11-MZ0-1 Tensile Module 38mm
11-MZ0-1 Tensile Module 38mm
11-MZ0-1 Tensile Module 38mm
1 Introduction 7
1.1 Shipment Volume ........................................................................................................... 8
1.2 Details to this User Manual ............................................................................................ 8
1.2.1 Illustration ................................................................................................................................................ 8
1.2.2 Symbols .................................................................................................................................................... 8
4 Operating Manual 45
4.1 Clamping a specimen in the micro-vise clamps ............................................................ 46
4.2 Clamping a Bending Specimen...................................................................................... 49
4.3 Use of Extensometer .................................................................................................... 50
4.4 Switching On ................................................................................................................. 50
4.5 MDS Software ............................................................................................................... 51
5 Further Information 53
5.1 Troubleshooting ............................................................................................................ 54
5.2 Technical Data............................................................................................................... 55
5.2.1 Tensile/Compression Module ................................................................................................................. 55
5.2.2 Load Cells ................................................................................................................................................ 55
5.2.3 Extensometers ........................................................................................................................................ 56
Figures
Modular Concept of the Module...................................................................................................... 17
End stop limiting the range of travel of the yokes during tensile testing ........................................ 18
Outer dimensions of the tensile/compression module (narrow version) ....................................... 19
Specimen mounted in the tensile/compression module ................................................................. 19
Schematic Setup of an Extensometer .............................................................................................. 22
Scope of delivery of an axial Extensometer ..................................................................................... 23
Tables
Specimens and Clamps ..................................................................................................................... 21
Troubleshooting ............................................................................................................................... 54
Technical Data of the Tensile/Compression Module ....................................................................... 55
Technical Data of the load cell ......................................................................................................... 56
Technical Data of the extensometer ................................................................................................ 57
Contact Information ......................................................................................................................... 58
1 Introduction
Overview
This chapter informs about:
- Shipment Volume
- Details to this User Manual
- Legal Matters
- Safety
1.2.1 Illustration
All illustrations in this manual are intended to assist the user in the understanding of described
tasks or working principles of the module. Certain aspects of illustrations may differ slightly from
some specific models of the module.
1.2.2 Symbols
The following symbols and notations are used in this manual and it is important that the user is
familiar with the meaning of the symbols and the associated instructions.
Symbols and associated instructions fall into three categories. Each category is color coded for
ease of recognition. Each category is explained below.
The terms DANGER, WARNING and CAUTION are used throughout this document. It is important
that the user is familiar with the meaning of those terms in the context of this document and the
meaning of the hazard symbols associated with potential hazards.
1.3.2 Copyright
This user manual and all drawings therein fall within the scope of Kammrath & Weiss Intellectual
Property Rights. Without written consent, neither any part, nor the entire contents are permitted
to be made available to third parties, regardless of the means of communication.
1.5 Safety
This Kammrath & Weiss module was designed and manufactured according to today’s
state-of-the-art and safety regulations. Besides all safety precautions described, any other
workplace hazards require close attention when working with the instrument.
The following general safety instructions are to be observed by all users of the instrument:
Do not attempt any repairs other than those described in this manual. If
any difficulties are encountered during operation, which cannot be
resolved with the instructions in this manual, contact Kammrath & Weiss
GmbH.
If the microscope is equipped with an immersion lens, ensure that there
are no loose components that could be attracted by the magnetic field of
the immersion lens and damage the microscope.
Switch off the controller after finishing work with the instrument.
1.6.1 Noise
The noise level of the overall module is <70 dB(A) and ear protection is not required during
operation.
Overview
This chapter informs about:
- Function of the instrument
- EBSD
- Module description
- Specimen types
- Extensometer
- Electrical connections
2.1 Functions
The tensile/compression module is suitable for material tests on samples using a tensile or
compressive load up to 10kN, depending on the load cell installed. Tests may be carried out under
light, scanning electron microscopy or force microscopy.
The following observations can be made on a material sample:
- Static and dynamic observation of surface changes under tensile/compressive loads
- Crack propagation
- Delamination phenomena
- Formation of sliding areas
The following applications are available:
- Testing in tensile and compression through zero,
- Load cell ranges: 1 to 10’000 N with corresponding clamping devices, appropriate to the
maximum required loads.
