Error - Code - List - Ver5.00 5.10
Error - Code - List - Ver5.00 5.10
10
ror Code List for Versi
Alert Window
ERR key
Message area #1
Message area #2
Message area #3
Message area #4
Reply keys
-1-
Error Code List for Version 5.00/5.10
ror Code List for Versi
1 Fatal Error
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-100 OK #1: Fatal Error
Asynchronous #2: Sensor Unit: SelfCheck Error
(SNS) #3:Detected an Error.
Fatal #4: ER (n) (xxxx)
xxxx: Error code
[Cause]
An undefined error has been found in the sensor A/D
board.
[Cause]
An error has been found in the capture board.
Undefined error
-2-
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-103 Shut Down #1: Fatal Error
Return value #2:DMA Timeout Error
(SNS) #3:Failed to transmit the data.
Fatal
[Cause]
DMA from the capture board has failed.
-3-
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-106 OK #1: Fatal Error
Asynchronous #2:Capture Board SelfCheck Error
(CAP) #3:Detected Capture Board Error.
Fatal
[Cause]
An error has been found in the capture board such as
one related to the data line.
[Cause]
For some reason, XXXX.BIN and XXXX.DCM files do
not exist. Or, they cannot be read.
-4-
Error Code List for Version 5.00/5.10
ror Code List for Versi
2 Alert
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-4 OK #1: Alert
Return value #2:System Info Error
Alert #3:Driver info is updated.
[Cause]
The driver information of NVRAM and the one obtained
from H8 are different from the former information. This
problem occurs when the version of the driver has been
updated or when the connected units have been
changed. If this error occurs when no such changes
have been made, it is attributed to another problem. In
this case, a service person needs to be called in.
-6 OK #1: Alert
Return value #2:System Info Error
Alert #3:A/D board info is updated.
[Cause]
The A/D board information of NVRAM and the one
obtained from H8 are different from the former
information. This problem occurs when the version of
the A/D board has been updated or when the connected
units have been changed. If this error occurs when no
such changes have been made, it is attributed to
another problem. In this case, a service person needs to
be called in.
-5-
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-7 OK #1: Alert
Return value #2:System Info Error
Alert #3:LANMIT info is updated.
[Cause]
The LANMIT information of NVRAM and the one
obtained from H8 are different from the former
information. This error occurs when the version of
LANMIT has been updated or when the connected units
have been changed. If this error occurs when no such
changes have been made, it can be attributed to
another problem. In this case, a service person needs to
be called in.
[Cause]
An undefined error has been found in the capture board.
[Cause]
An undefined error has been found in the sensor A/D
board.
-6-
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-5002 - #1: Alert
Return value #2:OPU-CCR RPC Error
Alert #3:OPU-CCR RPC Error.
[Cause]
For some reason, the CCR software has shut down.
[Cause]
In the RPC interface between the CCR and OPU, the
size of the structure specified by the OPU was wrong.
-7-
Error Code List for Version 5.00/5.10
ror Code List for Versi
3 Error
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
--- OK #1: Error
#2:File Open Error:filename
[Cause]
The magnetic card reader is not connected properly.
Connection or serial communication settings is wrong.
[Cause]
Undefined error. This error indicates that it is not defined
by the CCR manager. The CCR manager controls
various tasks. If an error not defined in tasks occurs,
error code “-1” is given to the error and thus displayed.
This error rarely appears because now most errors are
defined. Take necessary actions referring to the system
log for the CCR.
Local undefined errors output by the OPU can be
identified since they are displayed in the messages box
of the OS.
[Cause]
The remaining space in the temporary storage directory
dttmp has become less than the specified MB.
-8-
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-3 OK #1: Error
#2:Insufficient que disk space. Output the saved data or
erase the backed up data.
[Cause]
An undefined error has been found in the sensor A/D
board.
n: Sensor Index (1 to 4)
xxxx: Error code
-9-
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-122 OK #1: Error
Asynchronous #2:Sensor Unit:Control Error
(SNS) #3:Detected Control Error. Call a service person.
Error
[Cause]
An error in the control line has been detected in the
sensor A/D board.
n: Sensor Index (1 to 4)
xxxx: Error code
-123 OK #1: Error
Asynchronous #2:Sensor Unit:PLD Error
(SNS) #3:Detected PLD Error. Call a service person
Error
[Cause]
A PLD error has been found in the sensor A/D board.
n: Sensor Index (1 to 4)
xxxx: Error code
- 10 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-124 OK #1: Error
Asynchronous Cancel #2:Sensor Unit: Grid Error
(SNS) #3:If you select [OK], the system will separate the Grid
Error component, then change to force-Grid mode. In this
case, the image quality cannot be guaranteed.
If you select [CANCEL], Grid component will be re-
initialize.
[Cause]
A grid error has been found in the sensor A/D board.
n: Sensor Index (1 to 4)
xxxx: Error code
- 11 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-125 OK #1: Error
#2:Sensor Unit:Grid Error
#3:Because of the Grid error, this image may have a
quality problem.
