PeakTech 1404
PeakTech 1404
PeakTech 1404
User Manual
i
6. Demonstration ................................................................................................ 51
Example 1: Measurement a Simple Signal .............................................................. 51
Example 2: Gain of a Amplifier in a Metering Circuit .............................................. 52
Example 3: Capturing a Single Signal ..................................................................... 53
Example 4: Analyze the Details of a Signal ............................................................. 54
Example 5: Application of X-Y Function .................................................................. 56
Example 6: Video Signal Trigger .............................................................................. 57
7. Troubleshooting ............................................................................................. 59
8. Technical Specifications ............................................................................... 60
General Technical Specifications ............................................................................ 63
9. Appendix ......................................................................................................... 64
Appendix A: Enclosure ............................................................................................. 64
Appendix B: General Care and Cleaning ................................................................. 64
ii
1. General Safety Requirements
This product complies with the requirements of the following directives of
the European Union for CE conformity: 2014/30/EU (electromagnetic
compatibility), 2014/35/EU (low voltage), 2011/65/EU (RoHS).
Before use, please read the following safety precautions to avoid any
possible bodily injury and to prevent this product or any other connected
products from damage. To avoid any contingent danger, ensure this product
is only used within the ranges specified.
Only a qualified person should perform internal maintenance.
To avoid Fire or Personal Injury:
Use Proper Power Cord. Use only the power cord supplied with the product
and certified to use in your country.
Connect or Disconnect Correctly. When the probe or test lead is connected
to a voltage source, please do not connect and disconnect the probe or test
lead.
Product Grounded. This instrument is grounded through the power cord
grounding conductor. To avoid electric shock, the grounding conductor must be
grounded. The product must be grounded properly before any connection with
its input or output terminals.
When the instrument is powered by AC, do not measure AC power
sources directly otherwise it will cause a short circuit. This is because
the testing ground and power cord ground conductor are connected.
Check all Terminal Ratings. To avoid fire or shock hazard, check all ratings
and markings on this product. Refer to the user manual for more information
about ratings before connecting to the instrument.
Do not operate without covers. Do not operate the instrument with covers or
panels removed.
Use the Proper Fuse. Use only the specified type and rating fuse for this
instrument.
Avoid exposed circuit. Be careful when working on exposed circuitry to avoid
risk of electric shock or other injury.
Do not operate if any damage. If you suspect damage to the instrument,
have it inspected by qualified service personnel before further use.
Use your Oscilloscope in a well-ventilated area. Make sure the instrument
installed with proper ventilation.
Do not operate in damp conditions.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
1
2. Safety Terms and Symbols
Safety Terms
Terms in this manual (The following terms may appear in this manual):
Safety Symbols
Symbols on the product. The following symbol may appear on the product:
Test Ground
2
To avoid body damage and prevent product and connected equipment damage,
carefully read the following safety information before using the test tool. This
product can only be used in the specified applications.
Warning:
The two channels of the oscilloscope are not electrically isolated. The
channels should adopt a common ground during measuring. To prevent
short circuits, the 2 probe grounds must not be connected to 2 different
non-isolated DC levels.
The diagram of the oscilloscope ground wire connection:
Signal Input
Power Cord
Ground Clip
Warning:
To avoid fire or electrical shock, when the oscilloscope input
signal connected is more than 42V peak (30Vrms) or on circuits of
more than 4800VA, please take note of below items:
Only use accessory insulated voltage probes and test lead.
Check the accessories such as probe before use and
replace it if there are any damages.
Remove probes, test leads and other accessories
immediately after use.
Remove USB cable which connects oscilloscope and
computer.
Do not apply input voltages above the rating of the
instrument because the probe tip voltage will directly
transmit to the oscilloscope. Use with caution when the
probe is set as 1:1.
Do not use exposed metal BNC or banana plug connectors.
Do not insert metal objects into connectors.
3
3. Quick Start
Introduction to the Structure of the Oscilloscope
This chapter makes a simple description of the operation and function of the front
panel of the oscilloscope, enabling you to be familiar with the use of the
oscilloscope in the shortest time.
Front Panel
The front panel has knobs and function buttons. The 5 buttons in the column on
the right side of the display screen are menu selection buttons, through which, you
can set the different options for the current menu. The other buttons are function
buttons, through which, you can enter different function menus or obtain a specific
function application directly.
1 2 3
9 8 7 4
6 5
Figure 3-1 Front panel
1. Display area
2. Menu selection buttons: Select the right menu item.
3. Control (button and knob) area
4. Probe Compensation: Measurement signal (5V/1kHz) output.
5. Run/Stop button:Enable or disable sampling on input signals.
Notice: When there is no sampling at STOP state, the vertical division and the
horizontal time base of the waveform still can be adjusted within a certain range,
in other words, the signal can be expanded in the horizontal or vertical direction.
When the horizontal time base is ≤50ms, the horizontal time base can be
expanded for 4 divisions downwards.
6. Function button area: Total 5 buttons.
4
7. Copy button: This button is the shortcut for Save function in the Utility
function menu. Pressing this button is equal to the Save option in the Save menu.
The waveform, configure or the display screen could be saved according to the
chosen type in the Save menu.
8. USB Host port: It is used to transfer data when external USB equipment
connects to the oscilloscope regarded as "host device". For example: Saving
the waveform to USB flash disk needs to use this port.
9. Power on/off
Rear Panel
1 2
3
4
5
Control Area
5
1
6
User Interface Introduction
1 2 3 4 5 64 7 8
9
22
23 10
22
11
21
20 12
19
18
17 16 15 14 13
Figure 3-4 Illustrative Drawing of Display Interfaces
1. Waveform Display Area.
2. Run/Stop
3. The state of trigger, including:
Auto: Automatic mode and acquire waveform without triggering.
Trig: Trigger detected and acquire waveform.
Ready: Pre-triggered data captured and ready for a trigger.
Scan: Capture and display the waveform continuously.
Stop: Data acquisition stopped.
4. The two blue dotted lines indicates the vertical position of cursor
measurement.
5. The T pointer indicates the horizontal position for the trigger.
6. The pointer indicates the trigger position in the record length.
7. It shows present triggering value and displays the site of present window in
internal memory.
