Zeiss Axiolab5 Manual
Zeiss Axiolab5 Manual
All names of companies and products mentioned in this manual may be trademarks or registered trademarks.
Third party products are cited for information purposes only and this does not represent approval or
recommendation of these products.
Carl Zeiss Microscopy GmbH assumes no liability for the performance or use of such products.
[email protected]
www.zeiss.com/microscopy
CONTENTS
Page
1 INTRODUCTION........................................................................................................... 8
1.1 Notes on instrument safety .................................................................................................... 8
1.2 Warning labels on the microscopes ...................................................................................... 12
1.3 Notes on the warranty ......................................................................................................... 14
3 START-UP.................................................................................................................. 51
3.1 Mounting standard components .......................................................................................... 51
3.1.1 Unpacking and setting up microscope stand ................................................................................. 51
3.1.2 Attaching the binocular tube/photo tube ...................................................................................... 53
3.1.3 Inserting eyepieces or auxiliary microscope or pinhole diaphragm .............................................. 54
3.1.4 Screwing in objectives .................................................................................................................... 55
3.1.5 Inserting and removing P&C reflector modules in/from the reflector turret ................................ 56
04/2020 430037-7444-001 3
Axiolab 5 Contents / List of Illustrations ZEISS
4 OPERATION ............................................................................................................... 79
4.1 Default setting of the microscope ......................................................................................... 79
4.1.1 Setting the inter-pupillary distance on the binocular tube ............................................................79
4.1.2 Setting the viewing height ..............................................................................................................79
4.1.3 Adjusting for ametropia (user's visual impairment) when using eyepiece reticles .......................80
4.2 Illumination and contrast methods in transmitted light ......................................................... 81
4.2.1 Configuring transmitted light brightfield microscopy using the KÖHLER method .........................81
4.2.2 Configuring transmitted light darkfield microscopy using the KÖHLER method ...........................84
4.2.3 Configuring transmitted light phase contrast microscopy .............................................................87
4.2.4 Configuring transmitted light polarization microscopy ..................................................................89
4.2.5 Configuring transmitted light polarization with the conoscopy stand ...........................................98
4.2.6 Configuring transmitted light polarization for conoscopic observation – determining the
optical character of crystals..........................................................................................................109
4.3 Illumination and contrast methods in reflected light ........................................................... 112
4.3.1 Configuring reflected light brightfield microscopy using the KÖHLER method ...........................112
4.3.2 Configuring reflected light darkfield microscopy .........................................................................114
4.3.3 Configuring reflected light polarization – Proof of bireflectance and reflexion pleochroism ......115
4.3.4 Setting reflected light fluorescence..............................................................................................116
04/2020 430037-7444-001 4
Axiolab 5 Contents / List of Illustrations ZEISS
LIST OF ILLUSTRATIONS
Fig. 1-1 Warning labels on the Axiolab 5 stand for transmitted light and reflected light ............................. 12
Fig. 1-2 Warning labels on the Axiolab 5 stand for transmitted light ........................................................... 13
Fig. 2-1 Interface diagram (Example: Axiolab 5 Mat-TL/RL stand) ................................................................ 20
Fig. 2-2 Axiolab 5 stand, Bio-TL...................................................................................................................... 23
Fig. 2-3 Axiolab 5 stand, Bio-TL/FL................................................................................................................. 25
Fig. 2-4 Axiolab 5 stand, Pol-TL ...................................................................................................................... 27
Fig. 2-5 Axiolab 5 stand, Pol-TL/Conoscopy................................................................................................... 29
Fig. 2-6 Axiolab 5 stand, Pol-TL/RL ................................................................................................................ 31
Fig. 2-7 Axiolab 5 stand, Mat-TL/RL ............................................................................................................... 33
Fig. 2-8 Binocular photo tube 30°/23 with fixed graduation 50:50............................................................... 35
Fig. 2-9 Binocular photo tube 20°/23 with toggle graduation 100:0/0:100 .................................................. 35
Fig. 2-10 Binocular photo tube 30°/23 with toggle graduation 100:0/0:100 .................................................. 36
Fig. 2-11 Binocular ergo photo tube 20°/23 with vertical adjustment ........................................................... 37
Fig. 2-12 Mechanical stage 75x50 R with specimen holder ............................................................................ 38
Fig. 2-13 Mechanical stage 75x50 R with specimen holder ............................................................................ 38
Fig. 2-14 Mechanical stage for reflected light 75x30 R with specimen-holding plate .................................... 39
Fig. 2-15 Pol rotary stage ................................................................................................................................. 39
Fig. 2-16 Filter mount on luminous-field diaphragm operating ring for filter d=32x4 mm ............................ 40
Fig. 2-17 Condenser 0.9/1.25 BF, DF, Ph1, Ph2, Ph3 with modulator disk ...................................................... 40
Fig. 2-18 Condenser 0.9/1.25 BF ..................................................................................................................... 40
Fig. 2-19 4-position reflector turret................................................................................................................. 41
Fig. 2-20 Low-power system ............................................................................................................................ 41
Fig. 2-21 Polarizers........................................................................................................................................... 42
Fig. 2-22 Nosepiece of the transmitted/ reflected light polarization stand with mount for compensators .. 42
04/2020 430037-7444-001 5
Axiolab 5 Contents / List of Illustrations ZEISS
04/2020 430037-7444-001 6
Axiolab 5 Contents / List of Illustrations ZEISS
Fig. 4-6 Centering the annular phase diaphragm (light-colored, in the condenser) and the phase ring
(dark-colored, in the object) ............................................................................................................. 88
Fig. 4-7 Components for transmitted light polarization ................................................................................ 90
Fig. 4-8 Determining the polarization direction n' using a synthetic fiber as an example ........................... 91
Fig. 4-9 Schematic diagram of the color charts developed by Michel-Lévy .................................................. 93
Fig. 4-10 Components for circular polarization contrast ................................................................................. 96
Fig. 4-11 Axiolab 5 for transmitted light conoscopy........................................................................................ 98
Fig. 4-12 Determining optical character ........................................................................................................ 100
Fig. 4-13 Components for transmitted light polarization on the conoscopy stand ...................................... 102
Fig. 4-14 Determining the polarization direction n' using a synthetic fiber as an example ......................... 103
Fig. 4-15 Schematic diagram of the color charts according to Michel-Lévy.................................................. 105
Fig. 4-16 Components for circular polarization contrast on conoscopy stand.............................................. 108
Fig. 4-17 Axiolab 5 for transmitted light conoscopy...................................................................................... 110
Fig. 4-18 Determining the optical character.................................................................................................. 111
Fig. 4-19 Microscope settings in reflected light brightfield microscopy ....................................................... 113
Fig. 4-20 Components for reflected light fluorescence ................................................................................. 118
Fig. 5-1 Replacing the fuses in the stand ..................................................................................................... 120
04/2020 430037-7444-001 7
INTRODUCTION
Axiolab 5 Notes on instrument safety ZEISS
1 INTRODUCTION
1.1 Notes on instrument safety
The Axiolab 5 microscopes were engineered, manufactured and tested in accordance with the DIN EN 61010-1
(IEC 61010-1) and IEC 61010-2-101 safety standards for electrical measuring, control and laboratory
equipment.
The microscopes fulfill the requirements as stated in Directive 98/79/EC on in vitro diagnostic medical devices
(IVDD) and bear the marking.
The present manual contains information and safety warnings with which the operator must comply.
The following warning, instructional and informational symbols are used in this manual:
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
Warns the operator to disconnect the instrument from the power supply before opening or doing
any work on the instrument.
ATTENTION
This symbol indicates the risk of damage to the instrument or the system.
NOTE
This symbol indicates information or instructions which must be followed especially carefully.
04/2020 430037-7444-001 8
INTRODUCTION
Axiolab 5 Notes on instrument safety ZEISS
Axiolab 5 microscopes and their original accessories may be used only for the microscopy procedures
described in this manual.
CAUTION
The microscope may only be plugged into an electrical outlet equipped with a safety contact. The
safety feature must not be disabled by using an extension cord which does not have a protective
ground conductor.
CAUTION
Whenever it becomes apparent that any of the safety mechanisms are out of order, the
microscope must be switched off and secured against any inadvertent use.
Please contact the ZEISS Service Department or the Carl Zeiss Microscopy Service before switching
the microscope on again.
CAUTION
The microscopes are equipped with integrated power supply units, which adapt to line voltages
ranging from 100 V to 240 V and frequencies ranging from 50 to 60 Hz. No voltage adjustment is
required on the microscopes themselves.
CAUTION
Always disconnect the microscope from the power supply before you open it or change a fuse!
Make sure that the fuses are suitable for the applied nominal current. Never use any makeshift
fuses and do not short-circuit the fuse holders.
CAUTION
The microscopes do not have any special safety devices to protect users from acid or from
potentially infectious, toxic, radioactive or other samples that may be hazardous to your health.
Compliance with all statutory requirements, especially national accident prevention regulations, is
required when handling such samples.
CAUTION
Only authorized personnel are permitted to operate the instruments. The personnel must have
been instructed about and be aware of the risks involved in using the microscope. The Axiolab is a
precision instrument whose functionality can easily be damaged or even destroyed when handled
improperly.
04/2020 430037-7444-001 9
INTRODUCTION
Axiolab 5 Notes on instrument safety ZEISS
CAUTION
CAUTION
The immersion fluid Immersol 518 N can cause skin irritations. Avoid any contact with skin, eyes
and clothes. Read the safety data sheets on Immersol 518 N, Immersol 518 F and Immersol W
carefully. If skin contact occurs, wash off immediately using lots of water and soap.
If eye contact occurs, flush with water immediately for at least 5 minutes. Seek medical assistance
if irritation continues.
CAUTION
Dispose of immersion fluid Immersol 518 N appropriately: Do not allow it to contaminate surface
water or enter drains or the sewage system.
CAUTION
Do not bring any flammable or easily combustible materials into the light beam.
CAUTION
This device belongs to LED Risk Group 2 as specified in IEC 62471 and emits LED radiation.
Never look into the LED beam of the illuminating device – either with or without optical
instruments. Failure to comply with this warning may result in eye injuries!
ATTENTION
The manufacturer assumes no liability for any use of the microscope, its assemblies or individual
parts if these are used in in a manner or for any purpose other than those described in this
operating manual. This also applies to any maintenance or repair work not performed by
authorized maintenance and repair personnel. Moreover, the warranty and any claims thereunder
will be voided.
ATTENTION
Dirt and dust may affect the microscope’s performance. Protect it by using a dust cover when it is
not in use. Always make sure that the instrument is switched off before covering it up.
04/2020 430037-7444-001 10
INTRODUCTION
Axiolab 5 Notes on instrument safety ZEISS
ATTENTION
Always position equipment in a manner that permits easy disconnection from the power supply.
ATTENTION
ATTENTION
Closing or covering the ventilation slits may result in heat accumulation which could damage the
instrument and even start a fire. Ensure that the ventilation slits are always kept open, uncovered
and clear. Do not insert or drop anything into them or let anything fall into them.
ATTENTION
Do not dispose of defective microscopes in the household waste. Comply with the applicable
statutory regulations for their disposal.
Specimens must also be disposed of appropriately in accordance with applicable statutory
regulations and internal work instructions.
