Rosemount 8700 Series Magnetic Flowmeter Systems

Download as pdf or txt
Download as pdf or txt
You are on page 1of 56

Product Data Sheet

00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series
Rosemount 8700 Series
Magnetic Flowmeter Systems
THE 8700 SERIES...
• Rosemount 8712 C/D/U Transmitter - Easiest
to use local operator interface and works with
any manufacturer’s flowtubes
• Rosemount 8712H/8707 High-Signal System
- Pulsed dc solutions for the most demanding
flow measurement applications
• Rosemount 8705 - Fully welded flowtube for
maximum protection (standard ISO lay length)
• Rosemount 8711 - Economical, compact, and
light weight flowtube, provided with alignment
rings for easy installation
• Rosemount 8732/8742 Transmitter -
Integral-mount design eliminates wiring error,
backlit display for easy viewing, and
explosion-proof housing for hazardous
environments
• Rosemount 8742 Transmitter - Utilizing
advanced diagnostics (high process noise
detection, grounding/wiring diagnostic, and
electrode signal fault) increase magmeter
performance

Content
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 4
Product Certifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 19
Dimensional Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 28
Magnetic Flowmeter Sizing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 36
Material Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 38
Ordering Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 39
Rosemount 8712 C/D/U/H and 8732C Configuration Data Sheet . . . . . . . . . . . . . . . . . . . .page 52
Rosemount 8742C Configuration Data Sheet. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 54

Rosemount Magnetic Flowmeters


powers PlantWeb through advanced
diagnostics including grounding and
wiring, high process noise, and electrode fault.

www.rosemount.com
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

Rosemount 8700 Series System Overview


Rosemount 8712C/D/U/H Sanitary Options
The remote mount Rosemount 8712 transmitter features The adapter flanges on the 3-A-compliant 8705 are
an easy to use operator interface. The intuitive 15 actuator designed to attach to a magnetic flowtube of smaller
keypad provides instant access to the most commonly diameter than the process piping to ensure a crevice-free
used functions, and the 2 line by 20 character display interface. The adapters provided are polished internally
provide clear indication. Together they provide fast, and externally. A more robust sanitary wafer flowtube
intuitive, and easy configuration. offering, the Rosemount 8721, has been released
(document number 00813-0100-4901).
The 8712 C/D/U transmitters operate with all Rosemount
magnetic flow meters and the D and U versions provide
universal functionality allowing them to operate other Wafer Flowtubes
manufacturer’s flow tubes. The 8712H high signal The flangeless design of the 8711 wafer flowtube makes it
transmitter provides unprecedented performance in high an economical, compact, and lightweight alternative to
process noise applications. flanged magnetic flowmeters. Alignment rings, provided
with every Rosemount 8711 flowtube, center the flowtube
Rosemount 8732C in the process line and makes the installation easier.
The integral mount Rosemount 8732C transmitter has an
explosion proof housing and full RFI/EMI protection that is
ideal for harsh environment installations where moisture
and contaminant in filtration are possible. With an optional
backlit 2 line by 16 character local operator interface, the
transmitter can be configured by optical switches to
simplify adjustments in hazardous environments without
removing the cover.

High-Signal Magmeter System (1)


The Rosemount 8707 High-Signal Flowtube, used in
conjunction with the 8712H High-Signal Transmitter, forms
the Rosemount High-Signal Magnetic Flowmeter System.
This system provides stable flow measurement in the most
difficult high-noise applications while maintaining the
Rosemount 8714D benefits of dc technology. The increased signal strength of
the Rosemount® high-signal system is made possible
The Rosemount 8714D Calibration Standard attaches to a through a combination of flowtube coil design that
8712C/U, 8732, or 8742C transmitter’s flowtube incorporates the most advance materials and an extremely
connections to ensure traceability to NIST standards and efficient and innovative coil drive circuit. The increased
long-term accuracy of the flowmeter system. The 8714D is signal strength of the Rosemount high-signal system,
not compatible with 8712H high-signal transmitter. coupled with advanced signal processing and superior
filtering techniques, provides the solution to demanding
Flanged Flowtubes flow measurement applications.
All flanged Rosemount flowtubes are fabricated from
stainless and carbon steel and welded to provide a
hermetic seal that protects against moisture and other
contaminants. Sizes range from 1/2 inch (15 mm) to 36 inch
(900 mm). The sealed housing ensures maximum flowtube
reliability by protecting all internal components and wiring (1) The high-signal magmeter system
from the most hostile environments. is not available in Europe.

2
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

Unleash the Power of Magmeter Diagnostics


with FOUNDATION™ Fieldbus
Increase process stability, reduce commissioning time, and FIGURE 1. Diagnostic Messaging Appears in
maintenance costs by unleashing the diagnostics found AMS and local display
only in the Rosemount 8742C magnetic flowmeter
transmitter. The AMSaware 8742 transmitter provides
useful information through a backlit display or AMSinside™
a DeltaV™ control system.
The grounding/wiring diagnostic in the 8742C reduces
commissioning time by detecting poor process grounding.
Proper installation techniques are critical to the success of
any in-line measurement device and proper process
grounding substantially impacts magnetic flowmeter
performance. Once the problem is identified by the

8742-8742011ad
transmitter, a red light is triggered in the transducer block
(Figure 1) and detailed informative help screens list
appropriate recommendations for the grounding/wiring
diagnostic (Figure 2).
Process noise caused by slurries, high consistency pulp
stock, or upstream chemical additions can result in
wavering, unsteady flow measurements. The 8742C high FIGURE 2. Informative help screens provide
process noise diagnostic monitors the signal to noise ratio direction to solve problems
and recommends steps to improve flow measurements. If
the transmitter detects a low signal-to noise ratio, the high Grounding/Wiring Fault
process noise diagnostic is triggered and suggests The transmitter detected high levels of 50/60Hz noise caused by
increasing the coil drive frequency to improve the improper wiring or poor process grounding.
signal-to-noise ratio (Figure 3). By changing the coil drive
1. Connect ground rings, grounding electrode, lining protector,
frequency, a two to four times improvement in signal or grounding straps. Grounding diagrams are included in
stability is typical and results in a more consistent flow the flowtube products manual.
measurement and thus creating an opportunity for tighter 2. Verify flowtube is full.
control.
3. Verify wiring between flowtube and transmitter is prepared
Flowtube failures caused by liner failure, faulty electrode properly. Shielding should be stripped back less than 2.5 cm
(1in.).
connections or internal moisture can result in expensive,
unscheduled shut-downs. In a standard magmeter 4. Use separate shielded twisted pairs for wiring between
flowtube and transmitter.
installation, this type of flowtube failure could drop the
output flowrate by 50% without notifying the user. The 5. Properly connect the wiring between the flowtube and the
transmitter on the flowtube.
8742C electrode fault diagnostic analyzes the electrode
signal to monitor status and then produces a detailed, To turn off grounding/wiring diagnostic, go to the diagnostic
informative help screen that lists the appropriate remedies screen in the transducer block properties.
when the diagnostic is triggered.
FIGURE 3. Increasing coil drive frequency increases the
Diagnostics that proactively suggest the optimum software signal to noise ratio
configuration, provide powerful information through AMS
help screens, or a local display have never been available
before. Each diagnostic has unique help screens. These
Flow Signal
diagnostics are only the beginning. The 8742C is the
Amplitude
standard of the future for magnetic flowmeter diagnostics
that increase process stability, reduce commissioning time,
Noise at 5 Hz
8742-8742_05

and maintenance costs.


Noise at
371/2 Hz

Start = 0.00 Hz Stop = 100.00 Hz

3
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

Rosemount 8712C/D/U/H Transmitter Specifications


Functional Specifications Supply Current Requirements
Units powered by 10-30 V dc power supply may draw up to 2 amp
Flowtube Compatibility of current.
Rosemount 8712C is compatible with all Rosemount flowtubes:
8705, 8707, 8711, 8721, and 570TM. The 8712D/U is also 2.0
compatible with AC and DC powered flowtubes of other
manufacturers. The 8712H is only compatible with 8707

Supply Current (Amps)


1.5
High-Signal flowtube.
1.0
Flowtube Coil Resistance
Rosemount 8712C: 25  maximum 0.5
Rosemount 8712D/U: 350  maximum
0
Rosemount 8712H: 12  maximum 10 20 30
Power Supply (Volts)
Flow Rate Range I = 20/V
Capable of processing signals from fluids that are traveling I = Supply current requirement (Amps)
between 0.04 and 30 ft/s (0.01 to 10 m/s) for both forward and V = Power supply voltage (Volts)
reverse flow in all flowtube sizes. Full scale continuously
adjustable between –30 and 30 ft/s (–10 to 10 m/s).
Installation Coordination
Conductivity Limits Installation (overvoltage) Category II
Process liquid must have a conductivity of 5 microsiemens/cm (5
micromhos/cm) or greater for Rosemount 8712C/U. Process liquid Power Consumption
must have a conductivity of 50 microsiemens/cm (50 8712C/U: 20 watts maximum
micromhos/cm) for the 8712H. Excludes the effect of
8712D: 10 watts maximum
interconnecting cable length in remote mount
transmitter installations. 8712H: 300 watts maximum

Power Supply Ambient Temperature Limits


8712C/U: 115 or 230 V ac ±10%, 50–60 Hz or 10–30 V dc Operating
8712D: 90-250 V ac ±10%, 50–60 Hz or 12–42 V dc 8712C/U: –20 to 140 °F (–29 to 60 °C)
8712H: 115 V ac ±10%, 50–60 Hz with local operator interface
–30 to 150 °F (–34 to 66 °C) without local
DC Load Limitations (Analog Output) operator interface
Maximum loop resistance is determined by the voltage level of the
8712D: –20 to 140°F (–29 to 60 °C) with local
external power supply, as described by:
operator interface
–40 to 165°F (–40 to 74°C) without local
1250 operator interface

1000 8712H: –20 to 130 °F (–29 to 54 °C) with or without


Load (Ohms)

local operator interface


Operating
500 Region Storage
–22 to 176 °F (–40 to 80 °C)
0
10.8 42 Humidity Limits
Power Supply (Volts) 0–100% RH at 120 °F (49 °C), decreases linearly to 10% RH at
130 °F (54 °C)
Rmax = 41.7(Vps – 10.8)
Vps = Power Supply Voltage (Volts) Enclosure Ratings
Rmax = Maximum Loop Resistance (Ohms) Enclosure Type 4X, IP65

NOTE
HART Communication requires a minimum loop resistance of 250
ohms.

4
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series
Output Signals Turn-on Time
8712C/U/H: 30 minutes to rated accuracy from power up,
Analog Output Adjustment
5 seconds from power interruption
4–20 mA, jumper-selectable as internally or externally
8712D: 5 minutes to rated accuracy from power up, 5 seconds
powered 5 to 24 V dc; 0 to 1000 Ω load.
from power interruption
Engineering units—lower and upper range values
are user-selectable. Start-up Time
Output automatically scaled to provide 4 mA at lower range 0.2 seconds from zero flow
value and 20 mA at upper range value. Full scale continuously
adjustable between -30 and 30 ft/s (-10 to 10 m/sec), 1 ft/s Low Flow Cutoff
(0.3 m/s) minimum span. Adjustable between 0.04 and 1 ft/s (0.01 and 0.3 m/s). Below
HART Communications, digital flow signal, superimposed on selected value, output is driven to the zero flow rate signal level.
4–20 mA signal, available for control system interface. 250 Ω (8712D is adjustable between 0.01 and 1 ft/s).
required for HART communications.
Overrange Capability
Scalable Frequency Adjustment Signal output will remain linear until 110% of upper range value.
8712C/U/H: 0-1000 Hz, externally powered at 5 to 24 V dc, The signal output will remain constant above these values. Out of
transistor switch closure up to 5.75 w. Pulse value can be set range message displayed on LOI and the HART Communicator.
to equal desired volume in selected engineering units. Pulse
width adjustable from 0.5 to 100 m/s. Local operator interface Damping
automatically calculates and displays maximum allowable 8712C/U/H: Adjustable between 0.2 and 256 seconds
output frequency. 8712D: Adjustable between 0.0 and 256 seconds
8712D: 0-10,000Hz, externally powered at 5 to 24 V dc,
transistor switch closure supports power loads up to 2W for Flowtube Compensation
frequencies up to 4000Hz, and 5 V dc at 0.1 W at maximum Rosemount flowtubes are flow-calibrated and assigned a
frequency of 10,000 Hz. Pulse can be set to equal desired calibration factor at the factory. The calibration factor is entered
velocity or volume in user selectable engineering units. Pulse into the transmitter, enabling interchangeability of flowtubes
width is adjustable from 1.5 to 500 msec, below 1.5 msec without calculations or a compromise in accuracy.
pulse width automatically switches to 50% duty cycle. 8712U/D transmitters and other manufacturer’s flowtubes can be
calibrated at known process conditions or at the Rosemount
Auxiliary Output Function NIST-Traceable Flow Facility. Transmitters calibrated on site
Externally powered at 5 to 24 V dc, transistor switch closure require a two-step procedure to match known flow rate.
up to 3 W to indicate either:
Reverse Flow: Activates switch closure output when reverse
flow is detected. The reverse flow rate is displayed.
Zero Flow: Activates switch closure output when flow
goes to 0 ft/s.

Positive Zero Return


Forces outputs of the transmitter to the zero flow rate signal
level. Activated by applying a contact closure.

Software Lockout
Security lockout jumper on the electronics board can be set to
deactivate all LOI and HART-based communicator functions to
protect configuration variables from unwanted or
accidental change.

Output Testing
Analog Output Test
Transmitter may be commanded to supply a specified current
between 3.75 and 23.25 mA

Pulse Output Test


8712C/U/H: Transmitter may be commanded to supply a
specified frequency between 1 and 1000 Hz
8712D: Transmitter may be commanded to supply a specified
frequency between 1 pulse/ day and 10,000 Hz

5
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

Performance Specifications Ambient Temperature Effect


8712C/U/H: ±1% per 100 °F (37.8 °C)
(System specifications are given using the frequency output and
with the unit at referenced conditions.) 8712D: 0.25% over operating temperature range

Accuracy RFI Effect


Class 1, A, B, C: ±0.5% of span at 3 V/m per SAMA PMC 33.1,
Rosemount 8712C/U with 8705 Flowtube wires and conduit
System accuracy is ±0.5% of rate from 1 to 30 ft/s (0.3 to 10 EN61326-1 1997 + A1/A2 (Industrial) electromagnetic
m/s); between 0.04 and 1.0 ft/s (0.01 and 0.3 m/s), the system compatibility (EMC) for process and laboratory apparatus.
has an accuracy of ±0.005 ft/s (0.0015 m/s). Analog output
has the same accuracy as frequency output plus an additional Supply Voltage Effect
0.1% of span. Transmitter meets supply voltage effect requirements of SAMA
PMC 31.1, Section 5.10.1 through 5.10.5. Transmitter withstands
Rosemount 8712H with 8707 Flowtube surges in supply voltage as specified in IEEE 472, 1974
System accuracy is ±0.5% of rate from 3 to 30 ft/s (1 to
10 m/s); between 0.04 and 3.0 ft/s (0.01 and 0.3 m/s), the Physical Specifications
system has an accuracy of ±0.005 ft/s (0.0015 m/s). Analog
output has the same accuracy as frequency output plus an
Materials of Construction
additional 0.1% of span.
Housing
Rosemount 8712C/U with 8711 Flowtube
Low-copper aluminum, NEMA 4X and IEC 60529 IP65
System accuracy is ±0.5% of rate from 3 to 30 ft/s (1 to 10
Pollution Degree 2
m/s); below 3 ft/s (1 m/s), the system has an accuracy of
±0.015 ft/s (0.005 m/s). Analog output has the same accuracy Paint
as frequency output plus an additional 0.1% of span.
Polyurethane
Rosemount 8712U/D with Other Manufacturers’
Cover Gasket
Flowtubes
Rubber
When calibrated in the Rosemount Flow Facility, system
accuracies as good as 0.5% of rate can be attained. Analog
Electrical Connections
output has the same accuracy as frequency output, plus an
Three ¾–14 NPT connections provided on the base of the
additional 0.1% of span.
transmitter. Screw terminals provided for all of the connections.
There is no accuracy specification for other manufacturers’ Power wiring connected to the transmitter only. Remote mounted
flowtubes calibrated in the process line. transmitters require only a single conduit connection to the
flowtube. Integrally mounted transmitters are factory wired to
Rosemount 8712D with 8705 and 8721 Flowtube
the flowtube.
System accuracy is ±0.5% of rate from 1 to 40 ft/s (0.3 to 12
m/s); between 0.04 and 1.0 ft/s (0.01 and 0.3 m/s), the system
has an accuracy of ±0.005 ft/s (0.0015 m/s). Analog output
Line Power Fuses
has the same accuracy as frequency output plus an additional
0.05% of span. 115 V ac systems
1 amp, Quick-acting Bussman AGCI or equivalent.
Rosemount 8712D with 8711 Flowtube
5 amp, Quick-acting Bussman AGCI or equivalent
System accuracy is ±0.5% of rate from 3 to 40 ft/s (1 to 12 (Rosemount 8712H only).
m/s); below 3 ft/s (1 m/s), the system has an accuracy of
±0.015 ft/s (0.005 m/s). Analog output has the same accuracy 230 V ac systems
as frequency output plus an additional 0.1% of span. ½ amp, Quick-acting Bussman AGCI or equivalent

Vibration Effect 90–250 V ac systems (8712D)


±0.1% of span per SAMA PMC 31.1, Level 2 1 amp, Quick-acting Bussman AGCI or equivalent

Repeatability 10–30 V dc systems


±0.1% of reading 3 amp, Quick-acting Bussman AGCI or equivalent

Response Time Transmitter Weight


0.2 seconds maximum response to step change in input Transmitter approximately 9 lb (4 kg). Add 1 lb (0.5 kg) for local
operator interface.
Stability
±0.1% of rate over six months

6
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

Rosemount 8732C Transmitter Specifications


Functional Specifications Supply Current Requirements
Units powered by 15-50 V dc power supply may draw up to 1 amp
Flowtube Compatibility of current.
Compatible with Rosemount 8705, 8711, and 8721 flowtubes. 1.0

Flowtube Coil Resistance

Supply Current (Amps)


0.75
25 Ω maximum
0.5
Flow Rate Range
Capable of processing signals from fluids that are traveling
between 0.04 and 30 ft/s (0.01 to 10 m/s) for both forward and 0.25
reverse flow in all flowtube sizes. Full scale continuously
0
adjustable between –30 and 30 ft/s (–10 to 10 m/s). 15 20 30 40 50
Power Supply (Volts)
Conductivity Limits I = 10/V
Process liquid must have a conductivity of 5 microsiemens/cm (5 I = Supply current requirement (Amps)
micromhos/cm) or greater for 8732C. Excludes the effect of V = Power supply voltage (Volts)
interconnecting cable length in remote mount
transmitter installations. Installation Coordination
Installation (overvoltage) Category II
Power Supply
85 -250 V ac ±10%, 50–60 Hz or 15-50 V dc Power Consumption
10 watts maximum
DC Load Limitations (Analog Output)
Maximum loop resistance is determined by the voltage level of the Ambient Temperature Limits
external power supply, as described by:
Operating
1250 –40 to 165 °F (–40 to 74 °C) without local operator interface
13 to 149 °F (–25 to 65 °C) with local operator interface
1000
Load (Ohms)

Storage
Operating –40 to 185 °F (–40 to 85 °C)
500
Region
Humidity Limits
0–100% RH to 150 °F (65 °C)
0
10.8 42
Power Supply (Volts) Enclosure Rating
NEMA 4X CSA Type 4X, IEC 60529, IP67 (transmitter),
Rmax = 41.7(Vps – 10.8) Pollution Degree 2
Vps = Power Supply Voltage (Volts)
Rmax = Maximum Loop Resistance (Ohms)

NOTE
HART Communication requires a minimum loop resistance of 250
ohms.

