100% found this document useful (1 vote)
156 views

Function Description System

This document provides an overview of the MAGNETOM ESSENZA magnetic resonance imaging system. It describes the system components, including the magnet, gradient, and radiofrequency systems. It also discusses the patient handling systems, system control components, power and cooling infrastructure, and software. The document is intended to familiarize service technicians with the overall structure and key parts of the MRI system.

Uploaded by

aytek ertunc
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
156 views

Function Description System

This document provides an overview of the MAGNETOM ESSENZA magnetic resonance imaging system. It describes the system components, including the magnet, gradient, and radiofrequency systems. It also discusses the patient handling systems, system control components, power and cooling infrastructure, and software. The document is intended to familiarize service technicians with the overall structure and key parts of the MRI system.

Uploaded by

aytek ertunc
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 20

g

tE
u
lo
a
c
d
M
s
n
m
ie
S

MAGNETOM
MR

Function Description
System

System

MAGNETOM ESSENZA
10130587

© Siemens 2006
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

Print No.: M4-060.850.10.01.02 English


Replaces: n.a. Doc. Gen. Date: 12.06
2 Revision / Disclaimer
1Revision / Disclaimer

Document revision level


The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.

Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise autho-
rized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with
or authorized by Siemens or one of its affiliates are directed to contact one of the local
offices of Siemens or one of its affiliates before attempting installation or service proce-
dures.

MAGNETOM M4-060.850.10.01.02 Page 2 of 20 Siemens


12.06 CS SD MR Medical Solutions
Table of Contents 3
0Table of Contents

1 _______ Introduction ____________________________________________________ 4

Documentation structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Hardcopy documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
CD-ROM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

2 _______ System introduction _____________________________________________ 6

System overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
System components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Magnet system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Gradient system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
RF system. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Patient handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
System control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
System supply (mains, cooling water) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

3 _______ List of acronyms and abbreviations _______________________________ 14

4 _______ Changes to previous version _____________________________________ 20

Siemens M4-060.850.10.01.02 Page 3 of 20 MAGNETOM


Medical Solutions 12.06 CS SD MR
4 Introduction
1-
Documentation structure
1Introduction

This prefix is intended to familiarize you with the structure of the MAGNETOM ESSENZA.
It explains the print format as well as the on-line documentation and how to use it.

Hardcopy documentation 0

The hardcopy service documentation will be issued in the same DIN A4 portrait format and
comprise a Functional Description (FUN), the Troubleshooting Guide (TSG) and the
Replacement of Parts (REP).

Changes to previous version


The last part of each chapter shows the changes to the previously released version. You
will find the topic, section or graphic which was changed together with the corresponding
page number.

CD-ROM 0

As a consistent feature, the documentation has been put onto electronic media
(CD/DVD-ROM) for convenience and fast access to the information needed by you.
The electronic documents are identical to the hardcopy versions and available in both
HTML and PDF formats. To read the HTML files, you will need a browser installed on your
service PC. The recommended browser is MS Internet Explorer. You have to be able to
read the PDF files in the Adobe Acrobat Reader. A copy of this reader is also included on
the service CD/DVD.
Navigation in and between the various electronic documents is made possible with hyper-
links provided either within the text or placed in many of the graphics. Text links are recog-
nizable by their BLUE color. If the cursor changes to a "hand" when placed over a graphic,
then you can click it - its a link.

Documents on the CD
Files included in PDF format:
• Functional Description
• Troubleshooting Guide
• Replacement of Parts
• Planning Guide
• Installation
• Start-up
• Diagrams
• Maintenance Instructions
• Safety Information

MAGNETOM M4-060.850.10.01.02 Page 4 of 20 Siemens


12.06 CS SD MR Medical Solutions
Introduction 5

Service PC
You might be asking yourself "how do we access all this information?" We think we have
good news for you. As a service engineer, you will be supplied with a portable service PC
(also known as the "notebook", or "laptop"). This service PC will have been especially pre-
pared for you to meet all your service needs. As a standard configuration, it includes a
CD/DVD-ROM drive as well as a graphics adapter with at least 800 by 600 pixel resolution.
You will also be supplied with all the software you need to perform your service responsi-
bilities: A modern, ultra-high-performance and stable operating system from Microsoft, an
Internet browser of your choice (as long as it is Internet Explorer), and a very user-friendly
text editor (Notepad or better), as well as administrative programs to help you keep track of
your hours and expenditures, and much, much more!

