Detective DX
Detective DX
Detective-EX/DX -100T ®
User’s Manual
WARRANTY
AMT warrants that the items will be delivered free from defects in material or workmanship. AMT makes no other warranties,
express or implied, and specifically NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE.
AMT’s exclusive liability is limited to repairing or replacing at AMT’s option, items found by AMT to be defective in
workmanship or materials within one year from the date of delivery. AMT’s liability on any claim of any kind, including
negligence, loss, or damages arising out of, connected with, or from the performance or breach thereof, or from the
manufacture, sale, delivery, resale, repair, or use of any item or services covered by this agreement or purchase order, shall in
no case exceed the price allocable to the item or service furnished or any part thereof that gives rise to the claim. In the event
AMT fails to manufacture or deliver items called for in this agreement or purchase order, AMT’s exclusive liability and buyer’s
exclusive remedy shall be release of the buyer from the obligation to pay the purchase price. In no event shall AMT be liable
for special or consequential damages.
Quality Control
Before being approved for shipment, each AMT instrument must pass a stringent set of quality control tests designed to expose
any flaws in materials or workmanship. Permanent records of these tests are maintained for use in warranty repair and as a
source of statistical information for design improvements.
Repair Service
If it becomes necessary to return this instrument for repair, it is essential that Customer Services be contacted in advance of
its return so that a Return Authorization Number can be assigned to the unit. Also, AMT must be informed, either in writing,
by telephone [(865) 482-4411] or by facsimile transmission [(865) 483-2133], of the nature of the fault of the instrument being
returned and of the model, serial, and revision (“Rev” on rear panel) numbers. Failure to do so may cause unnecessary delays
in getting the unit repaired. The AMT standard procedure requires that instruments returned for repair pass the same quality
control tests that are used for new-production instruments. Instruments that are returned should be packed so that they will
withstand normal transit handling and must be shipped PREPAID via Air Parcel Post or United Parcel Service to the designated
AMT repair center. The address label and the package should include the Return Authorization Number assigned. Instruments
being returned that are damaged in transit due to inadequate packing will be repaired at the sender’s expense, and it will be the
sender’s responsibility to make claim with the shipper. Instruments not in warranty should follow the same procedure and AMT
will provide a quotation.
Damage in Transit
Shipments should be examined immediately upon receipt for evidence of external or concealed damage. The carrier making
delivery should be notified immediately of any such damage, since the carrier is normally liable for damage in shipment.
Packing materials, waybills, and other such documentation should be preserved in order to establish claims. After such
notification to the carrier, please notify AMT of the circumstances so that assistance can be provided in making damage claims
and in providing replacement equipment, if necessary.
ORTEC® is a registered trademark of Advanced Measurement Technology, Inc. All other trademarks used herein are the property of their
respective owners.
NOTICE OF PROPRIETARY PROPERTY — This document and the information contained in it are the proprietary property of
AMETEK Inc. It may not be copied or used in any manner nor may any of the information in or upon it be used for any purpose without the
express written consent of an authorized agent of AMETEK Inc.
ADDITIONAL WARRANTY STATEMENT
Please note that the computer that controls the Detective-EX/DX-100T is
intended exclusively for the tasks detailed in this manual. Using this
computer for any other purpose could void your warranty.
If you have any questions about the use or maintenance of this instrument,
please contact your ORTEC representative or our Global Service Center
first.
IMPORTANT
The Detective-EX/DX-100T is designed to function with the Regional
Settings in Microsoft® Windows® Mobile™ set to the English (United
States) region and all of its default settings, as discussed in Section 7.12.
Using other than the factory default settings can cause operational
problems.
iii
[Intentionally blank]
iv
TABLE OF CONTENTS
WARRANTY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii
Cleaning Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ix
1. INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1. Monitoring Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2. Hardware Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3. How This Manual Is Organized . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.4. Spectrum and Search File Storage, Transfer, and Viewing . . . . . . . . . . . . . . . . . . . . . . 5
1.4.1. Saving Spectra and Searches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.4.2. File Formats and Filenaming Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.4.2.1. Identification Mode Spectra . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.4.2.2. Search Mode “Strip Charts” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.4.2.3. The Background File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.5. A Note About the DX and DX-100 User Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.6. The “Launcher” Interface for Multiple Identifier Applications . . . . . . . . . . . . . . . . . . . 8
2. THE DETECTIVE-EX/DX-100T . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1. Unpacking and Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2. The Detective-EX/DX-100T . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.2.1. Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.2.2. Rear-Panel Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.2.3. The Touchscreen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.2.4. Attaching the Carrying Strap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.3. The Docking Station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.3.1. Connectors and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.3.2. Strapping the Detective-EX to the Docking Station . . . . . . . . . . . . . . . . . . . . . 15
2.4. The PAC-II Charger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.5. The Automobile Accessory Adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.6. Internal and External Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.6.1. A Note on Internal vs External Battery Power . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.6.2. Replacing the Internal Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.7. Operating Cautions and Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.8. Startup and Cooldown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2.8.1. Connecting the Detective-EX to the Charger Base and Powering It On . . . . . . 20
2.8.2. Turning On the Mechanical Cooler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
v
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
3. RADIATION MONITORING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.1. Survey (Count-Rate/Dose-Rate) Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.2. Identification Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.2.1. Start the Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
3.2.2. Extending the Count-to-Preset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.2.3. Save . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.2.4. View Found Nuclides List for Classify ID Display Mode . . . . . . . . . . . . . . . . 33
3.2.5. Intense (View Highest-Intensity Spectrum Lines) . . . . . . . . . . . . . . . . . . . . . . . 34
3.2.6. Display Spectrum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
3.2.6.1. Display Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
3.3. Search Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
3.4. Monitor Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
3.5. SNM Search Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5. ADVANCED SETUP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.1. Advanced Setup — Page 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.1.1. Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.1.2. Turn Cooler Off/On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5.1.3. Calibrate (Manual) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5.1.3.1. Start the Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5.1.3.2. Apply the Improved Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
5.1.3.3. Perform a Background Measurement . . . . . . . . . . . . . . . . . . . . . . . . . 51
5.1.3.4. Restore Default . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
5.1.3.5. Back . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
5.1.4. Background . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
5.1.5. List Nuclides (View Analysis Library) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
5.1.6. Recall Spectra . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
5.1.6.1. Viewing Stored Spectra . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
5.1.6.2. Viewing Stored Spectrum Reports . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
vi
931040P / 1014 TABLE OF CONTENTS
5.1.7. About . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
5.2. Advanced Setup — Page 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
5.2.1. Calibration Settings and Auto-Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
5.2.1.1. Setup Requirements for Successful Auto-Calibration . . . . . . . . . . . . 58
5.2.1.2. Auto-Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.2.2. Background Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.2.3. Identify (Identification Mode) Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
5.2.3.1. Real Time Preset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
5.2.3.2. ID Display Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
5.2.3.3. LCX (Low-Confidence Expert) ID Mode . . . . . . . . . . . . . . . . . . . . . . 60
5.2.3.4. Color Coded IDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
5.2.4. Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
5.2.4.1. Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
5.2.4.2. Allow Display of Spectrum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
5.2.4.3. GPS Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
5.2.5. Search Mode Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
5.2.6. SNM Mode Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
5.2.7. Dose-Rate Units and Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
5.2.8. Audio Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
5.3. Advanced Setup — Page 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
5.3.1. Remote . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
5.3.2. Exit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
5.3.3. Spectra . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
5.3.3.1. Ask for Sample Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
5.3.3.2. Default Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
5.3.3.3. Data Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
5.3.3.4. File Save Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
5.3.3.5. Delete Stored Spectra . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
7. TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
7.1. “Error Saving .SPC File” Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
7.2. If the DetectiveEX Program Stops . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
7.3. If You Drop the Detective-EX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
7.4. If the Detective-EX Will Not Turn On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
7.5. If the Docking Station Spontaneously Shuts Down and Restarts . . . . . . . . . . . . . . . . . 76
7.6. If the Display is Lost or Posts a “Display Disabled” Message . . . . . . . . . . . . . . . . . . . 76
vii
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
9. SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
9.1. Electrical and Mechanical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
9.2. Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
9.3. Docking Station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
9.4. PAC-II Power Adapter/Charger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
9.5. Computer Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
9.6. Shipping the Detective-EX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
9.7. Long-Term Shutdown/Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
INDEX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
viii
Safety Instructions and Symbols
This manual contains up to three levels of safety instructions that must be observed in order to avoid
personal injury and/or damage to equipment or other property. These are:
DANGER Indicates a hazard that could result in death or serious bodily harm if the safety instruction
is not observed.
WARNING Indicates a hazard that could result in bodily harm if the safety instruction is not observed.
CAUTION Indicates a hazard that could result in property damage if the safety instruction is not
observed.
Please read all safety instructions carefully and make sure you understand them fully before attempting to
use this product.
Cleaning Instructions
To clean the instrument exterior:
! Disconnect the instrument from the power source.
! Remove loose dust on the outside of the instrument with a lint-free cloth.
! Remove remaining dirt with a lint-free cloth dampened in a general-purpose detergent and water
solution. Do not use abrasive cleaners.
CAUTION To prevent moisture inside of the instrument during external cleaning, use only
enough liquid to dampen the cloth or applicator.
! Allow the instrument to dry completely before reconnecting it to the power source.
ix
x
1. INTRODUCTION
The ORTEC Detective-EX®-100T portable nuclide identifier uses both a high-purity germanium
(HPGe) detector for identification using gamma-ray emissions and a moderated neutron detector
for detection of the neutron emissions. The Detective-DX-100T is identical to the -EX-100T but
has no neutron detector.1 The HPGe detector is cooled by a miniature, high-reliability mechani-
cal cooler — no liquid nitrogen is required. The user interface is a high-resolution color touch-
screen. The result is a lightweight and rugged instrument that delivers unmatched identification
and detection for the needs of nuclear security.
The Detective-EX identifies and classifies radionuclides according to the requirements of ANSI
N42.34 and other standards. The radionuclide library for this release of the Detective-EX is
detailed in Chapter 4.2
The Detective-EX can make an accurate and fast determination without the need for expert
human intervention. Nuclide identifications are based on the gamma-ray peaks in the spectrum
and on other aspects of the total spectrum.
! Survey Mode — In this “home screen” mode, the Detective-EX displays gamma count-
rate, gamma dose-rate meter (and neutron dose-rate meter, where applicable), a battery-
life indicator, and an icon indicating the current power source (e.g., internal battery,
power adapter/charger), detector dead (busy) time, and available disk storage.
1
In this manual, the term Detective-EX refers to all members of this instrument family, exceptions noted. See also
Section 1.5.
2
Other lists of nuclides can be supplied on special order.
3
Measurements with a portable identifiers cannot take into account the counting geometry and source shielding,
so the Detective-EX-100T does not determine the quantity of the detected nuclides.
1
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
! Monitor Mode — Monitor Mode is used to identify radionuclides while moving relative
to object(s) of interest. In this mode, the Detective-EX collects and analyzes data in rol-
ling 8-second windows, making it particularly suited to situations where a completely
different set of objects may be presented to the instrument in that time period. An exam-
ple application is an ad hoc portal monitor where packages or people pass by the instru-
ment. This mode is also useful as a search tool where the instrument is carried by the user
in an effort to locate and identify a radioactive source. IDs persist until you tap to
acknowledge them; this ensures none go unnoticed.
! Search Mode — Search Mode makes it easy to find the areas of highest radionuclide
activity, that is, places where Identification Mode should be used. The display shows the
total gamma-ray count rate and total neutron count rate in each time period of the last 221
time periods in a rolling “strip chart” display. The time period is selectable. The neutron
count rates and gamma dose rate are also displayed. These strip-chart data can be saved
on disk in the ORTEC .CHN file format.
! SNM Search Mode4 — This mode lets you quickly sense the possible presence of
special nuclear materials 235U and 239Pu, even if other radioactive materials are present. It
4
Patent pending.
2
931040P / 1014 1. INTRODUCTION
is designed for rapid area searches as well as fast scanning of containers or vehicles. Use
it to locate possible SNM, then test more rigorously using Identification Mode. SNM
Search Mode also takes advantage of the fact that the comparatively innocuous 133Ba can
be used as a plutonium surrogate. Advanced users can choose to enable a “search for
133
Ba” feature within SNM Search Mode both for training purposes and for developing a
site’s concept of operations.
