Prelim To Finals rt11
Prelim To Finals rt11
Prelim To Finals rt11
The Processing Room 1.3m(3ft) from the working bench or the processing tanks
This Chapter covers the introduction to the processing room and ⮚ constructed with an air inlet and outlet with exhaust fan – these
the accessories required inside it. This contains the areas found in openings must be designed so that no light shall enter the
the radiology department, processing room design and the darkroom while the processing is done
minimum requirements in putting up a processing room based on
⮚ preferably must have louver blocks, painted black to absorb
the Department of Health’s Administrative Order No. 35, series
whitelight, placed on the lower portion of the door for ventilation
of 1994, “Requirements for the Control of Radiation Hazards
from Clinical Diagnostic X-ray Facilities”.
⮚ adequate supply of water
Major requirements inside the darkroom are likewise found in ⮚ good drainage system
this chapter which are the loading bench, the sink, the different ⮚ contains master tank, for manual processing, or an automatic
types of lightproof entrances, types of illumination inside the processor or both
darkroom and the appropriate filters. Sections in the darkroom is ⮚ luminous interval timer
also discussed here, as well as the characteristicsof a good ⮚ metallic stem dial type thermometer
darkroom. ⮚ mercury or alcohol thermometer with metallic casing
⮚ standard manual processing technique chart that indicates time
development, temperature of the solution and the number of films
AREAS IN RADIOLOGY DEPARTMENT
developed
⮚ walls painted white or placed with white tiles on floors and
1. INFORMATION AREA OR RECEIVING AREA
walls –
⮚ A place where patient complete data receipted gathered and
⮚ Non Reflective Paint.
schedule for examination.
⮚ sink and a working dry bench
2. DRESSING ROOM ⮚ automatic electronic lock to avoid accidental opening of the
⮚ An area where the patient change their dress to laboratory room during processing
gown for examination. ⮚ must contain film hangers and film cassettes of different sizes
⮚ a small processing room measures 15 x 9 ½ feet that permits a
3. EXPOSURE ROOM daily production of about 200 radiographs.
⮚ An area where x-ray examination is conducted for routine and
special radiography, as well as, fluoroscopic procedure of Work surface (Loading Bench)
diagnosing certain disease. ⮚ at least 1.3 m long for the loading and unloading of cassettes
⮚ MOST important room ⮚ should be constructed of a hard, anti-static material that is
easily cleaned
4. DARK ROOM OR PROCESSING ROOM ⮚ A light color is appropriate for all surfaces in the room, as this
⮚ A room devoid of all white light in which the processing cycle reduces the safe-lighting wattage required and allows for the
of theexposed x –ray film takes place. detection of light leaks.
⮚ The most SECOND important room in Radiology Department. ⮚ Beige or mid-grey is a poor choice for darkroom counters or
floors since film emulsion is this color, and a film may not be
5. READING ROOM OR INTERPRETATION ROOM seen if it is placed on the counter.
⮚ An area where Radiologist makes diagnostic report or findings
tointerpret or rule-out pathologic conditions based on the patient Sink
request. ⮚ A large sink is required, installed at waist height to
⮚ Radiograph is viewed by the aid of NEGATOSCOPE or accommodate the cleaning of the rollers from the processor.
VIEWBOX. ⮚ This sink should be at least 60 cm wide by 45 cm deep and be
provided with both hot and cold water.
6. SORTING AREA ⮚ There should be a laundry tub hose at least long enough to
⮚ An area where patient’s diagnostic report were stored. reach into every part of the processor connected to a mixing
⮚ Sorting out or classifying name of film number and the period faucet
were examination was taken
Sections in the Darkroom
7. FILING ROOM 1. DRY SECTION
⮚ Where x – ray films/radiographs are stored. 2. WET SECTION
Processing Room/Darkroom Design ⮚ The minimum distance between these areas should be 4 Feet
⮚ there shall be a separate darkroom constructed near the x-ray LIGHTPROOF ENTRANCE
examination room - An entrance that is easily accessible while providing complete
Requirements protection from the outside.
Based on Department of Health A.O. # 35, Series of 1994
⮚ well-ventilated GENERAL TYPES OF LIGHTPROOF ENTRANCE
1. SINGLE DOOR
⮚ light-tight
⮚ The Radiologist and his Technologist are the only person who
⮚ minimum dimension of 2.0m x 1.5 m
routinely utilizes the processing with an inside bolt
⮚ must be far or away from steam pipes and direct source of heat or lock will suffice.
⮚ must be away or far from any source of radiation ⮚ It is appropriate for small x-ray section
wherein few persons only use the room.
⮚ The door must have an inside lock to ⮚ It is a form of fluorescent bulb or small bulb that gives enough
prevent any accidental opening while light that we need to do such activities like cleaning the room,
working on the said film. washing the tanks, mixing a solution, placing film in the dryer
and unloading films in the hanger.
