Operator Manual
Operator Manual
CT/e
Operator Manual
CT/e
Operator Manual, English
5126652-100
Revision:6
© 2006 General Electric Company
All rights reserved.
WARNING
X-RAY EQUIPMENT IS DANGEROUS TO BOTH PATIENT AND OPERATOR
UNLESS MEASURES OF PROTECTION ARE STRICTLY OBSERVED
Though this equipment is built to the highest standards of electrical and mechanical safety, the useful
x-ray beam becomes a source of danger in the hands of the unauthorized or unqualified operator.
Excessive exposure to x-radiation causes damage to human issue.
Therefore, adequate precautions must be taken to prevent unauthorized or unqualified persons from
operating this equipment or exposing themselves or others to its radiation.
Before operation, persons qualified and authorized to operate this equipment should be familiar with
the Recommendations of the International Commission on Radiological Protection, contained in
Annals Number 26 of the ICRP, and with applicable national standards.
Chapter 1
SYSTEM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Overview of CT System and Examination . . . . . . . . . . . . . . . . . . . 1
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
General Procedures of CT Examination : . . . . . . . . . . . . . . . . 1
Main Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Gantry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Gantry Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Gantry Display Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Operator Console (OC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
User Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Mouse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Standard Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Optional Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Head Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Body Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
System Power On/Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
HIPAA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Shut Down and Start Up the System . . . . . . . . . . . . . . . . . . . . . . 26
Login and Logout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Configure Users for the System . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Chapter 2
SCAN TOP LEVEL SCREEN . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Icon Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Function of Each Icon 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
New Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Patient Schedule 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Select Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
View More Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Add Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Edit Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Delete Selected . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Delete All . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Preferences . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Protocol Management . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Auto Voice Record 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Protocol Management 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Retro Recon 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Chapter 3
DISPLAY TOP LEVEL SCREEN . . . . . . . . . . . . . . . . . . . . . ... 1
Primary/Secondary Viewport . . . . . . . . . . . . . . . . . . . . . . . . ... 2
Focus Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 2
Image Display Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 3
Layouts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 3
Autoview Layouts 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 4
Review Layouts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 6
Multiple Image Display (MID) . . . . . . . . . . . . . . . . . . . . . ... 6
List/Select 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 7
Routine Display 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 9
Roam / Zoom . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 10
Explicit Magnify . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 11
Flip/Rotate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 12
ProView . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 13
GSE (Gray Scale Enhancement) . . . . . . . . . . . . . . . . . . ... 14
Display Normal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 14
List / Select . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 15
Ellipse ROI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 15
Measure Distance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 16
Grid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 16
Cross Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 17
User Annotation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 18
Exam Pg / Series Pg . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 19
Hide/Show Graphics . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 20
Erase . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 20
Screen Save . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 20
Chapter 4
SCAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Patient Positioning 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
New Patient 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Scout Scan 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Icons on Scout prescription screen . . . . . . . . . . . . . . . . . . . . . 7
Auto Store . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Auto Transfer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Axial/Helical Scan Prescription 1 . . . . . . . . . . . . . . . . . . . . . . . . . 13
Axial/Helical Prescription (View/Edit) Screen . . . . . . . . . . . . . . 13
Continuous Scan Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Prescribe Scan Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Prospective Multiple Reconstruction . . . . . . . . . . . . . . . . . . . . 21
Prospective Multiple Reconstruction . . . . . . . . . . . . . . . . . . . . 22
Icons of the Axial/Helical Scan Prescription screen (1) . . . . . . 23
Icons of the Axial/Helical Scan Prescription screen (2) . . . . . . 28
Show Localizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Show Localizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Auto mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Auto mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Auto mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Auto mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Auto mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Axial/Helical Scan in Progress . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Axial/Helical Scan End . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Priority Recon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Next Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Repeat Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
One More . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Repeat Last Group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Chapter 5
FILMING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 1
AutoFilm 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 1
Start New Sheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 5
Continue Same Sheet . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 5
Cancel Film Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 5
AutoFilm Composer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 10
Manual Film . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 11
Manual Film Composer 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 12
Image removal from Film Composer . . . . . . . . . . . . . . . . ... 16
Load images by drag and drop . . . . . . . . . . . . . . . . . . . . ... 16
Load images by F1 function key . . . . . . . . . . . . . . . . . . . ... 16
Page filming by F2 function key . . . . . . . . . . . . . . . . . . . ... 17
MID (Multiple Image Display) filming by F3 function key ... 17
Series filming by F4 function key . . . . . . . . . . . . . . . . . . ... 18
Chapter 6
QUALITY ASSURANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 1
Phantom Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 2
QA Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 2
Phantom Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 3
Scan the QA Phantom . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 4
Test and Analysis of the Phantom Images . . . . . . . . . . . ... 4
Contrast Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 5
High Contrast Spatial Resolution . . . . . . . . . . . . . . . . . . . . . ... 6
Slice Thickness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 7
Positioning Light Accuracy (optional) . . . . . . . . . . . . . . . . . . ... 8
Low Contrast Detectability . . . . . . . . . . . . . . . . . . . . . . . . . . ... 9
Noise and Uniformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 10
Typical Results and Allowable Variations . . . . . . . . . . . . . . . ... 11
Contrast Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 11
High Contrast Spatial Resolution . . . . . . . . . . . . . . . . . . ... 11
Nominal Slice Thickness . . . . . . . . . . . . . . . . . . . . . . . . . ... 11
Low Contrast Detectability . . . . . . . . . . . . . . . . . . . . . . . ... 11
Noise and CT Number of Water . . . . . . . . . . . . . . . . . . . ... 11
Weighted CTDI100 (CTDIW) . . . . . . . . . . . . . . . . . . . . . . . . . ... 12
Dose and Performance 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 13
Chapter 7
TECHNICAL SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . 1
Component Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Component Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
X-Ray Tube Assembly Information . . . . . . . . . . . . . . . . . . . . . . . . 3
Diagnostic Source Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Leakage Technique Factors . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Minimum Inherent Filtration . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
X-Ray Tube IEC Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
X-ray Tube Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
X-ray Tube Insert . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
X-ray Tube Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
X-ray Tube Insert Information . . . . . . . . . . . . . . . . . . . . . . . . . 5
Nominal Anode Input Power . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Table 1
General Notice
o Designation of Peripheral Equipment Connectable to This Product
Avoid using other equipment than designated. Failure to comply with this
instruction may result in poor EMC performance of the product.
o Notice against User Modification
Never modify this product. Unilateral user modification may cause degradation
in EMC performance. Modifications of product include ;
• Changes in cable ( length, material, wiring etc.)
• Changes in system installation/layout
• Changes in system configuration/components
• Changes in means of fixing the system/parts ( cover open/close, cover screw-
ing )
o Operate the system with all covers closed. If you open any cover for some
reason, be sure to shut it before starting/resuming operation. Operating the
system with any cover open could affect EMC performance.
General countermeasures
Electromagnetic interference with other equipment
• Electromagnetic interference may be alleviated by positioning other equip-
ment far from the system.
• Electromagnetic interference could be mitigated by changing the relative loca-
tion/installation angle between the system and other equipment.
• Electromagnetic interference may be eased by changing wiring locations of
power/signal cables of other equipment.
• Electromagnetic influence could be reduced by altering the path of power sup-
ply for other equipment.
• Electromagnetic environment specified below Table 3 and Table 4.
3,5
d =[ ] P 80 MHz to 800 MHz (see Table 4)
3
Notice on Service
o Ensure all screws are tight after servicing.
Loose screws may cause the degradation in EMC performance.
o In case the high frequency gasket of this system is broken, replace it with a
new one immediately.
WARNING!
FAILURE TO FOLLOW THE OPERATING INSTRUCTIONS AND SAFETY
PRECAUTIONS COULD RESULT IN INJURY TO THE PATIENT, YOURSELF,
OR OTHERS.
General Safety
o Keep the patient in view at all times. Never leave the patient unattended.
Always stay alert to safety concerns involving the patient's condition and
equipment operation.
o Check the lengths of all patient health lines (IV tubing, oxygen lines, etc.)
and make sure they accommodate cradle travel. Position these lines so
they cannot catch on anything within the patient vicinity or between the
table and gantry during cradle travel or gantry tilt.
o Always follow the exam procedures provided in the operator manual.
Verify correct entry of identification, patient positioning, and other patient
data before proceeding with the exam. Incorrect procedures or patient
data entry could cause misinterpretation of the exam results.
o Conduct Image Quality checks and follow the maintenance schedule
outlined in your CT/e operator manual. Discontinue use of the equipment
whenever you notice equipment damage, or a malfunction occurs. Do not
use the equipment until qualified service personnel correct the problem.
Never use the equipment unless all the protective covers are in place.
Radiation Safety
o Use of controls or adjustments, or performance of procedures other than
those specified herein may result in hazardous radiation exposure.
o Always use proper technique factors for each procedure to minimize X-
Ray exposure while still producing the best diagnostic results.
o Be sure to have an operator wear X-ray protection vest when a work is
needed near the gantry while X-ray is being emitted.
Software Safety
o Should a malfunction occur, or a patient condition develop that requires
interrupting a scan series, push (Abort) on the right end of the OC
scan panel to stop X-Ray.
o Wait for image annotation to complete before filming. If you press the
exposure button too soon, the film records a composite of two different
scans.
o DICOM protocol has a "dialect", which may cause some troubles like
disappearances of some portion of image annotations, when connecting to
the station where data transfer is not confirmed.
Emergency Stop
o In the event of a hardware failure that could cause serious damage, such as
Motion artifacts in large vessels may be caused when using any manufacturer’s
CT scanner capable of scanning with 1 second or less rotation times. The
pulsation of the vessel creates a double margin of the vessel emulating a
dissection. The motion artifact is an interaction between the pulsation of the
vessel and the rotation time of the scan.
Please ensure that the appropriate personnel in your CT Department are made
aware of this notice. If you have any questions regarding this notice, please
contact your local GE representative.
If you would like to review additional information in the clinical literature please
review the following publication: Gotway, Michael: Helical CT evaluation of
the thoracic aorta. Applied Radiology: Sept. 2000, 7-28.
Labels on Keyboard
o The following labels are attached to the upper side of the keyboard.
CAUTION
CAUTION
WARNING
This X-ray unit may be dangerous to patient and operator
unless safe exposure factors, instructions and maintenance
schedules are observed. To be used by authorized
personnel only.
MODEL xxxxxxx
S/N yyyyyyy CLASS 1
MANUFACTURED
DESC GANTRY
SOURCE: 30A/200V~50/60Hz
SOURCE: 30A/115V~50/60Hz
SOURCE: 15A/115V~50/60Hz
SOURCE: 50A/380~480V3~50/60Hz
CAUTION
Accessory Caution
o The following labels are attached to the standard headholder and shallow
head holder.
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
Packing Materials
The materials used to pack our equipment are recyclable. They must be
collected and processed in accordance with the regulations in force for the
country where the machines or accessories are unpacked.
2. As the operator initiates the scan procedures with the patient in the gantry
aperture, the X-ray unit inside the gantry rotates around the patient body
emitting the fan-shaped X-ray beam to the body. The X-ray penetrates the
body and is received by the detector, then it is converted to the electrical
data.
3. The data acquired in the above step is processed into the images by the
computer at the OC. Then, the images are initially displayed on the Monitor
of the OC and subject to manipulation for analysis. The images can also be
filmed later for diagnoses by a physician.
4. The image data can be saved in some electronic media like optical disk for
later analysis and also be transferred to some other stations for remote
diagnoses.
Gantry
Breath Navi
Control Panel
Positioning Positioning
Light Light
Function
The gantry incorporates the X-ray tube unit, the HiLight detector and DAS (Data
Acquisition System) inside. It also provides the following functions.
• Display Panel
The display panel shows the readings of the gantry tilt, table height, position of
land mark, latch status, scannable range and tilt range.
Refer to the Gantry Display Panel page for the function descriptions.
• Emergency Button
Pressing the emergency button stops every mechanical movement and the X-
ray emission.
• Control Panel
The control panel incorporates the several buttons to mainly control the
movements of the gantry and table. Each front and rear cover has two control
panels. Refer to the Gantry Control Panel page for the function descriptions.