Before using SNM Search Mode in the field, you must develop your concept of opera-
tions for most effective use. The Detective-EX allows you to adjust the length of the
SNM data sampling (“dwell”) time. ORTEC specialists are available to help you optimize
the instrument settings; contact your ORTEC representative or our Global Service
Center.
In all modes, when the gamma dose rate exceeds 10,000 µSv/hr (1 rem/hr) or the neutron count
rate exceeds 500 cps, both audible and visual alarms warn you immediately of the high radiation
field. You can optionally set the alarm threshold(s) to lower settings.
Search and identification data can be viewed on the display for further analysis. For work in
sensitive areas, the spectrum display can be disabled.
! The unit comes with either a docking station or the more compact PAC-II power adapter/
charger (collectively referred to as charger bases). Both types of charger base contain the
charger for the internal battery. The docking station’s fans both cool the charger unit and
dissipate heat from the Detective-EX when the unit is docked. The docking station can
also hold a weak (~37 kBq) 137Cs source used to verify the energy calibration.
The Detective-EX can operate for about 3 hours on its internal lithium-ion battery. The
car accessory connector (included) allows you to use 12-V power from an automobile
battery. In addition, ORTEC offers optional battery belts to extend the remote operation
time.
3
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
! Checking the Detective-EX’s calibration is easy, whether you have a docking station
with its built-in 137Cs source; a PAC-II charger, which does not include a calibration
source; or are working in the field under battery power. The Detective-EX allows you to
choose the calibration source that best suits your needs — even the 40K naturally present
in concrete or low-sodium salt substitute.
! The Detective-EX includes global positioning (GPS) hardware, and where GPS signals
can be detected, the upper right corner of the screen includes the current latitude and
longitude, as shown in Fig. 14 on page 22. The GPS coordinates are captured and stored
at the start of each identification acquisition, saved as part of .SPC or .N42 spectrum files,
and available as one of reports discussed in Section 5.1.6.2.
Before using the Detective-EX, it must be set up and cooled; and the calibration should be
checked according to the instructions in Chapter 2.8. The instrument is shipped fully calibrated
and is likely to still be correctly calibrated, but checking the calibration is considered good
practice.
“Operator-oriented” chapters:
! Chapter 1 — System features, spectrum storage, file formats, and filename conventions.
! Chapter 2 — Unpacking, startup, operating limits, initial configuration, and user-
interface overview.
! Chapter 3 — Using the Detective-EX to monitor and identify radiation sources.
! Chapter 4 — Radionuclide library.
“Administrator-oriented” chapters:
! Chapter 5 — Administrator-level setup features.
! Chapter 6 — Transferring survey files to a computer.
! Chapter 7 — Troubleshooting and error messages.
! Chapter 8 — Using the Detective-EX with a computer and ORTEC spectroscopy
software.
4
931040P / 1014 1. INTRODUCTION
! Chapter 9 — Specifications.
! Chapter 10 — Changing the internal battery.
! Chapter 11 — Calibration source statement.
The files can be stored on user-supplied, removable Secure-Digital (SD) memory cards,5 or in
the computer’s My Documents folder. The number of spectra stored on an SD card depends on
the card’s capacity. For SD cards, the DetectiveEX program automatically creates a Data folder
in which spectrum and search files will be stored. If the card fills to capacity, the Detective will
automatically begin storing spectra in its computer’s My Documents folder.
The background.chn file is stored on the computer in the \Program Files\DetectiveEX folder. If
it is moved or deleted, you must perform a new background measurement before using the
instrument for radionuclide identification.
If you store the spectrum files on SD cards, they can be transferred directly to your computer
with an SD card reader and Windows Explorer. Alternatively, files stored on SD card or in the
computer’s My Documents folder can be downloaded via the USB port to any computer running
under Windows 8 or 7 via the Windows Mobile Device Center, or Windows XP Professional
SP3 via Active-Sync®. See Chapter 6 for detailed instructions.
IMPORTANT Before connecting the Detective-EX to a computer for the first time, install
ActiveSync or Windows Mobile Device Center first. Otherwise, the Detec-
tive-EX may not be able to communicate properly with the computer. Be sure
to read Chapter 6.
The accompanying MAESTRO® MCA Emulator Software (A65-BW) can be used to view the
search and identification spectrum files from the Detective-EX. MAESTRO can also be used to
operate the Detective-EX as a standard multichannel buffer (MCB) and store the resulting spec-
tra to disk as ORTEC spectrum files. For more information on these operations, see Chapter 8.
5
Note that the computer also has a CF type II CompactFlash slot. However, the dust cover — which must be in
place to prevent contaminating the card slots — blocks full insertion of the CF card.
5
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
To save the current gamma-ray spectrum or search strip chart in stored-spectrum memory, tap
the Save button.
The Spectra dialog allows you to select the sample description options, and file formats and
storage location; and view the contents of the stored-spectrum memory. See also Section 5.3.3.
The .SPC file contains three parameters in addition to the contents specified in the ORTEC
Software File Structure Manual for DOS and Windows Systems (Part No. 753800; hereinafter
called the File Structure Manual):
! The neutron count rate at the time the spectrum file is saved (zero for Detective-
DX/DX-100).
! A “normal” or “100" entry that indicates whether the spectrum was collected using a
standard Detective-EX/DX or a Detective-EX/DX-100.
YYYY_MM_DD_hh_mm_sss
6
931040P / 1014 1. INTRODUCTION
The .CHN format is simpler than the .SPC format, and is described in the File Structure Manual.
The .N42 file includes the sample description and a count rate log for the gamma strip chart; the
count rate log for the neutron strip chart (if the unit is so equipped); and GPS location log (if so
equipped).
YYYY_MM_DD_HH_MM_SSStype
When saving in the ORTEC .CHN format, two files are saved for each search, one for the gamma
count and one for the neutron count; type is either G for the gamma-ray chart or N for the
neutron chart.
When saving in ANSI .N42 format, only one file is generated; type is B indicating that it
contains both the gamma-ray chart and (for units so equipped) the neutron chart.
7
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
IMPORTANT If you use the trans-SPEC program to change any of the hardware settings,
restarting the Detective program will return the identifier to all Detective
factory calibration, gain, and rise-time settings, ensuring the instrument will
operate as described in the rest of this manual.
6
If, on reboot, the Microsoft® Windows® Mobile desktop is displayed instead of the Launcher or a spectroscopy
application, tap Start then Launcher. Alternatively, you can bypass the Launcher and start either program by
tapping Start, Programs, and the application name.
8
2. THE DETECTIVE-EX/DX-100T
This chapter covers the steps you will need to prepare the Detective-EX for use:
Note also that where the spectrum should not be viewed by the operator, the spectrum display
can be disabled (see Section 5.2.4.2).
Any options will be packaged separately, and will include instructions for that option.
7
Subject to change without notice.
8
The identifier is factory pre-calibrated using a multi-nuclide source. The calibration can be updated using a only
small 137Cs source. The instrument is usually supplied with an exempt-quantity 137Cs source, with a nominal
activity of 1.0 µCi (37 kBq), built into the docking station source tower. If your unit was not supplied with a 137Cs
source, you can use naturally occurring 40K or else define other calibration sources, as described in Section 5.2.1.
9
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
! SD Card Slot This is located on the front panel under the protective dust cover. To
release the cover, press and rotate by a quarter turn the two captured, bayonet-mount
screws. The SD slot is keyed so the memory card can only be inserted in one orientation.
Insert the card and close the protective cover.
10
931040P / 1014 2. THE DETECTIVE-EX/DX-100
NOTE We strongly recommend that the dust cover be fastened in place at all times to protect
the card slots from moisture and particulates.
Figures 4 and 5 show the PAC-II charger, and the docking station, respectively. The hardware
specifications are in Chapter 9. Place the Detective-EX on the docking station or beside the
PAC-II.
11
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
! Headphone jack ( ) — For dose and count rate warning sounds. The headphone volume
can be adjusted as described in Section 5.2.8.
NOTE We strongly recommend that the dust cover be fastened in place at all times to protect
the card slots from moisture and particulates.
To select a function, tap it with your finger or a stylus (pens, pencils, and other objects will
scratch the touch-screen and should not be used). In fields where alphanumeric input is needed,
tap to open the “soft keyboard” at the bottom of the screen (see Fig. 56 on page 57).
12
931040P / 1014 2. THE DETECTIVE-EX/DX-100
13
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
Place the docking station on a hard, level surface. The underside of the docking station should
be kept clear so air can circulate to help dissipate the heat from the cooler. For more information
on proper placement and treatment of the docking station, see the setup instructions in
Section 2.8.1 and the cautions in Section 2.7.
! Integral front-panel external power and battery charger cable that attaches to the Detec-
tive-EX INPUT POWER connector.
! Front-panel red and green CHARGE STATUS indicators. While the Detective-EX’s
internal battery is charging, the red indicator is continuously lit. When the battery is fully
charged, the green indicator turns on and the red indicator turns off.
14
931040P / 1014 2. THE DETECTIVE-EX/DX-100
Place the PAC-II on a hard, level surface where air can adequately circulate around it. For more
information on proper placement and treatment of the docking station or PAC-II, see the setup
instructions in Section 2.8.1 and the cautions in Section 2.7.
To use the PAC-II, attach its OUTPUT cable to the Detective-EX INPUT POWER connector,
then connect either the ac or dc power cord (both supplied) to the appropriate rear-panel connec-
tor. While the Detective-EX’s internal battery is charging, the red LED on the PAC-II front
panel is continuously lit. The red indicator flashes if you disconnect the Detector before its bat-
tery is fully charged, or if the Detective-EX is connected but the internal battery is faulty. When
the battery is fully charged, the green LED turns on and the red LED turns off.
15
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
NOTE If the detector is warm, the power required to cool the detector can significantly
discharge a standard automobile battery.
! Connect the cable to the appropriate connector. If using the adapter cable that connects to
the docking station or PAC-II, plug it into the charger base’s POWER IN connector. The
internal battery will begin charging; if completely discharged, it will typically reach full
charge in 3–4 hours.
If using the adapter cable that connects directly to the Detective-EX, unscrew the dust
cap from the INPUT POWER connector on the rear panel; slide the coupling ring on the
male (cable-end) connector fully open; align the 5 keys on the male connector with the
slots on the rear-panel female connector; press to fully seat the male connector, then hold
it in place; and rotate the coupling ring approximately 1/8th of a turn clockwise until it
seats securely.
! The internal battery provides approximately 3 hours of operation. The internal battery
cannot be used to start the cooler or cool the detector from room-temperature; external
power is required. The transition from internal to external power is automatic and does
not interfere with data collection. Cooler startup without external power cause opera-
tional anomalies; these are discussed in troubleshooting Sections 7.5 and 7.6. An
exhausted internal battery should be charged for 2 hours before the Detective-EX is
cooled using external power.
16
931040P / 1014 2. THE DETECTIVE-EX/DX-100
! The docking station or PAC-II provides external power to simultaneously start and
operate the Detective-EX, as well as charge the internal battery. In addition, the docking
station fans help dissipate heat from the cooler. Both charger bases can be powered from
mains or a 12 V source.
! Other external power (supplied by external battery belt or 12 V automobile battery) can
start, cool, and operate the Detective-EX, but does not charge the internal battery. The
internal battery is charged only if the external source also powers the docking station or
PAC-II. If extended remote operate is needed, use the external battery first, then
the internal battery.
The icon on the right side of the screen above the buttons
(Fig. 12) shows the current type of power connection. These
are, clockwise from upper left: (1) internal battery, (2) unit
properly connected to the docking station or power adapter/
charger, (3) external power to operate unit, and (4) battery
charging.
Figure 12. Power Source Icons.
2.6.1. A Note on Internal vs External Battery Power
The recharging circuit for the internal battery is separate from external power so the external bat-
teries are not consumed charging the internal battery. The internal battery is only recharged
through the use of a special lithium-ion battery charger unit built into the docking station and
PAC-II using connections in the multi-pin power plug. The charging base will simultaneously
supply both “external power” and connect to the battery charger to recharge the internal battery.