04/2020 430037-7444-001 11
INTRODUCTION
Axiolab 5 Warning labels on the microscopes ZEISS
NOTE
Fig. 1-1 Warning labels on the Axiolab 5 stand for transmitted light and reflected light
04/2020 430037-7444-001 12
INTRODUCTION
Axiolab 5 Warning labels on the microscopes ZEISS
Fig. 1-2 Warning labels on the Axiolab 5 stand for transmitted light
04/2020 430037-7444-001 13
INTRODUCTION
Axiolab 5 Notes on the warranty ZEISS
The manufacturer guarantees that the instrument is free of any material and workmanship defects upon
delivery. If you become aware of any deficiencies, please contact us immediately and take all necessary
precautions in order to avoid further damage. Upon notice of deficiencies, the manufacturer may choose to
correct the deficiencies or to deliver a defect-free instrument at his discretion. Defects due to ordinary wear
and tear (especially on wearing parts) and to improper handling are not covered by our warranty.
The manufacturer is not liable for damage to the instrument due to incorrect operation, negligence or any
other manipulation of the instrument, in particular due to the removal or replacement of instrument parts or
due to the use of accessories from other manufacturers. This will immediately void the warranty.
No maintenance or repair work, except for the instances mentioned in the manual, may be performed on the
microscopes. Only ZEISS service personnel or personnel authorized especially by ZEISS may perform repair
work on the microscopes. Should your instrument malfunction, please contact the ZEISS Microscopy Service
Department (see page 126) or the ZEISS agency assigned to your country.
04/2020 430037-7444-001 14
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Intended use ZEISS
Axiolab 5 microscopes were designed as all-purpose microscopes for biological and medical applications as
well as materials analyses.
Depending on the microscope stand selected, they may also be used as true transmitted or reflected light
microscopes or as combined transmitted/reflected light fluorescence microscopes.
Typical biomedical applications of Axiolab 5 microscopes include:
− medical analysis in laboratories, clinics and medical practices
− science and research (colleges, universities) in the fields of medicine and biology
− industrial applications (pharmacology, food technology)
− analysis of human blood and tissue samples.
Axiolab 5 microscopes are typically used for materials analysis in the following areas, among others:
− metallographic laboratories
− automotive industry
− microsystems engineering
− geoscientific institutes
− mineral exploration industry
Depending on the instrument configuration, the following microscopy and contrasting techniques can be used:
Transmitted light Reflected light
− Brightfield (BF) − Brightfield (BF)
− Darkfield (DF) − Darkfield (DR)
− Phase contrast (Ph) − Polarization (Pol)
− Polarization (Pol) − Fluorescence (FL)
− Polarization (Conoscopy) − Differential Interference Contrast (DIC)
− Polarization (C-Pol) − Circular differential interference contrast/total
interference contrast
(C-DIC/TIC)
The binocular photo tubes and suitable adapters permit one microscope camera, one reflex camera or one
digital/video camera to be attached for documentation purposes.
The Axiolab 5 was specifically designed and developed for ergonomic use in lengthy routine applications, e.g.
hematological, histological and cytological laboratory analyses.
Using a ZEISS Axiocam 202 mono or Axiocam 208 color camera, the microscope can be connected to an
external HD monitor via an HDMI connection or to a PC/laptop via a USB connection to control camera
functionality. Advanced analysis functions for recorded images are available if the ZEN software is installed on
04/2020 430037-7444-001 15
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Intended use ZEISS
the PC. Furthermore, the camera functions can be controlled via WiFi connection from a tablet, PC or
smartphone using the Labscope/Matscope app.
04/2020 430037-7444-001 16
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Technical data ZEISS
The other stand models differ slightly in depth and significantly in height, depending on the tube used. An
overview of viewing heights (eyepoint heights) of the various tubes can be found on page 19.
An estimate of the height of the stand with the respective tube can be obtained by adding:
− 10 mm to the viewing height in the lower position of the binocular section for tubes with a fixed viewing
angle
− 10 mm to the viewing height of the upper limit for ergo tubes
Weight
Axiolab 5 microscope stand (depending on version and accessories) ......................................... approx. 8 to 20 kg
Ambient conditions
Operational specifications
Operational area ................................................................................................................................. Closed rooms
Protective class ......................................................................................................................................................... I
Protection type .................................................................................................................................................. IP 20
Electrical safety ......................................................................... in accordance with DIN EN 61010-1 (IEC 61010-1)
.............................................................................................................. in compliance with CSA and UL regulations
Overvoltage category .............................................................................................................................................. II
RFI suppression ...................................................................................................... complies with EN 55011 Class B
Noise immunity ..................................................................................................... complies with DIN EN 61326/A1
Mains voltage for the Axiolab 5 ........................................................................................................... 100 to 240 V
Mains frequency ...................................................................................................................................... 50 to 60 Hz
Power consumption of the Axiolab 5 ............................................................................................................ 100 VA
04/2020 430037-7444-001 17
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Technical data ZEISS
Light sources
LED illumination with transmitted light/reflected light
Power consumption............................................................................................................................ max. 10 W
Adjustment of light source ............................................................................. continuous approx. 10 to 800 mA
Halogen lighting with transmitted light/reflected light
Power consumption............................................................................................................................ max. 35 W
Adjustment of light source ................................................................................ continuous, approx. 0.5 to 12 V
LED lighting with reflected light fluorescence with replaceable LED modules
Wavelengths optional...................................................................................................... 385, 470, 565, 625 nm
LED classification....................................................................................................... LED Risk Group 2 to IEC 62471
Axiolab 5
Stand with manual stage focusing
Coarse focusing............................................................................................................ approx. 4 mm/revolution
Fine focusing ............................................................................ approx. 0.4 mm/revolution; 4 µm scale interval
Lifting range ......................................................................................... depending on the stand, 15 mm/30 mm
Height stop ................................................................................................................................... factory pre-set
Condenser 0.9/1.25 BF with optional
modulator disk ........................................................................................................ for brightfield, darkfield and
........................................................................................................................................... phase contrast 1, 2, 3
Manual objective change.................................... depending on the nosepiece, 5x BF Pol or 5x BF DF, M27, coded
Manual reflector module change .................................................................. for 4-position reflector turret, coded
04/2020 430037-7444-001 18
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Technical data ZEISS
425520-9010-000 Binocular photo tube 30°/23 (50:50) 30° - None - 449 / 485
425520-9020-000 Binocular photo tube 30°/23 (100:100) Bio 30° - None - 449 / 485
425520-9030-000 Binocular photo tube 20°/23 (100:100) 20° - None - 442 / 481
425520-9050-000 Binocular ergo tube 15°/23 (50/50), 15° Height, 410 - 509
telescopic, height, upright image telescopic
425520-9100-000 Binocular photo tube 20°/23 Pol (100:100) 20° 442 / 481
425512-0000-000 Binocular ergo tube 20°/23 (100/100), 20° Height 457 - 574
reverse image, 44 mm height
* Viewing heights:
Tubes with fixed viewing angle without ergo function:
Binocular part, lower/upper e.g. 442 / 481 → 442 to 481 mm
04/2020 430037-7444-001 19
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Interface diagram ZEISS
The following figure shows a diagram of the interfaces of the microscope stand. The Axiolab 5 Mat-TL/RL stand
is used here as an example.
04/2020 430037-7444-001 20
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on the microscope ZEISS
04/2020 430037-7444-001 21
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on the microscope ZEISS
Transmitted light stand for bioscience (exemplarily: 430037-9011-000) equipped with the following main
components:
1. LED 10W-TL illuminator, optional with halogen reflector lamp 12 V 35 W
2. Nosepiece with 5 positions BF, coded
3. Mechanical stage, 75x50 R
4. Binocular photo tube, 30x/23 (50:50), reversed image
5. Eyepiece E-PL 10x/22, GW, focusable
6. Condenser 0.9/1.25H
04/2020 430037-7444-001 22
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on the microscope ZEISS
04/2020 430037-7444-001 23
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on the microscope ZEISS
Transmitted light and reflected light fluorescence stand for bioscience (exemplarily: 430037-9021-000)
equipped with the following main components:
1. LED 10W-TL illumination, optional with halogen reflector lamp 12 V 35 W
2. FL-LED reflected light illumination with 3-position mount for LED
3. Nosepiece with 5 positions BF, coded
4. Mechanical stage, 75x50 R
5. 4-position reflector turret for P&C modules, coded
6. Binocular photo tube, 30x/23 (100:0/0:100), reversed image
7. Eyepiece E-PL 10x/22, GW, focusable
8. Condenser 0.9/1.25 BF
04/2020 430037-7444-001 24
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on the microscope ZEISS
04/2020 430037-7444-001 25
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on the microscope ZEISS
Transmitted light stand for polarization (430037-9130-000) with the following main components:
1. LED 10W-TL illumination, optional with halogen reflector lamp 12 V 35 W
2. Nosepiece with 5 positions BF Pol (with 4x centerable, 1x fixed), coded
3. Rotary stage Pol, 360° with clamping device click stop per 45°
4. Polarizer D, fixed, removable
5. Binocular photo tube 30x/23 (100:0/0:100), reversed image
6. Eyepiece E-PL 10x/22, GW, focusable, Pol
7. Condenser 0.9 BF Pol
04/2020 430037-7444-001 26
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on the microscope ZEISS
04/2020 430037-7444-001 27
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on the microscope ZEISS
Transmitted light stand for polarization/conoscopy (430037-9042-000) with the following main components:
1. LED 10W-TL illumination, optional with halogen reflector lamp 12 V 35 W
2. Nosepiece with 5 positions BF Pol (with 4x centerable, 1x fixed), coded
3. Rotary stage Pol, 360° with clamping device click stop per 45°
4. Polarizer D, fixed, removable
5. Binocular photo tube 30x/23 (100:0/0:100), reversed image
6. Eyepiece E-PL 10x/22 GW, focusable, Pol
7. Condenser 0.9 BF Pol
04/2020 430037-7444-001 28
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on the microscope ZEISS
ATTENTION
The movements of rotary knobs A and BL (Fig. 2-5/9 and 10) and the respective setting wheels
(Fig. 2-5/33 and 32) are coupled with each other. This means that only one control element should
be operated at a time and the movement of the other should not be inhibited or blocked.
Otherwise, mechanical damage may occur.
04/2020 430037-7444-001 29
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on the microscope ZEISS
Transmitted light and reflected light stand for polarization (430037-9032-000) with the following main
components:
1. LED 10W-TL illumination, optional with halogen reflector lamp 12 V 35 W
2. LED 10W-RL illumination
3. Nosepiece with 5 positions BF DF Pol (with 4x centerable, 1x fixed), coded
4. Rotary stage Pol, 360° with clamping device click stop per 45°
5. Polarizer D, fixed, removable
6. Polarizer slider A, 6x30mm, 90° rotatable
7. 4-position reflector turret for P&C modules, coded
8. Binocular photo tube 30°/23 (50/50), reversed image
9. Eyepiece E-PL 10x/22 GW, focusable, Pol
10. Condenser 0.9 BF Pol
04/2020 430037-7444-001 30
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on the microscope ZEISS
04/2020 430037-7444-001 31
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on the microscope ZEISS
Transmitted light and reflected light materials stand (430037-9052-000) with the following main components:
1. LED 10W-TL illumination, optional with halogen reflector lamp 12 V 35 W
2. LED 10W-RL illumination
3. Nosepiece with 5 positions BF DF, coded
4. Mechanical stage, 75x30 R
5. 4-position reflector turret for P&C modules, coded
6. Binocular photo tube 20°/23 (100:0/0:100), upright image with sliding prism
7. Eyepiece E-PL 10x/22 GW, focusable
8. Condenser 0.9/1.25 BF
04/2020 430037-7444-001 32
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on the microscope ZEISS
04/2020 430037-7444-001 33
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on the microscope ZEISS
For the location of the keys and display elements at your stand, refer to section 2.4.2 to 2.4.7.
