MOCVD Manual v100
MOCVD Manual v100
Manual
for
Room EP 240
at
Overview of System Operation and Features ................................................................................................................ 6 Operational features ................................................................................................................................................. 6 Sources .................................................................................................................................................................. 6 Carrier gases .......................................................................................................................................................... 6 Input gas control .................................................................................................................................................... 6 Gas delivery plumbing ........................................................................................................................................... 6 Reactor chamber ................................................................................................................................................... 7 Exhaust gas control ................................................................................................................................................ 7 Operating controls ................................................................................................................................................. 7 Safety features........................................................................................................................................................... 8 Valve switching features assuring safe operation ................................................................................................. 8 Vacuum system operational features .................................................................................................................... 8 Interlock system .................................................................................................................................................. 10 Control panel design criteria ................................................................................................................................... 11 Gas system criteria .................................................................................................................................................. 12 Computer interface ................................................................................................................................................. 14 Emergency Procedures ................................................................................................................................................ 15 Background .............................................................................................................................................................. 15 Action to be taken ................................................................................................................................................... 16 Level of gas below TLV ......................................................................................................................................... 16 Level of gas above TLV......................................................................................................................................... 16 Brief summary of emergency procedures ............................................................................................................... 17 Interlock System .......................................................................................................................................................... 18 Alarm Inputs ............................................................................................................................................................ 18 Interlock Structure ................................................................................................................................................... 20 ABORT Status ....................................................................................................................................................... 20 A-Level Status ...................................................................................................................................................... 20 B-Level Status ...................................................................................................................................................... 20 C-Level Status ...................................................................................................................................................... 20 D-Level Status ...................................................................................................................................................... 21 Hydrogen/Nitrogen ..................................................................................................................................................... 22
Table of Contents
Gas Distribution ....................................................................................................................................................... 22 Hazards .................................................................................................................................................................... 22 Precautions .............................................................................................................................................................. 22 Normal Operation .................................................................................................................................................... 23 Start of day .......................................................................................................................................................... 23 End of day ............................................................................................................................................................ 23 Interconnections ...................................................................................................................................................... 24 Interlocks ................................................................................................................................................................. 24 Hydrogen purification system ................................................................................................................................. 24 Hydrogen purifier start-up procedure ................................................................................................................. 24 Hydrogen purifier shut-down procedure............................................................................................................. 25 Hydrogen purifier idling procedure ..................................................................................................................... 26 Hydrogen cylinder replacement .............................................................................................................................. 26 Hydrogen cylinder change with gas manifold already purged ............................................................................ 26 Initial connection of hydrogen cylinders. ............................................................................................................ 27 Purging the hydrogen gas manifold ..................................................................................................................... 28 Nitrogen purification system ................................................................................................................................... 28 Purging the nitrogen purification system ............................................................................................................ 28 Nitrogen cylinder replacement ................................................................................................................................ 29 Nitrogen cylinder replacement without purge .................................................................................................... 29 Nitrogen cylinder replacement with purge ......................................................................................................... 29 Trimethylaluminum (TMA) .......................................................................................................................................... 30 Gas Distribution ....................................................................................................................................................... 30 Hazards .................................................................................................................................................................... 30 Precautions .............................................................................................................................................................. 30 Normal Operation .................................................................................................................................................... 30 Start-of-day: ......................................................................................................................................................... 30 End of day: ........................................................................................................................................................... 31 Interconnections ...................................................................................................................................................... 31 Interlocks ................................................................................................................................................................. 31 Programming commands for remote operation ..................................................................................................... 32 Removing and replacing source bubbler ................................................................................................................. 32 Bubbler removal .................................................................................................................................................. 32 Bubbler installation.............................................................................................................................................. 33
Ethyldimethylindium (EDMln) ..................................................................................................................................... 35 Gas Distribution ....................................................................................................................................................... 35 Hazards .................................................................................................................................................................... 35 Precautions .............................................................................................................................................................. 35 Normal Operation .................................................................................................................................................... 35 Interconnections ...................................................................................................................................................... 36 Interlocks ................................................................................................................................................................. 36 Programming commands for remote operation ..................................................................................................... 37 Removing and replacing source bubbler ................................................................................................................. 37 Bubbler removal .................................................................................................................................................. 37 Bubbler installation.............................................................................................................................................. 38 Tertiarybutylarsine (TBAs) ........................................................................................................................................... 40 Gas Distribution ....................................................................................................................................................... 40 Hazards .................................................................................................................................................................... 40 Precautions .............................................................................................................................................................. 40 Normal Operation .................................................................................................................................................... 40 Interconnections ...................................................................................................................................................... 41 Interlocks ................................................................................................................................................................. 41 Programming commands for remote operation ..................................................................................................... 42 Removing and replacing source bubbler ................................................................................................................. 42 Bubbler removal .................................................................................................................................................. 42 Bubbler installation.............................................................................................................................................. 43 Tertiarybutylphosphine (TBP) ...................................................................................................................................... 45 Gas Distribution ....................................................................................................................................................... 45 Hazards .................................................................................................................................................................... 45 Precautions .............................................................................................................................................................. 45 Normal Operation .................................................................................................................................................... 46 Interconnections ...................................................................................................................................................... 46 Interlocks ................................................................................................................................................................. 46 Programming commands for remote operation ..................................................................................................... 47 Removing and replacing source bubbler ................................................................................................................. 47 Bubbler removal .................................................................................................................................................. 47 Bubbler installation.............................................................................................................................................. 48 Iron carbonyl (FE) ........................................................................................................................................................ 50
Gas Distribution ....................................................................................................................................................... 50 Hazards .................................................................................................................................................................... 50 Precautions .............................................................................................................................................................. 50 Normal Operation .................................................................................................................................................... 50 Operation without additional hydrogen dilution: ............................................................................................... 51 Operation with hydrogen dilution: ...................................................................................................................... 51 Interconnections ...................................................................................................................................................... 53 Interlocks ................................................................................................................................................................. 53 Programming commands for remote operation ..................................................................................................... 54 Removing and replacing source bubbler ................................................................................................................. 54 Bubbler removal .................................................................................................................................................. 54 Bubbler installation.............................................................................................................................................. 54 Trimethylgallium (TMG)............................................................................................................................................... 56 Vacuum Control ........................................................................................................................................................... 95 Front Panel Functions .............................................................................................................................................. 95 Indicator Lamps ................................................................................................................................................... 95 Switches ............................................................................................................................................................... 97 Normal Operation .................................................................................................................................................... 98 Interconnections ...................................................................................................................................................... 99 Interlocks ............................................................................................................................................................... 100 Programming commands for remote operation ................................................................................................... 101 Nitrogen Flush System ........................................................................................................................................... 102 Setting the flow rates for the nitrogen flush system ......................................................................................... 102 Substrate Control Panel ............................................................................................................................................. 105 Front Panel Switches ............................................................................................................................................. 105 Programming commands for remote operation ................................................................................................... 106 Induction Heater .................................................................................................................................................... 107
Operational features
Sources
1. The group V sources are provided in the form of gases by the hydrides AsH3 and PH3 diluted. Quantities are limited to 27 cu. ft. cylinders to reduce the consequences of a catastrophic release of the contents. These sources may also be provided by alkyl bubblers of TertiaryButylArsine (TBAs) or TertiaryButylPhosphine (TBP) Group III sources are provided by alkyl bubblers of EthylDiMethylIndium (EDMln), TriMethylGallium (TMG), and TriTethyAluminum (TMA). N-dopants are provided either by silane diluted to 10 percent in hydrogen or by a TertaEthylTin (TESn) bubbler. P-dopants are provided either by DiMethylZinc (DMZ) diluted to 0.1 percent in hydrogen or by a carbon tetrachloride bubbler. The dopant for growing semi-insulating material is iron provided by a ferrocene bubbler.
2. 3. 4. 5. 6. 7.
Carrier gases
8. 9. Hydrogen from a palladium purifier is used as the carrier for all the alkyl sources and as the primary diluent for the reactor gases. Nitrogen purified by a Nanochem purifier is used as an alternative diluent and purge gas for displacement of hydrogen in the reactor.
Reactor chamber
17. The circular cross section reaction chamber is part of a vertical flow system in which gases enter from the top and impinge on a horizontal substrate supported on a graphite susceptor which is heated by inductive coupling from an RF magnetic field. 18. A high velocity air stream flows over the walls of the quartz upper deposition chamber to keep the walls cool and prevent deposition on the quartz. The lower portion of the chamber which is constructed of stainless steel is water cooled. 19. The system is operated at pressures in the range 0.01-0.1 atmosphere.
Operating controls
26. The reactor can be operated by manual controls on the reactor panel. All valving is done with switches, and the gas flow of the individual gas source is digitally displayed. 27. Alternatively the reactor can be computer operated. Interfacing of the valve and flow controllers is provided via a Hewlett Packard Multiprogrammer controlled by a PC running LabView.
Safety features
28. Normally-off bellows valves are used for fail-safe protection from toxic gases. 29. Pneumatic valve operation assures positive, leak-tight shut-off. 30. All-welded stainless steel plumbing with Nupro VCR compression gasket fittings meet requirements for hazardous production gases. 31. A ventilated reactor cabinet with separate sections for the chamber, gas control and the electronics contains all of the hazardous gas components. 32. Excess flow shut-off valves are used on all the gas cylinders. 33. The reactor will shut down for: a. Failure of cabinet exhaust. b. Detection of PH3 or AsH3 leaks. c. Detection of hydrogen leaks. d. Loss of pneumatic valve air pressure. e. Low hydrogen pressure. f. Low nitrogen pressure. g. Power failure. h. Loss of chamber vacuum.
10
Interlock system
The system is shut down and latched-off if any of the conditions listed below arise. 48. Power failure (system power must be reset to operate any sources; only N2 flow can turn on when power is resumed). 49. AsH3/PH3 monitor alarm (set to monitor for the TLV level in both the gas cabinet exhaust and in the adjacent lab space). 50. H2 at 20% of the explosive limit alarm. 51. Exhaust ventilation failure alarm. 52. Low H2 supply pressure alarm. 53. Low N2 supply pressure alarm. 54. Low of vacuum in chamber alarm. 55. Loss of pneumatic valve air supply pressure. 56. Emergency stop button activated.
11
12
13
103.A separate sample-entry chamber and load lock system allows samples to be inserted and removed from the deposition chamber without the necessity of letting air into the reaction chamber. This improves the purity of the grown material by eliminating water vapor absorption on the chamber walls. Hazardous H2 and source gases are prevented from venting to atmosphere for operator safety. 104.The exhaust valves are interlocked to prevent simultaneous operation. 105.A 1 psig pressure relief valve prevents inadvertent over pressure of the reactor if the exhaust valve is closed.
14
Computer interface
The operations necessary for layer growth can be remotely programmed by a desktop computer. The features of this system are; 106.Uses LabView for simplified operation. 107.Magnetic disk mass storage for retraining a library of operating programs. 108.Printout of process parameters. 109.All low level functions such as keeping track of relay numbers and programming commands for the D/A converters are performed in subroutine modules which have been previously debugged and tested. The programming statements used for a growth sequence are character strings which have a logical relationship to the function being performed. This simplifies the programming and greatly minimizes the possibility of error in writing programs for growth sequences. It is not possible to program the wrong valve to be open by entering an incorrect decimal or binary number in a growth program since numbers are not used in the high-level programming commands. 110.Each gas circuit must first be set in the REMOTE mode by computer control before any other circuit functions can be programmed by computer. This two-step sequence minimizes the possibility that a valve will be accidentally turned on under computer control. 111.The gas flows necessary for a growth sequence can be programmed by set-point voltages applied by D/A converters under computer control. The set-point voltages can also be applied from potentiometers on the front panels. 112.The actual flow rates through the mass flow controllers, the pressure in the deposition chamber, and the susceptor temperature can be measured under computer control to provide automatic data logging during a growth sequence.
15
Emergency Procedures
Background
The DOD ChemLogic CL8 toxic gas monitor which monitors the MOeVD cabinet and the laboratory space in Room 130 is now connected to the building Evacuation alarm. The system is connected so that if a level of toxic gas in excess of the TLV (threshold limit value) occurs in either the MOeVD cabinet, the exhaust gas outlet, or the laboratory space in Room 130,then an alarm on the MOeVD system will sound and the the interlock circuits for the MOeVD system will go into the abort mode. All of the pneumatic valves for the MOeVD system which carry hydrogen and the source gases will close automatically. These valves cannot reopen until the system is manually reset. If the alarm is triggered by a gas release in the MOVeD cabinet then the orange light on top of the gas monitor will be on. If the alarm is triggered by a gas release from the scrubber outlet into the exhaust vent then the blue light on top of the gas monitor will be on. These two events are not life threatening and do not trigger the building alarm. If the alarm is triggered by a gas release in Room 130 then the red light on top of the gas monitor will flash, an audio alarm on the MoeVD cabinet will sound, the source and hydrogen valves will close, the Evacuation alarm for the building will sound, and the fire department will be notified. The alarm to the fire department will not indicate to origin of the alarm (Le. whether it is a fire or a toxic gas detection). The Fire Annunciator Panel on the west end of the Engineering/Physics Building, second floor, will indicate that the source of the alarm is from the toxic gas monitor in Room 131 in the event of a gas release, and a red light outside Room 118 by the door on the northeast corner of the building will come on. The fire alarm system latches on when triggered and stays on until it is reset, even if the triggering event is no longer active. However, the warning light outside Room 118 is not latched and will go out once the gas level falls below the TLV. Arsine and the hydrogen carrier gas are flammable. Arsine is a toxic colorless gas with a garlic-like odor. The odor threshold is 0.5 ppm. The OSHA TWA (time-weighted average for an 8-hour shift) ceiling is 0.05 ppm. So, you cannot smell arsine at the TWA ceiling level. The DOD ChemLogic CL8 toxic gas monitor is set to give an alarm at the TWA ceiling which is the same as the TLV. Pure arsine has a specific gravity of 2.695 (air=1). However, a mixture of 10% AsH3 in hydrogen has a specific gravity of 0.33 (lighter than air). Pure phosphine is pyrophoric. The hydrogen diluent gas is flammable. Phosphine is a toxic colorless gas with a disagreeable characteristic fishy, garlic-like, or acetylene-like odor. The odor threshold is 0.021 ppm. The OSHA TWA ceiling is 0.3 ppm so phosphine can be detected by smell at the TWA ceiling. The DOD ChemLogic CL8 toxic gas monitor is set to give an alarm at 0.05 ppm. Pure phosphine has a specific gravity 1.17 (air=1). However, a mixture of 10% PH3 in hydrogen has a specific gravity of 0.18 (lighter than air). The liquid alkyl sources used in the MOeVD system are all pyrophoric (they burn spontaneously when exposed to air). If a spill of one of these liquids occurs, the material will slowly bum as it is oxidized by the air. Some of these liquids react violently when mixed with water which should never be applied to the burning material. Use either a dry chemical fire extinguisher or an absorbent clay to smother the fire. The fumes from the burning material are in some cases toxic. Do not inhale them. If the burning material is presenting a hazard to personnel it is best to leave the area and let the material burn until it is consumed. If the smoke is escaping from Room 240 and the building Evacuation alarm has not already sounded, then activate one of the building alarm stations to evacuate the building.
16
Action to be taken
If the building Evacuation alarm sounds while the the MOCVD system is in operation, the operator will know that it is due to a toxic gas release if the alarm within the laboratory space also activates. Since the MOCVD system will shut down automatically, no corrective action should be taken by the operator. The operator should immediately leave the laboratory space and alert others in the vicinity to also leave immediately. Once outside the building, the operator should use a telephone in another building, the Physical Plant is closest, to dial 911 and alert the fire department that a response by the Hazardous Materials Team will be required. The operator should then wait outside the building to provide assistance to fire department team as needed. The system operator is a valuable resource person since he/she knows better than anyone what the fire department team will encounter. There are two possible scenarios at this point. Either the level of toxic gas in Room 130 will be above the TLV, or it won't be.
17
18
Interlock System
A series of system wide interlocks sense the states of a number of transducers which can detect conditions which require immediate intervention. Some of these conditions present immediate or potential operating hazards and some of the conditions jeopardize the integrity of the source materials. The interlocks have been connected to system controls in a manner such that appropriate action is taken without operator intervention. The system will be shut down, an alarm will sound, and a light on the ABORT panel will come on, indicating the nature of the malfunction. If a transducer, say for example the smoke detector, sends out an alarm which shuts down the system and at later time the smoke goes away, the system cannot restart unless it is manually reset. Also, the lamp on the ABORT panel will stay on indicating the nature of the fault even though it has cleared. The transducers are all wired in such a manner that their normal (or no alarm) state is a closed switch. If wires to the transducer are broken or become disconnected, an alarm will be indicated. The ABORT panel contains a key-operated switch which deactivates the alarm for servicing purposes. During normal operation the alarm should enabled and the key removed. The key can be removed only when the alarm is enabled.
Alarm Inputs
The transducer inputs and their functions are as follows: AsH3/PH3 MONITOR An output signal from the DOD Toxic Gas Monitor indicating the presence of toxic gas at either the scrubber output, the MOCVD cabinet or in room 240. If the abort panel activates and the building alarm does not, then the gas is contained within the cabinet. H2 LEAK DETECTOR An output signal from the Matheson hydrogen gas purifier indicating the presence of hydrogen gas above the level of 4000 ppm in the MOCVD cabinet. EXHAUST AIR FLOW An output signal from a differential manometer checks for loss of air flow through the venting system. LOW H2 SOURCE PRESSURE A pressure switch which checks for sufficient hydrogen pressure for proper operation of the bubbler sources. LOW N2 SOURCE PRESSURE A pressure switch which checks for sufficient nitrogen pressure for the nitrogen diluent system. CHAMBER VAC LOSS A signal indicating that the vacuum deposition chamber has come up to atmospheric pressure in an unnormal fashion. In order for this circuit to be activated the exhaust valve between the chamber and the vacuum pump must be open and the system must have first been evacuated to a pressure of 100 Torr or less. This circuit would be activated if the quartz chamber were to break in the middle of a run or if the vacuum pump failed in the middle of a run. The three conditions necessary for a chamber vacuum loss signal are (1) the exhaust valve must be open, (2) the vacuum latch must be off, and (3) the chamber must be at atmospheric pressure. See the section on vacuum control for further details. SMOKE DETECTOR A combined signal from a smoke detector and a heat detector in the MOCVDcabinet. Both are located near the vent outlet for greatest sensitivity. HIGH H2 HUMIDITY An output signal from the Panametrics hygrometer indicating high humidity in the hydrogen gas coming from the Matheson hydrogen purifier. This would probably be caused by a rupture in the internal membrane of the purifying cell. The system is shut down automatically so that the moisture in the hydrogen does not contaminate the bubbler sources.
19 HIGH N2 HUMIDITY
An output signal from the Panametrics hygrometer indicating high humidity in the nitrogen gas coming from the dessicator NP1. This would indicate that the dessicator cartridge should be replaced.
20 Indicates that the red EMERGENGYSTOP button has been pressed. This is the preferred method for shutting down the system quickly since a nitrogen flow into the vacuum pump inlet is maintained. Indicates that the power from the 24 VDC supply has been interrupted either by a failure of the power supply or an interruption of the electrical service.
POWER FAIL
Interlock Structure
The interlock system operates by disconnecting the +24 v power lead from circuits which are to be shut down. Since all of the pneumatic systems are designed to be in the closed state when power is removed, disconnecting the power provides a safe shutdown of the system. An alarm signal from any of the above interlocks de-energizes relay ABRTK12 which removes power from the A level interlock system. If no alarms are being received from any of the above sensors, then all of the A-level systems are powered up. If the A-level system is up and the vacuum valve (EXPV2) between the deposition chamber and the vacuum pump is open then all B-Ievel systems are powered up. If the B-Ievel system is up and the vacuum chamber is evacuated (pressure < 400 Torr) and the SCRUBBER OXIDATION switch on the VACUUM CONTROL panel is off, then the all Clevel systems are powered up. If the C-Ievel system is up and either the nitrogen or the hydrogen manifold diluent gas is turned on, then all DIevel systems are powered up.
ABORT Status
If the A-level system is powered down (ABORT status) then the following components are active: 117.lf the vacuum pump is on, then EXSV5 is energized which turns on EXPV5. This feeds nitrogen into the inlet of the vacuum pump to maintain a pressure of 0.1 Torr and prevent the upstream plumbing from being coated with vacuum pump oil. The green DEP CHAMBER light for the NITROGEN FLUSH system will be on. The other NITROGEN FLUSHfeeds will be disabled. 118.The N2 CARRIER function on the HYDROGEN/NITROGENcontrol panel is enabled. The H2 CARRIER function is disabled.
A-Level Status
If the A-level system is powered up then the following components are active: 119.The REMOTE/LOCALfunction is activated on the ALUMINUM,.INDIUM, ARSENIC LIQUID, PHOSPHORUS LIQUID, ARSINE, PHOSPHINE, IRON, GALLIUM, N DOPANT, and PDOPANT control panels. 120. The ACTIVE/INACTIVE function is activated on the ALUMINUM,.INDIUM, ARSENIC LIQUID, PHOSPHORUS LIQUID, ARSINE, PHOSPHINE, IRON, GALLIUM, N DOPANT, and PDOPANT control panels. 121.The substrate spinner can be activated. 122.The substrate heater can be activated 123.The substrate can be raised or lowered. 124.The nitrogen manifold diluent can be turned on if there is vacuum in the depositon chamber. 125.Solenoid valve AUXSV1 is energized which opens ASGPV20, PGPV20, NDOPPV20, and PDOPPV20.
B-Level Status
If the B-Ievelsystem is powered up then all A-level components are active in addition to the following: 126.The nitrogen manifold diluent can be turned on regardless of whether or not there is vacuum in the depositon chamber.
C-Level Status
If the C-Ievel system is powered up then all B-Ievel components are active in addition to the following:
D-Level Status
If the D-Ievel system is powered up then all C-Ievel components are active in addition to the following: 128.All valves for the ALUMINUM system can be opened. 129.All valves for the INDIUM system can be opened. 130.All pneumatic valves for the ARSENIC LIQUID system can be opened. 131.All pneumatic valves for the PHOSPHORUS LIQUID system can be opened. 132.All pneumatic valves for the IRON system can be opened. 133.All pneumatic valves for the GALLIUM system can be opened. 134.All pneumatic valves for the N-DOPANT system can be opened. 135. All pneumatic valves for the P-DOPANT system can be opened.
22
Hydrogen/Nitrogen
Gas Distribution
The hydrogen and nitrogen gas flows are controlled from a common panel on the electronics cabinet. Nitrogen and/or hydrogen gas must flow through the Swan fast switching manifold in order for the source gas controllers to be enabled. The flow rates for nitrogen and hydrogen manifold diluent gases can be controlled separately. In addition either nitrogen or hydrogen gas can be used for the transport gas through the source bubblers. Nitrogen is used only for purging purposes and hydrogen is used both for both source purging and vapor transport during growth of layers. The hydrogen and nitrogen flows for the vapor carrier system cannot be turned on simultaneously.
Hazards
136.Nitrogen is an asphyxiant. If the oxygen in the room were displaced with nitrogen, suffocation could occur. With the cabinet ventilation (approximately 500 cubic feeVminute)turned on, this poses a small hazard since the total volume of nitrogen in the high-pressure cylinder is 230 cubic feet. If a liquid source of nitrogen is used, then the room should be evacuated if a massive leak of nitrogen gas or liquid occurs which cannot be stopped by turning off the valve on the source cylinder. 137.Hydrogen is flammable and potentially explosive. Hydrogen mixtures between the limits of 4.0% and 75% by volume in air are flammable. It is very important that hydrogen-air mixtures not be present in the components of the reactor. For example, if hydrogen mixed with air were present in the deposition chamber with the substrate heater on, an explosion could result, shattering the quartz deposition chamber. 138.The Nanochem nitrogen purification cell contains a lithium resin which reacts with oxygen and water vapor. Excessive heating of the cell would occur if air were to flow through the cell. The lines should be carefully purged before opening the inlet and outlet valves.
Precautions
139.Each hydrogen cylinder is equipped with an excess flow shutoff valve which closes if the flow exceeds 5 liters/minute. This prevents a high volume release of hydrogen in the event of rupture of one of the gas lines. The excess flow valves must be manually reset once they are tripped. 140.The hydrogen line contains an over pressure release valve (HPRV1) to protect the components in the event of failure of one of the tank regulators. If the pressure in the line exceeds 150 psig, then the hydrogen is dumped directly into the exhaust vent. 141.The hydrogen supply is enabled by a pneumatic valve (HPV17) which closes if the interlock system is in Abort status. 142.A hydrogen sensor in the MOCVD cabinet activates the interlock system and places the system in Abort status. 143.The air bleed system for scrubber oxidation cannot be turned if the hydrogen manifold diluents is on, or if any of the source gases are turned on. 144.System shutdown occurs if the pressure in the nitrogen supply line falls below 10 psi or the pressure in the hydrogen supply line falls below 50 psi.
23
Normal Operation
When the reactor is in an idle state, a small quantity of nitrogen flows through the system to continuously purge the vacuum pump. There is no hydrogen flow in the idle state. At the end of the day the hydrogen supply should be turned off at the high pressure cylinders and the nitrogen supply should be left on.
Start of day
145.Check that the nitrogen supply is on and that there is sufficient nitrogen pressure in the cylinder for the days operation; about 400 psig for a 216 cu. ft. high pressure cylinder. 146.Turn on the valves for both hydrogen cylinders. The outputs of both hydrogen regulators feed into a common supply line for the MOCVD reactor. If the pressure regulators for the tanks are set correctly, when the main tank is empty, the backup tank will automatically come on. The procedure for doing this is in the section on cylinder replacement. The outlet pressure should be 120 psig. 147.During a normal growth run the carrier gas should be hydrogen. Nitrogen should be used in the carrier gas circuit only during purging of the lines and changing cylinders. On the HYDROGEN/NITROGEN control panel the switch which selects the carrier gas has no off position. Either hydrogen is selected or nitrogen is selected. When the switch is in the CARRIER N2 position then valve NPV13 is open and HPV13 is closed. When the switch is in the CARRIER H2 position then valve HPV13 is open and NPV13 is closed. Two panel lamps indicate the gas selected. The selection of the carrier gas cannot be implemented from the remote computer. 148.The manifold diluent gas may be either hydrogen or nitrogen or a combination of the two. The selection of the gases and the flow rates may be made remotely from the computer. When the MANIFOLD N2 switch is turned on, NPV4 opens and nitrogen begins to flow through NMFC1, through the Swan manifold and into the deposition chamber. The MANIFOLD N2 DVM on the control panel indicates the nitrogen flow through NMFC1on the HYDROGEN/NITROGENcontrol panel. When the DVM1 switch is pressed in, the set point for NMFC1 can be adjusted using the FLOW SET 1 potentiometer. When the MANIFOLD H2 switch is turned on, HPV4 opens and nitrogen begins to flow through HMFC1, through the Swan manifold and into the deposition chamber. The MANIFOLD H2 DVM on the control panel indicates the hydrogen flow through HMFC1 on the HYDROGEN/NITROGENcontrol panel. When the DVM2 switch is pressed in, the set point for NMFC2 can be adjusted using the FLOW SET 2 potentiometer.
End of day
149.The MANIFOLD N2 and the MANIFOLD H2 switches should both be off. 150.The control panel switches for all of the source materials should be off. 151.Close the valves on both of the hydrogen cylinders. 152.Leave the nitrogen cylinder turned on unless the system is to be completely turned off with the vacuum pump off. If the vacuum pump is turned off, then the valve for the nitrogen cylinder may be turned off.
24
Interconnections
153.N2 CARRIER controls NSV2 which drives NPV13. 154.H2 CARRIER controls HSV2 which drives HPV13. 155.H2 ON controls relay HNK1which controls HSV1 which drives HPV4. 156.N2 ON controls relay HNK2 which controls NSV1 which drives NPV4.