When connected to an external battery, the Detective-EX relies on external power until the
external voltage drops below 9 V. It then switches to the internal battery. The instrument makes
the transition smoothly, and there is no effect on the energy resolution. If you begin with both
batteries fully charged, the sum of the two battery capacities will be achieved. If the external
battery has its own charge indicator, this can be used to determine when the changeover will
occur. The Detective-EX battery monitor indicates the available capacity in the internal battery,
and when the switch over occurs, this indicator runs down in the usual manner.
NOTE When all power is exhausted, the high voltage and cooler will shut down and the
detector will begin to warm. The Detective-EX cannot be used again until the unit has
cooled enough that the high voltage automatically switches on and the Status
indicator says READY.
17
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
! The Detective-EX is splash-resistant, but not watertight. To avoid equipment failure, pro-
tect it from prolonged or heavy rainfall and from immersion. Heat is dissipated from the
cooler via the Detective-EX baseplate; therefore, do not place the instrument in mud,
loose soil or sand, standing water, dense grass; or on thick carpet. Wrap or cover the unit
to keep it clean and dry. Where possible, avoid placing it on warm surfaces, such as sun-
heated pavement, for extended periods. The front-panel memory card slots and rear panel
connectors include dust caps; keep the card slots and all unused connectors capped.
! The docking station and PAC-II contain a portion of the battery-recharging system and
should not be exposed to water. When using the docking station, be sure to keep the
underside clear of obstructions so the fans can function properly. Both charger bases
should ideally be placed on a dry, flat, level, and dust-free surface. When not in use, the
charger base can be disconnected from the ac power supply, then covered to protect it
from moisture and dirt. Any covering must be completely removed before reconnecting
the docking station to either charger base to the input power source. Make sure the
recharging connector on both the Detective-EX and charger base are clean and dry before
connecting to an the input power supply.
! The Detective-EX is designed for use at temperatures between 0 and +40°C and at a
relative humidity of <90% at 35°C, non-condensing.
18
931040P / 1014 2. THE DETECTIVE-EX/DX-100
! To avoid damaging the cooler (and, where applicable, to comply with transportation
regulations), be sure to turn the cooler off before shipping the Detective-EX (see
Section 2.8.2). The unit can be shipped as soon as the cooler is turned off; there is no
need to wait until the detector warms up.
! For best results, we recommend that you not allow the instrument’s internal battery to
become exhausted, and that you read Section 2.6 on the Detective-EX’s internal and
external power sources.
! When monitoring very high-activity sources (whether the source is an innocent or threat
nuclide), if the detector Dead Time exceeds 70%, the detector cannot process gamma
radiation and generate IDs according to its performance specifications. A red indicator
in the upper right corner of the screen tells you to MOVE BACK from the source until the
dead time drops below 70% so the instrument can perform as specified. See Fig. 15
(page 22) and its associated discussion.
! When the most recent background reading is no longer valid, the Background Invalid
message is displayed on the Survey Mode screen. The Identify and Monitor buttons
become inactive and the system will not post radionuclide IDs until a new background is
accepted (see Section 2.12). However, the gamma and neutron count-rate meters on the
Survey, Search, and SNM Mode screens continue to function normally, so the unit is still
passively monitors.
! Connect it to the power adapter/charger or other external power source. This will boot up
the computer, start the cooler, and start the DetectiveEX software application at the
Survey Mode screen.
! Wait for the detector to cool down to operating temperature. Typically, this takes
<12 hours for the Detective-EX.
! Depending on your power source options, charge the internal battery for 3–4 hours.
19
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
This section will help you prepare the Detective-EX for use as quickly as possible. Therefore,
some of the steps are not explained in detail; however, references to additional information are
provided.
! Place the charger base on a hard, level surface. If using the docking station, its underside
must be kept clear so air can circulate to help dissipate heat from the cooler. See the
cautions in Section 2.7 (page 18).
! AC — Connect the charger base’s output cable to the front-panel POWER IN connector,
then connect the mains power adapter cord to mains power.
or
DC — Connect the dc power cord to the front-panel POWER IN connector, then to a
suitable dc power source.
! Place the Detective-EX on the docking station or beside the PAC-II, and attach the
charger base’s output cable to the Detective-EX INPUT POWER connector. To attach
the cable to the Detective-EX, unscrew the dust cap from the INPUT POWER connector
on the Detective-EX’s rear panel; slide the coupling ring on the male (cable-end)
connector fully open; align the 5 keys on the male connector with the slots on the rear-
panel female connector; press to fully seat the male connector, then hold it in place; and
rotate the coupling ring approximately 1/8th of a turn clockwise until it seats securely.
The internal battery will begin charging and will typically reach full charge in 3–4 hours
(well before the detector has cooled to operating temperature range).
! As soon as power is applied, the computer will automatically begin its boot-up sequence,
and within 20–40 seconds the cooler will start running. After the computer initializes, the
DetectiveEX software application will start up in Survey Mode. As noted above, the
internal battery will not charge until the detector is cooled to operating temperature. After
that, the battery will typically reach full charge in 3–4 hours. If the cooler does not start
by itself, see the next section.
! Note the Status indicator on the display. On startup (Fig. 13), it will read:
20
931040P / 1014 2. THE DETECTIVE-EX/DX-100
The Detective-EX’s internal battery does not have enough power to cool down the detector
from room temperature, so external power is required for cooldown. If you are using a standard
automobile battery as the external power source, note that detector cooling can significantly
discharge the battery.
When the detector reaches operating temperature, the Detective-EX bias voltage will automati-
cally turn on. To see if the bias voltage status is On or Off, tap on Advanced, then Status. Until
the detector has cooled and the HV has automatically turned on, Detector Temp, Cold Tip
Temp, and HV Volt should read ERR, then switch to OK. At this point, the status indicator on
the Survey Mode screen will show Ready.
NOTE You can turn the cooler off for a brief period (e.g., 10 to 20 minutes) without signifi-
cantly warming the detector. As soon as the cooler is restarted, the Detective-EX
typically shows a Ready status within a few seconds. In addition, the cooler can be
restarted at any time, without harming the detector crystal, regardless of whether the
detector is warm, partially cooled, or cooled to the operating temperature range.
21
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
! When close to “hot” gamma-ray sources, the HPGe detector can become so busy proces-
sing radiation pulses that its analysis algorithm does not function as designed and speci-
fied. To avoid this problem, when the detector dead (busy) time >70%, a red warning
message is displayed in the upper right corner of the screen (Fig. 15), and persists until
you move away from the hot source.
! Instrument Status Area — The center of the Survey Mode screen provides brief indi-
cator’s of the unit’s readiness for data acquisition. The Status is either READY (Fig. 23) or
Detector is warm. In addition, warnings such as the high gamma dose rate message are
posted in this part of the screen.
22
931040P / 1014 2. THE DETECTIVE-EX/DX-100
! To return to the preceding dialog, tap Back. For example, tapping Back from the
Advanced Setup dialog (Fig. 17 on page 24) returns you to the Survey Mode screen
(Fig. 23).
! On any entry, if you just tap Back without tapping any other keys, the current value will
be used; that is, no change will be made. Therefore, if you open a dialog by mistake, you
can just tap Back to return to the previous screen without changing any values.
! If a text or number entry is needed, a soft keyboard will be displayed at the bottom of the
screen. Tap on the letters or numbers desired and then tap OK.
! The Identification, Search, and Display screens have a Save button. In Identification
Mode, this saves the current energy histogram (spectrum) in memory and identification
results. In Search Mode, a histogram of counts vs time is saved. The storage location is
selected on the Spectra screen (Section 5.3.3.3). The Spectra screen also includes an Ask
for Sample Description option that allows you to enter a description of this data
acquisition.
NOTE Before using the Calibrate feature, the calibration source must be selected on the Cal.
Settings dialog under the Advanced Settings. The Detective-EX allows you to select
from two predefined calibration sources, 137Cs or 40K; or define two of your own
sources. See Sections 5.2.1 and 5.1.3.
23
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
If the docking station is not available or if you are using the PAC-II, a small calibration source
can be used in conjunction with the manual calibration check to ensure the unit remains in
calibration.
! On the Survey Mode screen, tap on Advanced (see Fig. 17), then Calibrate.
! Position the indicated source (in this case, 60Co) in front of the detector window, then tap
the Start Calibration Now (Fig. 18) button.
! While calibration is in progress (Fig. 19), the unit will show the Time remaining. This
interval will depend on the strength of your calibration source.
! At the end of the calibration period, the Use Improved Calibration button will be
displayed, along with the calibration and peak-shape indicators for the new measurement
(Fig. 20).
Figure 17. On the Advanced Figure 18. Tap the “Start Calibra-
Menu, Tap the Calibrate Button. tion Now” Button.
The choice is yours whether to use this new calibration or not. To accept it, tap Use
Improved Calibration. Usually, the proposed change in calibration is of the order of the
one shown in Fig. 20. If it is larger than 0.3%, the cause should be investigated (see
Section 7.8).
24
931040P / 1014 2. THE DETECTIVE-EX/DX-100
25
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
NOTE This section discusses only the mandatory background measurement required to
complete the manual calibration. For the full discussion of Detective-EX background
measurements, see Section 5.2.2. As noted elsewhere, until you complete the
background measurement, the Identify and Monitor Modes will be disabled.
The minimum acceptable data collection time for the background meas is set on the Bkg
Settings screen (see Fig. 61 on page 60). Note that, for optimum counting statistics, we
recommend a minimum 30-minute background acquisition. However, the background settings
screen also allows minimum intervals of 1, 5, or 15 minutes.
! All sources should be removed from the area before you tap the Run a Background but-
ton at the end of the manual calibration routine. If you leave the source in place, a “high
count rate” message will remind you to remove the source and restart the calibration.
! Tapping the Run a Background button opens the screen shown in Fig. 22 and automati-
cally starts the background acquisition.
26
931040P / 1014 2. THE DETECTIVE-EX/DX-100
NOTE Passwords are case-sensitive and can include “space” characters. Note that there is no
master password so be sure to record it. If you lose the password, you must contact
your ORTEC representative or our Global Service Center for assistance.
27
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
[Intentionally blank]
28
3. RADIATION MONITORING
Once the Detective-EX displays the Ready status, the calibration has been checked, and the
background has been measured, the instrument is ready for use. This chapter tells you how to
use the Survey, Identification, Search, Monitor, and SNM Modes.
Some users will find enough information in this chapter to begin using the
Detective-EX right away — operation is very straightforward.
! Chapter 4 lists the messages the Detective-EX displays when it identifies radionuclides.
IMPORTANT You can set a password to limit access to the second and third pages of
the Advanced menu; see Section 5.2.4.1 for instructions.
The three rate indicators show the count rate or dose as both a value and a bar graph. The
color of the bar changes with increasing rate. The lowest rate is green, then yellow, then red.
The γ (gamma) Count Rate bar changes to yellow at~6300 counts per second (cps), and red
at ~13300 cps. At ~20,000 cps, the red bar begins flashing. These various states are illustrated in
29
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
Fig. 24. The factory-set alarm limits are 10000 µSv/h for the Gamma Dose Rate and 500 cps for
the Neutron Count Rate bar. To set lower limits, see Section 5.2.7.
When the gamma dose rate goes above the limit, the WARNING!!! High Gamma Dose Rate!
message is displayed in the Instrument Status Area and an alarm sounds. You can optionally
choose to set lower gamma dose rate and neutron count rate alarms using the Dose Alarm setup
feature under the Advanced Settings screen (Section 5.2.7).
NOTE If a blue simulated LED ( ) is displayed in the upper left, beside the screen title, the
instrument is in the LCX radiation detection mode, which we strongly recommend for
use only by experienced gamma spectroscopists. Be sure to read Section 5.2.3.3 and/or
refer to your organization’s operating procedures before using LCX Mode.