Code Messages
0x0a03, "Overload with motor (n)(code)"
0x0a04
0x0b03, "Rotation in motor (n)(code)"
0x0b04
0x0b13, "Initial stop fail (n)(code)"
0x0b14
0x0c03, "Initial slit pos. fail (n)(code)"
0x0c04
0x0b23, "Unexpected sequence about grid-related actuators
0x0b24 (n)(code)"
0x0b31 "Fail to detect grid Id (n)(code)"
n: Sensor Index (1 to 4)
xxxx: Error code
- 12 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-126 OK #1: Error
Asynchronous #2:Sensor Unit:Temperature Error
(SNS) #3:The internal temperature of the Sensor Unit
Error exceeded the temperature limit.
Cannot capture images until the temperature goes
below the limit.
#4:TEMP OVER (n)(0901)
n: Sensor Index (1 to 4)
[Cause]
The sensor temperature has exceeded 45°C . Thus, the
error for aborting the operation has been detected. The
system cannot be used until the temperature lowers,
which will take several tens of minutes.
[Cause]
In the case of 0x531:
This error occurs when the grid unit has been unlocked.
- 13 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-128 OK #1: Error
Asynchronous #2:Sensor Unit:Communication Error.
(SNS) #3:Sensor Unit: Signal Line Error Problem has occurred
Error between sensor unit and EO box. Call a service person.
#4:Sensor Unit – EO Box(n)(0611)
n: Sensor Index (1 to 4)
[Cause]
Failure in image signal line of the cable between sensor
unit and EO box, or failure in
related ICs or connectors. (Communication problem of
image signal line)
[Note]
This error occurs even in sleep mode.
This is an error in CXDI-31 cassette sensor.
(The possibility to be generated is in other sensors in
the future.)
-130 OK #1: Error
Asynchronous #2:Capture Board Error
(CAP) #3:Detected Capture Board Error.
Error #4:ER(0) (xxxx)
xxxx: Error code
[Cause]
An undefined error has been found in the capture board.
- 14 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-131 OK #1: Error
#2:Capture Board Control Error
Code Messages
0x21ff "COM TRNS IL(ER 21ff)"
0x21fa "COM TRNS(ER 21fa)"
0x21f9 "COM DEC(ER 21f9)"
0x21f8 "COM RCV (ER 21f8)"
0x21f7 "IL BU RCV (ER 21f7)"
0x21f0 "COM CRC (ER 21f0)"
0x21ef "COM RCV IL RTN (ER 21ef)"
-132 OK #1: Error
Asynchronous #2:Capture Board Error
(CAP) #3:Detected Capture Board Error. Call a service person
Error
[Cause]
An error other than the errors described herein has
been found in the capture board.
- 15 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-133 OK #1: Error
#2:Capture Board Access Error
[Cause]
The system and the Imaging unit cannot be
synchronized.
[Cause]
The X-ray monitor have not detected the beginning of
irradiation.
- 16 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-137 OK #1: Error
Asynchronous #2:Capture Board: Control Line Error(-137)
(CAP) #3:Control line error has occurred in capture board.
Error Call service person.
[Cause]
Cable is cut or connection failure
- 17 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-143 OK #1: Error
Asynchro0000 #2:Calibration Table Error
0000nous #3:Perform Calibration.
Error
[Cause]
No gain calibration table has been prepared. Or, the
serial number of the connected sensor unit is different.
[Cause]
Excessive dose of radiation was used in exposure of
image for white correction.
[Cause]
Insufficient dose of radiation was used in exposure of
image for white correction.
- 18 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-304 OK #1: Error
Return value #2:Calibration Error
Error #3:Invalid Data. Or collimator is used.
[Cause]
Too much difference in dose acquired by the panels.
[Cause]
Feature extraction may have failed.
[Cause]
This error occurs when the REX value acquired by
image processing after the imaging is less than the
limitation value.
[Note]
999: REX value obtained by image processing
9999: lower limit value
- 19 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-311 OK #1: Error
Error #2:REX Value Warning
#3:REX is 9999. It is under the maximum limitation
value 999.
Please reconfirm the generator setting or exposure
condition.
If the condition is correct, adjust the image processing.
[Cause]
This error occurs when the REX value acquired by
image processing after the imaging is more than the
limitation value.
[Note]
9999: REX value obtained by image processing
999: upper limit value
-402 OK #1: Error
Asynchronous #2:Incomplete End Error
Error #3:The last image in the last study was not saved
completely.
[Cause]
Failed to write in the last image of the previous imaging.
[Cause]
The last image(s) was not queued due to incomplete
end in the previous imaging.
- 20 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-509 OK #1: Error
Return value #2:External Storage Error:
Error #3:Selected directory or disk for storage can only be
read.
[Cause]
The external storage medium for the stored data is
read-only.
[Cause]
No medium is found in the removable disk.
- 21 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-514 OK #1: Error
Return value #2:External Storage Error:
Error #3:The exposed images cannot be stored because the
size of folder partition is smaller than the exposed
images. Use a new disk, erase the folder contents or
enlarge the folder size and retry.
[Cause]
An image cannot be stored into a folder in the external
storage medium when the size of the folder is smaller
than the image. Enlarge the folder size in the system
setting.