Documentation access at the MR system


The electronic documentation can also be used at the MR system after you insert your
CD/DVD into the CD/DVD-ROM drive of the host.

Siemens M4-060.850.10.01.02 Page 5 of 20 MAGNETOM


Medical Solutions 12.06 CS SD MR
6 System introduction
2-
System overview
2System introduction

The MAGNETOM ESSENZA was developed from the ground up to be an affordable, reli-
able powerhouse that will support both clinical and financial success.
This ultra-short, lightweight 1.5T system delivers the flexibility, accuracy and speed of Tim
(Total imaging matrix), as well as a powerful new innovation: the IsoCenter Matrix coil,
which is a spine array fixed in the bottom part of the body coil.
The key features of the MAGNETOM ESSENZA are:
• 1.5 Tesla
• Most affordable, all new
• Tim [25x8]
• Head to toe applications, with a great bottom line

The block diagram below shows an overview of the MAGNETOM ESSENZA hardware
components.
The main functions of the MAGNETOM ESSENZA are as follows:
• Generation of the basic magnetic field (magnet)
• Generation of the gradient field (gradient)
• Generation and measurement of the radio-frequency field (RF)
• Patient handling and patient monitoring
• Measurement and reconstruction system (MARS)
• System supply (mains, cooling water)
• Software
The functions described above are executed via the respective system components.

MAGNETOM M4-060.850.10.01.02 Page 6 of 20 Siemens


12.06 CS SD MR Medical Solutions
System introduction 7
Block diagram MAGNETOM ESSENZA
Fig. 1:
Siemens M4-060.850.10.01.02 Page 7 of 20 MAGNETOM
Medical Solutions 12.06 CS SD MR
8 System introduction

System components 2.1

Magnet system 0

The magnet system generates the basic magnetic field. The components required are
described below. The electrical and mechanical shim are integrated into the gradient coil.

Magnet - MAG
The magnet system comprises the superconductive magnet, including the systems for
cooling (interface for helium fill/refill, cold head), energizing (fixed current leads) and mon-
itoring the magnet during operation.
The magnet is cooled by a conventional 4K cold head which allows helium to recondense
and reduces helium boil-off to zero during operation.

Magnet supervision - MSUP


This component provides the following: it supervises and controls the magnet during oper-
ation, it ramps it down in emergency situations (ERDU), and it provides user-accessible
operating controls for the magnet system.

Magnet covers - COV


The magnet covers consist of the front, back and side covers, the turret covers, the aper-
tures including all mechanisms required for installing the covers, as well as the components
located directly underneath the covers. The magnet cover is designed to provide the mag-
net with a self-supporting, noise absorbent enclosure. The equipment required for the tun-
nel lighting, patient ventilation, and light localizer is also incorporated into the magnet
covers.

E-shim
The component required to supply the shim coil for adjusting the A(2,0) term is located in
the E-shim power supply (1channel). Please be aware that only one channel is available
(not five, as in other systems). The E-shim is a standard component in all MAGNETOM
ESSENZA systems.

Passive shim
The passive shim includes the shim trays, the hardware to attach and install them, as well
as the set of shim plates and the insulation between the iron shims.

Magnet power supply - MPS


The magnet power supply consists of the unit which energizes and de-energizes the mag-
net, including the charge cable to the magnet, the cabling for the voltage supply, and Mag-
net Supervision. As a service tool, the MPS is not part of the delivery volume of the MR
system and is brought on-site only as required.

MAGNETOM M4-060.850.10.01.02 Page 8 of 20 Siemens


12.06 CS SD MR Medical Solutions
System introduction 9

Array shim device - ASD


The array shim device is a service and installation tool for magnet shimming. It consists of
such components as mechanical supports, MR probes, preamplifier, multiplexer, etc., used
to measure the magnetic field via the shim plot. As a service tool, the array shim device is
not part of the delivery volume of the MR system and is brought on-site only as required.