Interlocks
157.All functions are disabled in Abort statuswith the exception of the N2 CARRIER. 158.In Abort status the N2 CARRIER function on the HYDROGEN/NITROGENcontrol panel is enabled. The H2 CARRIER function is disabled. 159.If the system is not in Abort status and the exhaust valve EXPV2 is closed, the nitrogen manifold diluent can be turned on only if there is vacuum in the deposition chamber. 160.If the system is not in Abort status and the exhaust valve EXPV2 is open, then the nitrogen manifold diluent can be turned on regardless of whether or not there is vacuum in the deposition chamber. 161.Only if the system in not in abort status and the exhaust valve EXPV2 is open and there is vacuum in the deposition chamber and the air bleed to the scrubber is off, can the hydrogen manifold diluent can be turned on. 162.None of the source gases or bubbler systems can be turned on unless either the hydrogen manifold diluent or the nitrogen manifold diluent is turned on. 163.If any source gas or bubbler system is turned on, then neither the hydrogen manifold diluents nor the nitrogen manifold diluent can be turned off.
25
controlled by the metering valve on the front panel of the hydrogen purifier. Adjust this valve for a flow of 200 sccm (Note this valve is somewhat unconventional in that the off position is fully clockwise). 166.Make sure that the power to the hydrogen purifier is off. 167.With the vacuum pump running, open the VAC EXHAUST switch on the VACUUM CONTROL panel. Make sure that the CONTROL VALVE OPEN light on the VACUUM CONTROL panel is on. If not press the OPEN button on the AdaptTorr ACR-26 pressure controller to fully open the pressure control valve. 168.On the HYDROGEN/NITROGEN control panel turn the MANIFOLD N2 switch to the on position and adjust the FLOW SET 1 potentiometer to maintain a minimum deposition chamber pressure of 1-2 Torr as indicated on the AdaptTorr pressure controller. Turn the MANIFOLD H2 switch on and adjust the FLOW SET 2 potentiometer a reading of 100 sccm. Wait for the deposition chamber pressure to reach the range of 2-4 Torr. This step will evacuate the UHP outlet of the purifier. 169.Make sure that the purifier is plugged in. If the red light on the front panel is steady, then the power switch can be turned on. To do this press in the red [RESET] button on the front panel and turn on the power switch. 170.The power up configuration for the temperature controller is on, with a setpoint (SV) of 400C. The setpoint can be adjusted with the up and down arrow keys on the front panel. If SV is flashing STBY or TUNE, hold down the rightmost key (three horizontal lines) until the setpoint value is steady. 171.When the temperature reaches 400C, note the reading of the Manifold H2 panel meter on the HYDROGEN/NITROGEN control panel. A flow > 25 sccm indicates a leak in the purifier cell. Do not proceed further until this is repaired. 172.Reset the latched low temperature alarm on the temperature controller by pressing the rightmost key (three horizontal lines) and then turn the H2 PURIFIER PURGE switch on the HYDROGEN/NITROGEN control panel to the off position. This will close valves NPV14 and HPV15 and open valves HPV14 and HPV16. Hydrogen gas will flow through HPV14 into the purifier and the cell temperature (PV) will rise, then fall. Some of the gas will leave the bleed port through HMV5 and HPV16. The remainder will leave through the UHP outlet. The gas flow through the bleed circuit is now controlled by the setting of HMV5, which is located to the left of the hydrogen purifier. Adjust this valve for a flow of approximately 200 sccm as indicated by the rotameter on the front panel of the purifier. 173.Observe the pressure on the output of pressure regulator PR1. It will rise to its setpoint as hydrogen displaces nitrogen in the cell and pure hydrogen appears at the output. Allow hydrogen to purge both inlet and outlet for 10 minutes. 174.Adjust HMV5 to reduce the flow through the bleed circuit to 10 sccm. 175.Adjust regulator PR1 for an outlet pressure of 20 psig. 176.Turn off the MANIFOLD H2 and MANIFOLD N2 switches on the HYDROGEN/NITROGEN control panel and continue as appropriate.
26
180.Turn the H2 PURIFIER PURGE switch on the HYDROGEN/NITROGEN control panel to the on position. Under this condition, valves NPV14 and HPV15 will be open and valves HPV14 and HPV16 will be closed. Nitrogen will flow through NPV14, through the purifier, through HPV15 and out the exhaust. The nitrogen flow rate will controlled by the metering valve on the front panel of the hydrogen purifier. Adjust this valve for a flow of 200 sccm (Note: this valve is somewhat unconventional in that the off position is fully clockwise). Allow the nitrogen flow to purge the hydrogen from the purifier cell for at least 10 minutes. It is very important that the cell not be allowed to cool with hydrogen in it. The cell can be damaged. 181.Hold down the right most key (three horizontal lines) down until the setpoint temperature flashes STBY. 182.When the cell temperature is < 100C, which takes approximately one hour, reduce the nitrogen flow by turning the valve on the front panel of the purifier counterclockwise. Set for approximately 5 sccm to conserve nitrogen. 183.Turn off the H2 MANIFOLD and N2 MANIFOLD switches on the HYDROGEN/NITROGEN control panel. Close the vacuum exhaust valve. 184.The purifier can be left in its current state or it can be unplugged. If it is left plugged in, the H2 detector inside the purifier will continue to be active.
27
192.Disconnect the VCR fitting between the plug valve and the manual valve on the outlet of the regulator. 193.Disconnect the fitting connecting the excess flow valve (either HEXFV1 or HEXFV2) to the hydrogen cylinder. Note; this connector has a left-hand thread so it should be turned clockwise (when facing the cylinder) to loosen it. 194.Replace the cap on the hydrogen cylinder and store it in a location where it is chained to the wall until picked up by the vendor. 195.Put the new tank in position, remove the cap on the cylinder, connect the regulator, and reconnect the VCR fitting. The self-centering gasket in the VCR fitting does not need to be replaced every time the tank is changed as long as it is in good condition. However, it probably should not be reused more than four or five times. 196.Turn the pressure control knob on the regulator counterclockwise two turns to reduce the pressure setting of the regulator. 197.Open the plug valve and the valve on the output of the regulator. 198.Set the knob on the excess flow valve to the reset or purge position. To do this, the black knob on the excess flow valve should be turned counterclockwise (when facing the body of the valve) to the stop. 199.Open the valve on the outlet of the hydrogen cylinder. The valve should be opened very slowly at first to avoid a sudden shock to the regulator. 200.Close the manual valve on the outlet of the other hydrogen gas regulator. 201.Insure that the tank valve for the nitrogen is on, that the pressure regulator on the tank gives a pressure of 20 psig, and that NMV1 is open. 202.With the vacuum pump running, open the VAC EXHAUST switch on the VACUUM CONTROL panel. Make sure that the CONTROL VALVE OPEN light on the VACUUM CONTROL panel is on. If not press the OPEN button on the AdaptTorr ACR-26 pressure controller to fully open the pressure control valve. 203.On the HYDROGEN/NITROGENcontrol panel turn the MANIFOLD N2 switch to the on position and adjust the FLOW SET 1 potentiometer to give a flow rate of 1000 sccm. Turn the MANIFOLD H2 switch on and adjust the FLOW SET 2 potentiometer to give a flow fate of 200 sccm. 204.The hydrogen pressure at the outlet of the regulators will begin to drop. When it drops below 140 psig, begin to adjust the pressure regulatoron the cylinder just replaced slowly upward (clockwise) until its pressure setting is 140 psig. The outlet pressure of the regulator should now remain steady. 205.Re-open the manual valve on the outlet of the regulator for the other cylinder. Turn the pressure on this regulator up to 160 psig. With the regulators set in this manner, gas in the new cylinder will not be used until the older one is empty. 206.Set the excess flow valve on the cylinder that was just replaced to the RUN position. Do this by turning the knob clockwise (as viewed facing the valve) until it is tight. 207.Turn off the hydrogen and nitrogen manifold gas flows on the HYDROGEN/NITROGEN control paned. 208.Close the VAC EXHAUST switch on the VACUUM CONTROL panel. 209.Shut down the system as appropriate.
28
214.Set the pressure on the regulator for the backup cylinder to 140 psig. 215.Set the pressure on the regulator for the 'in use' cylinder to 160 psig. 216.Open HMV1 and HMV3. 217.Set the excess flow valve for each cylinder to the RUN position. Do this by turning the knob 218.clockwise (as viewed facing the valve) until it is tight. 219.Proceed with the purging of the hydrogen gas manifold as described below.
29 230.Open valves NMV6 and NMV7. Close valve NMV8 (lever should be horizontal) 231.Shut down the system as appropriate.
30
Trimethylaluminum (TMA)
Gas Distribution
245.Trimethylaluminum is a liquid at room temperature with a vapor pressure of about 9 Torr. This means that the pressure of the vapor is 9/760 x 14.7 psia; very much less than atmospheric pressure. The vapor is transported from the stainless steel canister which contains the liquid, into the reaction chamber by bubbling hydrogen gas through the liquid. 246.The pressure of the hydrogen gas inside the liquid container is determined by the pressure setting of PR2. This is normally set at 10 psig. 247.The precise vapor pressure of the TMA is determined by the temperature of the liquid. This is controlled by immersing the liquid in a coolant bath which is controlled at a precise temperature by a small Brinkmann refrigeration unit. 248.Pressure drop to the reaction chamber occurs across the mass flow controller, ALMFC1. 249.The alkyl is exposed to the heated sensor element of the mass flow controller (110C); thus it must be stable at this temperature. 250.Loss of liquid from the bubbler into the gas lines is unlikely because pressure in the gas line is the same as the bubbler. (Closure of the bypass valve, ALPV2, is delayed to allow bubbler inlet and outlet pressure to equilibrate.) 251.Upon power failure or other abort conditions, contamination of the upstream H2lines is prevented by closure of ALPV12. 252.Adiabatic cooling could possibly cause alkyl condensation in the mass flow controller valve. It is advisable to keep the alkyl temperature below room temperature. 253.A purge mode overrides some interlocks to allow Evacuation and backfill of the bubbler lines
Hazards
254.TMA is pyrophoric. 255.Aluminum oxide (from burning TMA) is highly toxic. 256.TMA reacts violently with water which should never be used to extinguish the burning alkyl. Use a dry chemical fire extinguisher only or else let it burn.
Precautions
257.Keep the manual valves on a newly installed bubbler closed until it is connected to the system, leak checked and purged. 258.Avoid gas pressure differentials that can force the alkyl from the bubbler back into the carrier gas piping.
Normal Operation
Start-of-day:
259. Since the pneumatic valves ALPV1 and ALPV2 seal off the bubbler from the rest of the system, the manual valves on the bubbler inlet and outlet can be left open except when the bubbler is being replaced or removed. At the start-of-day the TMA is therefore ready to operate from the control panel. 260.During a normal growth sequence the switches on the front panel are turned on in a left to right sequence and turned off in a right to left sequence. With all switches closed, ALMFC1 is commanded closed, ALPV1, ALPV2, ALPV3, and ALPV4 are closed and ALPV5 is in the vent position. When the ACTIVE switch is turned on, the
31
signal to command closure of ALMFC1 is removed, ALPV2, ALPV4-and ALPV12 open and the green ACTIVE light comes on. At this point hydrogen carrier gas will flow through the bypass valve ALPV2 and into the vent manifold at a rate determined by the current value of the voltage input to ALMFC1 as determined either by the front panel potentiometer or the D/A converter output for this flow controller. The red PURGE BUBBLER light on the control panel will be on indicating that the bypass valve ALPV2 is open. When the BUBBLER IN switch is turned on, ALPV1 and ALPV3 open immediately and the green light BUBBLER IN turns on. After a twosecond delay the red PURGE BUBBLER light goes out and ALPV2 closes. This insures that the downstream pressure is not greater than the upstream pressure (a condition which would force the liquid alkyl back into the hydrogen carrier gas piping) when ALPV1 and ALPV3 open. When ALPV2 closes, the hydrogen carrier gas is no longer allowed to bypass the bubbler, and instead must flow through the bubbler transporting the alkyl vapor to the vent/run valve. When the RUN switch is turned on, the yellow RUN light comes on and the gas flow is switched from the vent manifold to the run manifold and must now pass through the deposition chamber. During manual operation from the front panel, the flow through ALMFC1 is controlled by the setting of the 10-turn DVM SELECT potentiometer on the control panel. The value of the flow setting can be read on the digital panel meter by pushing in on the black DVM SELECT button. When the button is released, the meter reads the value of the actual flow. When the gas circuit is operated remotely from the computer, the input voltage to ALMFC1 is supplied by a D/A converter in the HP Multiprogrammer and can be remotely programmed. The meter on the panel will continue to indicate the actual flow, and when the DVM SELECT button is pushed in, the meter will read the input voltage to the flow controller from the D/A converter. Shutdown of the system should proceed by reversing the sequence of the above steps. 261.The BUBBLER IN switch should be turned off during standby periods to both conserve the source and to minimize the load of hazardous chemical which enters the scrubber. The alkyl vapor will continue to be transported from the bubbler as long as the BUBBLER IN switch is on and there is gas flow through ALMFC1.
End of day:
The alkyl vapors will accelerate the deterioration of some of the elastomers from which the o-rings are fabricated. Elastomer seals have been eliminated whenever possible in the construction of this system but current technology does not allow them to be completely eliminated. They are present in the mass flow controller and in the vent/run valve. For this reason it is good practice to thoroughly flush the alkyl vapor from ALMFC1 and ALPV5 before the circuit is shut down. This can be accomplished by allowing hydrogen to flow through the system for approximately 5 minutes after the BUBBLER IN switch is closed. After this, the ACTIVE switch can be closed, sealing off the circuit.
Interconnections
262.ACTIVE controls ALSV1 which drives ALPV4 and ALPV12. 263.BUBBLER IN controls ALSV2 which drives ALPV1 and ALPV3. 264.Relays ALK1 and ALK2 control ALSV3 which drives ALPV2.
Interlocks
265.All source valves are deactivated if the following conditions exist: 266.System is in abort status. 267.The vacuum exhaust valve, EXPV2, is closed. 268.The deposition chamber is up to atmosphere. 269.The manifold diluent gas is off. 270.ALPV1 and ALPV3 cannot be opened unless ALPV4 and ALPV12 are open.
32
271.With the PURGE switch on the front panel in the locked position (key removed), ALPV2 cannot be opened if ALPV1 and ALPV3 are open. 272.The bypass valve (ALPV2) is open if the BUBBLER IN switch is closed (handle down) and the ACTIVE switch is on (handle up). 273.The ACTIVE, BUBBLER IN, and RUN switches on the front panel are disabled when the computer switches the circuit into the REMOTE mode. 274. When the bubbler is connected to the circuit by opening the BUBBLER IN switch (handle up), valves ALPV1 and ALPV3 open immediately. Valve ALPV2 closes after a 2 second time delay to insure that the pressure on each side of the bubbler is equalized. 275.The manifold diluent gas cannot be turned off if the ACTIVE switch is on.
Bubbler removal
276.Insure that the tank valve for the nitrogen is on, that the pressure regulator on the tank gives a pressure of 20 psig, and that NMV1 is open. 277.Connect the solenoid valve from the BUBBLER PURGE kit to the pneumatic control inlet of ALPV12. The connector on the 'A' port of the solenoid valve should be attached to the pneumatic inlet of ALPV12. The connector on the 'P' port of the solenoid valve should be attached to the 1/8" 0.D. nylon tubing leading to ALPV12. Insure that the switch on the control box for the BUBBLER PURGE kit is closed and plug its line cord into a 120 VAC outlet. 278.Close the manual valves on the inlet and outlet of the TMA bubbler. 279.With the vacuum pump running, open the VAC EXHAUST switch on the VACUUM CONTROL panel. Make sure that the CONTROL VALVE OPEN light on the VACUUM CONTROL panel is on. If not press the OPEN button on the AdaptTorr ACR-26 pressure controller to fully open the pressure control valve. Make sure the MANIFOLD H2 switch on the HYDROGEN/NITROGEN control panel is off. Turn the MANIFOLD N2 switch on the HYDROGEN/NITROGEN control panel to the on position and adjust the FLOW SET 1 potentiometer to maintain a minimum deposition chamber pressure of 1-2 Torr as indicated on the AdaptTorr pressure controller. 280.On the HYDROGEN/NITROGEN panel set the N2 CARRIERlH2 CARRIER switch to the N2 CARRIER position. 281.Evacuation: On the ALUMINUM control panel turn the ACTIVE switch and the BUBBLER IN switch on. Insert a key and turn the PURGE BUBBLER switch clockwise. The red PURGE BUBBLER light should be on. Push in the black DVM SELECT button and adjust the FLOW SET potentiometer for ALMFC1 to a flow of 100 sccm. As the gas in the line is evacuated, the flow rate will drop.
33
282.Backfill: When the flow rate for ALMFC1 reaches zero as indicated on the panel meter, the circuit is completely evacuated. Turn the switch for the BUBBLER PURGE kit to the ON position for 10 seconds, then turn it off. This will allow the bubbler circuit to fill with nitrogen. 283.Repeat steps 6 and 7 twenty times to completely purge the circuit of alkyl vapor. Finish with a backfill step. 284.Turn the BUBBLER IN switch and the ACTIVE switch to the off position. 285.Disconnect and remove the bubbler. CAUTION: This is a potentially hazardous step if the manual bubbler valves have failed to seal. The pyrophoric liquid will ignite when exposed to air. The quantities are usually small enough to easily contain the burning alkyl and the oxidation is limited by the diffusion rate of oxygen into the tubing, which is rather slow. Avoid breathing the oxide fumes. Have a dry chemical or non-flammable dry absorbent medium available to smother the chemical fire should it be necessary. In most cases probably the best remedy is to attach the stainless steel VCR caps to seal the inlet and outlet of the bubbler. DO NOT USE WATER!! 286.Install the stainless steel VCR caps on the bubbler ports. Use new gaskets unless the gaskets are of the self centering type. 287.If a new bubbler is being installed immediately, proceed to the next section. Otherwise remove the bubbler purge kit, reconnect the pneumatic control line to ALPV12 and shut down the system as appropriate.
Bubbler installation
288.Insure that the tank valve for the nitrogen is on, that the pressure regulator on the tank gives a pressure of 20 psig, and that NMV1 is open. 289.Insure that one of the hydrogen cylinders is actively connected to the hydrogen supply line, that the pressure regulator is properly set and that the hydrogen purifier is turned on and up to temperature. 290.Insure that PR2 is at its normal setting of 10 psig. 291.Connect the solenoid valve from the BUBBLER PURGE kit to the pneumatic control inlet of ALPV12. The connector of the 'A' port of the solenoid valve should be attached to the pneumatic inlet of ALPV12. The connector on the 'P' port of the solenoid valve should be attached to the 1/8" 0.D. nylon tubing leading to ALPV12. Insure that the switch on the control box for the BUBBLER PURGE kit is closed and plug its line cord into a 120 VAC outlet. 292.With the vacuum pump running, open the VAC EXHAUST switch on the VACUUM CONTROL panel. Make sure that the CONTROL VALVE OPEN light on the VACUUM CONTROL panel is on. If not press the OPEN button on the AdaptTorr ACR-26 pressure controller to fully open the pressure control valve. Make sure the MANIFOLD H2 switch on the HYDROGEN/NITROGEN control panel is off. Turn the MANIFOLD N2 switch on the HYDROGEN/NITROGEN control panel to the on position and adjust the FLOW SET 1 potentiometer to maintain a minimum deposition chamber pressure of 1-2 Torr as indicated on the AdaptTorr pressure controller. 293.On the HYDROGEN/NITROGEN panel set the N2 CARRIER/H2 CARRIER switch to the H2 CARRIER position. 294.Insure that the manual valves on the bubbler are closed. Uncap the inlet and outlet ports on the bubbler and connect the bubbler to the circuit using new gaskets if the previously used gaskets were not of the selfcentering type. 295.Evacuation: On the ALUMINUM control panel turn the ACTIVE switch and the BUBBLER IN switch on. Insert a key and turn the PURGE BUBBLER switch clockwise. The red PURGE BUBBLER light should be on. Push in the black DVM SELECT button and adjust the FLOW SET potentiometer for ALMFC1 to a flow of 100 sccm. As the gas in the line is evacuated, the flow rate will drop. 296.Backfill: When the flow rate for ALMFC1 reaches zero as indicated on the panel meter, the circuit is completely evacuated. Turn the switch for the BUBBLER PURGE kit to the ON position for 10 seconds, then turn it off. This will allow the bubbler circuit to fill with hydrogen. 297.Repeat steps 6 and 7 twenty times to completely purge the circuit of air. Finish with a backfill step.
34
298.Open the manual valve on the outlet of the bubbler first. If the pressure in the bubbler should happen to be greater than 10 psig (the setting of PR2) and the inlet valve were opened first, the liquid alkyl would be forced back into the inlet plumbing. 299.Open the manual valve on the inlet of the bubbler. 300.Turn the BUBBLER IN switch and the ACTIVE switch to the off position. Turn the PURGE key lock switch off (counter-clockwise) and remove the key. 301.Remove the BUBBLER PURGE kit and reconnect the 1/8" 0.D. nylon supply line to ALPV12. 302.Shut down the system as appropriate.
35
Ethyldimethylindium (EDMln)
Gas Distribution
303.Ethyldimethylindium is a liquid at room temperature with a vapor pressure of about 1 Torr. This means that the pressure of the vapor is 1/760 x 14.7 psia; very much less than atmospheric pressure. The vapor is transported from the stainless steel canister which contains the liquid, into the reaction chamber by bubbling hydrogen gas through the liquid. 304.The pressure of the hydrogen gas inside the liquid container is determined by the pressure setting of PR2. This is normally set at 10 psig. 305.The precise vapor pressure of the EDMln is determined by the temperature of the liquid. This is controlled by immersing the liquid in a coolant bath which is controlled at a precise temperature by a small Brinkmann refrigeration unit. 306.Pressure drop to the reaction chamber occurs across the mass flow controller, INMFC1. 307.The alkyl is exposed to the heated sensor element of the mass flow controller (110 C); thus it must be stable at this temperature. 308.Loss of liquid from the bubbler into the gas lines is unlikely because pressure in the gas line is the same as the bubbler. (Closure of the bypass valve. INPV2, is delayed to allow bubbler inlet and outlet pressure to equilibrate.) 309.Upon power failure or other abort conditions, contamination of the upstream H2 lines is prevented by closure of INPV12. 310.Adiabatic cooling could possibly cause alkyl condensation in the mass flow controller valve. It is advisable to keep the alkyl temperature below room temperature. 311.A purge mode overrides some interlocks to allow Evacuation and backfill of the bubbler lines.
Hazards
312.EDMln is pyrophoric. 313.Indium oxide (from burning EDMln) is an irritant. 314.EDMln reacts violently with water which should never be used to extinguish the burning alkyl. Use a dry chemical fire extinguisher only or else let it burn.
Precautions
315.Keep the manual valves on a newly installed bubbler closed until it is connected to the system, leak checked and purged. 316.Avoid gas pressure differentials that can force the alkyl from the bubbler back into the carrier gas piping.
Normal Operation
317.Start-of-day: Since the pneumatic valves INPV1 and INPV2 seal off the bubbler from the rest of the system, the manual valves on the bubbler inlet and outlet can be left open except when the bubbler is being replaced or removed. At the start-of-day the EDMln is therefore ready to operate from the control panel. 318.During a normal growth sequence the switches on the front panel are turned on in a left to right sequence and turned off in a right to left sequence. With all switches closed, INMFC1 is commanded closed, INPV1, INPV2, INPV3, and INPV4 are closed and INPV5 is in the vent position. When the ACTIVE switch is turned on, the signal to command closure of INMFC1 is removed, INPV2, INPV4 and INPV12 open and the green ACTIVE light comes
36
on. At this point hydrogen carrier gas will flow through the bypass valve INPV2 and into the vent manifold at a rate determined by the current value of the voltage input to INMFC1 as determined either by the front panel potentiometer or the D/A converter output for this flow controller. The red PURGE BUBBLER light on the control panel willbe on indicating that the bypass valve INPV2 is open. When the BUBBLER IN switch is turned on, INPV1 and INPV3 open immediately and the green light BUBBLER IN turns on. After a two-second delay the red PURGE BUBBLER light goes out and INPV2 closes. This insures that the downstream pressure is not greater than the upstream pressure (a condition which would force the liquid alkyl back into the hydrogen carrier gas piping) when INPV1 and INPV3 open. When INPV2 closes, the hydrogen carrier gas is no longer allowed to bypass the bubbler, and instead must flow through the bubbler transporting the alkyl vapor to the vent/run valve. When the RUN switch is turned on, the yellow RUN light comes on and the gas flow is switched from the vent manifold to the run manifold and must now pass through the deposition chamber. During manual operation from the front panel, the flow through INMFC1 is controlled by the setting of the 10-turn DVM SELECT potentiometer on the control panel. The value of the flow setting can be read on the digital panel meter by pushing in on the black DVM SELECT button. When the button is released, the meter reads the value of the actual flow. When the gas circuit is operated remotely from the computer, the input voltage to INMFC1 is supplied by a DIA converter in the HP Multiprogrammer and can be remotely programmed. The meter on the panel will continue to indicate the actual flow, and when the DVM SELECT button is pushed in, the meter will read the input voltage to the flow controller from the D/A converter. Shutdown of the system should proceed by reversing the sequence of the above steps. 319.The BUBBLER IN switch should be turned off during standby periods to both conserve the source and to minimize the load of hazardous chemical which enters the scrubber. The alkyl vapor will continue to be transported from the bubbler as long as the BUBBLER IN switch is on and there is gas flow through INMFC1. 320.End of day: The alkyl vapors will accelerate the deterioration of some of the elastomers from which the o-rings are fabricated. Elastomer seals have been eliminated whenever possible in the construction of this system but current technology does not allow them to be completely eliminated. They are present in the mass flow controller and in the vent/run valve. For this reason it is good practice to thoroughly flush the alkyl vapor from INMFC1 and INPV5 before the circuit is shut down. This can be accomplished by allowing hydrogen to flow through the system for approximately 5 minutes after the BUBBLER IN switch is closed. After this, the ACTIVE switch can be closed, sealing off the circuit.
Interconnections
321.ACTIVE controls INSV1 which drives INPV4 and INPV12. 322.BUBBLER IN controls INSV2 which drives INPV1 and INPV3. 323.Relays INK1 and INK2 controilNSV3 which drives INPV2.
Interlocks
324.All source valves are deactivated if the following conditions exist: 325.System is in abort status. 326.The vacuum exhaust valve, EXPV2, is closed. 327.The deposition chamber is up to atmosphere. 328.The manifold diluent gas is off. 329.INPV1 and INPV3 cannot be opened unless INPV4 and INPV12 are open. 330.With the PURGE switch on the front panel in the locked position (key removed), INPV2 cannot be opened if INPV1 and INPV3are open.
37
331.The bypass valve (INPV2) is open if the BUBBLER IN switch is closed (handle down) and the ACTIVE switch is on (handle up). 332.The ACTIVE, BUBBLER IN and RUN switches on the front panel are disabled when the computer switches the circuit into the REMOTE mode. 333.When the bubbler is connected to the circuit by opening the BUBBLER IN switch (handle up), valves INPV1 and INPV3 open immediately. Valve INPV2 closes after a 2 second time delay to insure that the pressure on each side of the bubbler is equalized. 334.The manifold diluent gas cannot be turned off if the ACTIVE switch is on.
Bubbler removal
335.Insure that the tank valve for the nitrogen is on, that the pressure regulator on the tank gives a pressure of 20 psig, and that NMV1 is open. 336. Connect the solenoid valve from the BUBBLER PURGE kit to the pneumatic control inlet of INPV12. The connector on the 'A' port of the solenoid valve should be attached to the pneumatic inlet of INPV12. The connector on the 'P' port of the solenoid valve should be attached to the 1/8" O.D. nylon tubing leading to INPV12. Insure that the switch on the control box for the BUBBLER PURGE kit is closed and plug its line cord into a 120VAC outlet. 337.Close the manual valves on the inlet and outlet of the EDMln bubbler. 338.With the vacuum pump running, open the VAC EXHAUST switch on the VACUUM CONTROL panel. Make sure that the CONTROL VALVE OPEN light on the VACUUM CONTROL panel is on. If not press the OPEN button on the AdaptTorr ACR-26 pressure controller to fully open the pressure control valve. Make sure the MANIFOLD H2 switch on the HYDROGEN/NITROGEN control panel is off. Turn the MANIFOLD N2 switch on the HYDROGEN/NITROGEN control panel to the on position and adjust the FLOW SET 1 potentiometer to maintain a minimum deposition chamber pressure of 1-2 Torr as indicated on the AdaptTorr pressure controller. 339.On the HYDROGEN/NITROGEN panel set the N2 CARRIERlH2 CARRIER switch to the N2 CARRIER position. 340.Evacuation: On the INDIUM control panel turn the ACTIVE switch and the BUBBLER IN switch on. Insert a key and turn the PURGE BUBBLER switch clockwise. The red PURGE BUBBLER light should be on. Push in the black DVM SELECT button and adjust the FLOW SET potentiometer for INMFC1 to a flow of 100 sccm. As the gas in the line is evacuated, the flow rate will drop. 341.Backfill: When the flow rate for INMFC1 reaches zero as indicated on the panel meter, the circuit is completely evacuated. Turn the switch for the BUBBLER PURGE kit to the ON position for 10 seconds, then turn it off. This will allow the bubbler circuit to fill with nitrogen.
