0% found this document useful (0 votes)
4K views35 pages

Error - Code - List - Ver5.00 5.10

This document contains a list of error codes and their associated messages and meanings for a medical imaging system. It includes fatal error codes from -100 to -107 that cause the system to shut down or enter a non-recoverable state. These errors involve serious issues like sensor or capture board failures that require a service person to address. The document provides the error code, message, cause, and recommended response upon receiving the error.

Uploaded by

Gulenay
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
4K views35 pages

Error - Code - List - Ver5.00 5.10

This document contains a list of error codes and their associated messages and meanings for a medical imaging system. It includes fatal error codes from -100 to -107 that cause the system to shut down or enter a non-recoverable state. These errors involve serious issues like sensor or capture board failures that require a service person to address. The document provides the error code, message, cause, and recommended response upon receiving the error.

Uploaded by

Gulenay
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 35

Error Code List for Version 5.00/5.

10
ror Code List for Versi

Error Code List for Version 5.00/5.10

Alert Window
ERR key

Message area #1

Message area #2

Message area #3

Message area #4

Reply keys

Seriousness Automati Blinking of


Meaning
(Message Area #1) c Display “ERR”
The whole screen blinks in red and the
alert window appears automatically.
1 Fatal Error Yes Yes
System cannot be recovered. A service
person must be called.
A red ERR key blinks and alert window
appears automatically.
2 Alert Yes Yes
Exposure cannot be continued unless the
user performs some operation.
A red ERR key is displayed and alert
window appears automatically. The ERR
3 Error Yes No key does not blink.
The possibility of recovering the system
by the user is higher than Alert.
An yellow ERR key blinks.
4 Warning No Yes Alert window does not appear unless the
ERR key is touched.
An yellow ERR key is displayed. It does
not blink.
5 Attention No No
Alert window does not appear unless the
ERR key is touched.

-1-
Error Code List for Version 5.00/5.10
ror Code List for Versi
1 Fatal Error

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-100 OK #1: Fatal Error
Asynchronous #2: Sensor Unit: SelfCheck Error
(SNS) #3:Detected an Error.
Fatal #4: ER (n) (xxxx)
xxxx: Error code

[Cause]
An undefined error has been found in the sensor A/D
board.

[Operation after Reply Key is touched]


Nothing
-101 OK #1: Fatal Error
Asynchronous #2:Capture Board SelfCheck Error
(CAP) #3:Detected Capture Board Error.
Fatal #4:ER(0) (xxxx)
xxxx: Error code

[Cause]
An error has been found in the capture board.
Undefined error

[Operation after Reply Key is touched]


Nothing
-102 OK #1: Fatal Error
Return value #2:Capture Device Open Error
(CAP) #3:Failed to open capture device.
Fatal
[Cause]
Capture board cannot be detected. Failed to
communicate with capture board H8.

[Operation after Reply Key is touched]


The system enters a fatal condition. All the sensor units
will therefore be disconnected. A service person needs
to be called in. However, you can transfer the image
and use the study list.
API code Cause of the error
-14 Error in starting the task in DLL (Failed to begin thread.)
-15 Error in DLL resource (Driver does not start.)
-95 Number of sensor units is incorrect (No sensor unit is
connected.)
-98 Capture device cannot be found.

-2-
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-103 Shut Down #1: Fatal Error
Return value #2:DMA Timeout Error
(SNS) #3:Failed to transmit the data.
Fatal
[Cause]
DMA from the capture board has failed.

[Operation after Reply Key is touched]


For the system enters a fatal condition, it must be shut
down.
-104 OK #1: Fatal Error
Return value #2:Requested I/O pending error
(CAP) #3:Failed to access the device.
Fatal
[Cause]
Error in making an access to the capture board

[Operation after Reply Key is touched]


The system enters a fatal condition. All the sensor units
will therefore be disconnected. A service person needs
to be called in. However, you can transfer the image
and use the study list.
-105 OK #1: Fatal Error
Return value #2:CXDCAP.INI Open Error
(CAP) #3:Install CXDCAP.INI file.
Fatal
[Cause]
The cxdcap.ini file has not been installed.

[Operation after Reply Key is touched]


The system enters a fatal condition. All the sensor units
will therefore be disconnected. A service person needs
to be called in. However, you can transfer the image
and use the study list.
API code Cause of the error
-97 No cxdcap.ini file is found.

-3-
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-106 OK #1: Fatal Error
Asynchronous #2:Capture Board SelfCheck Error
(CAP) #3:Detected Capture Board Error.
Fatal
[Cause]
An error has been found in the capture board such as
one related to the data line.

