Description and Intended Use


I-View Silver is a digital imaging solution for intraoral dental radiography based on CMOS
(Complementary Metal-oxide Semiconductor) technology. The sensor is a receptor which
captures the images obtained when using an X-Ray generator for diagnostic purposes.

State-of-the-art technology reduces the patients' exposure and provides excellent image
quality with no missing data due to the single-chip CMOS. Images are obtained from the
exposure of the tissues of the mouth to a radiation source or X-ray generator. By interposing
the tissue between the radiation source and the receptor, the densest parts appear with
different tones within greyscale: structures that are dense, such as bones or metal, will block
most x-ray particles and appear white, air-containing structures will look black, and muscles,
fat and fluids will appear as shadows of grey color.

The sensor transmits the acquired image to the computer, instantly displaying it on the
screen without additional processes. Subsequently, the specific software facilitates the
identification of the images and allows to improve their analysis.

Product Presentation
I-View Silver is available in two sizes:
Size 1: surface of 25 x 36.88 x 4.4 mm and active area 20 x 30 mm
Size 2: surface of 30.4 x 41.9 x 4.4 mm and active area 26 x 36 mm
The sensor comes in a solid, black cardboard box, with the Trident logo, accompanied
by the USB key with the Deep-View v4.0 software and a dongle with the user license.

Main physical characteristics


. Rounded edges
. Flat and smooth coverage
. Very resistant cable
. Direct connection to the computer with USB cable

Sensor advantages
. Immediate images, without the use of films or chemicals
. Perfect exposure
. High resolution of images with the lowest radiation exposure
. Safe and efficient storage of information in digital archives
. Workspace optimization

Product code
IS034 -DL: I-View Silver Size 1
IS035 -DL: I-View Silver Size 2

2
Trident S.r.l.
I-View Silver
Technical Info

General Characteristics
Unit name I-View Silver
Trident Srl
Manufacturer
Via Artigiani 4, Castenedolo, 25014 (BS) Italy
Classification according to Directive EC
Class IIa with Type BF applied parts
93/42
IEC 60601-1:2005/A1:2012
IEC 60601-1-2:2014
IEC 60601-1-3:2008
Applicable Standards IEC 60601-1-6:2010
IEC 62366:2007/AMD1:2014
IEC 62304+AMD1:2015
ISO 15223-1:2016

Technical Information
Parameter Unit Size 1 Size 2
Detector Technology -- APS CMOS
Sensitive area structure -- CsI directly deposited (Ti)
Protection -- Equivalent to IP68
Dimensions W x L x H mm 25,4 x 36,88 x 4,4 30,4 x 41,9 x 4,4
Sensitive area (active) mm 20 x 30 26 x 36
Pixel size μm 20
Pixel shape -- squared
Number of active pixels pixel 1000 x 1500 1300 x 1800
Data output --- 16 bits ADC

Electrical and Optical Specifications


Parameter Value
Supply voltage 5 V DC
Power supply Directly through a USB connection
Typical absorbed maximum 250 mA
Dark noise < 18lb
Maximum linear dose ≥ 600μGy
Response linearity ≥ 0,99
Theoretical resolution 25 lp/mm
Measured resolution ≥ 20 lp/mm
Dynamic range ≥ 70 dB
Global non-uniformity ≤ 5%
Total dose irradiation ≥50 Gy
X-ray shooting mode AED
AED activation threshold 400μGy/s
X-ray energy 55 to 100 kV
Exposure window Max. 5s, and programmable

3
Trident S.r.l.
I-View Silver
Technical Info

Mechanical Characteristics
Housing material ABS
X-ray shielding Shield Pb behind the active area
Sensor input protection Equivalent to IP68
Sensor color Grey
USB cable length 3m

Recommended operating Conditions


10 - 40 ° C (in operation)
Operating temperature
-10 - 60 ° C (stored)
Operating humidity 20-90% RH (in operation, non-condensing)
10-95% RH (storage and transport, non-
Pressure
condensing)
Vibration 700 - 1060mbar (operation and storage)
Shock IEC 60068-2-6

Recommended Exposure Parameters


High voltage From 60 to 70 kV
Focal point (according to IEC 62336) 0.5 mm suggested, 0.7 mm
Focal point to sensor 22 cm
Beam limiting device Max. 6 cm
Between 300 and 500 µGy for incisors
Adult patient dose
Between 400 and 600 µGy for molars
Dosage for pediatric use Reduce the dose about 20%

Operating System requirements


Windows 8
Operating System
Windows 10
CPU Intel i5-2520M 2.5 GHz or superior
Main memory 1 GB (ideally 2 GB)
Hard drive 10 GB RAM
USB Port 2.0 type A
1024x768 resolution in 65,000 colors (ideally 1280x1024 - 16
Video
million colors, 32 bit)

4
Trident S.r.l.
I-View Silver
Technical Info

Measures

5
Trident S.r.l.
I-View Silver
Technical Info

Labels
The label with the information is located at the base of the sensor packaging. On the outside of the
black box is fixed a silver label that contains all the necessary data to identify the sensor and
facilitate its traceability. This label shows the following data:

Label for size 1 Label for size 2

• The company name in lowercase letters, font Century Gothic, with three points representing
the "i" which identifies the company logo.
• #: Unit Type I-View Silver
• Reference: code composed of six alphanumeric characters.

REF NAME SIZE SOFTWARE
IS034-S I-View Silver 1
No Software
IS035-S I-View Silver size 2 2
IS034-DL I-View Silver 1
Software Deep-View
IS035-DL I-View Silver size 2 2

• Serial number

TRS X XXXXXXNNNN

Identifies the X=1 size 1 The first six numbers identify the Batch
Manufacturer number; the following ones, show the
X=2 size 2
progressive number of the sensor
Trident-Silver

6
Trident S.r.l.
I-View Silver
Technical Info

• The unique device identifier (UDI) that consists of the following:


(01): Manufacturer ID

(21): Device type and S/N

(11): Manufacturing date

• Voltage, maximum current and type of USB connector.

Other information shown in the label :

ICON MEANING

Production date YYYY/MM/DD

Risk warning for the patient and the operator.

The sensor has some electro-static sensitive parts.


Do not touch the sensor and computer screen at the same time.
Do not touch the USB connector.
Come into physical contact with the patient in order for the device to carry
out its intended function.

Double insulation
Device in II security class
Selective collection of electrical or electronic devices. At the end of its useful
life do not throw this device in household waste

Read instructions

Manufacturer address, legal seat and production facilities

MD Medical Device

CE Certificate issued by IMQ certification body No. 0051

7
Trident S.r.l.
I-View Silver
Technical Info

Packaging
Weight Size
Box
(Kg) (Cm)

0.4 Kg 18 x 13,5 x h 4

You might also like