Gradient system 0

In addition to the basic magnetic field, the gradient system generates linear dynamic mag-
netic field gradients in the three spatial orientations. The gradient system includes the gra-
dient power amplifier (GPA) and the gradient coil.

Gradient power amplifier - GPA


The gradient power amplifier consists of the components necessary for generating and reg-
ulating the gradient currents. This includes the power line filter, the gradient power supply
lines, and the electronics for monitoring the temperature of the gradient coils. Unlike in the
previous 1.5T systems, the GPA forms part of the Electronic CAbinet (ECA) rather than
using a cabinet of its own.

Gradient coil - GC
The gradient coil comprises the coil systems for generating the gradient fields, the system
for cooling the coil, as well as the sensors for temperature monitoring plus the hardware for
installing the coil in the magnet. The gradient coil also includes the compartments for pas-
sive shimming of the magnet's main field and the A(2,0) shim coil.

RF system 0

The RF system generates the RF-fields and transmits them to the patient (transmit func-
tion). It also receives the MR signals emitted by the patient and provides the signals in dig-
ital format for image reconstruction (receive system).

RF small signal init - RFSU


The RFSU converts the digitally modulated transmit signal from the PCI_STAR (control
unit) into the required system frequency in the range of 63.6MHz and sends it to the RFPA.
Parallel to this, the RFSU receives the analog MR signals from the examination, mixes
them down to an intermediate frequency of about 1MHz, and performs an analog-to-digital
conversion before they are sent to the PCI_RX.

RF power amplifier
The RFPA amplifies the RF signal generated in the RF small signal unit to the amplitude
necessary for exciting hydrogen nuclei (1H). The RFPA generates the RF power required
for the body resonator. For diagnostic applications, the defining parameters are peak and
continuous output as well as the reproducibility of pulse patterns, i.e., temporal stability,
since these parameters represent the prerequisite technical conditions for realizable exam-
ination protocols and the reproducibility of examination results. Currently, it is identical to
the DORA used in the Symphony and Avanto systems.

Siemens M4-060.850.10.01.02 Page 9 of 20 MAGNETOM


Medical Solutions 12.06 CS SD MR
10 System introduction

RF application system
The RF application system functions as a switchboard in the examination room for the var-
ious RF channels in the system. It also functions as a preamplifier for the MR signals
(RCCS_60) and decouples the RF measurement signals at the resonator via the integrated
measurement unit. The RF cables routed to the resonators as well as the transmit lines
from the filter plate to the BCCS (which is identical to Avanto) are also part of the RF appli-
cation system. The functional scope of the RFAS includes the measurement unit for decou-
pling the RF measurement signals at the resonator and the TAS on the filter plate.

RF infrastructure - RFIS_60
Within the scope of the RF system, the water-cooled RFIS_60 provides the RF compo-
nents in the examination room with the necessary control signals and operating voltages.
The RFIS_60 is located directly at the magnet.

RF filter - RFFIL
MR examinations are performed in an RF-shielded room or cabin which protects against
external interference and unacceptable RF exposure. This room as well as the RF-sealed
feed-throughs for the cables are part of the RF filter plate.

Body coil - BC
The system includes a whole-body coil for emitting RF to the patient and for receiving MR
signals from the patient. The body coil also includes the tuning unit for adjusting the reso-
nator to the power amplifier based on the frequency. The support tube of the body resona-
tor also serves as the inner lining of the magnet bore.

IsoCenter matrix coil- ICM


This new design of a fixed spine array coil is permanently mounted into the body coil bore
tube and will be underneath the patient table when the patient is positioned inside the mag-
net.
Subsequently, no large spine array coil has to be carried around by the radiographer and
the coil is always perfectly centered.

Local coils - LC
RF local coils improve image quality in examinations of specific body regions. The range of
available coils is constantly expanded as new applications are developed. Approximately
80% of all examinations are performed with head and spine coils. However, orthopedic
coils as well as special coils are utilized when examining other areas, such as the rectum
or peripheral vessels.

Patient handling 0

The system handling function covers all requirements and includes such components as
the patient table, lighting, ventilation, and patient monitoring. Prior to the MR examination,
the patient is positioned on the table; the positioning accessories are used for additional
comfort. The patient is then moved into the magnet bore. During the examination, the
patient is monitored accordingly, via video camera or other means.