[Operation after Reply Key is touched]


The system enters a fatal condition. All the sensor units
will therefore be disconnected. A service person needs
to be called in. However, you can transfer the image
and use the study list.
Code Messages
0x21FE "COM TRNSQ ADD(ER 21FE)"
0x21FD "COM TRNSQ RMV(ER 21FD)"
0x21FC "COM RCVQ ADD(ER 21FC)"
0x21FB "COM RCVQ RMV(ER 21FB)"
0x71F9 "CAP FLASHROM ERR(ER 71F9)"
-107 Shut Down #1: Fatal Error
Asynchronous #2:Power Supply Error
(CAP) #3:Shuts down after 10 seconds.
Fatal
[Cause]
Error in the power supply or fan

[Operation after Reply Key is touched]


Shutting down starts. The system automatically shuts
down after 10 seconds even if the SHUT DOWN key is
not touched. A service person needs to be called in.
Code Messages
0x61FE None
-522 OK #1: Fatal Error
Asynchronous #2:File Open Error:filename
Fatal #3:File is not found.

[Cause]
For some reason, XXXX.BIN and XXXX.DCM files do
not exist. Or, they cannot be read.

[Operation after Reply Key is touched]


Shutting down starts.
Although this problem can be solved by using the
service tool, users must not use the service tool on their
own. Take necessary actions in accordance with the
service person’s instruction.

-4-
Error Code List for Version 5.00/5.10
ror Code List for Versi
2 Alert

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-4 OK #1: Alert
Return value #2:System Info Error
Alert #3:Driver info is updated.

[Cause]
The driver information of NVRAM and the one obtained
from H8 are different from the former information. This
problem occurs when the version of the driver has been
updated or when the connected units have been
changed. If this error occurs when no such changes
have been made, it is attributed to another problem. In
this case, a service person needs to be called in.

[Operation after Reply Key is touched]


The driver information of NVRAM is renewed to the one
obtained from H8.
-5 OK #1: Alert
#2:System Info Error

-6 OK #1: Alert
Return value #2:System Info Error
Alert #3:A/D board info is updated.

[Cause]
The A/D board information of NVRAM and the one
obtained from H8 are different from the former
information. This problem occurs when the version of
the A/D board has been updated or when the connected
units have been changed. If this error occurs when no
such changes have been made, it is attributed to
another problem. In this case, a service person needs to
be called in.

[Operation after Reply Key is touched]


The A/D board information of NVRAM is renewed to the
one obtained from H8.

-5-
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-7 OK #1: Alert
Return value #2:System Info Error
Alert #3:LANMIT info is updated.

[Cause]
The LANMIT information of NVRAM and the one
obtained from H8 are different from the former
information. This error occurs when the version of
LANMIT has been updated or when the connected units
have been changed. If this error occurs when no such
changes have been made, it can be attributed to
another problem. In this case, a service person needs to
be called in.

[Operation after Reply Key is touched]


The LANMIT information of NVRAM is renewed to the
one obtained from H8.
-8 OK #1: Alert
#2:System SelfCheck Error
#3:Call a service person and let him/her know the error
code.

-115 OK #1: Alert


Asynchronous #2:Capture Board Error
(CAP) #3:Detected Capture Board Error.
Alert #4:ER(0) (xxxx)
xxxx: Error code

[Cause]
An undefined error has been found in the capture board.

[Operation after Reply Key is touched]


Nothing
-116 OK #1: Alert
Asynchronous #2:Sensor Unit:SelfCheck Error
(SNS) #3:Detected an Error.
Alert #4:ER(n) (xxxx) n: Sensor Index (1 to 4)
xxxx: Error code

[Cause]
An undefined error has been found in the sensor A/D
board.

[Operation after Reply Key is touched]


Nothing

-6-
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-5002 - #1: Alert
Return value #2:OPU-CCR RPC Error
Alert #3:OPU-CCR RPC Error.

[Cause]
For some reason, the CCR software has shut down.

[Operation after Reply Key is touched]


-6001 OK #1: Alert
Return value #2:OPU-CCR RPC Error
Alert #3:Operation has been canceled.

[Cause]
In the RPC interface between the CCR and OPU, the
size of the structure specified by the OPU was wrong.

[Operation after Reply Key is touched]


Touch OK after you confirm it.

-7-
Error Code List for Version 5.00/5.10
ror Code List for Versi
3 Error

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
--- OK #1: Error
#2:File Open Error:filename

--- OK #1: Error


Return value #2:
Error #3:Card Reader Connection Error. Call service

[Cause]
The magnetic card reader is not connected properly.
Connection or serial communication settings is wrong.

[Operation after Reply Key is touched]


Settings change for not using magnetic card reader.
-1 OK #1: Error
Return value #2: Error
Error #3:

[Cause]
Undefined error. This error indicates that it is not defined
by the CCR manager. The CCR manager controls
various tasks. If an error not defined in tasks occurs,
error code “-1” is given to the error and thus displayed.
This error rarely appears because now most errors are
defined. Take necessary actions referring to the system
log for the CCR.
Local undefined errors output by the OPU can be
identified since they are displayed in the messages box
of the OS.

[Operation after Reply Key is touched]


After you confirm it, touch OK.
-2 OK #1: Error
Return value #2:Insufficient tmp disk space. Output the saved data or
Error erase the backed up data.
#3:

[Cause]
The remaining space in the temporary storage directory
dttmp has become less than the specified MB.