38
342.Repeat steps 6 and 7 twenty times to completely purge the circuit of alkyl vapor. Finish with a backfill step. 343.Turn the BUBBLER IN switch and the ACTIVE switch to the off position. 344.Disconnect and remove the bubbler. CAUTION: This is a potentially hazardous step if the manual bubbler valves have failed to seal. The pyrophoric liquid will ignite when exposed to air. The quantities are usually small enough to easily contain the burning alkyl and the oxidation is limited by the diffusion rate of oxygen into the tubing, which is rather slow. Avoid breathing the oxide fumes. Have a dry chemical or non-flammable dry absorbent medium available to smother the chemical fire should it be necessary. In most cases probably the best remedy is to attach the stainless steel VCR caps to seal the inlet and outlet of the bubbler. DO NOT USE WATER!! 345.Install the stainless steel VCR caps on the bubbler ports. Use new gaskets unless the gaskets are of the self centering type. 346.If a new bubbler is being installed immediately, proceed to the next section. Otherwise remove the bubbler purge kit, reconnect the pneumatic control line to INPV12 and shut down the system as appropriate.
Bubbler installation
347.Insure that the tank valve for the nitrogen is on, that the pressure regulator on the tank gives a pressure of 20 psig, and that NMV1 is open. 348.Insure that one of the hydrogen cylinders is actively connected to the hydrogen supply line, that the pressure regulator is properly set and that the hydrogen purifier is turned on and up to temperature. 349.Insure that PR2 is at its normal setting of 10 psig. 350.Connect the solenoid valve from the BUBBLER PURGE kit to the pneumatic control inlet of INPV12. The connector of the 'A' port of the solenoid valve should be attached to the pneumatic inlet of INPV12. The connector on the 'P' port of the solenoid valve should be attached to the 1/8" 0.D. nylon tubing leading to INPV12. Insure that the switch on the control box for the BUBBLER PURGE kit is closed and plug its line cord into a 120VAC outlet. 351.With the vacuum pump running, open the VAC EXHAUST switch on the VACUUM CONTROL panel. Make sure that the CONTROL VALVE OPEN light on the VACUUM CONTROL panel is on. If not press the OPEN button on the AdaptTorr ACR-26 pressure controller to fully open the pressure control valve. Make sure the MANIFOLD H2 switch on the HYDROGEN/NITROGEN control panel is off. Turn the MANIFOLD N2 switch on the HYDROGEN/NITROGEN control panel to the on position and adjust the FLOW SET 1 potentiometer to maintain a minimum deposition chamber pressure of 1-2 Torr as indicated on the AdaptTorr pressure controller. 352.On the HYDROGEN/NITROGEN panel set the N2 CARRIER/H2 CARRIER switch to the H2 CARRIER position. 353.Insure that the manual valves on the bubbler are closed. Uncap the inlet and outlet ports on the bubbler and connect the bubbler to the circuit using new gaskets if the previously used gaskets were not of the selfcentering type. 354.Evacuation: On the INDIUM control panel turn the ACTIVE switch and the BUBBLER IN switch on. Insert a key and turn the PURGE BUBBLER switch clockwise. The red PURGE BUBBLER light should be on. Push in the black DVM SELECT button and adjust the FLOW SET potentiometer for INMFC1 to a flow of 100 sccm. As the gas in the line is evacuated, the flow rate will drop. 355.Backfill: When the flow rate for INMFC1 reaches zero as indicated on the panel meter, the circuit is completely evacuated. Turn the switch for the BUBBLER PURGE kit to the ON position for 10 seconds, then turn it off. This will allow the bubbler circuit to fill with hydrogen. 356.Repeat steps 6 and 7 twenty times to completely purge the circuit of air. Finish with a backfill step. 357.Open the manual valve on the outlet of the bubbler first. If the pressure in the bubbler should happen to be greater than 10 psig (the setting of PR2) and the inlet valve were opened first, the liquid alkyl would be forced back into the inlet plumbing.
39
358.Open the manual valve on the inlet of the bubbler. 359.Turn the BUBBLER IN switch and the ACTIVE switch to the off position. Turn the PURGE key lock switch off (counter-clockwise) and remove the key. 360.Remove the BUBBLER PURGE kit and reconnect the 1/8" 0.0. nylon supply line to INPV12. 361.Shut down the system as appropriate.
40
Tertiarybutylarsine (TBAs)
Gas Distribution
362.Tertiarybutylarsine is a liquid at room temperature with a vapor pressure of 81 Torr at 10 C. This means that the pressure of the vapor is 81/760 x 14.7 psia; slightly more than 1/10 of atmospheric pressure. The vapor is transported from the stainless steel canister which contains the liquid, into the reaction chamber by bubbling hydrogen gas through the liquid. 363.The pressure of the hydrogen gas inside the liquid container is determined by the pressure setting of PR5. This is normally set at 10 psig. 364.The precise vapor pressure of the TBAs is determined by the temperature of the liquid. This is controlled by immersing the liquid in a coolant bath which is controlled at a precise temperature by a small Brinkmann refrigeration unit. 365.Pressure drop to the reaction chamber occurs across the mass flow controller, ASLMFC1. 366.The alkyl is exposed to the heated sensor element of the mass flow controller (110C); thus it must be stable at this temperature. 367.Loss of liquid from the bubbler into the gas lines is unlikely because pressure in the gas line is the same as the bubbler. (Closure of the bypass valve, ASLPV2, is delayed to allow bubbler inlet and outlet pressure to equilibrate.) 368.Upon power failure or other abort conditions, contamination of the upstream H2 lines is prevented by closure of ASLPV12. 369.Adiabatic cooling could possibly cause alkyl condensation in the mass flow controller valve. It is advisable to keep the alkyl temperature below room temperature. 370.A purge mode overrides some interlocks to allow evacuation and backfill of the bubbler lines
Hazards
371.TBAs is a colorless oil which is pyrophoric. In case of skin contact, immediately flush skin with plenty of water for at least 15 minutes while removing contaminated clothing and shoes. 372.Combustion may produce carbon monoxide and arsenic oxides. 373.The vapor has a very unpleasant odor suggesting rotten onions. The limits of olfactory detection are unknown. The OSHA TWA Ceiling for the vapor is 0.5 mg/m3 (0.1 ppm). The DOD ChemLogic CL8 toxic gas monitor which is set to give an alarm at the lower TWA Ceiling of 0.05 ppm for arsine will detect TBAs at the above ceiling. In the event of a TBAs spill, all personnel not equipped with respirators should immediately leave the vicinity of the spill. 374.The most effective fire extinguishing agent is a dry chemical powder pressurized with nitrogen. Sand, absorbent clay or CO2 may be used. If the fire cannot be controlled with an extinguishing agent, keep a safe distance and allow the material to burn until extinguished.
Precautions
375.Keep the manual valves on a newly installed bubbler closed until it is connected to the system, leak checked and purged. 376.Avoid gas pressure differentials that can force the alkyl from the bubbler back into the carrier gas piping.
Normal Operation
41
377.Start-of-day: Since the pneumatic valves ASLPV1 and ASLPV2 seal off the bubbler from the rest of the system, the manual valves on the bubbler inlet and outlet can be left open except when the bubbler is being replaced or removed. At the start-of-day the TBAs is therefore ready to operate from the control panel. 378.During a normal growth sequence the switches on the front panel are turned on in a left to right sequence and turned off in a right to left sequence. With all switches closed, ASLMFC1 is commanded closed, ASLPV1, ASLPV2, ASLPV3, and ASLPV4 are closed and ASLPV5 is in the vent position. When the ACTIVE switch is turned on, the signal to command closure of ASLMFC1 is removed, ASLPV2, ASLPV4 and ASLPV12 open and the green ACTIVE light comes on. At this point hydrogen carrier gas will flow through the bypass valve ASLPV2 and into the vent manifold at a rate determined by the current value of the voltage input to ASLMFC1 as determined either by the front panel potentiometer or the D/A converter output for this flow controller. The red PURGE BUBBLER light on the control panel will be on indicating that the bypass valve ASLPV2 is open. When the BUBBLER IN switch is turned on, ASLPV1 and ASLPV3 open immediately and the green light BUBBLER IN turns on. After a two-second delay the red PURGE BUBBLER light goes out and ASLPV2 closes. This insures that the downstream pressure is not greater than the upstream pressure (a condition which would force the liquid alkyl back into the hydrogen carrier gas piping) when ASLPV1 and ASLPV3 open. When ASLPV2 closes, the hydrogen carrier gas is no longer allowed to bypass the bubbler, and instead must flow through the bubbler transporting the alkyl vapor to the venVrunvalve. When the RUN switch is turned on, the yellow RUN light comes on and the gas flow is switched from the vent manifold to the run manifold and must now pass through the deposition chamber. During manual operation from the front panel, the flow through ASLMFC1 is controlled by the setting of the 10-turn DVM SELECT potentiometer on the control panel. The value of the flow setting can be read on the digital panel meter by pushing in on the black DVMSELECT button. When the button is released, the meter reads the value of the actual flow. When the gas circuit is operated remotely from the computer, the input voltage to ASLMFC1 is supplied by a D/A converter in the HP Multiprogrammer and can be remotely programmed. The meter on the panel will continue to indicate the actual flow, and when the DVM SELECT button is pushed in, the meter will read the input voltage to the flow controller from the D/A converter. Shutdown of the system should proceed by reversing the sequence of the above steps. 379.The BUBBLER IN switch should be turned off during standby periods to both conserve the source and to minimize the load of hazardous chemical which enters the scrubber. The alkyl vapor will continue to be transported from the bubbler as long as the BUBBLER IN switch is on and there is gas flow through ASLMFC1. 380.End of day: The alkyl vapors will accelerate the deterioration of some of the elastomers from which the o-rings are fabricated. Elastomer seals have been eliminated whenever possible in the construction of this system but current technology does not allow them to be completely eliminated. They are present in the mass flow controller and in the vent/run valve. For this reason it is good practice to thoroughly flush the alkyl vapor from ASLMFC1 and ASLPV5 before the circuit is shut down. This can be accomplished by allowing hydrogen to flow through the system for approximately 5 minutes after the BUBBLER IN switch is closed. After this, the ACTIVE switch can be closed, sealing off the circuit.
Interconnections
381.ACTIVE controls ASLSV1 which drives ASLPV4 and ASLPV12. 382.BUBBLER IN controls ASLSV2 which drives ASLPV1 and ASLPV3. 383.Relays ASLK1 and ASLK2 control ASLSV3 which drives ASLPV2.
Interlocks
384. All source valves are deactivated if the following conditions exist: 385.System is in abort status.
42
386.The vacuum exhaust valve, EXPV2, is closed. 387.The deposition chamber is up to atmosphere. 388.The manifold diluent gas is off. 389.ASLPV1 and ASLPV3 cannot be opened unless ASLPV4 and ASLPV12 are open. 390.With the PURGE switch on the front panel in the locked position (key removed), ASLPV2 cannot be opened if ASLPV1 and ASLPV3 are open. 391.The bypass valve (ASLPV2) is open if the BUBBLER IN switch is closed (handle down) and the ACTIVE switch is on (handle up). 392.The ACTIVE, BUBBLER IN, and RUN switches on the front panel are disabled when the computer switches the circuit into the REMOTE mode. 393.When the bubbler is connected to the circuit by opening the BUBBLER IN switch (handle up), valves ASLPV1 and ASLPV3 open immediately. Valve ASLPV2 closes after a 2 second time delay to insure that the pressure on each side of the bubbler is equalized. 394.The manifold diluent gas cannot be turned off if the ACTIVE switch is on.
Bubbler removal
395.Insure that the tank valve for the nitrogen is on, that the pressure regulator on the tank gives a pressure of 20 psig, and that NMV1 is open. 396.Connect the solenoid valve from the BUBBLER PURGE kit to the pneumatic control inlet of ASLPV12. The connector on the 'A' port of the solenoid valve should be attached to the pneumatic inlet of ASLPV12. The connector on the 'P' port of the solenoid valve should be attached to the 1/8" 0.D. nylon tubing leading to ASLPV12. Insure that the switch on the control box for the BUBBLER PURGE kit is closed and plug its line cord into a 120 VAC outlet. 397.Close the manual valves on the inlet and outlet of the TBAs bubbler. 398.With the vacuum pump running, open the VAC EXHAUST switch on the VACUUM CONTROL panel. Make sure that the CONTROL VALVE OPEN light on the VACUUM CONTROL panel is on. If not press the OPEN button on the AdaptTorr ACR-26 pressure controller to fully open the pressure control valve. Make sure the MANIFOLD H2 switch on the HYDROGEN/NITROGEN control panel is off. Turn the MANIFOLD N2 switch on the HYDROGEN/NITROGEN control panel to the on position and adjust the FLOW SET 1 potentiometer to maintain a minimum deposition chamber pressure of 1-2 Torr as indicated on the AdaptTorr pressure controller. 399.On the HYDROGEN/NITROGEN panel set the N2 CARRIER/H2 CARRIER switch to the N2 CARRIER position.
43
400.Evacuation: On the ARSENIC LIQUID control panel turn the ACTIVE switch and the BUBBLER IN switch on. Insert a key and turn the PURGE BUBBLER switch clockwise. The red PURGE BUBBLER light should be on. Push in the black DVM SELECT button and adjust the FLOW SET potentiometer for ASLMFC1 to a flow of 100 sccm. As the gas in the line is evacuated, the flow rate will drop. 401.Backfill: When the flow rate for ASLMFC1 reaches zero as indicated on the panel meter, the circuit is completely evacuated. Turn the switch for the BUBBLER PURGE kit to the ON position for 10 seconds, then turn it off. This will allow the bubbler circuit to fill with nitrogen. 402.Repeat steps 6 and 7 twenty times to completely purge the circuit of alkyl vapor. Finish with a backfill step. 403.Turn the BUBBLER IN switch and the ACTIVE switch to the off position. 404.Disconnect and remove the bubbler. CAUTION: This is a potentially hazardous step if the manual bubbler valves have failed to seal. The pyrophoric liquid will ignite when exposed to air. The quantities are usually small enough to easily contain the burning alkyl and the oxidation is limited by the diffusion rate of oxygen into the tubing, which is rather slow. Avoid breathing the oxide fumes. Have a dry chemical or non-flammable dry absorbent medium available to smother the chemical fire should it be necessary. In most cases probably the best remedy is to attach the stainless steel VCR caps to seal the inlet and outlet of the bubbler. DO NOT USE WATER!! 405.Install the stainless steel VCR caps on the bubbler ports. Use new gaskets unless the gaskets are of the self centering type. 406. If a new bubbler is being installed immediately, proceed to the next section. Otherwise remove the bubbler purge kit, reconnect the pneumatic control line to ASLPV12and shut down the system as appropriate.
Bubbler installation
407.Insure that the tank valve for the nitrogen is on, that the pressure regulator on the tank gives a pressure of 20 psig, and that NMV1 is open. 408.Insure that one of the hydrogen cylinders is actively connected to the hydrogen supply line, that the pressure regulator is properly set and that the hydrogen purifier is turned on and up to temperature. 409.Insure that PR5 is at its normal setting of 10 psig. 410.Connect the solenoid valve from the BUBBLER PURGE kit to the pneumatic control inlet of ASLPV12. The connector of the 'A' port of the solenoid valve should be attached to the pneumatic inlet of ASLPV12. The connector on the 'P' port of the solenoid valve should be attached to the 1/8" 0.D. nylon tubing leading to ASLPV12. Insure that the switch on the control box for the BUBBLER PURGE kit is closed and plug its line cord into a 120 VAC outlet. 411.With the vacuum pump running, open the VAC EXHAUST switch on the VACUUM CONTROL panel. Make sure that the CONTROL VALVE OPEN light on the VACUUM CONTROL panel is on. If not press the OPEN button on the AdaptTorr ACR-26 pressure controller to fully open the pressure control valve. Make sure the MANIFOLD H2 switch on the HYDROGEN/NITROGEN control panel is off. Turn the MANIFOLD N2 switch on the HYDROGEN/NITROGEN control panel to the on position and adjust the FLOW SET 1 potentiometer to maintain a minimum deposition chamber pressure of 1-2 Torr as indicated on the AdaptTorr pressure controller. 412.On the HYDROGEN/NITROGEN panel set the N2 CARRIER/H2 CARRIER switch to the H2 CARRIER position. 413.Insure that the manual valves on the bubbler are closed. Uncap the inlet and outlet ports on the bubbler and connect the bubbler to the circuit using new gaskets if the previously used gaskets were not of the selfcentering type. 414.Evacuation: On the ARSENIC LIQUID control panel turn the ACTIVE switch and the BUBBLER IN switch on. Insert a key and turn the PURGE BUBBLER switch clockwise. The red PURGE BUBBLER light should be on. Push in the black DVM SELECT button and adjust the FLOW SET potentiometer for ASLMFC1 to a flow of 100 sccm. As the gas in the line is evacuated, the flow rate will drop.
44
415.Backfill: When the flow rate for ASLMFC1 reaches zero as indicated on the panel meter, the circuit is completely evacuated. Turn the switch for the BUBBLER PURGE kit to the ON position for 10 seconds, then turn it off. This will allow the bubbler circuit to fill with hydrogen. 416.Repeat steps 6 and 7 twenty times to completely purge the circuit of air. Finish with a backfill step. 417.Open the manual valve on the outlet of the bubbler first. If the pressure in the bubbler should happen to be greater than 10 psig (the setting of PR5) and the inlet valve were opened first, the liquid alkyl would be forced back into the inlet plumbing. 418.Open the manual valve on the inlet of the bubbler. 419.Turn the BUBBLER IN switch and the ACTIVE switch to the off position. Turn the PURGE key lock switch off (counter-clockwise) and remove the key. 420.Remove the BUBBLER PURGE kit and reconnect the 1/8" 0.D. nylon supply line to ASLPV12. 421.Shut down the system as appropriate.
45
Tertiarybutylphosphine (TBP)
Gas Distribution
1. Tertiarybutylphosphine is a liquid at room temperature with a vapor pressure of 286 Torr at room temperature. This means that the pressure of the vapor is 286/760 x 14.7 psia; or about 1/3 atmospheric pressure. The vapor is transported from the stainless steel canister which contains the liquid, into the reaction chamber by bubbling hydrogen gas through the liquid. The pressure of the hydrogen gas inside the liquid container is determined by the pressure setting of PRS. This is normally set at 10 psig. The precise vapor pressure of the TBP is determined by the temperature of the liquid. This is controlled by immersing the liquid in a coolant bath which is controlled at a precise temperature by a small Brinkmann refrigeration unit. Pressure drop to the reaction chamber occurs across the mass flow controller, PLMFC1. The alkyl is exposed to the heated sensor element of the mass flow controller (110 C); thus it must be stable at this temperature. Loss of liquid from the bubbler into the gas lines is unlikely because pressure in the gas line is the same as the bubbler. (Closure of the bypass valve, PLPV2, is delayed to allow bubbler inlet and outlet pressure to equilibrate.) Upon power failure or other abort conditions, contamination of the upstream H2 lines is prevented by closure of PLPV12. Adiabatic cooling could possibly cause alkyl condensation in the mass flow controller valve. It is advisable to keep the alkyl temperature below room temperature. A purge mode overrides some interlocks to allow evacuation and backfill of the bubbler lines
2. 3.
4. 5. 6.
7. 8. 9.
Hazards
1. 2. TBP is a colorless oil which is pyrophoric. In case of skin contact, immediately flush skin with plenty of water for at least 1Sminutes while removing contaminated clothing and shoes. Combustion may produce carbon monoxide, phosphine oxides, phosphinic acid, phosphonic acid, and P205. The products formed by the thermal decomposition in an anaerobic atmosphere are elemental phosphorus, isobutane, isobutylene, and phosphine. The vapor has a very unpleasant odor. The limits of olfactory detection are unknown. Animal studies have indicated that the toxicity is about 20 times less than for phosphine. The OSHA TWA (time-weighted average for an 8 hour shift) ceiling for phosphine is 0.3 ppm. The DOD ChemLogic CL8 toxic gas monitor which is set to give an alarm at the lower TWA ceiling of O.O5ppm for arsine will detect TBP at the above ceiling. In the event of a TBP spill, all personnel not equipped with respirators should immediately leave the vicinity of the spill. The most effective fire extinguishing agent is a dry chemical powder pressurized with nitrogen. Sand, absorbent clay or CO2 may be used. If the fire cannot be controlled with an extinguishing agent, keep a safe distance and allow the material to burn until extinguished.
3.
4.
Precautions
1. 2. Keep the manual valves on a newly installed bubbler closed until it is connected to the system, leak checked and purged. Avoid gas pressure differentials that can force the alkyl from the bubbler back into the carrier gas piping.
46
Normal Operation
3. Start-of-day: Since the pneumatic valves PLPV1 and PLPV2 seal off the bubbler from the rest of the system, the manual valves on the bubbler inlet and outlet can be left open except when the bubbler is being replaced or removed. At the start-of-day the TBP is therefore ready to operate from the control panel. During a normal growth sequence the switches on the front panel are turned on in a left to right sequence and turned off in a right to left sequence. With all switches closed, PLMFC1 is commanded closed, PLPV1, PLPV2, PLPV3, and PLPV4 are closed and PLPV5 is in the vent position. When the ACTIVE switch is turned on, the signal to command closure of PLMFC1 is removed, PLPV2, PLPV4 and PLPV12 open and the green ACTIVE light comes on. At this point hydrogen carrier gas will flow through the bypass valve PLPV2 and into the vent manifold at a rate determined by the current value of the voltage input to PLMFC1 as determined either by the front panel potentiometer or the D/A converter output for this flow controller. The red PURGE BUBBLER light on the control panel will be on indicating that the bypass valve PLPV2 is open. When the BUBBLER IN switch is turned on, PLPV1and PLPV3 open immediately and the green light BUBBLER IN turns on. After a twosecond delay the red PURGE BUBBLER light goes out and PLPV2 closes. This insures that the downstream pressure is not greater than the upstream pressure (a condition which would force the liquid alkyl back into the hydrogen carrier gas piping) when PLPV1and PLPV3 open. When PLPV2closes, the hydrogen carrier gas is no longer allowed to bypass the bubbler, and instead must flow through the bubbler transporting the alkyl vapor to the vent/run valve. When the RUN switch is turned on, the yellow RUN light comes on and the gas flow is switched from the vent manifold to the run manifold and must now pass through the deposition chamber. During manual operation from the front panel, the flow through PLMFC1 is controlled by the setting of the 10-turn DVM SELECT potentiometer on the control panel. The value of the flow setting can be read on the digital panel meter by pushing in on the black DVM SELECT button. When the button is released, the meter reads the value of the actual flow. When the gas circuit is operated remotely from the computer, the input voltage to PLMFC1 is supplied by a D/A converter in the HP Multiprogrammer and can be remotely programmed. The meter on the panel will continue to indicate the actual flow, and when the DVM SELECT button is pushed in, the meter will read the input voltage to the flow controller from the D/A converter. Shutdown of the system should proceed by reversing the sequence of the above steps. The BUBBLER IN switch should be turned off during standby periods to both conserve the source and to minimize the load of hazardous chemical which enters the scrubber. The alkyl vapor will continue to be transported from the bubbler as long as the BUBBLER IN switch is on and there is gas flow through PLMFC1. End of day: The alkyl vapors will accelerate the deterioration of some of the elastomers from which the o-rings are fabricated. Elastomer seals have been eliminated whenever possible in the construction of this system but current technology does not allow them to be completely eliminated. They are present in the mass flow controller and in the vent/run valve. For this reason it is good practice to thoroughly flush the alkyl vapor from PLMFC1 and PLPV5 before the circuit is shut down. This can be accomplished by allowing hydrogen to flow through the system for approximately 5 minutes after the BUBBLER IN switch is closed. After this, the ACTIVE switch can be closed, sealing off the circuit.
4.
5.
6.
Interconnections
7. 8. 9. ACTIVE controls PLSV1 which drives PLPV4 and PLPV12. BUBBLER IN controls PLSV2 which drives PLPV1and PLPV3. Relays PLK1 and PLK2 control PLSV3 which drives PLPV2.
Interlocks
47 1.
2. 3. 4. 5. 6.
7.
Allsource valves are deactivated if the following conditions exist: a. System is in abort status. b. The vacuum exhaust valve, EXPV2, is closed. c. The deposition chamber is up to atmosphere. d. The manifold diluent gas is off. PLPV1 and PLPV3 can not be opened unless PLPV4 and PLPV12 are open. With the PURGE switch on the front panel in the locked position (key removed), PLPV2 cannot be opened if PLPV1and PLPV3are open. The bypass valve (PLPV2) is open if the BUBBLER IN switch is closed (handle down) and the ACTIVE switch is on (handle up). The ACTIVE, BUBBLER IN, and RUN switches on the front panel are disabled when the computer switches the circuit into the REMOTE mode. When the bubbler is connected to the circuit by opening the BUBBLER IN switch (handle up), valves PLPV1 and PLPV3 open immediately. Valve PLPV2closes after a 2 second time delay to insure that the pressure on each side of the bubbler is equalized. The manifold diluent gas cannot be turned off if the ACTIVE switch is on.
Bubbler removal
1. 2. Insure that the tank valve for the nitrogen is on, that the pressure regulator on the tank gives a pressure of 20 psig, and that NMV1 is open. Connect the solenoid valve from the BUBBLER PURGE kit to the pneumatic control inlet of PLPV12. The connector on the 'A' port of the solenoid valve should be attached to the pneumatic inlet of PLPV12. The connector on the 'P' port of the solenoid valve should be attached to the 1/8" O.D. nylon tubing leading to PLPV12. Insure that the switch on the control box for the BUBBLER PURGE kit is closed and plug its line cord into a 120 VAC outlet. Close the manual valves on the inlet and outlet of the TBP bubbler. With the vacuum pump running, open the VAC EXHAUST switch on the VACUUM CONTROL panel. Make sure that the CONTROL VALVE OPEN light on the VACUUM CONTROL panel is on. If not press the OPEN button on the AdaptTorr ACR-26pressure controller to fully open the pressure control valve. Make sure the MANIFOLD H2 switch on the HYDROGEN/NITROGEN control panel is off. Turn the MANIFOLD N2 switch on the HYDROGEN/NITROGEN control panel to the on position and adjust the FLOW SET 1 potentiometer to maintain a minimum deposition chamber pressure of 1-2 Torr as indicated on the AdaptTorr pressure controller.
3. 4.
48
On the HYDROGEN/NITROGEN panel set the N2 CARRIERlH2 CARRIER switch to the N2 CARRIER position. Evacuation: On the PHOSPHORUS LIQUID control panel turn the ACTIVE switch and the BUBBLER IN switch on. Insert a key and turn the PURGE BUBBLER switch clockwise. The red PURGE BUBBLER light should be on. Push in the black DVM SELECT button and adjust the FLOW SET potentiometer for PLMFC1to a flow of 100 sccm. As the gas in the line is evacuated, the flow rate will drop. 7. Backfill: When the flow rate for PLMFC1 reaches zero as indicated on the panel meter, the circuit is completely evacuated. Turn the switch for the BUBBLER PURGE kit to the ON position for 10 seconds, then turn it off. This will allow the bubbler circuit to fill with nitrogen. 8. Repeat steps 6 and 7 twenty times to completely purge the circuit of alkyl vapor. Finish with a backfill step. 9. Turn the BUBBLER IN switch and the ACTIVE switch to the off position. 10. Disconnect and remove the bubbler. CAUTION: This is a potentially hazardous step if the manual bubbler valves have failed to seal. The pyrophoric liquid will ignite when exposed to air. The quantities are usually small enough to easily contain the burning alkyl and the oxidation is limited by the diffusion rate of oxygen into the tubing, which is rather slow. Avoid breathing the oxide fumes. Have a dry chemical or non-flammable dry absorbent medium available to smother the chemical fire should it be necessary. In most cases probably the best remedy is to attach the stainless steel VCR caps to seal the inlet and outlet of the bubbler. DO NOT USE WATER!! 11. Install the stainless steel VCR caps on the bubbler ports. Use new gaskets unless the gaskets are of the self centering type. 12. If a new bubbler is being installed immediately, proceed to the next section. Otherwise remove the bubbler purge kit, reconnect the pneumatic control line to PLPV12 and shut down the system as appropriate.