Left Snap button or Snaps an image; after snap is finished, the monitor display appears
Short press *:
Right Snap button in BLACK for 50 ms.
(see Fig. 2-3/17) Starts video recording; another short press is required to stop
(only if Axiocam 202 or Long press **: recording. After recording is finished, the monitor display appears in
208 is attached) BLACK for 300ms.
Enables/disables Light Manager (LM):
- Disabling: The indicator light blinks
Snap button +
Long press **: GREEN / ORANGE / GREEN in sequence.
Intensity/LM knob
- Enabling: The indicator light blinks
GREEN / GREEN / GREEN in sequence.
04/2020 430037-7444-001 34
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on microscope components ZEISS
NOTE
For polarization microscopy, we recommend using the Pol photo tube with upright image and one
eyepiece reticle (graticule).
04/2020 430037-7444-001 35
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on microscope components ZEISS
04/2020 430037-7444-001 36
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on microscope components ZEISS
− The adjustment range can be read off the vertical scale (Fig. 2-11/4).
70H
The ergo photo tube has two switch settings (Graduation: 100:0/0:100).
− Push-pull rod (Fig. 2-11/2) pushed in:
708H 100% to the eyepieces
− Push-pull rod (Fig. 2-11/2) pulled out:
709H 100% to the camera port
04/2020 430037-7444-001 37
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on microscope components ZEISS
Fig. 2-12 Rackless stage 75x30 R with specimen − Vernier scale for display of the adjustment
holder range in X (Fig. 2-12/4) and Y direction
(Fig. 2-12/5).
04/2020 430037-7444-001 38
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on microscope components ZEISS
04/2020 430037-7444-001 39
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on microscope components ZEISS
2.5.4 Condensers
Condenser 0.9/1.25 BF
Condenser 0.9/1.25 BF (Fig. 2-18/1) with aperture
diaphragm (Fig. 2-18/2) for brightfield microscopy.
This condenser is also available with a modulator
disk.
04/2020 430037-7444-001 40
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on microscope components ZEISS
− Filter slider for reflected light with two positions for filters with a diameter of 25 mm (neutral and color
filters, white balance filter)
− Insert the filter slider from the left and operate (Fig. 2-7/5)
ATTENTION
If the condenser carrier with mounted low-power system is moved too far down, the low-power
system may collide with the field diaphragm and damage it!
04/2020 430037-7444-001 41
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on microscope components ZEISS
2.5.8 Polarizer
ATTENTION
If the condenser carrier with mounted polarizer is moved too far down, the polarizer may collide
with the field diaphragm and damage it!
ATTENTION
04/2020 430037-7444-001 42
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on microscope components ZEISS
The microscope with Axiocam 208/202 can be used in standalone mode. The camera acts as the control
interface and is powered by microscope via the USB (Commercial Micro-D power) cable. A USB Type-C drive is
included in the package and can be connected via the USB slot at the back of the camera for storing data. Then
images are recorded and saved to the USB drive. Functions of the microscope stand such as the Light Manager
and encoding are automatically launched. The camera is equipped with image enhancement functions such as
true color and noise reduction.
NOTE
Please note that the focus plane of the camera must be adjusted to the focus plane of the
eyepieces via the camera adapter.
Functionality:
− Light Manager
− Coded components
− Image enhancement
(true color, noise
reduction)
− Snap and save image on
USB drive
− Record and save movie
on USB drive
NOTE
− Currently the snap button on the stand functions only with the Axiocam 208/202. It cannot be
used with any other cameras.
− The snap button on the stand works only when a USB drive is detected by the camera port.
04/2020 430037-7444-001 43
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on microscope components ZEISS
A monitor can be connected to the camera via an HDMI cable. The camera is powered by the microscope via
the USB (Commercial Micro-D power) cable. A USB hub can be connected via the USB port on the camera. A
wireless or wired mouse and keyboard can be connected to the camera via the USB hub, which together with
the monitor, act as the control interface. Functions such as the Light Manager, encoding and image
enhancement are automatically deployed. Live images can be viewed on the monitor display and advanced
features are available in OSD (on-screen display). With Axiolab 5 TL/FL, One-key fluorescence function can be
used. Images can be snapped and saved into the USB Type-C drive, which is connected via the USB hub.
Functionality :
− Light Manager
− Coded components
− Image enhancement
− Observe live image on
display
− Snap and save image on
the USB drive
− Record and save movie
on the USB drive
− One-key fluorescence*
− Advanced features in
OSD
*: This works only with
Axiolab TL/FL
04/2020 430037-7444-001 44
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on microscope components ZEISS
2.5.10.3 Using the microscope imaging system with Labscope/Matscope via a Wi-Fi dongle connection
The camera is powered by the microscope via a USB (Commercial Micro-D power) cable. An optional monitor
can be connected to the camera via an HDMI cable. The recommended USB Wi-Fi dongle can be connected to
the camera via the USB hub. The control interface can be a PC or portable electronic device that uses Wi-Fi.
Functions such as the Light Manager, encoding, ECO mode and image enhancement are automatically
launched. When a monitor is connected, live images can be viewed on the monitor display. Live images can
also be viewed on PC or portable devices and advanced features in Labscope/Matscope are available. With
Axiolab 5 TL/FL, the one-key fluorescence function can be used.
Functionality:
− Light Manager
− Coded components
− ECO mode
− Image
enhancement
− Observe live image
− Snap and save
image via software
− One-key
fluorescence*
− Advanced features
in Labscope/
Matscope
*: Only with specific
microscopy configuration
1 USB Wi-Fi dongle (please see the ZEISS website for the recommended
model)
04/2020 430037-7444-001 45
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on microscope components ZEISS
2.5.10.4 Using the microscope imaging system with Labscope/Matscope via a WLAN router connection
The camera is powered by the microscope via a USB (Commercial Micro-D power) cable. An optional monitor
can be connected to the camera via an HDMI cable. A router is connected to the camera via Ethernet. The
control interface can be a PC or portable electronic device controlled via Ethernet or Wi-Fi. Functions such as
the Light Manager, encoding, ECO mode and image enhancement are automatically launched. When a monitor
is connected, live images can be viewed on the monitor display. Live images can also be viewed on a PC or a
portable device and advanced features in Labscope/Matscope are available. With Axiolab 5 TL/FL, the one-key
fluorescence function can be used.
Functionality:
− Light Manager
− Coded components
− ECO mode
− Image enhancement
− Observe live image
− Snap and save
image via software
− One-key
fluorescence*
− Advanced features
in Labscope/
Matscope
*: Only with specific
microscopy configuration
04/2020 430037-7444-001 46
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on microscope components ZEISS
2.5.10.5 Using the microscope imaging system with Labscope/Matscope via a USB connection
The camera is powered by the microscope via a USB (Commercial Micro-D power) cable. An optional monitor
can be connected to the camera via an HDMI cable. A PC or Windows Surface can be connected to the camera
via a USB cable. Functions such as the Light Manager, encoding, ECO mode and image enhancement are
automatically launched. When a monitor is connected, live images can be viewed on the monitor display. Live
images can also be viewed on a PC or Surface and advanced features in Labscope/Matscope are available.
With Axiolab TL/FL, the one-key fluorescence function can be used.
Functionality:
− Light Manager
− Coded components
− ECO mode
− Image enhancement
− Observe live image
− Snap and save
image via software
− One-key
fluorescence*
− Advanced features
in Labscope/
Matscope
*: Only with specific
microscopy configuration
04/2020 430037-7444-001 47
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on microscope components ZEISS
2.5.10.6 Using the microscope imaging system with ZEN software via a USB connection
The camera is powered via a USB (Commercial Micro-D power) cable connected to an external power socket. A
workstation can be connected to the camera and the microscope stand via USB cables at the same time.
Functions such as the Light Manager, encoding and ECO mode are automatically launched. Live images can
also be viewed on the workstation and basic features in ZEN are available.
Functionalitiy:
− Light Manager
− Coded components
− ECO mode
− Image enhancement
− Observe live image
− Snap and save
image via software
− Basic features in ZEN
04/2020 430037-7444-001 48
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on microscope components ZEISS
4 status LED
NOTE
04/2020 430037-7444-001 49
DESCRIPTION OF THE INSTRUMENT
Axiolab 5 Control and functional elements on microscope components ZEISS
NOTE
04/2020 430037-7444-001 50
START-UP
Axiolab 5 Mounting standard components ZEISS
3 START-UP
The Axiolab 5 microscopes can be independently installed, converted and started up by the customer.
On request, the microscope can also be installed or converted by ZEISS Service for an extra charge.
NOTE
Before installing and starting up the microscope, read the Notes on instrument safety (see section
1.1) carefully and thoroughly.
The assembly activities described in the following section are illustrated using examples for one microscope
stand type. However, they apply similarly to other microscope stand models. Special features are described
separately.
04/2020 430037-7444-001 51
START-UP
Axiolab 5 Mounting standard components ZEISS
04/2020 430037-7444-001 52
START-UP
Axiolab 5 Mounting standard components ZEISS
For tubes mounted with an intermediate plate on the Axiolab 5 stand, Pol-TL/conoscopy (430037-90042-000),
proceed as follows:
• Loosen the screw (Fig. 3-3/6) with an Allen wrench (AF 3). Remove dust caps (Fig. 3-3/8, 11) from the
underside of the tube and the dovetail ring mount on the stand side.
• Insert the intermediate plate (Fig. 3-3/10) with its dovetail ring into the stand mount (Fig. 3-3/7) and
tighten Allen screw (Fig. 3-3/6).
• Insert the binocular tube/photo tube (Fig. 3-3/12) into the intermediate plate, align and tighten the screw
(Fig. 3-3/9) with the Allen wrench.
04/2020 430037-7444-001 53
START-UP
Axiolab 5 Mounting standard components ZEISS
NOTE
• Instead of an eyepiece you may insert an auxiliary microscope (Fig. 3-4/3) into one of the binocular
301H
eyepiece tubes in order to observe aperture, phase and darkfield diaphragms and to center phase and
darkfield diaphragms. These diaphragms can be focused with the adjustable eye lens of the auxiliary
microscope.
• The auxiliary microscope (Fig. 3-4/3) or pinhole diaphragm (Fig. 3-4/4) can be used to observe conoscopic
images.
04/2020 430037-7444-001 54
START-UP
Axiolab 5 Mounting standard components ZEISS
The slight image shift caused by the additional path through glass is taken into account on the diopter scale by
the fact that the zero point position is indicated not by the white dot (Fig. 3-5/W), but the red dot (Fig. 3-5/R).