7
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

Output Signals Software Lockout


Security lockout switch on the electronics board can be set to
Analog Output Adjustment
deactivate all LOI and HART-based communicator functions to
4–20 mA, jumper-selectable as internally or externally protect configuration variables from unwanted or
powered 5 to 24 V dc; 0 to 1000 Ω load. accidental change.
Engineering units—lower and upper range values
are user-selectable. Display Lockout
Output automatically scaled to provide 4 mA at lower range All optical switches on the display can be locked locally from the
value and 20 mA at upper range value. Full scale continuously display layout configuration screen by holding the upper right
adjustable between -30 and 30 ft/s (-10 to 10 m/sec), 1 ft/s optical switch for 20 seconds. The display can be reactivated
(0.3 m/s) minimum span. holding the same switch for 20 seconds.
HART Communications, digital flow signal, superimposed on
4–20 mA signal, available for control system interface. 250 Ω
Output Testing
required for HART communications. Analog Output Test
Transmitter may be commanded to supply a specified current
Scalable Frequency Adjustment
between 3.75 and 23.25 mA.
0-1000 Hz, externally powered at 5 to 24 V dc, translator
switch closure up to 5.75 w. Pulse value can be set to equal Pulse Output Test
desired volume in selected engineering units. Pulse width
Transmitter may be commanded to supply a specified
adjustable from 0.5 to 100 m/s. Local operator interface frequency between 1 and 1000 Hz.
automatically calculates and displays maximum allowable
output frequency. Turn-on Time
30 minutes to rated accuracy from power up;
Auxiliary Output Function
5 seconds from power interruption
Externally powered at 5 to 24 V dc, transistor switch closure
up to 3 W to indicate either: Start-up Time
Reverse Flow: Activates switch closure output when reverse 0.2 seconds from zero flow
flow is detected. The reverse flow rate is displayed.
Zero Flow: Activates switch closure output when flow Low Flow Cutoff
goes to 0 ft/s. Adjustable between 0.001 and 1 ft/s (0.0003 and 0.3 m/s). Below
selected value, output is driven to the zero flow rate signal level.

Overrange Capability
Signal output will remain linear until 110% of upper range value or
33 ft/s. The signal output will remain constant above these values.
Out of range message displayed on LOI and the
HART Communicator.

Damping
Adjustable between 0.2 and 256 seconds

Flowtube Compensation
Rosemount flowtubes are flow-calibrated and assigned a
calibration factor at the factory. The calibration factor is entered
into the transmitter, enabling interchangeability of flowtubes
without calculations or a compromise in accuracy.

8
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

Performance Specifications Physical Specifications


(System specifications are given using the frequency output and
with the unit at reference conditions.) Materials of Construction
Housing
Accuracy
Low copper aluminum, NEMA 4X and IEC 60529 IP66
Rosemount 8732C with 8705 or 8721 Flowtubes: Pollution Degree 2
System accuracy is ±0.5% of rate from 1 to 30 ft/s (0.3 to 10
m/s); below 1.0 ft/s (0.3 m/s), the system has an accuracy of Paint
±0.005 ft/s (0.0015 m/s). Analog output has the same Polyurethane
accuracy as frequency output plus an additional 0.1% of span.
Cover Gasket
Rosemount 8732C with 8711 Flowtube: Rubber
System accuracy is ±0.5% of rate from 3 to 30 ft/s (0.9 to 10
m/s); below 3 ft/s (0.9 m/s), the system has an accuracy of Electrical Connections
±0.015 ft/s (0.005 m/s). Analog output has the same accuracy Two ¾–14 NPT connections provided on the transmitter housing.
as frequency output plus an additional 0.1% of span. PG13.5 and CM20 adapters are available. Screw terminals
provided for all connections. Power wiring connected to transmitter
Vibration Effect only. Integrally mounted transmitters are factory wired to the
IEC 60770-1 flowtube.

Repeatability Mounting
±0.1% of reading Transmitter is mounted integrally with the flowtube and does not
require interconnecting cables.
Response Time
0.2 seconds maximum response to step change in input Transmitter Weight
Approximately 7 pounds (3.2 kg). Add 1 pound (0.5 kg) for Option
Stability Code M4.
±0.1% of rate over six months

Ambient Temperature Effect


±0.25% change over operating temperature range

EMC Compliance
EN61326-1 1997 + A1/A2 (Industrial) electromagnetic
compatibility (EMC) for process and laboratory apparatus.

9
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

Rosemount 8742C Transmitter Specifications


Functional Specifications Enclosure Rating
NEMA 4X CSA Type 4X, IEC 60529, IP67 (transmitter),
Flowtube Compatibility Pollution Degree 2
Compatible with Rosemount 8705, 8711, and 8721 flowtubes.
Output Signal
Conductivity Limits Manchester-encoded digital signal that conforms to IEC 1158-2
Process liquid must have a conductivity of 5 microsiemens/cm (5 and ISA 50.02
micromhos/cm) or greater for 8742C. Excludes the effects of
interconnecting cable length in remote mount transmitter FOUNDATION Fieldbus Specifications
installations.
Schedule Entries
Flowtube Coil Resistance Seven (7)
25  maximum
Links
Flow Rate Range Twenty (20)
Capable of processing signals from fluids that are traveling
between 0 and 40 ft/s (0 to 12 m/s) for both forward and reverse Virtual Communications Relationships (VCRs)
flow in all flowtube sizes. One (1) predefined (F6, F7)
Nineteen (19) configurable (see Table 1)
Power Supply
90–250 V ac, 50–60 Hz, 15–50 V dc TABLE 1. Block Information
Execution Time
Supply Current Requirements Block Base Index (Milliseconds)
Units powered by 15-50 V dc power supply may draw up to 1 amp
of current. Resource (RB) 300 —
Transducer (TB) 400 —
1.0 Analog Input (AI) 1,000 15
Proportional/Integral/ 10,000 25
Supply Current (Amps)

0.75
Derivative (PID)
Intergrator 12,000 20
0.5
Reverse Flow
0.25 Detects and reports reverse flow
0
15 20 30 40 50 Software Lockout
Power Supply (Volts) A write-lock switch and software lockout are provided in the
I = 10/V resource function block.
I = Supply current requirement (Amps)
V = Power supply voltage (Volts) Turn-on Time
30 minutes to rated accuracy from power up;
Installation Coordination 10 seconds communication from power interruption
Installation (overvoltage) Category II
Start-up Time
Power Consumption 0.2 seconds from zero flow
10 watts maximum
Low Flow Cutoff
Ambient Temperature Limits Adjustable between 0 and 1 ft/s. Below selected value, output is
driven to the zero flow rate.
Operating
–40 to 165 °F (–40 to 74 °C) without local operator interface Overrange Capability
–13 to 147 °F (–25 to 65 °C) with local operator interface Signal output continues to 110% of upper range value setting, then
remains constant. Messages are supplied to the fieldbus network.
Storage
–40 to 185 °F (–40 to 85 °C) Damping
Adjustable between 0 and 255 seconds
Humidity Limits
0–100% RH to 150 °F (65 °C)

10
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

Performance Specifications Physical Specifications


(System specifications are given using the unit at referenced
conditions.) Materials of Construction (Transmitter)
Housing
Accuracy
Low-copper aluminum
Rosemount 8742C Transmitter with
8705 and 8721 Flowtubes Paint
System accuracy is ±0.5% of rate from 1 to 40 ft/s (0.3 to 12 Polyurethane
m/s); includes combined effects of linearity, hysteresis,
Cover Gasket
repeatability, and calibration uncertainty; below 1.0 ft/s (0.3
m/s), the system has an accuracy of ±0.005 ft/s (0.0015 m/s). Rubber

Rosemount 8742C Transmitter with 8711 Flowtube Electrical Connections


System accuracy is ±0.5% of rate from 3 to 40 ft/s (0.9 to 12 Two ¾–14 NPT with number 8 screw terminal connections are
m/s); includes combined effects of linearity, hysteresis, provided for electrical wiring. PG13.5 and CM20 adapters are
repeatability, and calibration uncertainty; below 3 ft/s (0.9 m/s), available. Screw terminals provided for all connections. Power
the system has an accuracy of ±0.015 ft/s (0.005 m/s). wiring connected to transmitter only. Integrally mounted
transmitters are factory wired to the flowtube.
Vibration Effect
IEC 60770-1 Mounting
Integrally mounted transmitters do not require interconnecting
Repeatability cables. The local display and transmitter can be rotated in 90°
±0.1% of reading increments. Remote mounted transmitters require only a single
conduit connection to the flowtube.
Response Time
0.2 seconds maximum response to step change in input Weight
Approximately 7 pounds (3.2 kg). Add 0.5 pounds (0.5 kg) for local
Stability display.
±0.1% of rate over six months

Ambient Temperature Effect


±0.25% of rate over operating temperature range

EMC Compliance
EN61326-1 1997 + A1/A2 3(Industrial) electromagnetic
compatibility (EMC) for process and laboratory apparatus.

Transient Protection
The 8742C transmitter prevents damage to the flowmeter from
transients compliant to:
IEC 6100 – 4-4 (for burst currents)
IEC 6100 – 4-5 (for surge currents)

Mounting Position Effect


None when installed to ensure flowtube remains full

11
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

Rosemount 8705 Flanged and 8707 High-Signal Flanged


Flowtubes Specifications
Functional Specifications Ambient Temperature Limits
–30 to 150 °F (–34 to 65 °C)
Service
Conductive liquids and slurries Pressure Limits
See Table 2 and Table 4
Line Sizes
1
/2–36 inch (15–900 mm) for Rosemount 8705 Vacuum Limits
3–36 inch (80–600 mm) for Rosemount 8707 Teflon (PTFE) Lining
Full vacuum to 350 °F (177 °C) through 4-inch (100 mm) line
Interchangeability sizes. Consult factory for vacuum applications with line sizes
Rosemount 8705 Flowtubes are interchangeable with 8712C/U, of 6 inches (150 mm) or larger.
8732, and 8742C Transmitters. Rosemount 8707 High-Signal
Flowtubes are interchangeable with 8712H High-Signal All Other Standard Flowtube Lining Materials
Transmitters. System accuracy is maintained regardless of line Full vacuum to maximum material temperature limits for all
size or optional features. Each flowtube nameplate has a available line sizes.
sixteen-digit calibration number that can be entered into a
transmitter through the Local Operator Interface (LOI) or the HART Submergence Protection
Communicator on the 8712C/U/H and the 8732C. In a IP 68. Continuous to 30 feet (10 meters)
FOUNDATION™ fieldbus environment, the 8742C can be configured
using the DeltaV™ fieldbus configuration tool or another Conductivity Limits
FOUNDATION fieldbus configuration device. No further calibration Process liquid must have a conductivity of 5 microsiemens/cm (5
is necessary. micromhos/cm) or greater for 8705. Process liquid must have a
conductivity of 50 microsiemens/cm (50 micromhos/cm) for 8707
Upper Range Limit when used with 8712H, 5 microsiemens/cm when used with other
30 ft/s (10 m/s) transmitters. Excludes the effect of interconnecting cable length in
remote mount transmitter installations.
Process Temperature Limits
Teflon (PTFE) Lining
–20 to 350 °F (–29 to 177 °C)

Tefzel (ETFE) Lining


–20 to 300 °F (–29 to 149 °C)

PFA Lining
-20 to 350 °F (-29 to 177°C)

Polyurethane Lining
0 to 140 °F (–18 to 60 °C)

Neoprene Lining
0 to 185 °F (–18 to 85 °C)

Linatex Lining
0 to 158 °F (–18 to 70°C)

12
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

TABLE 2. Temperature vs. Pressure Limits(1)


Flowtube Temperature vs. Pressure Limits for ASME B16.5 Class Flanges (1/2- to 36-inch line sizes) (2)
Pressure
@ -20 to 100 °F @ 200 °F @ 300 °F @ 350 °F
Flange Material Flange Rating (-29 to 38 °C) (93 °C) (149 °C) (177 °C)
Class 150 285 psi 260 psi 230 psi 215 psi
Class 300 740 psi 675 psi 655 psi 645 psi
Carbon Steel
Class 600 (3) 1000 psi 800 psi 700 psi 650 psi
Class 600 (4) 1480 psi NA NA NA
Class 900 2220 psi
Class 150 275 psi 235 psi 205 psi 190 psi
Class 300 720 psi 600 psi 530 psi 500 psi
304 Stainless Steel
Class 600 (5) 1000 psi 800 psi 700 psi 650 psi
Class 600 (6) 1440 psi NA NA NA
Class 900 2160 psi
(1) Liner temperature limits must also be considered. Polyurethane, Linatex, and Neoprene have temperature limits of 140°F, 158°F, and 185°F, respectively.

(2) 30- and 36-inch AWWA C207 Table 3 Class D rated to 150 psi at atmospheric temperature.

(3) Option Code C6

(4) Option Code C7

(5) Option Code S6

(6) Option Code S7

TABLE 3. Temperature vs. Pressure Limits (1)


Flowtube Temperature vs. Pressure Limits for AS2129 Table D and E Flanges (4- to 24-inch line sizes)
Pressure
@ -200 to 50 °F @ 100 °F @ 150°F @ 200 °F
Flange Material Flange Rating (-320 to 122 °C) (212 °C) (302 °C) (392 °C)
D 101.6 psi 101.6 psi 101.6 psi 94.3 psi
Carbon Steel
E 203.1 psi 203.1 psi 203.1 psi 188.6 psi
(1) Liner temperature limits must also be considered. Polyurethane, Linatex, and Neoprene have temperature limits of 140°F, 158°F, and 185°F, respectively.

TABLE 4. Temperature vs. Pressure Limits (1)


Flowtube Temperature vs. Pressure Limits for DIN Flanges (15 to 600 mm line sizes)
Pressure
@ -196 to 50 °C @ 100 °C @ 150°C @ 175°C
Flange Material Flange Rating (-320 to 122 °F) (212 °F) (302 °F) (347 °F)
PN 10 10 bar 10 bar 9.7 bar 9.5 bar
PN 16 16 bar 16 bar 15.6 bar 15.3 bar
Carbon Steel
PN 25 25 bar 25 bar 24.4 bar 24.0 bar
PN 40 40 bar 40 bar 39.1 bar 38.5 bar
PN 10 9.1 bar 7.5 bar 6.8 bar 6.5 bar
PN 16 14.7 bar 12.1 bar 11.0 bar 10.6 bar
304 Stainless Steel
PN 25 23 bar 18.9 bar 17.2 bar 16.6 bar
PN 40 36.8 bar 30.3 bar 27.5 bar 26.5 bar
(1) Liner temperature limits must also be considered. Polyurethane, Linatex, and Neoprene have temperature limits of 140°F, 158°F, and 185°F, respectively.

