Defibrillator Phase3VLM

The Phase 3 is a defibrillator/pacer analyzer designed to test multiphasic defibrillators. It can test monophasic, biphasic and multiphasic defibrillators, AEDs and pacemakers with over 1% accuracy on battery or line power. It is portable with a detachable paddle plate and provides up to 24 hours of battery life. It can capture waveforms and transfer data to a PC via USB for analysis. It provides storage for up to 50 test records and 10 defibrillator waveforms.

Uploaded by

ivan gare
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
49 views2 pages

Defibrillator Phase3VLM

The Phase 3 is a defibrillator/pacer analyzer designed to test multiphasic defibrillators. It can test monophasic, biphasic and multiphasic defibrillators, AEDs and pacemakers with over 1% accuracy on battery or line power. It is portable with a detachable paddle plate and provides up to 24 hours of battery life. It can capture waveforms and transfer data to a PC via USB for analysis. It provides storage for up to 50 test records and 10 defibrillator waveforms.

Uploaded by

ivan gare
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Phase 3

D ef i b r i l l at o r / Tr an s c u t a n e o us P a c e r An a l yz e r
Phase 3 is the first Defibrillator / Pacer Analyzer

Innovation by design
specifically designed to test Pulsed Multiphasic
defibrillators.

Versatility – Phase 3 tests Monophasic, Biphasic


and Pulsed Multiphasic Defibrillators, AEDs and
Transcutaneous Pacers with better than 1%
accuracy on either line or battery power.

Portability – Detachable Paddle Plate provides easy


hands-free testing. Battery provides up to 24 hours
of continuous operation.

USB Connectivity – USB Communication capability


allows real-time waveform capture and efficient data
transfer to your PC.
Phase 3
Real-Time Wave Capture – Capture high-resolution signal
data to a PC running Phase3pc software immediately after discharge.

Test Record Storage – Phase 3 provides storage for up to 50 Test Records


and up to 10 defibrillator discharge waveforms as Wave Records.

Variable Load Module – Our optional, exclusive VLM allows testing defibrillators at
maximum energy with selectable loads from 25 to 175 ohms, in accordance with
AAMI DF-80 and IEC 60601-2-4. High-resolution wave graphics and test data are
obtained with Phase 3VL software — a standard accessory to the VLM.

Phase 3pc Companion Software – Included as a standard accessory, Phase 3pc allows
you to create and edit autosequences, download and store test data and wave graphics
to your PC.