NOTE
The eyepiece reticles must be inserted under dust-free conditions. This should be carried out only
by ZEISS Service.
NOTE
Always replace the dust protection caps on any empty positions on the nosepiece.
NOTE
04/2020 430037-7444-001 55
START-UP
Axiolab 5 Mounting standard components ZEISS
3.1.5 Inserting and removing P&C reflector modules in/from the reflector turret
The reflector turret with four positions is firmly installed in the reflected light illumination module for
fluorescence or materials stands.
The modules must be inserted and removed from the front after removing the cover cap.
Inserting a module:
• Remove the cover cap (Fig. 3-7/4) from the stand
towards the front.
• Insert the module (Fig. 3-7/2) as illustrated
together with the retaining brackets on the right
and left (Fig. 3-7/3) diagonally from below into the
upper spring clips (Fig. 3-7/1) of the reflector
turret.
• Then apply pressure to the module from below
until it also securely engages with the lower spring
clips of the reflector turret.
The position number of the P&C module is shown
on the right hand side of the reflector turret
adjacent to the position of the respective P&C
module.
Fig. 3-7 Replacing the reflector module • Apply the supplied stickers with the filter
combination data of the respective module to the
corresponding field of the cover cap (Fig. 3-7/5,
positions 1 to 4).
Removing a module:
• Slightly tilt the module in order to detach it from
the lower spring clips, then from the upper spring
clips of the reflector turret.
• Once the reflector modules have been installed /
removed, refit the cover cap. The cover cap
should be fitted onto the stand as straight as
possible to avoid the knurled ring of the reflector
turret becoming jammed and damaged.
• Apply pressure to the cover cap until the retaining
brackets have engaged.
04/2020 430037-7444-001 56
START-UP
Axiolab 5 Mounting standard components ZEISS
• Place the stage (Fig. 3-8/3) on the stage carrier (Fig. 3-8/4) so that the threaded holes on the bottom of the
stage (Fig. 3-8/8) are positioned above the stage carrier openings (Fig. 3-8/7).
• Insert four fastening screws (Fig. 3-8/6) through the stage carrier from below and screw them into the
bottom of the stage.
• Turn the stage to orient it in an XY direction and tighten the fastening screws.
• Place the specimen holder (Fig. 3-8/2) on the stage and fasten the two clamping screws (Fig. 3-8/1).
The length of the X and Y drives can be extended by shifting the respective coaxial knurled knob (Fig. 3-9/4 or
1) axially within a range of approx. 15 mm.
04/2020 430037-7444-001 57
START-UP
Axiolab 5 Mounting standard components ZEISS
(1) X drive
• Push the coaxial knurled knob for the X
adjustment (Fig. 3-9/4) all the way to the bottom.
• Remove the supplied adjusting pin (Fig. 3-9/5)
from the coaxial knurled knob for the Y
adjustment (Fig. 3-9/1) and insert it into one of
the holes of the lower hole nut(Fig. 3-9/3).
• Hold the coaxial knurled knob for the X
adjustment (Fig. 3-9/4) and turn the hole nut with
the adjusting pin clockwise (small friction
adjustment: –) or counter-clockwise (large friction
adjustment: +) until the desired freedom of
Fig. 3-9 Friction adjustment movement has been achieved (see Fig. 3-9).
• It should not be shifted more than one revolution.
(2) Y drive
• Push the coaxial knurled knob for the Y adjustment (Fig. 3-9/1) all the way to the top.
• Insert the supplied adjusting pin (Fig. 3-9/5) into the hole of the upper hole nut (Fig. 3-9/2).
• Hold the coaxial knurled knob for the Y adjustment (Fig. 3-9/1) and turn the hole nut with the adjusting pin
clockwise (small friction adjustment: –) or counter-clockwise (large friction adjustment: +) until the desired
freedom of movement has been achieved.
• It should not be shifted more than one revolution.
• Re-insert the adjusting pin into the coaxial knurled knob for the Y adjustment (Fig. 3-9/1).
NOTE
Set the friction adjustment on the mechanical stage with the ergonomic, stationary XY drive
analogously. No tool is required for the purpose. The lock-nut (silver) of the respective drive can be
adjusted manually; hold the coaxial knurled knob tight.
04/2020 430037-7444-001 58
START-UP
Axiolab 5 Mounting standard components ZEISS
NOTE
The rotary stage must be mounted so that the vernier scale is on the left side and the clamp is on
the right side.
3.1.7.3 Dismantling the detachable specimen guide and mounting stage clips
• Loosen the two clamp screws (Fig. 3-10/1) on the Pol specimen guide with an Allen wrench (AF 2). Remove
the Pol specimen guide (Fig. 3-10/2) by lifting it upwards.
• Insert the stage clips (Fig. 3-10/9) into the holes provided.
04/2020 430037-7444-001 59
START-UP
Axiolab 5 Mounting standard components ZEISS
3.1.7.4 Removing the stage clips and mounting the detachable Pol specimen guide
• Remove the stage clips (Fig. 3-10/9) from the Pol rotary stage.
• Insert the Pol specimen guide (Fig. 3-10/2) with the two cylindrical pins on the underside into the holes
provided (Fig. 3-10/3) and tighten the two clamp screws (Fig. 3-10/1) with an Allen wrench (AF 2).
Fig. 3-11 Centering the Pol rotary stage • For centering the stage, use a contrasting
specimen and an eyepiece with a crossline reticle.
• Loosen the stage clamping screw (Fig. 3-11/1) and
the screw cap on the stage carrier (Fig. 3-11/3).
• Rotate the stage to determine the position of maximum offset of the specimen feature (Fig. 3-11/5, origin
of arrow) from the center of the eyepiece reticle.
• Reset the two centering screws on the stage carrier (Fig. 3-11/2) using an Allen wrench (AF 1.5) (Fig. 3-11/4)
to move the specimen detail by half an arrow length in the direction of the crossline center. Check whether
the specimen detail moves when the stage is rotated again; repeat the procedure, if required.
NOTE
The Allen wrenches (AF 1.5) are located in the storage compartment at the back of the microscope
stand.
04/2020 430037-7444-001 60
START-UP
Axiolab 5 Mounting standard components ZEISS
NOTE
To maintain this centering accuracy, when replacing an objective, do not hold the objective itself;
hold the knurled ring of the nosepiece to rotate it.
04/2020 430037-7444-001 61
START-UP
Axiolab 5 Mounting standard components ZEISS
ATTENTION
NOTE
04/2020 430037-7444-001 62
START-UP
Axiolab 5 Mounting standard components ZEISS
NOTE
04/2020 430037-7444-001 63
START-UP
Axiolab 5 Mounting standard components ZEISS
04/2020 430037-7444-001 64
START-UP
Axiolab 5 Mounting standard components ZEISS
• Loosen the three side screws (Fig. 3-18/4a, 4b, 4c) with an Allen wrench (AF 2.5) (Fig. 3-18/3) and remove
the old LED illuminator (Fig. 3-18/2) from the adapter (Fig. 3-18/1).
• Insert the new LED illuminator into the adapter and tighten the three side screws on the adapter.
04/2020 430037-7444-001 65
START-UP
Axiolab 5 Mounting standard components ZEISS
• Insert the new LED illuminator with the adapter (Fig. 3-19/3) into the holding tube up to the stop.
• Position the new LED illuminator with the pinhole (Fig. 3-19/5) at the top, or with the adapter lower front
edge (Fig. 3-19/4) aligned between the contact surface and securing clips.
• Plug the illuminator plug (Fig. 3-19/2) into the connector of the stand.
• Lift the securing clips (Fig. 3-19/1) of the lamp holder until they are completely enclosed in the lamp holder.
In doing so, slightly press the ends of the securing clips together and guide past the two upper retaining
elements. Release the pressure until the clamping bracket opens and engages on both sides in the retaining
elements.
04/2020 430037-7444-001 66
START-UP
Axiolab 5 Mounting standard components ZEISS
04/2020 430037-7444-001 67
START-UP
Axiolab 5 Mounting standard components ZEISS
04/2020 430037-7444-001 68
START-UP
Axiolab 5 Mounting standard components ZEISS
• Loosen the three side screws (Fig. 3-24/4a, 4b, 4c) with an Allen wrench (AF 2.5) (Fig. 3-24/3) and remove
the old LED illuminator (Fig. 3-24/2) from the adapter (Fig. 3-24/1).
• Insert the new LED illuminator into the adapter and tighten the three side screws on the adapter.
• Insert the new LED illuminator with adapter (Fig. 3-23/4) into the holding tube (Fig. 3-23/1) up to the stop.
• Position the new LED illuminator with the pinhole (Fig. 3-23/3) at the left, or with the adapter front edge
(Fig. 3-23/2) aligned to the right.
• Plug the lamp plug (Fig. 3-23/5) into the connector of the stand (Fig. 3-23/6).
• Lift the securing clips (Fig. 3-22/1) of the lamp holder until they are completely enclosed in the lamp holder.
In doing so, slightly press the ends of the securing clips together and guide past the two upper retaining
elements. Release the pressure until the clamping bracket opens and engages on both sides in the retaining
elements.
04/2020 430037-7444-001 69
START-UP
Axiolab 5 Mounting standard components ZEISS
• Pull the old LED modules (Fig. 3-26/1.3) out of the holding positions of the LED tube (Fig. 3-26/1.1).
04/2020 430037-7444-001 70
START-UP
Axiolab 5 Mounting standard components ZEISS
• Insert the new LED modules into the LED tube up to the stop. Ensure that the modules are inserted in the
right holding positions (Fig. 3-26/).
NOTE
The LED positions , and in Fig. 3-26 correspond to the UV, B and G labels on the right side
of the microscope (see section 2.4.3, Fig. 2-3/4).
• Plug the module plugs into the connectors (Fig. 3-26/1.2) of the stand. Ensure that the plugs are connected
to the right connectors (Fig. 3-26/).
• Screw the module holder (Fig. 3-26/1) into the LED tube to fix the modules in place.
NOTE
Due to the limited space in the LED module socket, the LED module in position needs to be
moved out first when replacing the LED module in position .
04/2020 430037-7444-001 71
START-UP
Axiolab 5 Mounting standard components ZEISS
• Mount the C-mount camera adapter (Fig. 3-27/2) on the Axiocam (Fig. 3-27/1).
• Attach the Axiocam with the adapter to the camera port (Fig. 3-27/4) of the tube.
• Orient the camera to the stand and fix it in position by tightening the ring nut (Fig. 3-27/3).
• Connect the camera to the stand (Fig. 3-27/7) via the USB (Commercial Micro-D) cable (Fig. 3-27/5).
• Connect the camera to an external monitor via an HDMI cable (Fig. 3-27/6).
• Alternatively, connect the camera to a WLAN router or PC.
NOTE
Fig. 3-27 Mounting the Axiocam 202 mono or Axiocam 208 color
04/2020 430037-7444-001 72
START-UP
Axiolab 5 Mounting optional components ZEISS
CAUTION
The microscope must be switched off and unplugged before starting work.
Upon completion of the work, the function of the respective modules must be restored (see
sections 3.1 to 3.4).