8. It indicates that there is a USB disk connecting with the oscilloscope.
9. Channel identifier of current menu.
10. The waveform of CH1.
11. Right Menu.
12. The waveform of CH2.
13. Current trigger type:
7
Rising edge triggering
Falling edge triggering
The reading shows the trigger level value of the corresponding channel.
14. It indicates the measured type and value of the corresponding channel. "T"
means period, "F" means frequency, "V" means the average value, "Vp"
the peak-peak value, "Vr" the root-mean-square value, "Ma" the maximum
amplitude value, "Mi" the minimum amplitude value, "Vt" the Voltage value
of the waveform's flat top value, "Vb" the Voltage value of the waveform's
flat base, "Va" the amplitude value, "Os" the overshoot value, "Ps" the
Preshoot value, "RT" the rise time value, "FT" the fall time value, "PW" the
+width value, "NW" the -Width value, "+D" the +Duty value, "-D" the -Duty
value, "PD" the Delay A->B value, "ND" the Delay A->B value, "TR"
the Cycle RMS, "CR" the Cursor RMS, "WP" the Screen Duty, "RP" the
Phase, "+PC" the +Pulse count, "-PC" the - Pulse count, "+E" the Rise
edge count, "-E" the Fall edge count, "AR" the Area, "CA" the Cycle area.
15. The readings show the record length.
16. The frequency of the trigger signal.
17. The readings show current sample rate.
18. The readings indicate the corresponding Voltage Division and the Zero
Point positions of the channels. "BW" indicates bandwidth limit.
The icon shows the coupling mode of the channel.
"—" indicates direct current coupling
"~" indicates AC coupling
" " indicates GND coupling
19. The reading shows the setting of main time base.
20. It is cursor measure window, showing the absolute values and the
readings of the cursors.
21. The blue pointer shows the grounding datum point (zero point position) of
the waveform of the CH2 channel. If the pointer is not displayed, it means
that this channel is not opened.
22. The two blue dotted lines indicate the horizontal position of cursor
measurement.
23. The yellow pointer indicates the grounding datum point (zero point position)
of the waveform of the CH1 channel. If the pointer is not displayed, it
means that the channel is not opened.
8
If it is found that the packaging carton or the foamed plastic protection cushion
has suffered serious damage, do not throw it away first till the complete device
and its accessories succeed in the electrical and mechanical property tests.
The instrument carries out all self-check items and shows the Boot Logo. Push
the Utility button, select Function in the right menu. Select Adjust in the left
menu, select Default in the right menu. The default attenuation coefficient set
value of the probe in the menu is 10X.
2. Set the Switch in the Oscilloscope Probe as 10X and Connect the
Oscilloscope with CH1 Channel.
Align the slot in the probe with the plug in the CH1 connector BNC, and then
tighten the probe with rotating it to the right side.
Connect the probe tip and the ground clamp to the connector of the probe
compensator.
3. Push the Autoset Button on the front panel.
The square wave of 1 KHz frequency and 5V peak-peak value will be
displayed in several seconds (see Figure 3-5).
9
Figure 3-5 Auto set
10
Figure 3-7 Adjust Probe
The set values of the probe switch are 1X and 10X (see Figure 3-8).
Caution:
When the attenuation switch is set to 1X, the probe will limit the
bandwidth of the oscilloscope in 5MHz. To use the full bandwidth of
the oscilloscope, the switch must be set to 10X.
11
How to Use the Probe Safely
The safety guard ring around the probe body protects your finger against any
electric shock, shown as Figure 3-9.
Warning:
To avoid electric shock, always keep your finger behind the safety
guard ring of the probe during the operation.
To protect you from suffering from the electric shock, do not touch any
metal part of the probe tip when it is connected to the power supply.
Before making any measurements, always connect the probe to the
instrument and connect the ground terminal to the earth.
12
Figure 3-10 Vertical Control Zone
1. Use the Vertical Position knob to show the signal in the center of the
waveform window. The Vertical Position knob functions the regulating of the
vertical display position of the signal. Thus, when the Vertical Position knob is
rotated, the pointer of the earth datum point of the channel is directed to move
up and down following the waveform.
Measuring Skill
If the channel is under the DC coupling mode, you can rapidly measure the DC
component of the signal through the observation of the difference between the
wave form and the signal ground.
If the channel is under the AC mode, the DC component would be filtered out.
This mode helps you display the AC component of the signal with a higher
sensitivity.
Vertical offset back to 0 shortcut key
Turn the Vertical Position knob to change the vertical display position of
channel and push the position knob to set the vertical display position back to
0 as a shortcut key, this is especially helpful when the trace position is far out
of the screen and want it to get back to the screen center immediately.
2. Change the Vertical Setting and Observe the Consequent State Information
Change.
With the information displayed in the status bar at the bottom of the waveform
window, you can determine any changes in the channel vertical scale factor.
Turn the Vertical Scale knob and change the "Vertical Scale Factor
(Voltage Division)", it can be found that the scale factor of the channel
corresponding to the status bar has been changed accordingly.
Push buttons of CH1, CH2 and Math, the operation menu, symbols,
waveforms and scale factor status information of the corresponding
channel will be displayed in the screen.
13
Introduction to the Horizontal System
Shown as Figure 3-11, there are a button and two knobs in the Horizontal
Controls. The following practices will gradually direct you to be familiar with
the setting of horizontal time base.
1. Turn the Horizontal Scale knob to change the horizontal time base setting
and observe the consequent status information change. Turn the
Horizontal Scale knob to change the horizontal time base, and it can be
found that the Horizontal Time Base display in the status bar changes
accordingly.
2. Use the Horizontal Position knob to adjust the horizontal position of the
signal in the waveform window. The Horizontal Position knob is used to
control the triggering displacement of the signal or for other special
applications. If it is applied to triggering the displacement, it can be
observed that the waveform moves horizontally with the knob when you
rotate the Horizontal Position knob.
Triggering displacement back to 0 shortcut key
Turn the Horizontal Position knob to change the horizontal position of
channel and push the Horizontal Position knob to set the triggering
displacement back to 0 as a shortcut key.
3. Push the Horizontal HOR button to switch between the normal mode and
the wave zoom mode.
14
Figure 3-12 Trigger Control Zone
1. Push the Trigger Menu button and call out the trigger menu. With the
operations of the menu selection buttons, the trigger setting can be
changed.