30
931040P / 1014 3. RADIATION MONITORING
! “ANSI” Mode — This is the default mode. Radionuclide IDs are listed on a background
color-coded according to their status as innocent (green), suspect (yellow; LCX Mode
only), or threat (red). The color background can be suppressed (Section 5.2.3.4).
! “Classify” Mode — In this mode, the list of detected radionuclides are listed according
to category instead of radionuclide name, e.g., Medical, Industrial, Nuclear Plutonium,
Other.
Chapter 4 lists all nuclide IDs along with “ANSI” their threat/suspect/innocent status and
“Classify” category.
Figures 25 and 26 compare the same four IDs as displayed in the “ANSI” and “Classify” modes,
respectively. In each figure, the Detective-EX has found one industrial nuclide (unshielded
241
Am; innocent), and two uranium isotopes (235U and 238U; threats).
! Display the real-time spectrum for the current data acquisition (Section 3.2.6). Note that
if the spectrum view has been disabled on the Settings dialog (Section 5.2.4.2), the
Display button is gray (inactive).
9
The expert LCX Mode (Section 5.2.3.3) also displays suspected IDs, i.e., nuclides that have been tentatively
identified at too low a confidence level to list them as definitively “found.”
31
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
! Switch to Search mode (Section 3.3). If you do not Save the current spectrum and
identifi-cation results data before going to Search Mode, these data will be lost.
NOTE When you start an identification or switch to Search Mode, the contents of the live-
spectrum memory are erased before the new data are collected. If you wish to save the
data from the most recent identification or search session, you must do so before
starting the new identification session (tap Save).
During data acquisition, a simulated green LED flashes at the top left of the screen. On units
with GPS capability, the top right of the screen alternates between the elapsed acquisition time
for the current spectrum and the GPS coordinates.10 In addition, the neutron count rate and the
gamma dose rate continually update.
10
If your GPS has not received signals from GPS satellites for >30 days (e.g., if it has not been in use or has been
used continuously indoors), it may take 10–15 minutes for the unit to update and begin displaying location
information.
32
931040P / 1014 3. RADIATION MONITORING
At any time you can Pause the count and either Save the spectrum, return to the Survey Mode
screen, or Continue counting.
NOTE Tapping Pause or Continue does not clear the spectrum. Pausing simply suspends data
collection. When you continue, subsequent data are added to the existing spectrum
until you tap Back. to return to the Survey Mode screen. The spectrum is retained in
memory until you start the next identification or search acquisition. If you have not
already viewed and/or saved the spectrum, you can do so from the Survey Mode screen.
3.2.3. Save
Tap Save to write the current identification spectrum and results in the selected data storage
location and file formats (Sections 5.3.3.3 and 5.3.3.4, respectively). You may save as many
times as you wish, during acquisition or pauses, and after acquisition.
The results are saved in the spectrum file(s) and can be viewed both on the Detective-EX
(Section 5.1.6) and with ORTEC spectroscopy applications such as MAESTRO. An identifying
text description can also be entered if the Ask for Sample Description feature is enabled on the
Spectra dialog (Fig. 74, page 66).
To end without creating a spectrum file, tap Back. This immediately returns you to the Survey
Mode. The current spectrum data is retained in memory until the next identification or search.
You can decide whether or not to save it from the Survey Mode screen.
33
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
LCX Mode). To view the found nuclides, simply tap on any of the ID messages on the ID Mode
screen. The first tap will open the Found Nuclides report (see Fig. 28). Each ID will have a red,
yellow, or green background corresponding to its threat level as described in Chapter 4. To
access the Suspect Nuclides report in LCX Mode, tap an ID message on the Found screen. The
report screens let you Save the current spectrum and identification results to disk and Display
the current spectrum. Tap Back to return to the main Identification Mode screen.
Figure 28. View Found Nuclides Figure 29. View Top Lines.
List.
The energies are based on the internal nuclide library. The rank is based on the confidence value
for the peak. When a detected energy matches an entry in the library, its tentative identity is
posted on this screen, by peak energy, in column 1; and ID confidence factor (also called the Q
value), in column 2. The higher the confidence value, the more likely that the peak is a real
gamma ray peak and not background noise. It should be understood, however, that a peak could
be seen in the top lines list but the corresponding nuclide not reported as found. This is because
there are many other tests applied before an identity or tentative identity is chosen. In general,
any Confidence value over 25 indicates a good peak.
34
931040P / 1014 3. RADIATION MONITORING
Tap Save to save the current spectrum and identification results in the selected data storage
location and file format(s). Tap Display to view the current spectrum (see Section 3.2.6).
The spectrum display has a vertical marker line. Just below the spectrum is shown the position
of the marker in the spectrum (in keV) and the counts in the marker channel. To move the
marker line to a different location, tap the desired position. The marker will jump to that location
and the values will update correspondingly. You can also press on the display area and drag the
marker left or right to the new location.
! Full/Zoom — In the expanded view, the Full button restores the display to 8000 chan-
nels. In the full view, the Full button changes to Zoom and zooms the display to the last
horizontal expansion before Full. In this way, you can easily look at the peaks in the
spectrum in detail (Zoom in) and move quickly from energy to energy in Full mode.
35
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
! Zoom In <> — The <> button expands the display by halving the number of channels, so
the peaks appear wider. The center channel of the new display will be the marker
position.
! Zoom Out >< — The > < button contracts the display by doubling the number of chan-
nels, so the peaks appear narrower. The center channel of the new display will be the
marker position or channel 0 is the leftmost corner of the display or channel 8192 is the
rightmost corner of the display.
! Lin/Log — The Lin button changes the vertical display to linear mode and automatic
vertical limits. The Lin button changes to Log in the linear mode. Tap Log to change the
vertical scale to log mode.
! Pause/Continue — If the Detective-EX is acquiring data, you can pause the data
collection and resume it later. Tap Pause to stop collection. The button will change to
Continue in the pause mode. Tap Continue to resume the data collection.
! To start a search session, tap Search in the Survey Mode or any other dialog with the
Search button. If you wish to save the spectral data from the preceding identification or
search session, you must do so before starting the new search (tap Display, then Save).
36
931040P / 1014 3. RADIATION MONITORING
The search will continue until you stop it, but only the most current count rate values
(the newest 221) are displayed and can be saved.
! To save the displayed search results, tap Save. If the Ask for Sample Description option
is enabled on the Spectra setup screen (see Fig. 74, page 66), a soft keyboard will open so
you can enter a description for this data acquisition; tap Set to complete the save. If sav-
ing in ORTEC .CHN format, two count-rate “strip chart” files are saved for each search,
one for the gamma count (labeled with a G suffix) and one for the neutron count (with an
N suffix). For ANSI .N42 files, one file with both gamma and neutron count rate (with a
B suffix) is created. The filenames are automatically generated (Section 1.4.2) and saved
according to the spectrum settings (Section 5.3.3).
! To end this search and return to the Survey Mode (Fig. 23), tap Back. The current search
data will be retained until the next identification or search. (You can display the most
recently acquired search or identification spectrum from the Survey Mode screen.)
The Detective collects one spectrum per second, and analyzes a rolling window of the eight
most recent 1-second spectra for radioisotope IDs. This makes it particularly suited to situations
where a completely different set of objects may be presented to the instru-ment in that time
interval.
Each ID’s background is color-coded according to its threat status (see the ID table in
Chapter 4): red for threat nuclides; green for innocent; and, in LCX Mode only, yellow for
suspects.
37
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
SNM searches cannot be saved; to collect data that can be saved and reviewed later, tap
Identify.
Figures 34 through 37 depict a search in which 133Ba is detected. At the start of the search, the
status message says Continue to search (Fig. 34). This message is displayed until the detection
threshold for any one of the radionuclides or neutrons is crossed. As a 133Ba source is
approached, the confidence bar associated with it begins to lengthen and approach the threshold
(Fig. 35).
As the Detective-EX comes close to the 133Ba source, the identification for the radionuclide
exceeds the threshold. The confidence bar enters the shaded part of its box, and the shading
turns red to indicate the radionuclide has been found (Fig. 36). Figure 37 shows a high-prob-
ability hit. To verify the presence of SNM, tap Identify to switch to Identification Mode.
38
931040P / 1014 3. RADIATION MONITORING
133
Figure 34. SNM Search Under Figure 35. Ba Threshold Almost
Way. Exceeded.
133
Figure 36. Ba Tentatively Figure 37. High Confidence That
133
Detected. Ba Has Been Detected. Use
Identification Mode to Verify.
39
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
[Intentionally blank]
40
4. IDENTIFICATION MESSAGES AND
RADIONUCLIDE LIBRARY
This chapter provides the list of identifications for the v8.6 nuclide library, and their categoriza-
tion in the “Classify” and “ANSI” ID Display Modes (Section 5.2.3).
Identification Category
Classify Mode ANSI Mode
186 keV peak present Suspect Uranium
2614 keV peak present Suspect Uranium
375/414 peak present Suspect Plutonium
Ac-225 Innocent Medical
Ac-227 Innocent Industrial
Ag-110m Innocent Industrial
Am-241 Threat Industrial
Am-241 (shielded) Threat Plutonium
Am-241 (unshielded) Innocent Industrial
Ar-41 Innocent Industrial
As-72 Innocent Industrial
As-74 Innocent Medical
At-211 Innocent Other
Au-198 Innocent Industrial
Ba-133 Innocent Industrial
Ba-140 Innocent Industrial
Be-7 Innocent Industrial
Beta emitter Innocent Bremsstrahlung
Bi-207 Innocent Other
Bi-212 (Th-232/U-232 daughter) Innocent Industrial
Bi-214 (Ra-226 daughter) Innocent NORM
Br-76 Innocent Other
Br-76 (heavily shielded) Innocent Other
Br-76 (shielded) Innocent Other
Br-77 Innocent Industrial
Ca-47 Innocent Industrial
Cd-109 Innocent Other
Cd-115 Innocent Industrial
Ce-139 Innocent Medical
Ce-141 Innocent Medical
Ce-144 Innocent Industrial
Cm-242 Innocent Industrial
Cm-243 Innocent Industrial
Cm-244 Innocent Industrial
41
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
Identification Category
Classify Mode ANSI Mode
Co-55 Innocent Industrial
Co-56 Innocent Other
Co-56 (shielded) Innocent Other
Co-57 Innocent Industrial
Co-57 (shielded) Innocent Industrial
Co-58 Innocent Medical
Co-60 Innocent Industrial
Cr-51 Innocent Other
Cs-131 Innocent Medical
Cs-134 Innocent Industrial
Cs-137 Innocent Industrial
Cu-64 Innocent Other
Cu-67/Ga-67 Innocent Medical
Enriched Uranium Threat Uranium
Eu-152 Innocent Other
Eu-154 Innocent Other
Eu-155 Innocent Other
Eu-156 Innocent Other
Fe-59 Innocent Other
Ga-64 Innocent Other
Ga-64 (shielded) Innocent Other
Ga-67 Innocent Medical
Ga-67 (shielded) Innocent Medical
Gd-153 Innocent Other
Gd-159 Innocent Other
Ge-68/Ga-68 Innocent Medical
HEU Threat Uranium
Hf-181 Innocent Industrial
Hg-203 Innocent Industrial
Ho-166 Innocent Other
Ho-166m Innocent Industrial
Ho-166m (shielded) Innocent Industrial
I-123 Innocent Medical
I-123 (shielded) Innocent Medical
I-124 Innocent Medical
I-125 Innocent Medical
I-126 Innocent Industrial
I-126 (shielded) Innocent Industrial
I-131 Innocent Medical
42
931040P / 1014 4. IDENTIFICATION MESSAGES AND RADIONUCLIDE LIBRARY
Identification Category
Classify Mode ANSI Mode
I-131 (shielded) Innocent Medical
I-132 Innocent Industrial
I-133 Innocent Industrial
I-134 Innocent Industrial
I-135 Innocent Industrial
In-111 Innocent Medical
Ir-192 Innocent Industrial
Ir-192 (shielded) Innocent Industrial
Ir-194 (shielded) Innocent Other
K-40 Innocent NORM
Kr-87 Innocent Industrial
Kr-88 Innocent Industrial
Kr-88 (shielded) Innocent Industrial
La-138 Innocent NORM
La-140 Innocent Other
Lu-172 Innocent Medical
Lu-176 Innocent NORM
Lu-177 Innocent Medical
Lu-177m Innocent Medical
Mn-52 Innocent Industrial
Mn-54 Innocent Other
Mn-56 Innocent Industrial
Mo-99 Innocent Medical
Na-22 Innocent Other
Na-24 Innocent Medical
Nb-92m Innocent Industrial
Nb-94 Innocent Industrial
Nb-95 Innocent Industrial
Nb-96 Innocent Industrial
Nb-96 (shielded) Innocent Industrial
Nd-147 Innocent Industrial
Neutrons Threat Other
Neutrons CR Threat Neutron
Neutrons on Fe Threat Other
Neutrons on Hydrogen Threat Other
Np-237 Threat Nuclear Neptunium
Os-194/Ir-194 Innocent Other
Pa-231 Innocent Industrial
Pb-203 Innocent Industrial
43
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
Identification Category
Classify Mode ANSI Mode
Pd-103 Innocent Medical
Po-210 Innocent Other
Pr-144 Innocent Industrial
Pu-238 Threat Nuclear Plutonium
Pu-239 Threat Nuclear Plutonium
Ra-223 Innocent Industrial
Ra-226 Innocent NORM
Rb-83 Innocent Medical
Rb-86 Innocent Medical
Rh-105 Innocent Industrial
Ru-103 Innocent Industrial
Ru-106/Rh-106 Innocent Medical
Ru-97 Innocent Industrial
Sb-124 Innocent Industrial
Sb-124 (shielded) Innocent Industrial
Sb-125 Innocent Industrial
Sb-127 Innocent Industrial
Sc-46 Innocent Medical
Se-75 Innocent Medical
Sm-153 Innocent Medical
Sm-153 (shielded) Innocent Medical
Sn-113 Innocent Other
Sr-82/Rb-82 Innocent Medical
Sr-85/Kr-85 Innocent Industrial
Sr-89 Innocent Medical
Ta-182 Innocent Other
Tc-96 Innocent Industrial
Tc-99m Innocent Medical
Te-132 Innocent Industrial
Th-229 Innocent NORM-Th
Th-230 Innocent NORM-Th
Th-232 Innocent NORM-Th
Tl-200 Innocent Other
Tl-201 Innocent Medical
Tl-202 Innocent Other
Tl-204 Innocent Medical
Tm-170 Innocent Medical
Tm-171 Innocent Industrial
U-232 Threat Nuclear Uranium
44
931040P / 1014 4. IDENTIFICATION MESSAGES AND RADIONUCLIDE LIBRARY
Identification Category
Classify Mode ANSI Mode
U-233 Threat Nuclear Uranium
U-235 Threat Nuclear Uranium
U-238 Threat Nuclear Uranium
Unknown Peak Threat Other
Unknown/Beta emitter Threat Other
V-48 Innocent Medical
W-187 Innocent Industrial
W-188/Re-188 Innocent Industrial
Xe-127 Innocent Industrial
Xe-131m Innocent Other
Xe-133 Innocent Medical
Xe-133m Innocent Medical
Xe-135 Innocent Other
Y-88 Innocent Other
Y-91 Innocent Industrial
Yb-169 Innocent Medical
Zn-62 Innocent Medical
Zn-65 Innocent Medical
Zr-95 Innocent Medical
45
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
emitter is present (beta emitters typically producing counts over a broad range of energies).