The light-intensive co-observer unit is mounted on the Axiolab 5 with a main observer and one or two co-
observers in accordance with the separate instructions for use for multi-conference facilities (order no.
425145-7144-001).
The other components shown in the system overview Fig. 3-28 Mounting the polarizer D
at this point must be mounted in a similar manner.
04/2020 430037-7444-001 73
START-UP
Axiolab 5 Mounting optional components ZEISS
the stop.
Fig. 3-29 Mounting the low-power system • Swing the low-power system into the beam path.
Make sure that the fixture is securely engaged.
• Open the aperture diaphragm and field diaphragm
completely.
• Adjust both adjustment screws (Fig. 3-29/5) using two Allen wrenches (AF 1.5) until the field of vision is
57H
optimally illuminated.
NOTE
The low-power system can be used only in combination with the condenser 0.9/1.25.
04/2020 430037-7444-001 74
START-UP
Axiolab 5 Connecting to the power supply ZEISS
04/2020 430037-7444-001 75
START-UP
Axiolab 5 Switching the microscope on/off ZEISS
04/2020 430037-7444-001 76
START-UP
Axiolab 5 Using the Light Manager function ZEISS
The Light Manager (LM) function saves the ratios of the set light intensities between different combinations of
objective and reflector turret positions for a given light source. When changing the light intensity of one
objective/reflector combination, the light intensities of other combinations will also change according to the
set ratios. This ensures that users don't need to repeatedly set up light intensities for each objective/reflector
combination when switching between samples which require different illumination intensity.
After switching on the microscope, the previous setting of the Light Manager will be restored.
04/2020 430037-7444-001 77
START-UP
Axiolab 5 Default factory settings of the microscope ZEISS
ATTENTION
Please use this function with caution as it will reset all existing configurations.
04/2020 430037-7444-001 78
OPERATION
Axiolab 5 Default setting of the microscope ZEISS
4 OPERATION
4.1 Default setting of the microscope
04/2020 430037-7444-001 79
OPERATION
Axiolab 5 Default setting of the microscope ZEISS
4.1.3 Adjusting for ametropia (user's visual impairment) when using eyepiece reticles
The prerequisite for correct use of an eyepiece reticle is two adjustable eyepieces to compensate for different
degrees of ametropia of the user.
• Focus on the line figure of the eyepiece reticle with the focusable eye lens of the adjustable eyepiece.
• Focus on the microscopic image of a loaded specimen with the focusing drive while observing with the
eyepiece containing the eyepiece reticle.
• When both the microscopic image and the eyepiece reticle are in focus, the image for the second eye is
brought into focus with the focusable eye lens of the second eyepiece.
Both microscopic images including the eyepiece reticle are thus in focus. From this point, use only the focusing
drive for any subsequent focusing activity.
04/2020 430037-7444-001 80
OPERATION
Axiolab 5 Illumination and contrast methods in transmitted light ZEISS
4.2.1 Configuring transmitted light brightfield microscopy using the KÖHLER method
(3) Configuring transmitted light brightfield microscopy using the KÖHLER method
− The Axiolab 5 has been started up correctly (see section 3).
− The Axiolab 5 is switched on.
• Adjust the image brightness using the Intensity/LM knob (Fig. 4-3/1) on the microscope stand.
• Insert a high-contrast specimen into the specimen holder of the mechanical stage.
• If condensers with a swiveling front lens are used, swivel these into the beam path with 10x objectives
and set the condenser with the knurled knob for vertical adjustment (Fig. 4-3/3 or Fig. 4-4/2) to the upper
stop. The stop must be adjusted so that the specimen is not lifted by the condenser (to set the condenser
stop, see section 4.2.1 (4)).
• When using condensers with a turret/modulator disk and knurled ring (Fig. 4-4/3), set position B (or H =
brightfield).
• Swivel in the 10x objective on the nosepiece (Fig. 4-3/6) and use the focus drive to focus on the specimen
(Fig. 4-3/2).
04/2020 430037-7444-001 81
OPERATION
Axiolab 5 Illumination and contrast methods in transmitted light ZEISS
NOTE
Every change of objective will result in a change in specimen field size and objective aperture,
together with a possible slight change in centering, so that for optimal results the luminous-field
and aperture diaphragm adjustments must be repeated.
With objectives < 10x, the front lens of the condenser (if swivelable) must be swiveled out of the
beam path and the aperture diaphragm completely opened. For better contrast with such large
object fields, the luminous-field aperture should be closed to a certain extent. Overclosing should
be avoided so as not to impair the uniformity of the illumination of the field of view.
04/2020 430037-7444-001 82
OPERATION
Axiolab 5 Illumination and contrast methods in transmitted light ZEISS
04/2020 430037-7444-001 83
OPERATION
Axiolab 5 Illumination and contrast methods in transmitted light ZEISS
4.2.2 Configuring transmitted light darkfield microscopy using the KÖHLER method
In darkfield microscopy, only the diffracted and scattered light components which are important for imaging
reach the objective, while the direct unchanged light bundles are routed past the objective. This is one of the
reasons why even fine structures that are sometimes below the resolving power of the light microscope can be
resolved and appear very bright on a dark background.
(2) Instrumentation
All Axiolab 5 microscopes, except stands for reflected light, are suitable for darkfield applications.
Condenser with darkfield stop in position D e.g.:
− Condenser 0.9/1.25 H with modulator disk BF, DF, Ph 1, Ph 2, Ph 3
− Condenser, achrom.-aplan. 0.9 H D Ph DIC
− Darkfield condenser with dry darkfield (465505-0000-000 applicable aperture from 0.6 – 0.75)
− Ultra condenser (465500-0000-000 applicable aperture from 0.75 – 1)
• To center the darkfield stop (not possible with all condensers), use two Allen wrenches (AF 1.5) (Fig. 4-5/1)
to turn the two centering screws (Fig. 4-5/2 and 3) until the exit pupil of the objective appears
homogeneously dark. After centering, remove both Allen wrenches (AF 1.5) from the condenser.
04/2020 430037-7444-001 84
OPERATION
Axiolab 5 Illumination and contrast methods in transmitted light ZEISS
NOTE
Since the apertures of objectives with an integrated aperture iris stop are too high for transmitted
light darkfield microscopy, the aperture iris stop must at least be closed to the limit aperture of
0.65.
The performance criterion for the darkfield method is always the darkest possible background of the field of
view.
• Reinsert the eyepiece into the tube.
• If the height of the darkfield condenser is set correctly and sensitively, it is possible to reduce any
brightening in the field of view left, and the luminous-field diaphragm image appears almost perfectly in
focus.
• Finally, match the size of the luminous-field diaphragm to the size of the field of view.
Darkfield microscopy requires specimens to be considerably cleaner than in other techniques. In particular,
fingerprints, dirt or dust particles have a negative effect, as they brighten the background of the field of view
and decrease the contrast of the object image.
04/2020 430037-7444-001 85
OPERATION
Axiolab 5 Illumination and contrast methods in transmitted light ZEISS
(5) Setting darkfield contrast with the immersion oil darkfield condenser
• If necessary, swivel out the low-power system, polarizer or λ plate.
• Move the condenser carrier down until the end stop.
• Place the immersion oil darkfield condenser in the condenser holder Z (see section 3.1.9).
• Place the condenser holder Z in the condenser carrier and center it roughly, so the condenser fits into the
opening of the mechanical stage without contact when moving upwards.
• Move the condenser up to the end stop.
• Place a drop of immersion oil (without bubbles, if possible) on the center of the condenser.
• Position a specimen. The immersion oil will disperse between the condenser and specimen holder.
• Move the mechanical stage back and forth slightly to dissipate any air bubbles in the immersion oil.
• Adjust the illumination intensity so that it is bright enough and open the field diaphragm completely.
• Swivel in a low-magnification objective (e.g. 10x) and focus the specimen, using the focusing drive.
• Center the field diaphragm on the condenser carrier with the adjustment screws and focus the image with
the condenser drive.
• Place a drop of immersion oil on the specimen holder, swivel in the immersion oil objective and focus the
specimen.
• Close the field diaphragm down to the end stop.
• Lower the condenser until the edge of the field diaphragm appears sharp.
• Center the field diaphragm image with both centering screws on the condenser carrier.
The luminous field diaphragm appears only as a circle segment on the edge of the viewing field due to the high
magnification of the immersion oil objective. As a result, the focusing and centering of the field diaphragm
must be repeated. If necessary, the luminous-field objective should be opened slightly if the light intensity is
too low.
The field diaphragm is centered properly when the edge of the luminous field diaphragm is centered or
equidistant from the edges of the viewing field.
• For a sharply focused specimen, open the sharply set field diaphragm enough to make the edge of the
diaphragm disappear from the field of view.
• You can improve the contrast of the microscope image by slightly adjusting the focus level of the condenser
with the condenser drive.
• The background of the eyepiece image should appear evenly dark.
• For immersion oil objectives with an iris diaphragm, the contrast can be further optimized by turning the
adjustment wheel of the iris diaphragm.
04/2020 430037-7444-001 86
OPERATION
Axiolab 5 Illumination and contrast methods in transmitted light ZEISS
(2) Instrumentation
All Axiolab 5 microscopes, except stands for reflected light, are suitable for phase contrast applications.
− Phase contrast objectives with phase rings Ph 1, Ph 2 or Ph 3 for different average numerical apertures
which can also be used in the brightfield.
− Condenser with turret/modulator disk containing centering phase stops Ph 1, Ph 2 and Ph 3 for different
average numerical apertures.
− The phase stop used on the condenser must correspond to the label on the objective used,
e.g. Ph 1.
04/2020 430037-7444-001 87
OPERATION
Axiolab 5 Illumination and contrast methods in transmitted light ZEISS
04/2020 430037-7444-001 88
OPERATION
Axiolab 5 Illumination and contrast methods in transmitted light ZEISS
(1) Application
The transmitted light polarization method is used for samples which change the polarization of the light. Such
samples are called birefringent. Examples include crystals, minerals or polymers. If such birefringent
substances are observed between crossed polarizers, the birefringent portion of the sample appears bright
while its surroundings remain dark.
A birefringent substance can be recognized by rotating the sample by 360° between crossed polarizers. The
sample should show four bright and four dark appearances during the rotation procedure. During the rotation
procedure, interference colors appear that range from gray (mostly for biological samples) through white,
yellow and red to blue, depending on the birefringence, thickness and orientation of the sample. The
interference colors may be of the first or a higher order.
(2) Instrumentation
Polarization methods can be used in the transmitted light on Axiolab 5 microscopes for transmitted light
polarization and conoscopy.
− Strain-free objectives
− Pol rotary stage
− Polarizer D (rotatable or fixed)
− Analyzer slider D, fixed, or lambda compensator or lambda/4 compensator
− Depolarizer (for screwing into Axiolab 5 tubes) to avoid undesirable polarization effects
NOTE
04/2020 430037-7444-001 89
OPERATION
Axiolab 5 Illumination and contrast methods in transmitted light ZEISS
04/2020 430037-7444-001 90
OPERATION
Axiolab 5 Illumination and contrast methods in transmitted light ZEISS
(1) Application
The determination of the polarization direction of n or n' respectively (polarization direction with the
absolute or relative largest index of refraction) and n or n' respectively (polarization direction with the
absolute or relative smallest index of refraction) relative to the morphological directions, e.g. of crystal
surfaces, crystal needles or fibers, provide an important signature of the material. This method is also used in
the diagnosis of bio-crystals (e.g. gout and pseudo-gout).