[Operation after Reply Key is touched]


Touch OK after you confirm it.

-8-
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-3 OK #1: Error
#2:Insufficient que disk space. Output the saved data or
erase the backed up data.

-120 OK #1: Error


Asynchronous #2:Sensor Unit: SelfCheck Error
(SNS) #3:Detected an Error.
Error #4:ER(n) (xxxx)
n: Sensor Index (1 to 4)
xxxx: Error code

[Cause]
An undefined error has been found in the sensor A/D
board.

[Operation after Reply Key is touched]


Nothing
-121 OK #1: Error
Asynchronous #2:Sensor Unit:Sequence Error
(SNS) #3:Sequence Error. Call a service person.
Error
[Cause]
An imaging sequence error has been found in the
sensor A/D board.

[Operation after Reply Key is touched]


When OK key is touched, the sensor sleep command
will be sent. Then, the screen will return to the normal
display.
A service person needs to be called in.
Code Messages
0x0602 "D_WAIT before D_RDY (n)(0602)"
0x0603 "D_WAIT timeout (n)(0603)"
0x0604 "Too Long D_WAIT (n)(0604)"
0x0701 "FAN failed (n)(0701)"
0x0801 "LAMP failed (n)(0801)"
0x0e02 "Photo before D_WAIT (n)(0e02)"

n: Sensor Index (1 to 4)
xxxx: Error code

-9-
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-122 OK #1: Error
Asynchronous #2:Sensor Unit:Control Error
(SNS) #3:Detected Control Error. Call a service person.
Error
[Cause]
An error in the control line has been detected in the
sensor A/D board.

[Operation after Reply Key is touched]


When OK key is touched, the sensor will be
disconnected. (Access will not be made, as the system
is in an “abnormal” state.)
A service person needs to be called in.
Code Messages
0x0601 "Unexpected D_WAIT (n)(0601)"
0x0521 "Internal Power Control (n)(0521)"
This error occurs when the sensing signal of the analog
power unit does not match the control signal.(Added
since Version 4.0)

n: Sensor Index (1 to 4)
xxxx: Error code
-123 OK #1: Error
Asynchronous #2:Sensor Unit:PLD Error
(SNS) #3:Detected PLD Error. Call a service person
Error
[Cause]
A PLD error has been found in the sensor A/D board.

[Operation after Reply Key is touched]


When OK key is touched, the sensor will be
disconnected. (Access will not be made, as the system
is in an "abnormal" state.)
If this error persists even after the system is rebooted, a
service person needs to be called in.
Code Messages
0x0101-0104 "PLD err (n)(0101-0104)"
0x0111-0114 "PLD Config err (n)(0111-0114)"

n: Sensor Index (1 to 4)
xxxx: Error code

- 10 -
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-124 OK #1: Error
Asynchronous Cancel #2:Sensor Unit: Grid Error
(SNS) #3:If you select [OK], the system will separate the Grid
Error component, then change to force-Grid mode. In this
case, the image quality cannot be guaranteed.
If you select [CANCEL], Grid component will be re-
initialize.

[Cause]
A grid error has been found in the sensor A/D board.

[Operation after Reply Key is touched]


When OK is chosen, the fixed-grid mode will be
selected, the sensor sleep command will be sent, and
the screen will return to the normal display.
When Cancel is chosen, the system will be re-initialized
by the sensor sleep command. If the error occurs again,
touch OK key to select the fixed grid mode.
As this mode is intended for emergency use, a service
person needs to be called in.
Calibration and self-test cannot be performed in the
fixed grid mode. (Specification)
Code Messages
0x0a01, 0a02 "Overload with motor (n)(code)"
0x0b01, 0b02 "Rotation in motor (n)(code)"
0x0b11, 0b12 "Initial stop fail (n)(code)"
0x0c01, 0c02 "Initial slit pos. fail (n)(code)"
0x0d01, 0d02 "Too wide initial slit (n)(code)"
0x0b21, 0b22 "Unexpected sequence about grid-related actuators
(n)(code)"

n: Sensor Index (1 to 4)
xxxx: Error code

- 11 -
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-125 OK #1: Error
#2:Sensor Unit:Grid Error
#3:Because of the Grid error, this image may have a
quality problem.

Code Messages
0x0a03, "Overload with motor (n)(code)"
0x0a04
0x0b03, "Rotation in motor (n)(code)"
0x0b04
0x0b13, "Initial stop fail (n)(code)"
0x0b14
0x0c03, "Initial slit pos. fail (n)(code)"
0x0c04
0x0b23, "Unexpected sequence about grid-related actuators
0x0b24 (n)(code)"
0x0b31 "Fail to detect grid Id (n)(code)"

n: Sensor Index (1 to 4)
xxxx: Error code

- 12 -
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-126 OK #1: Error
Asynchronous #2:Sensor Unit:Temperature Error
(SNS) #3:The internal temperature of the Sensor Unit
Error exceeded the temperature limit.
Cannot capture images until the temperature goes
below the limit.
#4:TEMP OVER (n)(0901)

n: Sensor Index (1 to 4)

[Cause]
The sensor temperature has exceeded 45°C . Thus, the
error for aborting the operation has been detected. The
system cannot be used until the temperature lowers,
which will take several tens of minutes.