Bubbler installation
1. 2. 3. 4. Insure that the tank valve for the nitrogen is on, that the pressure regulator on the tank gives a pressure of 20 psig, and that NMV1 is open. Insure that one of the hydrogen cylinders is actively connected to the hydrogen supply line, that the pressure regulator is properly set and that the hydrogen purifier is turned on and up to temperature. Insure that PR5 is at its normal setting of 10 psig. Connect the solenoid valve from the BUBBLER PURGE kit to the pneumatic control inlet of PLPV12. The connector of the 'A' port of the solenoid valve should be attached to the pneumatic inlet of PLPV12. The connector on the 'P' port of the solenoid valve should be attached to the 1/8" 0.D. nylon tubing leading to PLPV12. Insure that the switch on the control box for the BUBBLER PURGE kit is closed and plug its line cord into a 120VAC outlet. With the vacuum pump running, open the VAC EXHAUST switch on the VACUUM CONTROL panel. Make sure that the CONTROL VALVE OPEN light on the VACUUM CONTROL panel is on. If not press the OPEN button on the AdaptTorr ACR-26 pressure controller to fully open the pressure control valve. Make sure the MANIFOLD H2 switch on the HYDROGEN/NITROGEN control panel is off. Turn the MANIFOLD N2 switch on the HYDROGEN/NITROGEN control panel to the on position and adjust the FLOW SET 1 potentiometer to maintain a minimum deposition chamber pressure of 1-2 Torr as indicated on the AdaptTorr pressure controller. On the HYDROGEN/NITROGEN panel set the N2 CARRIER/H2 CARRIER switch to the H2 CARRIER position. Insure that the manual valves on the bubbler are closed. Uncap the inlet and outlet ports on the bubbler and connect the bubbler to the circuit using new gaskets if the previously used gaskets were not of the selfcentering type. Evacuation: On the PHOSPHORUS LIQUID control panel turn the ACTIVE switch and the BUBBLER IN switch on. Insert a key and turn the PURGE BUBBLER switch clockwise. The red PURGE BUBBLER light should be on. Push
5.
6. 7.
8.
49
9.
10. 11.
in the black DVM SELECT button and adjust the FLOW SET potentiometer for PLMFC1to a flow of 100 sccm. As the gas in the line is evacuated, the flow rate will drop. Backfill: When the flow rate for PLMFC1 reaches zero as indicated on the panel meter, the circuit is completely evacuated. Turn the switch for the BUBBLER PURGE kit to the ON position for 10 seconds, then turn it off. This will allow the bubbler circuit to fill with hydrogen. Repeat steps 6 and 7 twenty times to completely purge the circuit of air. Finish with a backfill step. Open the manual valve on the outlet of the bubbler first. If the pressure in the bubbler should happen to be greater than 10 psig (the setting of PR5) and the inlet valve were opened first, the liquid alkyl would be forced back into the inlet plumbing. Open the manual valve on the inlet of the bubbler. Turn the BUBBLER IN switch and the ACTIVE switch to the off position. Turn the PURGE key lock switch off (counter-clockwise)and remove the key. Remove the BUBBLER PURGE kit and reconnect the 1/8" O.D. nylon supply line to PLPV12. Shut down the system as appropriate.
50
2.
3.
4. 5. 6.
7. 8. 9.
Hazards
1. 2. 3. FE is pyrophoric. The oxides formed from the burning of FE are toxic. FE reacts violently with water which should never be used to extinguish the burning alkyl. Use a dry chemical fire extinguisher only or else let it burn.
Precautions
1. 2. Keep the manual valves on a newly installed bubbler closed until it is connected to the system, leak checked and purged. Avoid gas pressure differentials that can force the alkyl from the bubbler back into the carrier gas piping.
Normal Operation
1. Start-of-day: Since the pneumatic valves FEPV1and FEPV2 seal off the bubbler from the rest of the system, the manual valves on the bubbler inlet and outlet can be left open except when the bubbler is being replaced or
51
2.
removed. At the start-of-day the FE is therefore ready to operate from the control panel. The operation of the controls for the cases without hydrogen dilution and with hydrogen dilution will be discussed separately. Normal growth sequence:
52
hydrogen carrier gas will flow through FEMFC2,through the bypass valve FEPV2, through FEMFC1, through FEPV4, throughFEPV5andintotheventmanifoldata rate determined by the current value of the voltage input to FEMFC1as determined either by the front panel potentiometer or the D/A converter output for this flow controller. The bubbler pressure will be determined by the setting of PR1 and the pressure drop from the bubbler to the chamber will occur across FEMFC1. During manual operation from the front panel, the flow through FEMFC1 is controlled by the setting of the 10-turn FLOW SET 1 potentiometer on the control panel. The value of the flow setting can be read on the digital panel meter by pushing in on the black FLOW SET 1 button. When the button is released, the meter reads the value of the actual flow. When the gas circuit is operated remotely from the computer, the input voltage to FEMFC1 is supplied by a D/A converter in the HP Multiprogrammer and can be remotely programmed. The meter on the panel will continue to indicate the actual flow, and when the FLOW SET 1 button is pushed in, the meter will read the input voltage to the flow controller from the D/A converter. With the switches in this configuration the red PURGE BUBBLER light on the control panel will be on indicating that the bypass valve FEPV2 is open. The H2 DILUTE switch becomes enabled when the ACTIVE switch is on. It may be turned at any time, however, in a normal growth sequence it would be turned on after the ACTIVE switch is turned on and before the BUBBLER IN switch is turned on. When the H2 DILUTE switch is turned on, FEPV6, FEPV7, and FEPV12 open, the positive closure signal on FEMFC3 is removed, and the +10 v control signal which had been keeping FEMFC2fully open is removed. The flow rates through FEMFC1, FEMFC2, and FEMFC3 can now be controlled independently. Hydrogen may be supplied to the circuit through FEMFC2 and/or FEPV12 and may leave through FEMFC1 and/or FEMFC3. If the flow meters are set correctly, then the flow through FEMFC2 is less than the sum of the flows through FEMFC1 and FEMFC3. The pressure in the circuit will then drop until it is less than the pressure at the outlet of PR4. At this point the check valve between FEPV6and FEPV12 will open and the circuit pressure will be maintained at the pressure of PR4. The difference between the exit rate of gas through FEMFC1 and FEMFC3 and the entrance rate through FEMFC2 will be maintained by the flow through the check valve. The equivalent flow through valve FEPV5 is 2 1 1 + 3 There is nothing to prevent the operator from setting the mass flow controllers such that the flow setting for FEMFC2 is greater than the combined flow through FEMFC1 and FEMFC3. In this case the control valve for FEMFC2 would remain fully open and the flow through the circuit would be controlled by the settings on FEMFC1 and FEMFC3. The pressure in the bubbler would remain at the outlet pressure of PR1 and flow through the check valve between FEPV6 and FEPV12 would not occur. Although no damage would occur from operating the system in this manner, there would be no advantage to doing it. In fact, the amount of alkyl flowing through FEMFC3 would be wasted with no improvement in the accuracy in controlling the flow of a small amount of alkyl through FEMFC1. When the BUBBLER IN switch is turned on, FEPV1 and FEPV3 open immediately and the green light, BUBBLER IN, turns on. After a two-second delay the red PURGE BUBBLER light goes out and FEPV2 closes. The delay on closure insures that the downstream pressure is not greater than the upstream pressure (a condition which would force the liquid alkyl back into the hydrogen carrier gas piping) when FEPV1and FEPV3open. When FEPV2 closes, the hydrogen carrier gas is no longer allowed to bypass the bubbler, and instead must flow through the bubbler transporting the alkyl vapor to the vent/run valve. When the RUN switch is turned on, the yellow RUN light comes on and the gas flow is switched from the vent manifold to the run manifold and must now pass through the deposition chamber. Shutdown of the system should proceed by turning off, in sequence, the RUN switch, the H2 DILUTE switch, the BUBBLER IN switch, and the ACTIVE switch. The BUBBLER IN switch should be turned off during standby periods to both conserve the source and to minimize the load of hazardous chemical which enters the scrubber. The alkyl vapor will continue to be
53
transported from the bubbler as long as the BUBBLER IN switch is on and there is gas flow through FEMFC1 and/or FEMFC3. 3. End of day: The alkyl vapors will accelerate the deterioration of some of the elastomers from which the o-rings are fabricated. Elastomer seals have been eliminated whenever possible in the construction of this system but current technology does not allow them to be completely eliminated. They are present in the mass flow controller and in the vent/run valve. For this reason it is good practice to thoroughly flush the alkyl vapor from FEMFC1, FEMFC2, FEMFC3, and FEPV5 before the circuit is shut down. This can be accomplished by allowing hydrogen to flow through FEMFC1, FEMFC2, and FEMFC3for approximately 10 minutes after the BUBBLER IN switch is closed. During the purging step the ACTIVE and H2 PURGE switches must be on and there should be a gas flow of at least 10 sccm through FEMFC1and 100 sccm through FEMFC2 and FEMFC3. After this, the ACTIVE switch can be closed, sealing off the circuit.
Interconnections
1. 2. 3. 4. ACTIVE controls FESV1 which drives FEPV4. BUBBLER IN controls FESV2 which drives FEPV1 and FEPV3. Relays FEK1 and FEK2 control FESV3which drives FEPV2. H2 DILUTE controls FESV4 which drives FEPV6, FEPV7, and FEPV12.
Interlocks
All source valves are deactivated if the following conditions exist: a. System is in abort status. b. The vacuum exhaust valve, EXPV2, is closed. c. The deposition chamber is up to atmosphere. d. The manifold diluent gas is off. 2. All switches are disabled if the ACTIVE switch is off. 3. FEPV1 and FEPV3 can not be opened unless FEPV4 is open. 4. With the PURGE switch on the front panel in the locked position (key removed), FEPV2 cannot be opened if FEPV1 and FEPV3 are open. 5. The bypass valve (FEPV2) is open if the BUBBLER IN switch is closed (handle down) and the ACTIVE switch is on (handle up). 6. The ACTIVE, BUBBLER IN, H2 DILUTE, and RUN switches on the front panel are disabled when the computer switches the circuit into the REMOTE mode. 7. When the bubbler is connected to the circuit by opening the BUBBLER IN switch (handle up), valves FEPV1 and FEPV3 open immediately. Valve FEPV2closes after a 2 second time delay to insure that the pressure on each side of the bubbler is equalized. 8. FEMFC1, FEMFC2, and FEMFC3 are commanded closed if the ACTIVE switch is off. The key lock switch CLOSE MFC2 when turned clockwise will close FEMFC2 under all conditions. 9. When the H2 DILUTE switch is off, FEMFC3 is commanded closed and the set point of FEMFC2 is at 200 seem. The purpose of setting the FEMFC2flow to 200 sccm is to insure that the flow setting for FEMFC2 is always greater that the flow setting for FEMFC1 when H2 DILUTE is off. This prevents the bubbler from becoming evacuated. When the H2 DILUTE switch in on, the closure signal is removed from FEMFC3 and the set point voltage for FEMFC2 is determined by either front panel potentiometer or remote D/A converter. A closure signal on a mass flow controller will over-ride the set point signal. 10. The manifold diluent gas cannot be turned off if the ACTIVE switch is on. 1.
54
Bubbler removal
1. Insure that the tank valve for the nitrogen is on, that the pressure regulator on the tank gives a pressure of 20 psig, and that NMV1 is open. 2. Close the manual valves on the inlet and outlet of the FE bubbler. 3. With the vacuum pump running, open the VAC EXHAUST switch on the VACUUM CONTROL panel. Make sure that the CONTROL VALVE OPEN light on the VACUUM CONTROL panel is on. If not press the OPEN button on the AdaptTorr ACR-26 pressure controller to fully open the pressure control valve. Make sure the MANIFOLD H2 switch on the HYDROGEN/NITROGEN control panel is off. Turn the MANIFOLD N2 switch on the HYDROGEN/NITROGEN control panel to the on position and adjust the FLOW SET 1 potentiometer to maintain a minimum deposition chamber pressure of 1-2 Torr as indicated on the AdaptTorr pressure controller. 4. On the HYDROGEN/NITROGEN panel set the N2 CARRIERlH2 CARRIER switch to the N2 CARRIER position. 5. On the IRON control panel turn the ACTIVE switch and the BUBBLER IN switch on. Insert a key and turn the PURGE BUBBLER switch clockwise. The red PURGE BUBBLER light should be on. Push in the black DVM SELECT button and adjust the FLOW SET potentiometer for FEMFC1 to a flow of 100 sccm. 6. Evacuation: Insert a key in the CLOSE MFC2 switch and turn it clockwise. This will shut off the gas flow through FEMFC2. As the gas in the line is evacuated, the flow rate will drop. When the flow rate for FEMFC1 reaches zero as indicated on the panel meter, the circuit is completely evacuated. 7. Backfill: Turn the key in the CLOSE MFC2 switch counter clockwise for 10 seconds. This will allow the bubbler circuit to fill with nitrogen. 8. Repeat steps 6 and 7 twenty times to completely purge the circuit of alkyl vapor. Finish with a backfill step. 9. Turn the BUBBLER IN switch and the ACTIVE switch to the off position. 10. Disconnect and remove the bubbler. CAUTION: This is a potentially hazardous step if the manual bubbler valves have failed to seal. The pyrophoric liquid will ignite when exposed to air. The quantities are usually small enough to easily contain the burning alkyl and the oxidation is limited by the diffusion rate of oxygen into the tubing, which is rather slow. Avoid breathing the oxide fumes. Have a dry chemical or non-flammable dry absorbent medium available to smother the chemical fire should it be necessary. In most cases probably the best remedy is to attach the stainless steel VCR caps to seal the inlet and outlet of the bubbler. DO NOT USE WATER!! 11. Install the stainless steel VCR caps on the bubbler ports. Use new gaskets unless the gaskets are of the self centering type. 12. If a new bubbler is being installed immediately, proceed to the next section. Otherwise, shut down the system as appropriate.
Bubbler installation
55 1. 2. 3. 4.
5. 6.
7.
8.
9. 10. 11.
Insure that the tank valve for the nitrogen is on, that the pressure regulator on the tank gives a pressure of 20 psig, and that NMV1 is open. Insure that one of the hydrogen cylinders is actively connected to the hydrogen supply line, that the pressure regulator is properly set and that the hydrogen purifier is turned on and up to temperature. Insure that PR4 is set at 10 psig and PR1 is set at 20 psig. With the vacuum pump running, open the VAC EXHAUST switch on the VACUUM CONTROL panel. Make sure that the CONTROL VALVE OPEN light on the VACUUM CONTROL panel is on. If not press the OPEN button on the AdaptTorr ACR-26 pressure controller to fully open the pressure control valve. Make sure the MANIFOLD H2 switch on the HYDROGEN/NITROGEN control panel is off. Turn the MANIFOLD N2 switch on the HYDROGEN/NITROGEN control panel to the on position and adjust the FLOW SET 1 potentiometer to maintain a minimum deposition chamber pressure of 1-2 Torr as indicated on the AdaptTorr pressure controller. On the HYDROGEN/NITROGEN panel set the N2 CARRIER/H2 CARRIER switch to the H2 CARRIER position. Insure that the manual valves on the bubbler are closed. Uncap the inlet and outlet ports on the bubbler and connect the bubbler to the circuit using new gaskets if the previously used gaskets were not of the selfcentering type. On the IRON control panel turn the ACTIVE switch and the BUBBLER IN switch on. Insert a key and turn the PURGE BUBBLER switch clockwise. The red PURGE BUBBLER light should be on. Push in the black DVM SELECT button and adjust the FLOW SET potentiometer for FEMFC1 to a flow of 100 sccm. Evacuation: Insert a key in the CLOSE MFC2 switch and turn it clockwise. This will shut off the gas flow through FEMFC2. As the gas in the line is evacuated, the flow rate will drop. When the flow rate for FEMFC1 reaches zero as indicated on the panel meter, the circuit is completely evacuated. Backfill: Turn the key in the CLOSE MFC2 switch counter clockwise for 10 seconds. This will allow the bubbler circuit to fill with nitrogen. Repeat steps 8 and 9 twenty times to completely purge the circuit of air. Finish with a backfill step. Remove the key to the CLOSE MFC2 switch. Open the manual valve on the outlet of the bubbler first. If the pressure in the bubbler should happen to be greater than 10 psig (the setting of PR4) and the inlet valve were opened first, the liquid alkyl would be forced back into the inlet plumbing. Open the manual valve on the inlet of the bubbler. Turn the BUBBLER IN switch and the ACTIVE switch to the off position. Turn the PURGE key lock switch off (counter-clockwise) and remove the key. Shut down the system as appropriate.
56
Trimethylgallium (TMG)
Gas Distribution
1. Trimethylgallium is a liquid at room temperature with a vapor pressure of about 178 Torr at room temperature. This means that the pressure of the vapor is 178/760 x 14.7 psia; about 1/5 atmospheric pressure. The vapor is transported from the stainless steel canister which contains the liquid, into the reaction chamber by bubbling hydrogen gas through the liquid. 2. The TMG plumbing system is a more complex arrangement which allows the vapor stream passing through the bubbler to be diluted with additional hydrogen carrier gas. In addition, a portion of the total volume of gas may be dumped into the scrubber before it enters the vent/run valve. This arrangement permits a wider range of TMG flow rates to be directed into the deposition chamber. The pressure of the hydrogen gas inside the liquid container is determined by the pressure setting of either PR1 or PR2 depending on the mode of operation. The proper functioning of the various components is discussed in detail in the section on operating 3. procedures. 4. The precise vapor pressure of the TMG is determined by the temperature of the liquid. This is controlled by immersing the liquid in a coolant bath which is controlled at a precise temperature by a small Brinkmann refrigeration unit. 5. Pressure drop to the reaction chamber occurs across the mass flow controller, GAMFC1. 6. The alkyl is exposed to the heated sensor element of the mass flow controller (110C); thus it must be stable at this temperature. 7. Loss of liquid from the bubbler into the gas lines is unlikely because pressure in the gas line is the same as the bubbler. (Closure of the bypass valve, GAPV2, is delayed to allow bubbler inlet and outlet pressure to equilibrate.) 8. Upon power failure or other abort conditions, contamination of the upstream H2 lines is prevented by closure of the control valve for mass flow controller GAMFC2. 9. Adiabatic cooling could possibly cause alkyl condensation in the mass flow controller valve. It is advisable to keep the alkyl temperature below room temperature. 10. A purge mode overrides some interlocks to allow evacuation and backfill of the bubbler lines.
Hazards
1. 2. 3. TMG is pyrophoric. The oxides formed from the burning of TMG are toxic. TMG reacts violently with water which should never be used to extinguish the burning alkyl. Use a dry chemical fire extinguisher only or else let it burn.
Precautions
1. 2. Keep the manual valves on a newly installed bubbler closed until it is connected to the system, leak checked and purged. Avoid gas pressure differentials that can force the alkyl from the bubbler back into the carrier gas piping.
Normal Operation
1. Start-of-day: Since the pneumatic valves GAPV1and GAPV2 seal off the bubbler from the rest of the system, the manual valves on the bubbler inlet and outlet can be left open except when the bubbler is being replaced
57
2.
or removed. At the start-of-day the FE is therefore ready to operate from the control panel. The operation of the controls for the cases without hydrogen dilution and with hydrogen dilution will be discussed separately. Normal growth sequence:
58
hydrogen carrier gas will flow through GAMFC2,through the bypass valve GAPV2, through GAMFC1, through GAPV4, throughGAPV5andintotheventmanifoldata rate determined by the current value of the voltage input to GAMFC1as determined either by the front panel potentiometer or the D/A converter output for this flow controller. The bubbler pressure will be determined by the setting of PR1 and the pressure drop from the bubbler to the chamber will occur across GAMFC1. During manual operation from the front panel, the flow through GAMFC1 is controlled by the setting of the 10-turn FLOW SET 1 potentiometer on the control panel. The value of the flow setting can be read on the digital panel meter by pushing in on the black FLOW SET 1 button. When the button is released, the meter reads the value of the actual flow. When the gas circuit is operated remotely from the computer, the input voltage to GAMFC1 is supplied by a D/A converter in the HP Multiprogrammer and can be remotely programmed. The meter on the panel will continue to indicate the actual flow, and when the FLOW SET 1 button is pushed in, the meter will read the input voltage to the flow controller from the D/A converter. With the switches in this configuration the red PURGE BUBBLER light on the control panel will be on indicating that the bypass valve GAPV2 is open. The H2 DILUTE switch becomes enabled when the ACTIVE switch is on. It may be turned at any time, however, in a normal growth sequence it would be turned on after the ACTIVE switch is turned on and before the BUBBLER IN switch is turned on. When the H2 DILUTE switch is turned on, GAPV6, GAPV7, and GAPV12 open, the positive closure signal on GAMFC3 is removed, and the +10 v control signal which had been keeping GAMFC2fully open is removed. The flow rates through GAMFC1, GAMFC2, and GAMFC3 can now be controlled independently. Hydrogen may be supplied to the circuit through GAMFC2 and/or GAPV12 and may leave through GAMFC1 and/or GAMFC3. If the flow meters are set correctly, then the flow through GAMFC2 is less than the sum of the flows through GAMFC1 and GAMFC3. The pressure in the circuit will then drop until it is less than the pressure at the outlet of PR4. At this point the check valve between GAPV6and GAPV12 will open and the circuit pressure will be maintained at the pressure of PR4. The difference between the exit rate of gas through GAMFC1 and GAMFC3 and the entrance rate through GAMFC2 will be maintained by the flow through the check valve. The equivalent flow through valve GAPV5 is 2 1 1 + 3 There is nothing to prevent the operator from setting the mass flow controllers such that the flow setting for GAMFC2 is greater than the combined flow through GAMFC1 and GAMFC3. In this case the control valve for GAMFC2 would remain fully open and the flow through the circuit would be controlled by the settings on GAMFC1 and GAMFC3. The pressure in the bubbler would remain at the outlet pressure of PR1 and flow through the check valve between GAPV6 and GAPV12 would not occur. Although no damage would occur from operating the system in this manner, there would be no advantage to doing it. In fact, the amount of alkyl flowing through GAMFC3 would be wasted with no improvement in the accuracy in controlling the flow of a small amount of alkyl through GAMFC1. When the BUBBLER IN switch is turned on, GAPV1 and GAPV3 open immediately and the green light, BUBBLER IN, turns on. After a two-second delay the red PURGE BUBBLER light goes out and GAPV2 closes. The delay on closure insures that the downstream pressure is not greater than the upstream pressure (a condition which would force the liquid alkyl back into the hydrogen carrier gas piping) when GAPV1and GAPV3open. When GAPV2 closes, the hydrogen carrier gas is no longer allowed to bypass the bubbler, and instead must flow through the bubbler transporting the alkyl vapor to the vent/run valve. When the RUN switch is turned on, the yellow RUN light comes on and the gas flow is switched from the vent manifold to the run manifold and must now pass through the deposition chamber. Shutdown of the system should proceed by turning off, in sequence, the RUN switch, the H2 DILUTE switch, the BUBBLER IN switch, and the ACTIVE switch.
59
The BUBBLER IN switch should be turned off during standby periods to both conserve the source and to minimize the load of hazardous chemical which enters the scrubber. The alkyl vapor will continue to be transported from the bubbler as long as the BUBBLER IN switch is on and there is gas flow through GAMFC1 and/or GAMFC3. 3. End of day: The alkyl vapors will accelerate the deterioration of some of the elastomers from which the o-rings are fabricated. Elastomer seals have been eliminated whenever possible in the construction of this system but current technology does not allow them to be completely eliminated. They are present in the mass flow controller and in the vent/run valve. For this reason it is good practice to thoroughly flush the alkyl vapor from GAMFC1, GAMFC2, GAMFC3, and GAPV5 before the circuit is shut down. This can be accomplished by allowing hydrogen to flow through GAMFC1, GAMFC2, and GAMFC3for approximately 10 minutes after the BUBBLER IN switch is closed. During the purging step the ACTIVE and H2 PURGE switches must be on and there should be a gas flow of at least 10 sccm through GAMFC1and 100 sccm through GAMFC2 and GAMFC3. After this, the ACTIVE switch can be closed, sealing off the circuit.
Interconnections
1. 2. 3. 4. ACTIVE controls GASV1 which drives GAPV4. BUBBLER IN controls GASV2 which drives GAPV1 and GAPV3. Relays GAK1 and GAK2 control GASV3 which drives GAPV2. H2 DILUTE controls GASV4 which drives GAPV6, GAPV7, and GAPV12.
Interlocks
1. All source valves are deactivated if the following conditions exist: a. System is in abort status. b. The vacuum exhaust valve, EXPV2, is closed. c. The deposition chamber is up to atmosphere. d. The manifold diluent gas is off. All switches are disabled if the ACTIVE switch is off. GAPV1 and GAPV3 can not be opened unless GAPV4 is open. With the PURGE switch on the front panel in the locked position (key removed), GAPV2 cannot be opened if GAPV1 and GAPV3 are open. The bypass valve (GAPV2) is open if the BUBBLER IN switch is closed (handle down) and the ACTIVE switch is on (handle up). The ACTIVE, BUBBLER IN, H2 DILUTE, and RUNswitches on the front panel are disabled when the computer switches the circuit into the REMOTE mode. When the bubbler is connected to the circuit by opening the BUBBLER IN switch (handle up), valves GAPV1and GAPV3 open immediately. Valve GAPV2 closes after a 2 second time delay to insure that the pressure on each side of the bubbler is equalized. GAMFC1, GAMFC2, and GAMFC3 are commanded closed if the ACTIVE switch is off. The key lock switch CLOSE MFC2 when turned clockwise will close GAMFC2 under all conditions. When the H2 DILUTE switch is off, GAMFC3 is commanded closed and the set point of GAMFC2 is at 200 sccm. The purpose of setting the GAMFC2flow to 200 sccm is to insure that the flow setting for GAMFC2 is always greater that the flow setting for GAMFC1 when H2 DILUTE is off. This prevents the bubbler from becoming evacuated. When the H2 DILUTE switch in on, the closure signal is removed from GAMFC3 and the set point voltage for GAMFC2 is determined by either the front panel potentiometer or the remote D/A converter. A closure signal on a mass flow controller will over-ride the set point signal.
2. 3. 4. 5. 6. 7.
8. 9.