2. Use the Trigger Level knob to change the trigger level setting.
By turning the Trigger Level knob, the trigger indicator in the screen will
move up and down. With the movement of the trigger indicator, it can be
observed that the trigger level value displayed in the screen changes
accordingly.
Note: Turning the Trigger Level knob can change trigger level value and
it is also the hotkey to set trigger level as the vertical mid point values of
the amplitude of the trigger signal.
3. Push the Force button to force a trigger signal, which is mainly applied to
the "Normal" and "Single" trigger modes.
15
4. Advanced User Guidebook
This chapter will deal with the following topics mainly:
16
How to Set the Vertical System
The VERTICAL CONTROLS includes three menu buttons such as CH1, CH2
and Math, and four knobs such as Vertical Position, Vertical Scale for each
channel.
Setting of CH1 and CH2
Each channel has an independent vertical menu and each item is set
respectively based on the channel.
To turn waveforms on or off (channel, math)
Pushing the CH1, CH2, or Math buttons have the following effect:
• If the waveform is off, the waveform is turned on and its menu is displayed.
• If the waveform is on and its menu is not displayed, its menu will be
displayed.
• If the waveform is on and its menu is displayed, the waveform is turned off
and its menu goes away.
The description of the Channel Menu is shown as the following list:
Function Setting Description
Menu
Pass both AC and DC components of the input
DC
signal.
Coupling AC
Block the DC component of the input signal.
Ground
Disconnect the input signal.
ON Display inverted waveform.
Inverted
OFF Display original waveform.
1X
10X Match this to the probe attenuation factor to have
Probe
100X an accurate reading of vertical scale.
1000X
OFF Turn measuring current off.
Turn measuring current on.
MeasCurr Turn the M knob to set the Amps/Volts ratio. The
V/A
range is 100 mA/V - 1 KA/V.
mV/A
Amps/Volts ratio = 1/Resistor value
Volts/Amp ratio is automatically calculated.
Limit (only for OFF Turn bandwidth off.
50M&100M 20M Limit the channel bandwidth to 20MHz to reduce
model) display noise.
17
signal containing the direct current bias. The operation steps are shown as
below:
(1) Push the CH1 button to show the CH1 SETUP menu.
(2) In the right menu, select Coupling as DC. Both DC and AC
components of the signal are passed.
(3) In the right menu, select Coupling as AC. The direct current
component of the signal is blocked.
2. To invert a waveform
Waveform inverted: the displayed signal is turned 180 degrees against the
phase of the earth potential.
Taking the Channel 1 for example, the operation steps are shown as
follows:
(1) Push the CH1 button to show the CH1 SETUP menu.
(2) In the right menu, select Inverted as ON, the waveform is inverted.
Push again to switch to OFF, the waveform goes back to its original
one.
3. To adjust the probe attenuation
For correct measurements, the attenuation coefficient settings in the
operating menu of the Channel should always match what is on the probe
(see "How to Set the Probe Attenuation Coefficient" on P11). If the
attenuation coefficient of the probe is 1:1, the menu setting of the input
channel should be set to1X.
Take the Channel 1 as an example, the attenuation coefficient of the probe
is 10:1, the operation steps are shown as follows:
(1) Push the CH1 button to show the CH1 SETUP menu.
(2) In the right menu, select Probe. In the left menu, turn the M knob to set
it as 10X.
4. To measure current by probing the voltage drop across a resistor
Take the Channel 1 as an example, if you are measuring current by
probing the voltage drop across a 1Ω resistor, the operation steps are
shown as follows:
(1) Push the CH1 button to show CH1 SETUP menu.
(2) In the right menu, set MeasCurr as V/A mV/A, turn the M knob to set
the Amps/Volts ratio. Amps/Volts ratio = 1/Resistor value. Here the
A/V radio should be set to 1.
18
addition, multiplication, division and subtraction operations between two
channels, or the FFT operation for a channel. Press the Math button to display
the menu on the right.
Taking the additive operation between Channel 1 and Channels 2 for example,
the operation steps are as follows:
1. Press the Math button to display the math menu in the right. The pink M
waveform appears on the screen.
2. In the right menu, select Type as Math.
3. In the right menu, select Factor1 as CH1.
4. In the right menu, select Sign as +.
5. In the right menu, select Factor2 as CH2.
6. Press Next Page in the right menu. Select Vertical (div), the symbol is
in front of div, turn the M knob to adjust the vertical position of Math
waveform.
7. Select Vertical (V/div) in the right menu, the symbol is in front of the
voltage, turn the M knob to adjust the voltage division of Math waveform.
19
signal into its frequency components mathematically (the record length should be
10K or above). The final frequency contains 1024 points ranging from 0Hz to
Nyquist frequency.
Press the Math button to display the menu on the right, select Type as FFT.
Function Menu Setting Description
Type FFT Display the FFT menu
CH1 Select CH1 as FFT source.
Source
CH2 Select CH2 as FFT source.
Hamming
Rectangle
Blackman
Window Select window for FFT.
Hanning
Kaiser
Bartlett
Vrms Select Vrms for Format.
Format
dB Select dB for Format.
Next Page Enter next page
frequency Switch to select the horizontal
Hori (Hz) frequency/di position or time base of the FFT
v waveform, turn the M knob to adjust it
Taking the FFT operation for example, the operation steps are as follows:
1. Press the Math button to display the math menu in the right.
2. In the right menu, select Type as FFT.
3. In the right menu, select Source as CH1.
4. In the right menu, select Window. Select the proper window type in the left
menu.
5. In the right menu, select Format as Vrms or dB.
6. In the right menu, press Hori (Hz) to make the symbol in front of the
frequency value, turn the M knob to adjust the horizontal position of FFT
waveform; then press to make the symbol in front of the frequency/div
below, turn the M knob to adjust the time base of FFT waveform.
7. Select Vertical in the right menu; do the same operations as above to set the
vertical position and voltage division.
20
To select the FFT window
■ There are 6 FFT windows. Each one has trade-offs between frequency
resolution and magnitude accuracy. What you want to measure and your source
signal characteristics help you to determine which window to use. Use the
following guidelines to select the best window.
21
Good for magnitude, but poorer frequency
resolution than Hamming.
Recommend to use for:
1. The Vertical Position knob is used to adjust the vertical positions of the
22
waveforms.