13
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

Performance Specifications Process Connections


(System specifications are given using the frequency output and ASME B16.5 (ANSI) Class 150, Class 300,
with the unit at referenced conditions.) Class 600, or Class 900
0.5- to 30-inch (Class 150)
Accuracy
0.5- to 24-inch (Class 300)
Rosemount 8705 with 8712D/C/U, 8732C, or 8742C 0.5- to 10-inch (Class 600 derated to 1000 psi max)
±0.5% of rate from 1 to 30 ft/s (0.3 to 10 m/s). Includes 1- to 8-inch (Full rated Class 600 and 900)
combined effects of linearity, hysteresis, repeatability, and
calibration uncertainty. Accuracy is ±0.005 ft/s (±0.0015 m/s) AWWA C207 Table 3 Class D
from low-flow cutoff to 1.0 ft/s (0.3 m/s). 30- and 36-inch

Rosemount 8707 with 8712D/C/U/H, 8732C, or 8742C DIN PN 10, 16, 25, and 40
±0.5% of rate from 3 to 30 ft/s (1 to 10 m/s). Include combined PN10: Not available for flange sizes from 15 to 150 mm
effects of linearity, hysteresis, repeatability and calibration PN16: Not available for flange sizes from 15 to 80 mm
uncertainty. Accuracy is ±0.015 ft/s (±0.0045 m/s) from
PN 25: Not available for flange sizes from 15 to 150 mm
low-flow cutoff to 3.0 ft/s (1 m/s).
PN40: Available for all flange sizes
Vibration Effect
AS 2129 Table D and E
IEC 60770-1
0.5- to 36 inch
Mounting Position Effect
AISI Type 304 SST Sanitary Tri-Clover
None when installed to ensure flowtube remains full
3-A-compliant quick disconnect ferrule-mounted to ASME
B16.5 (ANSI) Class 150 flange; 0.5- to 3-inch.
Physical Specifications
Electrical Connections
Non-Wetted Materials Two ¾–14 NPT connections with number 8 screw terminals are
Flowtube provided in the terminal enclosure for electrical wiring.
AISI Type 304 SST
Grounding Electrode
Flanges A grounding electrode is installed similarly to the measurement
electrodes through the flowtube lining on 8705 flowtubes. It is
Carbon steel, AISI Type 304/304L SST,
available in all electrode materials.
or Type 316/316L SST

Housing Grounding Rings


Grounding rings are installed between the flange and the tube face
Welded steel
on both ends of the flowtube. Single ground rings can be installed
Paint on either end of the flowtube. They have an I.D. slightly larger than
the flowtube I.D. and an external tab to attach ground wiring.
Polyurethane
Grounding rings are available in 316L SST, Hastelloy-C, titanium,
and tantalum.
Process Wetted Materials
Lining Lining Protectors
PFA, Teflon (PTFE), Tefzel (ETFE), polyurethane, neoprene, Lining protectors are installed between the flange and the tube
Linatex face on both ends of the flowtube. The leading edge of lining
material is protected by the lining protector; lining protectors
Electrodes cannot be removed once they are installed. Lining protectors are
316L SST, Hastelloy C-276, tantalum, 90% platinum-10% available in 316L SST, Hastelloy-C, and titanium.
iridium, titanium
Dimensions
See Figure 6, Figure 7, and Figure 8 and Table 12, Table 13, and
Table 14.

Weight
See Table 5 and Table 6

14
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

TABLE 5. Flowtube Weight (ASME)


Flowtube Flange Rating
Flowtube Weight
Nominal Line Size(1) Inches (mm) ASME B16.5 (ANSI) DIN lb (kg)
½ (15) 150 PN 40 20 (9)
½ (15) 300 22 (10)
1 (25) 150 PN 40 20 (9)
1 (25) 300 22 (10)
1½ (40) 150 PN 40 22 (10)
1½ (40) 300 24 (11)
2 (50) 150 PN 40 26 (12)
2 (50) 300 28 (13)
3 (80) 150 PN 40 40 (18)
3 (80) 300 47 (21)
4 (100) 150 PN 16 48 (22)
4 (100) 300 65 (30)
6 (150) 150 PN 16 81 (37)
6 (150) 300 93 (42)
8 (200) 150 PN 10 110 (50)
8 (200) 300 162 (74)
10 (250) 150 PN 10 220 (98)
10 (250) 300 300 (136)
12 (300) 150 PN 10 330 (150)
12 (300) 300 435 (197)
14 (350) 150 PN 10 370 (168)
16 (400) 150 PN 10 500 (227)
18 (450) 150 PN 10 600 (272)
20 (500) 150 PN 10 680 (308)
24 (600) 150 PN 10 1,000 (454)
30 (750) 150 - 1,747 (792)
36 (900) 125 - 1,975 (898)
(1) 30- and 36-inch AWWA C207 Table 2 Class D rated to 150 psi at atmospheric temperature.

TABLE 6. Flowtube weights (AS2129)


Nominal Line Size Inches (mm) AS2129 Flowtube Weight lb (kg)
4 (100) D 33 (15)
4 (100) E 37 (17)
6 (150) D 66 (30)
6 (150) E 71 (32)
8 (200) D 86 (39)
8 (200) E 88 (40)
10 (250) D 187 (85)
10 (250) E 201 (91)
12 (300) D 273 (124)
12 (300) E 284 (129)
14 (350) D 293 (133)
14 (350) E 317 (144)
16 (400) D 386 (175)
16 (400) E 430 (195)
18 (450) D 516 (234)
18 (450) E 569 (258)
20 (500) D 569 (258)
20 (500) E 626 (284)
24 (600) D 855 (388)
24 (600) E 974 (442)

15
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

Rosemount 8711 Wafer Flowtube Specifications


SPECIFICATIONS Performance Specifications
(System specifications are given using the frequency output and
Functional Specifications with the unit at referenced conditions.)

Service Accuracy
Conductive liquids and slurries Rosemount 8711 with 8712C/U, 8732C, or
8742C Transmitters
Line Sizes
±0.5% of rate from 3 to 30 ft/s (1 to 10 m/s)
0.15- through 8-inch (4 through 200 mm)
±0.015 ft/s (0.0045 m/s) from low-flow cutoff to 3 ft/s (1 m/s)
Interchangeability
Rosemount 8711 Flowtubes are interchangeable with 8712C/U, Vibration Effect
8732, and 8742C Transmitters. System accuracy is maintained IEC 60770-1
regardless of line size or optional features. Each flowtube
nameplate has a sixteen-digit calibration number that can be Mounting Position Effect
entered into a transmitter through the Local Operator Interface No effect when installed to ensure flowtube remains full
(LOI) or the HART Communicator on the Rosemount 8712C/U/H
and the 8732C. In a FOUNDATION fieldbus environment, the 8742C Physical Specifications
can be configured using the DeltaV fieldbus configuration tool or
another FOUNDATION fieldbus configuration device. No further Non-Wetted Materials
calibration is necessary.
Flowtube
Upper Range Limit 303 SST (ASTM A-743)
30 ft/s (10 m/s)
Coil Housing
Process Temperature Limits Investment cast steel (ASTM A-27)

Tefzel (ETFE) Lining Paint


–20 to 300 °F (–29 to 149 °C) for 0.5- through 8-inch Polyurethane
(15–200 mm) line sizes
–20 to 200 °F (–29 to 93 °C) for 0.15- and 0.3-inch Process-Wetted Materials
(4 and 8 mm) line sizes
Lining
Teflon (PTFE) Lining Tefzel (ETFE), Teflon (PTFE)
-20 to 350 °F (-29 to 177 °C)
Electrodes
Ambient Temperature Limits 316L SST, Hastelloy C-276, tantalum,
–30 to 150 °F (–34 to 65 °C) 90% platinum—10% iridium, titanium

Maximum Safe Working Pressure at 100 °F (38 °C)


Tefzel (ETFE) Lining
Full vacuum to 740 psi (5.1 MPa) for 0.5- through 8-inch
(15 through 200 mm) flowtubes
285 psi (1.96 MPa) for 0.15- and 0.30-inch
(4 and 8 mm) flowtubes

Teflon (PTFE) Lining


Full vacuum through 4-inch (100 mm) line sizes. Consult
factory for vacuum applications with line sizes of 6 inches (150
mm) or larger.

Conductivity Limits
Process liquid must have a conductivity of 5 microsiemens/cm (5
micromhos/cm) or greater for 8711. Excludes the effect of
interconnecting cable length in remote mount
transmitter installations.

16
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series
Process Connections
Mounts between these Flange Configurations
ASME B16.5 (ANSI): Class 150, 300
DIN: PN 10 and 25
BS: 10 Table D, E, and F

Studs, Nuts, and Washers(1)


ASME B16.5 (ANSI)
0.15- through 1-inch (4 through 25 mm):
316 SST, ASTM A193, Grade B8M, Class 1 threaded
mounting studs; ASTM A194, Grade 8M heavy hex nuts; SAE
per ANSI B18.2.1, Type A, Series N flat washers.
1.5- through 8-inch (40 through 200 mm):
CS, ASTM A193, Grade B7, Class 1 threaded mounting studs;
ASTM A194, Grade 2H heavy hex nuts; SAE per ANSI
B18.2.1, Type A, Series N flat washers; all items clear,
chromate zinc-plated.
DIN
4 through 25 mm (0.15- through 1-inch):
316 SST ASTM A193, Grade B8M Class 1 threaded mounting
studs; ASTM A194, Grade 8M, DIN 934 H=D, metric heavy
hex nuts; 316 SST, A4, DIN 125 flat washers.
40 through 200 mm (1.5- through 8-inch):
CS, ASTM A193, Grade B7 threaded mounting studs; ASTM
A194, Grade 2H, DIN 934 H=D, metric heavy hex nuts; CS,
DIN 125 flat washers; all items yellow zinc-plated.

Electrical Connections
Two ¾–14 NPT connections with number 8 screw terminals are
provided in the terminal enclosure for electrical wiring.

Grounding Electrode
A grounding electrode is installed similarly to the measurement
electrodes through the flowtube lining. It is available in all
electrode materials.

Grounding Rings
Grounding rings are installed between the flange and the tube face
on both ends of the flowtube. They have an I.D. slightly larger than
the flowtube I.D. and an external tab to attach ground wiring.
Grounding rings are available in 316L SST, Hastelloy C-276,
titanium, and tantalum.

Dimensions and Weight


See Figure 9, Figure 10, and Table 15

(1) 0.15 and 0.30 inch (4 and 80 mm) flowtubes mount


between 1/2-inch flange.

17
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

Rosemount 8714D Reference Calibration Standard


Specifications
Functional Specifications
Ambient Temperature Limits
Operating
–30 to 140 °F (–34 to 60 °C)

Storage
–40 to 140 °F (–40 to 60 °C)

Humidity Limits
0 to 95% relative humidity

Performance Specifications
Accuracy
±0.05% of rate at 30 ft/s at 25°C
±0.10% of rate at 10 ft/s and 3 ft/s

Warm-up Time
30 minutes

Ambient Temperature Effect


< 0.015% of rate per 10 °F (< 0.027% per 10 °C)

Humidity Effect
No effect from 0 to 60% relative humidity
< 0.10% of rate from 60 to 90% relative humidity

Long-Term Stability
< 0.10% of rate shift in one year

Physical Specifications
Electrical Connections
Electrical connections are compatible with 8712C/U or 8732C
terminal blocks. Electrical connections are not compatible with
8712H terminal block.

Mounting
Any position is acceptable

Materials of Construction
Housing
Extruded aluminum

Covers
Stamped aluminum, silk-screened

Paint
Polyurethane

Weight
Approximately 3 lb (2 kg)

18
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

Product Certifications
Approved Manufacturing Locations Electro Magnetic Compatibility (EMC)
Rosemount Inc. — Eden Prairie, Minnesota, USA (89/336/EEC)
Fisher-Rosemount Technologias de Flujo, S.A. de C.V. — All Models EN 50081-1: 1992, EN 50082-2: 1995,
Chihuahua, Chihuahua, Mexico EN 61326: 1997/ A1:1998 / A2:2000
Installed signal wiring should not be run together and should not
EUROPEAN DIRECTIVE INFORMATION be in the same cable tray as AC power wiring.

The EC declaration of conformity for all applicable European Device must be properly grounded or earthed according to local
electric codes.
directives for this product can be found on our website at
www.rosemount.com. A hard copy may be obtained by contacting To improve protection against signal interference, shielded cable is
our local sales office. recommended.

ATEX Directive Low Voltage Directive (93/68/EEC)


Rosemount Inc. complies with the ATEX Directive. 8712C/U and All Models 8732, Model 8742C - AC
Type n protection type in accordance with EN50 021 EN 61010-1: 1995
• Closing of entries in the device must be carried out using the
appropriate EExe or EExn metal cable gland and metal Other important guidelines
blanking plug or any appropriate ATEX approved cable gland Only use new, original parts.
and blanking plug with IP66 rating certified by an EU approved To prevent the process medium escaping, do not unscrew or
certification body. remove process flange bolts, adapter bolts or bleed screws during
operation.
European Pressure Equipment Directive Maintenance shall only be done by qualified personnel.
(PED) (97/23/EC)
CE CE Marking (8712C/U Only)
Model 8705 Magnetic Flowmeter flowtubes in line size Compliance with European Union EMC and Low Voltage
and flange combinations: Directives. (Note: CE Marking is not available on Model
8712H).
Line Size: 1 1/2 inch - 3 inch with all flanges available.
Line Size: 4 inch - 24 inch with all DIN flanges and ANSI 150 and
ANSI 300 flanges.
Line Size: 30 inch - 36 inch with AWWA 125 flanges
QS Certificate of Assessment - EC No. PED-H-20
Module H Conformity Assessment

Model 8711 Magnetic Flowmeter Flowtubes


Line Sizes: 1.5, 2, 3, 4, 6, and 8 inch
QS Certificate of Assessment - EC No. PED-H-20
Module H Conformity Assessment

Model 8721 Sanitary Magmeter Flowtubes


in line sizes of 11/2 inch and larger:
Module A Conformity Assessment

All other Model 8705/8711/8721 Flowtubes —


Sound Engineering Practice
Flowtubes that are SEP are outside the scope of PED and cannot
be marked for compliance with PED.
Mandatory CE-marking for flowtubes in accordance with Article 15
of the PED can be found on the flowtube body (CE 0575).
Flowtube category I is assessed for conformity per module A
procedures.
Flowtube categories II – IV, use module H for conformity
assessment procedures.

19
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

HAZARDOUS LOCATION CERTIFICATIONS


Equivalent Hazardous Location Certifications for flowtube and
transmitter must match in integrally-mounted magnetic flowmeter
systems. Remote-mounted systems do not require matched
hazardous location certification option codes.

Transmitter Approval Information


TABLE 7. Transmitter Option Codes
Rosemount 8742 Transmitter
Rosemount Rosemount
Approval Codes 8712D/C/U 8712H Rosemount 8732 Fieldbus Output I.S. fieldbus Output
(1)
CE •
N0 • • • •
N5 • • • •
E1 • •
E5 • •
ED • •
K0 • •
K1 •
K5 • •
KD(2) • •
N1(3) •
(1) CE Marking is standard on a Rosemount 8712D, 8742 and 8732. Not available on Rosemount 8712H.

(2) Refer to Table 9 on page 25 for relation between ambient temperature, process temperature, and temperature class.

(3) Only available for the 8712D.

North American Certifications K5 Explosionproof Approval with


Factory Mutual (FM) Intrinsically Safe Output (8742 only)

N0 Division 2 Approval (All transmitters) Reference Rosemount Control Drawing 08742-1051

Reference Rosemount Control Drawing 08742-1051 Explosion-Proof for Class I, Division 1, Groups C, D with IS
output for Class I, Division 1, Groups A, B, C, D.
(8742C) or 08732-1052 (8732C).
Temp Code – T4 at 60°C
Class I, Division 2, Groups A, B, C, D
Temp Codes – T4 (8712 at 40°C, 8742 at 60°C), Dust-ignition proof Class II/III, Division 1, Groups E, F, G
Temp Code – T6 at 60°C
T5 (8732 at 60°C)
Class I, Division 2, Groups A, B, C, D
Dust-ignition proof Class II/III, Division 1, Groups E, F, G
Temp Codes – T4 (8712 at 40°C), T6 (8732/8742 at 60°C) Temp Codes – T4 (8742 at 60°C)

Enclosure Type 4X Enclosure Type 4X

K0 Division 2 Approval with N5 Division 2 Approval (All Transmitters)


Intrinsically Safe Output (8742 Only) For flowtubes with IS electrodes only

Reference Rosemount Control Drawing 08742-1051 Reference Rosemount Control Drawing 08742-1051
(8742C) or 08732-1052 (8732C).
Class I, Division 2, Groups A, B, C, D with IS output for
Class I, Division 2, Groups A, B, C, D
Class I, Division 1, Groups A, B, C, D.
Temp Code – T4 at 60°C Temp Codes – T4 (8712 at 40°C, 8742 at 60°C),
T5 (8732 at 60°C)
Dust-ignition proof Class II/III, Division 1, Groups E, F, G
Temp Code – T6 at 60°C Dust-ignition proof Class II/III, Division 1, Groups E, F, G
Temp Codes – T4 (8712 at 40°C), T6 (8732/8742 at 60°C)
Enclosure Type 4X
Enclosure Type 4X

20
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series
E5 Explosion-Proof Approval (8732 and 8742 only) European Certifications
Reference Rosemount Control Drawing 08732-1052 E1 ATEX Flameproof
Explosion-Proof for Class I, Division 1, Groups C, D Hydrogen gas group
Temp Code – T6 at 60°C 8732 - Certificate No.: 03ATEX2052X II 2G
Dust-ignition proof Class II/III, Division 1, Groups E, F, G EEx d IIB + H2 T6 (-20°C ≤ Ta ≤ +65°C)
Temp Code – T6 at 60°C 8742 - Certificate No.: 03ATEX2159X II 2G
Class I, Division 2, Groups A, B, C, D EEx de IIB + H2 T6 (-20°C ≤ Ta ≤ +65°C)
Temp Codes – T4 (8742 at 60°C), T5 (8732 at 60°C) When installed per drawing 08732-1050
Vmax = 250 V AC or 50 V DC
Enclosure Type 4X
0575
Canadian Standards Association (CSA) ED ATEX Flameproof
N0 Division 2 Approval Reference Rosemount Control Drawing 08732-1052
Reference Rosemount Control Drawing 08732-1051 8732 - Certificate No.: KEMA03ATEX2052X II 2G
(8732C or 8742C only) EEx d IIB T6 (Ta = -20°C to +65°C)
Vmax = 250 V AC or 50 V DC
Class I, Division 2, Groups A, B, C, D
0575
Temp Codes – T5 (8732 at 60°C), T4 (8742 at 60°C)
Dust-ignition proof Class II/III, Division 1, Groups E, F, G
Enclosure Type 4X SPECIAL CONDITIONS FOR SAFE USE (X (03ATEX2052X):
If the Model 8732 Flow Transmitter is used integrally with the
K0 Division 2 Approval with
Model 8705 or 8711 Flowtubes, it shall be assured that the
Intrinsically Safe Output (8742 Only)
mechanical contact areas of the Flowtube and Flow Transmitter
Reference Rosemount Control Drawing 08742-1052 comply with the requirements for flat joints according to standard
Class I, Division 2, Groups A, B, C, D with IS output for EN 50018, clause 5.2.
Class I, Division 1, Groups A, B, C, D. Temp Code – T4 at
60°C The relation between ambient temperature, process temperature,
Dust-ignition proof Class II/III, Division 1, Groups E, F, G and temperature class is to be taken from the table under (15 -
Temp Code – T6 at 60°C description) above. (See Table 9)
Enclosure Type 4X
The electrical data is to be taken from the summary under (15 -
electrical data) above. (See Table 10)

If the Model 8732 Flow Transmitter is used integrally with the


Junction Box, it shall be assured that the mechanical contact
areas of the Junction Box and Flow Transmitter comply with the
requirements for flanged joints according to standard EN 50018,
clause 5.2.