Fig. 4-8 Determining the polarization direction n' using a synthetic fiber as an example
(2) Instrumentation
− Eyepiece with crossline reticle
− Strain-free objectives
− Pol rotary stage (Fig. 4-7/1)
− Polarizer D (rotatable or fixed)
− Screw-in fixed analyzer slider D or lambda compensator or lambda/4 compensator combined with
analyzer (in Axiolab 5 tubes)
− Pol adjustment tool sample for polarization microscope (453679-0000-000)
04/2020 430037-7444-001 91
OPERATION
Axiolab 5 Illumination and contrast methods in transmitted light ZEISS
• Remove the analyzer from the beam path and align the reticle along the split cracks of the sample.
• Subsequently reinsert the analyzer and remove the Pol adjustment tool sample. The pass directions of the
polarizer and analyzer will now be parallel to the reticle (Polarizer EW, Analyzer NS).
NOTE
It is not necessary to adjust the reticle when working with the intermediate plate and the
binocular photo tube Pol (425520-9100-000).
• Rotate the Pol rotary stage with the sample, e.g. a synthetic fiber, until the sample appears as dark as
possible. In this position, the fiber extends parallel to one of the two directions of the crossline reticle.
NOTE
Do not change the inter-pupillary distance on the binocular tube, as the angle of the crossline
reticle to the fiber will be changed.
• Now rotate the stage by 45° so that the longitudinal axis of the fiber is oriented NE-SW (Fig. 4-9). The
sample will display the greatest brightness here (diagonal position). In this position the sample may have
any color.
• Insert the lambda compensator (possible only if used with screw-in analyzer in tube or intermediate plate).
Like the sample, the lambda compensator is a birefringent object, albeit with a defined path difference of
550 nm and the principal polarization direction n definitely oriented in a NE-SW direction.
When the lambda compensator is moved into the beam path, the sample changes its color. The type of color
change depends on the orientation of the sample (NE-SW or NW-SE).
The changes in color are attributable to optical interference. The interference colors (path differences) in both
diagonal positions (NE-SW and NW-SE) of the sample must be compared in this connection.
The path difference results from the superposition (interference) of the polarization direction of the sample
over the polarization direction of the lambda compensator.
The largest path difference occurs when the polarization direction of the sample with the absolutely or
relatively highest refractive index (n or n’) is parallel to the principal polarization direction of the lambda
compensator . The sample will then appear greenish-blue, for example (Fig. 4-8/2).
The smallest path difference occurs when the polarization direction of the sample with the absolutely or
relatively lowest refractive index (n or n’) is perpendicular to the polarization direction of the lambda
compensator . The sample will then appear yellow, for example (Fig. 4-8/3).
04/2020 430037-7444-001 92
OPERATION
Axiolab 5 Illumination and contrast methods in transmitted light ZEISS
(4) Conclusions
The gray-white color appearing first in the bright
position in the above example (Fig. 4-8/1)
corresponds to a path difference of 150 nm according
to the Michel-Lévy color chart (Fig. 4-9).
If the polarization direction of the birefringent sample to be examined is perpendicular to the principal
polarization direction of the lambda compensator, i.e. in the NW-SE direction, the path difference of the
sample (e.g. gray-white: 150 nm) is subtracted from the path difference of the compensator (red: 550 nm). In
this case, the interference color of the sample visibly changes from gray-white to orange (resulting path
difference = 400 nm).
NOTE
The Michel-Lévy color tables are available in the literature catalog no. 42-312.
04/2020 430037-7444-001 93
OPERATION
Axiolab 5 Illumination and contrast methods in transmitted light ZEISS
The measurement compensators are required for the exact measurement of path differences. These return,
i.e. compensate, the path difference created by the specimen to zero (black of the first order).
Whereas in the above-described methods the addition or subtraction position was of interest, only the
subtraction position is of interest in the measurement.
Path differences in the specimen can assume very small values (1/50 or 10 nm) and very large values (greater
than 10 or approx. 5500 nm and higher) and thus determine the compensator appropriate for the
measurement.
The suitable compensator is determined as follows:
• Configure the microscope as for transmitted light brightfield microscopy (see section 4.2.1), taking care to
ensure the correct inter-pupillary distance in the binocular tube (see section 4.1.1).
• Accurately position the specimen to be examined on the center of the reticle.
• Limit the aperture to a value of about 0.2.
• Turn the Pol rotary stage until the specimen is almost extinguished, i.e. completely dark, and set the 45°
locking position.
• Rotate the stage once (by 45°) so that the specimen is in a diagonal position (sample becomes bright).
NOTE
A prerequisite for the occurrence of the color change effect may be the evaluation in two
specimen positions rotated at an angle of 90° from one another, plus a centered stage.
NOTE
The DE SENARMONT compensation method requires the use of the rotatable analyzer.
− After insertion of the lambda compensator and rotation of the specimen by 90°, the interference color
remains white; in this case, however, it is a "higher-order white" and thus the path difference is > 5 .
The suitable compensator is:
K 0-30 tilting compensator (Accessory 000000-1115-698)
04/2020 430037-7444-001 94
OPERATION
Axiolab 5 Illumination and contrast methods in transmitted light ZEISS
− A dark gray appearing as the interference color indicates a very small path difference (/10 or 54.6 nm).
The accompanying instructions must be observed for the measurement preparation and procedure.
(1) Application
Unlike standard polarization contrast, circular polarization contrast does not show any dark (extinction)
positions that depend on the angle of rotation (azimuth) of the specimen relative to the polarizer or analyzer.
This means that when the stage is rotated the same image impression remains, as the light/dark positions are
omitted. With optical anisotropy, all transparent specimens display the characteristic interference colors.
(2) Instrumentation
− Strain-free objectives
− Pol rotary stage
− Circular polarizer D (no polarizers may be adapted on the condenser) including corresponding
lambda/4 plate.
− Stationary analyzer slider D or screw-in analyzer (in Axiolab 5 tubes).
04/2020 430037-7444-001 95
OPERATION
Axiolab 5 Illumination and contrast methods in transmitted light ZEISS
• Then swivel the upper part of the circular polarizer D (Fig. 4-10/4) into the beam path.
• Rotate the lever of the lambda/4 plate of the circular polarizer D (Fig. 4-10/3) until maximum extinction is
achieved (dark-gray field of view) (lever points 45° to the right).
• An (anisotropic) specimen should not be examined until after the above adjustment.
• Reinsert the specimen to be examined.
The interference colors – which depend on the material, specimen thickness and orientation – of the
specimens appear constant and independent of stage rotation.
04/2020 430037-7444-001 96
OPERATION
Axiolab 5 Illumination and contrast methods in transmitted light ZEISS
NOTE
For a high-contrast image with higher-magnification objectives (from approx. 20x) the illumination
aperture must be reduced to a value between 0.15 and 0.20, i.e. the aperture diaphragm must be
closed accordingly.
The effect of the lambda/4 plate (Fig. 4-10/4) can be undone by either swiveling it out of the beam
path or turning it with the lever (Fig. 4-10/3) into one of its two click-stop positions.
(5) Analysis
• If the crystal needles oriented parallel to the gamma direction of the lambda plate are yellow, and the
crystal needles lying at a right angle to the gamma direction are blue, the crystals are monosodium urate
crystals (gout).
• If the crystal needles oriented parallel to the gamma direction of the lambda plate are blue, and the crystal
needles lying at a right angle to the gamma direction are yellow, the crystals are calcium pyrophosphate
crystals (pseudo-gout).
This analysis is also possible using a polarizer with a fixed lambda plate which can be placed on the luminous-
field diaphragm. In that case, the lambda plate does not have to be rotated.
04/2020 430037-7444-001 97
OPERATION
Axiolab 5 Illumination and contrast methods in transmitted light ZEISS
For the classification (and thus identification) of crystalline material – instead of the observation of the
specimen itself – the analysis of an interference image in the objective pupil provides the more valuable
information. This image is visible in the eyepiece when an additional lens (so-called Bertrand lens) is switched
on. Alternatively, the auxiliary microscope or a diopter may be used to view the interference image.
In contrast to orthoscopy, this is referred to as conoscopy, because the illumination is ideally provided by a
wide open cone. In practice this means that the aperture diaphragm is fully open and the objective should
likewise have a high aperture.
(1) Application
Crystal analysis is used to determine the optical character of transparent and weakly absorbent crystals. This
method is also referred to as conoscopy.
Its main application is classic mineral microscopy. However, synthetic crystals, industrial minerals and plastics
(e.g. films) can also be identified and characterized.
(2) Instrumentation
Conoscopic viewing is preferably carried out on the Axiolab 5 microscope for transmitted light conoscopy.
− Strain-free objectives; recommended:
N-Achroplan 50x/0.8 Pol objective or
EC Plan-Neofluar 40x/0.9 Pol objective
− Pol rotary stage
− Polarizer D (rotatable or fixed)
− 0.9 Pol condenser
04/2020 430037-7444-001 98
OPERATION
Axiolab 5 Illumination and contrast methods in transmitted light ZEISS
• The polarization direction can be changed using the setting wheel (Fig. 4-11/3) of the analyzer.
ATTENTION
The movements of rotary knobs A and BL and the respective setting wheels are coupled with each
other. Only one control element should therefore be operated at a time and the movement of the
other should not be inhibited or blocked. Otherwise, mechanical damage may occur.
NOTE
If the rotary knob BL is set to the On position, rotary knob A is automatically moved into the On
position if it is not there already.
If, on the other hand, rotary knob A is set to the Off position, rotary knob BL is automatically
moved into the Off position if it is not there already.
(4) Evaluation
Crystalline anisotropic specimens can be separated into optical uni- and biaxial specimens, in each case with
an "optically positive" or "negative" character.
Uniaxial crystals display a black cross when the optical axis is parallel to the direction of view. Depending on
the level of birefringence and specimen thickness, concentrically arranged colored interference rings (so-
called isochromes) may appear (see also Fig. 4-12, second row).
The lines of this black cross remain closed when the stage is rotated. Depending on the section it may lie
within or outside the displayed objective pupil.
With optically biaxial crystals, the cross resolves into two dark hyperbola branches (the so-called isogyres)
depending on stage rotation, which are surrounded by colored interference patterns depending on the
amount of birefringence and specimen thickness (suggestive of the figure "8").
04/2020 430037-7444-001 99
OPERATION
Axiolab 5 Illumination and contrast methods in transmitted light ZEISS
Lambda/4 plate
(position of black
spots)
In the case of less favourable sections in which the cross-hair center is optically uniaxial or the isogyres are
optically biaxial specimens outside the objective pupil, an assessment is possible as follows:
• If the black isogyres are straight and they run parallel to the pupil (in relation to the cross-hairs), the
specimen is optically uniaxial.
• If the black isogyres are curved lines which wander on a circular path through the pupil, the specimen is
optically biaxial.