MAGNETOM M4-060.850.10.01.02 Page 10 of 20 Siemens


12.06 CS SD MR Medical Solutions
System introduction 11

Patient table - PTAB


The patient table is used to position as well as move the patient into the magnet bore. The
motor-assisted table movement is controlled via the front control unit located at the left side
of the magnet. The patient table is designed for partial body applications of up to 1.4m. The
total horizontal movement range is 2.05m.
The MAGNETOM ESSENZA patient table is available in two versions:
• Standard version without vertical drive
• Optional version with vertical drive

Intercom - COM
Since the patient is usually alone in the examination room during the measurements, the
patient and the physician communicate via the intercom. The intercom consists of the the
examination room components, the operating console components, and the necessary
cabling.

Patient monitoring - VID


Whenever required (when there is no direct face-to-face interaction between physician and
patient such as through a window), the patient is monitored during the examination via a
video camera. The setup consists of the camera, the monitor, and the cabling.

System control 0

System control comprises all of the system's computer functions: host computer, image
reconstruction, measurement control, and physiological signal acquisition. The system
control is the main communications interface for the overall system. It converts examination
protocols or user input to the corresponding control signals and transmits these signals to
the relevant components via a main control bus. It also monitors the functional status of the
various subcomponents and processes these accordingly for system monitoring.

Host computer - HOST


The hardware comprising the host computer includes all processor components required
for patient management, image handling (image computation, display, post-processing,
evaluation, hardcopy documentation, and archiving), and measurement sequence pro-
gramming. This includes all elements of the processor's user interface: monitors, satellite
console, power distribution MRC, keyboards, mouse, data carriers, etc.

Measurement And Reconstruction System - MARS


This new component contains the MR Image Reconstruction system (MRIR) and the
Advanced Measurement Control (AMC) in one single computer system.
The image reconstruction system computes the final images from the digitized receive sig-
nals of the RF small signal unit.
The measurement control controls the peripheral components within the time frame set for
the MR measurement.

Siemens M4-060.850.10.01.02 Page 11 of 20 MAGNETOM


Medical Solutions 12.06 CS SD MR
12 System introduction

Physiological Measurement Unit - PMU


The optional PMU acquires the ECG signals, the pulse, and the respiration of the patient
and forwards them to the MARS. The PMU electronics are installed within the patient table
electronics and comprise all sensors, voltage supplies, and cables for communication with
the MARS. The sensors are connected to the PMU via a wireless connection.

In-room MRC
With the optional in-room MRC, you can start measurements inside the examination room
and check the images immediately.

System supply (mains, cooling water) 0

Power distribution system - PDS


The power distribution system supplies the components with electrical power. The power
distributor ensures that the system is connected as specified at a central point to the power
supply side of the system. The power distributor provides the different components with
individually fuse-protected supply voltages. The main power distributor is located in the
main electronics cabinet, which includes space for installing additional components.

Integrated cooling system - ICS


The main power-consuming components of the system are water-cooled. Cooling is per-
formed with water at a temperature of 20 C. This keeps condensation from building up on
the cooled electronics components and eliminates the need for complicated insulation
measures, as long as basic values for temperature and humidity are maintained.
The ICS exists in two versions:
• ICS passive: a regulated water-water heat exchanger which discharges lost heat via
an existing cold water pipe network (on-site primary cold water).
• ICS active: the water can be tempered using a specially designed water cooler with an
external fan unit (EFU) which does not require any water cooling supply by the cus-
tomer.
An air conditioning system (ACS) built into the ICS provides air cooling for the components
in the ECA and a dedicated air cooling supply for the GPA.

System cabling - CABLE


The system cabling includes all control, signal, data, communications, and supply cables
that have to be attached to the system during on-site installation.