The analytic resolution of this control knob changes with the vertical
division.
2. The Vertical Scale knob is used to regulate the vertical resolution of the
wave forms. The sensitivity of the vertical division steps as 1-2-5.
The vertical position and vertical resolution is displayed at the left bottom
corner of the screen (see Figure 4-1).
23
In normal mode, the Horizontal Position and Horizontal Scale knobs are
used to adjust the horizontal position and time base of the Main window.
In wave zoom mode, the Horizontal Position and Horizontal Scale knobs are
used to adjust the horizontal position and time base of the Zoom window.
Trigger Control
The oscilloscope provides two trigger types: single trigger, alternate trigger.
Each type of trigger has different sub menus.
Single trigger: Use a trigger level to capture stable waveforms in two
channels simultaneously.
Alternate trigger: Trigger on non-synchronized signals.
The Single Trigger, Alternate Trigger menus are described respectively as
24
follows:
Single Trigger
Single trigger has two types: edge trigger, video trigger.
Edge Trigger: It occurs when the trigger input passes through a specified voltage
level with the specified slope.
Video Trigger: Trigger on fields or lines for standard video signal.
The two trigger modes in Single Trigger are described respectively as follows:
1. Edge Trigger
An edge trigger occurs on trigger level value of the specified edge of input signal.
Select Edge trigger mode to trigger on rising edge or falling edge.
Push the Trigger Menu button to display the Trigger menu on the right. Select
Type as Single in the right menu. Select Single as Edge in the right menu.
In Edge Trigger mode, the trigger setting information is displayed on bottom right
25
Trigger Level: trigger level indicates vertical trig position of the channel, rotate trig
level knob to move trigger level, during setting, a dotted line displays to show trig
position, and the value of trigger level changes at the right corner, after setting,
dotted line disappears.
2. Video Trigger
Choose video trigger to trigger on fields or lines of NTSC, PAL or SECAM standard
video signals.
Push the Trigger Menu button to display the Trigger menu on the right. Select
Type as Single in the right menu. Select Single as Video in the right menu.
In Video Trigger mode, the trigger setting information is displayed on bottom right
26
trigger.
CH1 Channel 1 as trigger signal.
Source
CH2 Channel 2 as trigger signal.
Next Page Enter next page
AC Block the direct current component.
Coupling
DC Allow all component pass.
Trigger on rising edge
Slope
Trigger on falling edge
27
Function Setting Description
Menu
Sample Normal sampling mode.
Persist
When the Persist function is used, the persistence display effect of the picture
tube oscilloscope can be simulated. The reserved original data is displayed in fade
color and the new data is in bright color.
(1) Push the Acquire button.
(2) In the right menu, press Persist to select the persist time, including OFF, 1
Second, 2 Seconds, 5 Seconds and Infinity. When the "Infinity" option is
set for Persist Time, the measuring points will be stored till the controlling value
is changed. Select OFF to turn off persistence and clear the display.
XY Format
This format is only applicable to Channel 1 and Channel 2. After the XY display
format is selected, Channel 1 is displayed in the horizontal axis and Channel 2 in
the vertical axis; the oscilloscope is set in the un-triggered sample mode: the data
28
are displayed as bright spots.
The operations of all control knobs are as follows:
The Vertical Scale and the Vertical Position knobs of Channel 1 are used to
set the horizontal scale and position.
The Vertical Scale and the Vertical Position knobs of Channel 2 are used to
set the vertical scale and position continuously.
The following functions can not work in the XY Format:
Reference or digital wave form
Cursor
Trigger control
FFT
Operation steps:
1. Push the Acquire button to show the right menu.
2. Select XY Mode as ON or OFF in the right menu.
Counter
It is a 6-digit single-channel counter. The counter can only measure the frequency
of the triggering channel. The frequency range is from 2Hz to the full bandwidth.
Only if the measured channel is in Edge mode of Single trigger type, the counter
can be enabled. The counter is displayed at the bottom of the screen.
Operation steps:
1. Push Trigger Menu button, set the trigger type to Single, set the trigger mode
to Edge, select the signal source.
2. Push the Acquire button to show the right menu.
29
CH1 Choose the waveform to be saved.
CH2 (Choose All to save all the waveforms
Source Math that are turned on. You can save into
All the current internal object address, or
into USB storage as a single file.)
The object Wave0 –Wave15 are listed
in the left menu, turn the M knob to
choose the object which the waveform
is saved to or recall from.
Recall or close the waveform stored in
the current object address. When the
ON
Object show is ON, if the current object
OFF
address has been used, the stored
waveform will be shown, the address
number and relevant information will be
displayed at the top left of the screen; if
the address is empty, it will prompt
"None is saved".
Next Page Enter next page
Close all the waveforms stored in the
Close All
object address.
BIN For internal storage, only BIN can be
File Format TXT selected. For external storage, the
CSV format can be BIN, TXT or CSV.
Save the waveform of the source to the
Save
selected address.
Save to internal storage or USB
storage. When External is selected, the
Internal file name is editable. The BIN
Storage
External waveform file could be open by
waveform analysis software (on the
supplied CD).
Prev Page Enter previous page
When the Type is selected as Configure, the menu is shown as the following
table:
Function Menu Setting Description
Function Save Display the save function menu
Configur
Type Choose the saving type as configure.
e
Setting1
Configure ….. The setting address
Setting8
Save the current oscilloscope configure
Save
to the internal storage
30
Recall the configure from the selected
Load
address
When the Type is selected as Image, the menu is shown as the following table:
Function Menu Setting Description
Function Save Display the save function menu
Type Image Choose the saving type as image.
Save the current display screen. The
file can be only stored in a USB
storage, so a USB storage must be
Save
connected first. The file name is
editable. The file is stored in BMP
format.
In order to save the waveform of CH1 and CH2 into the USB storage as a BIN file,
the operation steps should be followed:
1. Turn on CH1 and CH2 channels, turn off the Math channel.
2. Push the Utility button, select Function in the right menu, select Save in the
left menu. In the right menu, select Type as Wave.