2. LIGHT LOCK / DOUBLE DOOR ⮚ The white light must be placed away from the storage of the
⮚ This type of entrance may be consisting of a small hall with films.
two doors. One opening and the other to the outside.
⮚ There must be a mechanical interlock so that both doors can be 2. SAFE LIGHT
opened of the same time to move supplies into processing room ⮚ Is design to give enough visibility to accomplish all necessary
or in case of emergency duties in the handling and processing x-ray film without harmful
⮚ Darkroom work can go on without interruption. effect to any unprocessed film
⮚ A safe light consist of all filter used in a lamp housing to
3. LABYRINTH OR MAZE provide light such as spectral quality and intensity that it does not
⮚ This is not provided with doors. fog film exposed to it for reasonable handling period.
⮚ It prevents the entrance of white light to the processing room, ⮚ It is a lamp with color filters that provide sufficient
yet provides air circulation. illumination in the darkroom to give enough visibility in the
⮚ It requires, however at least twice as much floor space as a processing room for the handling and processing of x-ray films.
light lock. ⮚ It consists of lamp housing with 15 - watts bulb and a filter
⮚ The passage way must not be narrower than 2 feet; 6 feet in (6Bfilter) to provide light of such spectral quality and intensity
height with an entrance should be limited to 7 feet. that does not fog the film exposed to it for reasonable handling
period.
⮚ The walls and the ceiling should be painted with flat non-
reflected paint and be illuminated by safe light. ⮚ The safelight must be mounted on the top of dry section.
⮚ a zigzag hall way ⮚ It should not be closer than 5 feet from the work surface.
⮚ without source of light which end up to the processing room ⮚ With blue-sensitive film, or film used with calcium-tungstate
screens, an amber filter is used.
4. REVOLVING DOOR ⮚ The amber filter transmits light that has wavelengths longer
⮚ the best type of door which consists of master door and inside that about 550nm.
door ⮚ The use of an amber filter would fog green-sensitive film (used
⮚ master door has two openings, one with rare- earth screens); therefore a red filter, which transmits
coming from the outside and other from the inside of the light above about 600nm, must be used.
darkroom ⮚ Photographic safelight filters are of the absorption gelatin type
⮚ inside door has only one opening and are designed so that darkrooms have as high degree of
illumination as is consistent with the safe handling of
ILLUMINATION photosensitive materials.
✔ Different lights used in a Darkroom ⮚ Manufacturers recommend replacement of the safelight filters
every two years.
✔ Two levels of lighting should be available in the darkroom.
⮚ At the time of replacement, document the date on the filter
✔ A strong white light should be provided by surface-mounted
with a marking pen.
incandescent fixtures.
⮚ The filter should be inserted in such a way that it will be
✔ Fluorescent bulbs tend to emit a long afterglow that may fog
possible to read the writing and easily determine when the next
films if they are switched off just before the films are unloaded or
replacement is due.
loaded. ✔ Safe lighting should be provided, with manufacturer-
recommended filters in ceiling mounted fixtures. 2 Types of Safelight
✔ 15W bulbs are the maximum recommended for direct safe-
lighting 1. Direct Type- is mounted on the dry side with the light emitted
✔ 25W bulbs may be used if the safe-lighting is indirect. toward the area.
✔ It is important to check current recommended safelight
wattages with the film provider as some films require much 2. Indirect Type- the light emitted by the lamp from the dry side
lower power ratings than these levels (e.g. Kodak recommends a bouncing toward the area.
7.5 watt bulb with their Min-R2000andEV films). ⮚ In the darkroom there must be an unlighted switch for the
✔ There should be a switched duplex receptacle mounted in the white light (1.6m off floor) and another unlighted switch for the
ceiling for the safelight. safelight (1.4 m off floor).
✔ The switches for the safelight and white light should be ⮚ safelight housing must be provided with a safelight filter made
located at the door, but at different heights above the floor to of glass or hard plate
avoid errors in the dark. ⮚ safelight filter must be tinted with amber or red
⮚ must not fog a pre-exposed film within 45 seconds
⮚ safelight bulb must have a wattage of not more than 15 watts
1. WHITELIGHT
⮚ Use to perform many activities, such as:
a. Mixing chemicals WRATTEN GB FILTER
b. Cleaning tanks ⮚ Is the recommended type of filter that the light produce is not
c. Unloading film hangers in the same spectral range as the film sensitivity
⮚ White light should have a sufficient intensity to afford a
general illumination. TYPES OF SAFELIGHT FILTER
1. Amber Filter - for blue sensitive film (Wratten GB) ⮚ The depth is determined by the distance a person can easily
reach on the surface without undue leaning or stretching, 24
2. Red Filter - for green and blue sensitive film (GBX 2) In inches is recommended.
addition, sodium vapor lamps are also used in darkroom.