- Testing under light (LM), scanning electron (SEM), focused ion-beam (FIB), scanning
ultrasonic (acoustic) (SAM) and atomic force microscopy, (AFM). Please consult Kammrath
& Weiss GmbH for information about the compatibility of specific microscope designs
with the tensile/compression modules.
In applications in which the module is mounted inside the SEM, the module is placed in the same
way that a large sample would be positioned on the sample stage. A special vacuum flange on the
chamber wall provides electrical connection between the module and the controller. All required
test settings and the performance of tests are carried out under computer control. When
operating at atmosphere, the flange can remain on the SEM and the module can be connected
directly to the control unit.
There are different modes of operation. For example, during tensile testing the specimen is pulled
apart until it breaks, allowing crack formation to be observed during the test. During a
tensile/compression test the specimen is stretched and compressed, a number of times, so that
fatiguing of the material can be observed. If testing is completed, the module can be removed,
permitting the microscope to return to its normal operating mode.
The module is suitable for sample dimensions of 50 mm to 60 mm overall length, including the
specimen ends, prepared for clamping. Specimens can be up to 10 mm wide. For applications
under heating and cooling, the sample thickness should be about 1mm. Specimens for tensile-
only experiments should have a mounting hole at each end. Other sample dimensions may be
considered. Please consult Kammrath & Weiss for information on sample compatibility.
The tensile movement is performed by two parallel lead screws, one with a left-hand thread pitch,
and one with a right-hand thread pitch. Bronze nuts in two opposing yokes thread onto these lead
screws and apply the compression or tensile load uniformly from both ends of the specimen.
Applying the force at both ends helps to keep the area of interest positioned under the beam.
Application of the force at both ends of the specimen means that the SEM stage, on which the
tensile/compression module is mounted, only needs minimal movement to keep the area of
observation in view during the tensile/compression test.
The load frame is very important to the mechanical performance of the assembly and it was
designed for maximum stiffness. The load frame contains the following elements:
- Two parallel yokes, on opposite sides of the force rectangle. The specimen is mounted
between them.
- Two left/right leadscrews. Their opposing pitch causes the yokes to move symmetrically
away from or toward each other during actuation.
The load cell measures the applied load and is firmly mounted to the load frame at the rear yoke.
Load cells of various force ranges can be installed. The load cell allows force measurements from
zero to 5,000 N, or from plus 5,000 N to minus 5,000 N, when the tension/compression option is
supplied. Higher range load cells up to 10,000 N may be installed.
The module is equipped with a linear encoder displacement gauge, covering a total displacement
range up to 100 mm, with a resolution of 100 nm. The linear encoder displacement gauge has a
laser-engraved linear scale on it. The linear encoding of the gauge has the advantage that it is
linear over the full range of elongation.
The leadscrews are driven by a DC Tacho-motor, followed by a self-locking, dual reduction gear
assembly with a ratio of 1:40:40. Other speed options are available upon request. The turning
direction determines whether tensile or compression motion is activated.
A mechanical end switch limits the total range of travel of the yokes away from each other during
a tensile test. This ensures that the module cannot be damaged by the yokes being driven beyond
the permitted range. When the end switch is initiated, the voltage to the motor is interrupted.
During compression, the force exerted by the yoke is measured by the load cell. When the load
cell force limit is reached, the motor will be stopped.
The tensile/compression module is designed in such a way that the tensile or compression forces
are applied from each side coaxially. This is particularly important when applying large forces, to
avoid applying any bending or torsion forces to the specimen or to the frame.
Consequently, the load cell and the sample constraints are co-linear with the lead screws, which
actuate movement from outside the yoke rectangle. The load force is concentrated on a very
small area, giving significant freedom in the design of the mechanical components outside the
yoke rectangle.
All mechanical parts outside the yoke force rectangle have no forces applied to them. All forces
are applied within the force rectangle.
Figure 2: End stop limiting the range of travel of the yokes during tensile testing
In some circumstances therefore, special adaption may need to be developed for certain
applications in some specific SEMs, whereby the module is restricted to occupy a limited space
within the SEM chamber, to avoid any collision with any SEM detector
The diagram below shows the dimensions of the SEM version of the module:
A typical tensile specimen has a hole on each end, into which pins are inserted for the tensile test.