[Operation after Reply Key is touched]


The sensors automatically enter the sleep mode. The
host computer does nothing. If you try to make the
system ready, the same error occurs again.
Press the [OK] to close the window, after few minutes
later operate again.
Note: The error might be able to be evaded by reducing
an environmental temperature.
-127 OK #1: Error
Asynchronous #2:Sensor Unit
(SNS) #3:Sensor Unit cannot become ready for exposure.
Error Check the following part:
#4:Slide out Adapter Lock Knob (n)(0531)
n: Sensor Index (1 to 4)

[Cause]
In the case of 0x531:
This error occurs when the grid unit has been unlocked.

[Operation after Reply Key is touched]


When OK key is touched, the screen returns to the
normal display without doing anything.

- 13 -
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-128 OK #1: Error
Asynchronous #2:Sensor Unit:Communication Error.
(SNS) #3:Sensor Unit: Signal Line Error Problem has occurred
Error between sensor unit and EO box. Call a service person.
#4:Sensor Unit – EO Box(n)(0611)
n: Sensor Index (1 to 4)

[Cause]
Failure in image signal line of the cable between sensor
unit and EO box, or failure in
related ICs or connectors. (Communication problem of
image signal line)

[Operation after Reply Key is touched]


When OK button is pressed, sensor sleep command will
be issued.
Serviceman call is required.

[Note]
This error occurs even in sleep mode.
This is an error in CXDI-31 cassette sensor.
(The possibility to be generated is in other sensors in
the future.)
-130 OK #1: Error
Asynchronous #2:Capture Board Error
(CAP) #3:Detected Capture Board Error.
Error #4:ER(0) (xxxx)
xxxx: Error code

[Cause]
An undefined error has been found in the capture board.

[Operation after Reply Key is touched]


Nothing

- 14 -
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-131 OK #1: Error
#2:Capture Board Control Error

Code Messages
0x21ff "COM TRNS IL(ER 21ff)"
0x21fa "COM TRNS(ER 21fa)"
0x21f9 "COM DEC(ER 21f9)"
0x21f8 "COM RCV (ER 21f8)"
0x21f7 "IL BU RCV (ER 21f7)"
0x21f0 "COM CRC (ER 21f0)"
0x21ef "COM RCV IL RTN (ER 21ef)"
-132 OK #1: Error
Asynchronous #2:Capture Board Error
(CAP) #3:Detected Capture Board Error. Call a service person
Error
[Cause]
An error other than the errors described herein has
been found in the capture board.

[Operation after Reply Key is touched]


All the sensor units are disconnected to prohibit access
to them. Otherwise, the system will enter a fatal
condition.
A service person needs to be called in.
API code Cause of the error
-204 Timeout while waiting for exposure
-401 READ error from the capture board
-410 READ timeout from the capture board
-421 READ error
-501 Exposure wait API error
Code Messages
0x11fe "IL SNS STS (ER 11fe)"
0x41ff "IL PARAM (ER 41ff)"
0x51ff "BU NO RCV (ER 51ff)"
0x51fe "SNS NOT CNN (ER 51fe)"
0x51fd "PWU NOT CNN (ER 51fd)"
0x51fc "IL BOOT MODE (ER 51fc)"
0x51fb "SNS RN (ER 51fb)"
0x51fa "SNS RN NO RCV (ER 51fa)"
0x51f9 "SNS NOT READY (ER 51f9)"

- 15 -
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-133 OK #1: Error
#2:Capture Board Access Error

API code Cause of the error


-105 An error has occurred during DPRAM communication
with the capture board.
-134 OK #1: Error
Asynchronous #2:Capture Board: Sequence Error
(CAP) #3:Failed to synchronize with the sensor unit. Call a
Error service person.

[Cause]
The system and the Imaging unit cannot be
synchronized.

[Operation after Reply Key is touched]


Nothing
Code
0x11ff DRDY_TMO
-135 OK #1: Error
Asynchronous #2:Capture Board: Sequence Error
(CAP) #3:X-ray monitor in sensor unit failed to detect the
Error beginning of irradiation.

[Cause]
The X-ray monitor have not detected the beginning of
irradiation.

[Operation after Reply Key is touched]


Nothing
Code
0x11f9 NO_ASRT_DPHT (for Level 2)
-136 OK #1: Error
Asynchronous #2:Capture Board: Data Transmit Error
(CAP) #3:Failed to transmit the data. Call a service person.
Error
[Cause]
The image data transfer operation has timed out.

[Operation after Reply Key is touched]


Nothing
Code
0x11fd NO_TRNS_BSY

- 16 -
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-137 OK #1: Error
Asynchronous #2:Capture Board: Control Line Error(-137)
(CAP) #3:Control line error has occurred in capture board.
Error Call service person.

