Chiralpak Ie
Chiralpak Ie
Chiralpak Ie
CHIRALPAK® IE COLUMNS
Please read this instruction sheet completely before using this column
IE columns can also be used in reversed phase mode.
Please refer to the corresponding instruction sheet for details.
Column description
CHIRALPAK® IE
OR
Cl
O
O
RO
R:
OR O N Cl
n H
silica gel
Operating Instructions
Should be maintained < 300 Bar (4350 psi) for maximum column life
Pressure limitation
Adapt flow rates to column size.
Temperature 0 to 40°C
The maximum flow rate depends on the mobile phase viscosity (mobile phase composition), and
should be adjusted in accordance with the pressure upper’s limit (i.e. 300 Bar).
Method Development / Normal Phase
A - Mobile phases
CHIRALPAK® IE can be used with all ranges of organic miscible solvents, progressing from the traditional
mobile phases used with other DAICEL columns (mixtures of alkanes/alcohol, pure alcohol or acetonitrile
(CH3CN)) to mobile phases containing methyl tert-butyl ether (MtBE), tetrahydrofuran (THF),
dichloromethane (DCM), chloroform (CHCl3), ethyl acetate (EtOAc) among others.
Primary solvent
Alkane/2-PrOH Alkane/EtOH Alkane/MtBE/EtOH Alkane/THF Alkane/DCM/EtOH
mixtures
Typical starting
80:20 80:20 0:98:2 70:30 50:50:2
conditions
Alkane = n-Hexane, iso-Hexane or n-Heptane. Some small selectivity differences may sometimes be found.
In absence of alkane, methanol is more efficient than ethanol when combined with MtBE.
In the case of no environmental restrictions, use of DCM is preferred to THF in terms of better enantioselectivity that the former may
induce.
For excessively retained samples, addition of ethanol up to 20% in pure DCM would be helpful.
If a suitable chiral separation is not found using the Immobilised Primary Screening strategy, we recommend
an Immobilised Secondary Screening to be applied using the following conditions:
Alcohols () or THF can be added into EtOAc to enhance the eluting strength for strongly retained compounds.
Alkane: n-Hexane, iso-Hexane or n-Heptane. Some small selectivity differences may sometimes be found.
Transfers between alkane mixtures and CH3CN are preferably made with a transition in alcohol in order to avoid miscibility issues.
Alcohol: MeOH, EtOH and 2-PrOH.
C – General Comments
Additional solvent combinations such as CHCl3/Alkane, 1,4-Dioxane/Alkane, Toluene/Alkane or
Acetone/Alkane can also be investigated with CHIRALPAK® IE column.
The typical starting conditions represent the mobile phases of upper middle eluting strength. Under such
conditions, most of the analytes can be eluted within a reasonable time range with a good probability of
full resolution of the enantiomers.
Toluene, MtBE and chlorinated solvents can be used in their pure form as the mobile phase.
For fast eluting solvents, such as THF, we recommend to add alkane in order to modulate the retention.
Detection with a regular UV detector may become difficult depending on a combination of sample and
mobile phase (e.g. EtOAc, high percentages of DCM). In these cases an alternative detector, such as RI
detector or ELSD (Evaporative Light Scattering Detector), may be more effective than the UV detector.
Samples should be dissolved in the mobile phase. The mobile phase and the sample solution should be
filtered through a membrane filter of approximately 0.5µm porosity to ensure that there is no
precipitate before using.
Following extensive use of the column in multiple solvents there may be a change in separation
reproducibility. In order to ensure consistent performance, a regeneration method may be implemented
to eliminate any change in chiral recognition due to the history of the column (mobile phases,
additives…).
• Flush with ethanol at 0.5 ml/min(*) for 30 min, followed by 100% N,N-dimethylformamide (DMF)
at 0.3 ml/min(*) for 3 hours.
• Flush with ethanol at 0.3 ml/min(*) for 50 min and then equilibrate with n-Hexane/IPA = 90/10
(v/v) prior to retesting the column.
(*)
Recommended flow rate for analytical columns (4.6mm ID).
