as shown in Figure4-24.
b) Press【▲】or【▼】key to select loading file No. Displayed “Y” means the file has existed,
and it will be loaded, and “N” means the file doesn’t exist.
c) Press【ENTER】key to perform the loading operation, and it automatically exits back to
measurement page after loading completes.
d) Press【ESC】to exit.
30
TH2817B Operation Manual
3) ERASE
Perform the following steps to erase file:
▼
a) Select “ERASE” at file operation page through【 】or【 】key, as shown in
Figure4-25.
b) Press【▲】or【▼】key to select erasing file No. Displayed “Y” means the file has existed,
and “N” means the file doesn’t exist.
c) Press【ENTER】key to perform the erasing operation, and hint “SURE” to ensure the
operation, press【ENTER】key to erase the file, otherwise, press【ERC】key to give up the
operation.
d) Press【ESC】to exit.
31
TH2817B Operation Manual
1) BUS
There are 3 BUS types:
OFF: communication interface off;
RS232: RS232C communication interface is turned on, and GPIB communication function
is turned off;
GPIB: GPIB (IEEE488) communication interface is turned on, and RS232C
communication function is turned off.
▼
a) At system setup page, select “BUS” through 【 】or【 】key, as shown in Figure
4-30.
b) Press 【▲】or【▼】key to select “RS232C”, “GPIB” or “OFF”.
c) Press 【ENTER】key or【ESC】key to exit.
2) ADDR
Address can be changed from 0 to 31.
Perform the following steps to select:
▲
a) At system setup page, select “ADDR” through 【 】or【 ▼ 】key, as shown in Figure
4-17.
b) Press 【▲】or【▼】key to add or minus address.
c) Press 【ENTER】key to enter data modification status and modifying data twinkles,
▲
then press【 】or【 】key to select the data that is going to be modified, press【▲】
or【▼】key to modify value. Then press【ENTER】key to ensure the modification, or【ESC】
key to give up it.
d) Press 【ESC】key to exit.
3) BAUD
TH2817B’s RS232C supports 8 common baud rates: 4800bps, 9600bps, 11520bps,
12800bps, 14400bps, 19200bps, 28800bps, and 38400bps.
a) At system setup page, select “BAUD” through 【 】or【 】key, as shown in Figure
4-28.
b) Press 【▲】or【▼】key to select baud rate.
c) Press 【ENTER】key or【ESC】key to exit.
32
TH2817B Operation Manual
4) DELIM
TH2817B supports to send several common ending symbols:
LF: line change symbol (ASCII ode 100);
CR: enter symbol (ASCII code 13);
CR+LF: enter to change line
▼
a) At system setup page, select “DELIM” through 【 】or【 】key, as shown in Figure
4-29.
b) Press 【▲】or【▼】key to select ending symbol that needs setting.
c) Press 【ENTER】key or【ESC】key to exit.
5) FETCH
The measurement result can be automatically sent or queried when the instrument
communicates with PC through RS232C or GPIB.
a) At system setup page, select “FETCH” through 【 】or【 】key, as shown in Figure
4-30.
b) Press 【▲】or【▼】key to select fetch mode.
c) Press 【ENTER】key or【ESC】key to exit.
6) TRGEG
There are two kinds of trigger edges:
RISE: Measurement starts at trigger pulse’s rising edge;
FALL: Measurement starts at trigger pulse’s falling edge.
33
TH2817B Operation Manual
▼
a) At system setup page, select “TRIGER” through 【 】or【 】key, as shown in Figure
4-31.
b) Press 【▲】or【▼】key to select trigger edge mode.
c) Press 【ENTER】key or【ESC】key to exit.
7) HDL
Perform the following steps set signal hold:
▼
a) At system setup page, select “HDL” through 【 】or【 】key, as shown in Figure
4-32.
b) Press 【▲】or【▼】key to select HOLD or CLEAR.
c) Press 【ENTER】key or【ESC】key to exit.
8) ALARM
TH2817B provides alarm function for sorting result.
OFF: alarm off;
NOGO: alarm at fail (no good);
AUX: alarm at auxiliary bin;
PASS: alarm at any pass bin
a) At system setup page, select “ALARM” through 【 】or【 】key, as shown in Figure
4-33.
b) Press 【▲】or【▼】key to select alarm mode.
c) Press 【ENTER】key or【ESC】key to exit.
9) SOUND
34
TH2817B Operation Manual
▼
a) At system setup page, select “SOUND” through 【 】or【 】key, as shown in Figure
4-34.
b) Press 【▲】or【▼】key to select sound.
c) Press 【ENTER】key or【ESC】key to exit.