Fast
Table Up
Cradle In
Positioning
Light
Table Down
Cradle Out
Fast Gantry Tilt
Lamp
Gantry Tilt
Lamp
Gantry Tilt
Function
The gantry control panel enables you to control the movement of the gantry,
table or cradle with the following buttons.
• Cradle In/Out
Use this to move the cradle toward or away from the gantry aperture.
• Table Up/Down
Use this to move the table up or down.
• Fast
Use this to increase the speed of cradle in/out. This does not apply to table up/
down.
WARNING!
The touch sensor on the gantry cover works only during the remote tilt
operation at the OC. It does not work when the tilt operation is done at the
Control Panel.
• Positioning Light
Use this to turn on or off the internal/external positioning lights.
• External Landmark
Use this to move the anatomy beneath the external positioning light.
• Internal Landmark
Use this to move the anatomy beneath the internal light to adjust scan location
zero.
• Practice
Use this to test the * Breath Navi function.
(*Breath Navi is the small device attached to the edge of the gantry aperture that
gives the visual breathing instructions to a patient with hearing problem.)
Interference
Function
The gantry display panel shows the following readings and indications.
• X-ray On
The X-ray On lamp illuminates during X-ray exposure.
• Interference
The Interference lamp turns on when some mechanical interference occurs.
Function
The table is used to load a patient for scanning.
• Cradle
The cradle moves into or out of the gantry aperture.
• Latch Button
The latch button is used to latch or unlatch the cradle. The unlatched cradle can
be manually slid. The Display Panel shows whether the cradle is latched.
• Speaker
The speaker is used to deliver oral instructions to a patient.
• Foot Switch
The Foot switch is placed at the foot of the table. When the operator steps on it,
the switch turns on and activates functional buttons on the gantry panel.
Main Switch
CD–ROM Drive
5” MOD Drive
3.5” MOD Drive
Function
The operator console (OC) is mainly used for the operator to set up the scan
procedures and process the resultant image data.
• Scan/Display Monitor
The 17-inch monitor (21-inch optional) on the OC can be mainly used for two
purposes, scanning patients and displaying images.
• Keyboard/Mouse
Please refer to the User Interface on page 1-10.
• Main Switch
Please refer to the System Power On/Off on page 1-22.
• CD-ROM Drive
This drive is dedicated to service or application software installation.
Main Switch
DVD-RAM Drive
5” MOD Drive(Option)
• DVD-RAM Drive
Raw data can be stored in DVD-RAM.
Mouse
Middle button
• "Click"
Press a button once then release it quickly.
• Left button
To select any of icon, item, function, etc.
• Middle button
To change window width and/or level by moving the mouse left/right and/or up/
down respectively with the button pressed.
• Right button
To scroll the image or modify the image zoom factor.
Volume Dial
Move to Stop Start Stop
Scan Move Scan Pause Scan Emergency
X–ray on Stop Mic.
Talk Button
Functions of each button and key are described on the following pages.
• Move to Scan
Press this button when lit to move the cradle to the prescribed scan start
location.
• Stop Move
Press this button to stop the cradle when it is traveling.
• Start Scan
Press this button when lit to start scans.
• Pause
Press this button to temporarily halt scans. Scans pause after completing the
current one scan. Press Start Scan button to resume scans.
• X-ray on
This button lights up during X-ray emission.
• Stop Scan
Press this button to stop the current or halted scan.
• Emergency Stop
This button cuts off the power to the equipment inside the scan room.All
mechanical movements and X-ray radiation are halted.
Note : This button affects the power neither to OC nor laser camera.
• Talk (Intercom)
Hold down this button to give some verbal instructions to the patient during an
exam.
• Volume
Use these dials to adjust the voice volume of patient or operator.
• Function key
F1, F2, F3 and F4 function keys are used for filming-related operations. Please
refer to Chapter 5 [FILMING] for detailed information.
F1 F2 F3 F4
Film Film Film Film
Image Page MID Series
F5 function key is used to return to the window width and level that have been
used most recently
F5
Previous
W/L
F5-11 function keys are used to preset the specific window width and level for
specific anatomical areas. The procedure is as follows.
F6 F7 F8 F9 F10 F11
Abdo– Head Lung Media– Spine Verte– Drop on
men stinum bra Auto Film
1. Firstly select the image as a primary one (green border) of which WW and
WL you wish to register as default.
Erase Disp
Norm
Ellipse Grid
Trace
ROI On/Off
• Key top
These keys are mainly used to input texts/numbers and display images.
Enter
Shift
Standard Accessories
o Cradle pad
o Cradle extender
o Standard head holder assembly
o Security straps
o Phantoms to calibrate and check performance
o Phantom holder
o Console chair
o Operator manual
Optional Accessories
o Coronal head holder assembly
o Shallow head holder assembly
o Arm Support
WARNING!
Do not use any accessories other than those provided by GE. Use of other
accessories may cause unexpected troubles or physical injuries.
Also, if anything other than GE accessories is placed in the path of x-ray
beams, it may affect image quality.
(Back side)
(Front side)
Label
MADE FOR GENERAL ELECTRIC CO.
MILWAUKEE, WISCONSIN BY
GE YOKOGAWA MEDICAL SYSTEMS, LTD.
7–127 ASAHIGAOKA 4–CHOME, HINOSHI TOKYO, JAPAN
MODEL 2201806
SERIAL
MANUFACTURED OCTOBER 1997
Label
MADE FOR GENERAL ELECTRIC CO.
MILWAUKEE, WISCONSIN BY
GE YOKOGAWA MEDICAL SYSTEMS, LTD.
7–127 ASAHIGAOKA 4–CHOME, HINOSHI TOKYO, JAPAN
MODEL 2183827
SERIAL
MANUFACTURED OCTOBER 1997
Head Strap
Chin Strap
• Head Band
Place the wide area of the head band between the head holder and head holder
cushion, with the straps hanging on either side of the head holder, before the
patient lays on the table. Move the patient's head into position on the cushion,
cross each end of the band over the forehead, and fasten the pad of Velcro
"loops" to the "hooks" on the back of the head holder. Use the head band to
position and stabilize the head during a study.
• Chin Band
Center the hole in the band over the patient's chin, and fasten the pad of Velcro
"loops" to the "hooks" on the back of the head holder. Use the chin band to
position and stabilize the head during a study.
• Neck Pad
Place this pad vertically in the Coronal head holder to cushion the rear head.
WARNING!
Be extremely careful to avoid contact between the coronal head holder and
the gantry while lowering the table in the aperture. Patient injury to the
neck or equipment damage could result.
• Cradle Pad
Use during standard studies; attaches to the cradle with Velcro "hooks."
• Cradle Extender
Attach the cradle extender to the end of the table nearest the gantry, to increase
the overall cradle length, and position the body area of interest inside the
scannable range. Use the extender to support the patient's head or feet during a
body scan. The material is metal-free.
• Body Strap A
Use the 60 mm wide strap to position and secure the patient's head to the Head
Cushion; or use as needed to secure blankets or clothing. Follow Body Strap B
directions to attach and secure the straps. One set provided with system.
• Body Strap B
Use the 150 mm wide straps to secure the body to the cradle, to reduce patient
movement; to secure blankets or clothing, to prevent contact with the cradle
surface; or, to cover and secure the patient's hands. Slide the nylon guide,
opposite the Velcro ends of the straps, into the raceway attached to both side
edges of the cradle. Slide the strap along the length of the cradle to position it
over the patient, and fasten the Velcro ends together to secure the patient.
Three sets provided with the system
• Body Strap C
Use the 370 mm wide strap in the same manner as Body Strap B.
One set provided with system.
• Extension Strap
Use the extension strap with the corresponding body straps to secure large or
obese patients. Attach Velcro "hooks" to "loops" on the body strap to increase
the length. Each set of body straps has an extension strap; five provided with
the system.
Cradle Pad
Cradle Extender
(Back side)
Label
MADE FOR GENERAL ELECTRIC CO.
MILWAUKEE, WISCONSIN BY
GE YOKOGAWA MEDICAL SYSTEMS, LTD.
7–127 ASAHIGAOKA 4–CHOME, HINOSHI TOKYO, JAPAN
MODEL 2201803
SERIAL
MANUFACTURED OCTOBER 1997
To attach an accessory:
• Make sure the clasp on the accessory is unlatched, and hanging freely.Hold
the clasp in the unlatched position while you attach the accessory.
• Align the accessory tongue with the pocket at the end of the cradle.Keep fin-
gers clear of the cradle. Slide the tongue all the way into the pocket until it
rests snugly against the angled face of the cradle.
• Gently pull the clasp toward the floor and away from the gantry until it hooks
the catch on the cradle. Push the hinged-end of the clasp toward the cradle
until it snaps into place.
To detach an accessory:
• Pull the hinged-end of the clasp toward the floor and away from the gantry
until it snaps open and clears the catch.
• Gently slide the accessory toward the gantry until it clears the cradle.If the
accessory offers any resistance, check to make sure the clasp cleared the
catch on the cradle.
1. Turn on the main switch located on the lower right side of the OC.
2. The system will start up and the Scan/Display monitor on the OC will
automatically proceed to its Top Level screen.
How to turn off the system power (Usually power is on 24 hours a day.)
1. Click on the [Shutdown] icon on the upper left corner of the screen.
Shutdown
2. The system will automatically proceed to the point where you can turn off
the main switch.
2. From the pop-up attention box, select the option you wish to do.
- Logout User allows enables you to login as a different user.
NOTE: For information on how to Login and Logout, refer to: Login and Logout.
- Restart brings the system all the way down to the power off point
and then re-boots the system.
- Shutdown brings the system all the way down to the power off
prompt. You can then turn power off to the system.
3. Click [OK].
4. If you selected Shutdown or Restart a series of menus and messages
appear.
- If Restart was selected the system automatically restarts the system.
- If Shutdown was selected, turn off the OC power, if not already off
with thepower switch on the front of the console and then turn the
power switch on torestart the system.
5. To power up the system push the power switch on the front of the operator
console and the system will automatically restart.
2. Select your user name from the Select User pull down menu.
- You have to be an assigned administer to add or remove users from
the system.
3. Enter your Password.
4. Click the [User, Groups, Permission] tab.
- This brings up the HIPAA User Management Screen Figure 3-1.
5. Click [New User].
6. From the pull down menu Click Add a New User.
- Type in the information.
- The user name must not contain any spaces.
NOTE: The initial password is the user name the first time the new user logs in.
The new user is then prompted to change their password.
7. Click all the boxes that apply to that user.
- This will give them permission for different areas of the scanner.
Scan
This icon leads to the Display screen.
The icons on the Display screen are described in the later chapter.
Display
This contains Reformation, 3D, Network and other advanced
image analysis applications. Refer to {Image Works} section.
Image Works
Shutdown
This is only for service purpose. Users are not allowed to touch
this.
Service
December 22 5:49 PM
512: 148 256: 595
To Current Messages
• Reconstruction Status
This area shows the exam, series and image number of the exam that is
currently being reconstructed.
• Archive Status
This area shows the status of images being archived.
• Network Status
This area shows the status of networking.
• Filming Status
This area shows the status of filming.
• Current Messages
Click here to open the Current Messages screen.
New Patient
1. Click on the [New Patient] icon to open the following New Patient screen.
Patient Information Protocol Selection
Exam Number
Accession Number
Patient ID
Format:LastName^FirstName^Middle^Prefix^Suffix
Patient Name
mon day yr
Birthdate Sex
Years Months Weeks Days
Age
Weight Kilograms Pounds
Referring Physician
Radiologist
Operator
History
Exam Description
Protocol Number
Req.Proc.ID
End
Exam
3. Choose a desired scan protocol from the Default Protocol area located on
the right side of the New Patient screen.
Patient Schedule
Function : Patient Schedule allows you to enter patient data ahead of time.
Once it is time to perform a specific patient exam, you can select from the
created list, enter the patient ID or requisition number to call up the patient data.