3. Saving: In the right menu, select Source as All.
31
4. In the right menu, press Next Page, and select File Format as BIN.
5. In the right menu, select Storage as External.
6. In the right menu, select Storage, an input keyboard used to edit the file name
will pop up. The default name is current system date and time. Turn the M
knob to choose the keys; press the M knob to input the chosen key. The length
of file name is up to 25 characters. Select the key in the keyboard to
confirm.
7. Recalling: The BIN waveform file could be open by waveform analysis
software (on the supplied CD).
Shortcut for Save function:
The Copy button on the bottom right of the front panel is the shortcut for Save
function in the Utility function menu. Pressing this button is equal to the Save
option in the Save menu. The waveform, configure or the display screen could be
saved according to the chosen type in the Save menu.
Save the current screen image:
The screen image can only be stored in USB disk, so you should connect a USB
disk with the instrument.
1. Install the USB disk: Insert the USB disk into the "8. USB Host port" of
"Figure 3-1 Front panel". If an icon appears on the top right of the screen,
the USB disk is installed successfully. If the USB disk cannot be recognized,
format the USB disk according to the methods in "USB disk Requirements" on
P32.
2. After the USB disk is installed, push the Utility button, select Function in the
right menu, select Save in the left menu. In the right menu, select Type as
Image.
3. Select Save in the right menu, an input keyboard used to edit the file name will
pop up. The default name is current system date and time. Turn the M knob to
choose the keys; press the M knob to input the chosen key. The length of file
name is up to 25 characters. Select the key in the keyboard to confirm.
32
2. Right click Computer- Manage to enter Computer Management interface.
3. Click Disk Management menu, and information about the USB disk will display
on the right side with red mark 1 and 2.
33
Figure 4-4: Formatting the USB disk setting
6. Formatting process.
34
and information about the USB disk will display on the right side with red mark
1 and 2.
35
Figure 4-9: Apply setting
7. Formatting process
36
Function Setting Description
Menu
Function Configure Show the configure menu
Lock all keys. Unlock method: push
Trigger Menu button in trigger control
KeyLock
area, then push Force button, repeat 3
times.
About Show the version and serial number
●Display
Push the Utility button, select Function in the right menu, select Display in the
left menu.
The description of Display Menu is shown as the follows:
●Adjust
Push the Utility button, select Function in the right menu, select Adjust in the
left menu.
The description of Adjust Menu is shown as the follows:
Function Menu Description
Self Cal Carry out the self-calibration procedure.
Default Call out the factory settings.
ProbeCh. Check whether probe attenuation is good.
37
from the input connector. Push the Utility button, select Function in the right
menu, the function menu will display at the left, select Adjust. If everything is
ready, select Self Cal in the right menu to enter the self-calibration procedure of
the instrument.
Probe checking
To check whether probe attenuation is good. The results contain three
circumstances: Overflow compensation, Good compensation, Inadequate
compensation. According to the checking result, users can adjust probe
attenuation to the best. Operation steps are as follows:
1. Connect the probe to CH1, adjust the probe attenuation to the maximum.
2. Push the Utility button, select Function in the right menu, select Adjust in the
left menu.
3. Select ProbeCh. in the right menu, tips about probe checking shows on the
screen.
4. Select ProbeCh. again to begin probe checking and the checking result will
occur after 3s; push any other key to quit.
● Save
You can save the waveforms, configures or screen images. Refer to "How to Save
and Recall a Waveform" on page 29.
● Update
Use the front-panel USB port to update your instrument firmware using a USB
memory device. Refer to "How to Update your Instrument Firmware" on page 40.
● AutoScale
This is a very useful function for first time users to carry out a simple and quick test
on the input signal. The function is applied to follow-up signals automatically even
if the signals change at any time. Autoscale enables the instrument to set up
trigger mode, voltage division and time scale automatically according to the type,
amplitude and frequency of the signals.
The menu is as follows:
Function Setting Instruction
Menu
ON Turn on Autoscale.
Autoscale
OFF Turn off Autoscale.
Follow-up and adjust both vertical and horizontal
settings.
Mode Follow-up and only adjust horizontal scale.
38
Show Multi-period waveforms.
Wave
Only show one or two periods.
If you want to measure the two-channel signal, you can do as the follows:
1. Push the Utility button, the function menu will appear.
2. In the left menu, Press the menu key on the right side of the screen orturn the
M knob to select Autoscale.
3. In the right menu, select ON in the Autoscale menu item.
4. In the right menu, select Mode, select .
5. In the right menu, select Wave, select .
39
How to Update your Instrument Firmware
Use the front-panel USB port to update your instrument firmware using a USB
memory device.
USB memory device requirements: Insert a USB memory device into the USB
port on the front panel. If the icon appears on the top right of the screen, the
USB memory device is installed successfully. If the USB memory device cannot be
detected, format the USB memory device according to the methods in "USB disk
Requirements" on P32.
Caution: Updating your instrument firmware is a sensitive operation, to prevent
damage to the instrument, do not power off the instrument or remove the USB
memory device during the update process.
To update your instrument firmware, do the following:
1. Push the Utility button, select Function in the right menu, select Configure in
the left menu, select About in the right menu. View the model and the currently
installed firmware version.
2. Check if the website offers a newer firmware version. Download the firmware
file. The file name must be Scope.update. Copy the firmware file onto the root
directory of your USB memory device.
3. Insert the USB memory device into the front-panel USB port on your
instrument.
4. Push the Utility button, select Function in the right menu, select Update in the
left menu.
5. In the right menu, select Start, the messages below will be shown.
6. In the right menu, select Start again, the interfaces below will be displayed in
sequence. The update process will take up to three minutes. After completion,
the instrument will be shut down automatically.
40
7. Press the button to power on the instrument.
41
Remove All Remove all the measures
Measure
Only if the waveform channel is in the ON state, the measurement can be
performed. The automatic measurement can not be performed in the following
situation: 1) On the saved waveform. 2) On the Dual Wfm Math waveform. 3) On
the Video trigger mode.
On the Scan format, period and frequency can not be measured.
Measure the period, the frequency of the CH1, following the steps below:
42
Figure 4-14
Mean: The arithmetic mean over the entire waveform.
PK-PK: Peak-to-Peak Voltage.
RMS: The true Root Mean Square voltage over the entire waveform.
Max: The maximum amplitude. The most positive peak voltage measured over
the entire waveform.
Min: The minimum amplitude. The most negative peak voltage measured over
the entire waveform.