INSTALLATION INSTRUCTIONS:
The cable and conduit entry devices and blanking elements shall
be of a certified flameproof type, suitable for the conditions of use
and correctly installed. With the use of conduit, a certified stopping
box shall be provided immediately to the entrance of the
enclosure.

21
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

ED 8742 - Certificate No: KEMA03ATEX2159X II 2G


EEx de IIB T6 (Ta = -20°C to +65°C) Remote Junction Box (8732C or 8742C)
Vmax = 250 V AC or 50 V DC Certificate No.: KEMA03ATEX2052x II 2G
0575 ATEX EEx e (1) T6 (Ta = -20°C to +65°C)
K1 ATEX Flameproof When installed per drawing 08732-1050
Hydrogen Gas Group with After de-energizing, wait 10 minutes before opening cover
Intrinsically safe output (8742 only) 0575
Certificate No.: KEMA03ATEX2159x II 2G
(1)
ATEX EEx de [ia] IIB + H2 T6 (-20°C ≤ Ta ≤ +65°C) IIB + H2 for E1, K1
Vmax = 250 V AC or 50 V DC IIB for ED, KD
0575
KD ATEX Flameproof with
Intrinsically Safe Output (8742 only)
Certificate No: KEMA03ATEX2159X II 2G
EEx de [ia] IIB T6 (Ta = -20°C to +65°C)
Vmax = 250 V AC or 50 V DC
0575
See Table 10 for Electrical Parameters

SPECIAL CONDITIONS FOR SAFE USE (X) (03ATEX2159X):


The relation between ambient temperature, process temperature
and temperature class is to be taken from the table under (15 -
description) above. (See Table 9).

If the Model 8742C Flow Transmitter is used integrally with the


Junction Box, it shall be assured that the mechanical contact
areas of the Junction Box and Flow Transmitter comply with the
requirements for flanged joints according to standard EN 50018,
clause 5.2.

INSTALLATION INSTRUCTIONS:
The cable and conduit entry devices and the closing elements
shall be of a certified increased safety type, suitable for the
conditions of use and correctly installed.

At ambient temperatures above 50°C, the flow meter shall be used


with heat resistant cables with a temperature rating of at least
90°C.

A Junction Box in type of explosion protection increased safety “e”


may be attached to the base of the Model 8742C Flow Transmitter,
permitting remote mounting of the Models 8705 and 8711
Flowtubes.

Ambient temperature range of the Junction Box: -20°C to +65°C.

The Junction Box is classified as II 2 G EEx e IIB T6 and certified


under KEMA 03ATEX2052X.

N1 ATEX (Pending)
8712D - ATEX Certificate No:
EEx nA nL IIC T4

22
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

Flowtube Approval Information


TABLE 8. Flowtube Option Codes(1)
Rosemount 8705 Flowtube Rosemount 8707 Flowtube Rosemount 8711 Flowtube
For Non-flammable For Flammable For Non-flammable For Flammable For Non-flammable For Flammable
Approval Codes Fluids Fluids Fluids Fluids Fluids Fluids
N0 • • •
N1 • • • •
N5 • • • • • •
E1 • • • •
E5 • •
CD(2) • •
KD(2) • •
(1) CE Marking is standard on Rosemount 8705 and 8711. No hazardous location certifications are available on the Rosemount 570TM.

(2) Refer to Table 9 on page 25 for relation between ambient temperature, process temperature, and temperature class.

North American Certifications Canadian Standards Association (CSA)


N0 Suitable for Class I, Division 2, Groups A, B, C, D
Factory Mutual (FM) Temp Code – T5 (8705/8711 at 60°C)
N0 Division 2 Approval for Temp Code – T3C (8707 at 60°C)
Non-Flammable Fluids (All Flowtubes) Dust-Ignition proof Class II/III, Division 1, Groups E, F, G
Class I, Division 2, Groups A, B, C, D Enclosure Type 4X
Temp Code – T5 (8705/8711 at 60°C)
Temp Code – T3C (8707 at 60°C)
Dust-Ignition proof Class II/III, Division 1, Groups E, F, G
Temp Code – T6 (8705/8711 at 60°C)
Temp Code – T3C (8707 at 60°C)
Enclosure Type 4X
N5 Division 2 Approval for Flammable Fluids
(All Flowtubes)
Class I, Division 2, Groups A, B, C, D
Temp Code – T5 (8705/8711 at 60°C)
Temp Code – T3C (8707 at 60°C)
Dust-Ignition proof Class II/III, Division 1, Groups E, F, G
Temp Code – T6 (8705/8711 at 60°C)
Temp Code – T3C (8707 at 60°C)
Enclosure Type 4X
E5 Explosion-Proof (8711 Only)
Explosion-Proof for Class I, Division 1, Groups C, D
Temp Code – T6 at 60°C
Dust-Ignition proof Class II/III, Division 1, Groups E, F, G
Temp Code – T6 at 60°C
Class I, Division 2, Groups A, B, C, D
Temp Code – T5 at 60°C
Enclosure Type 4X

23
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

European Certifications E1, ATEX Increased Safety (Zone 1)


N1 ATEX Non-Sparking/Non-incendive (8705/8711 Only) KD with IS Electrodes (8705 only)
Certificate No: KEMA02ATEX1302X II 3G Certificate No. KEMA 03ATEX2052X II 1/2G
EEx nA [L] IIC T3... T6 EEx e ia IIC T3... T6 (Ta = -20 to 60°C) (See Table 9)
Ambient Temperature Limits -20 to 65°C 0575
Special conditions per drawing 08732-1050 When installed per drawing 08732-1050
0575 Vmax = 40 V DC (pulsed)

SPECIAL CONDITIONS FOR SAFE USE (X): SPECIAL CONDITIONS FOR SAFE USE (X):
If the Model 8732 Flow Transmitter is used integrally with the If the Model 8732 Flow Transmitter is used integrally with the
Model 8705 or Model 8711 Flowtubes, it shall be assured that the Model 8705 or Model 8711 Flowtubes, it shall be assured that the
mechanical contact areas of the Flowtube and Flow Transmitter mechanical contact areas of the Flowtube and Flow Transmitter
comply with the requirements for flat joints according to standard comply with the requirements for flat joints according to standard
EN 50018, clause 5.2. The relation between ambient temperature, EN 50018, clause 5.2. The relation between ambient temperature,
process temperature and temperature class is to be taken from the process temperature and temperature class is to be taken from the
table under (15-description) above. - (See certificate) The table under (15-description) above. - (See certificate) The
electrical data is to be taken from the summary under (15-electrical electrical data is to be taken from the summary under (15-electrical
data above). data above).

E1, ATEX Increased Safety (Zone 1)


CD with IS Electrodes (8711 only) INSTALLATION INSTRUCTIONS:
At ambient temperature above 50°C, the flowmeter shall be used
Certificate No: KEMA03ATEX2052X II 1/2G
with heat resistant cables with a temperature rating of at least
EEx e ia IIC T3... T6 (Ta = -20 to +60°) (See Table 9)
90°C.
0575
When installed per drawing 08732-1050
A fuse with a rating of maximum 0,7 A according to IEC 60127-1
Vmax = 40 V DC (pulsed)
shall be included in the coil excitation circuit if the flowtubes are
used with other flow transmitters (e.g. Model 8712).
SPECIAL CONDITIONS FOR SAFE USE (X):
If the Model 8732 Flow Transmitter is used integrally with the
Model 8705 or Model 8711 Flowtubes, it shall be assured that the
mechanical contact areas of the Flowtube and Flow Transmitter
comply with the requirements for flat joints according to standard
EN 50018, clause 5.2. The relation between ambient temperature,
process temperature and temperature class is to be taken from the
table under (15-description) above. - (See certificate) The
electrical data is to be taken from the summary under (15-electrical
data above).

INSTALLATION INSTRUCTIONS:
At ambient temperature above 50°C, the flowmeter shall be used
with heat resistant cables with a temperature rating of at least
90°C.

A fuse with a rating of maximum 0,7 A according to IEC 60127-1


shall be included in the coil excitation circuit if the flowtubes are
used with other flow transmitters (e.g. Model 8712).

24
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

TABLE 9. Relation between ambient temperature, process temperature, and temperature class(1)
Meter Size (Inches) Maximum Ambient Temperature Maximum Process Temperature Temperature Class
1/2 149°F (65°C) 297°F (147°C) T3
1 149°F (65°C) 318°F (159°C) T3
1 95°F (35°C) 232°F (111°C) T4
11/2 149°F (65°C) 297°F (147°C) T3
11/2 140°F (60°C) 171°F (77°C) T4
2 149°F (65°C) 289°F (143°C) T3
2 149°F (65°C) 163°F (73°C) T4
2 104°F (40°C) 144°F (62°C) T5
3 - 36 149°F (65°C) 351°F (177°C) T3
3 - 36 149°F (65°C) 210°F (99°C) T4
3 - 36 131°F (55°C) 144°F (62°C) T5
3 - 36 104°F (40°C) 144°F (62°C) T6
(1) This table is applicable for CD and KD option codes only.

TABLE 10. Electrical Data


Model 8732 Flow Transmitter
Power supply: 250 Vac, 1 A or 50 Vdc, 2,5 A, 20 W maximum
Pulsed output circuit: 30 V dc (pulsed), 0,25 A, 7,5 W maximum
4-20 mA output circuit: 30 V dc, 30 mA, 900 mW maximum
Model 8705 and 8711 Flowtubes
Coil excitation circuit: 40 V dc (pulsed), 0,5 A, 20 W maximum
Electrode circuit: in type of explosion protection intrinsic safety EEx ia IIC, 5 V, 1 mW maximum, Um = 250 V
Model 8742C Flow Transmitter (EEx de
version):
Power supply: 250 Vac, 1 A, 40 VA or 50 Vdc, 2,5 A, 15 W maximum
Foundation Fieldbus output: 30 Vdc, 30 mA, 1 W maximum
Model 8742C Flow Transmitter (EEx de [ia]
version):
Power supply: 250 Vac, 1 A, 40 VA or 50 Vdc, 2,5 A, 15 W maximum
Foundation Fieldbus output: in type of explosion protection intrinsic safety EEx ia IIB, only for connection to a certified intrinsically safe
(terminals + and -) circuit according to the FISCO model, with the following maximum values:
Ui = 30 V
Ii = 380 mA
Pi = 5,32 W
Ci = 4,4 nF
Li = 0 mH

25
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

TABLE 11. Relation between the maximum ambient temperature, the maximum process temperature,
and the temperature class(1)
Maximum process temperature °F (°C) per temperature class
Maximum Ambient
Temperature T3 T4 T5 T6
0.5 inch flowtube size
149°F (65°C) 297°F (147°C) 138°F (59°C) 54°F (12°C) 18°F (-8°C)
140°F (60°C) 309°F (154°C) 151°F (66°C) 66°F (19°C) 28°F (-2°C)
131°F (55°C) 322°F (161°C) 163°F (73°C) 79°F (26°C) 41°F (5°C)
122°F (50°C) 334°F (168°C) 176°F (80°C) 90°F (32°C) 54°F (12°C)
113°F (45°C) 347°F (175°C) 189°F (87°C) 102°F (39°C) 66°F (19°C)
104°F (40°C) 351°F (177°C) 199°F (93°C) 115°F (46°C) 79°F (26°C)
95°F (35°C) 351°F (177°C) 212°F (100°C) 127°F (53°C) 90°F (32°C)
86°F (30°C) 351°F (177°C) 225°F (107°C) 138°F (59°C) 102°F (39°C)
77°F (25°C) 351°F (177°C) 237°F (114°C) 151°F (66°C) 115°F (46°C)
68°F (20°C) 351°F (177°C) 248°F (120°C) 163°F (73°C) 127°F (53°C)
1.0 inch flowtube size
149°F (65°C) 318°F (159°C) 158°F (70°C) 72°F (22°C) 34°F (1°C)
140°F (60°C) 331°F (166°C) 171°F (77°C) 84°F (29°C) 46°F (8°C)
131°F (55°C) 343°F (173°C) 183°F (84°C) 97°F (36°C) 59°F (15°C)
122°F (50°C) 351°F (177°C) 196°F (91°C) 109°F (43°C) 72°F (22°C)
113°F (45°C) 351°F (177°C) 207°F (97°C) 122°F (50°C) 84°F (29°C)
104°F (40°C) 351°F (177°C) 219°F (104°C) 135°F (57°C) 97°F (36°C)
95°F (35°C) 351°F (177°C) 232°F (111°C) 145°F (63°C) 109°F (43°C)
86°F (30°C) 351°F (177°C) 244°F (118°C) 158°F (70°C) 122°F (50°C)
77°F (25°C) 351°F (177°C) 257°F (125°C) 171°F (77°C) 135°F (57°C)
68°F (20°C) 351°F (177°C) 270°F (132°C) 183°F (84°C) 145°F (63°C)
1.5 inch flowtube size
149°F (65°C) 297°F (147°C) 160°F (71°C) 88°F (31°C) 55°F (13°C)
140°F (60°C) 307°F (153°C) 171°F (77°C) 97°F (36°C) 66°F (19°C)
131°F (55°C) 318°F (159°C) 181°F (83°C) 108°F (42°C) 77°F (25°C)
122°F (50°C) 329°F (165°C) 192°F (89°C) 118°F (48°C) 88°F (31°C)
113°F (45°C) 340°F (171°C) 203°F (95°C) 129°F (54°C) 97°F (36°C)
104°F (40°C) 351°F (177°C) 214°F (101°C) 140°F (60°C) 108°F (42°C)
95°F (35°C) 351°F (177°C) 223°F (106°C) 151°F (66°C) 118°F (48°C)
86°F (30°C) 351°F (177°C) 234°F (112°C) 160°F (71°C) 129°F (54°C)
77°F (25°C) 351°F (177°C) 244°F (118°C) 171°F (77°C) 140°F (60°C)
68°F (20°C) 351°F (177°C) 255°F (124°C) 181°F (83°C) 151°F (66°C)
Continued on Next Page

26
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series
TABLE 11. Relation between the maximum ambient temperature, the maximum process temperature,
and the temperature class(1)
Maximum process temperature °F (°C) per temperature class
Maximum Ambient
Temperature T3 T4 T5 T6
2.0 inch flowtube size
149°F (65°C) 289°F (143°C) 163°F (73°C) 95°F (35°C) 66°F (19°C)
140°F (60°C) 300°F (149°C) 172°F 78(°C) 104°F (40°C) 75°F (24°C)
131°F (55°C) 309°F (154°C) 183°F (84°C) 115°F (46°C) 84°F (29°C)
122°F (50°C) 318°F (159°C) 192°F (89°C) 124°F (51°C) 95°F (35°C)
113°F (45°C) 329°F (165°C) 201°F (94°C) 135°F (57°C) 104°F (40°C)
104°F (40°C) 338°F (170°C) 212°F (100°C) 144°F (62°C) 115°F (46°C)
95°F (35°C) 349°F (176°C) 221°F (105°C) 153°F (67°C) 124°F (51°C)
86°F (30°C) 351°F (177°C) 232°F (111°C) 163°F (73°C) 135°F (57°C)
77°F (25°C) 351°F (177°C) 241°F (116°C) 172°F (78°C) 144°F (62°C)
68°F (20°C) 351°F (177°C) 252°F (122°C) 183°F (84°C) 153°F (67°C)
3 to 60 inch flowtube size
149°F (65°C) 351°F (177°C) 210°F (99°C) 117°F (47°C) 75°F (24°C)
140°F (60°C) 351°F (177°C) 223°F (106°C) 129°F (54°C) 90°F (32°C)
131°F (55°C) 351°F (177°C) 237°F (114°C) 144°F (62°C) 102°F (39°C)
122°F (50°C) 351°F (177°C) 250°F (121°C) 156°F (69°C) 117°F (47°C)
113°F (45°C) 351°F (177°C) 264°F (129°C) 171°F (77°C) 129°F (54°C)
104°F (40°C) 351°F (177°C) 266°F (130°C) 183°F (84°C) 144°F (62°C)
95°F (35°C) 351°F (177°C) 266°F (130°C) 198°F (92°C) 156°F (69°C)
86°F (30°C) 351°F (177°C) 266°F (130°C) 203°F (95°C) 171°F (77°C)
77°F (25°C) 351°F (177°C) 266°F (130°C) 203°F (95°C) 176°F (80°C)
68°F (20°C) 351°F (177°C) 266°F (130°C) 203°F (95°C) 176°F (80°C)
(1) This table is applicable for N1 option codes only.