A compression specimen is normally mounted in a vice and does not need a hole. A compact
tension (CT) specimen is mounted on both ends and a force is applied to both ends of the
specimen. For thin steel specimens, a notch is pre-fabricated in the specimen, so that crack
propagation and elongation will occur in the region of interest. A bending specimen is a piece of a
specific material that is clamped in a bending clamp. The clamps are available for 3-point and 4-
point tests. Some specimens and the corresponding clamps are shown in Table 1. More clamping
devices are available on request. Please get in touch with Kammrath & Weiss for more
information.
Thickness = 1 to 4 mm
Thickness = 1 to 4 mm
2.6 Extensometer1
An Extensometer are used for the measurement of elongation during tensile and compression
tests under static, quasi-static and dynamic load acting on a specimen. Each transducer consists of
a basic unit and accessories to suit the application. The basic units are identified by their nominal
gauge length as specified in the Table 6. Easy-to-apply clamping springs ensure reliable fitting
even to specimens with rough and uneven surfaces.
The strain transducers operate on the ‘transverse beam measuring principle’. Strain gauges are
applied to the cross member of a U-shaped, precision-manufactured measuring spring, and
connected into a Wheatstone bridge. The two legs of the measuring spring are elastically
deformable in such a way that despite the very small overall dimensions, long measuring
distances may be obtained. Increase or reduction of the distance between the ends of the legs
causes bending of the cross member of the measuring spring, which results in a proportional
electrical change of resistance in the strain gauges. This is output in the form of an electrical signal
by a signal conditioner.
The accessories of the axial extensometer consist of two edges, two stroke limiting blocks, a set of
clamping springs and a gauge block or locking pin. The 60° edges are hardened and precision-
ground. They are used for round and flat specimens and suitable for most materials. Special edges
are available for specific testing problems. All edges have longitudinal bores for accommodation
of a clamping spring. In addition, there are flats at the edges for attaching the stroke limiting block
and the guides for the gauge block. Together with stroke limiting blocks, which are to restrict the
stroke in both directions, and thus to avoid a mechanical overload of the transducer, the edges
are screwed onto the legs of the measuring spring.
For attachment of the transducer onto the specimen, clamping springs are moved onto the
longitudinal bores of the edges, and around the specimen. This is considered to be very easy and
safe method of fastening. The standard delivery contains a set of clamping springs for a thickness
of specimen of 1-3 mm, 3-6 mm, 6-10 mm, and 10-18 mm.
1
Extracted from Instruction Manual of Sandner
Test Frequency
Small design and low weight results in a high natural frequency, and thus, in a high test frequency.
Natural frequencies cause interfering signals which depends on the accelerating forces acting on
the transducer. Usually, test frequencies from 0.2 to 0.3 of natural frequency may be obtained
without noticeable disturbance of signals.
Overview
This chapter informs about:
- Initial installation and commissioning
- Maintenance
- Transport and storage
- The load cell
- The 1N load cell
- The specimen clamping devices
- Specimen geometry
The flange can remain permanently on the microscope after the initial installation.
3.2 Maintenance
Various mechanical components and assemblies require periodic lubrication. It is very important
to the performance of the system that the correct lubricants are used:
Component Lubrication
Do not make any modifications to the mounting of the yoke, the gears
and lead screws. Careful alignment was carried out prior to delivery of
the system. Any changes can lead to serious damage to the system,
especially when specimens are under tensile or compressive load.
Do not disable, over-ride or modify the mechanical end stops, which
limit the travel of the yokes during the test run.
Only install load cells that operate within the stated operating range.
Ensure that the individual load limits for any installed load cell are never
exceeded. Overloading can cause irreversible damage to load cells,
gearing, lead screws and to the frame of the module.
Never disconnect the load cell from the load cell holder. If the load cell
unit is removed from the holder, or if the securing nut fastening the load
cell into the holder is loosened, the calibration data may be invalidated
and the accuracy of the load cell may be affected.
The controller must be switched off before exchanging the sensor plug.
When the load cell is being exchanged for a load cell with a different measurement range, it is of
vital importance to exchange the sensor plug. Both the load cell and the sensor plug have a serial
number. Both serial numbers must match; otherwise the load cell is not calibrated. The load cell
and sensor plug have been pre-calibrated together in the factory.
The following figures show the tension/compression module with the load cell unit mounted and
disassembled from the module.
1. Insert and clamp a rigid specimen and 2. Loosen the nut about 1 turn, with the
apply the maximum tensile force for the supplied 3 mm tool.
built-in load cell.