60
10. The manifold diluent gas cannot be turned off if the ACTIVE switch is on.
Bubbler removal
1. Insure that the tank valve for the nitrogen is on, that the pressure regulator on the tank gives a pressure of 20 psig, and that NMV1 is open. 2. Close the manual valves on the inlet and outlet of the TMG bubbler. 3. With the vacuum pump running, open the VAC EXHAUST switch on the VACUUM CONTROL panel. Make sure that the CONTROL VALVE OPEN light on the VACUUM CONTROL panel is on. If not press the OPEN button on the AdaptTorr ACR-26 pressure controller to fully open the pressure control valve. Make sure the MANIFOLD H2 switch on the HYDROGEN/NITROGEN control panel is off. Turn the MANIFOLD N2 switch on the HYDROGEN/NITROGEN control panel to the on position and adjust the FLOW SET 1 potentiometer to maintain a minimum deposition chamber pressure of 1-2 Torr as indicated on the AdaptTorr pressure controller. 4. On the HYDROGEN/NITROGEN panel set the N2 CARRIER/H2 CARRIER switch to the N2 CARRIER position. 5. On the GALLIUM control panel turn the ACTIVE switch and the BUBBLER IN switch on. Insert a key and turn the PURGE BUBBLER switch clockwise. The red PURGE BUBBLER light should be on. Push in the black DVM SELECT button and adjust the FLOW SET potentiometer for GAMFC1 to a flow of 100 sccm. 6. Evacuation: Insert a key in the CLOSE MFC2 switch and turn it clockwise. This will shut off the gas flow through GAMFC2. As the gas in the line is evacuated, the flow rate will drop. When the flow rate for GAMFC1 reaches zero as indicated on the panel meter, the circuit is completely evacuated. 7. Backfill:Turn the key inthe CLOSEMFC2switch counter clockwise for 10 seconds. This will allow the bubbler circuit to fill with nitrogen. 8. Repeat steps 6 and 7 twenty times to completely purge the circuit of alkyl vapor. Finish with a backfill step. 9. Turn the BUBBLER IN switch and the ACTIVE switch to the off position. 10. Disconnect and remove the bubbler. CAUTION: This is a potentially hazardous step if the manual bubbler valves have failed to seal. The pyrophoric liquid will ignite when exposed to air. The quantities are usually small enough to easily contain the burning alkyl and the oxidation is limited by the diffusion rate of oxygen into the tubing, which is rather slow. Avoid breathing the oxide fumes. Have a dry chemical or non-flammable dry absorbent medium available to smother the chemical fire should it be necessary. In most cases probably the best remedy is to attach the stainless steel VCR caps to seal the inlet and outlet of the bubbler. DO NOT USE WATER!! 11. Install the stainless steel VCR caps on the bubbler ports. Use new gaskets unless the gaskets are of the self centering type. 12. If a new bubbler is being installed immediately, proceed to the next section. Otherwise, shut down the system as appropriate.
61
Bubbler installation
1. 2. 3. 4. Insure that the tank valve for the nitrogen is on, that the pressure regulator on the tank gives a pressure of 20 psig, and that NMV1 is open. Insure that one of the hydrogen cylinders is actively connected to the hydrogen supply line, that the pressure regulator is properly set and that the hydrogen purifier is turned on and up to temperature. Insure that PR2 is set at 10 psig and PR1 is set at 15 psig. With the vacuum pump running, open the VAC EXHAUST switch on the VACUUM CONTROL panel. Make sure that the CONTROL VALVE OPEN light on the VACUUM CONTROL panel is on. If not press the OPEN button on the AdaptTorr ACR-26 pressure controller to fully open the pressure control valve. Make sure the MANIFOLD H2 switch on the HYDROGEN/NITROGEN control panel is off. Turn the MANIFOLD N2 switch on the HYDROGEN/NITROGENcontrol panel to the on position and adjust the FLOW SET 1 potentiometer to maintain a minimum deposition chamber pressure of 1-2 Torr as indicated on the AdaptTorr pressure controller. On the HYDROGEN/NITROGEN panel set the N2 CARRIER/H2 CARRIER switch to the H2 CARRIER position. Insure that the manual valves on the bubbler are closed. Uncap the inlet and outlet ports on the bubbler and connect the bubbler to the circuit using new gaskets if the previously used gaskets were not of the selfcentering type. On the GALLIUM control panel turn the ACTIVE switch and the BUBBLER IN switch on. Insert a key and turn the PURGE BUBBLER switch clockwise. The red PURGE BUBBLER light should be on. Push in the black DVM SELECT button and adjust the FLOW SET potentiometer for GAMFC1 to a flow of 100 sccm. Evacuation: Insert a key in the CLOSE MFC2 switch and turn it clockwise. This will shut off the gas flow through GAMFC2. As the gas in the line is evacuated, the flow rate will drop. When the flow rate for GAMFC1 reaches zero as indicated on the panel meter, the circuit is completely evacuated. Backfill: Turn the key in the CLOSE MFC2 switch counter clockwise for 10 seconds. This will allow the bubbler circuit to fill with nitrogen. Repeat steps 8 and 9 twenty times to completely purge the circuit of air. Finish with a backfill step. Remove the key to the CLOSE MFC2 switch. Open the manual valve on the outlet of the bubbler first. If the pressure in the bubbler should happen to be greater than 10 psig (the setting of PR2)and the inlet valve were opened first, the liquid alkyl would be forced back into the inlet plumbing. Open the manual valve on the inlet of the bubbler. Turn the BUBBLER IN switch and the ACTIVE switch to the off position. Turn the PURGE key lock switch off (counter-clockwise) and remove the key. Shut down the system as appropriate.
5. 6.
7.
8.
9. 10. 11.
62
A. Gas Distribution
1. A mixture of 10% arsine in hydrogen at 10 psig output pressure of the gas cylinder regulator is fed to a mass flow controller supplying this gas to the reactor venVrun manifold. The pressure differential between the source and the reactor chamber is accommodated across the flow controller. Alternatively the ASH3 can be valved off and hydrogen passed through the flow controller to purge the toxic arsine. 2. A purge-mode switchable from the front panel allows Evacuation and backfill of the delivery lines with H2 purge gas. Valving at the regulator allows purging of the cylinder lines for changing the cylinder.
B. Hazards
1. Arsine is very toxic. 2. Release of high pressure arsine can occur by operator carelessness. 3. Arsine and the hydrogen carrier gas are flammable. 4. Arsine is a colorless gas with a garlic-like odor. The odor threshold is 0.5 ppm. The OSHA TWA (time-weighted average for an 8-hour shift) ceiling is 0.05 ppm. The DOD ChemLogic CL8 toxic gas monitor is set to give an alarm at this value. If the level of arsine detected in the laboratory space, exceeds 0.05 ppm the building Evacuation alarm will be activated. All personnel not equipped with respirators should immediately leave the building. 5. Pure arsine has a specific gravity of 2.695 (air=1). However, a mixture of 10% AsH3 in hydrogen has a specific gravity of 0.33 (lighter than air).
C. Precautions
1. Source volume limit: Safety dictated use of a limited volume of AsH3 as a means of controlling a catastrophic failure. We have limited ourselves to a 26 cu. ft. cylinder of 10% AsH3 in hydrogen. 2. Ventilation: The AsH3 cylinder is in an enclosed, ventilated MOCVD reactor cabinet. 3. Toxic gas monitoring: a. The DOD ChemLogic CL8toxic gas monitor samples the MOCVDcabinet atmosphere. An alarm will activate at the AsH3 TLV and shut down the system. b. The DOD ChemLogic CL8 monitor also samples the immediate laboratory air. An alarm will activate at the AsH3 TLV and shut down the system. At the TLV the building Evacuation alarm will sound. (In case of alarm - leave the area and assist others in leaving the building.) c. The DOD ChemLogic CL8 monitors the outlet of the scrubber. If the gas concentration exceeds the arsine TLV threshold, the sources automatically shut down and an alarm on the operator panel sounds. 4. System operation features: a. Interlocks prevent the use of AsH3 except under normal operation of the sealed gas system. D. Opening the chamber for loading and unloading samples disables the AsH3' c. Removal of the AsH3 cylinder requires purging of the gas regulator inlet line to remove the AsHgpriorto disconnecting.
55
d. Replacement of AsH3 cylinder requires purging of the gas regulator inlet line to remove air contamination. e. Gas cylinder change should be done with an observer to double-check procedures and assist in an emergency.
D. Normal Operation
Normal daily operation once a cylinder has been installed and the delivery lines and regulator are filled. Caution: The regulator and delivery line contain arsine and are not purged during overnight shutdown. Since the cylinder valve is closed at days end, the volume of toxic gas left in the lines represents a small risk in case of line rupture. The alternative of purging the line each day would waste a significant fraction of the small cylinder volume and would add to the load of arsine to be scrubbed from the exhaust gas. 1. Start of day: a. Open the cylinder valve (Normal operating pressure is 1a psig). b. Open the regulator low pressure outlet valve, AGMV1.
63
c. Use AsH3 from the control panel as appropriate. 2. During a normal growth sequence the switches on the front panel are turned on in a left to right sequence and turned off in a right to left sequence. With all switches closed, AGMFC1 is commanded closed, AGPV4, AGPV1a and AGPV11 are closed and AGPV5 is in the vent position. When the ACTIVE switch is turned on, the signal to command closure of AGMFC1 is removed, AGPV4 opens and the green ACTIVE light comes on. When the GAS SOURCE IN switch is turned on, AGPV1a opens, gas flow as determined by the input voltage to AGMFC1 begins to flow into the vent manifold, and the green GAS SOURCE IN light comes on. During manual operation from the front panel, the flow through AGMFC1 is controlled by the setting of the 1a-turn FLOW SET potentiometer on the control panel. The value of the flow setting can be read on the digital panel meter by pushing in on the black DVM SELECT button. When the button is released, the meter reads the value of the actual flow. When the gas circuit is operated remotely from the computer, the input voltage to AGMFC1 is supplied by a D/A converter in the HPMultiprogrammer andcanbe remotelyprogrammed.Themeteron the panelwill continueto indicate the actual flow, and when the DVM SELECT button is pushed in, the meter will read the input voltage to the flow controller from the D/A converter. When the RUN switch is turned on, the flow at the outlet of AGPV5 is switched from the vent manifold to the run manifold and the yellow RUN light comes on. The system would be shut down in the reverse order. 3. End of day. a. Secure the AsH3 flow from the control panel. b. Close the cylinder valve. c. Close the valve, AGMV1, at the output of the regulator.
pneumaticvalves
1. ACTIVE controls AGSV1 which drives AGPV4. 2. GAS SOURCE IN controls AGSV2 which drives AGPV1a. 3. H2 FLUSH controls AGSV3 which drives AGPV11.
F. Interlocks
1. All source valves are deactivated if the following conditions exist: a. System is in abort status. b. The vacuum exhaust valve, ~XPV2,is closed.
56
c. The deposition chamber is up to atmosphere. d. The manifold diluent gas is off. 2. If the GAS SOURCE IN switch is on (handle up), then the H2 FLUSH switch is disabled. This prevents AGPV11 from opening if AGPV10 is open. 3. If the H2 FLUSH switch is on (turned clockwise, the red H2 FLUSH light will be on), then the GAS SOURCE IN switch is disabled. This prevents AGPV10 from opening if AGPV11 is open. 4. Interlocks2 and 3 are disabled when the key lock PURGE switch is in the on position (In this position the key is in the switch and cannot be removed) The red PURGE light in the panel will be on. This allows the gas circuit to be flushed with either hydrogen or nitrogen during either startup or after repairs have been made. For normal operation the PURGE switch should be in the off position and the key removed. 5. All front panel switches with the exception of the PURGE switch are disabled when the computer commands the circuit to the REMOTE mode.
64
actual flow in sccm A typical programming sequence would be as follows: As_g_remote_on As_g_active_on As_g_gas_on SeCas_g(230) PAUSE 5 As_g_run_on !Program the flow rate to 230 sccm !Allow for the flow rate to settle for 5 seconds !AsH3 begins to enter the chamber Deposition occurs As_g_run_off SeCas_g(O.O) As_g_gas_off As_g_active_off As_g_remote_off
65
f. Backfill the regulator inlet: First insure the tank valve for AGPURGEPR1 is open, that the outlet of regulator AGPURGEPR1 is at least 15 psig, and that AGPURGEMV1 is fully open (clockwise). Backfill the regulator input with nitrogen by opening AGMV3 for at least 2 seconds. Close AGMV3. g. Repeat the purge-backfill cycle of steps e and f twenty times. Finish with a backfill step, close AGMV3 and AGMV4, and disconnect and cap the cylinder. The handheld Mathesontoxic gas monitor should be used when unscrewing the cylinder connection. The CGA 350 connection on the AsH3 cylinder has a left-hand thread. If you are facing the cylinder, the nut on the excess flow valve must be turned clockwise to loosen it. It is Ym-importantthat the brass cap for the cylinder valve be replaced. The cylinder valves sometimes have small leaks even when tightly closed. 2. Cylinder installation: a. With the new AsH3 cylinder sitting inside the MOCVDcabinet and the exhaust fan on, slightly loosen the brass cap which is plugging the end of the CGA 350 connection on the cylinder valve. The cap has a left-hand thread. Check for the presence of AsH3 using the portable Matheson toxic gas detector. If a high level release of AsH3 is detected, the cap on the valve should be retightened
58
due to a small amount of arsine trapped between the closed valve on the cylinder and the brass cap. With the brass cap left slightly unscrewed close the door to the MOCVD cabinet and let the tank set for several hours. If no AsH3 is detected at that point, it is safe to completely unscrew the cap and install the cylinder. Otherwise the cap should be re-tightened and the tank returned to the vendor for replacement. b. Connect the excess flow valve, AGEFV1, to the cylinder. Keep the regulator inlet valve, AGMV2, closed as well as the nitrogen backfill valve AGMV3. Hold the excess flow valve, AGEFV1, open by temporarily holding in the red button with a piece of tape. c. Purge the regulator input: Turn on the nitrogen flow for the vacuum eductor, by opening NMV3. Open the purge valve, AGMV4, wait 5 seconds, and then close AGMV4 and NMV3. d. Backfill the regulator input: First insure the tank valve for AGPURGEPR1 is open, that the outlet of regulator AGPURGEPR1 is at least 15 psig, and that AGPURGEMV1 is fully open (clockwise). Backfill the regulator input with nitrogen by opening AGMV3 for at least 2 seconds. Close AGMV3. e. Repeat the backfill-purge cycle 20 times finishing with a purge step. Insure that AGMV3, AGMV4, and NMV3 are closed. f. Remove the tape from the excess flow valve, AGEFV1. g. Open the manual valve on the tank and check for leaks. h. Reset the excess flow valve, AGEFV1, by pushing in the red button and check for leaks. i. Open the manual valve on the regulator input, AGMV2 and check for leaks. j. Close the tank valve on the nitrogen cylinder supplying gas to regulator AGPURGEPR1.
A. Gas Distribution
1. A mixture of 10% phosphine in hydrogen at 10 psig output pressure of the gas cylinder regulator is fed to a mass flow controller supplying this gas to the reactor vent/run manifold. The pressure differential between the source and the reactor chamber is accommodated across the flow controller. Alternatively the PH3 can be valved off and hydrogen passed through the flow controller to purge the toxic phosphine. 2. A purge-mode switchable from the front panel allows Evacuation and backfill of the delivery lines with H2 purge gas. Valving at the regulator allows purging of the cylinder lines for changing the cylinder.
B. Hazards
1. Phosphine is very toxic. 2. Phosphine is pyrophoric. 3. Release of high pressure phosphine can occur by operator carelessness. 4. The hydrogen carrier gas is flammable 5. Phosphine is a colorless gas with a disagreeable characteristic fishy, garlic-like, or acetylenelike
66
odor. The odor threshold is 0.021 ppm. The OSHA TWA (time-weighted average for an 8hour shift) ceiling is 0.3 ppm. The DOD ChemLogic CL8 toxic gas monitor is set to give an alarm at 0.05 ppm. If the level of phosphine detected in the laboratory space, exceeds 0.1 ppm the building Evacuation alarm will be activated. All personnel not equipped with respirators should immediately leave the building. 6. Pure phosphine has a specific gravity 1.17 (air=1). However, a mixture of 10% PH3 in hydrogen has a specific gravity of 0..18(lighterthan air).
C. Precautions
1. Source volume limit: Safety dictated use of a limited volume of PH3 as a means of controlling a catastrophic failure. We have limited ourselves to a 26 cu. ft. cylinder of 10% PH3 in hydrogen. 2. Ventilation: The PH3 cylinder is in an enclosed, ventilated MOCVD reactor cabinet. 3. Toxic gas monitoring: a. The DOD ChemLogic CL8toxic gas monitor samples the MOCVD cabinet atmosphere. An alarm will activate at the AsH3 TLV and shut down the system. The more sensitive limit for arsine provides extra margin for phosphine detection. b. The DOD ChemLogic CL8 monitor also samples the immediate laboratory air. An alarm will activate at the AsH3 TLV and shut down the system. At the TLV the building Evacuation alarm will sound. (In case of alarm -leave the area and assist others in leaving the building.) c. The DOD ChemLogic CL8 monitors the outlet of the scrubber. If the gas concentration exceeds the arsine TLV threshold, the sources automatically shut down and an alarm on the operator panel sounds. 4. System operation features: a. Interlocks prevent the use of PH3 except under normal operation of the sealed gas system. b. Opening the chamber for loading and unloading samples disables the PH3' c. Removal of the PH3 cylinder requires purging of the gas regulator inlet line to remove the PH3 prior to disconnecting. 60 "
~~--~--- ~- ---~-
d. Replacement of PH3 cylinder requires purging of the gas regulator inlet line to remove air contamination. e. Gas cylinder change should be done with an observer to double-check procedures and assist in an emergency.
D. Normal Operation
Normal daily operation once a cylinder has been installed and the delivery lines and regulator are filled. Caution: The regulator and delivery line contain phosphine and are not purged during overnight shutdown. Since the cylinder valve is closed at days end, the volume of toxic gas left in the lines represents a small risk in case of line rupture. The alternative of purging the line each day would waste a significant fraction of the small cylinder volume and would add to the load of phosphine to be scrubbed from the exhaust gas. 1. Start of day: a. Open the cylinder valve (Normal operating pressure is 10 psig). b. Open the regulator low pressure outlet valve, PGMV1. c. Use PH3 from the control panel as appropriate. 2. During a normal growth sequence the switches on the front panel are turned on in a left to right sequence and turned off in a right to left sequence. With all switches closed, PGMFC1 is commanded closed, PGPV4, PGPV10and PGPV11 are closed and PGPV5 is in the vent position. When the ACTIVE switch is turned on, the signal to command closure of PGMFC1 is removed, PGPV4 opens and the green ACTIVE light comes on. When the GAS SOURCE IN switch is turned on, PGPV10 opens, gas flow as determined by the input voltage to PGMFC1 begins to flow into the vent manifold, and the green GAS SOURCE IN light comes on. During manual operation from the front panel, the flow through PGMFC1 is controlled by the setting of the 10-turn FLOW SET potentiometer on the control panel. The value of the flow setting can be read on the digital panel meter by pushing in on the black DVMSELECT button. When the button is released, the meter reads the value of the actual flow. When the gas circuit is operated
67
remotely from the computer, the input voltage to PGMFC1is supplied by a D/A converter in the HP Multiprogrammer and can be remotely programmed. The meter on the panel willcontinue to indicate the actual flow, and when the DVM SELECT button is pushed in, the meter will read the input voltage to the flow controller from the D/A converter. When the RUN switch is turned on, the flow at the outlet of PGPV5 is switched from the vent manifold to the run manifold and the yellow RUN light comes on. The system would be shut down in the reverse order. 3. End of day. a. Secure the PH3 flow from the control panel. b. Close the cylinder valve. c. Close the valve, PGMV1,at the output of the regulator.
F. Interlocks
1. All source valves are deactivated if the following conditions exist: 61 a. System is in abort status. b. The vacuum exhaust valve, EXPV2, is closed. c. The deposition chamber is up to atmosphere. d. The manifold diluent gas is off. 2. If the GAS SOURCE IN switch is on (handle up), then the H2 FLUSH switch is disabled. This prevents PGPV11from opening if PGPV10 is open. 3. If the H2 FLUSH switch is on (key turned clockwise and not removable from switch, the red H2 FLUSH light will be on), then the GAS SOURCE IN switch is disabled. This prevents PGPV10from opening if PGPV11 is open. 4. Interlocks 2 and 3 are disabled when the key lock PURGEswitch is in the on position (In this position the key is ill the switch and cannot be removed) The red PURGE light in the panel will be on. This allows the gas circuit to be flushed with either hydrogen or nitrogen during either startup or after repairs have been made. For normal operation the PURGE switch should be in the off position and the key removed. 5. All front panel switches with the exception of the PURGEswitch are disabled when the computer commands the circuit to the REMOTE mode.
!Program the flow rate to 230 sccm !Allow for the flow rate to settle for 5 seconds !PH3 begins to enter the chamber
Deposition occurs
68
62
-- -- n- - - _n_---
lines allow cycle times of a few minutes. The regulator inlet lines can be independently purged for changing cylinders;
69
closed. 2. Cylinder installation: a. With the new PH3 cylinder sitting inside the MOCVD cabinet and the exhaust fan on, slightly loosen the brass cap which is plugging the end of the CGA 350 connection on the cylinder valve. 63 The cap has a left-hand thread. Check for the presence of PH3 using the portable Matheson toxic gas detector. If a high level release of PH3 is detected, the cap on the valve should be retightened and the cylinder returned to the vendor. If a low level release is detected, it may be due to a small amount of phosphine trapped between the closed valve on the cylinder and the brass cap. With the brass cap left slightly unscrewed close the door to the MOCVDcabinet and let the tank set for several hours. If no PH3 is detected at that point, it is safe to completely unscrew the cap and install the cylinder. Otherwise the cap should be re-tightened and the tank returned to the vendor for replacement. b. Connect the excess flow valve, PGEFV1,to the cylinder. Keep the regulator inlet valve, PGMV2, closed as well as the nitrogen backfill valve PGMV3. Hold the excess flow valve, PGEFV1, open by temporarily holding in the red button with a piece of tape. c. Purge the regulator input: Turn on the nitrogen flow for the vacuum eductor, by opening NMV5. Open the purge valve, PGMV4,wait 5 seconds, and then close PGMV4 and NMV5. d. Backfill the regulator input: First insure the tank valve for PGPURGEPR1 is open, that the outlet of regulator PGPURGEPR1 is at least 15 psig, and that PGPURGEMV1 is fully open (clockwise). Backfill the regulator input with nitrogen by opening PGMV3for at least 2 seconds. Close PGMV3. e. Repeat the backfill-purge cycle 20 times finishing with a purge step. Insure that PGMV3, PGMV4, and NMV5 are closed. 1. Remove the tape from the excess flow valve, PGEFV1. g. Open the manual valve on the tank and check for leaks. h. Reset the excess flow valve, PGEFV1, by pushing in the red button and check for leaks. i. Open manual valve on the regulator input, PGMV2 and check for leaks. j. Close the tank valve on the nitrogen cylinder supplying gas to regulator PGPURGEPR1.
70
4. Pressure drop to the reaction chamber occurs across the mass flow controller, NDOPMFC1. 5. The alkyl is exposed to the heated sensor element of the mass flow controller (110C); thus it must be stable at this temperature. 6. Loss of liquid from the bubbler into the gas lines is unlikely because pressure in the gas line is the same as the bubbler. (Closure of the bypass valve, NDOPPV2, is delayed to allow bubbler inlet and outlet pressure to equilibrate.) 7. Upon power failure or other abort conditions, contamination of the upstream H2lines is prevented by closure of the control valve for mass flow controller NDOPMFC2. 8. Adiabatic cooling could possibly cause alkyl condensation in the mass flow controller valve. It is advisable to keep the alkyl temperature below room temperature. 9. A purge mode overrides some interlocks to allow Evacuation and backfill of the bubbler lines
B. Hazards
1. TMSn is pyrophoric 2. The oxides formed from the burning of TMSn are toxic. 3. TMSn reacts violently with water which should never be used to extinguish the burning alkyl. Use a dry chemical fire extinguisher only or else let it burn. 4. The dust (silicon dioxide) formed when silane reacts with air is toxic. 5. Release of high pressure silane can occur by operator carelessness.
65
--- -- - - - -- - -
c. Precautions
1. Source volume limit: Safety dictated use of a limited volume of SiH4 as a means of controlling a catastrophic failure. We have limited ourselves to a 26 cu. ft. cylinder of 5% SiH4 in argon. 2. Ventilation: The SiH4 cylinder is in an enclosed, ventilated MOCVD reactor cabinet. 3. Toxic gas monitoring: a. The DOD ChemLogic CL8toxic gas monitor samples the MOCVDcabinet atmosphere. An alarm will activate at the AsH3 TLV and shut down the system. The more sensitive limit for arsine provides extra margin for silane detection. b. The DOD ChemLogic CL8 monitor also samples the immediate laboratory air. An alarm will activate at the AsH3 TLV and shut down the system. At the TLV the building Evacuation alarm will sound. (In case of alarm -leave the area and assist others in leaving the building.) c. The DOD ChemLogic CL8 monitors the outlet of the scrubber. If the gas concentration exceeds the arsine TLV threshold, the sources automatically shut down and an alarm on the operator panel sounds. 4. System operation features: a. Interlocks prevent the use of SiH4 except under normal operation of the sealed gas system. b. Opening the chamber for loading and unloading samples disables the SiH4' c. Removal of the SiH4 cylinder requires purging of the gas regulator inlet line to remove the SiH4 prior to disconnecting. d. Replacement of SiH4 cylinder requires purging of the gas regulator inlet line to remove air contamination. e. Gas cylinder change should be done with an observer to double-check procedures and assist in an emergency. 5. The gaseous source is a mixture of 5% silane (SiH4) in argon. At this concentration, the gas is labeled flammable (not pyrophoric). Silane is especially reactive with air, forming silicon dioxide (Si02)' The silicon dioxide forms as a fine white dust which can deposit in the plumbing and valves if air and silane are allowed to mix. The dust will prevent valve seats from forming a complete seal and can cause the pressure control valve inside the gas regulator to fail. When this happens the pressure downstream of the gas regulator will creep upward; if left without an excape route the downstream pressure could presumably rise to the full value of the pressure in the source cylinder (=2000 psi). For this reason a relief valve (NDOPPRV1) on the downstream side of the regulator has been installed to vent the gas to the scrubber if the line pressure exceeds 150 psi. This will prevent components from bursting if the regulator should fail. It is therefore extremely important to purge all the air from the components before installing a silane cylinder and, likewise, very important to purge all the silane from the lines before removing a silane cylinder. If this is not done carefully, some of the gas control
71
components could be ruined and others might require a time consuming cleanup process. An excess flow valve (NDOPEFV1) is installed on the silane tank. If the gas flow through the valve exceeds 1000 sccm, the valve will trip shutting off the flow. It then stays closed unless it is manually reset by pushing in on the red button. The next valve after the excess flow valve is a pneumatically controlled valve (NDOPPV20) held open by the abort interlock system. This valve automatically closes in the event of power failure, loss of ventilation, operator intervention, fire, low hydrogen pressure, low nitrogen pressure, or detection of toxic gas in the cabinet, room, or at the scrubber outlet. 6. Keep the manual valves on a newly installed bubbler closed until it is connected to the system, leak checked and purged. 7. Avoid gas pressure differentials that can force the alkyl from the bubbler back into the carrier gas piping.