27
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

Dimensional Drawings
FIGURE 4. Rosemount 8712C/U/H Transmitter

WITH LOI COVER WITH STANDARD


COVER
9.01
4.31 (229)
LOI Cover 0.44
(109) 2.81 3.11
Standard 3.51 (11)
(71) (79)
Cover (89)

11.15
LOI Keypad
(283)
Cover 12.02
(305)

2.96
(75)

17.70 (450)

¾–14 NPT
Conduit 11.37
Connection (289)
(3 Places) 8712-12A01A, 8712B01A, 8712C01A, 8712D01A

Ground Lug

NOTE
Dimensions are in inches (millimeters)

28
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

FIGURE 5. Rosemount 8732C and 8742C Transmitter

7.49 (190)

6.48 (165)
1.94 LOI Cover
(49)

3
/4”-14 NPT Electrical
5.10 3.00 Conduit Connections
(130) (76) (2 places)

8.81
10.5 (224)
(267)

3.07
(78)

3.43
3
(87) /4”-14 NPT
Flowtube Conduit
Connections
(2 places)

11.02
(280)

5.82
(148)
8742-8742_06A, 8742_06B, 1002G01A

Note
Dimensions in inches (millimeters)

29
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

FIGURE 6. Rosemount 8705 and 8707 Flowtubes, Typical of 0.5- through 4-inch (15 through 100 mm) Line Sizes
4.02 5.00
(102) (127)
2.00
(50.8)

2.6
1.8
2.6 (66)
(46)
(66)

3 H
/4-14 NPT
Conduit
Connection C
Relief

8705-1002A05A, 1002B05A
Valve

Notes A L
Dimensions are in inches (millimeters)
See Table 13 and Table 14 for variable dimensions
W1 Housing Configuration (Sealed, welded housing with pressure relief valve)

FIGURE 7. Rosemount 8705 and 8707 Flowtubes, Typical of 6- through 36-inch (150 through 900 mm) Line Sizes

4.02
(102)
2.00
(51) 5.00
1.8 (127)
2.6 (46)
2.66
(66)
(66)

8705-1002A06A, 1002B06A

W3 Housing
Configuration

L
A Notes
Dimensions are in inches (millimeters)
D See Table 13 and Table 14 for variable dimensions
W3 Housing Configuration (Sealed welded housing with
separate electrode compartments)

30
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

FIGURE 8. Rosemount 8705 Sanitary Flowtubes, Typical of 1/2- through 3-inch (15 through 86 mm) Line Sizes

4.02
(102)
2.00
5.00
(50.8)
(127)
1.8
2.6 (46) 2.6
(66) (66)

3/4-14
Grounding
NPT Clamp
E C
Conduit
Connection

8705-1002A05A, 1002B05D
Notes
L
Dimensions are in inches (millimeters)
See Table 14 for available dimensions
W0 Housing Configuration (Sealed welded housing)

TABLE 12. Sanitary Rosemount 8705 Flowtube Dimensions in inches (millimeters)


Nominal Tri-Clamp
Line Size and Flange Rating Diameter Process Flange Rad. “B” Body Height “C” Max Overall Flowtube Length “L”
0.5–150 lb. 1.00 (25) 1.75 (44) 8.88 (226) 13.78 (350)
1–150 lb. 1.50 (40) 2.13 (54) 8.88 (226) 13.78 (350)
1.5–150 lb. 2.00 (50) 2.50 (64) 9.50 (242) 13.78 (350)
2–150 lb. 3.00 (80) 3.00 (76) 9.50 (242) 13.78 (350)
3–150 lb. 4.00 (100) 3.75 (95) 12.50 (318) 13.78 (350)
Dimensions with ASME B16.5 (ANSI) Flanges and Tri-Clamp Adapters.

31
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

TABLE 13. Rosemount 8705 and 8707 Dimensions in Inches (Millimeters) (Dimensions with ASME B16.5 (ANSI) Flanges) (1)
Refer to Dimensional Drawings, Figure 6, Figure 7, and Figure 8
Liner Face Overall Flowtube
Line Size(2) and Body Height Diameter Length Flange Diameter
Flange Rating “H” “A” “L”(3) “D” Liner Thickness Inside Diameter
0.5–150 6.75 (171) 1.38 (35) 7.88 (200) 3.50 (89) 0.09 (2.3) 0.49 (12.5)
0.5–300 6.75 (171) 1.38 (35) 7.88 (200) 3.75 (95) 0.09 (2.3) 0.49 (12.5)
0.5–600 6.75 (171) 1.38 (35) 8.67 (220) 3.75 (95) 0.09 (2.3) 0.49 (12.5)
1–150 6.75 (171) 2.00 (51) 7.88 (200) 4.25 (108) 0.09 (2.3) 0.91 (23)
1–300 6.75 (171) 2.00 (51) 7.88 (200) 4.88 (124) 0.09 (2.3) 0.91 (23)
1–600 6.75 (171) 1.62 (41) 8.67 (220) 4.88 (124) 0.09 (2.3) 0.91 (23)
1–900 6.75 (171) 1.62 (41) 9.66 (245) 5.88 (149) 0.13 (3.3) 0.80 (20)
1.5–150 7.10 (180) 2.88 (73) 7.88 (200) 5.00 (127) 0.12 (3.1) 1.44 (37)
1.5 –300 7.10 (180) 2.88 (73) 7.88 (200) 6.12 (155) 0.12 (3.1) 1.44 (37)
1.5–600 7.10 (180) 2.88 (73) 8.63 (219) 6.12 (155) 0.13 (3.3) 1.36 (35)
1.5–900 7.10 (180) 2.50 (64) 9.52 (242) 7.00 (178) 0.13 (3.3) 1.25 (32)
2–150 7.10 (180) 3.62 (92) 7.88 (200) 6.00 (152) 0.12 (3.1) 1.91 (49)
2–300 7.10 (180) 3.62 (92) 7.88 (200) 6.50 (165) 0.12 (3.1) 1.91 (49)
2–600 7.10 (180) 3.62 (92) 8.78 (223) 6.50 (165) 0.13 (3.3) 1.82 (46)
2–900 7.10 (180) 3.25 (83) 10.26 (261) 8.50 (216) 0.13 (3.3) 1.69 (43)
3–150 8.10 (206) 5.00 (127) 7.88 (200) 7.50 (191) 0.15 (3.8) 2.96 (75)
3–300 8.10 (206) 5.00 (127) 8.63 (219) 8.25 (210) 0.15 (3.8) 2.96 (75)
3–600 8.10 (206) 5.00 (127) 12.4 (315) 8.25 (210) 0.13 (3.3) 2.76 (70)
3–900 8.10 (206) 4.63 (118) 12.8 (326) 9.50 (241) 0.13 (3.3) 2.37 (60)
4–150 8.45 (215) 6.19 (157) 9.84 (250) 9.00 (229) 0.15 (3.8) 3.96 (101)
4–300 8.45 (215) 6.19 (157) 10.88 (276) 10.00 (254) 0.15 (3.8) 3.96 (101)
4–600 8.45 (215) 6.19 (157) 12.83 (326) 10.75 (273) 0.13 (3.3) 3.72 (95)
4–900 8.45 (215) 5.81 (148) 13.89 (353) 11.50 (292) 0.13 (3.3) 3.37 (86)
6–150 9.45 (240) 8.50 (216) 11.81 (300) 11.00 (279) 0.19 (4.8) 5.98 (152)
6–300 9.45 (240) 8.50 (216) 13.06 (332) 12.50 (318) 0.19 (4.8) 5.69 (144)
6–600 9.45 (240) 8.50 (216) 14.23 (361) 14.00 (356) 0.19 (4.8) 5.69 (144)
6–900 9.45 (240) 8.00 (203) 15.51 (394) 15.00 (381) 0.16 (4.1) 4.86 (123)
8–150 10.42 (265) 10.62 (270) 13.78 (350) 13.50 (343) 0.19 (4.8) 7.94 (202)
8–300 10.42 (265) 10.62 (270) 15.60 (396) 15.00 (381) 0.17 (4.3) 7.64 (194)
8–600 10.42 (265) 10.62 (270) 16.72 (428) 16.50 (419) 0.17 (4.3) 7.64 (194)
8–900 10.42 (265) 10.00 (254) 18.47 (469) 18.50 (470) 0.17 (4.3) 6.65 (169)
10–150 11.78 (299) 12.75 (324) 15.00 (381) 16.00 (406) 0.26 (6.5) 9.87 (251)
10–300 11.78 (299) 12.75 (324) 17.13 (435) 17.50 (444) 0.26 (6.5) 9.48 (241)
10–600 11.78 (299) 12.75 (324) 19.54 (496) 20.00 (508) 0.26 (6.5) 9.21 (234)
12–150 12.86 (327) 15.00 (381) 18.00 (457) 19.00 (483) 0.26 (6.7) 11.87 (301)
12–300 12.86 (327) 15.00 (381) 20.14 (512) 20.50 (52) 0.26 (6.7) 11.48 (292)
14–150 13.92 (354) 16.25 (413) 21.00 (533) 21.00 (533) 0.19 (4.8) 13.16 (334)
14–300 13.92 (354) 16.25 (413) 23.25 (591) 23.00 (584) 0.19 (4.8) 12.79 (325)
16–150 14.93 (379) 18.50 (470) 24.00 (610) 23.50 (597) 0.19 (4.8) 15.12 (384)
16–300 14.93 (379) 18.50 (470) 26.25 (667) 25.50 (648) 0.19 (4.8) 14.75 (375)
18–150 16.19 (411) 21.00 (533) 27.00 (686) 25.00 (635) 0.19 (4.8) 17.09 (434)
18–300 16.19 (411) 21.00 (533) 30.12 (765) 28.00 (711) 0.19 (4.8) 16.35 (415)
20–150 17.20 (437) 23.00 (584) 30.00 (762) 27.50 (698) 0.19 (4.8) 18.96 (482)
20–300 17.20 (437) 23.00 (584) 33.25 (845) 30.50 (774) 0.19 (4.8) 18.21 (463)
24–150 19.48 (495) 27.25 (692) 36.00 (914) 32.00 (813) 0.19 (4.8) 22.94 (583)
24–300 19.48 (495) 27.25 (692) 39.64 (1007) 36.00 (914) 0.19 (4.8) 22.06 (560)
30 22.23 (565) 33.80 (859) 37.25 (946) 38.75 (984) 0.19 (4.8) 28.75 (730)
36 26.10 (663) 40.27 (1023) 40.88 (1038) 46.00 (1168) 0.19 (4.8) 29.08 (739)
(1) AS2129 Table D and E flange dimensions match ANSI 150# dimensions.

(2) 30- and 36-inch AWWA C207 Table 2 Class D rated to 150 psi at 150 °F.

(3) When 2 grounding rings are specified, add 0.25 inch (6.35 mm) for 0.50- through 14-inch (15 through 350 mm) flowtubes, add 0.50 inch (12.7 mm) for
16-inch (400 mm) and larger. When lining protectors are specified, add 0.25 inch (6.35 mm) for ½- through 12-inch (15 through 300 mm) flowtubes, add
0.50 inch (12.7 mm) for 14- through 36-inch (350 through 900 mm) flowtubes.

32
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

TABLE 14. Rosemount 8705 Flowtube Dimensions with DIN Flanges in Millimeters (Inches)
Line Size(1) Liner Face
and Flange Body Height Diameter Overall Flowtube Flange Diameter
Rating “H” “A” Length “L”(2) “D” Liner Thickness Inside Diameter
15 mm PN 10–40 171 (6.75) 45 (1.77) 200 (7.88) 95 (3.74) 2.3 (0.09) 12.5 (.49)
25 mm PN 10–40 171 (6.75) 68 (2.68) 200 (7.88) 115 (4.53) 2.3 (0.09) 23.1 (.91)
40 mm PN 10–40 180 (7.10) 88 (3.46) 200 (7.87) 150 (5.91) 3.1 (0.12) 37 (1.44)
50 mm PN 10–40 180 (7.10) 102 (4.02) 200 (7.87) 165 (6.50) 3.1 (0.12) 49 (1.91)
80 mm PN 10–40 206 (8.10) 138 (5.43) 200 (7.87) 200 (7.87) 3.8 (0.15) 75.2 (2.96)
100 mm PN 10–16 215 (8.45) 162 (6.38) 250 (9.84) 220 (8.66) 3.8 (0.15) 100.6 (3.96)
100 mm PN 25–40 215 (8.45) 162 (6.38) 250 (9.84) 235 (9.25) 3.8 (0.15) 100.6 (3.96)
150 mm PN 10 240 (9.45) 212 (8.35) 300 (11.81) 285 (11.22) 4.7 (0.19) 152 (5.98)
150 mm PN 16 240 (9.45) 215 (8.46) 300 (11.81) 220 (8.66) 4.7 (0.19) 152 (5.98)
150 mm PN 25 240 (9.45) 218 (8.58) 300 (11.81) 300 (11.81) 4.7 (0.19) 152 (5.98)
150 mm PN 40 240 (9.45) 218 (8.58) 332 (13.07) 300 (11.81) 4.7 (0.19) 144 (5.67)
200 mm PN 10 265 (10.42) 268 (10.55) 350 (13.78) 240 (13.39) 4.9 (0.19) 202 (7.94)
200 mm PN 16 265 (10.42) 268 (10.55) 350 (13.78) 340 (13.39) 4.9 (0.19) 202 (7.94)
200 mm PN 25 265 (10.42) 278 (10.94) 350 (13.78) 360 (14.17) 4.9 (0.19) 202 (7.94)
200 mm PN 40 265 (10.42) 285 (11.22) 396 (15.60) 375 (14.76) 4.3 (0.17) 194 (7.64)
250 mm PN 10 299 (11.78) 320 (12.60) 381 (15.00) 395 (15.55) 6.6 (0.26) 251 (9.88)
250 mm PN 16 299 (11.78) 320 (12.60) 381 (15.00) 405 (15.94) 6.6 (0.26) 251 (9.88)
250 mm PN 25 299 (11.78) 335 (13.19) 381 (15.00) 425 (16.73) 6.6 (0.26) 251 (9.88)
250 mm PN 40 299 (11.78) 345 (13.58) 435 (17.13) 450 (17.72) 6.6 (0.26) 240 (9.45)
300 mm PN 10 327 (12.86) 370 (14.57) 457 (18.00) 445 (17.52) 6.6 (0.26) 302 (11.87)
300 mm PN 16 327 (12.86) 378 (14.88) 457 (18.00) 460 (18.11) 6.6 (0.26) 302 (11.87)
300 mm PN 25 327 (12.86) 395 (15.55) 457 (18.00) 485 (19.09) 6.6 (0.26) 302 (11.87)
300 mm PN 40 327 (12.86) 410 (16.14) 512 (20.14) 515 (20.28) 6.6 (0.26) 292 (11.48)
350 mm PN 10 354 (13.92) 430 (16.93) 534 (21.03) 505 (19.88) 7.4 (0.19) 334 (13.16)
350 mm PN 16 354 (13.92) 438 (17.24) 534 (21.03) 520 (20.47) 7.4 (0.19) 334 (13.16)
350 mm PN 25 354 (13.92) 450 (17.72) 534 (21.03) 555 (21.85) 7.4 (0.19) 334 (13.16)
350 mm PN 40 354 (13.92) 465 (18.31) 591 (23.25) 580 (22.83) 7.4 (0.19) 12.79 (325)
400 mm PN 10 379 (14.93) 482 (18.98) 610 (24.00) 565 (22.24) 7.4 (0.19) 384 (15.12)
400 mm PN 16 379 (14.93) 490 (19.29) 610 (24.00) 580 (22.83) 7.4 (0.19) 384 (15.12)
400 mm PN 25 379 (14.93) 505 (19.88) 610 (24.00) 620 (24.41) 7.4 (0.19) 384 (15.12)
400 mm PN 40 379 (14.93) 535 (21.06) 667 (26.25) 660 (25.98) 7.4 (0.19) 375 (14.75)
500 mm PN 10 437 (17.20) 585 (23.03) 762 (30.00) 670 (26.38) 7.4 (0.19) 482 (18.96)
500 mm PN 16 437 (17.20) 610 (24.02) 762 (30.00) 715 (28.15) 7.4 (0.19) 482 (18.96)
500 mm PN 25 437 (17.20) 615 (24.21) 762 (30.00) 730 (28.74) 7.4 (0.19) 482 (18.96)
500 mm PN 40 437 (17.20) 615 (24.21) 845 (33.25) 755 (29.72) 7.4 (0.19) 463 (18.21)
600 mm PN 10 495 (19.48) 685 (26.97) 914 (36.00) 780 (30.71) 7.4 (0.19) 583 (22.94)
600 mm PN 16 495 (19.48) 725 (28.54) 914 (36.00) 840 (33.07)) 7.4 (0.19) 583 (22.94)
600 mm PN 25 495 (19.48) 720 (28.35) 914 (36.00) 845 (33.27) 7.4 (0.19) 581 (22.87)
Dimensions with DIN Flanges

(1) Consult factory for larger line sizes.

(2) When 2 grounding rings are specified, add 6.35 mm (0.25 in.) for 15 mm through 350 mm (½- through 14 in.) flowtubes or 12.7 mm (0.50 in.) for 400 mm
(16 in.) and larger. When lining protectors are specified, add 6.35 mm (0.25 in.) for 15 mm through 300 mm (½- through 12-in.) flowtubes, 12.7 mm (0.50
in.) for 350 mm through 900 mm (14- through 36-in.) flowtubes.

33
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

FIGURE 9. Rosemount 8711 Dimensional Drawings (0.15-inch through 1-inch line sizes)

4.02 (102) 5.00


(127)

1.8
2.6 (46)
(66)

Grounding
A Clamp
¾–14 NPT
Conduit
Connection B”

8711-1012B03B, 1012A03B
(2 places)

Flow

NOTE
Dimensions are in
C
inches (millimeters) D
See Table 15 for
variable dimensions

FIGURE 10. Rosemount 8711 Dimensional Drawings (1.5-inch through 8-inch line sizes)

4.02 (102) 5.00


(127)

1.8
2.6
(46)
(66)

B
Grounding
¾–14 NPT A Clamp
Conduit
Connection
(2 places)
8711-1012B04B, 1012A04B

Flow

C
NOTE D
Dimensions are in inches (millimeters)
See Table 15 for variable dimensions

34
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

TABLE 15. Rosemount 8711 Flowtube Dimensions and Weight


Nominal Flowtube Housing Dimensions
Line Size Flowtube Length Weight
Inches (mm) “A” Max. “B” “C” “D” Inside Diameter lb (kg)
0.15(1) (4) 4.00 (102) 5.44 (138) 3.56 (90) 2.17 (55) .165 (4) 4 (2)
0.30(1) (8) 4.00 (102) 5.44 (138) 3.56 (90) 2.17 (55) .287 (7) 4 (2)
0.5 (15) 4.00 (102) 5.44 (138) 3.56 (90) 2.17 (55) .595 (15) 4 (2)
1 (25) 4.31 (109) 6.06 (154) 4.50 (114) 2.17 (55) .959 (24) 5 (2)
1.5 (40) 4.42 (112) 7.41 (188) 3.28 (83) 2.73 (69) 1.50 (38) 5 (2)
2 (50) 4.64 (118) 7.94 (202) 3.91 (99) 3.26 (83) 1.95 (50) 7 (3)
3 (80) 5.26 (134) 9.19 (233) 5.16 (131) 4.68 (119) 2.98 (76) 13 (6)
4 (100) 5.87 (149) 10.41 (264) 6.38 (162) 5.88 (149) 3.90 (99) 22 (10)
6 (150) 6.97 (177) 12.60 (320) 8.56 (217) 6.87 (174) 5.825 (148) 35 (16)
8 (200) 8.00 (2003) 14.66 (372) 10.63 (270) 8.86 (225) 7.87 (200) 60 (27)
(1) 0.15 and 0.30 inch (4 and 8 mm) flowtubes mount between ½-inch (13 mm) flange.