Vtop: Voltage of the waveform's flat top, useful for square/pulse waveforms.
Vbase: Voltage of the waveform's flat base, useful for square/pulse waveforms.
Vamp: Voltage between Vtop and Vbase of a waveform.
OverShoot: Defined as (Vmax-Vtop)/Vamp, useful for square and pulse
waveforms.
PreShoot: Defined as (Vmin-Vbase)/Vamp, useful for square and pulse
waveforms.
Cycle RMS: The true Root Mean Square voltage over the first entire period of
the waveform.
Cursor RMS: The true Root Mean Square voltage over the range of two cursors.
43
Figure 4-15
Rise Time: Time that the leading edge of the first pulse in the waveform takes
to rise from 10% to 90% of its amplitude.
Fall Time: Time that the falling edge of the first pulse in the waveform takes to
fall from 90% to 10% of its amplitude.
+D width: The width of the first positive pulse in 50% amplitude points.
-D width: The width of the first negative pulse in the 50% amplitude points.
+Duty: +Duty Cycle, defined as +Width/Period.
-Duty:-Duty Cycle, defined as -Width/Period.
Delay A→B : The delay between the two channels at the rising edge.
Delay A→B : The delay between the two channels at the falling edge.
Screen Duty: Defines as (the width of the positive pulse)/(Entire period)
Phase: Compare the rising edge of CH1 and CH2, calculate phase difference
of two channels.
Phase difference=(Delay between channels at the rising
edge÷Period)×360°.
Other measurements
+PulseCount : The number of positive pulses that rise above the mid
reference crossing in the waveform.
-PulseCount : The number of negative pulses that fall below the mid
reference crossing in the waveform.
RiseEdgeCnt : The number of positive transitions from the low
reference value to the high reference value in the waveform.
FallEdgeCnt : The number of negative transitions from the high
44
reference value to the low reference value in the waveform.
Area : The area of the whole waveform within the screen and the unit is
voltage-second. The area measured above the zero reference (namely the
vertical offset) is positive; the area measured below the zero reference is
negative. The area measured is the algebraic sum of the area of the whole
waveform within the screen.
Cycle Area : The area of the first period of waveform on the screen and
the unit is voltage-second. The area above the zero reference (namely the
vertical offset) is positive and the area below the zero reference is negative.
The area measured is the algebraic sum of the area of the whole period
waveform.
Note: When the waveform on the screen is less than a period, the period area
measured is 0.
45
Perform the following operation steps for the time and voltage cursor
measurement of the channel CH1:
1. Push Cursor to display the cursor menu.
2. In the right menu, select Source as CH1.
3. Press the first menu item in the right menu, select Time&Voltage for Type,
two blue dotted lines displayed along the horizontal direction of the screen,
two blue dotted lines displayed along the vertical direction of the screen.
Cursor measure window at the left bottom of the screen shows the cursor
readout.
4. In the right menu, select Line Type as Time to make the vertical cursors
active. If the Line in the right menu is select as a, turn the M knob to move
line a to the right or left. If b is selected, turn the M knob to move line b.
5. In the right menu, select Line Type as Voltage to make the horizontal
cursors active. Select Line in the right menu as a or b, turn the M knob to
move it.
6. Push the horizontal HOR button to enter wave zoom mode. Push Cursor
to show the right menu, select Window as Main or Extension to make the
cursors shown in the main window or zoom window.
Auto Cursor
For the AutoCursr type, the horizontal cursors are set as the intersections of the
vertical cursors and the waveform.
46
The Cursor Measurement for FFT mode
In FFT mode, push the Cursor button to turn cursors on and display the cursor
menu.
The description of the cursor menu in FFT mode is shown as the following table:
Function
Setting Description
Menu
Vamp Display the Vamp measurement cursor and
Freq menu.
Freq&Vamp Display the Freq measurement cursor and
menu.
Type AutoCursr Display the Freq and Vamp measurement
cursor and menu.
The horizontal cursors are set as the
intersections of the vertical cursors and the
waveform
Line Type Freq Makes the vertical cursors active.
(Freq&Vam
p type) Vamp Makes the horizontal cursors active.
Window Main Measure in the main window.
(Wave zoom
mode) Extension Measure in the FFT extension window.
a Turn the M knob to move line a.
b Turn the M knob to move line b.
Line
ab Two cursors are linked. Turn the M knob to
move the pair of cursors.
Display the channel to which the cursor
Source Math FFT
measurement will be applied.
Perform the following operation steps for the amplitude and frequency cursor
measurement of math FFT:
1. Press the Math button to display the right menu. Select Type as FFT.
47
2. Push Cursor to display the cursor menu.
3. In the right menu, select Window as Extension.
4. Press the first menu item in the right menu, select Freq&Vamp for Type,
two blue dotted lines displayed along the horizontal direction of the screen,
two blue dotted lines displayed along the vertical direction of the screen.
Cursor measure window at the left bottom of the screen shows the cursor
readout.
5. In the right menu, select Line Type as Freq to make the vertical cursors
active. If the Line in the right menu is select as a, turn the M knob to move
line a to the right or left. If b is selected, turn the M knob to move line b.
6. In the right menu, select Line Type as Vamp to make the horizontal
cursors active. Select Line in the right menu as a or b, turn the M knob to
move it.
7. In the right cursor menu, you can select Window as Main to make the
cursors shown in the main window.
48
Judge waveform type by Autoset
Five kinds of types: Sine, Square, video signal, DC level, Unknown
signal.
Menu as follow:
Waveform Menu
Sine Multi-period, Single-period, FFT, Cancel Autoset
Square Multi-period, Single-period, Rising Edge, Falling Edge,
Cancel Autoset
Video signal Type (line, field), Odd, Even, Line NO., Cancel
Autoset
DC level/Unknown Cancel Autoset
signal
Description for some icons:
Multi-period: To display multiple periods
Single-period: To display single period
FFT: Switch to FFT mode
Rising Edge: Display the rising edge of square waveform
Falling Edge: Display the falling edge of square waveform
Cancel Autoset : Go back to display the upper menu and waveform
information
Note: The Autoset function requires that the frequency of signal should be no
lower than 20Hz, and the amplitude should be no less than 5mv. Otherwise,
the Autoset function may be invalid.