NOTE
Axial figures can often be better displayed with circular polarization. In particular, the axis angle of
optically biaxial specimens (quasi the distance between the isogyres) can be determined more
clearly. The optical character can also be determined. The lambda compensator (6x20mm),
arranged in the compensator slot, is used for this purpose.
NOTE
Two storage compartments for 6x20mm slides are located on the back side of the conoscopy
stand.
(1) Application
The transmitted light polarization method is used for samples which change the polarization of the light. Such
samples are called birefringent. Examples include crystals, minerals or polymers. If such birefringent
substances are observed between crossed polarizers, the birefringent portion of the sample appears bright
while it’s surrounding remains dark.
A birefringent substance can be recognized by rotating the sample by 360° between crossed polarizers. The
sample should show four bright and four dark appearances during the rotation. During the rotation procedure,
interference colors appear that range from gray (mostly for biological samples) through white, yellow and red
to blue, depending on the birefringence, thickness and orientation of the sample. The interference colors may
be of the first or a higher order.
(2) Instrumentation
On the Axiolab 5 microscope for transmitted light conoscopy:
− Strain-free objectives
− Pol rotary stage
− Polarizer D (rotatable or fixed)
− Analyzer slider D
− Lambda compensator or lambda/4 compensator
NOTE
A depolarizer (quartz depolarizer) should be installed in all microscopes used for examining mineral/geological
specimens.
A depolarizer extinguishes undesirable polarization effects (e.g. false or pseudo-pleochroism) that may occur
behind the analyzer (e.g. on prism surfaces in the tube), or shifts them to higher orders.
Fig. 4-13 Components for transmitted light polarization on the conoscopy stand
(1) Application
The determination of the polarization direction of n or n' respectively (polarization direction with the
absolute or relative largest index of refraction) and n or n' respectively (polarization direction with the
absolute or relative smallest index of refraction) relative to the morphological directions, e.g. of crystal
surfaces, crystal needles or fibers, provide an important signature of the material. This method is also used in
the diagnosis of bio-crystals (e.g. gout and pseudo-gout).
Fig. 4-14 Determining the polarization direction n' using a synthetic fiber as an example
NOTE
An adjustment of the crossline reticle is not necessary when working with the intermediate plate
and the binocular photo tube Pol (425520-9100-000).
• Rotate the rotary stage Pol with the sample, e.g. a synthetic fiber, until the sample appears as dark as
possible. In this position, the fiber extends parallel to one of the two directions of the crossline reticle.
NOTE
Do not change the inter-pupillary distance on the binocular tube, as the angle of the crossline
reticle to the fiber will be changed.
• Now turn the stage on by 45° so that the longitudinal axis of the fiber is oriented NE-SW (Fig. 4-15). The
sample will display the greatest brightness here (diagonal position). In this position the sample may have
any color.
• Insert the lambda compensator (473704-0000-000).
Like the sample, the lambda compensator is a birefringent object, albeit with a defined path difference of
550 nm and the principal polarization direction n definitely oriented in a NE-SW direction.
When the lambda compensator is moved into the beam path, the sample changes its color. The type of color
change depends on the orientation of the sample (NE-SW or NW-SE).
The changes in color are attributable to optical interference. The interference colors (path differences) in both
diagonal positions (NE-SW and NW-SE) of the sample must be compared in this connection.
The path difference results from the superimposition (interference) of the polarization direction of the sample
over the polarization direction of the lambda compensator.
The largest path difference occurs when the polarization direction of the sample with the absolute or relative
highest refractive index (n or n’) is parallel to the principal polarization direction of the lambda compensator.
The sample will then appear greenish-blue, for example (Fig. 4-14/2).
The smallest path difference occurs when the polarization direction of the sample with the absolute or relative
lowest refractive index (n or n’) is perpendicular to the polarization direction of the lambda compensator.
The sample will then appear yellow, for example (Fig. 4-14/3).
(4) Conclusions
The gray-white color appearing first in the bright
position in the above example (Fig. 4-14/1)
corresponds to a path difference of 150 nm according
to the Michel-Lévy color chart (Fig. 4-15).
When the lambda compensator is brought into the
beam path, the non-birefringent "surroundings" of
the synthetic fiber appear dark red, which
corresponds to the path difference of the
compensator of 550 nm (1st order interference color
for the path difference of 550 nm corresponds to
1 ).
If the polarization direction (n or n') of the
birefringent sample to be examined is parallel to the
principal polarization direction (n) of the lambda
compensator, i.e. in the NE-SW direction, the path
difference of the sample (e.g. gray-white: 150 nm)
and the path difference of the lambda compensator
(red: 550 nm) add up. This results in a color change of Fig. 4-15 Schematic diagram of the color charts
the sample from grayish white to greenish-blue according to Michel-Lévy
(resulting path difference = 700 nm).
If the polarization direction of the birefringence
sample to be examined is perpendicular to the principal polarization direction of the lambda compensator, i.e.
in the NW-SE direction, the path difference of the sample (e.g. gray-white: 150 nm) is subtracted from the
path difference of the compensator (red: 550 nm). In this case, the interference color of the sample visibly
changes from gray-white to orange (resulting path difference = 400 nm).
NOTE
The Michel-Lévy color charts are available in the literature catalog no. 42-312.
The measurement compensators are required for exact measurement. These return, i.e. compensate, the path
difference created by the specimen to zero (black of the first order).
Whereas in the above-described methods the addition or subtraction position was of interest, only the
subtraction position is of interest in the measurement.
Path differences in the specimen can assume very small values (1/50 or 10 nm) and very large values (more
than 10 or approx. 5500 nm and more) and thus determine the compensator appropriate for the
measurement.
NOTE
A prerequisite for the occurrence of the color change effect may be the evaluation in two
specimen positions rotated at an angle of 90° from each other, plus a centered stage.
NOTE
The DE SENARMONT compensation method requires the use of the rotatable analyzer.
− After insertion of the lambda compensator and rotation of the specimen by 90°, the interference color
remains white; in this case, however, it is a "higher-order white" and thus the path difference is > 5 .
The suitable compensator is:
the K 0-30 tilting compensator (Accessory 000000-1115-698)
− A dark gray as the interference color indicates very small path differences (/10 or 54.6 nm).
The accompanying instructions must be followed for the measurement preparation and procedure.
(1) Application
Unlike standard polarization contrast, circular polarization contrast does not show any dark (extinction)
positions that depend on the angle of rotation (azimuth) of the specimen relative to the polarizer or analyzer.
This means that when the stage is rotated, the image impression remains the same, as the light/dark positions
are omitted. With optical anisotropy, all transparent specimens display the characteristic interference colors.
(2) Instrumentation
− Strain-free objectives
− Pol rotary stage
− Circular polarizer D (no polarizers may be adapted on the condenser) including corresponding the
lambda/4 plate.
• An (anisotropic) specimen should not be examined until after the above adjustment has been completed.
• Reinsert the specimen to be examined.
The interference colors – which depend on the material, specimen thickness and orientation – of the
specimens appear constant and independent of stage rotation.
NOTE
For a high-contrast image with higher-magnification objectives (from approx. 20x) the illumination
aperture must be reduced to a value between 0.15 and 0.20, i.e. the aperture diaphragm must be
closed accordingly.
The effect of the lambda/4 plate (Fig. 4-16/4) can be undone by either swiveling it out of the beam
path or turning it with the lever (Fig. 4-16/3) to one of its two click-stop positions.
4.2.6 Configuring transmitted light polarization for conoscopic observation – determining the optical
character of crystals
For the classification (and thus identification) of crystalline material, the analysis of an interference image in
the objective pupil provides more valuable information than the examination of the specimen itself. This
image is visible in the eyepiece when an additional lens (known as a Bertrand lens) is switched on.
Alternatively, the auxiliary microscope or a diopter may be used to view the interference image.
In contrast to orthoscopy, this is referred to as conoscopy, because the illumination is ideally provided by a
wide open cone. In practice this means that the aperture diaphragm is fully open and the objective should
likewise have a large aperture.
4.2.6.1 Application
Crystal analysis is used to determine the optical character of transparent and weakly absorbent crystals. This
method is also referred to as conoscopy.
Its main application is classic mineral microscopy. However, synthetic crystals, industrial minerals and plastics
(e.g. films) can also be identified and characterized.
(1) Instrumentation
Conoscopic viewing is preferably carried out on the Axiolab 5 microscope for transmitted light conoscopy.
− Strain-free objectives; recommended:
N-Achroplan 50x/0.8 Pol objective or
EC Plan-Neofluar 40x/0.9 Pol objective
− Pol rotary stage
− Polarizer D (rotatable or fixed)
− 0.9 Pol condenser
ATTENTION
The movements of rotary knobs A and BL and the respective setting wheels are coupled to each
other. Only one control element should therefore be operated at a time and the movement of the
other should not be inhibited or blocked. Otherwise, mechanical damage may occur.
NOTE
If rotary knob BL is set to the On position, rotary knob A will be automatically set to the On if it is
not already there.
If, on the other hand, rotary knob A is set to the Off position, rotary knob BL will be automatically
set to the Off position if it is not already there.
(3) Evaluation
Crystalline anisotropic specimens can be separated into optical uni- and biaxial, in each case with an "optically
positive" or "negative" character.
Uniaxial crystals display a black cross when the optical axis is parallel to the direction of view. Depending on
the level of birefringence and the specimen thickness, concentrically arranged colored interference rings (so-
called isochromes) may appear (see also Fig. 4-12, second row).
The lines of this black cross remain closed when the stage is rotated. Depending on the section it may lie
within or outside the displayed objective pupil.
With optically biaxial crystals, the cross resolves into two dark hyperbola branches (the so-called isogyres)
depending on stage rotation, which are surrounded by colored interference patterns depending on the
amount of birefringence and specimen thickness (suggestive of the figure "8").
Inserting a lambda compensator (473704-0000-000) or a lambda/4 (473714-0000-000) or a 0-4 lambda wedge
compensator (000000-1140-663) in the compensator slot with the initial state of the axial figure being as
illustrated in Fig. 4-18 results in the following changes in color (shown schematically in the blue and yellow
areas) of the axial figure, thus allowing differentiation in "optically positive" and "optically negative".
Lambda/4 plate
(position of black
spots)
In the case of less favorable sections in which the cross-hair center is optically uniaxial or the isogyres are
optically biaxial specimens outside the objective pupil, an assessment can be done as follows:
• If the black isogyres are straight and they run parallel to the pupil (in relation to the cross-hairs), the
specimen is optically uniaxial.
• If the black isogyres are curved lines which wander on a circular path through the pupil, the specimen is
optically biaxial.
NOTE
Axial figures can often be better displayed with circular polarization. In particular, the axis angle of
optically biaxial specimens (quasi the distance between the isogyres) can be better determined.
The optical character can also be determined. the lambda compensator (6x20mm), arranged in the
compensator slot, is used for this purpose.
NOTE
Two storage compartments for 6x20mm slides are located at the back of the conoscopy stand.
4.3.1 Configuring reflected light brightfield microscopy using the KÖHLER method
(1) Application
Reflected light brightfield microscopy is the simplest and most common optical microscopy method for
examining opaque samples or specimens, e.g. material sections or wafers.