66
H- -n_-nn_-n-n
D. NormalOperation
The N-Dopant circuit can provide source material from either the bubbler or the high-pressure cylinder. These flows may be operated with or without additional hydrogen dilution. Each mode of operation will be described in a separate section. 1. Start-of-day: Since the pneumatic valves NDOPPV1and NDOPPV2seal off the bubbler from the rest of the system, the manual valves on the bubbler inlet and outlet can be left open except when the bubbler is being replaced or removed. At the start-of-day the TMSn is therefore ready to operate from the control panel. The regulator and delivery line contain silane and are not purged during overnight shutdown. Since the cylinder valve is closed at days end, the volume of toxic gas left in the lines represents a small risk in case of line rupture. The alternative of purging the line each day would waste a significant fraction of the small cylinder volume and would add to the load of silane to be scrubbed from the exhaust gas. 2. Normal growth sequence: Operation of bubbler source without additional hydrogen dilution: With all switches closed, NDOPMFC1, NDOPMFC2, NDOPMFC3,and NDOPMFC4 are commanded closed, NDOPPV1, NDOPPV2, NDOPPV3, NDOPPV4, NDOPPV6, NDOPPV7, NDOPPV8, NDOPPV9, NDOPPV10, and NDOPPV11 are closed and NDOPPV5 is in the vent position. When the ACTIVE switch is turned on, the signals to command closure of NDOPMFC1 and NDOPMFC2 are removed, NDOPPV2, NDOPPV4, and NDOPPV9open and the green ACTIVE light comes on. A voltage equal to +10 v is applied to the control inputs of NDOPMFC2 and NDOPMFC4 which programs their flow settings to 200 seem. This insures that the flow setting through NDOPMFC2 or NDOPMFC4 will always be larger than the flow through NDOPMFC1and the circuit will not become evacuated; the control valves contained within NDOPMFC2 and NDOPMFC4 will remain wide open. With the switches in this configuration, the hydrogen carrier gas will flow through NDOPMFC2,through the bypass valve NDOPPV2, through NDOPPV9, through NDOPMFC1,through NDOPPV4, through NDOPPV5and into the vent manifold at a rate determined by the current value of the voltage input to NDOPMFC1as determined either by the front panel potentiometer or the D/A converter output for this flow controllers. The circuit pressure will be determined by the setting of PR1 and the pressure drop from the circuit to the chamber will occur across NDOPMFC1. During manual operation from the front panel, the flow through NDOPMFC1 is controlled by the setting of the 10-turn FLOW SET 1 potentiometer on the control panel. The value of the flow setting can be read on the digital panel meter by pushing in on the black FLOW SET 1 button. When the button is released, the meter reads the value of the actual flow. When the gas circuit is operated remotely from the computer, the input voltage to NDOPMFC1 is supplied by a D/A converter in the HP Multiprogrammer and can be remotely programmed. The meter on the panel will continue to indicate the actual flow, and when the FLOW SET 1 button is pushed in, the meter will read the input voltage to the flow controller from the D/A converter. With the switches in the present configuration, the red PURGE BUBBLER light on the control panel will be on indicating that the bypass valve NDOPPV2 is open. When the BUBBLER INswitchisturnedon,NDOPPV1 and NDOPPV3open immediately and the green light,
72
BUBBLER IN, turns on. After a two-second delay the red PURGE BUBBLER light goes out and NDOPPV2 closes. This insures that the downstream pressure is not greater than the upstream pressure (a condition which would force the liquid alkyl back into the hydrogen carrier gas piping) when NDOPPV1and NDOPPV3open. When NDOPPV2closes, the hydrogen carrier gas is no longer allowed to bypass the bubbler, and instead must flow through the bubbler transporting the alkyl vapor to the vent/run valve. When the RUN switch is turned on, the yellow RUN light comes on and the gas flow is switched from the vent manifoldto the run manifold and must now pass 67
- -- - -- -- -- -
through the deposition chamber. Shutdown of the system should proceed by reversing the sequence of the above steps. The BUBBLER IN switch should be turned off during standby periods to both conserve the source and to minimize the load of hazardous chemical which enters the scrubber. The alkyl vapor will continue to be transported from the bubbler as long as the BUBBLER IN switch is on and there is gas flow through NDOPMFC1. Operation of bubbler source with additional hydrogen dilution: With all switches closed, NDOPMFC1, NDOPMFC2, and NDOPMFC3 are commanded closed, NDOPPV1, NDOPPV2, NDOPPV3, NDOPPV4, NDOPPV6, NDOPPV7, NDOPPV8, NDOPPV9, NDOPPV10, and NDOPPV11 are closed and NDOPPV5 is in the vent position. When the ACTIVE switch is turned on, the signals to command closure of NDOPMFC1 and NDOPMFC2 are removed, NDOPPV2, NDOPPV4,and NDOPPV9open and the green ACTIVE light comes on. A voltage equal to +10 v is applied to the control inputs of NDOPMFC2 and NDOPMFC4 which programs their flow settings to 200 sccm. This insures that the flow setting for NDOPMFC2 will always be larger than the flow setting for NDOPMFC1 and the circuit will not become evacuated; the control valves contained within NDOPMFC2 and NDOPMFC4will remain wide open. With the switches in this configuration, the hydrogen carrier gas will flow through NDOPMFC2,through the bypass valve NDOPPV2, through NDOPPV9,through NDOPMFC1, through NDOPPV4, through NDOPPV5and into the vent manifold at a rate determined by the current value of the voltage input to NDOPMFC1as determined either by the front panel potentiometer or the D/A converter output for this flow controller. The circuit pressure will be determined by the setting of PR1 and the pressure drop from the bubbler to the chamber will occur across NDOPMFC1. During manual operation from the front panel, the flow through NDOPMFC1 is controlled by the setting of the 10-turn FLOW SET 1 potentiometer on the control panel. The value of the flow setting can be read on the digital panel meter by pushing in on the black FLOW SET 1 button. When the button is released, the meter reads the value of the actual flow. When the gas circuit is operated remotely from the computer, the input voltage to NDOPMFC1is supplied by a D/A converter in the HP Multiprogrammer and can be remotely programmed. The meter on the panel will continue to indicate the actual flow, and when the FLOW SET 1 button is pushed in, the meter will read the input voltage to the flow controller from the D/A converter. With the switches in this configuration the red PURGE BUBBLER light on the control panel will be on indicating that the bypass valve NDOPPV2 is open. The H2 DILUTE switch becomes enabled when the ACTIVE switch is on. It may be turned at any time, however, in a normal growth sequence it would be turned on atter the ACTIVE switch is turned on and before the BUBBLER IN switch is turned on. When the H2 DILUTE switch is turned on, NDOPPV6, and NDOPPV7 open, the positive closure signal on NDOPMFC3 is removed, and the +10 v control signal which had been keeping NDOPMFC2 fully open is removed. The flow rates through NDOPMFC1, NDOPMFC2, and NDOPMFC3 can now be controlled independently. Hydrogen may be supplied to the circuit through NDOPMFC2 and/or NDOPPV6 and may leave through NDOPMFC1 and/or NDOPMFC3. If the flow meters are set correctly, then the flow through NDOPMFC2 is less than the sum of the flows through NDOPMFC1 and NDOPMFC3. The pressure in the circuit will then drop until it is less than the pressure at the outlet of PR6. At this point the check valve between NDOPPV6and PR6 will open and the circuit pressure will be maintained at the pressure of PR6. The difference between the exit rate of gas through NDOPMFC1 and NDOPMFC3 and the entrance rate through NDOPMFC2 will be maintained by the flow through the check valve. The equivalent flow through valve NDOPPV5 is
73
Flow2 Flow1
Flow 1 + Flow 3
68
- U H - - U U U -U- -----
- ----
There is nothing to prevent the operator from setting the mass flow controllers such that the flow setting for NDOPMFC2 is greater than the combined flow through NDOPMFC1 and NDOPMFC3. In this case the control valve for NDOPMFC2 would remain fully open and the flow through the circuit would be controlled by the settings on NDOPMFC1and NDOPMFC3. The pressure in the bubbler would remain at the outlet pressure of PR1 and flow through the check valve between NDOPPV6 and PR6 would not occur. Although no damage would occur from operating the system in this manner, there would be no advantage to doing it. In fact, the amount of alkyl flowing through NDOPMFC3would be wasted with no improvement in the accuracy in controlling the flow of a small amount of alkyl through NDOPMFC1. When the BUBBLER IN switch is turned on, NDOPPV1and NDOPPV3open immediately and the green light, BUBBLER IN, turns on. After a two-second delay the red PURGE BUBBLER light goes out and NDOPPV2 closes. The delay on closure insures that the downstream pressure is not greater than the upstream pressure (a condition which would force the liquid alkyl back into the hydrogen carrier gas piping) when NDOPPV1 and NDOPPV3open. When NDOPPV2closes, the hydrogen carrier gas is no longer allowed to bypass the bl,Jbbler, and instead must flow through the bubbler transporting the alkyl vapor to the vent/run valve. When the RUN switch is turned on, the yellow RUN light comes on and the gas flow is switched from the vent manifold to the run manifold and must now pass through the deposition chamber. Shutdown of the system should proceed by turning off, in sequence, the RUN switch, the BUBBLER IN switch, the H2 DILUTE switch, and the ACTIVE switch. The BUBBLER IN switch should be turned off during standby periods to both conserve the source and to minimize the load of hazardous chemical which enters the scrubber. The alkyl vapor will continue to be transported from the bubbler as long as the BUBBLER IN switch is on and there is gas flow through NDOPMFC1 andfor NDOPMFC3. Operation of gaseous source without additional hydrogen dilution: The valve on the top of the silane cylinder should be opened and the downstream pressure of regulator NDOPPR1should be checked to see that it is 10 psig. With all switches closed, NDOPMFC1, NDOPMFC2, and NDOPMFC3 are commanded closed, NDOPPV1, NDOPPV2, NDOPPV3, NDOPPV4, NDOPPV6, NDOPPV7, NDOPPV8, NDOPPV9, NDOPPV10,and NDOPPV11 are closed and NDOPPV5 is in the vent position. When the ACTIVE switch is turned on, the signals to command closure of NDOPMFC1 and NDOPMFC2 are removed, NDOPPV2, NDOPPV4,and NDOPPV9 open and the green ACTIVE light comes on. A voltage equal to +10 v is applied to the control inputs of NDOPMFC2 and NDOPMFC4 which programs their flow setting to 200 seem. This insures that the flow setting for NDOPMFC2 and NDOPMFC4 will always be larger than the flow setting for NDOPMFC1 and the circuit will not become evacuated; the control valves contained within NDOPMFC2 and NDOPMFC4 will remain wide open. With the switches in this configuration, the hydrogen carrier gas will flow through NDOPMFC2,through the bypass valve NDOPPV2, through NDOPPV9, through NDOPMFC1,through NDOPPV4, through NDOPPV5 and into the vent manifold at a rate determined by the current value of the voltage input to NDOPMFC1 as determined either by the front panel potentiometer or the D/A converter output for this flow controller. The circuit pressure will be determined by the setting of PR1 and the pressure drop from the bubbler to the chamber will occur across NDOPMFC1. During manual operation from the front panel, the flow through NDOPMFC1 is controlled by the setting of the 10-turn FLOW SET 1 potentiometer on the control panel. The value of the flow setting can be read on the digital panel meter by pushing in on the black FLOW SET 1 button. When the button is released, the meter reads the value of the actual flow. When the gas circuit is operated remotely from the computer, the input voltage to NDOPMFC1 is supplied by a D/A converter in the HP Multiprogrammer and can be remotely programmed. The meter on the panel will continue to indicate the actual flow, and when the FLOW SET 1 button is pushed in, the meter will read the input voltage to the flow controller from the D/A converter. With the switches in this configuration the red PURGE BUBBLER light on the control panel will be on indicating that the bypass valve
74
NDOPPV2 is open.
69
~~ -~ -- - ~---~--~ -- ~
~ ~~--~-~- - ~
The GASEOUS SOURCE IN switch is disabled when the BUBBLER IN switch is on. When the GASEOUS SOURCE IN switch is turned on (with the BUBBLER IN switch off), NDOPPV9 closes, NDOPPV8and NDOPPV10 open and the signal to command closure is removed from NDOPMFC4. Since a +10 v control signal is applied to NDOPMFC4, its control valve will go to a wide open state. Dopant gas from the high-pressure cylinder will begin to flow through the circuit at a rate determined by the setting for NDOPMFC1. The pressure in the ciruit will be determined by the pressure setting of NDOPPR1and the pressure drop will occur across NDOPMFC1.When the RUN switch is turned on, the flow at the outlet of PGPV5 is switched from the vent manifold to the run manifold and the yellow RUN light comes on. The system would be shut down by reversing the sequence of the above steps. Operation of gaseous source with additional hydrogen dilution: The valve on the top of the silane cylinder should be opened and the downstream pressure of regulator NDOPPR1should be checked to see that it is 10 psig. With all switches closed, NDOPMFC1, NDOPMFC2, and NDOPMFC3 are commanded closed, NDOPPV1, NDOPPV2, NDOPPV3, NDOPPV4, NDOPPV6, NDOPPV7, NDOPPV8, NDOPPV9, NDOPPV10,and NDOPPV11 are closed and NDOPPV5 is in the vent position. When the ACTIVE switch is turned on, the signals to command closure of NDOPMFC1 and NDOPMFC2 are removed, NDOPPV2, NDOPPV4, and NDOPPV9 open and the green ACTIVE light comes on. A voltage equal to +10 v is applied to the control inputs of NDOPMFC2 and NDOPMFC4 which programs their flow setting to 200 sccm. This insures that the flow setting for NDOPMFC2 and NDOPMFC4 will always be larger than the flow setting for NDOPMFC1 and the circuit will not become evacuated; the control valves contained within NDOPMFC2 and NDOPMFC4will remain wide open. With the switches in this configuration, the hydrogen carrier gas will flow through NDOPMFC2,through the bypass valve NDOPPV2, through NDOPPV9, through NDOPMFC1,through NDOPPV4,through NDOPPV5and into the vent manifold at a rate determined by the current value of the voltage input to NDOPMFC1as determined either by the front panel potentiometer or the D/A converter output for this flow controller. The circuit pressure will be determined by the setting of PR1 and the pressure drop from the bubbler to the chamber will occur across NDOPMFC1. During manual operation from the front panel, the flow through NDOPMFC1 is controlled by the setting of the 10-turn FLOW SET 1 potentiometer on the control panel. The value of the flow setting can be read on the digital panel meter by pushing in on the black FLOW SET 1 button. When the button is released, the meter reads the value of the actual flow. When the gas circuit is operated remotely from the computer, the input voltage to NDOPMFC1 is suppliedbya D/Aconverterinthe HP Multiprogrammer and can be remotely programmed. The meter on the panel will continue to indicate the actual flow, and when the FLOW SET 1 button is pushed in, the meter will read the input voltage to the flow controller from the D/A converter. With the switches in this configuration the red PURGE BUBBLER light on the control panel will be on indicating that the bypass valve NDOPPV2 is open. The GASEOUS SOURCE IN switch is disabled when the BUBBLER IN switch is on. When the GASEOUS SOURCE IN switch is turned on (with the BUBBLER IN switch off), NDOPPV9 closes, NDOPPV8and NDOPPV10 open and the signal to command closure is removed from NDOPMFC4. Since a +10 v control signal is applied to NDOPMFC4, its control valve will go to a wide open state. Dopant gas from the high-pressure cylinder will begin to flow through the circuit at a rate determined by the setting for NDOPOMFC1. The pressure in the ciruit will be determined by the pressure setting of NDOPPR1and the pressure drop will occur across NDOPMFC1. The H2 DILUTE switch becomes enabled when the ACTIVE switch is on. It may be turned at any time, however, in a normal growth sequence it would be turned on after the ACTIVE and the GASEOUS SOURCE IN switches are turned on. When the H2 DILUTE switch is turned on, NDOPPV6, and NDOPPV7open, the positive closure signal on NDOPMFC3 is removed, and the +10 v control signal which had been keeping NDOPMFC4fully open is removed.The flow rates through NDOPMFC1, NDOPMFC3,and NDOPMFC4 can now be
75
controlled independentlyH.ydrogenmaybesuppliedtothecircuithroughNDOPMFCa4nd/or
70
NDOPPV6 and may leave through NDOPMFC1 and/or NDOPMFC3. If the flow meters are set correctly, then the flow through NDOPMFC4 will be less than the sum of the flows through NDOPMFC1 and NDOPMFC3. The pressure in the circuit will then drop until it is less than the pressure at the outlet of PR6. At this point the check valve between NDOPPV6 and PR6 will open and the circuit pressure will be maintained at the pressure of PR6. The difference between the exit rate of gas through NDOPMFC1 and NDOPMFC3 and the entrance rate through NDOPMFC4 will be maintained by the flow through the check valve. The equivalent flow through valve NDOPPV5 is
-- -- - ~
F. Interlocks
1. All source valves are deactivated if the following conditions exist: a. System is in abort status. b. The vacuum exhaust valve, EXPV2, is closed. c. The deposition chamber is up to atmosphere. d. The manifold diluent gas is off. 2. The BUBBLER IN, GAS SOURCE IN, RUN, and H2 DILUTE switches are disabled if the ACTIVE switch is off.
76
3. The BUBBLER IN switch is disabled if the GAS SOURCE IN switch is on. This implies that NDOPPV1 and NDOPPV3 can not be opened unless NDOPPV4 is open and NDOPPVa and NDOPPV10 are closed. 4. The GAS SOURCE IN switch is disabled if the BUBBLER IN switch is on. This implies that NDOPPV8 and NDOPPV10 can not be opened unless NDOPPV4 is open and NDOPPV1and NDOPPV3 are closed. 5. With the PURGE switch on the front panel in the locked position (key removed), NDOPPV2 cannot be opened if NDOPPV1 and NDOPPV3are open. 6. The bypass valve (NDOPPV2) is open if the BUBBLER IN switch is closed (handle down) and the ACTIVE switch is on (handle up). 7. The ACTIVE, BUBBLER IN, GAS SOURCE IN, H2 DILUTE, and RUN switches on the front panel are disabled when the computer switches the circuit into the REMOTE mode. a. When the bubbler is connected to the circuit by opening the BUBBLER IN switch (handle up), valves NDOPPV1and NDOPPV3open immediately. Valve NDOPPV2 closes after a 2 second time delay to insure that the pressure on each side of the bubbler is equalized 9. NDOPMFC1, NDOPMFC2, NDOPMFC3,and NDOPMFC4 are commanded closed if the ACTIVE switch is off. The key lock switch CLOSE MFC2 when turned clockwise will close NDOPMFC2 under all conditions. The key lock switch CLOSE MFC4 when turned clockwise will close NDOPMFC4 under all conditions. 10. When the H2 DILUTE switch is off, NDOPMFC3 is commanded closed and the set points of NDOPMFC2 and NDOPMFC4 are at 200 seem. The purpose of setting the NDOPMFC2 and NDOPMFC4 flows to 200 seem is to insure that their flow settings are always greater that the flow setting for NDOPMFC1 when H2 DILUTE is off. This prevents the circuit from becoming evacuated. When the H2 DILUTE switch in on, the closure signal is removed from NDOPMFC3 and the set point voltages for NDOPMFC2 and NDOPMFC4 are determined by either their front panel potentiometers or their remote D/A converters. A closure signal on a mass flow controller will over-ride the set point signal. 11. The manifold diluent gas cannot be turned off if the ACTIVE switch is on. 12. If the GAS SOURCE IN switch is on (handle up), then the H2 FLUSH switch is disabled. This prevents NDOPPV11 from opening if NDOPV10 is open. 13 All front panel switches with the exception of the PURGE BUBBLER and H2 FLUSH switches are disabled when the computer commands the circuit to the REMOTE mode.
72
7. Programming of mass flow controllers: SeCn_dop_da !The same DA converters are used for both the n-dopant and p-dopant !systems. This command connects the DA converters to the n-dopant !system. SeCndop_1(Flow) Flow given in sccm (0~Flow~200) SeCndop_2(Flow) Flow given in sccm (0~Flow~200) SeCndop_3(Flow) Flow given in sccm (0~Flow~2000) SeCndop- 4(Flow) Flow given in sccm (0~Flow~200 To read flow; Read_ndop(Flow) Returns a floating point number,Flow, equal to the actual flow in sccm Read_ndop_1(Flow) Read_ndop_2(Flow) Read_ndop_3(Flow) Read_ndop-4(Flow) To set flow: A typical programming sequence for flow from the bubbler with no hydrogen dilution would be as
77
follows: Ndop_remote_on SeCndop_1(6.5) Ndop_active_on Ndop_bub_on Ndop_run_on !Program the flow rate for NDOPMFC1to 6.5 sccm !TMSn vapor and hydrogen carrier begins to enter the chamber Deposition occurs Ndop_run_off SeCndop_1(0.0) Ndop_bub_off Ndop_active_off Ndop_remote_off A typical programming sequence for flow from the bubbler with additional hydrogen dilution would be: Ndop_remote_on SeCndop_1(6.5) SeCndop_2(15) Set_ndop_3(100) Ndop_active_on Ndop_h2_on !Program the flow rate for NDOPMFC1 to 6.5 secm !Program the flow rate for NDOPMFC2 to 15 sccm !Program the flow rate for NDOPMFC3 to 100 sccm !Pure hydrogen will flow through NDOPMFC1 and NDOPMFC2 !at a flow rate determined by the flow setting of NDOPMFC1 and !exit through the vent manifold. !Pure hydrogen will flow through NDOPMFC1, NDOPMFC2, and !NDOPMFC3at the set flow rates and exit through the vent !manifold and the exhaust line connected to the outlet of !NDOPMFC3 !The hydrogen flow through NDOPMFC2 will be diverted through !the bubbler !TMSn vapor and hydrogen carrier begins to enter the chamber Ndop_bub_on Ndop_run_on
73
Deposition occurs Ndop_run_off !The flow to the deposition chamber is diverted to the vent !manifold Ndop_bub_off Ndop_h2_off Ndop_active_off SeCndop_1(0.0) SeCndop_2(0.0) SeCndop_3(0.0) Ndop_active_off Ndop_remote_off !Turns off all gas flows !These commands are mostly for housekeeping so that a mass !flow controller isn't later used at an unintended setting A typical programming sequence for flow from the high-pressure source without hydrogen dilution would be: Ndop_remote_on Ndop_active_on
78
Ndop_gas_on SeCndop_1(230) Ndop_run_on !Program the flow rate to 230 sccm !SiH4 begins to enter the chamber Deposition occurs Ndop_run_off SeCndop_1(0.0) Ndop_gas_off Ndop_active_off Ndop_remote_off A typical programming sequence for flow from the high-pressure source with additional hydrogen dilution would be: Ndop_remote_on Set_ndop_1(6.5) SeCndop- 4(15) SeCndop_3(100) Ndop_active_on !Program the flow rate for NDOPMFC1 to 6.5 sccm !Program the flow rate for NDOPMFC4to 15 sccm !Program the flow rate for NDOPMFC3,to 100 sccm !Pure hydrogen will flow through NDOPMFC1 and NDOPMFC2 !at a flow rate determined by the flow setting of NDOPMFC1 and !exit through the vent manifold. INDOPPV9 closes and NDOPPV8and NDOPPV10 !open. Dopant gas flows through NDOPPV10 and into the !vent manifold at a rate determined byNDOPMFC1 !Gas flows through NDOPMFC1, NDOPMFC3, and !NDOPMFC4at the set flow rates and exits through the vent !manifold and the exhaust line connected to the outlet of !NDOPMFC3 !SiH4 and hydrogen carrier begins to enter the chamber Ndop_gas_on Ndop_h2_on Ndop_run_on 74 Deposition occurs Ndop_run_off !The flow to the deposition chamber is diverted to the vent !manifold Ndop_h2_off Ndop_gas_off Ndop_active_off Set_ndop_1(0.0) SeCndop_3(0.0) SeCndop- 4(0.0) Ndop_active_off Ndop_remote_off
79
them with hydrogen. During the Evacuation cycle it is necessary to open NDOPPV4 and keep NDOPMFC2 closed. For this purpose, a key lock switch, CLOSE MFC2, is installed on the front panel. When the key position is fully clockwise, NDOPMFC2will be closed regardless of the positions of the other switches.. Bubbler removal 1. Insure that the tank valve for the nitrogen is on, that the pressure regulator on the tank gives a pressure of 20 psig, and that NMV1 is open. 2. Close the manual valves on the inlet and outlet of the TMSn bubbler. 3. With the vacuum pump running, open the VAC EXHAUST switch on the VACUUM CONTROL panel. Make sure that the CONTROL VALVE OPEN light on the VACUUM CONTROL panel is on. If not press the OPEN button on the AdaptTorr ACR-26 pressure controller to fully open the pressure control valve. Make sure the MANIFOLD H2 switch on the HYDROGEN/NITROGEN control panel is off. Turn the MANIFOLD N2 switch on the HYDROGEN/NITROGENcontrol panel to the on position and adjust the FLOW SET 1 potentiometer to maintain a minimum deposition chamber pressure of 1-2 Torr as indicated on the AdaptTorr pressure controller. 4. On the HYDROGEN/NITROGEN panel set the N2 CARRIER/H2 CARRIER switch to the N2 CARRIER position. 5. On the N-DOPANTcontrol panel turn the ACTIVE switch and the BUBBLER IN switch on. Insert a key and turn the PURGE BUBBLER switch clockwise. The red PURGE BUBBLER light should be on. Push in the black DVM SELECT button and adjust the FLOW SET potentiometer for NDOPMFC1 to a flow of 100 sccm. 6. Evacuation: Insert a key in the CLOSE MFC2 switch and turn it clockwise. This will shut off the gas flow through NDOPMFC2. As the gas in the line is evacuated, the flow rate will drop. When the flow rate for NDOPMFC1 reaches zero as indicated on the panel meter, the circuit is completely evacuated. 7. Backfill: Turn the key in the CLOSE MFC2 switch counter clockwise for 10 seconds. This will allow the bubbler circuit to fill with nitrogen. 8. Repeat steps 6 and 7 twenty times to completely purge the circuit of alkyl vapor. Finish with a backfill step. 9 Turn the BUBBLER IN switch and the ACTIVE switch to the off position. 75 10. Disconnect and remove the bubbler. CAUTION: This is a potentially hazardous step if the manual bubbler valves have failed to seal. The pyrophoric liquid will ignite when exposed to air. The quantities are usually small enough to easily contain the burning alkyl and the oxidation is limited by the diffusion rate of oxygen into the tubing, which is rather slow. Avoid breathing the oxide fumes. Have a dry chemical or non-flammable dry absorbent medium available to smother the chemical fire should it be necessary. In most cases probably the best remedy is to attach the stainless steel VCR caps to seal the inlet and outlet of the bubbler. DO NOT USE WATER!! 11. Install the stainless steel VCR caps on the bubbler ports. Use new gaskets unless the gaskets are of the self centering type. 12. If a new bubbler is being installed immediately, proceed to the next section. Otherwise, shut down the system as appropriate. Bubbler installation 1. Insure that the tank valve for the nitrogen is on, that the pressure regulator on the tank gives a pressure of 20 psig, and that NMV1 is open. 2. Insure that one of the hydrogen cylinders is actively connected to the hydrogen supply line, that the pressure regulator is properly set and that the hydrogen purifier is turned on and up to temperature. 3. Insure that PR6 is set at 10 psig and PR1 is set at 15 psig. 4. With the vacuum pump running, open the VAC EXHAUST switch on the VACUUM CONTROL panel. Make sure that the CONTROL VALVE OPEN light on the VACUUM CONTROL panel is on. If not press the OPEN button on the AdaptTorr ACR-26 pressure controller to fully open the pressure control valve. Make sure the MANIFOLD H2 switch on the HYDROGEN/NITROGEN control panel is off. Turn the MANIFOLD N2 switch on the HYDROGEN/NITROGENcontrol panel to the on position and adjust the FLOW SET 1 potentiometer to maintain a minimum deposition chamber pressure of 1-2 Torr as indicated on the AdaptTorr pressure controller.