[Run/Stop] button
Enable or disable sampling on input signals.
Notice: When there is no sampling at STOP state, the vertical division and the
horizontal time base of the waveform still can be adjusted within a certain
range, in other words, the signal can be expanded in the horizontal or vertical
direction.
When the horizontal time base is ≤50ms, the horizontal time base can be
expanded for 4 divisions downwards.
[Copy] button
This button is the shortcut for Save function in the Utility function menu.
Pressing this button is equal to the Save option in the Save menu. The
waveform, configure or the display screen could be saved according to the
chosen type in the Save menu. For more details, please see "How to Save and
Recall a Waveform" on P29.
49
5. Communication with PC
The oscilloscope supports communications with a PC through USB. You can use
the Oscilloscope communication software to store, analyze, display the data and
remote control.
To learn about how to operate the software, you can push F1 in the software to
open the help document.
Here is how to connect with PC via USB port.
(1) Install the software: Install the Oscilloscope communication software on the
supplied CD.
(2) Connection: Use a USB data cable to connect the USB Device port in the
right panel of the Oscilloscope to the USB port of a PC.
(3) Install the driver: Run the Oscilloscope communication software on PC, push
F1 to open the help document. Follow the steps of title "I. Device connection"
in the document to install the driver.
(4) Port setting of the software: Run the Oscilloscope software; click
"Communications" in the menu bar, choose "Ports-Settings", in the setting
dialog, choose "Connect using" as "USB". After connect successfully, the
connection information in the bottom right corner of the software will turn
green.
50
6. Demonstration
Example 1: Measurement a Simple Signal
The purpose of this example is to display an unknown signal in the circuit, and
measure the frequency and peak-to-peak voltage of the signal.
1. Carry out the following operation steps for the rapid display of this
signal:
(1) Set the probe menu attenuation coefficient as 10X and that of the switch in
the probe switch as 10X (see "How to Set the Probe Attenuation
Coefficient" on P11).
(2) Connect the probe of Channel 1 to the measured point of the circuit.
(3) Push the Autoset button.
The oscilloscope will implement the Autoset to make the waveform optimized,
based on which, you can further regulate the vertical and horizontal divisions
till the waveform meets your requirement.
2. Perform Automatic Measurement
The oscilloscope can measure most of the displayed signals automatically. To
measure the period, the frequency of the CH1, following the steps below:
(1) Push the Measure button to show the right menu.
(2) Select AddCH1 in the right menu.
(3) In the left Type menu, turn the M knob to select Period.
(4) In the right menu, select AddCH1. The period type is added.
(5) In the left Type menu, turn the M knob to select Frequency.
(6) In the right menu, select AddCH1. The frequency type is added.
The measured value will be displayed at the bottom left of the screen
automatically (see Figure 6-1).
51
Figure 6-1 Measure period and frequency value for a given signal
52
Figure 6-2 Waveform of Gain Measurement
53
(10) In the right menu, select Slope as (rising).
(11) Turn the Trigger Level knob and adjust the trigger level to the roughly
50% of the signal to be measured.
(12) Check the Trigger State Indicator on the top of the screen, if it is not Ready,
push down the Run/Stop button and start acquiring, wait for trigger to
happen. If a signal reaches to the set trigger level, one sampling will be
made and then displayed in the screen. By using this approach, a random
pulse can be captured easily. For instance, if we want to find a burst burr of
high amplitude, set the trigger level to a slightly higher value of the average
signal level, push the Run/Stop button and wait a trigger. Once there is a
burr occurring, the instrument will trigger automatically and record the
waveform during the period around the trigger time. By turning the
Horizontal Position knob in the horizontal control area in the panel, you
can change the horizontal triggering position to obtain the negative delay,
making an easy observation of the waveform before the burr occurs (see
Figure 6-3).
54
(2) In the right menu, select Acqu Mode as Peak Detect.
The signal displayed on the screen containing some noise, by turning on Peak
Detect function and changing time base to slow down the incoming signal, any
peaks or burr would be detected by the function (see Figure 6-4).
55
Figure 6-5 Reduce Noise level by using Average function
56
(7) Turn the Vertical Scale and Vertical Position knobs, optimizing the
waveform.
(8) With the elliptical oscillogram method adopted, observe and calculate the
phase difference (see Figure 6-6).
Based on the expression sin (q) =A/B or C/D, thereinto, q is the phase
difference angle, and the definitions of A, B, C, and D are shown as the graph
above. As a result, the phase difference angle can be obtained, namely, q =±
arcsin (A/B) or ± arcsin (C/D). If the principal axis of the ellipse is in the I and
III quadrants, the determined phase difference angel should be in the I and IV
quadrants, that is, in the range of (0 - π /2) or (3π / 2 - 2π). If the principal axis
of the ellipse is in the II and IV quadrants, the determined phase difference
angle is in the II and III quadrants, that is, within the range of (π / 2 - π) or (π -
3π /2).
57
(1) Push the Trigger Menu button to display the right menu.
(2) In the right menu, select Type as Single.
(3) In the right menu, select Single as Video.
(4) In the right menu, select Source as CH1.
(5) In the right menu, select Modu as NTSC.
(6) In the right menu, press Next Page, select Sync as Field.
(7) Turn the Vertical Scale, Vertical Position and Horizontal Scale knobs to
obtain a proper waveform display (see Figure 6-7).
58
7. Troubleshooting
1. Oscilloscope is powered on but no Display.
Check whether the power connection is connected properly.
Restart the instrument after completing the checks above.
If the problem persists, please contact us and we will be under your
service.
2. After acquiring the signal, the waveform of the signal is not displayed in
the screen.
Check whether the probe is properly connected to the signal connecting
wire.
Check whether the signal connecting wire is correctly connected to the
BNC (namely, the channel connector).
Check whether the probe is properly connected with the object to be
measured.
Check whether there is any signal generated from the object to be
measured (the trouble can be shot by the connection of the channel from
which there is a signal generated with the channel in fault).
Make the signal acquisition operation again.
3. The measured voltage amplitude value is 10 times or 1/10 of the actual
value.
Look at the attenuation coefficient for the input channel and the attenuation
ration of the probe, to make sure they are match (see "How to Set the Probe
Attenuation Coefficient" on P11).