Software 0

syngo MR is the software application package. The software is based on syngo, the com-
mon software for all Siemens Medical systems. The software includes the system, applica-
tion, and measurement controller software. Only firmware modules directly assigned to
hardware components are not included. The system software includes functions required
by the user and technical service. These functions are:

MAGNETOM M4-060.850.10.01.02 Page 12 of 20 Siemens


12.06 CS SD MR Medical Solutions
System introduction 13

• Execute and interpret dialog-driven interactions in multi-user and multi-tasking func-


tions
• Organize and control the measurement system
• Acquire measurement data as well as preprocess and reconstruct images
• Display image data as well as results on the monitor of the operating console
• Post-process images for statistical and geometric evaluation
• Document and archive images and post-processing results
• Test system functions and monitor individual system components

Siemens M4-060.850.10.01.02 Page 13 of 20 MAGNETOM


Medical Solutions 12.06 CS SD MR
14 List of acronyms and abbreviations
3-
3List of acronyms and abbreviations

1000Base-T 1000 Mbit per second, Base-d on twisted pair cable


100Base-TX 100 Mbit per second, Base-d on twisted pair cable
10Base-T 10 Mbit per second, Base-d on twisted pair cable
3D 3 (three) dimensional
ABX Alarm box
ACS Air Conditioning System
ADC Analog-to-Digital Converter
AGP Accelerated Graphics Port
AS Active Shielded (Coil)
ASD Array Shim Device
ATMU Air Temperature Measurement Unit (Patient Table)
AVA Actual Value Amplifier (Gradient)
BC Body Coil
BCCS Body Coil Channel Selector
BLOB Binary Large OBjects
BIOS Basic Input Output System
BPL BackPLane
BU Business Unit
CAN Controller Area Network
CABLE system cabling
CB-DOC Computer Based DOCumentation
CCU Current Converter Unit
CCW Counter ClockWise
CD-R CD Recorder
COB Can Open Bridge
Codec Compressor and decompressor (of data)
COM COMponent communication
COM Patient interCOM
CONT CONTroller (Patient Table)
COOL COOLing system
COSLIO Can (CAN) Open SLIO
COV COVers

MAGNETOM M4-060.850.10.01.02 Page 14 of 20 Siemens


12.06 CS SD MR Medical Solutions
List of acronyms and abbreviations 15

cPCI Compact PCI


CPU Central Processing Unit
CS Customer Service
CSE Customer Service Engineer
CSI Chemical Shift Imaging
CV Context Vision (software option) or Coil Voltage (in RF-System)
CW Control Word or ClockWise
DAC Digital-to-Analog Converter
DCF Device Configuration File
DDR SDRAM Double Data Rate Synchronous Dynamic RAM
DDS Direct Digital Synthesizer
DIMM Dual Inline Memory Modules
DICO DIrectional COupler
DORA DOuble resonant Radio-frequency Amplifier
DSL Digital Subscriber Line
DSP Digital Signal Processor
DSV Diameter Spherical Volume
DUCY DUty CYcle (Gradient)
DVD Digital Versatile Disk
DYSCON Dynamic Switch Control
ECA Electronic CAbinet
ECC Eddy Current Compensation
ECG ElectroCardioGram
ECL Enhanced Current Lead (Magnet)
EFU External Fan Unit
EIS External Interference Shield (Magnet)
EPR Electronic Patient Record
ERDU Emergency Run Down Unit
E-Shim Electronic Shim (only A(2,0) term)
ETL Erdu (ERDU) Test Load
FCL Fixed Current Lead (Magnet)
FOC Fiber Optic Cable

Siemens M4-060.850.10.01.02 Page 15 of 20 MAGNETOM


Medical Solutions 12.06 CS SD MR
16 List of acronyms and abbreviations

FOV Field of View


FPGA Field Programmable Gate Array
FRU Field Replacable Unit
FSA Fixed Spine Array --> ICM
FUN FUnction(al) Description
GC Gradient Coil
GPA Gradient Power Amplifier
GSSU Gradient Small Signal Unit
GSWD Gradient Safety WatchDog (stimulation monitor)
HDD Hard Disk Drive
HIPAA Health Insurance Portability and Accountability Act
HSC Headquarters Support Center
ICS Integrated Cooling System
ICE Image Calculation Environment
ICM IsoCenter Matrix (=FSA)
IDE Integrated Device Equipment
IDEA Integrated Development Environment for Application
IEEE Institute of Electric and Electronic Engineers
IGBT Insulated Gate Bipolar Transistor
iPAT integrated Parallel Acquisition Technique
ISDN Integrated Services Digital Network
LAN Local Area Network
LC Local Coil
LCD Liquid Crystal Display
LPD Line Power Distribution
LVD Low Voltage Differential
LWL Line With Light (i.e. fiber optic cable) --> FOC
MAG MAGnet
MARS Measurement And Reconstruction System
MC Measurement Control
MDH Measurement Data Header
ME Magnet Electronics