Move closer to the source and count longer to determine the nature of the suspect item.
46
5. ADVANCED SETUP
The three-page Advanced Setup menu covers tasks
that are not typically required every work session and
that require some knowledge of gamma-ray spectro-
metry and/or radiation monitoring. The Detective saves
all settings and uses them until the next time they are
changed. Figure 38 shows the first page.
! The Dead Time is related to the current gamma count rate, and indicates the percentage
of time the Detector is busy (i.e., not available to accept another pulse).11
NOTE When the Dead Time exceeds 70% because you are near a very high-activity
source (whether the source is an innocent or threat nuclide), the detector is not
able to process gamma radiation according to its performance requirements.
A red indicator in the upper right corner of the screen tells you to MOVE
BACK from the source until the dead time drops below 70% so that the
11
This release of the Detective-EX software does not use the Live Time and Real Time fields.
47
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
instrument can perform as specified. See Fig. 15 (page 22) and its associated
text.
! The remaining indicators are live (i.e., they continuously update to reflect current hard-
ware conditions). For the first few seconds after you open this screen, the nine live indi-
cators typically read Waiting, after which they acquire their actual current values or status
messages (e.g., OK, Error, Waiting, OFF).
On startup and until the detector has cooled and the HV has automatically turned on,
Detector Temp, Cold Tip Temp, and Bias Voltage read Error. Once cooldown is com-
plete and the HV automatically turns on, these indicators change to OK. When the cooler
is switched off, the Cold Tip Temp and Cooler Drive indicators read OFF.
The OK indicator means the parameter is within the allowed limits. The Error indicator
means the parameter is outside the limits and must be corrected before the Detective-EX
can be used. If the Waiting indicator persists beyond a few seconds or if you receive an
Error reading for any of these parameters after the nominal cooldown time has elapsed,
contact your ORTEC representative or our Global Service Center.
When all parameters are within allowed limits, the Survey Mode screen displays the Ready
status message.
48
931040P / 1014 5. ADVANCED SETUP
You can turn the cooler off for 10–20 minutes without significantly warming the detector. As
soon as the cooler is restarted, the Detective-EX typically shows a Ready status within a few
seconds.
The cooler can be restarted at any time, without harming the detector crystal, regardless of
whether the detector is warm, partially cooled, or cooled to the operating temperature range.
You can abort the calibration process at any time; no changes will be made to the calibration
coefficients (i.e., the existing calibration will still be in effect).
49
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
At the end of the calibration period, the Use Improved Calibration button will be displayed
(Fig. 44), along with the actual energy of the calibration source peak. For the factory-defined
sources, the Centroid values should ideally be:
! 137
Cs — 661.62 keV ! 40K — 1460.75 keV
In addition, the screen will display the full width at half-maximum (FWHM), an indicator of
peak shape; and the calibration adjustment (Cal. Adjust). The calibration adjustment is the
difference between the spectrum position of the calibration source’s peak in the new calibration
and the position of the same peak in the current calibration, expressed in energy, as a percentage
of the actual peak energy.
12
If the calibration fails, an error message will be displayed. For more detailed calibration troubleshooting, see
Section 7.8.
50
931040P / 1014 5. ADVANCED SETUP
NOTE When the previous background is no longer valid, the Background Invalid message is
displayed on the Survey Mode screen. The Identify and Monitor buttons become inac-
tive and the system will not post radionuclide IDs. However, the gamma and neutron
count rate meters on the Survey, Search, and SNM Mode screens continue to function
normally.
51
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
5.1.3.5. Back
If you do not want to apply the new calibration, tap Back to return to the Advanced Setup
dialog. (However, if you have run this background as the final step of calibration, remember that
you must run a new background before the Identify and Monitor Modes will function.)
5.1.4. Background
This command begins the background radiation measurement pre-configured with the Bkg
Settings command discussed in Section 5.2.2.
! Tapping the Run a Background button, at the end of the manual calibration routine,
opens the screen shown in Fig. 46 and automatically starts the background acquisition.
! Acquisition continues until you tap either Save or Cancel, which will return you to the
Advanced menu. If the measurement is canceled before the minimum time has elapsed,
the software does not accept (i.e., save and use) the new background spectrum. Instead, it
reverts to the existing back-ground file.13
13
As noted elsewhere, if you are running this background as the last step of the manual calibration routine, the
Identify and Monitor Modes will be disabled until the background measurement is completed and accepted.
52
931040P / 1014 5. ADVANCED SETUP
Figure 46. Collect Background for Figure 47. View the Radionuclide
At Least the Specified Minimum Library.
Time.
53
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
54
931040P / 1014 5. ADVANCED SETUP
Figure 52. Suspect Nuclide Report Figure 53. Top Spectral Lines.
(for LCX Mode spectra only).
55
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
5.1.7. About
This dialog (Fig. 55) shows the instrument type,
presence or absence of neutron detector, firmware
and application software version numbers, instru-
ment serial number, and nuclide library version.
Tap Back to return to the Advanced Setup menu.
If the password was incorrect, the Password invalid message will be displayed. You may try to
enter the password again or tap Back to return to the previous menu page. If no password was
set, the Detective-EX will go directly to the second Advanced Setup menu screen.
56
931040P / 1014 5. ADVANCED SETUP
The calibration routine checks and adjusts the Detective-EX’s electronics so that it correctly
identifies the nuclides in the area. This function searches for the calibration source peak in a
window 30 keV wide, centered on the peak channel for the selected source. Once the peak is
located and its net area reaches 4000 counts, the detector amplifier’s fine gain setting is adjusted
for optimum performance. Unlike manual calibration, auto-calibration does not delete the
existing background.chn file so there is no need to take a new background measurement
afterwards.
! To choose a calibration source, tap the source list at top left and make your selection.
! To define a calibration source (up to two),14 tap the Setup... button to open the screen
shown in Fig. 59. Tap in the Source Name field and use the soft keyboard to enter the
name, tap in the Energy Peak field and enter the peak energy in keV, then tap Back
14
If you wish, you may delete the factory-defined sources; they can be restored to the source list at any time.
57
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
to return to the calibration settings screen. The new entry will be available on the source
list.
Figure 58. Set Up the Auto- Figure 59. Enter the Peak Energy
Calibration Feature. for a Calibration Source.
! To use the auto-calibration feature, mark the Enable box by tapping it. To disable auto-
calibration, simply unmark the box.
Set the auto-calibration interval by tapping the Schedule lists and selecting the desired
time of day as well as one of the following intervals:
— Daily.
— Weekly; choose the Day of the week.
— Monthly; tap in the Day of the month field and use the soft keyboard to enter the day.
58
931040P / 1014 5. ADVANCED SETUP
5.2.1.2. Auto-Calibration
! During auto-calibration, the screen displays the
time remaining to completion of the calibration,
as shown in Fig. 60. This number is the real time
for the calibration spectrum acquisition and is
based on the peak count rate. Auto-calibration
will not succeed until the calibration peak passes
the quality test.
! If you disconnect the instrument from the external power source during auto-calibration,
the routine will attempt to run to completion. Tap on the Abort button to interrupt auto-
calibration, then Back to return to Survey Mode. When the unit is reconnected to external
power, auto-calibration will start at the next scheduled interval.
The minimum acceptable data collection time for the background measurement is set on the Bkg
Settings screen, shown in Fig. 61.
59
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
In preset mode, the Pause/Continue button allows you to extend the count. For more informa-
tion and an example, see Section 3.2.2.
60
931040P / 1014 5. ADVANCED SETUP
is arguably improved in LCX mode, the false alarm rate will increase to a level unusable for
many applications. Be aware that the instrument will issue approximately two false suspect
threat alarms every hour while in LCX mode. Therefore, we recommend LCX mode only for
experienced spectroscopists.
To enable LCX Mode, mark the checkbox and click on Back. A blue simulated LED ( ) will be
displayed in the upper left of all screens, beside the screen title (Figs. 63 and 64). This shows
you, at a glance, which radiation monitoring mode is currently selected.
Figures 65 and 66 compare how suspect identifications in LCX mode are displayed in the
“Classify” and “ANSI” ID Display Modes (both screens reflect the same set of found and
suspect substances). Note that in LCX mode, the Color coded IDs option is always on.
Figure 65. Suspect IDs in “Classify” Figure 66. Suspect IDs in “ANSI”
Display Mode. Display Mode.
5.2.4. Settings
5.2.4.1. Password
The Settings screen allows you to set a password to prevent unintended changes to the
Advanced Setup parameters. The password is case sensitive, alpha-numerical, can include
spaces, and can be of any length. There is no factory default.
61
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
! To prevent spectra from being viewed on the Detective screen, unmark the Allow
Display of Spectrum checkbox, then set a password in the Advanced Setup dialog
(Section 5.2.4.1).