On the left side of the column under Status, there are two letters. C stands for
Completed and N for New. The completed exam at the top of the list is the last
completed exam. The only active functions at this point are [Quit], [Add Patient],
[Delete All], [Preferences] and Next/Prior arrows.
Note : Active icons have clear black characters on them whereas inactive icons
have dim characters.
Select Patient
Select
Patient
4. Select [Select Patient]. The patient data will then be placed in the New
Patient data area.
Note : The patient data will also be called up by typing in the patient ID.
While your CT/e system is connected with HIS/RIS and if your CT/e system has
Connect Pro option, select [View More Info] to get more information on the
patient.
Add Patient
Add
Patient
Select [Add Patient] to open the following menu to add patients to the schedule.
Patient data, date and time of the exam and the protocol can be entered here.
Schedule Patient Protocol Selection
Accession Number
Patient ID
Format:LastName^FirstName^Middle^Prefix^Suffix Anatomical Selector
Patient Name
mon day yr
Birthdate Sex
Years Months Weeks Days
Age
Weight Kilograms Pounds
Referring Physician
Radiologist
Operator
History
Exam Description
Protocol Number
Cancel Accept
Edit Patient
Edit
Patient
Select [Edit Patient] to bring up the Patient Information screen to edit patient
data.
Delete Selected
Delete
Selected
[Delete Selected] allows you to delete specific patients from the Schedule list.
1. To delete only one patient, click on that patient's information, then select
[Delete Selected].
2. To delete multiple patients in order, click on one patient, then go to the last
patient and press and hold the Shift key, then click on that patient. Then,
select [Delete Selected].
To delete multiple patients out of order, press and hold the Control key, then
click on the desired patients and select [Delete Selected].
3. A confirmation message "Are you sure you want to delete selected patients
?" appears. If OK, click on [OK].
Delete All
Delete
All
Delete All
OK Cancel
Preferences
Preferences
[Preferences] allows you to set the sort order and delete time.
Preferences
Edit Modality No
Use Study UID? Yes No Yes
worklist?
All Dates
OK Cancel
3. If you want the system to automatically delete completed exams, you can
click in the Days box and type in the number of days desired. The default is
0 (zero).
Protocol Management
Function : Protocol Management includes two functions, which are [Auto Voice
Record] and [Protocol Management].
Quit
Quit
2. The following Auto Voice Record screen appears upon the selection of
[Auto Voice Record].
Twenty protocols in all are available. The first three protocols are preset in the
system and can not be changed. Other 17 protocols are programmable.
Korean Chinese
Cancel
You can select any one of them to listen to it. In order to select any one of them
for the scan, you are required to do so through the [Auto Voice No.] button on
the Scout (or Axial/Helical) Prescription Screen. (Refer to Chapter 4.)
3. To create a new Auto Voice instruction message, point and click in the
Name area where you wish to put the protocol. Type in the name and press
the Enter key.
After pressing the Enter key, the Pre and Post Message areas become
active.
6. When finished with recording, select [Stop]. Select [Play] to listen to what
was recorded.
Message Management
Done
To play a message, simply select the message box you wish to play and
select [Play].
Quit
Quit
3. Select a protocol or an anatomic region to open the menu on the next page.
Done Cancel
5. To duplicate (copy) an existing protocol, first choose the protocol and select
[Duplicate]. Then, you may slightly modify it to create a new protocol that
will have the similar parameters to the original protocol.
6. To edit or delete an existing protocol, first choose the protocol and select
[Edit] or [Delete] respectively.
8. After selecting [New], enter the protocol name on the highlighted line and
press the Enter key. To set a protocol as the default for the region, select
the [Default] button next to the protocol. Only one protocol can become the
default for the region.
9. Next, select the Series Type that will begin the protocol and click on [OK].
Scout Axial
OK Cancel
Protocol : Series:
Anatomical
Reference
Series Description
Auto Show
Store Auto mA OFF Localizer
Smart
Prep
RX
Create
Cancel Prior Next Repeat Delete
New Accept
Series Series Series Series
Series
10. In the above screen, the [Scan] icon parameters (Thick, Image Int'val,
pitch, etc.) are supposed to be filled. In order to enter Timing, Recon and
Film parameters, click on the [Timing], [Recon] and [Film] icon,
respectively. The menus for Timing, Recon and Film are shown on the
following pages.
Prior/Next : Prior or Next key appears on the screen only when more than
three groups are prescribed for one screen.
The following Recon menu will appear when you click on the [Recon] icon.
The following Film menu will appear when you click on the [Film] icon.
Scan Timing Recon Film
11. After all the scout view parameters have been entered, select [Create New
Series] button on the front page screen to create the next series. Then, the
following screen appears.
Scout Axial
Create Create
Before After Cancel
14. After all the series and the parameters have been created, touch [Accept]
button to save the protocol.
15. 15. You will then return to the beginning so that you may choose to delete,
edit or create a new protocol. Or you can select [Done] and then [Quit] to
exit the protocol section.
Retro Recon
1. First click on the [Retro Recon] icon. The following screen appears listing
up to 1000 scan data. Click on the [Update] button to update the list or
select [Quit] to exit.
Select Update
Quit Series
2. Point and click on the exam and series you wish to reconstruct
retrospectively and then choose [Select Series] at the bottom center.
The system automatically proceeds to the Retro Recon View/Edit screen.
List
Quit Exams Confirm
When the [Recon] tab is selected, the right side of the screen is as follows.
Scan Scan Scan R/L A/P Recon Image Motion Recon Matrix Special Graphic
Retro Start End DFOV Center Center
Type (cm.) Type Filter Correct Mode Size Filter Retro
Location Location (mm) (mm)
Y N
Y N
Y N
List
Quit Exams Confirm
Image FLR
[ANR 2] has a stronger effect than [ANR 1]. Also, [AAR 2] has a stronger
effect than [AAR 1]. Select either of them or select [None] not to use these
filters.
5. Select the [Recon Mode] button to open the following screen. (*Smart
Helical Option is needed to have [Plus] button.)
Recon Mode section allows you to edit the scan type and the amount of
data for reconstruction.
Recon Mode
Axials: Full 2 3 4 5 6
OK Cancel
Recon Mgmt
Function : Within the Recon Management function, you can have access to
Suspended Entries and Recon Queue Management.
Recon Management
Suspended Entries
Patient ID Exam # Series # Image # Type
Delete
Suspended Unsuspended Update
Entries Entries List Prior Next
Quit
Suspended Entries
These are basically images that were not reconstructed for some reason and
were placed in a Queue.
• [Unsuspended Entries]
Select this button to try to reconstruct the images again after selecting image
entries.
• [Update List]
Select this button to update the Unsuspended Entries List if any entries need to
be added while Recon Management was open.
• [Restart Queue]
Select this button to restart the reconstruction queue after you stopped it for
some reason.
Tube Warmup
Function : Tube Warmup allows you to have access to the Tube Warm-up and
Daily Calibration procedure.
Tube Warmup
Daily Calibration
Tube Warmup
Warming up the tube as recommended will provide the high quality
Menu1 images and extend the life of the tube.
Make sure that no one is in the scan room and the gantry aperture is
clear during warmup.
Cancel OK
2. Select the [Tube Warmup] button, the[OK]to proceed to the tube warmup
procedure or select the [Cancel] button to exit.
Service Scan
Remaining scan : 9
Press <Confirm> and <Start scan> to start scan
Cancel Confirm
Warning
Yes No
Service Scan
Remaining scan : 73
Press <Confirm> and <Start scan> to start scan
Cancel Confirm
Upon selection of [Confirm], the system first performs the same scan
protocol as Tube Warmup and daily calibration later.
Scanner Utilities
Function : From Scanner Utilities you can have access to either Rawdata
Funtion or User Calibration.
Note : Make sure which drive do you installed in the system. If you install the
MOD drive, Rawdata will save to 3.5” MOD; If you install the DVD drive,
Rawdata will save to the DVD.
Scanner Utilities
Rawdata User
Functions Calibration
Quit
Series No:
Acqs # Type Rotation Azimuth Location Scan Time kV mA Thick Focus SFOV Reserve Date/Time
Start Save
Series No:
Acqs # Type Rotation Azimuth Location Scan Time kV mA Thick Focus SFOV Reserve Date/Time
Start Save
1. Select any of raw data that you wish to save to MOD (Magnetic Optical
Disk). You can also use [Whole Exam], [Whole Series] or [Individual Raw]
button from the [Rawdata Selection] pulldown menu.
2. After selecting the raw data, click on the [Start Save] button to start saving
the raw data into the MOD.
Note : Image data must be stored in other MOD(s) than the one for raw data.
Series No:
Acqs # Type Rotation Azimuth Location Scan Time kV mA Thick Focus SFOV Reserve Date/Time
Restore Delete
1. Select any of raw data that you wish to restore from MOD to the system
disk, or that you wish to delete from MOD. You can also use [Whole Exam],
[Whole Series] or [Individual Raw] button from the [Rawdata Selection]
pulldown menu.
Reserve/Release Rawdata
When you select [Reserve/Release] from the [Function Selection] pulldown
menu, the following screen will appear.
Series No:
Acqs # Type Rotation Azimuth Location Scan Time kV mA Thick Focus SFOV Reserve Date/Time
Reserve Release
Initialize MOD
Use this function to initialize completely new MODs. (Previously initialized MODs
for image storage can not be initialized by this.)
When you select [Initialize MOD] from the [Function Selection] pulldown menu,
the following screen will appear.
Initialize MOD
Yes No
Scanner Utilities
Rawdata User
Functions Calibration
Quit
Series No:
Acqs # Type Rotation Azimuth Location Scan T ime kV mA Thick Focus SFOV Reserve Date/T ime
Start Save
Series No:
Acqs # Type Rotation Azimuth Location Scan T ime kV mA Thick Focus SFOV Reserve Date/T ime
Start Save
1. Select any of raw data that you wish to save to DVD (Digital Video Disk).
You can also use [Whole Exam], [Whole Series] or [Individual Raw] button
from the [Rawdata Selection] pulldown menu.
2. After selecting the raw data, click on the [Start Save] button to start saving
the raw data into the DVD.
Note : Image data must be stored in other MOD(s) than the one for raw data.
Series No:
Acqs # Type Rotation Azimuth Location Scan Time kV mA Thick Focus SFOV Reserve Date/Time
Restore Delete
1. Select any of raw data that you wish to restore from DVD to the system
disk, or that you wish to delete from DVD. You can also use [Whole Exam],
[Whole Series] or [Individual Raw] button from the [Rawdata Selection]
pulldown menu.
Reserve/Release Rawdata
When you select [Reserve/Release] from the [Function Selection] pulldown
menu, the following screen will appear.
Series No:
Acqs # Type Rotation Azimuth Location Scan Time kV mA Thick Focus SFOV Reserve Date/Time
Reserve Release
Initialize DVD
When you select [Initialize DVD] from the [Function Selection] pulldown menu,
the following screen will appear.
Initialize DVD
Yes No
Calibration
Sequence kV
Cancel OK
Display
Image Area
Focus Selection
"Focus" means which viewport is active for image display. There are two kinds of
focus, which are primary and secondary focuses. Primary focus viewport has a
blue border, whereas the secondary focus viewport has a yellow border.
Every image analysis or manipulation function is available in primary focus.
Image analysis or manipulation functions available in the secondary focus are as
follows: Window W/L, Magnification, Pan/Zoom, Flip/Rotate, Display Normal,
and Erase Graphics.
You can select the primary or secondary focus by clicking on the desired
viewport as follows:
• Double click
To select primary focus viewport and deselect other viewports
• Single click
To select a new primary focus viewport and change the previous primary to
secondary
• Triple click
To select a new primary focus viewport and change all the other viewports to
secondary
Layouts
On the Display desktop, there is a choice between two layouts for viewing
images. They are the Autoview Layouts and the Review Layouts. During
scanning the Autoview Layouts will primarily be used.
Click on the left side of this button to select Au-
toview Layouts and on the right side for Review
Layouts.
Autoview Layouts
Autoview Layouts
Auto Auto
Link View
Back
• Autoview
This viewport always displays in real time the most recently reconstructed
image. Next/Prior buttons do not work here.