FIGURE 11. Rosemount 8714D Calibration Standard

2.69
(68.3)

4.13
(104.9)

7.19
(182.6)

Rosemount 8732C
Connector Type
Note:
The Rosemount 8714D is shipped with
8714-8714A

both the 8712C/U and 8732C Connector


Types Rosemount 8712C/U Connector Type

35
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

Magnetic Flowmeter Sizing


Flowmeter Sizing
TABLE 17. Line Size vs. Conversion Factor
Because of its effect on flow velocity, flowtube size is Nominal Line Size Gallons Per Minute Liters Per
an important consideration. It may be necessary to Inches (mm) Factor Minute Factor
select a magnetic flowmeter that is larger or smaller 0.15 (4) 0.055 0.683
than the adjacent piping to ensure the fluid velocity is 0.30 (8) 0.220 2.732
in the specified measuring range of the flowtube. ½ (15) 0.947 11.745
Suggested guidelines and examples for sizing 1 (25) 2.693 33.407
normal velocities in different applications are listed in 1½ (40) 6.345 78.69
Table 16, Table 17, and Table 18. Operation outside 2 (50) 10.459 129.7
3 (80) 23.042 285.7
these guidelines may also give acceptable
4 (100) 39.679 492.0
performance.
6 (150) 90.048 1,116
TABLE 16. Sizing Guidelines 8 (200) 155.94 1,933
10 (250) 245.78 3,048
Velocity Range
12 (300) 352.51 4,371
Application Velocity Range (ft/s) (m/s)
14 (350) 421.70 5,229
Normal Service 2–20 0.6–6.1
16 (400) 550.80 6,830
Abrasive Slurries 3–10 0.9–3.1 18 (450) 697.19 8,645
Non-Abrasive 5–15 1.5–4.6 20 (500) 866.51 10,745
Slurries 24 (600) 1,253.2 15,541
30 (750) 2,006.0 24,877
To convert flow rate to velocity, use the appropriate 36 (900) 2,935.0 36,398
factor listed in Table 16 and the following equation:

Flow Rate
Velocity =
Factor

Example: SI Units
Magmeter Size: 100 mm (factor from Table 17 = 492.0)
Normal Flow Rate: 800 L/min
800 (L/min)
Velocity =
492.0

Velocity = 1.7 m/s

Example: English Units


Magmeter Size: 4 inch (factor from Table 17 = 39.679)
Normal Flow Rate: 300 GPM

Velocity = 300 (gpm)


39.679

Velocity = 7.56 ft/s

36
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

TABLE 18. Line Size vs. Velocity/Rate


Minimum/Maximum Flow Rate
Gallons per Minute Liters per Minute
Nominal
Line Size in at 0.04 ft/s at 1 ft/s at 30 ft/s at 0.012 m/s at 0.3 m/s at 10 m/s
Inches (Low-flow (Min Range (Max Range (Low-flow (Min Range (Max Range
(mm) Cutoff) Setting) at 3 ft/s Setting) Cutoff) Setting) at 1 m/s Setting)
.15 (4) 0.002 0.055 0.16 1.65 0.01 0.21 0.68 6.83
.30 (8) 0.009 0.220 0.66 6.60 0.03 0.83 2.73 27.32
1/2 (15) 0.038 0.947 2.84 28.412 0.14 3.58 11.74 117.45
1 (25) 0.108 2.694 8.08 80.813 0.41 10.18 33.40 334.07
11/2 (40) 0.254 6.345 19.03 190.36 0.96 23.98 78.69 786.9
2 (50) 0.418 10.459 31.37 313.77 1.58 39.54 129.7 1,297
3 (80) 0.922 23.042 69.12 691.26 3.49 87.10 285.7 2,857
4 (100) 1.588 36.679 119.0 1,190.4 6.00 138.6 492.0 4,920
6 (150) 3.600 90.048 270.1 2,701.4 13.61 340.3 1,116 11,167
8 (200) 6.240 155.94 467.7 4,677.8 23.59 589.4 1,933 19,337
10 (250) 9.840 245.78 737.3 7,373.4 37.20 929.0 3,048 30,480
12 (300) 14.200 352.51 1,059 10,575 53.68 1,332 4,371 43,715
14 (350) 16.800 421.70 1,265 12,651 63.50 1,594 5,230 52,296
16 (400) 22.000 550.80 1,652 16,524 83.16 2,082 6,830 68,304
18 (450) 27.800 697.19 2,091 20,916 105.0 2,635 8,646 86,459
20 (500) 34.600 866.51 2,599 25,995 130.7 3,275 10,740 107,457
24 (600) 50.200 1,253.2 3,759 37,596 189.7 4,737 15,540 155,414
30 (750) 80.200 2,006.0 6,018 60,180 303.1 7,582 24,880 248,773
36 (900) 117.40 2,935.0 8,805 88,050 443.7 11,094 36,390 363,983

Upstream/Downstream Piping Length


TABLE 19. Grounding Options
To ensure specification accuracy over widely varying Grounding Options General Characteristics
process conditions, install the flowtube with a No Grounding Options • Acceptable for conductive unlined pipe
minimum of five straight pipe diameters upstream (grounding straps) • Grounding straps provided by
and two straight pipe diameters downstream from the Rosemount Inc.
electrode plane. See Figure 12. This procedure Grounding Electrode • Same material as measurement
should adequately allow for disturbances created by electrodes
elbows, valves, and reducers. • Sufficient grounding option when
process fluid conductivity is greater
FIGURE 12. Upstream and Downstream Straight Pipe than 100 microsiemens/cm
Diameters • Not recommended in electrolysis or
galvanic corrosion applications
5 Pipe Diameters 2 Pipe Diameters
Grounding Rings • Low conductivity process fluids
8705-0281F02A

• Cathodic or electrolysis applications


Flow that may have stray currents in or
around the process
• Variety of materials for process fluid
compatibility
Lining Protectors • Protects upstream edge of flowtube
Flowtube Grounding
from abrasive fluids
A reliable ground path is required between the • Permanently installed on flowtube
flowtube and the process fluid. Optional grounding • Protects liner material from over
torquing of flange bolts
rings, grounding electrodes, and lining protectors are
• Provides ground path and eliminates
available with 8700 Series flowtubes to ensure need for grounding ring or grounding
proper grounding. See Table 19 and Table 20. electrode

37
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

TABLE 20. Grounding Installation


Grounding Options
Type of Pipe No Grounding Options Grounding Rings Grounding Electrodes Lining Protectors
Conductive Unlined Acceptable Not Required Not Required Acceptable
Pipe
Conductive Lined Pipe Not Acceptable Acceptable Acceptable Acceptable
Non-Conductive Pipe Not Acceptable Acceptable Acceptable Acceptable

Material Selection
Several liner types, electrode materials, and TABLE 22. Electrode Material
electrode types are available on Rosemount Electrode Material General Characteristics
Magnetic Flowtubes to ensure compatibility with 316L Stainless Steel • Good corrosion resistance
virtually any application. See Table 21 for information • Good abrasion resistance
on liner types, Table 22 for information on electrode • Not recommended for sulfuric
materials, and Table 23 for information on electrode or hydrochloric acids
types. Other material options not mentioned here are Hastelloy C-276 • Better corrosion resistance
available. Contact your local sales representative for • High strength
• Good in slurry applications
material selection. For further guidance on selecting
• Effective in oxidizing fluids
materials, refer to the Magnetic Flowmeter Material
Tantalum • Better chemical resistance
Selection Guide located on Rosemount.com
• Not recommended for
(document number 00816-0100-3033). hydrofluoric acid, fluosilic
acid, or sodium hydroxide
TABLE 21. Lining Material
90% Platinum - 10% Iridium • Best chemical resistance
Lining Material General Characteristics
• Expensive material
PFA • Highly chemical-resistant • Not recommended for aqua
• Excellent high temperature regia
capabilities
Titanium • Better chemical resistance
Teflon (PTFE) • Highly chemical-resistant • Better abrasion resistance
• Excellent high temperature • Good for sea water
capabilities applications
Tefzel (ETFE) • Highly chemical-resistant • Not recommended for
• Better abrasion resistance hydrofluoric or sulfuric acid
than Teflon
Polyurethane • Excellent abrasion resistance TABLE 23. Electrode Type
• Limited chemical resistance Electrode Type General Characteristics
Neoprene • Very good abrasion Standard Measurement • Costs least
resistance • Good for most applications
• Better chemical resistance
Bulletnose Electrodes • Slightly more expensive
than polyurethane
• Best option for coating
Linatex Rubber • Very good abrasion processes
resistance
Removable Electrode • Most expensive
• Limited chemical resistance
especially in acids • Better than standard
• Softer material than electrode for coating
polyurethane and neoprene processes

38
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

Ordering Information
ROSEMOUNT 8712C/U/H ORDERING INFORMATION

Availability
Model Product Description C U H
8712C Magnetic Flowmeter Transmitter • NA NA
8712U Universal Magnetic Flowmeter Transmitter NA • NA
8712H High-Signal Magnetic Flowmeter Transmitter (For use with 8707 High-Signal Flowtube only.) NA NA •
Code Transmitter Style C U H
R Remote (2-inch pipe or surface mounting) • • •
Code Power Supply Voltage C U H
03 10–30 V dc • • NA
12 115 V ac, 50–60 Hz • • •
24 230 V ac, 50–60 Hz • • NA
Code Product Certifications C U H
CE CE Marking • • NA
N0 Factory Mutual (FM) Class I, Division 2 Approval for nonflammable fluids; • • •
Canadian Standards Association (CSA) Class I, Division 2 Approval
N5 Factory Mutual (FM) Class I, Division 2 Approval for flammable fluids • • •
Code Options C U H
B6 Stainless Steel 4-bolt Kit for 2-inch Pipe Mount • • •
C1 Custom Configuration (Completed CDS required with order) • • •
C4 Analog Output Levels Compliant with NAMUR recommendations NE43, 18-January-1994, • • NA
and high alarm level(1)
CN Analog Output Levels Compliant with NAMUR recommendations NE43, 18-January-1994, • • NA
and low alarm level(1)
D1 High Accuracy Calibration [0.25% of rate from 3 to 30 ft/s (0.9 to 10 m/s)] matched flowtube • • •
and transmitter system(2)
M4 Local Operator Interface (LOI) • • •
T1 Battery-backed Totalizer • • NA
J1 CM 20 conduit adapter • • NA
J2 PG 13.5 conduit adapter • • NA
Code Quick Installation Guide (QIG) Language Options (Default is English)
YA Danish QIG • • •
YD Dutch QIG • • •
YF French QIG • • •
YG German QIG • • •
YH Finnish QIG • • •
YI Italian QIG • • •
YN Norwegian QIG • • •
YP Portuguese QIG • • •
YS Spanish QIG • • •
YW Swedish QIG • • •
Typical Model Number: 8712C R 12 N0 M4

(1) NAMUR compliant operation and the Alarm latch options are preset at the factory and can not be changed to standard operation in the field.

(2) Option Code must be selected for both flowtube and transmitter.

39
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

ROSEMOUNT 8712D ORDERING INFORMATION

Model Product Description Availability


8712D Magnetic Flowmeter Transmitter •
Code Transmitter Style
R Remote (2-inch pipe or surface mounting) •
Code Power Supply Voltage
03 12–42 V dc •
12 90-250 V ac, 50–60 Hz •
Code Product Certifications
N0 Factory Mutual (FM) Class I, Division 2 Approval for nonflammable fluids; •
Canadian Standards Association (CSA) Class I, Division 2 Approval; CE Marking
N5 Factory Mutual (FM) Class I, Division 2 Approval for flammable fluids •
Code Options
B6 Stainless Steel 4-bolt Kit for 2-inch Pipe Mount •
C1 Custom Configuration (Completed CDS required with order) •
C4 Analog Output Levels Compliant with NAMUR recommendations NE43, 18-January-1994, •
and high alarm level(1)
CN Analog Output Levels Compliant with NAMUR recommendations NE43, 18-January-1994, •
and low alarm level(1)
D1 High Accuracy Calibration [0.25% of rate from 3 to 30 ft/s (0.9 to 10 m/s)] matched flowtube •
and transmitter system(2)
M4 Local Operator Interface (LOI) •
J1 CM 20 conduit adapter •
J2 PG 13.5 conduit adapter •
Code Quick Installation Guide (QIG) Language Options (Default is English)
YA Danish QIG
YD Dutch QIG •
YF French QIG •
YG German QIG •
YH Finnish QIG •
YI Italian QIG
YN Norwegian QIG •
YP Portuguese QIG
YS Spanish QIG
YW Swedish QIG
Typical Model Number: 8712D R 12 N0 M4

(1) NAMUR compliant operation and the Alarm latch options are preset at the factory and can not be changed to standard operation in the field.

(2) Option Code must be selected for both flowtube and transmitter.

40
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

ROSEMOUNT 8732C ORDERING INFORMATION

Model Product Description Availability


8732C Magnetic Flowmeter Transmitter •
Code Transmitter Style
T Integral (mounted to flowtube) •
R Remote (2-inch pipe mount)
Code Power Supply Voltage
03 15–50 V dc •
12 85–250 V ac, 50–60 Hz •
Code Product Certifications
N0 Factory Mutual (FM) Class I, Division 2 Approval for nonflammable fluids; •
Canadian Standards Association (CSA) Class I, Division 2 Approval;
CE Marking
N5 Factory Mutual (FM) Class I, Division 2 Approval for flammable fluids •
E1 EEx d IIB + H2 T6, Hydrogen gas flameproof approval •
E5 Factory Mutual (FM) Class I, Division 1, Explosion Proof Approval (1) •
ED ATEX EEx d IIB T6 Flameproof Approval; •
Code Options
C1 Custom Configuration – Completed CDS required with order •
C4 Analog Output Levels Compliant with NAMUR recommendations NE43, 18-January-1994, and high alarm level(2) •
CN Analog Output Levels Compliant with NAMUR recommendations NE43, 18-January-1994, and low alarm level(2) •
D1 High Accuracy Calibration [0.25% of rate from 3 to 30 ft/s (0.9 to 10 m/s)] •
matched flowtube and transmitter system(3)
J1 CM 20 Conduit Adaptor •
J2 PG 13.5 Conduit Adaptor •
L1 Transient Protection Circuitry •
M4 Local Operator Interface •
T1 Non-volatile Totalizer •
Code Quick Installation Guide (QIG) Language Options (Default is English)
YA Danish QIG •
YD Dutch QIG •
YF French QIG •
YG German QIG
YH Finnish QIG •
YI Italian QIG
YN Norwegian QIG •
YP Portuguese QIG •
YS Spanish QIG •
YW Swedish QIG •
Typical Model Number: 8732C T 12 N0 M4

(1) Only available with Rosemount 8711 Integral Mount Flowtube

(2) NAMUR compliant operation and the Alarm latch options are preset at the factory and can not be changed in the field.

(3) Option Code must be ordered for both flowtube and transmitter.