For a true-to-object imaging, indirect ray bundles, i.e. ray bundles diffracted and scattered on the specimen
details, are of major importance in addition to the so-called direct ray bundles. The higher the proportion of
indirect bundles of rays (aperture), the more realistic the ABBE microscopic image will be.
The cone of light emitted by the reflected light unit is reflected on a color-neutral beam splitter and then
passes through the objective, which focuses the beam on the surface of the sample (so-called condenser
function). The objective collects the light reflected by the specimen, and together with the tube lens it
generates the microscopic intermediate image which can then be visually observed or objectively
documented.
(2) Instrumentation
Reflected light brightfield viewing is possible only with the stand for reflected light.
− P&C ACR brightfield reflector module for reflected light in the reflector turret
• To adjust the aperture diaphragm (image contrast), remove an eyepiece from the eyepiece tube and look
into the tube with the naked eye, or use the auxiliary microscope instead of the eyepiece.
This functions only with sufficiently reflective specimens.
• For specimens with medium contrast characteristics, set the aperture diaphragm with the knurled wheel
(Fig. 4-19/1) to between 2/3 and 4/5 of the exit pupil diameter of the objective.
In most applications, this aperture diaphragm setting provides optimal contrast at almost ideal resolution, and
is therefore the best compromise for the human eye.
• Then reinsert the eyepiece, adjust the focus with the coaxial coarse and fine focusing knobs and adjust the
brightness to the reflected light specimen. Readjust aperture stop diameter after each objective change.
NOTE
Never use the aperture diaphragm for controlling image brightness. Use the Intensity/LM knob
(Fig. 4-19/4) for illumination intensity!
(1) Application
Reflected light darkfield microscopy is used for examining incompletely reflecting surfaces with different
degrees of reflectivity (ideal reflected light brightfield specimens), i.e. with scratches, ruptures, pores or other
disruptions to the even surface. All these light-scattering details light up brightly in the darkfield, whereas the
even surface remains dark.
(2) Instrumentation
Observations in the reflected light darkfield can be made only on Axiolab 5 microscopes for reflected light.
− Epiplan-Neofluar, EC Epiplan-Neofluar, Epiplan objectives with the additional designation “HD”
− ACR P&C darkfield reflector module for reflected light
NOTE
The stand for reflected light is equipped with a built-in darkfield stop.
4.3.3 Configuring reflected light polarization – Proof of bireflectance and reflexion pleochroism
(1) Application
Reflected light polarization is a further contrasting method for cut surfaces of mineral ore, coal, ceramic
products, certain metals and alloys. Depending on the orientation of the crystals and specimen details, the cut
surfaces often react differently when reflected in linear polarized light
The illumination light is linearly polarized by the polarizer before passing through the objective onto the
specimen surface, where it is reflected. Here the beam parts experience path differences depending on the
structure and polarization of optical rotations which, when passing through the analyzer, are displayed in
different shades of gray. The gray can be converted into a color contrast with the aid of a compensator
equipped with a lambda plate.
With objectives of very low magnification, a rotatable lambda/4 plate arranged in front of the objective
(Antiflex cap) permits the reflections to be eliminated even with "dark" specimen surfaces, which otherwise
would be unavoidable.
(2) Instrumentation
Observations in the reflected light darkfield can be made only on Axiolab 5 microscopes for reflected light.
− Pol rotary stage
− Epiplan-Neofluar Pol, EC Epiplan-Neofluar Pol, Epiplan Pol objectives
− C DIC/DIC/TIC ACR P&C or DIC/Pol ACR P&C or DIC Red I ACR P&C reflector module
or Pol ACR P&C reflector module in reflector turret
− Analyzer slider D, fixed or lambda compensator, 6x20mm or Lambda/4, 6x20mm
A specimen is bireflective when the details display differences in brightness and color which change when the
stage is rotated.
For samples with low bireflectance it is advisable to use the rotatable analyzer equipped with a lambda plate.
Pleochroism can be detected when the color of the specimen changes when the stage is rotated (reflected
light polarizer turned on, analyzer turned off).
Only powerful LEDs are supplied as FL excitation light sources in the Axiolab 5 program with the following
options:
LED module 385nm for Axio 423052-9593-000
LED module 470nm for Axio 423052-9573-000
LED module 505nm for Axio 423052-9562-000
LED module 565nm for Axio 423052-9602-000
LED module 625nm for Axio 423052-9522-000.
(2) Instrumentation
Observations in reflected light fluorescence can be made only on Axiolab 5 microscopes for reflected light and
reflected light fluorescence.
− Recommended objectives: EC Plan-Neofluar or Fluar (UV excitation)
− LED modules for FL excitation (installed in FL stand)
− FL P&C reflector modules equipped with respective filter sets
− Fluorescence protection shield
NOTE
If the lambda compensator used for the transmitted light polarization method is still in its slot
above the nosepiece, it must be removed before configuring reflected light fluorescence.
• Slide the fluorescence protection shield (Fig. 4-20/8) into the compensator slot above the nosepiece.
• Swivel in the 20x/0.50 EC Plan-Neofluar objective on the nosepiece (Fig. 4-20/2).
• Initially set transmitted light illumination by pushing the TL button (Fig. 4-20/5).
• If necessary, turn the condenser turret (Fig. 4-20/7) to the BF position for transmitted light brightfield (or
phase contrast if using a Ph objective) and seek the specimen detail to be examined.
• Adjust the light intensity by turning the Intensity/LM knob (Fig. 4-20/3) and focus (Fig. 4-20/6).
• On the reflector turret (Fig. 4-20/9), select the FL P&C reflector module with the desired fluorescence filter
combination (depending on the excitation mode) and swivel it into the beam path.
• Use the LED selection knob (Fig. 4-20/1) to swivel the desired LED (UV, B or G) into the beam path.
ATTENTION
To avoid dazzling when switching between the LEDs, the brightness should be reduced slightly
beforehand.
NOTE
When switching among the three LEDs, the current brightness setting is adopted.
1 LED selection knob for swiveling in the LED UV (385 nm) or LED B (475 nm) or LED G (555 nm)
2 Nosepiece
3 Intensity/LM knob
4 Reflected light button (RL)
5 Transmitted light button (TL)
6 Focusing drive
7 Condenser turret
8 Fluorescence protection shield
9 Reflector turret
ATTENTION
Do not use acetone to clean the front lens of the Pol condenser.
Please comply with the following guidelines if the microscope is to be used in a hot and humid climate:
• Keep the instrument in bright, dry and well-aired rooms, with humidity ≤ 75%; in particular, sensitive parts
such as objectives and eyepieces should be kept in special dry closets.
Precision optical instruments are always susceptible to mold if they are kept and used under the following
conditions:
− If the relative humidity > 75% over periods exceeding three days at temperatures between +15 °C and
+35 °C.
− if they are set up in dark rooms without sufficient ventilation.
− if there are dust and fingerprints on optical surfaces.
CAUTION
5.3 Troubleshooting
5.3.1 Microscope
Repairs of mechanical, optical or electronic components inside the instrument and electrical components of
Axiolab 5 microscopes may be performed only by ZEISS service staff or specially authorized personnel.
To ensure optimal configuration and trouble-free function of your microscope over a longer period of time, we
recommend that you enter into a service/maintenance agreement with ZEISS.
For subsequent orders or when service is required, please get in touch with your local ZEISS representative.
If servicing is required, please contact your local representative or
[email protected]
www.zeiss.com/microscopy
For updating the firmware, please use the following link for downloading the latest version:
https://www.zeiss.com/microscopy/us/downloads.html
6 APPENDIX
6.1 List of abbreviations
AC Alternating current
ACR Automatic Component Recognition
AF Width across flats
B/BF Brightfield
BL Bertrand lens
CSA Canadian Standards Association
C DIC Differential Interference Contrast in circular polarized light
CGT Cover glass thickness
D/DF Darkfield
d Diameter (e.g. filter)
DC Direct Current
DIC Differential Interference Contrast
DIN Deutsches Institut für Normung (German Institute for Standardization)
EC European Community
EN Euronorm (European standard)
Ergo Ergonomic/Ergonomics
FL Fluorescence
foc. focusable
GW Suitability for glasses wearers
HDMI High Definition Multimedia Interface
IEC International Electrotechnical Commission
IP Internal protection (through housing)
ISO International Standardization Organization
L Left (coaxial knurled knob left on mechanical stage)
LED Light emitting diode
LM Light manager
OSD On-screen display
Ph Phase contrast
phot photographic
PL Plan
Pol Polarization
P&C Push&Click
6.2 Index
Page
A
Ametropia.............................................................................................................................................................. 80
Analyzer ................................................................................................................................... 28, 95, 102, 103, 107
Analyzer slider ................................................................................................................................................. 90, 91
Aperture diaphragm ............................................................................................................................ 30, 32, 40, 82
Auxiliary microscope ............................................................................................................................................. 54
Axiocam ..................................................................................................................................................... 43, 49, 72
Axiolab 5 stand, Bio-TL .......................................................................................................................................... 22
Axiolab 5 stand, Bio-TL/FL ..................................................................................................................................... 24
Axiolab 5 stand, Mat-TL/RL ................................................................................................................................... 32
Axiolab 5 stand, Pol-TL .......................................................................................................................................... 26
Axiolab 5 stand, Pol-TL/Conoscopy ....................................................................................................................... 28
Axiolab 5 stand, Pol-TL/RL ..................................................................................................................................... 30
B
Bertrand lens ........................................................................................................................................... 28, 98, 109
Binocular photo tube............................................................................................................................................. 36
Binocular tube ................................................................................................................................................. 35, 53
Bireflexion ........................................................................................................................................................... 115
Birefringence ................................................................................................................................................. 89, 101
Bright field ..................................................................................................................................................... 81, 112
C
Cable holder ............................................................................................................................ 22, 24, 26, 28, 30, 32
Care ..................................................................................................................................................................... 119
Carrying handle ......................................................................................................................................... 22, 26, 28
Centering screw for condenser ............................................................................................... 22, 24, 26, 28, 30, 32
Checking the instrument ..................................................................................................................................... 120
Circular polarization contrast ........................................................................................................................ 95, 107
Coarse adjustment .................................................................................................................. 22, 24, 26, 28, 30, 32
Color glass carrier .................................................................................................................................................. 88
Color tables ........................................................................................................................................................... 93
Comfortable ergo tube .......................................................................................................................................... 24
Components .......................................................................................................................................................... 35
Optional ............................................................................................................................................................. 73
Standard ............................................................................................................................................................ 51
Condenser ........................................................................................... 21, 22, 24, 26, 28, 30, 32, 40, 62, 75, 81, 84
Condenser carrier ........................................................................................................ 21, 22, 24, 26, 28, 30, 32, 62
Condenser, darkfield ............................................................................................................................................. 63
Conoscopy ............................................................................................................................................... 28, 98, 109
Contrasting techniques ......................................................................................................................................... 15
Control and functional elements....................................................................................... 21, 24, 26, 28, 30, 32, 35
Co-observer unit .................................................................................................................................................... 73
D
Dark field ............................................................................................................................................................. 114
Dark field stop ....................................................................................................................................................... 84
Default factory settings ......................................................................................................................................... 78
Instruments, instrument components and methods described in this manual are protected by the following
patents:
− see label on microscope stand