• Auto Reference
This viewport displays a Scout image with reference lines that show the
locations of Autoview images. Front Scout and side Scout can be toggled by
Next/Prior buttons.
• Auto Film
This viewport displays an image for auto filming and occupies the upper left
quadrant. This can be done by selecting the folded corner at the top right of the
viewport. To return the viewport to the cross reference display, select the folded
corner again.
• AL (Auto Link)
While reconstruction is active, the lower right viewport will automatically display
the first image of the series being currently reconstructed. The viewport will be
annotated with AL to indicate that Auto Link is on. All display features and Next/
Prior can be used with Auto Link.
• Free
This viewport can be used to display any image from other exam or series.
2. Select Review Layouts at the Exam Rx Display menu to open the Review
Layouts menu.
Review Layouts
[Full Screen
Format]
[Viewport
Format]
3. Select [Full Screen Format] to display the selected layout on the entire
screen. Select [Viewport Format] to display the selected layout in primary
focus viewport.
List / Select
Sort
Examinations : Exam #47, May 05 92, SMITH, JON
Exam Name Date Description Mod Fmt A Ser Type Imgs Description Mod
3145 J.Herman Jan 08 98 CT 1 PROSP 18 CT
3512 B.Fox Dec 23 97 CT
C
A : Examination List
B : Series List
C : Image List
Routine Display
Hide Screen
Erase
Graphics Save
Back
Roam / Zoom
1. Place the desired image in primary focus viewport.
2. Select [Routine Display] at the Exam Rx menu to open the Routine Display
menu.
4. After selecting [Roam], place the mouse cursor over the image you wish to
roam and hold down the right mouse button, then move the mouse to roam
the image. When you release the mouse button, the image will be
deposited on that point. Select [Display Normal] to return the image to its
original position.
After selecting [Zoom], place the mouse cursor over the image you wish to
zoom and hold down the right mouse button, then move the mouse right (or
up) to magnify or left (or down) to minify the image. When you release the
mouse button, the image will retain the zoom and the magnification factor
will be displayed at the bottom of the annotation on the upper right of the
image. The valid magnification ranges are from 0.5 to 8 and from1.0 to 4.0
in a 1024 matrix viewport.
Select [Display Normal] to return the image to its original size.
Explicit Magnify
This function allows you to numerically set an image magnification factor.
Cancel
Flip/Rotate
1. Place the desired image in primary focus and select [Routine Display].
FLIP/ROTATE
FLIP:
ROTATE:
RR RL
Cancel
4. Select one from FLR (Flip Left to Right), FTB (Flip Top to Bottom), FTB/
FLR, RR (Rotate to Right) or RL (Rotate to Left).
ProView
This function allows you to select an edge enhancement or smoothing filter to be
applied to images.
Cancel
Note : These filters can also be selected on the Auto Film Tab card.
Note : These filters can not be applied to Scout images.
Gray Scale
G1 G2 G3 Off
Cancel
Note : These filters can also be selected on the Auto Film Tab card.
Note : These filters can not be applied to Scout images.
Display Normal
1. Images must be in primary focus.
List / Select
1. When you select [List/Select], the browser appears.
Ellipse ROI
1. Images must be in primary focus.
3. The ellipse ROI cursor will be displayed in blue (active) on the image. To
move the ROI cursor, place the mouse cursor in the center of the ellipse,
then click and drag it with the left mouse button. The other way to move the
ROI cursor is to place the mouse cursor on the number next to the ellipse or
on the line, then click and drag it with the left mouse button. To change the
size, click and drag the small box outside the ROI cursor. To rotate the ROI
cursor, click and rotate the solid cursor on the line.
4. Place the mouse cursor out of the ellipse, then, click the left mouse button
once to calculate the ROI of the ellipse. The ellipse ROI cursor turns white
(inactive) and is deposited there.
Note : Once the ellipse ROI cursor turns white, it can not be modified, but you
can turn it blue again by placing the mouse cursor inside the ROI cursor and
click the left mouse button once. Then, you can change the size and location of
the ROI cursor again.
The figures of the Mean, Standard Deviation and Area are displayed at the
bottom right of the image. The ROI measurements can be done many times, but
the system will display only the last three results on the image.
Select [Erase] to erase the ROI.
Measure Distance
1. Images must be in primary focus.
3. A line with two small boxes at each end will appear in blue (active).
4. To move the line, place the mouse cursor on the line and hold down the left
mouse button, then drag the line to a new location. To change the length of
the line, place the mouse cursor over the small box and hold down the left
mouse button, then drag the box until you get a desired length.
5. Place the mouse cursor out of the line, then, click the left mouse button
once to calculate the length of the line. The line turns white (inactive) and is
deposited there.
Note : Once the line turns white, it can not be modified, but you can turn it blue
again by placing the mouse cursor on the line and click the left mouse button
once. Then, you can change the length and location of the line again.
At the bottom right of the image area, a real time readout of the distance will be
shown. The Distance measurement can be done many times, but the system will
display only the last three results on the image.
Select [Erase] to erase the line.
Grid
1. Images must be in primary focus.
4. To move the grid, first click left on the center of the grid. It becomes active
(blue), then drag it to the desired position. To deposit the grid, click outside
the grid.
Cross Reference
This function allows you to display the cross reference lines on a Scout image.
Series Number
Increment
OK Cancel
4. Select [OK] to display the image lines on the Scout or select [Cancel] to
exit.
User Annotation
1. Images must be in primary focus.
2. Select [User Annotation] to show the following active (blue) arrow cursor on
the image.
3. You can move both the box and arrow at once by placing the mouse cursor
anywhere on the blue line, then click and hold down the left mouse button,
then move it. You can move just the arrow or box by the same way except
you would place the mouse cursor on just the area you wish to move
instead of the blue line.
4. Type in the necessary information, then click the left mouse button outside
the image area to conclude the type entry. Make sure when typing the
annotation that the mouse cursor is inside the image area.
Exam Pg / Series Pg
1. Select [Exam Pg] or [Series Pg] to open the exam page or series page of
the exam in primary focus.
3. Select [Manual Film] to put the page into Manual Film Composer or select
[Auto Film] to put the page into Auto Film Composer.
Hide/Show Graphics
1. Select [Hide Graphics] to temporarily hide any graphics, annotation or
measurement.
2. The [Hide Graphics] button works as a toggle. Once you have selected
Hide, then the button changes to [Show Graphics], allowing you to show
any graphics, annotation or measurement.
Erase
The Erase function allows you to erase annotations, cursors and measurements.
To erase, simply click on [Erase]. If you have multiple annotations or cursors on
the image, the erase function will erase the active one (blue).
Screen Save
Screen Save will save everything that is on the image in primary focus.
Measurements
Report
MIROI
Pixels
Back
MIROI Analysis
OK Cancel
3. The first selection on the menu is Image Range. The default is All.If you
wish to select certain image range, first click on [All] to turn it off, then type
in the number range.
5. Define Region selection allows you to choose the Ellipse, Box or Trace
ROI.
Report Pixels
This function allows you to get individual pixel values readings from a certain
small area.
2. An 11x11 pixel box will appear along with the Report Pixels menu.The size
of the pixel box can not be changed but you can move the box to the
desired location by clicking and dragging on the letters RP or on the box
itself.
Report Pixels
OK Cancel
A graph like the following example will be displayed showing the pixel values
within the box.
Report Pixels
ID: 8120
I 84 I 85 I 86 I 87 I 88 I 89 I 90 I 91 I 92 I 93 I 94
I 165 20 21 23 23 19 20 21 21 20 25 26
I 166 20 22 24 23 20 20 22 25 22 24 26
I 167 22 23 24 23 24 25 25 25 22 23 23
I 168 21 20 20 20 24 25 23 23 22 23 23
I 169 18 18 19 21 22 24 22 21 20 20 21
I 170 17 18 19 21 20 21 22 21 20 20 21
I 171 18 18 19 21 20 20 20 21 20 20 21
I 172 17 17 17 21 20 19 20 21 20 20 20
I 173 17 17 18 21 21 20 20 21 21 20 20
I 174 18 17 18 21 21 19 20 21 21 20 20
I 175 17 17 18 21 21 19 20 21 21 20 20
Display Preference
Maintain
Annotation
User
Levels Annotation
(Not available)
Auto
Auto Minify Enlarge
Next/Prior
Inverse Each
Video Viewport
Next/Prior Continuous
Series Report
Binding Cursor
Back
Annotation Levels
This function allows you to change the annotation levels of the screen and film.
Annotation Levels
Screen Film
No Annotation No Annotation
Customize Customize
Save as
Cancel
default
2. You can change the annotation level by clicking on the diamond or the
words for Screen and Film. The selected choice will be highlighted in blue.
- No Annotation : Only window width and level
- Partial Annotation : Exam number, image number, anatomic reference,
table location, patient name, date of exam, RAS coordinate, and the
measuring scale
- Full Annotation : All image annotations (Default)
- Custom Annotation : Customized annotation
Customize
1. Hospital Name
2. Patient Name
3. Patient ID
4. ESI Information
5. Scan Time
6. Scan Parameters
7. View Parameters
8. Top Marker
9. Bottom Marker
10. Left Marker
11. Right Marker
W. Windowing
G. Graphic Annotation
14. Display Parameters
15. Filter
16. Date of Birth
W. Windowing
G. Graphic Annotation
OK Cancel
[Manual Film
Composer]
Film Composer
Options
Current status is :
Clear
[Auto Film
Composer]
Clear
Accelerator Bar
WARNING!
Avoid patient injury ! Use the security straps provided with the system to
immobilize the patient. The security straps secure the patient to the table
and prevent the patient from falling to the floor. Never leave a patient
unattended while the patient is in the scan room. A violent patient may
break the straps or become unsecured.
Be extremely careful with patient IV or oxygen lines, patient's hands and
feet and loose clothing to avoid patient injury during table travel.
The cradle has a maximum distribution load capacity of 400 pounds (180
Kg). Exceeding this load limit could affect some mechanical performance.
Interference
External 190 mm
beam
Internal
beam
Tomographic
plane
Patient positioning ends when the table location figure is displayed on the
display panel.
GB
OM
SN
EM
CM
XY
UB IC
SP
GB : Glabella
OM : Orbital Meatal line
EM : External Auditory Meatus
SN : Sternal Notch
XY : Xyphoid
CM : Costal Margin
IC : Iliac Crest
UB : Umbilicus
SP : Symphysis Pubis
New Patient
mon day yr
Birthdate Sex
Years Months Weeks Days
Age
Weight Kilograms Pounds
Referring Physician
Radiologist
Operator
History
Exam Description
Protocol Number
Req.Proc.ID
End
Exam
Name : ID :
Anatomical
Reference
Image Area
Series Description
Auto
Store
Auto T ransfer
Add Delete
Scout Selected
Scout
Scout Auto
Scout Scan Start End mA
kV Plane
Voice
Num Type Loc. Loc. No.
1 Scout
2 Scout
Note : Make sure the patient's position in the protocol you have selected
matches that on the CT table. Compare your patient's position with the on-
screen picture in the upper left corner of the screen. Changes to the patient's
position in the protocol can be made by Arrow icons on the picture. In order to
reverse the position of Head-first or Feet-first, point and click on the icon. In
order to change the position of Decubitus left/right or Supine/Prone, point
and click on the or icon.
Before entering or modifying any starting or ending locations for a Scout view,
the system will determine if an anatomic reference point has been established. If
it has not, the starting and ending location parameters will be highlighted in red.
To proceed, you must establish an anatomic reference point by using the internal
or external landmark buttons on the gantry control.
Add Scout
Add
Scout
End Exam
End
Exam
Select this to return to the Protocol Selection screen and choose a new protocol
for a different region.
Next Series
Next
Series
Repeat Series
Repeat
Series
One More
One
More
Confirm
Confirm
Select this icon to accept the prescription and proceed to X-ray radiation. This
icon will change to the following [Pause] icon during the scan.
Pause
Pause
Select this icon to temporarily halt the scan. This icon will change to the
[Resume] icon shown below during the pause.