Variable Load Module

Accur acy M att er s • www.datr end.com


Phase 3 – Performance Specifications
Tests Performed ECG Simulator
Defibrillator Energy, Charge Time and Cardioversion Performance Test Waveforms
Automated External Defibrillator (AED) Performance DC Pulse, 4 seconds, Square Wave, 2 Hz, Triangle Wave, 2 Hz
ECG Monitor Performance Sine Wave @ 0.1, 0.5, 10, 20, 40, 50, 60, 70, or 100 Hz
Pacer Sensitivity, Refractory Periods and Noise Immunity
Normal Sinus Rhythm
Pacer Pulse Characteristics
30, 60, 90, 120, 150, 180, 240 or 300 BPM
Energy Measurement, General
Cardioversion, Shock Advisory and AED Test Waveforms
Load resistance: 50 ohms ±1%, non-inductive
Atrial Fibrillation, Coarse
ECG amplitude at defib pads: 1 mV QRS
Atrial Fibrillation, Fine
WAVEFORM (oscilloscope) Output
Asystole 1 (random, low-frequency baseline fluctuation)
High Range: 1000:1 amplitude attenuation
Asystole 2 (flat line/zero volts)
Low Range: 200:1 amplitude attenuation
Supraventricular Tachycardia (SVT-140)
Waveform Playback: 200: 1 time base expansion
VTACH @ 140, 160, 190 BPM
First Phase Pulse Width: 0.1 - 58 msec
Torsades de Pointe @ 200 BPM
Second Phase Pulse Width 0.1 - 58 msec
Coarse Ventricular Fibrillation (CVF) / Fine Ventricular Fibrillation (FVF)
Interphase Delay 0.1 - 58 msec
Square Pulse 1ms to 60 BPM
Tilt 0 - 99.9%
Modulation Frequency 1000 - 8000 Hz Arrhythmia Simulations
Modulation Duty Cycle 0 - 99.9% Second Degree A-V Block, Premature Atrial Contraction (PAC), R-on-T
Test Pulse: 46 Joules ±10% PVC,
Right Bundle Branch Block (RBBB), Premature Ventricular Contraction
Defibrillator High Range Energy Test
(PVC)
Energy Measurement: 0.0 to 600.0 Joules (±1% ±2 LSD)
Multifocal PVC, Run of 5 PVC, Bigeminy, Trigeminy
Voltage Measurement: 0 to 5000 Volts (±1% ±2 LSD)
Current Measurement: 0.0 to 100.0 Amps (±1% ±2 LSD) Pacemaker Test Waveforms
Pulse Width Measurement: Range: 0.5 to 58.36 msec. SQR (square) Pacer Trigger, width = 2, 25, 40, 100 or 200 msec
(±1% ±2 LSD) TRI (triangle) Pacer Trigger, width = 10, 25, 40, 100 or 200 msec
Trigger Level: 80 Volts SSQ (haversine) Pacer Trigger, width = 10, 25, 40, 100 or 200 msec
Playback Amplitude: 1 mV per 1000 volts on Lead II;
1 mV per 2000 volts at defib pads Performance Specifications
Test Pulse: 126 Joules ±10% Output Level: Selectable, 1 mV, 2 mV or 0.5 mV into ECG Lead II
Impedance: 500 ohms (±0.2%), Amplitude: ±2%
Defibrillator Low Range Energy Test
Energy Measurement: 0.0 to 50.0 Joules (±1% ±2 LSD) Non-Volatile Memory
Voltage Measurement: 0 to 1000 Volts (±1% ±2 LSD) Data Capacity
Current Measurement: 0.0 to 20.0 Amps (±1% ±2 LSD) 50 Test Records, 10 Defibrillator Waveform Records, 32 Autosequences
Pulse Width Measurement: 0.5 to 58.36 msec. (±1% ±2 LSD) Test Record Content
Trigger Level: 16 Volts Device ID, Time/date of test, Test type (Manual or Auto), Device type
Playback Amplitude: 1 mV per 200 volts on Lead II; (defib. or AED)
1 mV per 400 volts at defib paddles Up to 10 defibrillator energy tests (or 32 AED energy tests)
Defibrillator Charge Time Test 1 defibrillator charge time test
Charge Time Measurement: 0.0 to 99.9 seconds (±1 LSD) Up to 4 defibrillator cardioversion tests
Up to 12 ECG performance tests
Defibrillator Cardioversion Test Up to 10 pacer pulse tests
Sync Delay Measurement: -200 to +800 msec. (±1 LSD) 1 pacer noise immunity test
Delay Target: 20 to 65 msec window when enabled Up to 2 pacer sensitivity tests
Sync Point: Selectable, peak of ECG Q or R wave Up to 2 pacer refractory period tests
AED Performance Test Interface
Test Method: Verify AED shock advisory for specified arrhythmia LCD (5.2" x 1.5"; 40 characters x 8 lines text; 240 x 64 pixel graphics)
Defibrillator Input: Molex 42820-3212
Pacemaker Pulse Test
Pacemaker Input: 2 x safety-style banana jack (red (+)/black (-))
Pulse Amplitude Measurement: 4 to 250 milliamps, all loads
ECG Simulator Outputs: 10 x safety banana jack (RA; RL; LA; LL; V1-V6)
(±1% ±1 LSD)
Defibrillator/Pacer Waveform Output: 1/8" mono phono jack
Pulse Rate Measurement: 20 to 220 PPM (±1% ±1 LSD)
High-Level ECG Output: 1/8" mono phono jack
Pulse Width Measurement: 0.5 to 58.36 msec. (±1% ±2 LSD)
USB Port: Type "B", USB 1.1 or USB 2.0 compatible.
Test Load Range: 50 to 1600 ohms, in 50 ohm steps
Serial (RS-232) Port: DB9 Male, RS-232C, bi-dir, CTS handshaking, 9600
WAVEFORM output: 50 milliamps per volt, all loads
baud, 8-N-1
Measurement Methods: Average, leading edge, trailing edge, peak
Keyboard Port: PS/2 (6-pin miniDIN female)
Test Pulse: 145 mA ±10%
Power Supply: Internal 12.5V/1.4A-h NiCad, 24 hours of use between
Pacemaker Noise Immunity Test
charges
Test Waveform: 50Hz or 60Hz sine wave
Noise Amplitude Range: 0.18 to 10.00 mV peak-to-peak Environment:
Noise Amplitude Resolution: 0.139 mV 15°C to 40°C, 10% to 90% RH, Indoor Use Only, Category II
Pacemaker Sensitivity Test Dimensions:
Test Waveform: Square (SQR), Triangle (TRI) or Haversine (SSQ) pulse 9.5" W x 8" H x 5.5" D (24cm W x 20cm H x 14cm D)
Waveform Width: 10, 25, 40, 100, or 200 msec.
Amplitude Range: 0.00 to 3.00 mV peak Weight: 3 lbs. (1.4 kg)

Pacemaker Refractory Period Test


Paced Refractory Period (PRP): 50 to 750 msec. (±1 LSD) All specifications subject to change without notice.
Sensed Refractory Period (SRP): 50 to 750 msec. (±1 LSD)

5355 Parkwood Place, Richmond, BC Canada V6V 2N1


Phone: 604.291.7747 • Fax: 604.294.2355
Toll-free (North America only): 800.667.6557
Email : [email protected] © DATREND SYSTEMS INC. JAN 2013