80
5. On the HYDROGEN/NITROGEN panel set the N2 CARRIER/H2 CARRIER switch to the H2 CARRIER position. 6. Insure that the manual valves on the bubbler are closed. Uncap the inlet and outlet ports on the bubbler and connect the bubbler to the circuit using new gaskets if the previously used gaskets were not of the self-centering type. 7. On the N-DOPANT control panel turn the ACTIVE switch and the BUBBLER IN switch on. Insert a key and turn the PURGE BUBBLER switch clockwise. The red PURGE BUBBLER light should be on. Push in the black DVM SELECT button and adjust the FLOW SET potentiometer for NDOPMFC1 to a flow of 100 sccm. 8. Evacuation: Insert a key in the CLOSE MFC2 switch and turn it clockwise. This will shut off the gas flow through NDOPMFC2. As the gas in the line is evacuated, the flow rate will drop. When the flow rate for NDOPMFC1 reaches zero as indicated on the panel meter, the circuit is completely evacuated. 9. Backfill: Turn the key in the CLOSE MFC2 switch counter clockwise for 10 seconds. This will allow the bubbler circuit to fill with nitrogen. 10. Repeat steps 8 and 9 twenty times to completely purge the circuit of air. Finish with a backfill step. Remove the key to the CLOSE MFC2 switch. 11. Open the manual valve on the outlet of the bubbler first. If the pressure in the bubbler should happen to be greater than 10 psig (the setting of PR6) and the inlet valve were opened first, the liquid alkyl would be forced back into the inlet plumbing. 12. Open the manual valve on the inlet of the bubbler. 13. Turn the BUBBLER IN switch and the ACTIVE switch to the off position. Turn the PURGE key lock switch off (counter-clockwise) and remove the key. 14. Shut down the system as appropriate. 76
81
Cylinder replacement requires purging of only the high-pressure inlet line to the regulator. The high-purity, low-pressure delivery lines are not exposed to air during the replacement operation. 1. Cylinder removal: a. Tightly close the SiH4 cylinder valve and prepare to remove gas from the inlet line. b. Flow SiH4 through the reactor to reduce the inlet pressure to ""0 psig. c. Close the regulator inlet valve NDOPMV2. d. Lock the excess flow valve NDOPEFV1 in the open position with tape. e. Purge the regulator inlet: Turn on the nitrogen supply valve, NMV3, to the vacuum eductor and open the SiH4 regulator vent valve, NDOPMV4. (DANGER: Be sure the SiH4 cylinder valve is closed.) After 5 seconds close NDOPMV4 and NMV3. f. Backfill the regulator inlet: First insure the tank valve for NDOPPURGEPR1 is open, that the outlet of regulator NDOPPURGEPR1 is at least 15 psig, and that NDOPPURGEMV1 is fully open (clockwise). Backfill the regulator input with nitrogen by opening NDOPMV3 for at least 2 seconds. Close NDOPMV3. g. Repeat the purge-backfill cycle of steps e and f twenty times. Finish with a backfill step, close NDOPMV3 and NDOPMV4,and disconnect and cap the cylinder. The handheld Matheson toxic gas monitorshouldbe usedwhenunscrewingthe cylinder connection. The CGA 350 connection
77
mn__- - --- - --- Hnn__-
on the SiH4 cylinder has a left-hand thread. If you are facing the cylinder, the nut on the excess flow valve must be turned clockwise to loosen it. It is ~important that the brass cap for the cylinder valve be replaced. The cylinder valves sometimes have small leaks even when tightly closed. 2. Cylinder installation: a. With the new SiH4 cylinder sitting inside the MOCVDcabinet and the exhaust fan on, slightly loosen the brass cap which is plugging the end of the CGA 350 connection on the cylinder valve. The cap has a left-hand thread. Check for the presence of SiH4 using the portable Matheson toxic gas detector. If a high level release of SiH4 is detected, the cap on the valve should be retightened and the cylinder returned to the vendor. If a low level release is detected, it may be due to a small amount of arsine trapped between the closed valve on the cylinder and the brass cap. With the brass cap left slightly unscrewed close the door to the MOCVD cabinet and let the tank set for several hours. If no SiH4 is detected at that point, it is safe to completely unscrew the cap and install the cylinder. Otherwise the cap should be re-tightened and the tank returned to the vendor for replacement. b. Connect the excess flow valve, NDOPEFV1,to the cylinder. Keep the regulator inlet valve, NDOPMV2, closed as well as the nitrogen backfill valve NDOPMV3. Hold the excess flow valve, NDOPEFV1, open by temporarily holding in the red button with a piece of tape. c. Purge the regulator input: Turn on the nitrogen flow for the vacuum eductor, by opening NMV4. Open the purge valve, NDOPMV4, wait 5 seconds, and then close NDOPMV4 and NMV4. d. Backfill the regulator input: First insure the tank valve for DOPPURGEPR1 is open, that the outlet of regulator DOPPURGEPR1 is at least 15 psig, and that DOPPURGEMV1 is fully open (clockwise). Backfill the regulator input with nitrogen by opening NDOPMV3 for at least 2 seconds. Close NDOPMV3. e. Repeat the backfill-purge cycle 20 times finishing with a purge step. Insure that NDOPMV3, NDOPMV4, and NMV3 are closed. f. Remove the tape from the excess flow valve, NDOPEFV1. g. Open the manual valve on the tank and check for leaks. h. Reset the excess flow valve, NDOPEFV1, by pushing in the red button and check for leaks. i. Open the manual valve on the regulator input, NDOPMV2and check for leaks. j. Close the tank valve on the nitrogen cylinder supplying gas to regulator DOPPURGEPR1.
- n u- -u- u_- - -- -
82
cylinder regulator, is fed to a mass flow controller supplying this gas to the reactor venVrun manifold. The pressure differential between the source and the reactor chamber is accommodated across the flow controller PDOPMFC1. Alternatively, the DMZ can be valved off and hydrogen passed through the flow controller to purge the dimethylzinc. 2. A purge-mode switchable from the front panel allows Evacuation and backfill of the delivery lines with H2 purge gas. Valving at the regulator allows purging of the cylinder lines for changing the cylinder. 3. The liquid source is dimethylcadmium (DMCd) which is a liquid at room temperature with a vapor pressure of 28 Torr at -200C. This means that the pressure of the vapor is 28/760 x 14.7 psia; about 1/20 atmospheric pressure. The vapor is transported from the stainless steel canister which contains the liquid, into the reaction chamber by bubbling hydrogen gas through the liquid. 2. The P-dopant plumbing system is a more complex arrangement which allows the gas streams from either the high pressure or liquid source to be diluted with additional hydrogen carrier gas. In addition, a portion of the total volume of gas may be dumped into the scrubber before it enters the venVrunvalve. This arrangement permits a wider range of dopant flow rates to be directed into the deposition chamber. The pressure of the hydrogen gas inside the liquid container is determined by the pressure setting of either PR1 or PR3 depending on the mode of operation. The proper functioning of the various components is discussed in detail in the section on operating procedures. 3. The precise vapor pressure of the DMCd is determined by the temperature of the liquid. This is controlled by immersing the liquid in a coolant bath which is controlled at a precise temperature by a small Brinkmann refrigeration unit. 4. Pressure drop to the reaction chamber occurs across the mass flow controller, PDOPMFC1. 5. The alkyl is exposed to the heated sensor element of the mass flow controller (1100C); thus it must be stable at this temperature. 6. Loss of liquid from the bubbler into the gas lines is unlikely because pressure in the gas line is the same as the bubbler. (Closure of the bypass valve, PDOPPV2, is delayed to allow bubbler inlet and outlet pressure to equilibrate.) 7. Upon power failure or other abort conditions, contamination of the upstream H2lines is prevented by closure of the control valve for mass flow controller PDOPMFC2. 8. Adiabatic cooling could possibly cause alkyl condensation in the mass flow controller valve. It is advisable to keep the alkyl temperature below room temperature. 9. A purge mode overrides some interlocksto allow Evacuation and backfill of the bubbler lines
B. Hazards
1. DMCd is pyrophoric 2. The oxides formed from the burning of DMCd are toxic. 3. DMCd reacts violently with water which should never be used to extinguish the burning alkyl. Use a dry chemical fire extinguisher only or else let it burn. 4. The dust formed when dimethylzinc reacts with air is toxic. 5. Release of high pressure DMZ can occur by operator carelessness. 79
C. Precautions
1. Source volume limit: Safety dictated use of a limited volume of DMZ as a means of controlling a catastrophic failure. We have limited ourselves to a 26 cu. ft. cylinder of 5% DMZ in hydrogen. 2. Ventilation: The DMZ cylinder is in an enclosed, ventilated MOCVD reactor cabinet. 3. Toxic gas monitoring: a. The DOD ChemLogic CL8 toxic gas monitor samples the MOCVDcabinet atmosphere. An alarm will activate at the AsH3 TLV and shut down the system. The more sensitive limit for arsine provides extra margin for dimethylzinc detection. b. The DOD ChemLogic CL8 monitor also samples the immediate laboratory air. An alarm will activate at the AsH3 TLV and shut down the system. At the TLV the building Evacuation alarm will sound. (In case of alarm - leave the area and assist others in leaving the building.) c. The DOD ChemLogic CL8 monitors the outlet of the scrubber. If the gas concentration exceeds the arsine TLV threshold, the sources automatically shut down and an alarm on the operator panel
83
sounds. 4. System operation features: a. Interlocks prevent the use of DMZ except under normal operation of the sealed gas system. b. Opening the chamber for loading and unloading samples disables the DMZ. c. Removal of the DMZ cylinder requires purging of the gas regulator inlet line to remove the DMZ prior to disconnecting. d. Replacement of DMZ cylinder requires purging of the gas regulator inlet line to remove air contamination. e. Gas cylinder change should be done with an observer to double-check procedures and assist in an emergency. 5. The gaseous source is a mixture of 5% dimethylzinc (DMZ) in hydrogen. At this concentration, the gas is labeled flammable (not pyrophoric). It is extremely important to purge all the air from the components before installing a dimethylzinc cylinder and, likewise, very important to purge all the dimethylzinc from the lines before removing a dimethylzinc cylinder. If this is not done carefully, some of the gas control components could be ruined and others might require a time consuming cleanup process. An excess flow valve (PDOPEFV1) is installed on the dimethylzinc tank. If the gas flow through the valve exceeds 1000 seem,the valve will trip shutting off the flow. It then stays closed unless it is manually reset by pushing in on the red button. The next valve after the excess flow valve is a pneumatically controlled valve (PDOPPV20) held open by the abort interlock system. This valve automatically closes in the event of power failure, loss of ventilation, operator intervention, fire, low hydrogen pressure, low nitrogen pressure, or detection of toxic gas in the cabinet, room, or at the scrubber outlet. 6. Keep the manual valves on a newly installed bubbler closed until it is connected to the system, leak checked and purged. 7. Avoid gas pressure differentials that can force the alkyl from the bubbler back into the carrier gas piping.
D. Normal Operation
The P-dopant circuit can provide source material from either the bubbler or the high-pressure cylinder. These flows may be operated with or without additional hydrogen dilution. Each mode of operation will be described in a separate section. 1. Start-of-day: Since the pneumatic valves PDOPPV1and PDOPPV2 seal off the bubbler from the rest of the system, the manualvalves on the bubbler inlet and outlet can be left open except when the bubbler is being replaced or removed. At the start-of-day the DMCd is therefore ready to operate from the control panel. 80
-- - ---m --n_- ------ - - ----
The regulator and delivery line contain dimethylzinc and are not purged during overnight shutdown. Since the cylinder valve is closed at days end, the volume of toxic gas left in the lines represents a small risk in case of line rupture. The alternative of purging the line each day would waste a significant fraction of the small cylinder volume and would add to the load of dimethylzinc to be scrubbed from the exhaust gas. 2. Normal growth sequence: Operation of bubbler source without additional hydrogen dilution: With all switches closed, PDOPMFC1, PDOPMFC2, PDOPMFC3, and PDOPMFC4are commanded closed, PDOPPV1, PDOPPV2, PDOPPV3, PDOPPV4, PDOPPV6, PDOPPV7, PDOPPV8, PDOPPV9, PDOPPV10, and PDOPPV11are closed and PDOPPV5 is in the vent position. When the ACTIVE switch is turned on, the signals to command closure of PDOPMFC1and PDOPMFC2 are removed, PDOPPV2, PDOPPV4,and PDOPPV9 open and the green ACTIVE light comes on. A voltage equal to +10 v is applied to the control inputs of PDOPMFC2and PDOPMFC4 which programs their flow settings to 200 seem. This insures that the flow setting through PDOPMFC2or PDOPMFC4will always be larger than the flow through PDOPMFC1and the circuit will not become evacuated; the control valves contained within PDOPMFC2and PDOPMFC4 will remain wide open. With the switches in this configuration, the hydrogen carrier gas will flow through PDOPMFC2,through the bypass valve PDOPPV2,through PDOPPV9,
84
through PDOPMFC1, through PDOPPV4,through PDOPPV5and into the vent manifold at a rate determined by the current value of the voltage input to PDOPMFC1as determined either by the front panel potentiometer or the D/A converter output for this flow controllers. The circuit pressure will be determined by the setting of PR1 and the pressure drop from the circuit to the chamber will occur across PDOPMFC1. During manual operation from the front panel, the flow through PDOPMFC1 is controlled by the setting of the 10-turn FLOW SET 1 potentiometer on the control panel. The value of the flow setting can be read on the digital panel meter by pushing in on the black FLOW SET 1 button. When the button is released, the meter reads the value of the actual flow. When the gas circuit is operated remotely from the computer, the input voltage to PDOPMFC1 is supplied by a D/A converter in the HP Multiprogrammer and can be remotely programmed. The meter on the panel will continue to indicate the actual flow, and when the FLOW SET 1 button is pushed in, the meter will read the input voltage to the flow controller from the D/A converter. With the switches in the present configuration, the red PURGE BUBBLER light on the control panel will be on indicating that the bypass valve PDOPPV2 is open. When the BUBBLER IN switch is turned on, PDOPPV1 and PDOPPV3 open immediatelyand the green light, BUBBLER IN, turns on. After a two-second delay the red PURGE BUBBLER light goes out and PDOPPV2 closes. This insures that the downstream pressure is not greater than the upstream pressure (a condition which would force the liquid alkyl back into the hydrogen carrier gas piping) when PDOPPV1 and PDOPPV3 open. When PDOPPV2 closes, the hydrogen carrier gas is no longer allowed to bypass the bubbler, and instead must flow through the bubbler transporting the alkyl vapor to the venVrunvalve. When the RUN switch is turned on, the yellow RUN light comes on and the gas flow is switched from the vent manifoldto the run manifold and must now pass through the deposition chamber. Shutdown of the system should proceed by reversing the sequence of the above steps. The BUBBLER IN switch should be turned off during standby periods to both conserve the source and to minimize the load of hazardous chemical which enters the scrubber. The alkyl vapor will continue to be transported from the bubbler as long as the BUBBLER IN switch is on and there is gas flow through PDOPMFC1. Operation of bubbler source with additional hydrogen dilution: With all switches closed, PDOPMFC1, PDOPMFC2, and PDOPMFC3are commanded closed, PDOPPV1, PDOPPV2, PDOPPV3, PDOPPV4, PDOPPV6, PDOPPV7, PDOPPV8, PDOPPV9, PDOPPV10,and PDOPPV11 are closed and PDOPPV5 is in the vent position. When the ACTIVE switch is turned on, the signals to command closure of PDOPMFC1 and PDOPMFC2 are removed, PDOPPV2, PDOPPV4,and PDOPPV9 open and the green ACTIV~ light comes on. A voltage equal to +10 v is applied to the control inputs of PDOPMFC2
81
"- n -- -- U --- n H- -U" H"- "-"-- ---n-U
and PDOPMFC4 which programs their flow settings to 200 sccm. This insures that the flow setting for PDOPMFC2 will always be larger than the flow setting for PDOPMFC1and the circuit will not become evacuated; the control valves contained within PDOPMFC2and PDOPMFC4will remain wide open. With the switches in this configuration, the hydrogen carrier gas will flow through PDOPMFC2, through the bypass valve PDOPPV2,through PDOPPV9,through PDOPMFC1, through PDOPPV4,through PDOPPV5 and into the vent manifold at a rate determined by the current value of the voltage input to PDOPMFC1as determined either by the front panel potentiometer or the D/A converter output for this flow controller. The circuit pressure will be determined by the setting of PR1 and the pressure drop from the bubbler to the chamber will occur across PDOPMFC1. During manual operation from the front panel, the flow through PDOPMFC1 is controlled by the setting of the 10-turn FLOW SET 1 potentiometer on the control panel. The value of the flow setting can be read on the digital panel meter by pushing in on the black FLOW SET 1 button. When the button is released, the meter reads the value of the actual flow. When the gas circuit is operated remotely from the computer, the input voltage to PDOPMFC1 is supplied by a D/A converter in the HP Multiprogrammer and can be remotely programmed. The meter on the panel will continue to indicate the actual flow, and when the FLOW SET 1 button is pushed in, the meter will read the input voltage to the flow controller from the D/A converter. With the switches in this configuration the red PURGE BUBBLER light on the
85
control panel will be on indicating that the bypass valve PDOPPV2 is open. The H2 DILUTE switch becomes enabled when the ACTIVE switch is on. It may be turned at any time, however, in a normal growth sequence it would be turned on after the ACTIVE switch is turned on and before the BUBBLER IN switch is turned on. When the H2 DILUTE switch is turned on, PDOPPV6,and PDOPPV7 open, the positive closure signal on PDOPMFC3 is removed, and the +10 v control signal which had been keeping PDOPMFC2fully open is removed. The flow rates through PDOPMFC1, PDOPMFC2, and PDOPMFC3 can now be controlled independently. Hydrogen may be supplied to the circuit through PDOPMFC2 and/or PDOPPV6 and may leave through PDOPMFC1and/or PDOPMFC3. If the flow meters are set correctly, then the flow through PDOPMFC2 is less than the sum of the flows through PDOPMFC1and PDOPMFC3. The pressure in the circuit will then drop until it is less than the pressure at the outlet of PR3. At this point the check valve between PDOPPV6and PR3 will open and the circuit pressure will be maintained at the pressure of PR3. The difference between the exit rate of gas through PDOPMFC1and PDOPMFC3and the entrance rate through PDOPMFC2 will be maintained by the flow through the check valve. The equivalent flow through valve PDOPPV5 is
Flow2 Flow1
Flow 1 + Flow 3
There is nothing to prevent the operator from setting the mass flow controllers such that the flow setting for PDOPMFC2 is greater than the combined flow through PDOPMFC1 and PDOPMFC3. In this case the control valve for PDOPMFC2would remain fully open and the flow through the circuit would be controlled by the settings on PDOPMFC1and PDOPMFC3. The pressure in the bubbler would remain at the outlet pressure of PR1 and flow through the check valve between PDOPPV6and PR3 would not occur. Although no damage would occur from operating the system in this manner, there would be no advantage to doing it. In fact, the amount of alkyl flowing through PDOPMFC3would be wasted with no improvement in the accuracy in controlling the flow of a small amount of alkyl through PDOPMFC1. When the BUBBLER IN switch is turned on, PDOPPV1 and PDOPPV3 open immediately and the green light, BUBBLER IN, turns on. After a two-second delay the red PURGE BUBBLER light goes out and PDOPPV2 closes. The delay on closure insures that the downstream pressure is not greater than the upstream pressure (a condition which would force the liquid alkyl back into the hydrogen carrier gas piping) when PDOPPV1and PDOPPV3 open. When
82
- --- ---on -u-n un-
- - --
PDOPPV2 closes, the hydrogen carrier gas is no longer allowed to bypass the bubbler, and instead must flow through the bubbler transporting the alkyl vapor to the venVrunvalve. When the RUN switch is turned on, the yellow RUN light comes on and the gas flow is switched from the vent manifold to the run manifold and must now pass through the deposition chamber. Shutdown of the system should proceed by turning off, in sequence, the RUN switch, the BUBBLER IN switch, the H2 DILUTE switch, and the ACTIVE switch. The BUBBLER IN switch should be turned off during standby periods to both conserve the source and to minimize the load of hazardous chemical which enters the scrubber. The alkyl vapor will continue to be transported from the bubbler as long as the BUBBLER IN switch is on and there is gas flow through PDOPMFC1and/or PDOPMFC3. Operation of gaseous source without additional hydrogen dilution: The valve on the top of the dimethylzinc cylinder should be opened and the downstream pressure of regulator PDOPPR1 should be checked to see that it is 10 psig. With all switches closed, PDOPMFC1,PDOPMFC2, and PDOPMFC3 are commanded closed, PDOPPV1, PDOPPV2, PDOPPV3, PDOPPV4, PDOPPV6, PDOPPV7, PDOPPV8, PDOPPV9, PDOPPV10,and PDOPPV11 are closed and PDOPPV5 is in the vent position. When the ACTIVE switch is turned on, the signals to command closure of PDOPMFC1 and PDOPMFC2 are removed, PDOPPV2, PDOPPV4,and PDOPPV9 open and the green ACTIVE light comes on. A voltage equal to +10 v is applied to the control inputs of PDOPMFC2 and PDOPMFC4 which programs their flow setting to 200 sccm. This insures that the flow setting for PDOPMFC2and PDOPMFC4will always be larger than the flow setting for
86
PDOPMFC1and the circuit will not become evacuated; the control valves contained within PDOPMFC2 and PDOPMFC4 will remain wide open. With the switches in this configuration, the hydrogen carrier gas will flow through PDOPMFC2,through the bypass valve PDOPPV2, through PDOPPV9, through PDOPMFC1,through PDOPPV4, through PDOPPV5 and into the vent manifold at a rate determined by the current value of the voltage input to PDOPMFC1 as determined either by the front panel potentiometer or the D/A converter output for this flow controller. The circuit pressure will be determined by the setting of PR1 and the pressure drop from the bubbler to the chamber will occur across PDOPMFC1. Duringmanualoperationfrom the front panel, the flow through PDOPMFC1is controlled by the setting of the 10-turn FLOW SET 1 potentiometer on the control panel. The value of the flow setting can be read on the digital panel meter by pushing in on the black FLOW SET 1 button. When the button is released, the meter reads the value of the actual flow. When the gas circuit is operated remotely from the computer, the input voltage to PDOPMFC1 is supplied by a D/A converter in the HP Multiprogrammer and can be remotely programmed. The meter on the panel will continue to indicate the actual flow, and when the FLOW SET 1 button is pushed in, the meter will read the input voltage to the flow controller from the D/A converter. With the switches in this configuration the red PURGE BUBBLER light on the control panel will be on indicating that the bypass valve PDOPPV2 is open. The GASEOUS SOURCE IN switch is disabled when the BUBBLER IN switch is on. When the GASEOUS SOURCE IN switch is turned on (with the BUBBLER IN switch off), PDOPPV9 closes, PDOPPV8 and PDOPPV10open and the signal to command closure is removed from PDOPMFC4. Since a +10 v control signal is applied to PDOPMFC4, its control valve will go to a wide open state. Dopant gas from the high-pressure cylinder will begin to flow through the circuit at a rate determined by the setting for PDOPMFC1. The pressure in the ciruit willbe determinedbythe pressuresettingof PDOPPR1 and the pressure drop will occur across PDOPMFC1.When the RUN switch is turned on, the flow at the outlet of PGPV5 is switched from the vent manifold to the run manifold and the yellow RUN light comes on. The system would be shut down by reversing the sequence of the above steps. Operation of gaseous source with additional hydrogen dilution: The valve on the top of the dimethylzinc cylinder should be opened and the downstream pressure of regulator PDOPPR1 should be checked to see that it is 10 psig. With all switches closed, PDOPMFC1, PDOPMFC2, and PDOPMFC3arecommandedclosed,PDOPPV1,PDOPPV2, PDOPPV3, PDOPPV4,
83
- --
PDOPPV6, PDOPPV7, PDOPPV8, PDOPPV9, PDOPPV10, and PDOPPV11 are closed and PDOPPV5 is in the vent position. When the ACTIVE switch is turned on, the signals to command closure of PDOPMFC1 and PDOPMFC2are removed, PDOPPV2, PDOPPV4,and PDOPPV9open and the green ACTIVE light comes on. A voltage equal to +10 v is applied to the control inputs of PDOPMFC2 and PDOPMFC4 which programs their flow setting to 200 sccm. This insures that the flow setting for PDOPMFC2and PDOPMFC4will always be larger than the flow setting for PDOPMFC1 and the circuit will not become evacuated; the control valves contained within PDOPMFC2 and PDOPMFC4will remain wide open. With the switches in this configuration, the hydrogen carrier gas will flow through PDOPMFC2, through the bypass valve PDOPPV2,through PDOPPV9, through PDOPMFC1, through PDOPPV4, through PDOPPV5 and into the vent manifold at a rate determined by the current value of the voltage input to PDOPMFC1as determined either by the front panel potentiometer or the DIA converter output for this flow controller. The circuit pressure will be determined by the setting of PR1 and the pressure drop from the bubbler to the chamber will occur across PDOPMFC1. During manual operation from the front panel, the flow through PDOPMFC1 is controlled by the setting of the 10-turn FLOW SET 1 potentiometer on the control panel. The value of the flow setting can be read on the digital panel meter by pushing in on the black FLOW SET 1 button. When the button is released, the meter reads the value of the actual flow. When the gas circuit is operated remotely from the computer, the input voltage to PDOPMFC1is supplied by a D/A converter in the HP Multiprogrammer and can be remotely programmed. The meter on the panel will continue to
87
indicate the actual flow, and when the FLOW SET 1 button is pushed in, the meter will read the input voltage to the flow controller from the D/A converter. With the switches in this configuration the red PURGE BUBBLER light on the control panel will be on indicating that the bypass valve PDOPPV2 is open. The GASEOUS SOURCE IN switch is disabled when the BUBBLER IN switch is on. When the GASEOUS SOURCE IN switch is turned on (with the BUBBLER IN switch off), PDOPPV9 closes, PDOPPV8 and PDOPPV10open and the signal to command closure is removed from PDOPMFC4. Since a +10 v control signal is applied to PDOPMFC4, its control valve will go to a wide open state. Dopant gas from the high-pressure cylinder will begin to flow through the circuit at a rate determined by the setting for PDOPOMFC1. The pressure in the ciruit will be determined by the pressure setting of PDOPPR1and the pressure drop will occur across PDOPMFC1. The H2 DILUTE switch becomes enabled when the ACTIVE switch is on. It may be turned at any time, however, in a normal growth sequence it would be turned on after the ACTIVE and the GASEOUS SOURCE IN switches are turned on. When the H2 DILUTE switch is turned on, PDOPPV6, and PDOPPV7open, the positive closure signal on PDOPMFC3 is removed, and the +10v control signal which had been keeping PDOPMFC4fully open is removed. The flow rates through PDOPMFC1,PDOPMFC3,and PDOPMFC4can now be controlled independently. Hydrogen may be supplied to the circuit through PDOPMFC4and/or PDOPPV6 and may leave through PDOPMFC1and/or PDOPMFC3. If the flow meters are set correctly, then the flow through PDOPMFC4will be less than the sum of the flows through PDOPMFC1and PDOPMFC3. The pressure in the circuit will then drop until it is less than the pressure at the outlet of PR3. At this point the check valve between PDOPPV6 and PR3 will open and the circuit pressure will be maintained at the pressure of PR3. The difference between the exit rate of gas through PDOPMFC1and PDOPMFC3and the entrance rate through PDOPMFC4 will be maintained by the flow through the check valve. The equivalent flow through
valve PDOPPV5 is
Flow4 Flow1
Flow 1 + Flow 3
84
- -- - -- no - n- - - no nn _nn_-n-n__- --
There is nothing to prevent the operator from setting the mass flow controllers such that the flow setting for PDOPMFC4 is greater than the combined flow through PDOPMFC1 and PDOPMFC3. In this case the control valve for PDOPMFC4would remain fully open and the flow through the circuit would be controlled by the settings on PDOPMFC1and PDOPMFC3. The pressure in the bubbler would remain at the outlet pressure of PR1 and flow through the check valve between PDOPPV6 and PR3 would not occur. Although no damage would occur from operating the system in this manner, there would be no advantage to doing it. In fact, the amount of source gas flowing through PDOPMFC3would be wasted with no improvement in the accuracy of controlling the flow through PDOPMFC1. When the RUN switch is turned on, the yellow RUN light comes on and the gas flow is switched from the vent manifold to the run manifold and must now pass through the deposition chamber. Shutdown of the system should proceed by turning off, in sequence, the RUN switch, the H2 DILUTE switch, the GAS SOURCE IN switch, and the ACTIVE switch. 4. End of day: The alkyl vapors will accelerate the deterioration of some of the elastomers from which the o-rings are fabricated. Elastomer seals have been eliminated whenever possible in the construction of this system but current technology does not allow them to be completely eliminated. They are present in the mass flow controller and in the venVrun valve. For this reason, if the bubbler source has been used, it is good practice to thoroughly flush the alkyl vapor from PDOPMFC1, PDOPMFC2, PDOPMFC3, and PDOPPV5 before the circuit is shut down. This can be accomplished by allowing hydrogen to flow through PDOPMFC1, PDOPMFC2, and PDOPMFC3for approximately 10 minutes after the BUBBLER IN switch is closed. During the purging step the ACTIVE and H2 PURGE switches must be on and there should be a gas flow of at least 10 secm through PDOPMFC1and 100 sccm through PDOPMFC2 and PDOPMFC3. After this, the ACTIVE switch can be closed, sealing off the circuit.
88
F. Interlocks
1. All source valves are deactivated if the following conditions exist: a. System is in abort status. b. The vacuum exhaust valve, EXPV2,is closed. c. The deposition chamber is up to atmosphere. d. The manifold diluent gas is off. 2. The BUBBLER IN, GAS SOURCE IN, RUN, and H2 DILUTE switches are disabled if the ACTIVE switch is off. 3. The BUBBLER IN switch is disabled if the GAS SOURCE IN switch is on. This implies that PDOPPV1 and PDOPPV3 can not be opened unless PDOPPV4 is open and PDOPPV8 and PDOPPV10 are closed. 4. The GAS SOURCE IN switch is disabled if the BUBBLER IN switch is on. This implies that PDOPPV8 and PDOPPV10 can not be opened unless PDOPPV4 is open and PDOPPV1 and PDOPPV3 are closed.