Resume
Resume
Select this icon to resume the scan.
Scan Progress
Exam:128 Series:1
0 10
Number Type kV mA Plane Start End
Seconds
S65.0 I120.0
001 Scout 120 80 0 S65.0 I120.0
Biopsy
Rx
Scanning Delay Timer Patient Handling
Move to Scan Stop Move Start Scan Pause X–ray on Stop Scan
5. After finishing the first Scout scan, you may select [Repeat Series] button
on the Scan Progress screen to repeat the same Scout series.
6. If there is an axial series in the protocol you used, you can proceed to Axial/
Helical View/Edit screen by clicking on [Next Series] button on the Scan
Progress screen.
7. If you need to return to the Scout prescription screen, select the following
[Return to Rx] icon.
MOD Drive
Name : ID :
Protocol : Exam: Series:
Anatomical
Reference
Series Description
2. When Auto Store is on, all series will be automatically stored into the MOD
after you click on [End Exam].
1. Click on [Auto Transfer] on the Axial View/Edit screen to open the Auto
Transfer Remote Host Selection menu.
AW 1
AW 2
AW 3
AW 4
MR 1
MR 2
OK Cancel
Prior Next
2. After clicking on [Host x], select host(s) from the list to which you wish to
transfer the data. You can select up to four hosts (Host1-Host4).
Anatomical
Reference
Image Area
Series Description
Auto Show
Store AutomA OFF Localizer
Preview
Biopsy
Rx
Smart
Prep
RX
Image Area : Images will be displayed here. This image area can be enlarged to
full screen by clicking on the small square icon located in the upper right corner.
Prior/Next : Prior or Next key appears on the screen only when more than three
groups are prescribed.
• Scan speed
Since there is no interscan delay between each slice acquisition, scan speed is
much faster than conventional axial scans.
Faster scan also helps reduce the total amount of contrast agent.
• No misregistration of anatomy
Since one scan sequence is performed during patient's breath hold, it
contributes to very little misregistration of anatomy.
Optimize
Images Scan Start End No. of Thick Image Gantry CTDIvol not needed
Type Loc. Loc. Images Pitch SFOV kV mA
(mm) Int val Tilt DLP
Preview
Biopsy
Rx
Smart
Prep
RX
Select Create
End Next Repeat One Priority
New New
Exam Protocol Series Series More Recon
Series
Confirm
• When you wish to use the common parameters for all scan groups as to Scan
Type, Thick, Image Int'val, Pitch, Gantry Tilt, SFOV, kV or mA, click on each
column head and then select the alternative.
• When you wish to select parameters separately for each group, click on each
data field.
• Each parameter is described on the following page.
OK
[ANR 2] has a stronger effect than [ANR 1]. Also, [AAR 2] has a stronger
effect than [AAR 1]. Select either of them or select [None] not to use this
filter.
- [Show Recon 1, 2, 3] : These buttons are used for prospective multiple
reconstruction.
[ANR 2] has a stronger effect than [ANR 1]. Also, [AAR 2] has a stronger
effect than [AAR 1]. Select either of them or select [None] not to use this
filter.
When you need to return to the parameters of Recon1 , simply select [Show
Recon1].
3. If you need to perform the third reconstruction, select [Show Recon3], then
take the same steps as Recon2.
Autofilm Setup
Autofilm
Setup
Autofilm Setup
Destination
Laser Camera
Formats Size Auto Start Auto Print
Normal Slide Yes Yes
Yes Yes
Scout XRef–Scout
Film Direction
Top to Bottom
e/s/i Yes
OK Cancel
Show Localizer
Show
Localizer
Click this icon to display the Scout image with graphical lines that show the scan
locations prescribed beforehand. You can graphically modify the start/end
location, interval, image number, or DFOV. Those changes will be reflected on
the scan prescription screen in orange color.
Please refer to the later pages for detailed descriptions.
Recon
Area
Select this to display a circle cursor over the image, then change the size and
location of the cursor to designate the recon area instead of typing in DFOV, R/L
Center and A/P Center at the Axial/Prescription screen.
Add Group
Add
Group
First select a certain scan group, then click on this to add that group just after the
current group.
Split Group
Enter location or image number
OK Cancel
3. Enter the location or image number at which you wish to split the group.
Then, select [OK].
Note : If you enter an invalid number, the warning message will appear. "Image
number is outside of group image range."
First, select a scan group that you wish to delete, then click on this.
Biospy Rx
Biopsy
Rx
When the system decides that some change in scan techniques is needed to
complete the entire scan, this icon lights up in red. At this point you may enter
the following Technique Optimize screen by clicking on it.
Technique Optimize screen example
Tube cooling is needed at image ; 26
Optimize Method :
Group
by
Group
Image
Upfront Group
Delay Delay 1 45
(sec) mA
(sec)
240 46 65
Tube cooling is needed at image : At the top of the screen the message
appears that informs you at which image the tube cooling will be needed with the
current scan techniques. In the above example it is 26.
Optimize Method : Optimize Method tells you how the change in technique will
occur. In the above example it says "Group by Group" meaning the technique
must be changed by scan group. Also, in the above example, since the image
26 belongs to the first group, all the images in the first group from 1 to 45 are
equally subject to changes in technique. The system will calculate and display
up to three factors for the scans to be able to continue. Those three factors are;
- Upfront Delay (sec) : Pre-scan delay
- mA : Tube current used for that scan group
- Group Delay (sec) : Delay time between scan groups
Rectifying any one of those three factors will display a message "Tube cooling is
No Longer Needed" enabling the system to continue scans.
Notice : When the system estimates the cooling time to be more than 600
seconds, it shows "--" in the [Upfront Delay] box.
Finally select [Optimize in Progress] highlighted in blue to exit.
Preview
This function enables you to graphically view prior to the scan the interaction
between scan process and the combination of Breath Hold and Breathe Time.
When you click on this button after axial scan prescriptions, the following screen
appears.
Preview Screen
Prior Next
0 10 20 30 40
Images Type kV mA Time Thk Tilt Start End
Seconds
S80.0 S30.0
001 010 Helical 120 150 10.0 5 S0.0 S80.0 S30.0
S10.0 I50.0
011 022 Helical 120 150 12.0 5 S0.0 S10.0 I50.0
I60.0 I85.0
023 027 Helical 120 150 5.0 5 S0.0 I60.0 I85.0 Biopsy
Rx
Step Scroll
If the prescribed scan sequence is too long for a displayed time scale, use Step
or Scroll button to view the entire sequence for the corresponding time scale.
(Step/Scroll buttons appear on the screen only when too long scan sequence is
prescribed.)
When more than three groups are prescribed, use Prior or Next key to go
through the pages. (Prior or Next key appears on the screen only when more
than three groups are prescribed.)
End Exam
End
Exam
Next Series
Next
Series
Click on this to select another series within the protocol. Please refer to the later
page for the function description.
Select this to open the following menu and create a new series that does not
exist in a protocol.
Scout Axial
OK Cancel
Repeat
Series
Select this to repeat the current series. Please refer to the later page for the
function description.
Priority Recon
Priority
Recon
Select this to give the priority to a certain image to be reconstructed earlier than
other images. Refer to the later page for the details.
One More
One
More
Select this to repeat the last one scan. Please refer to the later page for the
function description.
Confirm
Confirm
Select this icon to proceed to X-ray radiation after confirming all scan
prescriptions. This icon will change to the [Pause] icon shown below during the
scan.
Pause
Pause
Select this icon to temporarily halt the scan. This icon will change to the
[Resume] icon shown on the next page during the pause.
Resume
IV
GI
1. Click on [Show Localizer] located on the upper left corner of the View/Edit
screen.
View/Edit screen
Name : ID :
Protocol : Exam: Series:
Anatomical
Reference
Show
Series Description
Localizer
Auto Show
Store AutomA OFF Localizer
2. The Scout image appears with lines that show prescribed scan locations.
4. In order to adjust the DFOV center, click and hold on the "X" mark in blue
with the Shift key pressed down, then drag the entire group vertically to the
desired location.
5. In order to adjust the location of only one group, first doubleclick on the
group to select, then do the above operation. (When you select a certain
group by doubleclicking on it, it turns blue.)
6. In order to add or delete a slice at the top of the group, click and hold on the
small blue solid square and move the mouse to the desired direction.
If you wish to add or delete a slice at the bottom of the group, click and hold
on the small blue hollow square and move the mouse to the desired
direction.
7. In order to adjust the size of DFOV, click and hold on the small diamond,
then extend it or contract it.
8. In order to adjust the gantry tilt, click and hold on the small circle, then tilt it
to the desired angle.
Tube
Start End
mA
0 Scan location
Start End
1. Click on the [AutomA OFF] located on the upper left corner of the View/
Edit screen. (Default setting is "OFF".)
View/Edit screen
Name : ID :
Protocol : Exam: Series:
Anatomical
Reference
Auto Show
Store AutomA OFF Localizer
Select one of the four modes, IQ, Normal, LowDose or MaxmA. The menu
disappears upon the selection.
- [IQ] : Select this mode when high image quality is needed.
The possible highest mA of this mode is the highest among IQ, Normal and
LowDose modes.
[AutomA OFF]will change to[AutomA IQ].
- [Normal] : This mode is most commonly recommended.
[AutomA OFF] will change to [AutomA Normal].
100
10
Scan location
250
200
mA
150
100
50
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
Slice
250
200
mA
150
100
50
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
Slice
3. The following menu appears upon the selection of [mA] on the View/Edit
screen.
When [AutomA OFF] is posted, the [mA Table] key looks dim indicating it
is not functionable.
Select or enter the desired maximum mA.
When any one of three modes (IQ, Normal, LowDose) is selected, all mA
choice boxes look dim, whereas the [mA Table] key looks solid. Click on
the [mA Table] key to open the following mA Table Information menu.
mA Table Information
Scan# mA
1 150
2 150
3 150
4 140
5 140
6 130
7 110
8 110
9 110
10 110
11 110
OK
(Example)
The above menu shows the expected mA at each scan location.
Click on [OK] to close the menu.
Select Create
End New Next Repeat One Priority
New
Exam Protocol Series Series Series More Recon
Confirm
View/Edit screen
Confirm
2. Press the [Move to Scan] button when lit, then, press the [Start Scan]
button to start a scan.
Scan Progress
Exam:285 Series:2
0 10 20 30 40
Images Type kV mA Time Thk Tilt Start End
S80.0 Seconds
S30.0
001 010 Helical 120 250 10.0 5 S0.0 S80.0 S30.0
S10.0 I50.0
011 022 Helical 120 250 12.0 5 S0.0 S10.0 I50.0
I60.0 I85.0
023 027 Helical 120 250 5.0 5 S0.0 I60.0 I85.0
Biopsy
Rx
Scanning Delay Timer Patient Handling
ProjectedSeriesDLP:
AccumulatedExamDLP:
Note : A warning message will appear in the Patient Handling box if the the
scanning object is off center.
Note : A scan will be automatically aborted unless you press [Start Scan]
within 30 seconds after it is lit.
4. The system automatically performs the prescribed axial or helical scan. You
can monitor the progress of the scan on the above Scan Progress screen.
1. After the scan, click on [End Exam] located on the lower left corner of the
Scan Progress screen.
Scan Progress
Exam:285 Series:2
0 10 20 30 40
Images Type kV mA Time Thk Tilt Start End
Seconds
I60.0 I85.0
023 027 Helical 120 250 5.0 5 S0.0 I60.0 I85.0
Biopsy
Rx
Scanning Delay Timer Patient Handling
ProjectedSeriesDLP: All scans completed.
AccumulatedExamDLP:
End Exam
2. Or, you may select "Return to View/Edit screen" icon located on the right
side of Scan Progress screen to return to the View/Edit screen.
3. Select [End Exam] on the lower left corner of the View/Edit screen to end
the current exam. The screen automatically returns to Top Level screen.
1. In the scan process, if you wish to reconstruct a certain image earlier than
any other image, select [Priority Recon] located on the bottom right of Scan
Progress screen.