! If the spectra must not be viewed at any time by anyone, we additionally recommend that
the field operator not save them.
NOTE If spectra are saved (whether or not the Display Mode is locked), a properly configured
computer and MAESTRO can still be used to view the saved spectra. In such cases, the
Detective-EX (and any memory cards containing spectra) must be secured from
unauthorized access according to your site operating procedures.
62
931040P / 1014 5. ADVANCED SETUP
To set the Q Alarm Threshold for each gamma emitter, tap the radioisotope name, tap in the
threshold field to open the soft keyboard, enter the desired value, and tap Set. Figure 69 shows
the setup dialog for Pu-239.
To set up the Neutron count rate (for units so equipped), in counts per second, tap the Neutron
tab, tap in the threshold field to open the soft keyboard, enter the desired value, and tap Set.
To enable the Ba-133 search (Fig. 70), mark the Include in Search checkbox and enter the
desired alarm threshold value, then tap Set. When this feature is disabled, the Ba-133 bar graph
is not displayed on the SNM Search Mode screen.
63
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
Figure 69. Configure the SNM Figure 70. Enable the Ba-133
Search. Search and Alarm Threshold.
To set either Threshold, tap in the field to display the soft keyboard. Enter the desired value and
tap Set.
You can choose display the gamma dose rate in microSieverts per hour (μSv/hr) or millirem per
hour (mrem/hr). The Dose Units status message updates immediately to reflect the new dose-
rate unit. To change units, tap either the μSv/h or mrem/h radio button to mark it.
64
931040P / 1014 5. ADVANCED SETUP
Figure 71. Advanced Setup, Figure 72. Set the Dose Rate Units
Second Page. and Alarm Thresholds.
65
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
5.3.2. Exit
The Exit button is used to close the DetectiveEX software application and go to the computer
operating system. This is not normally used, but could be needed from time to time. A confirma-
tion dialog is displayed before the program exits. To return to the DetectiveEX program without
exiting, tap Cancel. To restart the Detective application, tap Start, then locate and tap on the
DetectiveEX program shortcut. (The application executable file is located in the My Device\
Program Files\DetectiveEX folder.)
5.3.3. Spectra
The Spectra command opens the screen shown in
Fig. 74, which lets you choose sample description
options, the file storage location, and the file save
formats. In addition, you can delete spectrum files.
66
931040P / 1014 5. ADVANCED SETUP
When the Ask dialog is shown, the default description is in the dialog. You can add to it or
change it completely. This is a good way to enter the main portion of the description ahead of
time, then add a unique identifier, such as a sequence number, to each spectrum as it is acquired.
The File Save Format feature (Fig. 76) lets you choose to generate these files as ORTEC file
formats, ANSI N42.42 format files (compatible with ANSI N42.42-2006), or both.
67
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
Figure 75. Select the Data Figure 76. Select the Data
Location. Format(s).
68
6. FILE TRANSFER TO COMPUTER
The Detective-EX stores Identification Mode spectra in the ORTEC .SPC file format, and Search
Mode spectra in the simpler .CHN format. Once you have downloaded the spectrum files, you
can view and analyze them with ORTEC spectroscopy programs such as MAESTRO, Gamma-
Vision, and ISOTOPIC. The file transfer method depends on your computer operating system
and where your spectrum files are stored on the Detective-EX.
! Spectra stored in the SD card’s Data folder — The simplest method is to move the card
to a computer with a memory card reader and Windows Explorer. Otherwise, transfer
files using the Windows Mobile Device Center or ActiveSync as described below.
— Windows XP: You must download and install Microsoft ActiveSync driver, as
described in Section 6.2, then copy files via Windows Explorer.
IMPORTANT Before connecting the Detective-EX to a computer for the first time, install
Windows Mobile Device Center or ActiveSync first. Otherwise, the unit may not
be able to communicate properly with the computer. Contact our Global Service
Center if you need assistance.
2) Connect the Detective-EX’s DATA port to the computer. If this is the first time the Detec-
tive-EX has been connected to this computer, Windows 8 and 7 will indicate that the hard-
ware driver has been installed. Windows Mobile Device Center will then take several
seconds to communicate with the Detective-EX and indicate the its computer is Connected
(Fig. 78). If Windows Mobile Device Center cannot connect to the trans-SPEC, see the
troubleshooting steps in Section 7.10.
3) Click Connect without setting up your device. This will prevent the host computer and
Detective-EX from automatically synchronizing (i.e., exchanging files). Do not modify the
Program and Services, transfer graphics or music to the trans-SPEC computer, or change
the Mobile Device Settings.
69
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
IMPORTANT Remember that the Detective-EX is not to be used for purposes beyond the
monitoring tasks described in the user manual, and no changes should be
made to the operating system settings.
4) Hover over File Management to expand its topics, and click Browse the contents of your
device (Fig. 79).
Figure 78. The trans-SPEC is Connected to the Figure 79. Choose the File Management Browse
Computer. Option.
5) Windows Explorer will open and the Detective-EX’s computer will be listed as shown in
Fig. 80. Stored search and identification files will be located in the SD card’s Data folder or
in the My Documents folder, depending on the location set on the Spectra dialog (see
Section ?).
6) If your files are stored on the SD card, select the Storage Card and open the Data folder. If
your files are in the My Documents folder, select the Documents item on the left sidebar.
Identification Mode spectra in the .SPC format have a “spectrum” icon with a green back-
ground. Search Mode files in the .CHN format have a “spectrum” icon with a gray back-
ground. Text files have the default file icon.
70
931040P / 1014 6. FILE TRANSFER TO COMPUTER
6.2. Windows XP
1) Download the free ActiveSync v4.5 or higher from the Microsoft Download Center. When
download is complete, select Run. This will automatically start installation. Follow the
wizard prompts and restart the computer.
2) Connect the Detective-EX’s DATA port to the computer. If this is the first time the unit has
been connected to this computer, Windows XP will display a series of “Found New Hard-
ware” messages indicating a mobile device (e.g., the Detective-EX computer) has been
detected. The Synchronization Setup Wizard will start. Follow the wizard prompts for a
standard partnership according to step (3).
71
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
IMPORTANT Remember that the Detective-EX is not to be used for purposes beyond the
monitoring tasks described in the user manual, and no changes should be
made to the operating system settings. Doing so could void your warranty.
4) Continue with installation as directed. At the end of the wizard, the ActiveSync dialog will
indicate the Detective-EX computer is Connected (Fig. 82). If ActiveSync cannot connect to
the Detective-EX, see the troubleshooting steps in Section 7.10.
5) You can now click the Explore toolbar button in the ActiveSync dialog to open Windows
Explorer and view the new connection and its menu tree.
From the host computer, the Detective-EX is displayed as My Windows Mobile-Based Device
in the Mobile Device folder under My Computer (Fig. 83). Stored search and identification
files will be located in the SD card’s Data folder or the My Documents folder or, depending
on the location set on the Spectra dialog (see Section ?). These locations are indicated by
arrows in Fig. 83.
Identification Mode spectra in the .SPC format have a “spectrum” icon with a green back-
ground. Search Mode files in the .CHN format have a “spectrum” icon with a gray back-
ground. Text files have the default file icon.
72
931040P / 1014 6. FILE TRANSFER TO COMPUTER
73
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
[Intentionally blank]
74
7. TROUBLESHOOTING
7.1. “Error Saving .SPC File” Message
This message is generated if you are saving spectrum and search files to SD card, and (1) the
card is removed during a Save operation; or the storage medium is defective. If you removed the
card, reinsert it and tap Save again. If the card is defective, replace it and tap Save. If you do not
have a replacement card, you must stop the current identification session without saving (in
which case, the spectrum will be discarded), go to the Spectra dialog (Section 5.3.3.3), and
select a valid location.
If this error occurs when saving files to the computer’s My Documents folder, contact your
ORTEC representative or our Global Service Center.
! Turn off the MCA electronics (not the cooler) by removing all USB cables and any
power cables, then pressing the on/off switch under the protective cover on the top of the
display unit. Wait 5 seconds, then turn the unit back on. This should reset any mechanical
relays that might have been deactivated by the drop.
! Listen to the cooler: If you hear any unusual noises (rattling or pinging) emanating from
the cooler, contact your ORTEC representative or our Global Service Center.
75
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
! Monitor the unit for the next few hours to ensure that it remains Ready (indicating that
the cooler and bias voltage are functioning properly), calibrates successfully, and cor-
rectly identifies the specified calibration source. If, on recalibration, a large adjustment is
proposed (>0.3% or more), this could indicate a problem.
If the unit fails in one or more of these respects or if one or more hardware diagnostic settings
(Section 5.1.1) indicate a hardware failure, contact your ORTEC representative or our Global
Service Center for further assistance. Remember that there are no user-serviceable parts inside
the Detective-EX, and opening the case will void the warranty.
! If the battery does not hold a charge, ensure that it is properly installed. If properly
installed, the battery must be replaced according to the instructions in Chapter 10.
! If the cooler does not automatically restart, go to the Advanced Setup menu and tap Turn
Cooler On.
If the problem persists, contact your ORTEC representative or our Global Service Center for
assistance.
76
931040P / 1014 7. TROUBLESHOOTING
the on/off button to restart the computer. When restart is completed, the DetectiveEX
program will automatically start and the unit will be ready for use.
For further assistance, contact your ORTEC representative or our Global Service Center.
! If the amount of time remaining (see Fig. 43) is lengthy, use a higher-activity calibration
source or position the source on the detector endcap.
! If the calibration routine does not complete (continues running beyond one minute), tap
STOP to abort it, correct the situation, and restart the calibration.
! If the energy calibration appears highly variable, contact our Global Service Center or
your ORTEC representative. Routine changes of the order of 0.05% are to be expected.
This might simply be due to counting statistics in the spectrum. A change much larger
than this, say 0.5%, bears investigation. Apparent changes in the reported FWHM cali-
bration are most likely to be statistical variations. If the (gain) calibration adjustment
value is stable, the unit is functioning properly.
! If large changes in the calibration continue to occur, contact our Global Service Center or
your ORTEC representative.
77
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
! Confirm the USB cable is good. If connecting via a USB hub, ensure that it is correctly
connected and functioning. Finally, make sure you are connecting to the computer via the
DATA port. If this does not resolve the issue, go to the next step.
! In Control Panel, open the Device Manager and check for an Unknown Device entry that
disappears when you disconnect the Detective-EX from the computer. If such an entry
exists, reconnect the Detective-EX, right-click on the Unknown Device, select Uninstall,
and confirm that you wish to uninstall the instrument. Disconnect the Detective-EX from
the computer.
! Download and install (or uninstall and reinstall) the ActiveSync or Windows Mobile
Device Center software, making sure you have downloaded the correct version; restart
the computer; and start the connectivity program.
! Restart the Detective-EX by pressing and holding the On/Off button for 5 seconds, until
the screen goes dark and the instrument reboots. When the DetectiveEX program has fully
reinitialized, reconnect to the computer. You should see one or more “new hardware”
bubble messages indicating the Detective-EX was successfully detected.
78
931040P / 1014 7. TROUBLESHOOTING
! If these steps do not resolve the issue, contact your ORTEC representative or our Global
Service Center. In the meantime, if your spectrum files are stored in the instrument’s My
Documents folder, you may wish to exit the DetectiveEX program, use the computer’s
Windows Explorer program to copy the spectra to an SD card, then mount the SD card in
the computer and transfer the files that way.
! Make sure the Detective-EX is turned on and connected to the computer via the
CONTROL port.
! Check the Master Instrument List on the computer to ensure that the Detective-EX is on
the list. If it is not listed, manually run the MCB Configuration program from the Win-
dows Start menu. See the MAESTRO User’s Manual for detailed instructions on this
operation.
79
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
! On the third screen of the Advanced menu (Section 5.3.2), tap Exit, then confirm that
you wish to end the Detective-EX program.
! On the Windows desktop, tap the Start square in the upper left corner of the screen, then
tap Settings.
! In the Settings dialog, tap the System tab, then select Regional Settings.