[Cause]
The remaining space in the internal disk is insufficient.
[Cause]
There is not enough space for storing DICOM image.
- 22 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-520 OK #1: Error
Return value #2:Internal Storage Error:
Error #3:Exposed image cannot be stored, because the
capacity of internal storage memory is insufficient.
[Cause]
The remaining space in the internal disk is insufficient.
[Cause]
Failed to open the queue file. This error occurs when no
queue file exists.
[Cause]
This error occurs when the system has failed to open
the file during RESEND.
[Cause]
The image cannot be resend as it was deleted for some
reason while being called out.
- 23 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-5100 OK #1: Error
#2:Sensor Unit:Detect Error
#3:Check the sensor power unit or cable connection. If
you want to recover the system, shut it down.
[Cause]
The cable is not connected though it is setting which
uses the sensor.
[Cause]
This error occurs when grids are detected in the
Buck/Cassette setting, where imaging is expected to be
performed without grids.
[Cause]
In the case of the sensor 22, this error occurs when the
loaded grid ID and the recommended grid ID do not
match.
- 24 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-5203 OK #1: Error
Error #2:Grid ID Error
[Cause]
This is displayed when exposure mode is selected with
the grid removed when "Cassette sensor" has been set
to "Fixed grid". OK after the check.
[Cause]
An undefined error has been found in the OPU.
- 25 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
4 Warning
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-150 OK #1: Warning
Asynchronous #2:Sensor Unit:SelfCheck Error
(SNS) #3:Detected an Error.
Warning #4:ER(n) (xxxx)
n: Sensor Index (1 to 4)
xxxx: Error code
[Cause]
An undefined error has been found in the sensor A/D
board.
[Cause]
A minor imaging sequence error has been found in the
sensor A/D board.
- 26 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-153 OK #1: Warning
Asynchronous #2:Sensor Unit:Temperature Warning
(SNS) #3:The internal temperature of the Sensor Unit
Warning approaches the temperature limit.
Do not leave in the READY condition of the
Sensor Unit..
#4:TEMP OVER (n)(0904)
n: Sensor Index (1 to 4)
[Cause]
The warning sign is detected because the internal
temparature of Imaging unit
has approached the upper bound value.
[Cause]
An undefined error has been found in the capture board.
- 27 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-161 OK #1: Warning
#2:Capture Board Error
[Cause]
X-ray monitor have not detected the beginning of
irradiation.
- 28 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-164 OK #1: Warning
Asynchronous #2:Image quality attention
Warning #3:CXDI has stopped irradiation. Check image.
(CAP)
Warning [Cause]
X ray irradiation has been suspended.
[Cause]
Cannot be connected to the destination.
[Cause]
Network was cut off during the communication.
- 29 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-504 RETRY #1: Warning
Asynchronous CANCEL #2:DICOM Parameter Error
Warning #3:DICOM transfer Parameter is wrong.
[Cause]
DICOM transfer parameters are not correct.
[Cause]
DICOM printer status was received properly. However,
the target reported a FAILURE as printer status, and so
the image transfer was cancelled.
- 30 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-509 RETRY #1: Warning
Asynchronous CANCEL #2:External Storage Error:
Warning #3:Selected directory or disk for storage can only be
read.
[Cause]
The external storage medium is read-only.
[Cause]
Disk is not inserted.
- 31 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-514 RETRY #1: Warning
Asynchronous CANCEL #2:External Storage Error:
Warning #3:The exposed images cannot be stored because the
size of folder partition is smaller than the exposed
images. Use a new disk, erase the folder contents or
enlarge the folder size and retry.
(The size of folder partition: xxxxMB)
(The size of exposed images: xxxxMB)
*Set the value in parentheses to the extended massage
area by user.
[Cause]
Images cannot be stored into the folder when the folder
size of the external storage medium is smaller than a
study.
- 32 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-516 RETRY #1: Warning
Asynchronous CANCEL #2:Internal Storage Error:
Warning #3:Selected directory for storage does not exist, or
attribution is incorrect.
[Cause]
There is no directory in the destination, or recording
cannot be performed due to system function error.
[Cause]
The space in the internal disk is insufficient.
[Cause]
There is not enough space for storing DICOM image.
- 33 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-519 RETRY #1: Warning
CANCEL #2:External Storage Error:
[Cause]
Space in the internal disk is insufficient.
[Cause]
This error occurs when calibration has not been
performed within the set time.
- 34 -
Error Code List for Version 5.00/5.10
ror Code List for Versi
5 Attention
Error Code
Message
(Asynchronous Reply
[Cause]
/Return value) Keys
[Operation after Reply Key is touched]
Seriousness
-506 OK #1: Attention
Asynchronous #2:DICOM Response Warning
Attention #3:Although transfer finished properly, receiving device
reported a warning in DICOM response. Check
receiving device.
[Cause]
DICOM response was received properly. However, the
target reported a warning as a response.
[Cause]
(Space x 100)/Total capacity of the external storage
medium has exceeded the XXX %
setindtdsRestPercent.
- 35 -