[Cause]
Cable is cut or connection failure

[Operation after Reply Key is touched]


Sensor sleep command will be issued.
0xe000-0xefff COM CONTROL LINE(Err Code)
-140 OK #1: Error
Return value #2:Acquisition Parameter Error
(CAP) #3:Detected Illegal Parameter Error.
Error
[Cause]
An error has been found in the parameters given to the
sensor and capture board.

[Operation after Reply Key is touched]


Nothing
API code Cause of the error
-99 Error in handling of device.
-100 Wrong parameters
-101 Reserved area has not been cleared.
-102 Sensor ID error
-141 OK #1: Error
Return value #2:LUT Write Error
(C1 only) #3:Quit the exam and perform calibration.
Error
[Cause]
The system has failed to verify DMA Write access to the
capture board. Capture failure.

[Operation after Reply Key is touched]


The system can be recovered by rebooting, changing
sensors, or running calibration.
Normally, this error does not occur.
-142 API code Cause of the error
Return value -511 Failed to perform "Cancel Wait Expose".
(CAP) -522 Error in "Cancel Wait Expose" API response
Not Displayed -205 "Cancel Wait Expose" has been rejected.

- 17 -
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-143 OK #1: Error
Asynchro0000 #2:Calibration Table Error
0000nous #3:Perform Calibration.
Error
[Cause]
No gain calibration table has been prepared. Or, the
serial number of the connected sensor unit is different.

[Operation after Reply Key is touched]


Perform calibration to make a gain calibration table.
-144 OK #1: Error
#2:Sensor Unit Configuration
#3:Incorrect Senor Unit Configuration.
Call a service person.
#4:Invalid Serial No (n)(0501)

-302 OK #1: Error


Return value #2:Calibration Error
Error #3:Too Much Dose.
Or collimator is used.

[Cause]
Excessive dose of radiation was used in exposure of
image for white correction.

[Operation after Reply Key is touched]


Perform imaging again.
-303 OK #1: Error
Return value #2:Calibration Error
Error #3:Too Little Dose.
Or collimator is used.

[Cause]
Insufficient dose of radiation was used in exposure of
image for white correction.

[Operation after Reply Key is touched]


Perform imaging again.

- 18 -
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-304 OK #1: Error
Return value #2:Calibration Error
Error #3:Invalid Data. Or collimator is used.

[Cause]
Too much difference in dose acquired by the panels.

[Operation after Reply Key is touched]


Perform imaging again.
-305 OK #1: Error
#2:Calibration Error

-306 OK #1: Error


Return value #2:QA Process Error
Error #3:Image Analysis Error. Adjust image process
parameter manually.

[Cause]
Feature extraction may have failed.

[Operation after Reply Key is touched]


The user must process the image manually.
-310 OK #1: Error
Error #2:REX Value Warning
#3:REX is 999. It is under the minimum limitation value
9999.
Please reconfirm the generator setting or exposure
condition.

[Cause]
This error occurs when the REX value acquired by
image processing after the imaging is less than the
limitation value.

[Operation after Reply Key is touched]


Nothing

[Note]
999: REX value obtained by image processing
9999: lower limit value

- 19 -
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-311 OK #1: Error
Error #2:REX Value Warning
#3:REX is 9999. It is under the maximum limitation
value 999.
Please reconfirm the generator setting or exposure
condition.
If the condition is correct, adjust the image processing.

[Cause]
This error occurs when the REX value acquired by
image processing after the imaging is more than the
limitation value.

[Operation after Reply Key is touched]


Nothing

[Note]
9999: REX value obtained by image processing
999: upper limit value
-402 OK #1: Error
Asynchronous #2:Incomplete End Error
Error #3:The last image in the last study was not saved
completely.

[Cause]
Failed to write in the last image of the previous imaging.

[Operation after Reply Key is touched]


If there remain the previous images, they will be
recovered.
-403 OK #1: Error
Asynchronous #2:Incomplete End Error
Error #3:Do you want store captured images into internal
storage in order to recover them later?
(Name of the patient)
(Number of images)

[Cause]
The last image(s) was not queued due to incomplete
end in the previous imaging.

[Operation after Reply Key is touched]


When Ok is chosen, the recovery process will start.
When Cancel is chosen, the last images will be deleted.
In this case, normal exposure for imaging will start
without recovering the last images. When CANCEL is
chosen, a confirmation message will appear before
deleting the image.

- 20 -
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-509 OK #1: Error
Return value #2:External Storage Error:
Error #3:Selected directory or disk for storage can only be
read.

[Cause]
The external storage medium for the stored data is
read-only.

[Operation after Reply Key is touched]


The system returns to the state before the occurrence of
the error.
-512 OK #1: Error
Return value #2:External Storage Error:
Error #3:Cannot recognize the removable disk. Set the disk
properly and retry.

[Cause]
No medium is found in the removable disk.

[Operation after Reply Key is touched]


The system returns to the state before the occurrence of
the error.
-513 OK #1: Error
#2:Media Control Error

- 21 -
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-514 OK #1: Error
Return value #2:External Storage Error:
Error #3:The exposed images cannot be stored because the
size of folder partition is smaller than the exposed
images. Use a new disk, erase the folder contents or
enlarge the folder size and retry.