Scan Progress
Exam:285 Series:2
0 10 20 30 40
Images Type kV mA Time Thk Tilt Start End
Seconds
I60.0 I85.0
023 027 Helical 120 250 5.0 5 S0.0 I60.0 I85.0
Biopsy
Rx
Scanning Delay Timer Patient Handling
ProjectedSeriesDLP:
AccumulatedExamDLP:
Pause
Priority Recon
The system starts to reconstruct the designated image just after completion
of the current reconstruction.
2. After completion of the whole scan, if you wish to give the reconstruction
priority to the lastly scanned image, select [Priority Recon] on the bottom
right of the View/Edit screen.
Select Create
End New Next Repeat One Priority
New
Exam Protocol Series Series Series More Recon
Confirm
Biopsy
Rx
Scanning Delay Timer Patient Handling
ProjectedSeriesDLP: All scans completed.
AccumulatedExamDLP:
Next Series
2. The system automatically proceeds to the View/Edit screen that shows the
parameters of the series in the protocol. You may modify parameters.
View/Edit Screen
Optimize
not needed
Images Scan Start End No. of Thick Image Gantry CTDIvol
SFOV kV mA
Type Loc. Loc. Images (mm) Int val Tilt DLP
Preview
Biopsy
Rx
ProjectedSeriesDLP: Smart
AccumulatedExamDLP: Pre
RX
Confirm
Biopsy
Rx
Scanning Delay Timer Patient Handling
Repeat Series
2. There appears the following Select Previous Series menu which lists all the
previously scanned series that have the same Patient Position /Orientation
(supine/prone, head first/feet first) as the most recently scanned one. The
most recently scanned series tops the list and gets highlighted.
OK Cancel
The list can show up to ten series on one screen. When it contains more
than ten series, there appear(s) Prior and/or Next button on the screen.
1. After the selection of [OK], the system automatically proceeds to the View/
Edit screen that shows the parameters of the series chosen in the Select
Previous Series menu. You may modify parameters here.
View/Edit Screen
Optimize
not needed
Images Scan Start End No. of Thick Image Gantry CTDIvol
SFOV kV mA
Type Loc. Loc. Images (mm) Int val Tilt DLP
Preview
Biopsy
Rx
ProjectedSeriesDLP: Smart
AccumulatedExamDLP: Pre
RX
Confirm
Scan Progress
Exam:285 Series:2
0 10 20 30 40
Images Type kV mA Time Thk Tilt Start End
Seconds
I60.0 I85.0
023 027 Helical 120 250 5.0 5 S0.0 I60.0 I85.0
Biopsy
Rx
Scanning Delay Timer Patient Handling
ProjectedSeriesDLP: All scans completed
AccumulatedExamDLP:
One More
Scan Progress
Exam:285 Series:2
0 10 20 30 40
Images Type kV mA Time Thk Tilt Start End
Seconds
I60.0 I85.0
023 027 Helical 120 250 5.0 5 S0.0 I60.0 I85.0
Biopsy
Rx
Scanning Delay Timer Patient Handling
ProjectedSeriesDLP: All scans completed
AccumulatedExamDLP:
Biopsy Rx prescription
1. Biopsy Rx can be accessed through the [Biopsy Rx] icon on the right side of
Axial/Helical Prescription screen.
Optimize
Thick Image Gantry CTDIw
SFOV kV mA Rx
(mm) Int’val Tilt (mGy)
Preview
Biopsy Biopsy
Rx Rx
Biopsy Rx
Get Alignment
Biopsy Location Internal External
Light Location
Thickness 1 2 3 5 7 10
Confirm
Cancel
Biopsy Rx
4. When the internal light is used, select [Internal], or, when the external light
is used, select [External] at Get Alignment Light Location field.
7. Press [Move to Scan] button when it lights up. Then, press [Start Scan] to
start the biopsy scans.
Name : ID :
Protocol : Exam: Series:
Anatomical
Reference
Series Description
Auto Show
Store AutomA OFF Localizer
Scan Optimize
Images Start End No. of Thick Interval Gantry CTDIvol
SFOV kV mA not needed
Type Loc. Loc. Images (mm) (mm) Tilt DLP
Preview
Smart
ProjectedSeriesDLP: Prep Smart Prep
AccumulatedExamDLP: RX
Select Create
End Next Repeat One Priority
New New
Exam Protocol Series Series More Recon
Series
Confirm
2. When the Smart Prep is not incorporated in the protocol,toggle the [Off]
button to [On] on the Smart Prep screen.
To [On]
Smart Prep screen
Smart Prep
Off
Accept Cancel
5. Click on [Confirm] and press [Start Scan] when lit to initiate the Baseline
scan.
Scan Progress screen shows the Baseline group,Monitor group and scan
prescription.
Scan Progress
Exam:285 Series:2
0 10 20 30 40 50 60
Images Type kV mA Time Thk Tilt Start End
Seconds
S0.0
001–001 Axial 120 140 2.0 10 S0.0 S0.0 S0.0
S0.0
ProjectedSeriesDLP:
AccumulatedExamDLP:
Pause
Note : When you select [Accept] on the Smart Prep prescription screen,the
following timing menu shows”SP”in the Prep Group(delay)field,meaning the
Smart Prep Monitoring delay has been set.
Scan Timing Recon Film
SP
SP
Display Ellipse
Zoom ROI
Normal
Hide Explicit
Erase
Graphics Mag
Baseline image
9. After you press the [Start Scan] button, the system will automatically
proceed to the Monitoring scans through the Monitoring delay.
• The Display desktop screen will now look similar to the following one.
Most recently
Graph reconstructed image
• In the lower right quadrant the Baseline image with ROIs is displayed.
• The lower left quadrant will be displaying in real time the time when each
monitoring scan is acquired, based on the onset of the monitoring delay.
It also displays each of the ROI values of that scan.
• The upper right quadrant will be displaying in real time the most recently
reconstructed image.
• The upper left quadrant will be displaying in real time the enhancement
threshold graph, comparing the ROI of each monitoring scan with the time
from the start of the monitoring delay. If you did not take any ROI on the
Baseline image,this quadrant will be blank.
Most recently
Graph
reconstructed image
Preview
10. As the monitoring scans are being acquired, you can watch the lines
climbing toward the enhancement threshold on the graph in the upper
left quadrant.
T
T: Threshold
0 4 2
11. When the line depicting the ROI gets close to the threshold
enhancement, select [Scan Phase] on the Scan Progress screen to
initiate the Scan Phase. See the Note below.
Note : If the scan location of Monitoring Phase does not match the start
location of Scan Phase, the scan start will be delayed by the time the cradle
takes to move. For example, it takes some four seconds for the cradle to
move 300 mm. It is highly recommended that the scan location of Monitoring
Phase match the start location of Scan Phase.
Note : When the system initiates the Scan Phase, the real time calculation of
the Smart Prep quadrant will stop. A screen save will be used for later
inspection.
Conditions/Restrictions
• The following conditions/restrictions must be met to use the Smart Addition
option.
• Scan Type : Axial or Helical (360-degree data)
• Segment and Helical Plus incompatible with this option
• Available Helical pitch : 1.0 - 3.0
• This option can be applied to group by group.
• Only contiguous images (Interval = Thickness) to be added
• Possible thickness of added image : 2, 3, 4, 5, 6, 7, 10 mm
• Possible number of images to be added : 2, 3, 4, 5, 6, 7, 10
• When Smart Addition is prescribed for [Recon 1], both [Recon 2] and [Recon
3] will be automatically designated for Smart Addition.
Also, when Smart Addition is not prescribed for [Recon 1], Smart Addition can
not be prescribed for [Recon 2] nor [Recon3].
Procedure
1. Click on the [ Thick (mm)] key in the Scan Tab menu.
The Image Thickness selection menu opens (See next page).
Thick
Scan Timing Recon Film
1 2 3 5 7 10 Cancel
Note : The Smart Addition key is not available unless your CT system has
Smart Addition option installed. Also, even if your CT system has the Smart
Addition option installed, the Smart Addition key remains dim (inactive)
unless the right Scan Type is selected.
2. In order to activate the Smart Addition option, click on the [Smart Addition]
key. Then, the Fig.2 Smart Addition Image Thickness selection menu
opens.
Fig.2 Smart Addition Image Thickness selection menu
2 3 4 5 6 7 10
1 2 3 5 5
OK Cancel
On this menu select Beam Thickness (original image thickness) and Image
Thickness (thickness of added images), then the Multiplication Factor
(number of images to be added) is automatically determined.
Since the Multiplication Factor is an integer, there are restrictions on the
combination of Beam Thickness and Image Thickness. For example, when
2mm is selected as Beam Thickness, Image Thickness of 3, 5, and 7 are
unavailable.
The example of Fig.2 shows 1mm and 5mm are selected as Beam
Thickness and Image Thickness, respectively. So, the Multiplication Factor
is automatically set to 5.
Example
5
1x5
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 Image number
Conditions/Restrictions
• Data obtained with 7mm or 10mm slice thickness can not be used.
• Recon Mode : Full
• Contiguous image data ( no overlap, no gap)
• Available Helical pitch : 1.0 - 3.0
• Two or more images data
When the above conditions/restrictions are not completely observed, the keys
under [ Thick (mm)] field in the Image tab are dimly displayed indicating Smart
Addition retro recon is unavailable.
Thick key
Images Recon
Scan Scan Scan
Retro Start End Retro Retro No. of Thick Image Gantry
Type Location Interval Tilt SFOV
Location Start End Images (mm)
List
Quit Exams Confirm
1 2 3 4 5 6 7 10 Cancel
On the real menu of Fig.1 the valid thickness numbers are displayed boldly
whereas the invalid thickness numbers are displayed dimly depending on
the beam thickness.
2. Select the desired image thickness from the Fig.1 menu or click on
[Cancel].
View/Edit screen
Name : ID :
Protocol: Exam: Series:
Anatomical
Reference
Series Description
Auto Show
Store Localizer
AutoFilm Setup
Autofilm Setup
Destination
Laser Camera
Formats Size Auto Start Auto Print
Normal Slide Yes Yes
Yes Yes
Scout XRef–Scout
Film Direction
Top to Bottom
e/s/i Yes
OK Cancel
Note : Once a scan has been acquired, you can not go back to the Autofilm
Setup page, unless you select a new series or a new protocol.
Description of each function is as follows;
• Format
There are 11 options for film format. Click on one of them.
• Film Direction
You can have the images filmed Top-to-Bottom or Bottom-to-Top. Click on the
blue highlighted arrow to toggle.
• Size
You can choose to film either in the normal setting or slide setting.
• Copies
You can choose how many copies you want the camera to print. Select the
number you want by either clicking in the box and typing in the number or
clicking on arrows to increase or decrease the number. The valid range is
from 1 to 99.
• Scout
This allows you to autofilm a Scout image. Select Scout icon to open the
following menu.
Accept
In order to autofilm a Scout image, select [Yes] first, then enter other
parameters. Lastly select [Accept].
Note : The valid range for Magnification Factor is from 0.5 to 8.0. If a Scout
image exceeds 500mm in length, the magnification factor must be less than 1.0
to view the entire Scout image.
Accept
At Image Range selection, select [All] for axial images and [First/Last] for the
first and last axial image.
Note : The valid range for Magnification Factor is from 0.5 to 8.0. If a Scout
image exceeds 500mm in length, the magnification factor must be less than 1.0
to view the entire Scout image.
After entering all parameters, select [Accept].
Film Set 1
No Auto start Auto start
Auto start New Sheet Same Sheet
Film Set 2
No Auto start Auto start
Auto start New Sheet Same Sheet
Accept Cancel
If you select [No Auto start] on the above menu, the system will not automatically
start filming. In this case the operator is to decide when to start filming.
If Autofilm is on and the Autofilm viewport is showing, once you start scanning,
you can choose from one of the three options at the bottom of the Autofilm
window. The following three icons are those options.
[Start New Sheet] will start a new sheet of film with the format that has been
selected from the Autofilm Setup page.
[Continue Same Sheet] will continue filming on the current sheet in the Autofilm
composer with the same format as the Autofilm composer.