MAGNETOM M4-060.850.10.01.02 Page 16 of 20 Siemens


12.06 CS SD MR Medical Solutions
List of acronyms and abbreviations 17

MH_SEL Measurement Hybrid SELect


MOCO MOtion COrrection
MOFI MOdulator FIlter (filter) Function
MPS Magnet power supply
MPSU Magnet Power Supply Unit
MR Magnetic Resonance
MRC MR main Console
MREF Magnet REFrigerator (i.e. helium compressor)
MRSC MR Satellite Console
MSUP Magnet SUPervision
MWS Mantelwellensperre (= "RF Trap" in English)
NCO Numerical Controlled Oscillator
NOE Nuclear Overhauser Effect
NTP Network Time Protocol
OEM Original Equipment Manufacturer
OEMLC OEM Local Coil
OR60 1.5T magnet type for Essenza systems (SMT nomenclature)
OS Operating System
OVC Outer Vacuum Chamber (Magnet)
PALI Power Absorption LImiter
PAT Parallel Acquisition Technique
PCB Printed Circuit Board
PCI Peripheral Component Interconnect bus
PCIe PCI Express bus
PCI-X eXtended PCI Bus
PDAU Physiological Data Acquisition Unit
PDS Power Distribution System
PERU Physiological Ecg (ECG) and Respiratory Unit
PID Proportional-, Integrational-, Differential- (Regulator-Types)
PMU Physiological Measurement Unit
PORA Power Optimized Radio-Frequency Amplifier
PPU Peripheral Pulse Unit

Siemens M4-060.850.10.01.02 Page 17 of 20 MAGNETOM


Medical Solutions 12.06 CS SD MR
18 List of acronyms and abbreviations

PS/2 Personal System /2


PSHIM Passive SHIM
psia Pounds per square inch absolute (pressure)
psig Pounds per square inch relative (pressure)
PTAB Patient TABle
PWRD Power Driver (Patient Table)
RAID Redundant Array of Independent Disks
RAM Random Access Memory
RCCS_60 Receive Coil Channel Selector for aXXess60 (=Essenza project name)
REP REplacement of Parts
RF Radio Frequency
RFAS RF Application System
RFFIL RF FILter
RFIS_60 RF InfraStructure for aXXess60 (=Essenza project name)
RFPA RF Power Amplifier
RFSU RF Signal Unit
RFSWD Radio Frequency Safety WatchDog
ROFT Read Out Fourier Transformation
RSC Regional Support Center
SCA Single Connection Adapter
SCSI Small Computer System Interface
SCT Sequence Control Task
SD Shim Device
SeSo Service Software
SGA Switched Gain Amplifier
SHARC Super Harvard ARchitecture Computer
SHS SIEMENS Healthcare System
SLIO Serial Linked Input/Output
SMT Siemens Magnet Technology
SODIMM Small Outline DIMM
S/PDIF Sony/Phillips Digital Interconnect Format
SRS Siemens Remote Services

MAGNETOM M4-060.850.10.01.02 Page 18 of 20 Siemens


12.06 CS SD MR Medical Solutions
List of acronyms and abbreviations 19

STIMO STImulation MOnitor


SUPF SUPplied Functions (Patient table electronics)
SVS Single Voxel Spectroscopy
SW SoftWare
SWT Switch Terminal (Patient Table)
TALES Transmit Antenna LEvel Sensor
TAS Transmit Antenna Selector
Tim Total imaging matrix
TSG Troubleshooting Guide
TULI TUnnel LIght (Patient Table)
USB Universal Serial Bus
USC Uptime Service Center
VCC Voltage Common Collector (+5V supply voltage)
VGA Video Graphics Array
VID VIDeo (camera for patient monitoring)
VPN Virtual Private Network

Siemens M4-060.850.10.01.02 Page 19 of 20 MAGNETOM


Medical Solutions 12.06 CS SD MR