85
-- - -- --- - -- u -u --u
-- - -- - - ----
5. With the PURGE switch on the front panel in the locked position (key removed), PDOPPV2 cannot be opened if PDOPPV1 and PDOPPV3 are open. 6. The bypass valve (PDOPPV2) is open if the BUBBLER IN switch is closed (handle down) and the ACTIVE switch is on (handle up). 7. The ACTIVE, BUBBLER IN, GAS SOURCE IN, H2 DILUTE, and RUN switches on the front panel are disabled when the computer switches the circuit into the REMOTE mode. 8. When the bubbler is connected to the circuit by opening the BUBBLER IN switch (handle up), valves PDOPPV1 and PDOPPV3 open immediately. Valve PDOPPV2 closes after a 2 second time delay to insure that the pressure on each side of the bubbler is equalized 9. PDOPMFC1, PDOPMFC2, PDOPMFC3, and PDOPMFC4are commanded closed if the ACTIVE switch is off. The key lock switch CLOSE MFC2 when turned clockwise will close PDOPMFC2 under all conditions. The key lock switch CLOSE MFC4 when turned clockwise will close PDOPMFC4 under all conditions. 10. When the H2 DILUTE switch is off, PDOPMFC3is commanded closed and the set points of PDOPMFC2 and PDOPMFC4are at 200 sccm. The purpose of setting the PDOPMFC2and PDOPMFC4flows to 200 secm is to insure that their flow settings are always greater that the flow setting for PDOPMFC1 when H2 DILUTE is off. This prevents the circuit from becoming evacuated. When the H2 DILUTE switch in on, the closure signal is removed from PDOPMFC3 and the set point voltages for PDOPMFC2and PDOPMFC4are determined by either their front panel potentiometers or their remote D/A converters. A closure signal on a mass flow controller will over-ride the set point signal. 11. The manifold diluent gas cannot be turned off if the ACTIVE switch is on. 12. If the GAS SOURCE IN switch is on (handle up), then the H2 FLUSH switch is disabled. This prevents PDOPPV11from opening if PDOPV10 is open. 13 All front panel switches with the exception of the PURGE BUBBLER and H2 FLUSH switches are disabled when the computer commands the circuit to the REMOTE mode.
89
4. RUN switch: (Pdop_run_on, Pdop_run_off) 5. H2 DILUTE switch: (Pdop_h2_on, Pdop_h2_off) 6. GAS SOURCE IN switch: (Pdop_gas_on, Pdop_gas_off) 7. Programming of mass flow controllers: Set-p_dop_da !The same DA converters are used for both the n-dopant and p-dopant !systems. This command connects the DA converters to the p-dopant !system. Set-pdop_1 (Flow) Flow given in sccm (0~Flow~200) SeCpdop_2(Flow) Flow given in sccm (O~Flow~200) SeCpdop_3(Flow) Flow given in sccm (0~Flow~2000) SeCpdop- 4(Flow) Flow given in sccm (0~Flow~200 To read flow; Read_pdop(Flow) Returns a floating point number,Flow, equal to the actual flow in sccm Read_pdop_1(Flow) Read_pdop_2(Flow) Read_pdop_3(Flow) Read_pdop-4(Flow) To set flow: A typical programming sequence for flow from the bubbler with no hydrogen dilution would be as follows:
86
Pdop_remote_on SetJXIop_1 (6.5) Pdop_active_on Pdop_bub_on Pdop_run_on !Program the flow rate for PDOPMFC1 to 6.5 sccm
90
Pdop_bub_on Pdop_run_on Deposition occurs Pdop_run_off !The flow to the deposition chamber is diverted to the vent !manifold Pdop_bub_off Pdop_h2_off Pdop_active_off SeCpdop_1(0.0) SeCpdop_2(0.0) SeCpdop_3(O.O) Pdop_active_off Pdop_remote_off !Turns off all gas flows !These commands are mostly for housekeeping so that a mass !flow controller isn't later used at an unintended setting A typical programming sequence for flow from the high-pressurE!source without hydrogen dilution would be:
87
Pdop_remote_on Pdop_active_on Pdop_gas_on Setpdop_1 (230) Pdop_run_on !Program the flow rate to 230 sccm !DMZbegins to enter the chamber
Depositionoccurs
Pdop_run_off Set_pdop_1 (0.0) Pdop_gas_off Pdop_active_off Pdop_remote_off
91
Pdop_run_on Deposition occurs Pdop_run_off !The flow to the deposition chamber is diverted to the vent !manifold Pdop_h2_off Pdop_gas_off Pdop_active_off SeCpdop_1 (0.0) SeCpdop_3(O.O) Set_pdop- 4(0.0) Pdop_active_off Pdop_remote_off !Turns off all gas flows !These commands are mostly for housekeeping so that a mass !flow controller isn't later used at an unintended setting 88
-- - -- --- --m_-
92
oxide fumes. Have a dry chemical or non-flammable dry absorbent medium available to smother the chemical fire should it be necessary. In most cases probably the best remedy is to attach the stainless steel VCR caps to seal the inlet and outlet of the bubbler. DO NOT USE WATER!! 11. Install the stainless steel VCR caps on the bubbler ports. Use new gaskets unless the gaskets are of the self centering type. 12. If a new bubbler is being installed immediately, proceed to the next section. Otherwise, shut down the system as appropriate. 89
-- --- -no-on ---on -n__--- n -- -- -- - -- --
Bubbler installation
1. Insure that the tank valve for the nitrogen is on, that the pressure regulator on the tank gives a pressure of 20 psig, and that NMV1 is open.
2. Insure that one of the hydrogen cylinders is activelyconnected to the hydrogen supply line,
that the pressure regulator is properly set and that the hydrogen purifier is turned on and up to temperature. 3. Insure that PR3 is set at 10 psig and PR1 is set at 15 psig. 4. With the vacuum pump running, open the VAC EXHAUST switch on the VACUUM CONTROL panel. Make sure that the CONTROL VALVE OPEN light on the VACUUM CONTROL panel is on. If not press the OPEN button on the AdaptTorr ACR-26 pressure controller to fully open the pressure control valve. Make sure the MANIFOLD H2 switch on the HYDROGEN/NITROGEN control panel is off. Turn the MANIFOLD N2 switch on the HYDROGEN/NITROGEN control panel to the on position and adjust the FLOW SET 1 potentiometer to maintain a minimum deposition chamber pressure of 1-2 Torr as indicated on the AdaptTorr pressure controller. 5. On the HYDROGEN/NITROGEN panel set the N2 CARRIER/H2 CARRIER switch to the H2 CARRIER position. 6. Insure that the manualvalves on the bubbler are closed. Uncap the inlet and outlet ports on the bubbler and connect the bubbler to the circuit using new gaskets if the previously used gaskets were not of the self-centering type. 7. On the P-DOPANTcontrol panel turn the ACTIVE switch and the BUBBLER IN switch on. Insert a key and turn the PURGE BUBBLER switch clockwise. The red PURGE BUBBLER light should be on. Push in the black DVM SELECT button and adjust the FLOW SET potentiometer for PDOPMFC1 to a flow of 100 sccm. 8. Evacuation: Insert a key in the CLOSE MFC2 switch and turn it clockwise. This will shut off the gas flow through PDOPMFC2. As the gas in the line is evacuated, the flow rate will drop. When the flow rate for PDOPMFC1 reaches zero as indicated on the panel meter, the circuit is completely evacuated. 9. Backfill: Turn the key in the CLOSE MFC2 switch counter clockwise for 10 seconds. This will allow the bubbler circuit to fill with nitrogen. 10. Repeat steps 8 and 9 twenty times to completely purge the circuit of air. Finish with a backfill step. Remove the key to the CLOSE MFC2 switch. 11. Open the manual valve on the outlet of the bubbler first. If the pressure in the bubbler should happen to be greater than 10 psig (the setting of PR3) and the inlet valve were opened first, the liquid alkyl would be forced back into the inlet plumbing. 12. Open the manual valve on the inlet of the bubbler. 13. Turn the BUBBLER IN switch and the ACTIVE switch to the off position. Turn the PURGE key lock switch off (counter-clockwise)and remove the key. 14. Shut down the system as appropriate.
93
90
--- -- -- --- ----
2. With the vacuum pump running, open the VAC EXHAUST switch on the VACUUM CONTROL panel and set the N2 manifold flow controller to maintain a minimum deposition chamber pressure of 1-2 Torr. 3. Evacuation: On the N DOPANT control panel open the ACTIVE switch, the GAS SOURCE IN switch, and the RUN switch. Adjust PDOPMFC1for maximumflow. Insert a key and turn the key lock PURGE MFC4 switch clockwise. Allow several minutes for the line to evacuate. 4. Backfill: Turn the GAS SOURCE IN switch off. Turn the H2 DILUTE switch on. Insert a key in the key lock H2 FLUSH switch and turn it clockwise for at least 10 seconds to backfill the line with hydrogen. Turn the H2 FLUSH switch counterclockwise and turn the H2 DILUTE switch off. Turn the GAS SOURCE IN switch on. 5. Repeat the evacuate cycle allowing several minutes for the line to evacuate. 6. Repeat the backfilVevacuatecycle a total of five times. After the last evacuate cycle, set PDOPMFC1to zero flow, turn the PURGE MFC4 switch to the off position and remove the key. Turn the H2 FLUSH switch to the off position and removethe key.. 7. Backfill with DMZ: a. Back-off the knob on pressure regulator PDOPPR1 (fully counterclockwise) b. Open the high-pressure regulator inlet valve PDOPMV2. c. Set the knob on pressure regulator PDOPPR1to 10 psig at the outlet. Check the excess-flow valve PDOPEFV1to see that it has not tripped by pushing in on the red button and observing the gauge on the outlet of PDOPPR1 for any changes in pressure. d. Close the RUN SWITCH, the GAS SOURCE IN switch and the ACTIVE switch. Cylinder Replacement Cylinder replacement requires purging of only the high-pressure inlet line to the regulator. The high-purity, low-pressure delivery lines are not exposed to air during the replacement operation. 1. Cylinder removal: a. Tightly close the DMZ cylinder valve and prepare to remove gas from the inlet line. b. Flow DMZ through the reactor to reduce the inlet pressure to '" 0 psig. c. Close the regulator inlet valve PDOPMV2. d. Lock the excess flow valve PDOPEFV1 in the open position with tape. e. Purge the regulator inlet: Turn on the nitrogen supply valve, NMV3, to the vacuum eductor and open the DMZ regulator vent valve, PDOPMV4. (DANGER: Be sure the DMZ cylinder valve is closed.) After 5 seconds close PDOPMV4 and NMV3. f. Backfill the regulator inlet: First insure the tank valve for PDOPPURGEPR1is open, that the outlet of regulator PDOPPURGEPR1 is at least 15 psig, and that PDOPPURGEMV1 is fully open (clockwise). Backfill the regulator input with nitrogen by opening PDOPMV3 for at least 2 seconds. Close PDOPMV3. g. Repeat the purge-backfill cycle of steps e and f twenty times. Finish with a backfill step, close PDOPMV3 and PDOPMV4, and disconnect and cap the cylinder. The handheld Matheson toxic gas monitor should be used when unscrewing the cylinder connection. The CGA 350 connection on the DMZ cylinder has a left-hand thread. If you are facing the cylinder, the nut on the excess flow valve must be turned clockwise to loosen it. It is ~important that the brass cap for the cylinder valve be replaced. The cylinder valves sometimes have small leaks even when tightly closed. 2. Cylinder installation: a. With the new DMZ cylinder sitting inside the MOCVD cabinet and the exhaust fan on, slightly loosen the brass cap which is plugging the end of the CGA 350 connection on the cylinder valve. The cap has a left-hand thread. Check for the presence of DMZ using the portable Matheson toxic gas detector. If a high level release of DMZ is detected, the cap on the valve should be retightened and the cylinder returned to the vendor. If a low level release is detected, it may be due to a small amount of arsine trapped between the closed valve on the cylinder and the brass cap. With the brass cap left slightly unscrewed close the door to the MOCVDcabinet and let the tank set for several hours. If no DMZ is detected at that point, it is safe to completely unscrew
91
the cap and install the cylinder. Otherwise the cap should be re-tightenedand the tank returned
94
to the vendor for replacement. b. Connect the excess flow valve, PDOPEFV1,to the cylinder. Keep the regulator inlet valve, PDOPMV2,closed as well as the nitrogen backfill valve PDOPMV3. Hold the excess flow valve, PDOPEFV1, open by temporarily holding in the red button with a piece of tape. c. Purge the regulator input: Turn on the nitrogen flow for the vacuum eductor, by opening NMV2. Open the purge valve, PDOPMV4,wait 5 seconds, and then close PDOPMV4 and NMV2. d. Backfill the regulator input: First insure the tank valve for DOPPURGEPR1 is open, that the outlet of regulator DOPPURGEPR1 is at least 15 psig, and that DOPPURGEMV1 is fully open (clockwise). Backfill the regulator input with nitrogen by opening PDOPMV3 for at least 2 seconds. Close PDOPMV3. e. Repeat the backfill-purge cycle 20 times finishing with a purge step. Insure that PDOPMV3, PDOPMV4, and NMV3 are closed. f. Remove the tape from the excess flow valve, PDOPEFV1. g. Open the manual valve on the tank and check for leaks. h. Reset the excess flow valve, PDOPEFV1, by pushing in the red button and check for leaks. i. Open the manualvalve on the regulator input, PDOPMV2 and check for leaks. j. Close the tank valve on the nitrogen cylinder supplying gas to regulator
95
Vacuum Control
The vacuum control system includes a number of pneumatic valves and interlocks which allow the chambers in the MOCVD system to be evacuated and which at certain times purge portions of the system with nitrogen to prevent backstreaming of the vacuum pump oil. This section begins with a description of the front panel controls and indicators.
VACUUM EXHAUST
VACUUM LATCH
ENTRY CHAMBER UP TO AIR ENTRY CHAMBER EXHAUST LOAD LOCK OPEN LOAD LOCK CLOSED SCRUBBER OXIDATION
96 Nitrogen gas automatically flows into the system at one of three selected ports to insure that the pressure at the vacuum pump inlet does not drop below 100 microns Hg. The purpose of this is to prevent the backstreaming of oil from the pump to the deposition and sample entry chambers. The port that is selected depends on the state of the vacuum exhaust valves and the pressure measured by the AdapTorr pressure controller. Indicates that nitrogen is flowing into the port at the base of the spinner shaft. This flow serves two functions. It insures a minimum flow through the deposition chamber to prevent oil backstreaming. It also floods the upper opening of the bellows assembly with nitrogen so as to prevent the flow of source gases into the bellows assembly. If the source gases were to enter the bellows, it is possible that material could be deposited inside the bellows assembly. This lamp will be on and EXPV3 will open if valve EXPV2to the vacuum pump is open and there is adequate pressure in the nitrogen supply line and the vacuum pump is turned on and the pressure is below the setting of Trip Point 2 (vacuum in chamber) and the interlock system is not in abort status. The flow rate is controlled by manual valve EXMV3. Indicates that nitrogen is flowing into the sample entry chamber. This lamp comes on and EXPV4 will be open if EXPV2 is closed and there is adequate pressure in the nitrogen supply line and the vacuum pump is turned on and valve EXPV6 to the vacuum pump is open and the interlock system is not in abort status. The flow rate is controlled by EXMV4. Indicates that nitrogen is flowing into the particle filter on the inlet of the vacuum pump. The lamp comes on and EXPV5 will be open if there is adequate pressure in the nitrogen supply line and the vacuum pump is on and EXPV2 is closed and EXPV6 is closed. The flow rate is controlled by EXMV5.
DEP CHAMBER
PUMP INLET
97
Switches
VAC EXHAUST ENTRY CHAMBER UP TO AIR This switch opens valve EXPV2 between the deposition chamber and the vacuum pump. The switch is disabled when the panel is in the REMOTE mode. This switch opens EXPV9 which allows a high flow of nitrogen gas to enter the sample entry to rapidly bring it up to atmospheric pressure. The gas flow rate is controlled by EXMV9. This switch is disabled if either valve EXPV6 between the sample entry chamber and the vacuum pump is open or the gate valve between the deposition chamber and the sample entry chamber is open. This switch opens EXPV6 between the sample entry chamber and the vacuum pump. EXPV6 will open only if valve EXPV2 between the vacuum pump and the deposition chamber is closed. However, once EXPV6 is opened, EXPV2 can be opened and EXPV6 will not close automatically. The reason for this interlock is to prevent air from being introduced into the deposition chamber. If it is desired to have both EXPV2 and EXPV6open at the same time, such as when a sample is being loaded into the deposition chamber, then it is necessary to first evacuate the sample entry chamber with the valve to the deposition chamber closed. When the sample entry chamber is evacuated then both valves can be opened simultaneously. This is a three position momentary contact switch (ON-OFF-ON, normally in the off state) which controls the position of the gate valve between the deposition chamber and the sample entry chamber. When used, the switch should be held in position until the appropriate lamp comes on. The gate valve opens when the switch is pushed upward. This function is disabled unless valve EXPV2 to the deposition chamber is open and valve EXPV6 to the sample entry chamber is open. This prevents the valve from being actuated if there is a pressure differential across it. The gate valve closes when the switch is pushed downward. This function is disabled unless the push rod for sample loading is pulled completely out of the path of the gate valve and hooked to its retainer. This switch opens AUXPV2 which allows air to enter the inlet of the scrubber so than the adsorbed gases and vapors can be oxidized. The switch is disabled if any liquid or vapor source is turned on or if the hydrogen manifold diluent is turned. This prevents hydrogen and air from flowing through the system at the same time. Conversely, if the SCRUBBER OXIDATION switch is on, the interlock system is in CLevel status which prevents any of the valves for the sources or the hydrogen manifold diluent from being turned on. The flow rate of air through AUXPV2 is controlled by the setting of AUXMV2 and indicated by the adjacent rotameter which gives a full scale indication of 300 sccm.
SCRUBBER OXIDATION
98
Normal Operation
A normal sequence of operations for loading a substrate, making a deposition, and removing the substrate is as follows. It is assumed that all switches on the VACUUM CONTROL panel are closed. 15. Turn on the ENTRY CHAMBER UP TO AIR switch. When the sample entry chamber is at atmospheric pressure the glass cover can be easily lifted off. Remove the cover and turn off the ENTRY CHAMBER UP TO AIR switch. 16. Place the sample on the loading surface and replace the cover on the chamber. 17. Turn on the ENTRY CHAMBER EXHAUST switch. Wait about 5 minutes for water vapor to be desorbed from the walls of the chamber. 18. Open the VAC EXHAUST switch. Wait one minute. 19. Push the LOAD LOCK OPEN/CLOSED switch upward to open the gate valve. 20. Load the sample onto the susceptor. 21. Remove the sample pushrod from the deposition chamber and make sure its end is attached to the retaining hook. 22. Push the LOAD LOCK OPEN/CLOSED switch downward to close the gate valve. 23. Close the ENTRY CHAMBER EXHAUST switch. 24. Proceed with the deposition. 25. When the deposition is finished, all gases are turned off, the susceptor has cooled, and the susceptor has been lowered; open the ENTRY CHAMBER EXHAUST switch. 26. Push the LOAD LOCK OPEN/CLOSED switch upward to open the gate valve. 27. Use the pushrod to transport the substrate from the deposition chamber to the sample entry chamber. Hook the end of the pushrod to its retainer. 28. Push the LOAD LOCK OPEN/CLOSED switch downward to close the gate valve. 29. Close the ENTRY CHAMBER EXHAUST switch and the VAC EXHAUST switch. 30. Open the ENTRY CHAMBER UP TO AIR switch, remove the cover on the sample entry chamber and turn off the ENTRY CHAMBER UP TO AIR switch. 31. Remove the sample, replace the cover on the chamber and open the ENTRY CHAMBER EXHAUST switch. 32. Wait 5 minutes and close the ENTRY CHAMBER EXHAUST switch.
99
Interconnections
33. 34. 35. 36. 37. 38. VAC EXHAUST controls EXSV2 which drives EXPV2. ENTRY CHAMBER UP TO AIR controls EXSV9 which drives EXPV9. ENTRY CHAMBER EXHAUST controls EXSV6 which drives EXPV6. LOAD LOCK OPEN controls EXSV7 which drives one side of the pneumatic cylinder for the gate valve. LOAD LOCK CLOSED controls EXSV8 which drives the other side of the pneumatic cylinder. SCRUBBER OXIDATION controls AUXSV2 which drives AUXPV2.
100
Interlocks
39. If the pressure in the deposition chamber falls below the setting of Trip Point 2 (vacuum in chamber), the vacuum latch relay EXK1 becomes activated and the front panel lamp VACUUM LATCH turns on. If the pressure in the chamber rises above the setting of Trip Point 2, the latch stays engaged until either the vacuum valve EXPV2 closes or the pressure control valve closes completely. 40. The CHAMBER VAC LOSS interlock will activate and place the interlock system in Abort status if the chamber pressure rises above the setting for Trip Point 1 (atmospheric pressure in chamber) with the VACUUM LATCH activated and the valve, EXPV2, between the deposition chamber and the vacuum pump open. 41. Nitrogen will automatically enter the deposition chamber, DEP CHAMBER lamp will be on and EXPV3 will open if valve EXPV2 to the vacuum pump is open and there is adequate pressure in the nitrogen supply line and the vacuum pump is turned on and the pressure is below the setting of Trip Point 2 (vacuum in chamber) and the interlock system is not in abort status. 42. Nitrogen will automatically enter the sample entry chamber, the SAMPLE ENTRY CHAMBER lamp will come on and EXPV4 will be open if EXPV2 is closed and there is adequate pressure in the nitrogen supply line and the vacuum pump is turned on and valve EXPV6 to the vacuum pump is open and the interlock system is not in abort status. 43. Nitrogen will automatically enter the particle filter on the inlet of the vacuum pump, the PUMP INLET lamp will come on and EXPV5 will be open if there is adequate pressure in the nitrogen supply line and the vacuum pump is on and EXPV2 is closed and EXPV6 is closed. 44. The VAC EXHAUST switch is disabled when the panel is in the REMOTE mode. 45. The ENTRY CHAMBER UP TO AIR switch is disabled if either valve EXPV6 between the sample entry chamber and the vacuum pump is open or the gate valve between the deposition chamber and the sample entry chamber is open. 46. The ENTRY CHAMBER EXHAUST switch which opens valve, EXPV6, will be enabled only if valve EXPV2 between the vacuum pump and the deposition chamber is closed. However, once EXPV6 is opened, EXPV2 can be opened and EXPV6 will not close automatically. 47. The LOAD LOCK OPEN function is disabled unless valve EXPV2 to the deposition chamber is open and valve EXPV6 to the sample entry chamber is open. 48. The LOAD LOCK CLOSED function is disabled unless the push rod for sample loading is pulled completely out of the path of the gate valve and hooked to its retainer. 49. The SCRUBBER OXIDATION switch is disabled if any liquid or vapor source is turned on or if the hydrogen manifold diluent is turned on. Conversely, if the SCRUBBER OXIDATION switch is on, the interlock system is in C-Level status which prevents any of the valves for the sources or the hydrogen manifold diluent from being turned on.
101
102
103
to both EXPV5 and EXPV2. This will keep EXPV2 open when nitrogen is flowing through EXPV5 so that the pressure can be measured. Turn on the power to the control cabinet and the vacuum pump. 57. Insure that the tank valve for the nitrogen is on, that the pressure regulator on the tank gives a pressure of 20 psig, and that NMV1 is open. 58. With the vacuum pump running, all the switches on the VACUUM CONTROL panel should be closed. Make sure that the CONTROL VALVE OPEN light on the VACUUM CONTROL panel is on. If not press the OPEN button on the AdaptTorr ACR-26 pressure controller to fully open the pressure control valve. Make sure the MANIFOLD H2 and MANIFOLD N2 switches on the HYDROGEN/NITROGEN control panel are off. 59. Close the needle valve EXMV5 located in the vacuum pump compartment. 60. Quickly make a note of the pressure reading on the AdapTorr pressure controller and proceed to Step 6. The total amount of time the pump is left running with no gas flow into the input should not be more than a minute or so. 61. Open EXMV5 until the pressure as indicated on the AdapTorr pressure controller increases by five Torr. Count the number of turns of the valve needed to do this. 62. Use the graph at the end of this section to determine the amount of valve opening (CV value) for the number of turns measured in Step 6. Divide this Cv value by 10 and determine from the graph determine the number of turns open needed to achieve this. Reset EXMV5 to this value. This should give a system pressure of about 0.5 Torr or 500 microns Hg. 63. Turn off the vacuum pump and the power to the control cabinet. Reconnect the control lines for EXPV2 and EXPV5 in their normal configuration. Turn the power to the control cabinet and the vacuum pump back on. Sample entry chamber: 64. Turn off the vacuum pump and the power to the control cabinet. Remove the 1/8" O.D. pneumatic control tubing from EXPV6. Use a 1/8" union tee and lengths of 1/8" O.D. tubing to connect the control line for EXPV4 to both EXPV4 and EXPV2. This will keep EXPV2 open when nitrogen is flowing through EXPV4 so that the pressure can be measured. Turn on the power to the control cabinet and the vacuum pump. 65. Insure that the tank valve for the nitrogen is on, that the pressure regulator on the tank gives a pressure of 20 psig, and that NMV1 is open. 66. With the vacuum pump running, open the ENTRY CHAMBER EXHAUST switch. Make sure that the CONTROL VALVE OPEN light on the VACUUM CONTROL panel is on. If not press the OPEN button on the AdaptTorr ACR26 pressure controller to fully open the pressure control valve. Make sure the MANIFOLD H2 and MANIFOLD N2 switches on the HYDROGEN/NITROGEN control panel are off. 67. Close the needle valve EXMV410 located in the main gas control cabinet. 68. Quickly make a note of the pressure reading on the AdapTorr pressure controller and proceed to Step 6. The total amount of time the pump is left running with no gas flow into the input should not be more than a minute or so. 69. Open EXMV4 until the pressure as indicated on the AdapTorr pressure controller increases by five Torr. Count the number of turns of the valve needed to do this. 70. Use the graph at the end of this section to determine the amount of valve opening (CV value) for the number of turns measured in Step 6. Divide this Cv value by 10 and determine from the graph determine the number of turns open needed to achieve this. Reset EXMV4 to this value. This should give a system pressure of bout 0.5 Torr or 500 microns Hg. 71. Turn off the vacuum pump and the power to the control cabinet. Reconnect the control lines for EXPV2 and EXPV4 in their normal configuration. Turn the power to the control cabinet and the vacuum pump back on.
104
105
SPINNER
RAISE SUSCEPTOR
LOWER SUSCEPTOR
106
107
Induction Heater
The substrate is heated by inductively coupling a 10kHz signal into a graphite susceptor. Approximately 1 kW can be fed into the susceptor at maximum power. The power supply unit has a 20 A circuit breaker on the front panel which allows it to be turned off independently of the controls on the substrate control panel. The output power from the induction heating power supply is proportional to the value of the control signal received from the temperature controller. The power supply and the coil which surrounds the susceptor are water cooled. The power supply unit contains a pressure switch which determines if the water pressure is adequate. If it is not, then the heating circuit will not turn on and the red front panel lamp will be on. The power supply also contains a solenoid valve which turns on the water automatically when the power is turned on. The relative amount of power delivered by the power supply is indicated on the front panel meter. This meter measures the power delivered by the capacitor filter circuit to the switching circuit. The total power delivered to load will be less than this by the amount of loss in the power switching transistors and in the work coil. The power supply does not contain an isolation transformer so it should be assumed that every point in the circuitry has the full 120 VAC line voltage applied. This includes the work coil which should always be covered with insulating material to reduce risk to the operator. The work coil carries a peak current of about 140 A at full power so it is essential that all connections be made securely. The substrate heating can be turned on and off remotely but the set point temperature must be entered manually into the front panel of the temperature controller. Assuming the tuning procedure for the controller has already been performed, the set point can be adjusted by pressing only the (YES), (NO) arrows on the controller to raise (lower) the setpoint. Consult the operator manual for the controller for further details on its operation and for tuning procedures. The temperature of the substrate can be measured by remote control and recorded in the program output. The same thermocouple used for temperature control is connected to Channel 1 of the low voltage A/D converter in slot 403C of the Multiprogrammer.