Exam: 4
Series: 1
is ready for auto filming
If you select [Show Film Viewport], the autofilm viewport will be displayed, then
you can select [Start New Sheet], [Continue Same Sheet] or [Cancel Film
Series].
Autofilm viewport
If you select [Autostart New Sheet], the Autofilm process automatically uses
[Start New Sheet] function.
If you select [Autostart Same Sheet], the Autofilm process automatically uses
[Continue Same Sheet] function.
After making all the necessary selections, select [Accept] to continue or select
[Cancel] to cancel. In both cases, the system returns to the Autofilm Setup page.
• Auto Print
[Auto Print] is a toggle button between Yes and No. If you select Yes, then the
last film of the exam will be automatically printed, whether the Autofilm
composer is filled or not. If you select No, then you will need to select Print on
the Autofilm composer.
2. On the Autofilm Setup page, select [OK] to accept all entries. Or, select
[Cancel] to cancel.
1. Click on the Film Tab Card to open the following menu for Autofilming
parameters entry. You may skip this step if you do not intend to perform
Autofilming.
• Auto Film
Select On or Off on the following menu after clicking on [Auto Film].
Auto Film
On Off Cancel
• Frame Format
Select one of the four formats in the following menu by clicking on it.
MID Format
Cancel
• Interval
Select one of the five choices as to which image to film.1= every image, 2=
every other image, 3= every third image, and so on.
Flip
• Width 1
The first choice for window width (1 ~ 4096)
• Level 1
The first choice for window level (-1024 ~ 3072)
• Mag Factor
Valid image magnification factor range is from 0.5 to 4.0
• Rotate
Right 90 degrees, left 90 degrees, or 180 degrees
Rotation
None Cancel
• User Anno(tation)
Maximum four lines
User Annotation
Accept Cancel
Gray Scale
G1 G2 G3 Off
Cancel
Auto Film
Pause Filming
Clear
Print
Film format reflects the one prescribed on the AutoFilm Setup page.
Each port can contain either an image or a set of exam, series and image
number depending on the prescription on the AutoFilm Setup page.
Click on the [AutoFilm Composer] icon again to remove the composer.
[Autoview
[Review Layouts] hjjkk
Layouts]
[Manual Film
Composer]
When you perform Manual Filming, you have to use one or two viewports in the
Review Layouts, or one or two viewports in the bottom of the Autoview Layouts.
Options
Clear
Current status is :
Print
• Laser Camera
The destination of printing will be changed. If you wish to have a destination
other than a laser camera, place the cursor over the word "Laser Camera",
press and hold down the left mouse button to open a pull-down menu, then
select a desired destination.
• Formats
Click on any one of the eleven formats. It will be reflected on the composer.
Print options
Slideformat: Greyscale:
Off Off
Slide format: Select On from the Slide format pull-down menu to employ a
slide format.
Off
On Note : If the selected printing machine does not support slide
format, the Slide format button is shown in gray, indicating it is
not usable.
When filming function keys are being used, this selection will
Expose order:
determine the order with which the images will be printed into
Left/Right
Top/Bottom the frames of the composer.
Right/Left
Bottom/Top
2. There appears a message "Do you really want to delete this image?"
1. Move the cursor onto the desired image and press the F1 key.
2. The image will be automatically placed into the next available frame
Note : This method may be faster than the drag & drop method, however, you
can not jump the frame.
1. Move the cursor over any one of the displayed images and press the F2
key.
2. The displayed images will be loaded onto the Film Composer in order of the
displayed images.
Note : In order to perform page filming, the formats must be the same between
the displayed images and the Film Composer. If the formats are different, the
format of the Film Composer automatically turns to that of the displayed images
as you press the F2 key.
Note : When you are using the Viewer or Mini Viewer, you can also use the [Film
Page <F2>] button to perform page filming.
1. Move the cursor over any one of the displayed images and press the F3
key.
2. The displayed images will be loaded onto the Film Composer in order of the
displayed images.
Note : When you are using the Viewer or Mini Viewer, you can also use the [Film
MID <F3>] button to perform MID filming.
• Format
- Use Film Composer : Select this to use the same format as the current Film
Composer.
- Viewer Format : Select this to change the Film Composer format to that of
Viewer Format.
• Image Selection
This determines the number of images in the series for filming.
Use the slide bar to set the number.
• Interval
Print all Images : all images
After completing all entries, select [Print Series] to start series filming.
Note : If you wish to cancel the filming after clicking on [Print Series], press F4
and click on [Cancel All].
QA Schedule
The most effective Quality Assurance program involves obtaining basic
performance data once a day, or at least 2-3 times per week. You must obtain
data frequently and on a regular basis in order to detect any changes in system
performance that might occur before it effects clinical image quality. At minimum,
acquire a single 10mm scan of Sections 1 and 3 of the Performance Phantom
each day.
Grayscale:
Select On from the Grayscale pull-down menu to print a
grayscale onto a film.
Off
On
Note : If the selected printing machine does not support
grayscale, the Grayscale button is shown in gray, indicating it is
not usable.
2. Align the coronal light to the horizontal lines on either side of the phantom.
3. Align the sagittal light (where it strikes the top of the phantom) to the
vertical line on the face of the phantom.
4. Position the phantom and press the Internal Land button on the gantry.
The performance phantom contains three sections. When you correctly follow
the positioning instructions listed above, Section 1 corresponds to 0.0 mm table
position, Section 2 (Low Contrast Detectability) to the 35.0 mm location and
Section 3 (Noise and Uniformity) corresponds to the 50.0 mm location.
Circumferential
reference line
Vertical reference line
Horizontal
reference line
Horizontal
reference line
TABLE 1
SOFTKEY PROMPT SCAN VALUE ECONSTRUCTION
[NEW PATIENT] kV - 120 CAL FOV - 25cm
[HEAD FIRST] mA - 100 Recon FOV - 25cm
Time - 3 sec centered
[HEAD] Thickness - 10mm*
[SINGLE SCAN] Scan Mode - Single Recon Mode - Standard
2. Position the cursor on the Plexiglass block and click the left mouse button
once to calculate the ROI. Record the mean CT number on the QA Data
Form. (Standard deviation record optional.)
3. Position the cursor over the water section and click the left mouse button
once to calculate the ROI. Record the mean CT number for water on the
QA Data form. (Standard deviation record optional.)
FIGURE 1
2. Display and position a box cursor from [Box ROI] over the largest (1.6 mm)
bar pattern. The cursor should fit within the bar pattern as shown in Figure
2. Adjust the size and position of the cursor as necessary.
3. Click the left mouse button once to calculate the ROI and record the
standard deviation on the QA data form.
4. (Optional) Repeat this procedure for the 1.3, 1.0, and 0.8 mm bar patterns.
Optional: repeat
for 1.3mm pat-
tern
Optional: repeat
for 1.0mm pat-
tern
Optional: repeat
for 0.8mm pat-
tern
FIGURE 2
Air filled
holes
FIGURE 3
The resolution block contains holes drilled 1 mm apart and aligned in the
direction of slice thickness (perpendicular to scan plane). Each visible hole in the
image represents 1 mm of beam thickness. The software assigns less negative
CT numbers to partial hole images or holes located on the edge of the slice
profile. To determine slice thickness, display the image at the
recommended window level and width, and count the visible holes. Black
holes in the image represent a full millimeter of slice thickness. Gray holes count
as fractions of a millimeter; two equally gray holes count as a single 1 mm slice
thickness.
Recommended window width : 300. Recommended window level : -100 for
3.0 mm slices, 0 for 5.0 mm, and +50 for 10.0 mm slices. Your image may show
less detail than this example.
FIGURE 5
2. Display and position a box cursor from [Box ROI] over the image. Adjust the
cursor to a rectangle, approximately 1/2 cm high by 5 cm long, as shown in
Figure 6.
3. First position the cursor over the polystyrene membrane above the holes.
Click the left mouse button once to calculate the ROI. Record the mean CT
number in the Low Contrast resolution box on the QA Data Form.
4. Next place the cursor in the water section above the membrane and click
the left mouse button once to calculate the ROI. Record the mean CT
number.
5. Subtract the CT number of the water from the CT number of the membrane
and record the difference.
7. Repeat steps 3, 4, and 5. This time position the cursor below the
membrane holes, then move it over the water area below the membrane.
3. Click the left mouse button once to calculate the ROI. Record the mean CT
number and standard deviation on the QA Data Form.
Optional: Take an
ROI at the 3 o’clock
position.
FIGURE 7
Contrast Scale
The difference in CT numbers between the Plexiglass resolution block and water
should equal 118, with a suggested allowable variation of 10%.
HEAD BODY
25 cm Field of View (FOV) 43 cm FOV
120 kVp 120 kVp
150 mA 130 mA
2.0 sec scan time 1.5 sec scan time
10 mm slice thickness 10 mm slice thickness
Large Focus Large Focus
HEAD BODY
25 cm Field of View (FOV) 43 cm FOV
120 kV 120 kV
150 mA 130 mA
2.0 sec scan time 1.5 sec scan time
10 mm slice thickness 10 mm slice thickness
10 mm interval 10 mm interval
If the Helical mAs, Slice Count and Intvl selection equals the Axial mAs, Slice
Count and Intvl selection, then Helical dose equals Axial dose.
HEAD BODY
25 cm Field of View (FOV) 43 cm FOV
120 kV 120 kV
150 mA 130 mA
2.0 sec scan time 1.5 sec scan time
10 mm slice thickness 10 mm slice thickness
Large Focus Large Focus
0.5
0
0 Position (mm) 70 140
Dose Profile Sensitivity Profile
1.0
Axial, Body, Center,
120 kV, 130 mA,
7 mm, 1.5 sec.
0.5
0
0 Position (mm) 70 140
1.0
Axial, Body, Center,
120 kV, 130 mA,
5 mm, 1.5 sec.
0.5
0
0 Position (mm) 70 140
1.0
Axial, Body, Center,
120 kV, 130 mA,
3 mm, 1.5 sec.
0.5
0
0 Position (mm) 70 140
Dose Profile Sensitivity Profile
1.0
Axial, Body, Center,
120 kV, 130 mA,
2 mm, 1.5 sec.
0.5
0
0 Position (mm) 70 140
1.0
Axial, Body, Center,
120 kV, 130 mA,
1 mm, 1.5 sec.
0.5
0
0 Position (mm) 70 140
1.0
Axial, Head, Center,
120 kV, 150 mA,
10 mm, 2.0 sec.
0.5
0
0 Position (mm) 70 140
Dose Profile Sensitivity Profile
1.0
Axial, Head, Center,
120 kV, 150 mA,
7 mm, 2.0 sec.
0.5
0
0 Position (mm) 70 140
1.0
Axial, Head, Center,
120 kV, 150 mA,
5 mm, 2.0 sec.
0.5
0
0 Position (mm) 70 140
0.5
0
0 Position (mm) 70 140
Dose Profile Sensitivity Profile
1.0
Axial, Head, Center,
120 kV, 150 mA,
2 mm, 2.0 sec.
0.5
0
0 Position (mm) 70 140
1.0
Axial, Head, Center,
120 kV, 150 mA,
1 mm, 2.0 sec.
0.5
0
0 Position (mm) 70 140
Noise
At Typical Technique In Center Of Phantom Using Standard Algorithm
HEAD BODY
σ = 0.49 % σ = 0.47 %
MTF
(same conditions as above)
HEAD BODY
100 100
M M
O 80 O 80
D D
U U
L 60 L 60
A A
T T
I 40 I 40
O O
N N
20 20
(%) (%)
0 2 4 6 8 10 0 2 4 6 8 10
Line pairs/cm Line pairs/cm
HEAD BODY
10.0 mm 10.0 mm
7.0 mm 7.0 mm
5.0 mm 5.0 mm
3.0 mm 3.0 mm
2.0 mm 2.0 mm
1.0 mm 1.0 mm
Sensitivity Profile
See previous pages
Dose
For best results, use the phantoms, dose profile and CTDI value calculation
procedures recommended in the CDRH final draft of "Routine Compliance
Testing for Computed Tomography X-Ray Systems" dated April 26, 1984.