41
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

ROSEMOUNT 8742C ORDERING INFORMATION

Model Product Description Availability


8742C Magnetic Flowmeter Transmitter with FOUNDATION fieldbus (flowtube ordered separately) •
Code Transmitter Output
F FOUNDATION fieldbus protocol, comes with standard Analog Input Integrator Function block and •
Backup LAS
Code Power Supply Voltage
AC 90–250 V ac, 50–60 Hz •
DC 15-50 V dc •
Code Product Certifications
N0 Factory Mutual (FM) Class I, Division 2, Class II/III Division 1, approval for nonflammable fluids; •
Canadian Standards Association (CSA) Class I, Division 2 Approval
CE Marking;
N5 Factory Mutual (FM) Class I, Division 2, Class II/III Division 1, approval for flammable fluids •
E1 EEx de IIB + H2 T6, Hydrogen gas, flameproof approval •
E5(1) Factory Mutual (FM) Class I, Division 1, Class II/III Division 1, explosion proof approval •
ED ATEX EEx d IIB T6, Flame-proof approval •
K0 Factory Mutual (FM) Class I, Division 2, Class II/III Division 1, approval and •
Canadian Standards Association (CSA) Class 1, Division 2,
Class II/III Division 1 with intrinsically safe fieldbus output
K1 ATEX EEx de [ia] IIB + H2 T6 •
Hydrogen gas flameproof approval with intrinsically safe fieldbus output and FISCO compliant
K5(1) Factory Mutual (FM) Class I, Division 1, Explosion Proof with intrinsically safe Fieldbus output •
KD ATEX EEx d [ia] IIB T6, Flame-proof approval with intrinsically safe Fieldbus output and FISCO compliant •
Code Options
PlantWeb Software Functions •
A01 Proportional/Integral/Derivative (PID) function blocks •
D01 Product and process diagnostics: grounding/wiring diagnostic, electrode fault diagnostic, •
and high process noise detection
D11 Product diagnostic: grounding/wiring diagnostic and electrode fault diagnostic •
D21 Process diagnostic: high process noise detection •
Transmitter Options •
B4 Remote Mount for 2” Pipe Mount Bracket (transmitter junction box and mounting bracket included)(1) •
C1 Custom Configuration (completed configuration data sheet (CDS) required with order) •
D1 High Accuracy Calibration [0.25% of rate from 3 to 30 ft/s (0.9 to 10m/s)] matched flowtube •
J1 CM 20 Conduit Adapter •
J2 PG 13.5 Conduit Adapter •
M5 Local Display •
Code Quick Installation Guide (QIG) Language Options (Default is English)
YA Danish QIG •
YD Dutch QIG •
YF French QIG •
YG German QIG •
YH Finnish QIG
YI Italian QIG •
YN Norwegian QIG •
YP Portuguese QIG •
YS Spanish QIG
YW Swedish QIG •
Typical Model Number: 8742C F AC N0 A01

(1) Only available with Rosemount 8711 Integral Mount flowtube

42
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

ROSEMOUNT 8705 ORDERING INFORMATION


Code Product Description Availability
8705 Magnetic Flowmeter Flowtube •
Code Lining Material
A PFA •
T Teflon (PTFE) •
F Tefzel (ETFE) •
P Polyurethane •
N Neoprene •
L Linatex natural rubber •
Other lining materials available upon request. Consult factory. For availability by Line Size; see chart below •
Code Electrode Material / Electrode Type
Two Measurement Electrodes
SA 316L Stainless Steel •
HA Hastelloy C-276 •
TA Tantalum •
PA 90% Platinum—10% Iridium •
NA Titanium •
Two Measurement Electrodes + Third Grounding Electrode
(Not available with W3 on electrode housing configuration on 8” (200 mm) and under line sizes.)
SE 316L Stainless Steel •
HE Hastelloy C-276 •
TE Tantalum •
PE 90% Platinum—10% Iridium •
NE Titanium •
Two Bulletnose Measurement Electrodes
(Available in 2-inch (50 mm) and larger sizes only)
SB 316L Stainless Steel •
HB Hastelloy C-276 •
Two Removable Measurement Electrodes
(Available in 2-inch (50 mm) and larger sizes only); requires W3 option
SR 316L Stainless Steel •
HR Hastelloy C-276 •
Other electrode materials and types available upon request. Consult factory.
Lining Material (from above)
Code
Code Line Size Code A Code T Code F Code P N, and L
005 ½ inch (15 mm) NA • • NA NA
010 1 inch (25 mm) • • • • •
015 1½ inch (40 mm) • • • • •
020 2 inch (50 mm) • • • • •
030 3 inch (80 mm) • • • • •
040 4 inch (100 mm) • • • • •
060 6 inch (150 mm) • • • • •
080 8 inch (200 mm) • • • • •
100 10 inch (250 mm) • • • • •
120 12 inch (300 mm) • • • • •
140 14 inch (350 mm) NA • • • •
160 16 inch (400 mm) NA • • • •
180 18 inch (450 mm) NA • NA • •
200 20 inch (500 mm) NA • NA • •
240 24 inch (600 mm) NA • NA • •
300 30 inch (750 mm) NA • NA • •
360 36 inch (900 mm) NA • NA • •
Continued on Next Page

43
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

Code Flange Material, Type, and Rating Availability


C1 Carbon Steel, ASME B16.5 (ANSI) Class 150 (30- and 36-inch AWWA C207 Table 3 Class D Flat Face) •
C2 Carbon Steel, MSS SP44 Class 150 (30-inch [750 mm] only) •
C3 Carbon Steel, ASME B16.5 (ANSI) Class 300 •
C6 Carbon Steel, ASME B16.5 (ANSI) Class 600 (1) •
(maximum pressure: 1000 psig; available in 0.5- through 10-inch [15 through 250 mm])
C7 Carbon Steel, ASME B16.5 (ANSI) Class 600 (2) •
(with lining material codes P, N, and L only; available in 1- through 8-inch [25 through 200 mm])
C9 Carbon Steel, ASME B16.5 (ANSI) Class 900 (2) (3) •
(with lining material codes, P, N, L only, available in 1- through 8- [25 through 200mm])
S1 304 Stainless Steel, ASME B16.5 (ANSI) Class 150 •
(30- and 36-inch AWWA C207 Table 3 Class D Flat Face)
S2 304 Stainless Steel, MSS SP44 Class 150 (30- through 36-inch [750 through 900 mm] only) •
S3 304 Stainless Steel, ASME B16.5 (ANSI) Class 300 •
S6 304 Stainless Steel, ASME B16.5 (ANSI) Class 600(1) •
(maximum pressure: 1000 psig; available in 0.5- through 10-inch [15 through 250 mm])
S7 304 Stainless Steel, ASME B16.5 (ANSI) Class 600(2) •
(lining material codes P, N, and L only; available in 1- through 8-inch [25 through 200 mm])
S9 304 Stainless Steel, ASME B16.5 (ANSI) Class 900 (2) (3) •
(with lining material codes, P, N, L only, available in 1- through 8- [25 through 200mm])
P1 316 Stainless Steel, ASME B16.5 (ANSI) Class 150 •
(1/2- and 24-inch AWWA C207 Table 2 Class D Flat Face), pipe and flange
P3 316 Stainless Steel, ASME B16.5 (ANSI) Class 300, pipe and flange •
CD Carbon Steel, DIN PN 10 (not available for flange sizes 15 to 150 mm) •
CE Carbon Steel, DIN PN 16 (not available for flange sizes 15 to 80 mm) •
CF Carbon Steel, DIN PN 25 (not available for flange sizes 15 to 150 mm) •
CH Carbon Steel, DIN PN 40 (not available for flange sizes 750 to 900 mm) •
CK Carbon Steel, AS2129 Table D (available flange sizes 15 to 900 mm) (4) •
CL Carbon Steel, AS2129 Table E (available flange sizes 15 to 900 mm) (4) •
SD Stainless Steel, DIN PN 10 (not available for flange sizes 15 to 150 mm) •
SE Stainless Steel, DIN PN 16 (not available for flange sizes 15 to 80 mm) •
SF Stainless Steel, DIN PN 25 (not available for flange sizes 15 to 150 mm) •
SH Stainless Steel, DIN PN 40 (not available for flange sizes 750 to 900 mm) •
Other flange materials, types, and ratings available upon request. Consult factory.
Code Electrode Housing Configuration
W0 Sealed, welded housing •
W1 Sealed, welded housing with pressure relief •
W3 Sealed, welded housing with separate electrode compartments •
Code Product Certifications
N0 Factory Mutual (FM) Class I, Division 2 Approval for nonflammable fluids; •
Canadian Standards Association (CSA) Class I, Division 2 Approval,
CE Marking
N5 Factory Mutual (FM) Class I, Division 2 Approval for flammable fluids •
E1(5) ATEX EEx e ia IIC T3...T6, Increased Safety Approval (with intrinsically safe electrode) •
KD ATEX EEx e ia IIC T3...T6, Increased Safety Approval (with intrinsically safe electrode) •
N1 ATEX EEx nA [L] IIC Type n Approval •
Continued on Next Page

44
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

Code Options Continued


Optional Grounding Rings (6)
G1 316L SST Grounding Rings •
G2 Hastelloy C-276 Grounding Rings (0.5- through 12-inch (15 through 300 mm) flowtube line sizes) •
G3 Titanium Grounding Rings (0.5- through 12-inch (15 through 300 mm) flowtube line sizes) •
G4 Tantalum Grounding Rings (0.5- through 8-inch (15 through 200 mm) flowtube line sizes) •
G5 Single 316L SST Grounding Ring •
G6 Single Hastelloy C-276 Grounding Ring (0.5- through 12-inch (15 through 300 mm) flowtube line sizes) •
G7 Single Titanium Grounding Ring (0.5- through 12-inch (15 through 300 mm) flowtube line sizes) •
G8 Single Tantalum Grounding Ring (0.5- through 8-inch (15 through 200 mm) flowtube line sizes) •
Optional Lining Protectors(6)
L1 316L SST Lining Protectors •
L2 Hastelloy C-276 Lining Protectors (0.5- to 12-inch (15 to 300 mm) flowtube line sizes) •
L3 Titanium Lining Protectors (0.5- to 12-inch (15 to 300 mm) flowtube line sizes) •
Optional Sanitary Connections(7)
A3 Sanitary 3-A (0.5- to 3-inch (15 to 80 mm) sizes only); •
ASME B16.5 (ANSI) Class 150 to Tri-Clamp Adapter
A4 Sanitary 3-A (2.5-inch (64 mm) sizes only); ASME B16.5 (ANSI) Class 150 to Tri-Clamp Adapter •
A5 Cherry Burrell Sanitary I-line (0.5- to 3-inch (15 to 80 mm) sizes only); •
ASME B16.5 (ANSI) Class 150 to I-line Adapter
A6 Cherry Burrell Sanitary I-line (2.5-inch (64 mm) sizes only); •
ASME B16.5 (ANSI) Class 150 to I-Line Adapter
Continued on Next Page

45
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

Code Options Continued


Other Options
B3 Integral Mount with Rosemount 8732C/8742C Transmitter •
D1 High Accuracy Calibration [0.25% of rate from 3-30 ft/s (0.9-10 m/s)] •
matched flowtube and transmitter system(8)
DT Heavy Duty Tagging •
DW NSF Drinking Water Certification 24 inch (600mm) (0.5- to 24-inch [15 to 600 mm] sizes only) •
PTFE Teflon or ETFE liners (0.5- through 24-inch [15 through 600mm]),
SS electrode material (9)
H1 Rosemount 8701 flowtube lay length (available for 0.5- to 12-inch (15 to 300 mm) line sizes); •
spool piece or spacer: ASME B16.5 (ANSI) Class 150 or Class 300 flange and 304 stainless steel pipe
H2 Rosemount 8701 Flowtube lay length (available for 0.5- to 16-inch (15-400 mm) line sizes.) •
H5 Foxboro® Model 2800 lay length (available for 3.0 to 18 inch (80 to 450mm line sizes)) spool piece: ASME •
B16.5 (ANSI) Class 150 flange and 304 stainless steel pipe.
H7 ABB Fischer & Porter® Model CopaX and MagX lay length (available for 0.5 to 12in (15 to 3000 mm line •
sizes)) spool piece: ASME B16.5 (ANSI) Class 150 flange and 304 stainless steel pipe.
J1 CM20 Conduit Adapter •
J2 PG13.5 Conduit Adapter •
Q4 Inspection Certificate-Calibration Data, ISO 10474 3.1B, (Flow Cal., 1-10ft/s [0.3-3 m/s]) •
Q5 Hydrostatic Testing Certification (place on sales order, not on model string) •
Q8 Material Traceability Certificate per DIN 3.1 B •
Q9 Material Traceability Certificate (electrodes only) per DIN 3.1B 337 •
Q15 NACE Certificate (10) •
Q66 Welding Procedure Qualification Record Documentation •
Q67 Welder Performance Qualification Record Documentation •
Q68 Welding Procedure Specification Documentation •
Q70 Inspection Certificate Weld Examination, ISO 10474 3.1B
– 0.5- to 12-inch [15-300 mm] flowtube line sizes
– 14- to 18-inch [350-450 mm] flowtube line sizes •
– 20- to 36-inch [500-900 mm] flow tube line sizes
Typical Model Number: 8705 T SA 040 C1 W0 N0

(1) Electrode options limited to two measurement electrodes or two measurement electrodes + third grounding electrode.

(2) Electrode options limited to two measurement electrodes or two measurement electrodes only.

(3) Lining Protectors not available.

(4) Option code limited to Liner material T, P, or F; cannot be ordered with ground rings, lining protectors AX or HX options.

(5) For use with E1 / K1 transmitter.

(6) Grounding Rings and Lining Protectors provide the same fluid grounding function. Lining Protectors available in Teflon (PTFE) and Tefzel (ETFE) only.

(7) Sealed, welded housing (Option Code W0 or W1) required. Only available with Teflon (PTFE) lining material (Option Code T) and 316L Stainless Steel,
Hastelloy C-276, and 90% Platinum-10% Iridium electrode material (Option Codes S, H, and P). Not available with integral mount Rosemount 8712
transmitter. Sanitary connection codes A4 and A6 only available in line size code 020.

(8) Option Code must be ordered for both flowtube and transmitter.

(9) Pending approval, consult factory for availability.

(10) Order as a separate line item. Not available with PFA liners. Electrode code SHN only.

46
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

ROSEMOUNT 8707 ORDERING INFORMATION

Code Product Description Availability


8707 High-Signal Magnetic Flowmeter Flowtube •
Code Lining Material
A PFA (1)
T Teflon (PTFE) •
F Tefzel (ETFE) •
P Polyurethane •
N Neoprene •
L Linatex natural rubber •
Other lining materials available upon request. Consult factory. For availability by Line Size, see chart below. •
Code Electrode Material / Electrode Type
Two Measurement Electrodes
SA 316L Stainless Steel •
HA Hastelloy C-276 •
TA Tantalum •
PA 90% Platinum —10% Iridium •
NA Titanium •
Two Measurement Electrodes + Third Grounding Electrodes (2)
SE 316L Stainless Steel •
HE Hastelloy C-276 •
TE Tantalum •
PE 90% Platinum —10% Iridium •
NE Titanium •
Two Bulletnose Measurement Electrodes (Available in 2-inch (50 mm) and larger line sizes only)
SB 316L Stainless Steel •
HB Hastelloy C-276 •
Lining Material (from above)
Code
Code Line Size Code A Code T Code F P, N, and L
030 3 inch (80 mm) • • • •
040 4 inch (100 mm) • • • •
060 6 inch (150 mm) • • • •
080 8 inch (200 mm) • • • •
100 10 inch (250 mm) • • • •
120 12 inch (300 mm) • • • •
140 14 inch (350 mm) NA • • •
160 16 inch (400 mm) NA • • •
180 18 inch (450 mm) NA • NA •
200 20 inch (500 mm) NA • NA •
240 24 inch (600 mm) NA • NA •
300 30 inch (750 mm) NA • NA •
360 36 inch (900 mm) NA • NA •
Code Flange Material, Type, and Rating
C1 Carbon Steel, ASME B16.5 (ANSI) Class 150 (30- and 36-inch AWWA C207 Table 2 Class D Flat Face) •
C2 Carbon Steel, MSS SP44 Class 150 (30-inch [750 mm] only) •
C3 Carbon Steel, ASME B16.5 (ANSI) Class 300 •
S1 304 Stainless Steel, ASME B16.5 (ANSI) Class 150 (30- and 36-inch AWWA C207 Table 2 Class D Flat Face) •
S2 304 Stainless Steel, MSS SP44 Class 150 (30- through 36-inch [750 through 900 mm] only) •
S3 304 Stainless Steel, ASME B16.5 (ANSI) Class 300 •
Continued on Next Page

47
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

Code Housing Configuration Availability


W0 Sealed, welded housing •
W1 Sealed, welded housing with pressure relief •
W3 Sealed, welded housing with separate electrode compartments •
Code Product Certifications
N0 Factory Mutual (FM) Class I, Division 2 Approval for nonflammable fluids; •
Canadian Standards Association (CSA) Class I, Division 2 Approval
N5 Factory Mutual (FM) Class 1, Division 2 Approval for flammable fluids •
Code Options
Optional Grounding Rings(3)
G1 316L SST Grounding Rings •
G2 Hastelloy C-276 Grounding Rings (3- through 12-inch (80 through 300 mm) flowtube line sizes) •
G3 Titanium Grounding Rings (3- through 12-inch (80 through 300 mm) flowtube line sizes) •
G4 Tantalum Grounding Rings (3- through 8-inch (80 through 200 mm) flowtube line sizes) •
G5 Single 316L SST Grounding Ring •
G6 Single Hastelloy C-276 Grounding Ring (3- through 12-inch (80 through 300 mm) flowtube line sizes) •
G7 Single Titanium Grounding Ring (3- through 12-inch (80 through 300 mm) flowtube line sizes) •
G8 Single Tantalum Grounding Ring (3- through 8-inch (80 through 200 mm) flowtube line sizes) •
Optional Lining Protectors(3)
L1 316L SST Lining Protectors
L2 Hastelloy C-276 Lining Protectors (3- through 12-inch (80 through 300 mm) flowtube line sizes) •
L3 Titanium Lining Protectors (3- through 12-inch (80 through 300 mm) flowtube line sizes) •
Other Options
B3 Integral Mount with Rosemount 8732C (8742C) transmitter •
D1 High Accuracy Calibration [0.25% of rate from 3-30 ft/s (0.9-10 m/s)] matched flowtube and transmitter system(4) •
D2 Dual Flowtube Calibration Numbers on Rosemount 8712H and 8712C Transmitters •
DT Heavy Duty Tagging •
H1 Rosemount 8703 flowtube lay length (available for 3- through 4-inch (80 through 100 mm) line sizes); •
spool piece: ASME B16.5 (ANSI) Class 150 or Class 300 flange and 304 stainless steel pipe
H2 Rosemount 8703 flowtube lay length (available for 3- through 16-inch (80 through 400 mm) line sizes •
H5 Foxboro® Model 2800 lay length (available for 3.0 through 18 inch (80 through 450mm line sizes)) spool piece: •
ASME B16.5 (ANSI) Class 150 flange and 304 stainless steel pipe.
H7 ABB Fischer & Porter® Model CopaX and MagX lay length (available for 0.5 through 12in (15 through 3000 mm •
line sizes)) spool piece: ASME B16.5 (ANSI) Class 150 flange and 304 stainless steel pipe or teflon spacer.
J1 CM20 Conduit Adapter •
J2 CM13.5 Conduit Adapter •
Q4 Inspection Certificate-Calibration Data, ISO 10474 3.1B, (Flow Cal., 1-10ft/s [0.3-3 m/s]) •
Q5 Hydrostatic Testing Certification •
Q8 Material Traceability Certificate per DIN 3.1 B •
Q9 Material Traceability Certificate (electrodes only) per DIN 3.1B 337 •
Q15 NACE Certification (5) •
Q66 Welding Procedure Qualification Record Documentation •
Q67 Welder Performance Qualification Record Documentation •
Q68 Welding Procedure Specification Documentation •
Q70 Inspection Certificate Weld Examination, ISO 10474 3.1B
– 3- to 12-inch [80-300 mm] flowtube line sizes •
– 14- to 18-inch [350-450 mm] flowtube line sizes •
– 20- to 36-inch [500-900 mm] flow tube line sizes •
Typical Model Number: 8707 T SA 040 C1 W0 N0

(1) Electrode options limited to two Hastelloy C-276 or two 90% Platinum-10% Iridium electrode material.
Cannot be ordered with ground rings or lining protectors.

(2) Available for 10 inch and larger line sizes only.