3. Relax the load on the specimen to 0 N and 4. Remove the assembly aid by unscrewing
remove the specimen from the restraints. the two M5 screws.
5. Remove the four fastening screws from the load cell holder. Hold the load cell unit in
position with one hand.
Switch off the controller before continuing with the next step.
The load cell unit should be stored in a clean, dry place, preferably in the
small case in which it was shipped, if available.
A sensor plug is supplied with the load cell, on which the calibrated sensitivity of the load cell is
stored. This sensor plug is exchanged together with the load cell and ensures that the correct
calibrated settings are automatically adjusted.
For further information about the sensor plug see the DDS3 manual.
The pre-adjustment of the load cell and the load cell holder is carried out
at Kammrath & Weiss prior to shipment. Both parts comprise the load
cell unit. This connection between the load cell and the load cell holder
should not be loosened by the user and should be viewed as a sealed
unit. The small sealing stamp should always be intact, indicating that this
unit has not been disconnected. If the cell and the holder are
disconnected, the unit could need to be shipped back to Kammrath &
Weiss for re-alignment.
If the wrong sensor plug is used with a load cell, the wrong adjustment
will be used and load values may be applied that distinctly differ from
displayed values. This can easily lead to damage to the load cell, lead
screws, gearing or the frame of the module. Both the load cell and the
sensor plug must have the same serial number.
Alignment pins are provided, to ensure that the load cell assembly is positioned precisely aligned
between the lead screws. Thus, the indicated load is a true representation of the applied load.
The alignment pins minimize friction effects, which would affect the quality of the load
measurement during a test run.
If the load cell is aligned out of axis, a bending force would be applied,
leading to an inaccurate measurement of the applied load. Great care is
taken at the factory to ensure precise alignment of the load cell. The
alignment pins supplied enable the user to place the load cell exactly
into its correct position.
The clamping device should move freely. There should be a very narrow gap all the way around
the clamping device to achieve frictionless movement. However, a small frictional force is
unavoidable. When high forces are applied to the specimen, the frictional force will clearly be
insignificant, relative to the applied force. For very low load cells, ball bearings are incorporated,
yielding even lower frictional forces.
The following steps should be followed to install a new load cell unit:
1. Screw in the load cell unit (marked in blue) clockwise, until the top surface is aligned
parallel with the top of the yoke, with a small gap remaining between the two parts. The
guide groove for the cables of the load cell must be at the bottom.
2. Insert the alignement pins into the holes 3. Attach the assembly aid by screwing in
provided. the two M5 screws.
4. Tighten the four mounting screws for 5. Plug the sensor connectors into the
the load cell assembly motor connector block.
Before installing a load cell assembly, ensure that the correct sensor plug
for the load cell is installed.
6. Insert and clamp the supplied rigid 7. Tighten the nut with the supplied 3 mm
specimen and apply the maximum tool.
permissible tensile force for the built-in
load cell.
8. Reduce the force to 0 N and remove the 9. Set the load value to 0 N (“Zero Adjust”)
specimen. to correct the offset value.
1. Insert the shaft into the pin holder and 2. Push the clamping bush onto the shaft and
tighten the fixing screw. tighten it with the grub screw.
3. When the shaft and the bush are correctly mounted in the pin holder, the ground ceramic
capillary tube can be inserted into the clamping bush (the hole on the tapered part of the
bush must face upwards). A drop of red sealing wax is placed into the hole in the bush fixing
the ceramic tube in position.
To loosen the tube again, hold a hot soldering iron tip to the sealing wax.
This softens the wax and the ceramic can be pulled out with tweezers.
1. Align the wire exactly centrally under the 2. Attach a standard weight guided over the
first clamping plate and fix it in place by pulley on the fulcrum. The wire must be
tightening the screws. exactly aligned with respect to the four
magnets – this can be achieved by gently
moving the fulcrum back and forth. The
weight produces a preload, necessary for
the measurement.
3. When the wire is aligned, fix it with the 4. Cut the wire behind the second clamping
second clamping plate (same procedure plate, when both clamping plates have
as with the screws of the first clamping been tightly secured on the taught wire.
plate), with the tension still applied. Cut the wire very close to the clamping.
When inserting the guide pins, make sure that the tip of the clamping
bush or the ceramic tube is centered through the hole provided.
Never tighten the M5 screws before the alignment pins are inserted.
Otherwise the load cell could be harmed.