b. FILM BIN
⮚ Sodium-vapor lamp - a gas-discharge lamp that uses sodium ⮚ It is customary to deep supply of opened boxes of film
in an excited state to produce light. sufficient for immediate needs, located underneath the loading
bench.
low pressure and high pressure ⮚ The most convenient method is to use a lightproof film bin
Low-pressure sodium lamps - highly efficient electrical light possessing compartment in which opened film boxes can be kept.
sources, but their yellow light restricts applications to outdoor
lighting such c. STORAGE COMPARTMENT
as street lamps; only give monochromatic yellow light (589nm) ⮚ The ideal way to store cassette and film holders is in vertical
and so inhibit color vision at night. compartment, so that they are readily accessible but not encroach
on the working surface of the loading bench.
High-pressure sodium lamps - have a broader spectrum of light
than the low-pressure lamps, but still poorer color rendering than d. STORAGE CABINET
other types of lamps
⮚ A cabinet which you can store, different accessories used in
processing such as:
⮚ 5 Watts, tungsten bulb is recommended which is widely used.
✔ Package chemicals
The particular wattage is enough to provide adequate
✔ Mixing accessories
illumination and not have any adverse effect on film.
✔ Towels and etc.
3. WET FILM ILLUMINATOR
⮚ It is mounted above and rears the washing tank which save for e. STORAGE HANGER
the inspection of radiograph during course of hardening and ⮚ Film hangers are usually stored above or below the loading
washing bench
⮚ This illuminator is being switch-on if it is necessary to check
the film within fixer solution during the course of hardening and f. CASSETTE TRANSFER CABINET
when washing a film. ⮚ The cassette transfer cabinet for cassette film holders should be
⮚ Before turning on the illuminator, be sure that there will be no open in to the loading bench, so that these items can be move to
open box of film on the dry side and no film is under the the bench top without undue lifting or carrying.
developer solution or in stop bath to avoid the exposure of the ⮚ The opening to the exposure room makes film folders readily
film. available to the technologist.
Adhesive layer- simply adheres one layer of the film to another 3. SPECTRAL EMISSION
Silver Halide Crystals - refers to the color of light produces by a particular IS, blue
- the active ingredient of the emulsion light-emitting—
- the material that is sensitive to radiation and light
- are photosensitive crystal composed primarily of IMPORTANT RELATIONSHIPS
a. Silver Bromide
b. Silver Chloride A. SILVER HALIDE AND FILM SENSITIVITY
c. Silver Iodide - as the number of silver halide crystals INCREASES, film
sensitivity or speed INCREASES
COMPOSITION OF SILVER HALIDE CRYSTAL - as the size of the silver halide crystals INCREASES, film
a. Silver Bromide sensitivity or speed INCREASES
- most commonly used silver crystal B. FILM SPEED AND FILM SENSITIVITY
- it is a combination of silver nitrate and potassium bromide - the faster the speed of a film, the less radiation exposure
b. Silver Chloride needed to produce a specific density
- an emulsion containing only silver chloride which has the C. SPECTRAL MATCHING AND DENSITY
advantages of rapid development and fixing its principal - to best use a film-screening system, the radiographer must
disadvantage is its lower photosensitivity match the color sensitivity of the film with the color emission of
- it is a combination of silver nitrate and potassium chloride the IS
- Failure to do so results in suboptimal density
c. Silver Iodide
- this is suspended to increase the speed of the emulsion and D. SPECTRAL EMISSION AND SPECTRAL
to increase the time taken from the film to clear in the fixer SENSITIVITY
- it is a combination of a silver nitrate and potassium iodide - the spectral emission of IS must be matched to the spectral
sensitivity of the film
3. ADHESIVE LAYER - the spectral emission of safelight filters in the darkroom
- a thin coating of material between the emulsion and base must be compatible with the spectral sensitivity of the film
- it ensures the uniform adhesion between the emulsion and - safelight filters are places in safelights to produce a
base particular color of light for illumination that matches the spectral
- allows the emulsion and base to maintain proper contact sensitivity of the film
during use and processing - failure to use the right type of safelight filters would result
in unusually high levels of safelight fogging on the film and it
THE RECIPROCITY LAW would appears as a radiographic image with increases density
- States that exposure of film depends in the intensity of x- and decreased contrast
ray of the exit beam and the time the film is exposed to those x-
ray a. GBX filter- used for both the blue- and green-sensitive
- only applicable on direct-exposure film film (GBX stands for green/blue x-ray)
- states the inverse relationship between exposure and time
- intensity x time= constant OD b. GBX Wratten 1A safelight filter is safe for green-