[Cause]
An image cannot be stored into a folder in the external
storage medium when the size of the folder is smaller
than the image. Enlarge the folder size in the system
setting.

[Operation after Reply Key is touched]


When RETRY is chosen, the image is stored again,
while selecting CANCEL cancels storing the image to
the external storage medium. When CANCEL is chosen,
a confirmation message appears before deleting the
image.
-517 OK #1: Error
Return value #2:Internal Storage Error:
Error #3:Processed image cannot be stored, because the
capacity of internal storage memory is insufficient.

[Cause]
The remaining space in the internal disk is insufficient.

[Operation after Reply Key is touched]


The system returns to the state before the occurrence of
the error.
-518 OK #1: Error
Return value #2:Internal Storage Error:
Error #3:DICOM image cannot be stored, because the
capacity of internal storage memory is insufficient.

[Cause]
There is not enough space for storing DICOM image.

[Operation after Reply Key is touched]


The system returns to the state before the occurrence of
the error.
-519 OK #1: Error
#2:External Storage Error:

- 22 -
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-520 OK #1: Error
Return value #2:Internal Storage Error:
Error #3:Exposed image cannot be stored, because the
capacity of internal storage memory is insufficient.

[Cause]
The remaining space in the internal disk is insufficient.

[Operation after Reply Key is touched]


The system returns to the state before the occurrence of
the error.
-521 OK #1: Error
Return value #2:Queue File Open Error:
Error #3:Queue file is not found.

[Cause]
Failed to open the queue file. This error occurs when no
queue file exists.

[Operation after Reply Key is touched]


* Under the current system, this error is unlikely to
occur.
-599 OK #1: Error
Return value #2:File Open Error:
Error #3:File is not found.

[Cause]
This error occurs when the system has failed to open
the file during RESEND.

[Operation after Reply Key is touched]


Just touch OK.
The system returns to the state before the occurrence of
the error.
-601 OK #1: Error
Return value #2:Resend Error
Error #3:File may have been automatically deleted.

[Cause]
The image cannot be resend as it was deleted for some
reason while being called out.

[Operation after Reply Key is touched]


The image being displayed on the re-print screen is
made invalid and the image can no longer be resend.

- 23 -
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-5100 OK #1: Error
#2:Sensor Unit:Detect Error
#3:Check the sensor power unit or cable connection. If
you want to recover the system, shut it down.

-5101 OK #1: Error


Asynchronous #2:Sensor Unit:Detect Error
Error #3:Check the sensor unit cable connection. If you want
to recover the system, shut it down.

[Cause]
The cable is not connected though it is setting which
uses the sensor.

[Operation after Reply Key is touched]


The system only requires your confirmation. Touch OK.
Only the problematic sensor is disabled.
-5102 OK #1: Error
Return value #2:Grid Error
Error #3:Remove Grid for this exposure.

[Cause]
This error occurs when grids are detected in the
Buck/Cassette setting, where imaging is expected to be
performed without grids.

[Operation after Reply Key is touched]


Touch OK after the check.
-5202 OK #1: Error
Error #2:Grid ID Error
#3:Change with the recommended grid(XXXX).
Installed grid: (YYYY).

[Cause]
In the case of the sensor 22, this error occurs when the
loaded grid ID and the recommended grid ID do not
match.

[Operation after Reply Key is touched]


After you confirm it, touch OK.

- 24 -
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-5203 OK #1: Error
Error #2:Grid ID Error

-5206 OK #1: Error


Error #2:Grid ID Error
#3:Although this is a fixed grid exposure mode, the
system is in none-grid exposure mode.

[Cause]
This is displayed when exposure mode is selected with
the grid removed when "Cassette sensor" has been set
to "Fixed grid". OK after the check.

[Operation after Reply Key is touched]


After you confirm it, touch OK.
OK #1: Error
Asynchronous #2: UNKNOWN ERROR
Error #3:

[Cause]
An undefined error has been found in the OPU.

[Operation after Reply Key is touched]

- 25 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
4 Warning

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-150 OK #1: Warning
Asynchronous #2:Sensor Unit:SelfCheck Error
(SNS) #3:Detected an Error.
Warning #4:ER(n) (xxxx)
n: Sensor Index (1 to 4)
xxxx: Error code

[Cause]
An undefined error has been found in the sensor A/D
board.

[Operation after Reply Key is touched]


Nothing.
-151 OK #1: Warning
Asynchronous #2:Sensor Unit:Flash ROM Error
(SNS) #3:Detected Flash ROM Error. Call a service person.
Warning
[Cause]
A warning on the flash ROM has been detected in the
sensor A/D board.