4. There is a waveform displayed, but it is not stable.
Check whether the Source item in the TRIG MODE menu is in conformity
with the signal channel used in the practical application.
Check on the trigger Type item: The common signal chooses the Edge
trigger mode for Type and the video signal the Video. If Alternate trigger is
selected, both of the channel 1 and channel 2 trigger levels should be
adjusted to the proper position. Only if a proper trigger mode is applied,
the waveform can be displayed steadily.
5. No Display Responses to the Push-down of Run/Stop.
Check whether Normal or Signal is chosen for Polarity in the TRIG MODE
menu and the trigger level exceeds the waveform range.
If it is, make the trigger level is centered in the screen or set the trigger mode as
Auto. In addition, with the Autoset button pressed, the setting above can be
completed automatically.
6. The displaying of waveform seems getting slow after increasing
AVERAGE value in Acqu Mode (see "How to Set the Sampling/Display" on
P27 ), or a longer duration is set in the Persist in Display (see "Persist" on
P28).
It's normal as the Oscilloscope is working hard on many more data points.
59
8. Technical Specifications
Unless otherwise specified, the technical specifications applied are for the
oscilloscope only, and Probes attenuation set as 10X. Only if the oscilloscope
fulfills the following two conditions at first, these specification standards can be
reached.
This instrument should run for at least 30 minutes continuously under the
specified operating temperature.
If change of the operating temperature is up to or exceeds 5℃, do a
"Self-calibration" procedure (see "How to Implement Self-calibration" on
P12).
All specification standards can be fulfilled, except one(s) marked with the word
"Typical".
Performance Characteristics Instruction
1400 5 MHz
1401 10 MHz
Bandwidth 1402 20 MHz
1403 50 MHz
1404 100 MHz
Channel 2 channels
Mode Normal, Peak detect, Averaging
1400
100 MS/s
1401
Acquisition Sample rate
1402 250 MS/s
(real time)
1403 500 MS/s
1404 1 GS/s
Input coupling DC, AC , Ground
Input impedance 1 MΩ±2%, in parallel with 20 pF±5 pF
Probe attenuation
1X,10X,100X,1000X
factor
Max. input voltage 400V (DC+AC, PK - PK)
Channel –channel 50Hz: 100 : 1
Input
isolation 10MHz: 40 : 1
1400
1401 Not support
Bandwidth limit 1402
1403
20 MHz, full bandwidth
1404
1400
0.5 S/s~100 MS/s
1401
Horizontal
Sampling rate range 1402 0.5 S/s~250 MS/s
System
1403 0.5 S/s~500 MS/s
1404 0.5 S/s~1 GS/s
60
Performance Characteristics Instruction
Interpolation (Sinx)/x
Max Record length 10K
1400
5 ns/div – 1000 s/div,
1401
Scanning speed step by 1 – 2 - 5
1402
(S/div)
1403 2 ns/div – 1000 s/div,
1404 step by 1 – 2 - 5
Sampling rate / relay
±100 ppm
time accuracy
Single:
±(1 interval time+100 ppm×reading+0.6
Interval(△T) accuracy ns);
(DC - 100MHz) Average>16:
±(1 interval time +100 ppm×reading+0.4
ns)
Vertical Resolution
8 bits (2 channels simultaneously)
(A/D)
Sensitivity 5 mV/div~5 V/div
1400
1401 ±2 V (5 mV/div – 200 mV/div)
Displacement 1402 ±200 V (500 mV/div – 5
1403 V/div)
1404
1400 5 MHz
1401 10 MHz
Analog bandwidth 1402 20 MHz
1403 50 MHz
1404 100 MHz
Vertical
Single bandwidth Full bandwidth
system
Low Frequency ≥10 Hz (at input, AC coupling, -3 dB)
1400 ≤ 70 ns
1401 ≤ 35 ns
Rise time (at input,
1402 ≤ 17.5 ns
Typical)
1403 ≤ 7.0 ns
1404 ≤ 3.5 ns
DC gain accuracy ±3%
Delta Volts between any two averages of
≥16 waveforms acquired with the same
DC accuracy
scope setup and ambient conditions
(average)
(△V):
±(3% reading + 0.05 div)
Waveform inverted ON/OFF
61
Performance Characteristics Instruction
△V, △T, △T&△V between cursors,
Cursor
auto cursor
Period, Frequency, Mean, PK-PK, RMS,
Max, Min, Top, Base, Amplitude,
Overshoot, Preshoot, Rise Time, Fall
Time, +Pulse Width, -Pulse Width, +Duty
Automatic Cycle, -Duty Cycle, Delay A→B , Delay
A→B , Cycle RMS, Cursor RMS,
Measurement Screen Duty, Phase, +Pulse Count,
-Pulse Count, Rise Edge Count, Fall
Edge Count, Area, and Cycle Area.
Waveform Math +, -, *, / ,FFT
Waveform storage 16 waveforms
Bandwidth Full bandwidth
Lissajous
Phase
figure ±3 degrees
difference
Communicatio
USB 2.0 (USB storage)
n port
Counter Support
Trigger:
Performance
Instruction
Characteristics
1400
Trigger level 1401 ±4 div from the screen center
Internal 1402
range 1403
±5 div from the screen center
1404
Trigger level
Accuracy Internal ±0.3 div
(typical)
Trigger
According to Record length and time base
displacement
Trigger Holdoff
100 ns – 10 s
range
50% level setting
Input signal frequency ≥ 50 Hz
(typical)
Edge trigger slope Rising, Falling
62
Performance
Instruction
Characteristics
Support standard NTSC, PAL and
Modulation
SECAM broadcast systems
Video Trigger
Line number 1-525 (NTSC) and 1-625
range (PAL/SECAM)
Mechanical Specifications
Dimension 301 mm× 152 mm×70 mm (L*H*W)
Weight About 1.1 kg
63
9. Appendix
Appendix A: Enclosure
(The accessories subject to final delivery.)
Standard Accessories:
64
All rights, also for translation, reprinting and copy of this manual or parts are reserved.
Reproduction of all kinds (photocopy, microfilm or other) only by written permission of the publisher.
This manual considers the latest technical knowing. Technical changings which are in the interest
of progress reserved.
We herewith confirm, that the units are calibrated by the factory according to the specifications as
per the technical specifications.
We recommend to calibrate the unit again, after one year.
65