sensitive only
FILM CHARACTERISTICS c. GBX Wratten 6B safelight filter is safe for blue-
1. FILM SPEED sensitive film only
- defined as the intensity of the screen film combination to
x-rays and light FILM HOLDERS
1. HANGERS AND CLIPS b. PHOSPHORESCENCE- occurs when screen phosphors
2. CARDBOARD continue to emit light after the x-ray exposure has stopped. Other
3. CASSETTE WITH IS name, screen lag and afterglow
In what way does the IS reduce patient dose? IS PHOSPHOR MATERIALS AND THEIR SPECTRAL
- the film is placed inside a light-tight cardboard holder then EMISSION
used as image receptor (direct exposure) Phosphors- Spectral Emission Calcium Tungstate (CaWO4)-
- the additions of IS allows the radiographer to use Blue Rare earth elements:
significantly less mAs compared with not using screens (direct
exposure) Lanthanum Oxybromide (LaOBr)- Blue Yttrium Tantalate
- less mAs decreases the patient dose and allows shorter (YTaO4)- Ultraviolet Blue Gadolinium Oxysulfide (Gd2O2s)-
exposure time to be used Green Others:
Barium Lead Sulfite (BaPbSO4)- Blue
LUMINESCENCE Barium Strontium Sulfate (BaSrSO4)- Blue
- the IS operates by a process known as luminescence- the
emission of light from the screen when stimulated by radiation SCREEN SPEED
- the capability of a screen to produce visible light with a
faster screen producing more light than a slower screen (given
- two ways that an IS may luminesce: the same exposure)
a. FLUORESCENCE- refers to the ability of phosphors to
emit visible light only while exposed to x-rays
INTENSIFICATION FACTOR
- it accurately represents the degree to which exposure
factors (and patient dose) are reduced when IS are used.
SCREEN CHARACTERISTICS
1. x-ray absorption
2. Screen conversion efficiency
3. image noise
4. quantum mottle
5. spatial resolution or image blue
6. line- pair test pattern
7. intensification factor
Automatic processing Diffusion- process by which washing works on the film
-paco ang ga develop
Drying- 10 to 15%
roller type - mutuyok ra diretso isman codac in 1956 (developed)
fixing (hypo)- make the image permanent making it available for 4 stages of automatic processing
a very long period of time (archival quality), to remove an - Developing, fixing, washing, drying
exposed silver halide crystals (main purpose) -archival quality
Transform system
washing- terminate the action of fixing, remove fixing chemical - We have the ruler
in order to prevent the film to turn into yellow or coloration - Transport ruler (transport the film, convey the film along its
(hypo rentention) 20secs path with the diameter of 1 in), master ruler (longest ruler and
can be seen at the bottom), planetary ruler (can be seen around
dying - leave at least 10% to 15% moisture in order to avoid the master ruler)
crack, 36 seconds remkve water substances
- Guide shoe (will guide the film as it turn around the master
Processing chemistry: walay explanation ruler and curve metal lip w smooth curves that guides the film
around the bend)
how can a certain image converted into manifest image?
- silver ion (ionic silver) (ang nagtapok sa center) (mga charged - Transport rack assembly (rulers that can be detached, a rack,
particles, positive charge) -(how is it produce?) because ang ang nagpa tuyok, provide for convenient maintenance such as
silber halide crystal na ionized, na detach sa latice paingon sa cleaning and has a guide shoe)
sensitivity center - Cross over rack (rollers found between two tanks that
- pag develop naa tay gitawag na europe (electron use in facilitates the film from one tank to another)
reduction oxidation produces electron) - Drive subsystem (provide power that turn rollers
Buffering agent maintain akalinity -sodium carbonate Developer is 60 to 70 ml for every 40 inc thickness of film
develop
Potassium bromide- anti fogging Fixer is 100 to 110 ml for every 35 cm of film
ARTIFACT
- Any optical density of the radiograph sophor position that can
cause or of the anatomy
- Any undesirable optical density of the radiograph that can
interfere in the diagnosis/image
- A mark which is foreign
Exposure artifact
- Because of exposure/during the xposure that happen inside the
xray room
Processing artifact
- Developed during film processing
P.I lines
- Cause by dirt and stain in the roller
Gelatin build up
- Dirty roller can cause emulsion pick of and gelatin build up
Static marks
- It causes build up electrons in the emulsion and is most
noticeable during the winder low humiditys
- Happens during low humidity
Crescent marks
- Half moon shape densities resukt to being subjected to pressure
Halo marks
- Brought by air bubbles tp the surface of thd film processing
Bleb
- Are drying marks, characterized by whitish on the film causes
by in a adequate film drainage and an increase temperature or
drying
Kiss marks
- Are greenish white smudge at the film surface brought about the
film touching each other in the area of smudge during processing
Streaks
- Either black or white on the radiograph