10) KBEEP
Perform the following steps to set key beep.
▲
a) At system setup page, select “KBEEP” through 【 】or【 】key, as shown in Figure
4-35.
b) Press 【▲】or【▼】key to turn on or off keylock.
c) Press 【ENTER】key or【ESC】key to exit.
11) KLOCK
Perform the following steps to set keylock.
▲
a) At system setup page, select “KLOCK” through 【 】or【 】key, as shown in Figure
4-36.
b) Press 【▲】or【▼】key to turn on or off keylock.
c) Press 【ENTER】key or【ESC】key to exit.
35
TH2817B Operation Manual
After keylock is set, function keys on panel can’t be available. Press SHIFT, then SYSTEM
to enter “KLOCK” operation, press 【▲】or【▼】to unlock.
12) SAVE
Perform the following steps to save setups:
▼
a) Select “SAVE” through 【 】or【 】key, as shown in Figure 4-37.
b) Press 【ENTER】to save setups.
36
TH2817B Operation Manual
The RS232C interface can be used to remotely control the TH2817B, and it also can be
used to control the DC current bias sources manufactured by our company such as
TH1773/TH1775 DC current bias source. All operations from the front panel can be
performed by a computer via the serial interface.
Signal Ground
GND 7 5
Common
TH2817B only uses the smallest subset of the RS232C standard, the signal are listed as
37
TH2817B Operation Manual
follows.
TH2817B’s RS232 interface may be different with the standard RS232C. The pin
configuration is shown as follows.
DTR(4)
DSR(6)
RXD(2) (2) RXD
Computer
(controller) TXD(3) (3) TXD TH2817B
There may be some difference between TH2817B’s RS232 interface and a standard
RS232C interface. You can make the connection cable by yourself according the diagram or
order one from our company.
Note: Pin 4 and 6, pin 7 and 8 are shorted respectively at the end of controller.
z When the RS232C interface is used to communicate with a controller, RS232 bus mode
should be selected as follows.
1) Press SYSTEM menu key, the System Config page will be displayed.
2) Move the cursor to the Bus Mode field.
3) Press RS232 to select the bus mode to RS232 bus mode.
38
TH2817B Operation Manual
z Software Protocol
1) The controller sends the command using the ASCII code with NL as the end
character. TH2817B executes the command after the end character NL is
received.
2) The character received by TH2817B will be sent back to the controller again. The
controller will not send the next character until the last return character is
received correctively from TH2817B. If the controller cannot receive the character
sent back by TH2817B, the reasons may be as follows.
A. The serial interface is not connected correctly.
B. Check if the RS232 function is turned on and TALK ONLY function is
turned off.
C. When TH2817B is executing a bus command, TH2817B will not accept
any character through the serial interface at the same time and the
character sent by controller will be ignored. In order to make sure the
whole command is sent and received correctly, the character without a
return character should be sent again by the controller.
3) TH2817B sends information under following two conditions. The first is when a
character is received normally; TH2817B will send the character back as response.
The second is when a query command is received; TH2817B will send the query
response information.
4) Once a query command is received, TH2817B will send the query response
information immediately even if the whole command has not been executed. So if
the command includes two queries, the controller should read the query
responses twice. One query is recommended to be included in a single command.
5) A query response is sent out in ASCII codes with NL as the terminal character.
6) Several query responses will be sent continuously with 1ms interval. The
controller should be ready to receive the responses; otherwise the response
information will be lost.
7) The controller should receive the query response terminal character NL.
Otherwise you will confuse a terminal character NL with a returned character. At
the same time the controller should receive the last returned character before
receiving a query response.
39
TH2817B Operation Manual
8) For some commands that will take a long time to execute, for example Correction
command, the controller should keep waiting to avoid the next command being
lost when TH2817B is executing the former command.
main()
{
port_init( PORT,0xe3 );/* initilize serial port:baud = 9600,no verify,1 bit stop,8 bit data
*/
string_wr( "trig:sour bus;*trg" );
string_rd( input );
printf( "\n%s",input );
string_wr( "freq 10khz" );
string_wr( "func:imp:apar cs;bpar d" );
string_wr( "voltage:level 0.3v" );
}
40
TH2817B Operation Manual
{ send_port( PORT,0xaa );
for( n = 100;n;n-- )
{
delay( 2 ); /* wait about 2ms */
if( kbhit() && ( getch() == 27 ) ) /* if escape key keypress */
{ printf( "\nE20:Serial Port Write Canceled!" );
exit(1);
}
if( check_stat(PORT) & 256 )
{ c = read_port( PORT );
break;
}
}
if( n ) break;/*TH2817B/C not ready for receive*/
}
if( c != 0xcc )/*check receive*/
{ printf( "\nE10:Serial Port HandShake Failure!" );
exit(1);
}
c = read_port( PORT );
if( c == 0xcc ) continue;/*skip the more handshake characters*/
41
TH2817B Operation Manual
42
TH2817B Operation Manual
{ union REGS r;
r.x.dx = port; /* serial port */
r.h.ah = 0; /* int14 function0:initial serial port */
r.h.al = code; /* initialization code */
int86( 0x14,&r,&r );
}
z The total length of cable in one bus system must be less than or equal to two meters
times the number of devices connected on the bus (the GPIB controller counts as one
device) and the total length of cable must not exceed 20 meters.