(3) Grounding Rings and Lining Protectors provide the same fluid grounding function. Lining Protectors available with Teflon (PTFE) and Tefzel (ETFE) lining
material only.

(4) Option Code must be ordered for both flowtube and transmitter.

(5) Order as a separate line item (not available with PFA liners). Electrode code SHN only.

48
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

ROSEMOUNT 8711 ORDERING INFORMATION

Model Product Description Availability


8711 Magnetic Flowmeter Flowtube (flangeless construction) •
Code Lining Material
T Tefzel (ETFE) •
S Teflon (PTFE) (not available with 0.15 and 0.30 inch [4 and 8 mm] line sizes) •
Code Electrode Material / Electrode Type
Two Measurement Electrodes
SA 316L Stainless Steel •
HA Hastelloy C-276 •
TA Tantalum •
PA 90% Platinum—10% Iridium •
NA Titanium •
Two Measurement Electrodes + Third Grounding Electrode
SE 316L Stainless Steel •
HE Hastelloy C-276 •
TE Tantalum •
PE 90% Platinum—10% Iridium •
NE Titanium •
Two Bulletnose Measurement Electrodes (Available in 2-inch (50 mm) and larger line sizes only)
SB 316L Stainless Steel •
HB Hastelloy C-276 •
Code Line Size
15F 0.15 inch (4 mm) (not available with Teflon (PTFE) lining material) •
30F 0.30 inch (8 mm) (not available with Teflon (PTFE) lining material) •
005 ½ inch (15 mm) •
010 1inch (25 mm) •
015 1½ inch (40mm) •
020 2 inch (50mm) •
030 3 inch (80 mm) •
040 4 inch (100 mm) •
060 6 inch (150 mm) •
080 8 inch (200 mm) •
Code Transmitter Mounting Configuration
R Remote •
U Integral, mounted to Rosemount 8732C/8742C transmitter •
Continued on Next Page

49
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

Code Mounting Kit Availability


Expanded Kit: includes two alignment rings (where applicable), threaded SST studs, and nuts
1 ASME B16.5 (ANSI) Class 150 •
2 DIN PN 10/16 (8 inch [200 mm] has a PN 10 mounting kit only) •
3 ASME B16.5 (ANSI) Class 300 •
4 DIN PN 25/40 (8 inch [200 mm] has a PN 25 mounting kit only) •
Standard Kit: includes two alignment rings (where applicable)
5 ASME B16.5 (ANSI) Class 150 •
6 DIN PN 10/16 (8 inch [200 mm] has a PN 10 alignment rings only) •
7 ASME B16.5 (ANSI) Class 300 •
8 DIN PN 25/40 (8 inch [200 mm] has a PN 25 alignment rings only) •
Code Product Certifications
N0 Factory Mutual (FM) Class I, Division 2 Approval for nonflammable fluids; •
Canadian Standards Association (CSA) Class I, Division 2 Approval;
CE Marking
N5 Factory Mutual (FM) Class 1, Division 2 Approval for flammable fluids •
N1 ATEX EEx nA [L] IIC Type n Approval •
E1(1) ATEX EEx e ia IIC T3...T6, increased safety approval with intrinsically safe electrodes •
E5 Factory Mutual (FM) Class 1, Division 1, Explosion-Proof Approval •
(available with integral mount Rosemount 8732C or remote mount transmitters)
CD ATEX EEx e ia IIC Increased Safety Approval with intrinsically safe electrodes •
Code Options
D1 High Accuracy Calibration [0.25% of rate from 3-30 ft/s (0.9-10m/s)] matched flowtube and transmitter •
DT Heavy Duty Tagging •
DW NSF Drinking Water Certification (Stainless steel electrode material only) •
G1 316L SST Grounding Rings •
G2 Hastelloy C-276 Grounding Rings •
G3 Titanium Grounding Rings •
G4 Tantalum Grounding Rings •
J1 CM20 Conduit Adapter •
J2 CM13.5 Conduit Adapter •
Q4 Inspection Certificate-Calibration Data, ISO 10474 3.1B, (Flow Cal., 1-10ft/s [0.3-3 m/s]) •
Q5 Hydrostatic Testing Certification (place on sales order, not on model string) •
Q8 Material Traceability Certificate per DIN 3.1B •
Q9 Material Traceability Certificate (Electrodes only) per DIN 3.1B •
Q15 NACE Certification (2) •
Q66 Welding Procedure Qualification Record Documentation (6- and 8-inch (150- and 200mm) line sizes only) •
Q67 Welder Performance Qualification Record Documentation •
Q68 Welding Procedure Specification Documentation •
Q70 Inspection Certificate Weld Examination, ISO 10474 3.1B •
–.15 through 8-inch [4- through 300 mm] flowtube line sizes •
Typical Model Number: 8711 T SA 030 U 5 N0

(1) For use with E1 / K1 transmitter.

(2) Order as a separate line item. Electrode code SHN only.

ROSEMOUNT 8714 DQ4 ORDERING INFORMATION


Model Description Availability
8714DQ4 Reference Calibration Standard •

50
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

Tagging Cable Requirements for Remote


The flowtube and transmitter will be tagged, at no Transmitters
charge, in accordance with customer requirements.
Description Length P/N
Transmitter tag character height is
Signal Cable (20 AWG) ft 08712-0061-0001
0.125 in. (3.18 mm).
Belden 8762, Alpha 2411 m 08712-0061-0003
Flowtube tag: 40 character maximum. equivalent
Transmitter tag: see Configuration Data Sheet for Coil Drive Cable ft 08712-0060-0001
character maximum. (14 AWG) Belden 8720, m 08712-0060-0003
Alpha 2442 equivalent
Ordering Procedure Combination Signal and ft 08712-0752-0001
Coil Drive Cable m 08712-0752-0003
To order, select the desired flowtube and/or (18 AWG)(1)
transmitter by specifying model codes from the (1) Combination signal and coil drive cable is not recommended for
ordering table. high-signal magmeter system. For remote mount installations,
combination signal and coil drive cable should be limited to less
For remote transmitter applications, note the cable than 100 ft (30 m).

specification requirements. Remote transmitter installations require equal


Flowtubes and transmitters must be selected from lengths of signal and coil drive cables. Integrally
Product Data Sheet 00813-0100-4727. mounted transmitters are factory wired and do not
require interconnecting cables.
Standard Configuration Lengths from 5 to 1,000 feet (1.5 to 300 meters) may
Unless the Configuration Data Sheet is completed, be specified, and will be shipped with the flowtube.
the transmitter will be shipped as follows: Cable longer than 100 feet (30 meters) is not
Engineering Units: ft/sec recommended for high-signal systems.
4 mA (1 V dc): 0
20 mA (5 V dc): 30 Custom Configuration (Option Code C1)
Tube Size: 3-inch If Option Code C1 is ordered, the Configuration Data
Empty Pipe: Off Sheet (CDS) must be submitted at the time of order.
Flowtube Calibration Number: 1000005010000000
The Rosemount 8732C and the 8742C Transmitter
are factory-calibrated with the attached flowtube size
and appropriate calibration number.

51
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

Rosemount 8712 C/D/U/H and 8732C


Configuration Data Sheet
CONFIGURATION DATA SHEET
(Unless otherwise specified, transmitter will be shipped as ★Default Factory Configuration).

Customer: ———————————————————————— P.O No: ————————————————————————

Transmitter Model No: —————————————————— Line Item: ————————————————————————


SST Transmitter Tag No:
(72 characters maximum)
|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|
|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|
Transmitter Tag (Software): |—|—|—|—|—|—|—|—| (8 characters maximum)
OUTPUT INFORMATION (Minimum Information Necessary for Startup)

Fluid Description = _______________________________________________________________________________________


Units = □ ★ Ft/Sec
□ Gal/Sec □ Gal/Min □ Gal/Hr □ Gal/Day □ Special (Fill out the information in the box below.
□ Liters/Sec □ Liters/Min □ Liters/Hr □ Liters/Day Examples on back side.)
□ Imp. Gal/Sec □ Imp. Gal/Min □ Imp. Gal/Hr □ Imp. Gal/Day
□ CuMeters/Sec □ CuMeters/Min □ CuMeters/Hr □ CuMeters/Day
□ Meters/Sec ➀Volume Units = |—|—|—|—|
(4 characters)
20 mA (Full Scale) Value = ★ 30 ft/sec __________
➁Base Units
4mA (Zero) Value = ★ 0.0 ft/sec __________
□ Gallons Imp. Gallons
Damping = ★ 2.0 seconds __________
□ ★ Eng Units □ Liters Cubic Meters
Flowrate Display □ Eng Units and □ % Span and
and % span Gross Total Gross Total □ Feet Meters
Totalizer Display □ ★ Net and Gross □ Forward and ➂Conversion No. __________________
Reverse ➃Time Base
Hardware Jumpers □ Hours Minute
A) Write Protect □★ Off □ On □ Day Second
B) Analog Loop Power □★ Internal □ External ➄Rate Units = |—|—|—|—|
(4 characters)
C) Alarm Option □★ High □ Low

Flowtube Size = ————— .15 – 60” ★ 3”

Flowtube Calibration No.: | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | 16 Digits from flowtube ★ 1000005010000000

SST Flowtube Tag No.: ——————————————————————

AUXILIARY OUTPUT INFORMATION (Not Required for Typical Start-up)


Empty Pipe Functionality: Operation Mode:
□ ★ Off □ ★ Normal
□ On □ Filter (Operating Mode for 8712H is not available.)
Pulse Scaling: Coil Pulse Mode:
□ ★.03 ft □ ★ 6 Hz
□ 1 Pulse = ————— units □ 30 Hz (Coil Pulse Mode for 8712H is preset at the factory.)
Auxiliary Output Control: Signal Processing:
□ ★ Zero Flow □ ★ Off
□ Reverse Flow □ On
Low Flow Cutoff: _________ ★ 90 No. Samples
_________ ★ 0.1 ft/sec _________ ★ 2 Max.% Limit %
Pulse Width: _________ ★ 2 Time Limit Sec
_________ ★ 0.5 m/s

52
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

FLOWTUBE INFORMATION (Data Only – Does Not Affect Transmitter Output)

Descriptor: | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | (16 characters maximum) Date (day/month/year): | — | — | |—|—| |—|—|

Message:
| — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | (32 characters maximum)

Flowtube Tag No. (Software): | — | — | — | — | — | — | — | — | (8 characters maximum)

Flowtube Serial No: | — | — | — | — | — | — | — | (7 characters maximum)

Enter either the Rosemount flowtube model number or select one option from each of the following groups of options:
Flowtube Model No.—————————————————————
Or:
Electrode Material: Electrode Type:
□ ★316 SST □ ★ Standard
□ Titanium □ Tantalum □ Bullet □ Standard, plus Grounding
□ Alloy-20 □ Platinum-Iridium □ Ultrasonic □ Special
□ Hastelloy C® □ Special
Flange Material: Liner Material:
□ ★Carbon Steel □ 304 SST □ ★Teflon® (PTFE)
□ Standard □ Special □ Tefzel® (ETFE) □ Ryton® □ Neoprene
□ Polyurethane □ Ceramic □ Special

□ ★ 4–20 mA, scaled pulse, and auxiliary output with simultaneous digital signal based on HART® protocol (Rosemount 8712C/U/H only)
□ ★ 4–20 mA and scaled pulse with simultaneous digital signal based on HART protocol (Rosemount 8732C only)
□ 4–20 mA and auxiliary output with simultaneous digital signal based on HART protocol (Rosemount 8732C only)
□ Burst mode of HART digital process variable

Burst mode output options:


□ Primary variable in engineering units.
□ Primary variable in percent of range.
□ All dynamic variables in engineering units.
□ All dynamic variables in engineering units and the primary variable mA value.

□ Multidrop Communications(1)
Choose transmitter address (1-15)(2): ———————
SPECIAL UNITS – OUTPUT INFORMATION EXAMPLE
(If applicable, complete the information on the front as shown in the examples below).

Example 1: Example 2:
User wants Barrels per day User wants Kgs per minute
(1 Barrel = 31 gallons): (Density = 1.05 Kg/Liter with 1 Kg = 0.9524 liters):

➀Vol Unit = BARL ➀Vol Unit = Kg


➁Base Units = Gallons ➁Base Units = Liters
➂Conversion No. = 31.0 ➂Conversion No. = 0.9524
➃Time Base = Day ➃Time Base = Minute
➄Rate Units = B/DY ➄Rate Units = Kg/m
(1) This option fixes the transmitter’s analog output at 4 mA.

(2) Default transmitter address is 1 if multidrop communication is selected.

53
Product Data Sheet
00813-0100-4727, Rev MA
Rosemount 8700 Series May 2004

Rosemount 8742C Configuration Data Sheet


Unless otherwise specified, transmitter will be shipped as ★ (default factory configuration).

CONFIGURATION DATA SHEET

Customer:——————————————————————— P.O. No.: ——————————————

Transmitter Model No.:—————————————————— Item:————————————————

Transmitter Tag (software): |—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|


(32 characters maximum)

OUTPUT INFORMATION (Minimum Information Necessary for Startup)

Fluid Description:————————————————————
Hardware Jumpers
Units = A) Write Protect □★ Off □ On
□ ★ Ft./Sec □ Ft/Min □ Ft/Hr □ Ft/d B) Simulate □★ Off □ On
□ m/Sec □ m/Min □ m/Hr □ m/d
□ Gal/Sec □ Gal/Min □ Gal/Hr □ Gal/d Flowtube Size = ————— 0.15 – 60” ★ 3”
□ m3/Sec □ m3/Min □ m3/Hr □ m3/d
3
□ cm /Sec □ cm3/Min □ cm3/Hr □ cm3/d Flowtube Calibration No. = ★ 1000005010000000
□ Ft3/Sec □ Ft3/Min □ Ft3/Hr □ Ft3/d
□ Liters/Sec □ Liters/Min □ Liters/Hr □Liters/d |—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|—|
□ Igal/Sec □ Igal/Min □ Igal/Hr □ Igal/d
□ bbl*/Sec □ bbl/Min □ bbl/Hr □ bbl/d SST Flowtube Tag No.:——————————————————
□ cF/Sec □ cF/Min □ cF/Hr □ cF/d
□ lb/Sec □ lb/Min □ lb/Hr □ lb/d Damping:———————————————————————
□ kg/Sec □ kg/Min □ kg/Hr □ kg/d ★ minimum 0 seconds
□ ton/Sec □ ton/Min □ ton/Hr □ ton/d
□ ston/Sec □ ston/Min □ ston/Hr □ ston/d Density: —————————————— □ ★ 62.4 lb/ft3
*bbl = 31.5 gal US (119 liters) □ kg/m3
(required for mass units only, mass units in italics)

AUXILIARY OUTPUT INFORMATION (Normally Not Required for Typical Start-up)

Coil Pulse Mode: □ ★ 5 Hz □ 37.5 Hz

Empty pipe functionality: □★ Off □ On Low Flow Cutoff—————ft/sec ★ 0.04


(minimum of 0 ft/sec for low flow cutoff)
Local Display Language: □ ★ English □ Spanish
□ French □ German

Continued On Next Page

54
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

FLOWTUBE INFORMATION (Data Only – Does Not Affect Transmitter Output)

Descriptor: | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — |
(32 characters maximum)

Flowtube Serial No.: | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — | — |


(32 characters maximum)

Enter either the Rosemount flowtube model number or select one option from each of the following groups of options:

Flowtube Model No.:—————————————————————


Or:
Electrode Material: Flange Material:
□ ★ 316 SST □ Tantalum □ 304 SST □★ Carbon Steel
□ Titanium □ Platinum-Iridium □ 316L SST □ Wafer
□ Alloy-20 □ Hastelloy® C-276 □ Other
□ Other
Flange Type:
Electrode Type: □ ★ ASME B16.5 (ANSI) Class 150 □ DIN PN 10
□ ★ Standard □ Removable □ ASME B16.5 (ANSI) Class 300 □ DIN PN 16
□ Bulletnose □ Other □ ASME B16.5 (ANSI) Class 600 □ DIN PN 25
□ Standard and Grounding Electrodes □ ASME B16.5 (ANSI) Class 900 □ DIN PN 40
□ Wafer □ DIN PN 64
□ Other
Liner Material:
□ ★ Teflon® (PTFE)
□ Tefzel® (ETFE)
□ PFA
□ Polyurethane
□ Natural Rubber
□ Linatex® Rubber
□ Neoprene
□ Other

55
Product Data Sheet
00813-0100-4727, Rev MA
May 2004 Rosemount 8700 Series

Rosemount and the Rosemount logotype are registered trademarks of Rosemount Inc.
PlantWeb is a mark of one of the Emerson Process Management companies.
All other marks are the property of their respective owners.
PlantWeb and DeltaV are trademarks of the Fisher-Rosemount Group of Companies.
HART is a registered trademark of the HART Communication Foundations.
Foundation is a trademark of the Fieldbus Foundation.
Hastelloy and Hastelloy C are registered trademarks of Haynes International.
Teflon and Tefzel are registered trademarks of E.I. du Pont de Nemours &Co.
Tri-Clamp is a registered trademark of Tri-Clover, Inc. of the Alfa-Laval Group.
Foxboro and I/A Series are registered trademarks of the Foxboro Company.
ABB Fischer & Porter is a registered trademark of ABB Company.
Cover Photo: 8700_01ba.tif

Emerson Process Management


Rosemount Inc. Emerson Process Management Flow Emerson Process Management
8200 Market Boulevard Groeneveldselaan 6-8 Singapore Pte Ltd.
Chanhassen, MN 55317 USA 3903 AZ Veenendaal 1 Pandan Crescent
T (U.S.) 1-800-999-9307 The Netherlands Singapore 128461
T (International) (952) 906-8888 Tel 31 (0) 318 549 549 Tel (65) 777-8211
F (952) 949-7001 Fax 31 (0) 318 549 559 Fax (65) 777-0947
Tel 0800-966 180 (U.K. only) AP.RMT-Specialist@emersonprocess.com
www.rosemount.com Fax 0800-966 181 (U.K. only)

¢00813-0100-4727>¤
© 2004 Rosemount Inc. All rights reserved.