1. Insert the two supplied guide pins into 2. Correctly position the 1N load cell on the yoke of
the holder block. Then remove the four the tensile module without any further
M5 screws. mechanical adjustments of the assembly using
the guide pins in the front. The four M5 screws
are now firmly tightened, alternating between
the screws, in opposite pairs, in small increments.
3. Loosen the clamping nut on the fixed side 4. Push the specimen clamp holder through the
of the specimen clamp on the outside of hole provided in the yoke and aligned with the
the frame. guide pin. Then tightly fasten the nut with the
supplied tool.
Do not turn the support
blocks, as the specimen
restraints for the small
load cell are extremely
sensitive and the load cell
can be damaged.
1. Apply a piece of the red sealing wax to 2. Heat the sealing wax with a thin soldering
the ground opening of the ceramic tip until it becomes liquid. Use tweezers
capillary tube at both specimen clamps. to insert the specimen into the ground
opening through the liquid sealing wax.
3. carefully and gently tension the specimen
using tweezers, when the varnish has
been cooled / cured.
4. Align the specimen so that it emerges as
straight as possible from the orifice.
1. Clamp a rigid specimen and use the 2. When the maximum compression load is
software to set up the maximum load applied (depending on the installed force
permitted by the installed load cell. cell), loosen the lock nut on the fixed side,
using the 3 mm tool.
3. When the maximum tensile load is applied 4. Relax the applied load to zero (0 Newton).
(depending on the installed force cell),
loosen the lock nut on the force measuring
side, using the 3 mm tool.
5. Remove the rigid specimen from the 6. Turn the lock nut on the fixed side
module. counter-clockwise, using the supplied
3 mm tool and remove the specimen
clamp.
7. Loosen the four M5 screws holding the 8. Pull out the connector block.
force measuring cell and remove the
mounting aid.
In the next step, ensure that there is no crimping or twisting of the cable.
9. Turn the load cell unit counterclockwise. 10. Remove the force measuring cell, and then
Pay particular attention to the cables remove the specimen clamp.
ensuring that they do not become tangled
or trapped.
1. Insert the specimen clamping through the 2. Screw the load cell unit clockwise into the
yoke on the load cell side. thread of the specimen clamp, paying
particular attention to the cables, ensuring
Ensure that there is no that they do not become tangled or
crimping or twisting of trapped. The correct position is reached
the cable. when the force load cell unit is as close as
possible to the yoke, such that the surfaces
are parallel and the cable of the load cell is
correctly positioned.
3. Insert the guide pins and screw the 4. Guide the specimen clamp through the
mounting aid beneath the load cell. Next, yoke on the fixed side and carefully align it
tighten the four M5 screws of the load cell with the center pin. Screw the
unit. Loosely screw in the lock nut towards corresponding lock nut hand-tight towards
the specimen clamp. The connector block the specimen clamp.
can now be plugged in. The outer pin holes are made for 20° tilt
needed for the use in an EBSD setup.
Remove the guide pins.
5. Clamp the sturdy specimen and use the 6. With the maximum tension load applied,
software to set the maximum load tighten the lock nut on the force
permitted by the installed load cell. measuring side, using the supplied 3 mm
tool. Position the four M5 screws of the
load cell unit.
7. With the maximum compression load 8. Relax the applied force to zero (0 Newton)
applied, tighten the lock nut on the fixed and remove the rigid specimen.
side, using the supplied 3 mm tool.
4 Operating Manual
Overview
This chapter informs about:
- Clamping the specimen
- Use of Extensometer
- Switching On
- MDS software
c. Calculation examples:
3. Start the motor and move the yoke in the corresponding direction until the specimen can
be placed in the specimen clamps.
4. Place spacers of the calculated thickness inside the specimen holder.
5. Insert the lower bite plates.
a. The rough surface of the bite plate should face towards the specimen.
b. The arrows correspond to the direction in which the force will be applied during
the test (Tensile test = away from each other – Compression test = towards each
other).
6. Insert the specimen in the clamps on the lower bite plates.
7. Insert the upper bite plates inside the clamps on the specimen.
a. Pay particular attention to the arrow direction.
b. The rough surface should face towards the specimen.
8. Place a spacer of the calculated thickness inside each clamp on the upper bite plate.
9. Place another 1.0 mm spacer thickness inside each clamp. The spacer can be selected
thinner, if the specimen does not become thinner when gripped. The specimen must still
be clamped correctly, even if it yields due to the stress during the experiment. You can
also put one single spacer with the sum of both thicknesses.