Column storage
For column storage, remove the acidic or basic additives by flushing the column with the same mobile
phase without the additive. Columns can be stored end capped with additive-free mobile phases.
If you have any questions about the use of these columns, or encounter a problem, please contact
DAICEL CORPORATION for assistance ([email protected]).
Operating this column in accordance with the guidelines outlined here will result in a long
column life.
CPI Company
2-18-1, Konan, Minato-ku, Tokyo 108-8230, Japan
TEL: +81-(0)3-6711-8222, FAX: +81-(0)3-6711-8228
http://www.daicelchiral.com/
E-mail: [email protected]
DAICEL CHIRAL TECHNOLOGIES(CHINA) CO., LTD. DAICEL CHIRAL TECHNOLOGIES(INDIA) PVT., LTD.
Part C, FL5 No.16, Xiya Road No.69, Waigaoqiao Free Laboratory No. 4A, Innovation Corridor 1,
Trade Zone, Shanghai, 200131, China Phase III, ICICI Knowledge Park, Turkapally Village,
TEL: +86-21-5046-0086, FAX: +86-21-5046-2321 Shameerpet Mandal, Ranga Reddy District,
http://www.daicelchina.com/ Hyderabad -500 078, INDIA
E-mail: [email protected] TEL: +91-40-2348-0103, FAX: +91-40-2348-0104
E-mail: dia.com" [email protected]
INSTRUCTION MANUAL FOR
CHIRALPAK® IE COLUMNS
Please read this instruction sheet completely before using this column
<REVERSED PHASE>
IE columns can also be used in normal phase mode.
Please refer to the corresponding instruction sheet for details.
To switch from reversed phase mode to normal phase mode, and vice versa, column should be carefully
flushed with miscible solvent.
It is highly recommended to apply the regeneration procedure described in the instruction sheet for
normal phase mode. Before applying this protocol, any traces of salts should be removed by flushing with a
mobile phase that does not contain any salts / buffers.
ACIDIC
NEUTRAL BASIC
(AMPHOTERIC)
Compounds Compounds
Compounds
20mM NH4HCO3 aq.
Aqueous HCOOH aq. pH 9.0
Water
solution pH 2.0 adjusted with a basic
additive
CHIRALPAK® IE Organic
CH3CN or MeOH or EtOH or IPA or THF
modifier
Typical starting Aqueous solutions 60%
conditions CH3CN 40%
NOTE 1: If you cannot achieve sufficient resolution, try the complementary aqueous solutions
NOTE 2: The concentration of all the buffering salt should be less than 500mM.
Refer to section C for preparation of aqueous solution and choice of basic additives.
It is recommended to use CH3CN to start the investigation
The elution power of organic modifiers for these columns is in the descending order of
CH3CN > EtOH > MeOH: 50%CH3CN ≈ 65-70%EtOH ≈ 75-80%MeOH.
The use of other organic solvents –except THF- has not been investigated and could be harmful
to the columns.
The use of alcohols causes the back pressure to be significantly higher compared to CH3CN due
to their high viscosity in mixtures with water.
Retention can be adjusted by changing the proportion of CH3CN. Retention may be very sensitive
to the amount of CH3CN present into the mobile phase.
Lowering the column temperature may increase the retention time and the selectivity.
Increasing the column temperature and decreasing the flow rate may increase the
resolution.
To maximize column life the use of a guard cartridge is essential when basic
conditions are employed.
The use of strong basic conditions (> pH 9) must be avoided, as they are known to damage
the silica gel matrix.
When these columns are used at pH > 7, the temperature should be maintained between
5°C and 25ºC for maximum column life.
High percentages of organic modifier in the mobile phase may precipitate the buffering salt from
the solution, and lead to consequent clogging of the column (refer to the table below).
90 / 10 to 0 / 100 90 / 10 to 15 / 85
Do not use the phosphate buffer for pH > 8. When pH 9 is necessary, use the ammonium
bicarbonate solution or borate buffer for maximum column life.
Any traces of salts should be removed before column storage and /or before switching to 100%
organic solvent (use Water/CH3CN 60:40 (v/v) for instance)