! On the Region tab, select English (United States) from the listbox.
! On the Number tab, choose the period ( . ) for the Decimal symbol and the comma ( , )
for the Digit grouping symbol.
! On the Time tab, select h:mm:ss tt as the Time style and the full colon ( : ) as the Time
separator.
! On the Date tab, use M/d/yy as the Short date, the forward slash ( / ) as the Date
separator, and the Long date format dddd, MMMM dd, yyyy.
! Tap the upper right OK circle, then the upper right X circle to return to the Windows
desktop.
For further assistance, contact your ORTEC representative or our Global Service Center.
80
8. USING THE DETECTIVE-EX WITH ORTEC
SPECTRUM ANALYSIS PROGRAMS
The Detective-EX is intended for completely independent operation, with no computer required.
However, when you connect the Detective-EX CONTROL port to a computer running ORTEC
spectroscopy applications such MAESTRO or GammaVision, you can use it as a high-perform-
ance HPGe spectrometer/digital MCB.
Be sure to install the accompanying versions of CONNECTIONS Driver Update Kit (P/N 797230)
and MAESTRO. These contain the most up-to-date drivers for the Detective-EX/DX-100T and
other ORTEC hardware. If you have any questions about software/hardware compatibility, con-
tact your ORTEC representative or our Global Service Center. The MAESTRO User’s Manual
contains complete instructions on software installation and configuration of the MCBs attached
to your computer. When installing CONNECTIONS, choose the USB-Based Instruments selection
to install the correct driver for the Detective-EX/DX-100T.
IMPORTANT
When both the Detective and trans-SPEC applications are installed on your identifier, the unit’s
operation depends on the application running when it is connected to the analysis computer.
If running the Detective application, the allowable changes to hardware settings within
MAESTRO are restricted. This includes the gain, rise time, calibration, and other settings that
will affect the location of gamma peaks (see Section for the allowable hardware settings). When
you disconnect the identifier’s CONTROL port from the computer, the unit automatically restores
all factory settings for the standalone Detective.
If the unit is running the trans-SPEC application, almost all hardware settings are adjustable (see
the trans-SPEC manual for complete details on the allowable hardware settings). Any changes
made in MAESTRO remain in effect when you disconnect from the computer and return to
standalone trans-SPEC operation.
The first time this Detective-EX is connected to a particular for use as an MCB, follow the
installation instructions in Chapter 6 for the host computer’s operating system.
81
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
Note, however, that the Detective-EX is not designed to store efficiency information in the .SPC
files it creates. Therefore, for quantitative determinations, special attention is needed to prepare
the calibration (.CLB) file so that GammaVision or ISOTOPIC will add the efficiency informa-
tion during processing.
82
931040P / 1014 8. USING THE DETECTIVE-EX WITH ORTEC SPECTRUM ANALYSIS PROGRAMS
To prepare this calibration file, load a Detective-EX calibration spectrum into GammaVision or
ISOTOPIC. Use the software application’s Calibration Wizard to obtain the additional
efficiency calibration, keeping the existing energy calibration. After the efficiency calibration is
computed, save the complete set of calibrations as a .CLB file.
To add the efficiency information to an existing .SPC file (typically from a standalone
Detective-EX):
! Transfer the .SPC file from the Detective-EX to the computer (see Chapter 6).
! Load the .CLB file containing the proper efficiency information. This can be done either
with the application’s Recall Calibration command; or by specifying the .CLB file as the
calibration override in the GammaVision analysis options (.SDF) file or the ISOTOPIC
configuration.
NOTE If the .CLB file is specified in the GammaVision .SDF file or ISOTOPIC
configuration, the .An1 file generated during analysis will contain the proper
calibration information but the original .SPC file (without the efficiency
information) will not change.
To load the efficiency information before acquiring a new spectrum (with the Detective-EX
attached to a computer):
! Load the .CLB file either with the Recall Calibration command; or by specifying the
.CLB file as the calibration override in the GammaVision .SDF file or ISOTOPIC
configuration.
83
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
and Identification Mode messages for spectra downloaded from the Detective-EX, move from
tab to tab. Click on Close when finished.
Note that as you enter characters in the data-entry fields, the characters will be underlined until
you move to another field or until 5 seconds have lapsed since a character was last entered.
During the time the entry is underlined, no other program or computer on the network can
modify this value. The changes you make on most property tabs take place immediately. There is
no cancel or undo for the dialog.
NOTE In MAESTRO, the Download Spectra and View ZDT Corrected commands on the
Acquire menu and the Smooth and Strip commands on the Calculate menu are
disabled for this instrument.
8.3.1. Amplifier
Figure 85 shows the Amplifier tab. This
tab displays the settings for the Coarse
and Fine Gain, which are factory preset
and cannot be changed. The fine gain
changes when the unit is recalibrated.
84
931040P / 1014 8. USING THE DETECTIVE-EX WITH ORTEC SPECTRUM ANALYSIS PROGRAMS
8.3.2. Amplifier 2
Figure 86 shows the Amplifier 2 tab,
which displays the fixed Rise Time
and Flattop Width settings. You can
also use the InSight Virtual Oscillo-
scope™ mode to view the digital signal
processor’s actual sampled waveform
on a reference graticule; however, the
shaping settings cannot be adjusted.
15
Patent pending.
85
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
8.3.4. ADC
This tab (Fig. 88) allows you to
monitor the current Real Time, Live
Time, and Count Rate. In addition,
you can view the fixed Conversion
Gain and the Lower and Upper
Level Discriminator settings.
16
A reference to our SMART-1™ detector technology. For more information on SMART-1, see the ORTEC
catalog or visit www.ortec-online.com.
86
931040P / 1014 8. USING THE DETECTIVE-EX WITH ORTEC SPECTRUM ANALYSIS PROGRAMS
8.3.6. Reports
This tab (Fig. 90) displays the
reports available for the current live
spectrum or spectrum file, including
the ID, Found Nuclides, Suspect
Nuclides, Top Lines, and (for instru-
ments so equipped) GPS reports. You
must be running MAESTRO or other
ORTEC spectroscopy applications.
8.3.7. About
This tab (Fig. 91) displays hardware
and firmware information about the
Detective-EX as well as the data
Acquisition Start Time (Sample
description is not used). In addition,
the Access field shows whether the
MCB is currently locked with a pass-
word by MAESTRO. Read/Write
indicates that the MCB is unlocked
and Read Only means it is locked.
See the note concerning passwords on
page 82, at the end of Section 8.1.
Use the Firmware Revision field to Figure 91. Detective-EX About Tab.
determine both the type of Detective-
EX (regular or -100 model, with or without neutron detector) and firmware version. This para-
meter is formatted DETX-nnn, where X is C for the regular Detective-EX, D for the Detective-
EX-100, F for the regular Detective-DX, and G for the Detective-DX-100; and nnn is the
firmware version. In the accompanying illustration, DETF-304 indicates a regular Detective-DX
87
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
running firmware V3.04. You can also use the About tab to view this information for .SPC
format spectra downloaded from the Detective-EX.
8.3.8. Status
Figure 92 shows the Status tab. Ten
parameters are continuously monitored
in real time. Satisfactory status is re-
ported as OK or as a numerical value.
A failure is reported as ERR or a
descriptive message. Use the listboxes
to select any six parameters to be dis-
played simultaneously on the Status
tab. You can change the selected para-
meters at any time.
! Gamma Dose Rate — reported in µSv/h, and indicating whether this reading was made
with the detector (Ge for low dose rates) or the Geiger-Müller tube (GM for high dose
rates). See Gamma Dose-Rate Determination, page 92, for further discussion of the two
dose-rate meters.
! HV Bias — in volts.
88
931040P / 1014 8. USING THE DETECTIVE-EX WITH ORTEC SPECTRUM ANALYSIS PROGRAMS
To resolve status problems, refer to the troubleshooting list in Chapter 7. For further assistance,
contact your ORTEC representative or our Global Service Center.
8.3.9. Presets
Figure 93 shows the Presets tab. The
presets can only be set when the
Detective-EX is not acquiring data
(during acquisition the preset field
backgrounds are gray indicating that
they are inactive). You can use any
or all of the presets at one time. To
disable a preset, enter a value of zero.
If you disable all of the presets, data
acquisition will continue until man-
ually stopped.
When more than one preset is enabled (set to a non-zero value), the first condition met during
the acquisition causes the MCB to stop. This can be useful when you are analyzing samples of
widely varying activity and do not know the general activity before counting. For example, the
Live Time preset can be set so that sufficient counts can be obtained for proper calculation of
the activity in the sample with the least activity. But if the sample contains a large amount of this
or another nuclide, the dead time could be high, resulting in a long counting time for the sample.
If you set the ROI Peak preset in addition to the Live Time preset, the low-level samples will
be counted to the desired fixed live time while the very active samples will be counted for the
ROI peak count. In this circumstance, the ROI Peak preset can be viewed as a “safety valve.”
The values of all presets for the currently selected MCB are shown on the Status Sidebar. These
values do not change as new values are entered on the Presets tab; the changes take place only
when you Close the Properties dialog.
Enter the Real Time and Live Time presets in units of seconds and fractions of a second. These
values are stored internally with a resolution of 20 milliseconds (ms) since the MCB clock
increments by 20 ms. Real time means elapsed time or clock time. Live time refers to the amount
89
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
of time that the MCB is available to accept another pulse (i.e., is not busy), and is equal to the
real time minus the dead time (the time the MCB is not available).
Enter the ROI Peak count preset value in counts. With this preset condition, the MCB stops
counting when any ROI channel reaches this value unless there are no ROIs marked in the MCB,
in which case that MCB continues counting until the count is manually stopped.
Enter the ROI Integral preset value in counts. With this preset condition, the MCB stops count-
ing when the sum of all counts in all channels for this MCB marked with an ROI reaches this
value. This has no function if no ROIs are marked in the MCB.
Marking the Overflow checkbox terminates acquisition when data in any channel exceeds 231!1
(over 2×109) counts.
90
9. SPECIFICATIONS17
System Conversion Gain The Detective-EX/DX operates at a conversion gain of 8192
channels.
Shaping-Time Constants
! Rise time: 3.4 µs, fixed.
! Flattop: 0.8 µs width, fixed.
Linearity
! Integral Nonlinearity: <±0.025% over top 99.5% of spectrum, measured with a mixed
source (55Fe at 5.9 keV to 88Y at 1836 keV).
! Differential Nonlinearity: <±1% (measured with a BNC pulser and ramp generator).
Overload Recovery At maximum gain, recovers to within 2% of rated output from ×1000
overload in 2.5 non-overloaded pulse widths (measured using our InSight™ Virtual
Oscilloscope).
Pulse Pileup Rejector Automatically set threshold. Pulse-pair resolution typically 500 ns.
Low Frequency Rejector (LFR) Filter19 The digital signal processor incorporates ORTEC’s
exclusive LFR filter, designed to reduce microphonic and low-frequency periodic noise (if
present) from germanium detector output signals, thereby producing significant improvement in
spectral resolution. The LFR is specifically beneficial in systems cooled by mechanical coolers,
which are known to often contribute microphonic noise to the spectrum (thereby degrading
spectral resolution). The LFR is also capable of reducing any periodic noise signal resulting
from surrounding electronics and equipment or ground loops.
17
Subject to change without notice.
18
Ron Jenkins, R. W. Gould, and Dale Gedcke, Quantitative X-Ray Spectrometry (New York: Marcel Dekker,
Inc.), 1981, pp. 266–267.
19
Patent pending.
91
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
Neutron Detector Internal, 4 each, 3He tubes, 1.3 cm × 10 cm active length, 20 atm, stainless
steel body, RS-P4-0404-214, inside 2.5 cm × 18 cm × 14 cm moderator.
Internal Compensated Geiger-Müller Tube LND 71322 Geiger Müller tube. GM tube
specifications are available on request.
The Detective-EX uses two detectors to determine the gamma dose rate over a wide range from
<0.05 µSv/h to >10000 µSv/h, a dose-rate range of around six decades. For low dose rates,
below ~20 µSv/h, the dose rate is determined from the Ge detector spectrum. For dose rates
above this value, the internal compensated GM tube is used. The Detective-EX switches
between the two automatically.