z There are no restrictions on how the cables are connected together. However, it is
recommended that no more than four piggyback connectors be stacked together on
any one device. The resulting structure could exert enough force on the connector
mounting to damage it.
43
TH2817B Operation Manual
44
TH2817B Operation Manual
Code Function
SH1 Complete Source Handshake capability
AH1 Complete Acceptor Handshake capability
T3 Basic Talker; Talk-Only; unaddressed if MLA; no serial poll
L3 Basic Listener; unaddressed if MTA; no Listen Only
RL1 Remote/Local capability
DC1 Device Clear capability
DT1 Device Trigger capability
C0 No Controller capability
E1 Drivers are open-collector
45
TH2817B Operation Manual
46
TH2817B Operation Manual
The handler interface outputs signals to indicate measurement completed, bin sorting
judgments of the comparator function, and Go/No-Go judgments of the list sweep
comparator function. The handler interface also has an input for an external trigger signal.
Using these signals, TH2817B can easily be combined with a component handler and a
system controller to fully automate component testing, sorting, and quality control data
processing to increase production efficiency. Table 9-1 lists the handler interface
specifications.
Specifications
Decision output for
Bin number, out of bins, and rejection status
comparator function
Decision output for
IN/OUT at each sweep point and pass/fail for
list sweep
Output signal the result of sequential comparisons
comparator function
(Negative true,
IDX Analog measurement complete
open collector,
Full measurement complete.
opto-isolated)
Display time is not included on the BinNo Disp
EOM page.
Display time is included on the List Sweep
page.
Input Signal
TRIG External Trigger, Pulse width≥1μs
(Opto-isolated)
47
TH2817B Operation Manual
8 NC No Connection
9 NC No Connection
48
TH2817B Operation Manual
33 COM2
34 COM1 Common for EXTV1
35 COM1
36 COM1
PHI(OUT)
Primary BIN 1
Parameter
SREJ(AUX) BIN 2 SREJ(AUX)
BIN 3
PLO(OUT)
Secondary Parameter
49
TH2817B Operation Manual
T1 T2 T3
/TRIG
/IDX
/EOM
50
TH2817B Operation Manual
the internally supplied voltage (+5V), or to an externally applied voltage ( EXTV: +5V to
+24V) by setting jumpers.
The electrical characteristics of the DC isolated outputs are divided into two types (See
Table 6-4).
Since the power source for the Comparison Output and Control Output signals are different,
two circuit commons (COM1, COM2) are made available.
51
TH2817B Operation Manual
30 /IDX
31 /EOM
32,33 COM2
*
J204
grounding
*Default jumper wire position
1 /BIN1
2 /BIN2
9 /BIN9
10 /OUT
11 /AUX
19 /PHI
20 /PLO
21 /SRE
34,35,36
JCOM1
*
J202
grounding
*Default jumper wire position
52
TH2817B Operation Manual
J203
* /TRI EXV
680 12,13
1.2k G
2.2k
trigger resistance trigger
with limit current J205 edge
COM
J203
1.2k G
2.2k
trigger
trigger resistance
J205 edge
with limit current COM
53
TH2817B Operation Manual
7.1 Packing
Check the contents when you receive the instrument. If the contents are incomplete, if
there is mechanical damage or defect, or if the instrument does not pass the power-on
self-tests, please contact our company or business department immediately (IEEE488
interface is optional).
7.2 Warranty
This Tonghui instrument product is warranted against defects in material and workmanship
for a period of two years from the date of shipment or purchase. During the warranty
period, Tonghui will, at its option, either repair or replace the products which prove to be
defective. And warranty card is required when the product needs repairing. And all Tonghui
products are warranted for service of lifetime maintenance.
The foregoing warranty shall not apply to defects resulting from improper or inadequate
maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification
or misuse, operation outside of the environmental specifications for the product, or
improper site preparation or maintenance.
54