[Operation after Reply Key is touched]


When OK key is touched, the screen returns to the
normal display. Imaging can be performed by touching
the exposure mode key again.
Although imaging can be performed, a service person
needs to be called in.
Code Messages
0x0201 "F-ROM ID err (n)(0201)"
0x0301 "F-ROM write err (n)(0301)"
0x0401 "F-ROM erase err (n)(0401)"
n: Sensor Index (1 to 4)
-152 OK #1: Warning
Asynchronous #2:Sensor Unit:Sequence Error
(SNS) #3:This image may have a problem.
Warning #4:Photo on FPN (n)(0e03)
n: Sensor Index (1 to 4)

[Cause]
A minor imaging sequence error has been found in the
sensor A/D board.

[Operation after Reply Key is touched]


When OK key is touched, the screen returns to the
normal display without doing anything. Imaging can be
performed by touching the exposure mode key again.

- 26 -
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-153 OK #1: Warning
Asynchronous #2:Sensor Unit:Temperature Warning
(SNS) #3:The internal temperature of the Sensor Unit
Warning approaches the temperature limit.
Do not leave in the READY condition of the
Sensor Unit..
#4:TEMP OVER (n)(0904)
n: Sensor Index (1 to 4)

[Cause]
The warning sign is detected because the internal
temparature of Imaging unit
has approached the upper bound value.

[Operation after Reply Key is touched]


The sensor selfishly enters the state of sleep. There is
nothing to operate at the host side.
The same error occurs when the user tries to put this
sensor into the state of READY.
In this case, press the [OK] to close the window and
continue to operate. This error might be able to be
evated by reducing the internal temparature or having
the Imaging unit READY STATE TIME shorten.
-160 OK #1: Warning
Asynchronous #2:Capture Board Error
(CAP) #3:Detected Capture Board Error.
Warning #4:ER(0) (xxxx)
xxxx: Error code

[Cause]
An undefined error has been found in the capture board.

[Operation after Reply Key is touched]


Nothing

- 27 -
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-161 OK #1: Warning
#2:Capture Board Error

API code Cause of the error


-601 Error in supervisory API
-621 Response error in supervisory API
-701 Error in sensor status API
-721 Response error in sensor status API
Code Messages
0x11fa "PASS XTMO (ER 11fa)"
0x31fe "DBG TRSQ ADD (ER 31fe)"
0x31fd "DBG TRSQ ADD (ER 31fd) "
0x31fc "DBG TRSQ ADD (ER 31fc) "
0x31fb "DBG TRSQ ADD (ER 31fb)"
-162 OK #1: Warning
Asynchronous #2:Capture Board: Sequence Error
(CAP) #3:X-ray monitor in sensor unit failed to detect the
Warning beginning of irradiation.

[Cause]
X-ray monitor have not detected the beginning of
irradiation.

[Operation after Reply Key is touched]


Nothing
Code
0x11F9 NO_ASRT_DPHT (For Level 1)
-163 OK #1: Error
Asynchronous #2:Capture Board: Sequence Error
Warning #3:X-ray monitor in sensor unit failed to detect the end
(CAP) of irradiation.
Warning
[Cause]
X-ray monitor have not detected the end of irradiation.

[Operation after Reply Key is touched]


Nothing
Code
0x11F8 ER_NONEGT_DPHT

- 28 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-164 OK #1: Warning
Asynchronous #2:Image quality attention
Warning #3:CXDI has stopped irradiation. Check image.
(CAP)
Warning [Cause]
X ray irradiation has been suspended.

[Operation after Reply Key is touched]


Nothing
Code
0x11F7 ER_CXDI_TER_XRAY
-502 RETRY #1: Warning
Asynchronous CANCEL #2:DICOM Connect Error
Warning #3:Cannot connect to target. Check network or port
number setting.

[Cause]
Cannot be connected to the destination.

[Operation after Reply Key is touched]


When RETRY is chosen, the image will be transferred
again.
When CANCEL is chosen, transfer will be canceled
after displaying a confirmation message. Re-transfer is
possible by sending the image from the study list.
-503 RETRY #1: Warning
Asynchronous CANCEL #2:DICOM Transfer Error
Warning #3:Error has occurred during association.

[Cause]
Network was cut off during the communication.

[Operation after Reply Key is touched]


When RETRY is chosen, the image will be transferred
again.
When CANCEL is chosen, transfer will be cancelled
after displaying a confirmation message. Re-transfer is
possible by transferring the image from the study list.

- 29 -
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-504 RETRY #1: Warning
Asynchronous CANCEL #2:DICOM Parameter Error
Warning #3:DICOM transfer Parameter is wrong.

[Cause]
DICOM transfer parameters are not correct.

[Operation after Reply Key is touched]


When RETRY is chosen, the image will be transferred
again.
When CANCEL is chosen, transfer will be canceled
after displaying a confirmation message. Enter the
correct parameters and transfer the image from the
study list again.
-505 RETRY #1: Warning
#2:DICOM Response Error

-507 RETRY #1: Warning


Asynchronous CANCEL #2:DICOM Printer Status Error
Warning #3:CXDI has aborted DICOM transmission because of a
FAILURE in DICOM printer status. Check receiving
device.