Also, for best results, use the phantoms and CTDIw value calculation
procedures recommended in the IEC committee draft for vote of IEC60601-2-44,
dated August 8, 1997.
Performance
Each test uses a 25 cm water-filled acrylic phantom
Noise
Noise equals the standard deviation of an array of pixels contained in 674 mm
square region of interest (ROI) for Head and 2696 mm square ROI for Body. The
software divides the standard deviation, expressed in Houndsfield Units, by
1000 (representing the contrast scale between air and water), then multiplies by
100 to give a value in percent.
Slice
Use a wire ramp section of Catphan phantom, inclined 23 from the scan plane.
Sensitivity
23° from the scan plane to obtain sensitivity profiles.
Dose Profile
Anticipate a "maximum deviation" of ± 30% or 2.0mm, whichever is larger,
relating to dose profiles (FWHM). This value includes variability inherent in the
measurement of dose profile with TLD chips.
Performance
Noise : The noise squared (σ2) in a CT image is inversely proportional to the x-
ray dose used to make the image. The maximum deviation anticipated for image
noise equals ± 30%. Expected deviation equals ± 10%.
MTF : Expect deviations within ± 10% for values on the MTF curve generated
with data gathered according protocol. Maximum deviations may reach ± 20%
for other methodologies.
Sensitivity Profile : Expect the full width slice half maximum sensitivity profiles
to vary ± 20% or 1.0 mm, whichever is larger, when measured with a wire ramp
section of Catphan phantom, inclined 23° from the scan plane. If you use other
methodologies, the maximum deviation may reach 1.5 mm for all slice
thicknesses, because these measurement errors have the greatest effect on thin
slices.
CTDIfree air
Statement of Typical Technique:
Head
77mGy ±40% Axial,Head sFOV,25cm dFOV,120kV,150mA,2s,10mm
CTDIfree air
Statement of Typical Technique:
Body
50mGy ±40% Axial, Large sFOV, 43cm dFOV, 120kV, 130mA, 1.5s, 10mm
68mGy ±40% Axial, Large sFOV, 43cm dFOV, 140kV, 130mA, 1.5s, 10mm
58mGy ±40% Axial, Large sFOV, 43cm dFOV, 120kV, 130mA, 1.5s, 1mm
58mGy ±40% Axial, Large sFOV, 43cm dFOV, 120kV, 130mA, 1.5s, 2mm
54mGy ±40% Axial, Large sFOV, 43cm dFOV, 120kV, 130mA, 1.5s, 3m
52mGy ±40% Axial, Large sFOV, 43cm dFOV, 120kV, 130mA, 1.5s, 5mm
50mGy ±40% Axial, Large sFOV, 43cm dFOV, 120kV, 130mA, 1.5s, 7mm
Spatial Resolution
Std
• 3.6 lp/cm±20% lp/cm@ 50%MTF Axial/1.5S/10mm /120kV/150mA/Head
• 6.4 lp/cm±20% lp/cm@ 10%MTF
sFOV/25cm dFOV/STD
HiRes
Axial/1.5S/10mm /120kV/150mA/Head
• 11.2 lp/cm±20% lp/cm@ 50%MTF
• 14.1 lp/cm±20% lpcm@ 10%MTF sFOV/10cm dFOV/PFRM
1.3mm MEMBRANE
WATER
1.0mm DIFFERENCE
DIFFERENCE HOLES
0.8mm
1.3mm MEMBRANE
WATER
1.0mm DIFFERENCE
DIFFERENCE HOLES
0.8mm
MODEL (Note 5)
MODEL (Note 1)
SERIAL
SERIAL
MANUFACTURED CLASS I
MANUFACTURED CLASS I
VOLTS (Note 6) V~
SOURCE: (Note 2) A/ (Note 3) V ~(Note 4) HZ
AMPS MOMENTARY (Note 7) A
SOURCE: (Note 2) A/ (Note 3) V ~(Note 4) HZ
AMPS CONTINUOUS (Note 8) A
SOURCE: (Note 2) A/ (Note 3) V ~(Note 4) HZ
kVA (Note 9) HZ (Note 10)
SOURCE: (Note 2) A/ (Note 3) V ~(Note 4) HZ
Improvements in the heat handling capabilities of this unit may result in new
catalog and Model number assignments
The following table lists cooling delay times (in seconds) used by the software
before each scan. These times assume maximum anode or tube unit heat
loading and include cycling the rotor for each scan.
Heating Curves
Stored Heat (KJ)
Time(sec) Time(min)
Stored Heat(KJ)
2500 2000
Storage (kJ)
2000 1500
1500 1kW
2 kW 1000
3.4 kW
1000
500
500
0
00 200 400 600 800 1000 1200 0 10 20 30 40 50 60
Note : Cooling and heating curves reflect maximum tube performance. System
software monitors and controls tube operation.
Classification
IEC publication 536 /1978 and IEC Publication 601-1 Class 1
Transport Packaging
Transport ONLY in packaging supplied by General Electric
Measurement Basis
Tube Potential : Precision voltage divider, Model No. 46-154966G1 reduces
high voltage generated across anode and cathode by 1000:1. Tube Potential
equals the average kilovoltage generated during the exposure, excluding
transients at the beginning and end. See Section 2 of the X-ray Alignment
procedure in the Service documentation for the complete measurement
procedure.
Tube Current : Tube current equals the average milliamperage generated
during x-ray exposure. See Section 2 of the X-Ray Alignment procedure in the
Service documentation for the complete procedure.
Scan Time : Exposure time intervals equal the time during which the kilovoltage
equals or exceeds 75% of its peak value. Measure kilovoltage with a precision
voltage divider and an oscilloscope. Use the time base of the oscilloscope to
measure exposure duration.
Priority Code
Pr (Priority) - Description
1 Safety and Rgulatory
2 Image Quality
3 Procedures that make the system unavailable for scanning
4 Procedures that can be performed while the customer is
scanning
Frequency Code
Fr (Frequency) - Description
S Semi-Annually (every 6 months)
A Annually (every 12 months)
System Performance
Annual PM
SUB-
ITEM DESCRIPTION Pr Fr Schedule
SYSTEM
A B
PM0106 System Perform Filter Curve Test 3 A 0:15
PM0107 System Check Ground Cable Terminals 3 A 0:10
PM0108 System Check DAS Count 3 A 0:15
PM0109 System Image Performance 2 A 1:00
PM0202 O.C Clean Air Filter 3 S 0:10 0:10
PM0203 O.C Inspect FANs 3 S 0:10 0:10
PM0204 O.C Verify Mouse for smooth operation 3 S 0:10 0:10
PM0205 O.C Clean Display monitor/Console exterior 4 S 0:15 0:15
PM0206 O.C Check Cables / Power Cable Terminals 3 S 0:10 0:10
PM0303 Gantry Verify cables and hardware are tight 3 S 0:20 0:20
PM0304 Gantry Gantry cover cleaning 4 S 0:10 0:10
PM0305 Gantry Check Positioning lights 3 S 0:10 0:10
PM0306 Gantry Check Gantry anchor / Gantry Isolation 3 A 0:05
PM0307 Gantry Main bearing grease-up 3 A 0:20
PM0308 Gantry Check drive belt for wear 3 A 0:05
PM0309 Gantry Check RF Shoe Position 3 A 0:10 0:10
PM0402 DAS Verify Detector Heater Control 3 S 0:05 0:05
PM0403 DAS Inspect FANs 3 S 0:05 0:05
PM0509 Table Clean Cradle tray / Table cover 4 S 0:15 0:15
PM0510 Table Check for Oil Leak 3 S 0:05 0:05
PM0511 Table Check Table anchors / Table isolation 3 A 0:05
PM0512 Table Inspect FANs 3 S 0:10 0:10
PM0513 Table Inspect Cradle Wire Tension 3 S 0:10 0:10
PM0601 XG Check/Clean Radiator FAN 3 S 0:10 0:10
PM0602 XG Check X-ray tube oil and HV connector 3 S 0:20 0:20
PM0603 XG Check HV Tank and HV connector 3 S 0:20 0:20
PM0604 XG Check Power cable connection 3 S 0:05 0:05
PM0605 XG Check Tube over heat safety 3 S 0:15 0:15
PM0606 XG Check KV and mA 3 A 1:00
PM0701 PDU Inspect FANs 3 S 0:05 0:05
PM0702 PDU Check Power Cable Terminals 3 A 0:05
PM0703 PDU Check Power Line Voltage 3 A 0:05
PM0801 MOD Dry Cleaning 3 S 0:05 0:05
Subtotal 4:15 7:00
Total 5:45 8:30
*1 : Option
~
Symbol Publication Description
~
3 335-1 Three-phase alternating current
3N
~ 335-1
Three-phase alternating current
with neutral conductor
Direct Current
Attention, consult
348
ACCOMPANYING DOCUMENTS
ON (Power: connection
417-5007
to the mains)
Dangerous voltage
Emergency Stop
Warning sign
RADIATION of LASER
APPARATUS
417-5009 Stand by
Start
Table Set
Abort
Intercom
Microphone (Mic)
Contrast
Brightness
70oCC
50
System storage prior to installation:
Maintain storage temperature between -20° C and
+70° C
-20oC
Humidity 10-90%
Excluding DO NOT store system longer than 90 days
Condensation
Air Pressure System storage and shipment
700-1060hPa Maintain Air Pressure between 700 and 1060hPa
Type B Equipment
CLASS I, II, or III EQUIPMENT or EQUIPMENT with INTERNAL ELECTRICAL
POWER SOURCES provide an adequate degree of protection against electric
shock arising from (allowable) LEAKAGE CURRENTS or a breakdown in the
reliability of the protective earth connection
Ordinary Equipment
Enclosed EQUIPMENT without protection against the ingress of water.
Operation Of Equipment
CONTINUOUS OPERATION WITH INTERMITTENT LOADING.
Operation in which EQUIPMENT is connected continuously to the SUPPLY
MAINS. The stated permissible loading time is so short that the long term on-
load operating temperature is not attained. The ensuing interval in loading is,
however, not sufficiently long for cooling down to the long term no-load operating
temperature.
EQUIPMENT not suitable for use in the presence of a FLAMMABLE
ANESTHETIC MIXTURE WITH AIR or WITH OXYGEN or NITROUS OXIDE.
P
G Patient Positioning, 4-1
Gantry, 1-2 Landmarks, 4-3
Control Panel, 1-2 Patient Schedule, 2-6
Detector Specifications, 1-3 Add Patient, 2-8
Display Panel, 1-2 Delete All, 2-10
Emergency Button, 1-2 Delete Selected, 2-10
Gantry Specifications, 1-3 Edit Patient, 2-9
Positioning Light & Breath Navi, 1-2 Preference, 2-11
X-ray Tube Specifications, 1-3 Select Patient, 2-7
View More Info, 2-7
Primary/Secondary Viewport, 3-2
Protocol Management, 2-12 ProView, 3-13
Auto Voice Record, 2-13 Roam/Zoom, 3-10
Protocol Management, 2-16 Screen Save, 3-20
[Film] icon, 2-18 User Annotation, 3-18
[Recon] icon, 2-18
[Scan] icon, 2-18
[Timing] icon, 2-18 S
Scanner Utilities, 2-28
Q Raw Data Function, 2-28
User Calibration, 2-34
QUALITY ASSURANCE, 6-1
Dose and Performance, 6-13, 6-15 Scout Scan, 4-6
High Contrast Spatial Resolution, Add Scout, 4-7
6-6 Confirm, 4-8
Image Performance, 6-21 Create New Series, 4-8
Low Contrast Detectability, 6-9 Delete Selected Scout, 4-7
Noise and Uniformity, 6-10 End Exam, 4-7
Phantom Description, 6-2 Next Series, 4-7
Phantom Setup, 6-3 One More, 4-8
Scan the QA Phantom, 6-4 Pause, 4-8
Slice Thickness, 6-7 Repeat Series, 4-8
Weighted CTDI, 6-12 Resume, 4-8
Scan Progress screen, 4-9
Select New Protocol, 4-7
R System Power On/Off, 1-22