For radiation fields from extended sources (e.g., from a container full of NORM) or far-field
point sources (>1 ft [0.3 m] from the instrument), the precise location of the detector in use is
not important. When point sources are close to the instrument, the detector readings might not be
in agreement because they are in different places within the instrument, and the dose rate is
varying rapidly with position. It will be different at these two detector locations.
NOTE The dose rate function in the Detective-EX should not be considered as adequate for
accurate dosimetry purposes. Its primary purpose is to protect the user of the Detective-
EX from high radiation fields.
If you must test the dose-rate measurement capability with a point source, especially at close
range, it is important to know the reference points of the instrument detectors, since test proce-
dures such as ANSI or IAEA prescribe use of point sources at specified distances from the
reference point of the instrument. For the Ge detector, the reference point is the geometric center
of the Ge detector endcap disk (see Fig. 94). For the GM and neutron detectors, the reference
point is shown in Fig. 94 and is marked on the case of the instrument.
92
931040P / 1014 9. SPECIFICATIONS
Figure 94. Reference Points for the HPGe and Neutron Detectors and the Geiger-Müller Tube (all
measurements in inches).
Presets Multiple presets can be set within MAESTRO for any or all of the following:
! Real Time Stops data collection when the real time reaches this value (in increments
of 1 s).
! Live Time Stops data collection when the live time reaches this value (in increments
of 1 s).
! ROI Integral Integral count stops data collection when the sum of all ROI channels
reaches this value.
93
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
! ROI Peak Peak count stops data acquisition when any ROI channel reaches this value
(maximum value of 232!1 counts [4 ×109]).
! Data Overflow Terminates acquisition when data in any channel exceeds 231!1
(2 ×109) counts.
!Weight
Detective-EX-100T: 12 kg (26.3 lb).
Detective-DX-100T: 11.3 kg (24.4 lb).
Maximum Shock Tolerance 20 g. An accelerometer inside the instrument tracks the maximum
shock to which the Detective-EX has been subjected, and a shock in excess of 20 g will void the
warranty.
External Power DC input and battery charge input, MS3112E12-10-S or Bendix PT02E-
12-10S connector.
Internal Battery Rechargeable, nominal 14.4-V lithium-ion battery pack. Contact ORTEC for
the appropriate replacement kit. Charging circuitry and battery management circuitry internal to
the docking station or PAC-II. Battery lifetime 3–5 years. Access door in right side panel of
case. The compartment is semi-sealed and protects against penetration of dirt and moisture. See
Chapter 10 for instructions on replacing the battery.
Battery Life >3 hours at 25°C when HPGe detector is cold. Battery life can be extended indefi-
nitely by the use of optional, external battery belt. The unit is expected to be kept running once
cold.
94
931040P / 1014 9. SPECIFICATIONS
Note that the internal battery charge is not sufficient to start the Detective-EX cooler or cool a
warm detector; external power is required. In addition, an exhausted internal battery should be
charged for 2 hours before the Detective-EX is cooled using external power. Cooler startup
without external power and/or with an exhausted internal battery can cause operational ano-
malies. For other startup issues, see Sections 7.5 and 7.6. Also note that using an automobile
battery to cool the detector could significantly discharge the auto battery.
CE Conforms to CE standards for radiated and conducted emissions, susceptibility, and low-
voltage power directives.
9.2. Connectors
DATA Rear-panel USB connector, with dust cover, for computer communications.
CONTROL Rear-panel USB connector, with dust cover, for MCA communications.
INPUT POWER Rear-panel external power input, with dust cover, 12–17 V dc, 60 W, or from
battery or auto-sensing ac-powered battery charger.
Headphones ( ) Rear-panel female connector, with dust cover, accepts standard male head-
phone jack. Headphone volume controllable from the Detective-EX button pad. Note that the
audible high-dose-rate alarm sounds whether or not headphones are connected to the Detective-
EX. This alarm is not routed through the headphones and its volume is independent of the head-
phone volume setting.
95
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
Charger Status LEDs Front-panel red and green LEDs. When the Detective-EX’s internal
battery is charging, the red indicator is continuously lit. The red indicator flashes if you discon-
nect the unit before its battery is fully charged, or if it is connected but the internal battery is
faulty. When the battery is fully charged, the green indicator turns on and the red indicator turns
off.
CAUTION
This instrument should be cooled and brought to operational status for at least 48 hours every
month. Failure to do this may result in degraded performance or cooling system failure.
This instrument uses a Sterling cycle refrigerator and all-metal-sealed cryostat designed for long
operational life. The vacuum integrity inside the cryostat is maintained by “getters” which
scavenge traces of residual material when cold, and by an ion pump which operates when the
96
931040P / 1014 9. SPECIFICATIONS
instrument is powered up. These two techniques maintain the system vacuum in peak condition
during normal use. A good vacuum yields rapid cooldown, and ensures operation at the high end
of the instrument’s operating temperature range.
Over time, in any cryostat system, residual contaminants can be released from the materials
inside the evacuated cryostat by a process called outgassing.20 It has become apparent that, in
extended storage, the vacuum in our mechanically cooled identifiers may degrade to the point
that, on trying to restart, the gas load inside the cryostat is too high and the unit fails to cool. The
remedy is reasonably straightforward, if inconvenient: a “pump and bake” of the instrument is
required at an ORTEC service center. Fortunately, you can easily avoid this problem:
We strongly recommend that instruments used only occasionally be turned on once a month,
fully cooled to Ready status, and allowed to run for at least 48 hours. Doing this starts the ion
pump and activates the internal getters, ensuring that the vacuum is maintained in good
condition.
As well as maintaining the vacuum, this procedure keeps the system battery “topped up” and
also provides peace of mind that the system will cool when required.
20
This is not to be confused with a vacuum leak, although both result in a degradation of the vacuum.
97
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
[Intentionally blank]
98
10. CHANGING THE INTERNAL BATTERY
The typical service life of the Detective-EX internal battery is 3–5 years. When the internal
battery no longer retains a charge, it must be replaced with ORTEC Part No. 798780. This is a
straight-forward operation that takes just a few minutes, so the detector does not significantly
warm up during the procedure. When connected to external power such as the docking station,
the new battery will typically reach full charge within 2–4 hours.
For more information, contact your ORTEC representative or our Global Service Center.
1) Turn off the cooler and the Detective-EX electronics by reversing the instructions in
Section 2.8.2.
2) Figure 96 shows the battery hatch on the side panel. Remove the five (5) Phillips or Torx®
T10 screws from the hatch (there is no sixth screw under the black rubber “foot”). Gently
pry up one end of the hatch, then slip it from behind the rubber foot and expose the battery.
99
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
100
11. CALIBRATION SOURCE STATEMENT
The quantity of radioactive material contained in this product is extremely small and presents no
known radiation hazard. However, it is always good practice to minimize exposure by following
the basic radiation principles of time, distance, and shielding. The solid sealed source in this
instrument requires no special handling.
Exempt quantity licensed products containing radioactive material should be used only as
intended by the manufacturer and in accordance with the instructions provided. All radioactive
materials should be securely stored when not in use.
This product may be disposed of in regular waste without regard to the radioactive content pro-
viding the customer is not a specific licensee and all radiation symbols have been removed or
defaced. If the customer receiving the exempt quantity is a specific licensee, then the customer is
subject to the requirements of 10 CFR Part 20 in areas where 10 CFR 30.18 is silent (e.g., waste
disposal).
This product conforms to the conditions and limitations specified in 49 CFR 173.424 for
radioactive material, excepted package-instruments or articles, UN2911.
101
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
[Intentionally blank]
102
INDEX
Administrator Mode power adapter/charger . . . . . . . . . . . . . . . . . 15
advanced setup . . . . . . . . . . . . . . . . . . . . . . 47 setup and power-up . . . . . . . . . . . . . . . . . . . 20
hiding the spectrum display . . . . . . . . . . . . . 62 Cooler
password . . . . . . . . . . . . . . . . . . . . . . . . . . . 61 cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Alarms detector warmup . . . . . . . . . . . . . . . . . . . . . 96
dose-rate and count-rate . . . . . . . . . . . . . . . . 64 power sources . . . . . . . . . . . . . . . . . . . . . . . 16
MOVE BACK . . . . . . . . . . . . . . . . . . . . . . . 22 power-up . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
RDD . . . . . . . . . . . . . . . . . . . . . . . . . . . 46, 64 restarting . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Audio settings (clicker) . . . . . . . . . . . . . . . . . . . 65 shipping . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Background measurements . . . . . . . . . . . . . . . . 59 shipping preparations . . . . . . . . . . . . . . . . . . 19
final (mandatory) step in calibration . . . . . . . 25 shock tolerance . . . . . . . . . . . . . . . . . . . . . . 18
ID and Monitor Modes disabled . . . . . . . . . . 19 shutdown for brief periods . . . . . . . . . . . . . . 49
invalid background . . . . . . . . . . . . . . . . . . . 19 shutdown for long periods/storage . . . . . 18, 96
minimum data collection time . . . . . . . . 52, 59 startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
performing the background measurement . . . 52 Detective-EX
required after manual calibration . . . . . . . . . 51 ADC setup . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16, 17 amplifier settings . . . . . . . . . . . . . . . . . . 84, 85
cannot cool detector . . . . . . . . . . . . . . . . 21, 49 carrying strap . . . . . . . . . . . . . . . . . . . . . . . . 13
charge status . . . . . . . . . . . . . . . . . . . . . 14, 15 docking station . . . . . . . . . . . . . . . . . . . . . . . 95
power source indicator . . . . . . . . . . . . . . 17, 22 features, illustrated . . . . . . . . . . . . . . . . . . . . 10
replacing . . . . . . . . . . . . . . . . . . . . . . . . 18, 99 GPS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Calibration check . . . . . . . . . . . . . . . . . . . . 23, 49 high-voltage setup . . . . . . . . . . . . . . . . . . . . 86
auto-calibration . . . . . . . . . . . . . . . . . . . . . . 57 operating cautions . . . . . . . . . . . . . . . . . . . . 18
manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 power adapter/charger . . . . . . . . . . . . . . . . . 96
new background required . . . . . . . . . . . . 25, 51 power-up . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 presets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
troubleshooting . . . . . . . . . . . . . . . . . . . . . . 77 setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
user-defined sources . . . . . . . . . . . . . . . . . . 57 shock tolerance . . . . . . . . . . . . . . . . . . . . . . 18
Calibration source . . . . . . . . . . . . . . . . . . . . 14, 24 software and firmware versions . . . . . . . . . . 56
user-defined . . . . . . . . . . . . . . . . . . . . . . . . . 57 status screen . . . . . . . . . . . . . . . . . . . . . . . . . 48
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 temperature and humidity specifications . . . . 18
Charger bases use as conventional MCA . . . . . . . . . . . . . . 79
cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Display Mode . . . . . . . . . . . . . . . . . . . . . . . . . . 35
docking station . . . . . . . . . . . . . . . . . . . . . . . 14 hiding the spectrum display . . . . . . . . . . . . . 62
power adapter/charger . . . . . . . . . . . . . . . . . 15 Docking station
Clicker cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
audio settings . . . . . . . . . . . . . . . . . . . . . . . . 65 connectors . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Connecting to a PC . . . . . . . . . . . . . . . . . . . . . . 81 features . . . . . . . . . . . . . . . . . . . . . . . . . 14, 15
using MAESTRO . . . . . . . . . . . . . . . . . . . . 83 indicators . . . . . . . . . . . . . . . . . . . . . . . . 14, 15
Connections placement . . . . . . . . . . . . . . . . . . . . . . . . 14, 15
auto battery adapter cables . . . . . . . . . . . . . . 16 power . . . . . . . . . . . . . . . . . . . . . . . . . . . 14, 16
Detective-EX . . . . . . . . . . . . . . . . . . . . . . . . 12 setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
docking station . . . . . . . . . . . . . . . . . . . . . . . 14 Dose- and count-rate limits . . . . . . . . . . . . . 30, 64
103
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual 931040P / 1014
104