[Cause]
DICOM printer status was received properly. However,
the target reported a FAILURE as printer status, and so
the image transfer was cancelled.

[Operation after Reply Key is touched]


When RETRY is chosen, the image will be transferred
again.
When CANCEL is chosen, the transfer will be cancelled
after displaying a confirmation message. Re-transfer is
possible by sending the image from the study list.

- 30 -
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-509 RETRY #1: Warning
Asynchronous CANCEL #2:External Storage Error:
Warning #3:Selected directory or disk for storage can only be
read.

[Cause]
The external storage medium is read-only.

[Operation after Reply Key is touched]


When RETRY is chosen, the image will be stored.
When CANCEL is chosen, the storage of image to
external storage medium will be cancelled after
displaying a confirmation message. Re-transfer is
possible by sending the image from the study list again.
-510 RETRY #1: Warning
CANCEL #2:External Storage Error:

-512 RETRY #1: Warning


Asynchronous CANCEL #2:External Storage Error:
Warning #3: Cannot recognize the removal disk. Set the disk
properly and retry.

[Cause]
Disk is not inserted.

[Operation after Reply Key is touched]


When RETRY is chosen, the image will be stored.
When CANCEL is chosen, storing of image will be
cancelled after displaying a confirmation message. Re-
transfer is possible by sending the image from the study
list.

- 31 -
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-514 RETRY #1: Warning
Asynchronous CANCEL #2:External Storage Error:
Warning #3:The exposed images cannot be stored because the
size of folder partition is smaller than the exposed
images. Use a new disk, erase the folder contents or
enlarge the folder size and retry.
(The size of folder partition: xxxxMB)
(The size of exposed images: xxxxMB)
*Set the value in parentheses to the extended massage
area by user.

[Cause]
Images cannot be stored into the folder when the folder
size of the external storage medium is smaller than a
study.

[Operation after Reply Key is touched]


When RETRY is chosen, the image will be stored.
When CANCEL is chosen, storing of image will be
cancelled after displaying a confirmation message. Re-
transfer is possible by transferring the image from the
study list.
-515 RETRY #1: Warning
#2:Internal Storage Error:

- 32 -
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-516 RETRY #1: Warning
Asynchronous CANCEL #2:Internal Storage Error:
Warning #3:Selected directory for storage does not exist, or
attribution is incorrect.

[Cause]
There is no directory in the destination, or recording
cannot be performed due to system function error.

[Operation after Reply Key is touched]


When RETRY is chosen, the image will be stored into
the internal disk.
When CANCEL is chosen, storing of image will be
cancelled after displaying a confirmation message.
Retransfer is possible by sending the image from the
study list.
-517 RETRY #1: Warning
Asynchronous #2:Internal Storage Error:
Warning #3:Processed image cannot be stored, because the
capacity of internal storage memory is insufficient.

[Cause]
The space in the internal disk is insufficient.

[Operation after Reply Key is touched]


QA process will be performed again.
-518 RETRY #1: Warning
Asynchronous CANCEL #2:Internal Storage Error:
Warning #3:DICOM image cannot be stored, because the
capacity of internal storage memory is insufficient.

[Cause]
There is not enough space for storing DICOM image.

[Operation after Reply Key is touched]


When RETRY is chosen, image will be stored into the
internal disk.
When CANCEL is chosen, storing of image will be
cancelled after displaying a confirmation message.

- 33 -
Error Code List for Version 5.00/5.10
ror Code List for Versi

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-519 RETRY #1: Warning
CANCEL #2:External Storage Error:

-520 OK #1: Warning


Asynchronous #2:Internal Storage Error:
Warning #3:Exposed image cannot be stored, because the
capacity of internal storage memory is insufficient.

[Cause]
Space in the internal disk is insufficient.

[Operation after Reply Key is touched]


When RETRY is chosen, the image will be stored into
the internal disk.
When CANCEL is chosen, storing of image will be
cancelled after displaying a confirmation message.
-6000 OK #1: Warning
Asynchronous #2:Calibration Error
Warning #3:STAND(SENSOR-ID:XXX) Time have passed since
the last calibration
Calibrate now.

[Cause]
This error occurs when calibration has not been
performed within the set time.

[Operation after Reply Key is touched]


The calibration screen appears to prompt you to perform
calibration.

- 34 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
5 Attention

Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-506 OK #1: Attention
Asynchronous #2:DICOM Response Warning
Attention #3:Although transfer finished properly, receiving device
reported a warning in DICOM response. Check
receiving device.

[Cause]
DICOM response was received properly. However, the
target reported a warning as a response.

[Operation after Reply Key is touched]


Checks only if the system is OK because the system is
normal.
-508 OK #1: Attention
#2:DICOM Printer Status Warning

-511 OK #1: Attention


Asynchronous #2:External Storage Attention:
Attention #3:Remaining space in external storage is less than
xxx%

[Cause]
(Space x 100)/Total capacity of the external storage
medium has exceeded the XXX %
setindtdsRestPercent.

[Operation after Reply Key is touched]


Only checks if the system is OK.

- 35 -

You might also like