10. Place the clamping blocks on both sides.
A small space must be visible between the lowest edge of the clamping
blocks and the corresponding edge of the clamps. This is an indication
for the correct selection of spacers. Place repeat the spacer calculation if
there is no space visible.
11. Use the controller or software to hold the load value constant. This saves the specimen
from stretching or compression caused by forces applied by clamping actions.
12. On the fixed side, tighten the two M5 screws of the clamping block to between 5 and
6 Nm.
1. Drive the bending points as far away as the specimen just fits in between them.
In the case of a specimen fracture, the user has to ensure that the
mechanics of the housing instrument (mciroscope) are not affected. In
case of disregarding the housing instrument could become damged.
4.4 Switching On
1. Ensure that all cables are connected. The module must be connected to the controller.
2. Switch on the controller.
Further details about the controller and its operation can be found in the separate controller
manual.
5 Further Information
Overview
This chapter informs about:
- Troubleshooting
- Temporary decommissioning
- Decommissioning
- Disposal
- Technical data
- Contact information
5.1 Troubleshooting
In this paragraph you find solutions to common errors. Please get in touch with Kammrath &
Weiss, if your issue is not listed in the table.
Cause
No. Trouble description Solution
Elongation gauge displays a sudden Plunger in the elongation gauge is out of its measuring range
1 change of sign or counts backwards Move the plunger back to the middle position (red ring)
Insufficient lubrication
2 Module does not reach 10 kN, anymore. Apply the supplied lubricant to the gears.
No voltage
Check mains connector and cable
4 Controller does not switch on.
Blown fuse
Disconnect mains connector and check fuse.
Controller switches on, but motor does Connector at rear of motor disconnected
5 not move. Plug in the connector.
Cable disconnected
Check cable connections.
7 No stable elongation or load display
Wrong configuration
Check software configuration
Table 2: Troubleshooting
Year of Manufacture
Outside Dimensions approx.. 220mm long, 120mm wide and 60mm tall
at maximum travel range
Available load cells 1, 10, 50, 100, 200, 500, 1000, 5000, 10000 N
Maximum Travel Range 40 mm for a specimen length about 5 mm between the clamps
0.1 to 20µm/s
Available speed 0.5 to 50 µm/s
Optional 2 to 140 µm/s (only for 1kN)
Max. Cycle Frequency < 0.4 Hz with standard motor, elongation about 7 µm
Motor Power 20 W
Screw M10 x 20 mm
Tungsten Wire 30 µm 50 µm
Preload 40 g 100 g
5.2.3 Extensometers
Name Extensometer
Type Number EXA 10-0.25 EXA 10-0.5 EXA 10-1 EXA 10-2
Gauge Length 10 mm
Weight 5 5 6 7
Length 20 mm
Width 10 mm
Height 15 mm 15 mm 21 mm 29 mm
1
Table 6: Technical Data of the extensometer
5.4 Decommissioning
When the instrument is placed out of operation permanently it is advisable to exchange the
Kammrath & Weiss flange for the original one.
5.5 Disposal
According to the European Guidelines 2002/96/EG, old electronic
devices must not be discarded to domestic waste. They have to be
collected separately and disposed appropriate.
Kammrath & Weiss takes back all electronic and mechanical
devices and takes care of the environmentally correct disposal.
E-mail [email protected]
Internet http://www.kammrath-weiss.com
EC Declaration of Conformity
The manufacturer:
Kammrath & Weiss GmbH
In addition, the agreement with the relevant conditions of the following EC-directives is confirmed:
EC Directive 2011/65/EU (RoHS)
Low Voltage Directive 2014/35/EU
EMC directive 2014/30/EU
The technical documentation is compiled in accordance with Annex VII A of the directive. The manufacturer
undertakes to transmit, in response to a reasoned request by the appropriate national authorities, relevant
information on the machinery identified above in paper copy. The technical documentation for the
machinery is available from the product management of the establishment, and bearing the same address
as the manufacturer.
This declaration is no longer valid once the product was modified without written consent of the
manufacturer.
The manufacturer hereby declares that all products of Kammrath and Weiss GmbH are in
accordance with the requirements of the EC directive 2006/95/EC. We are in close contact with
our suppliers, to ensure the RoHS conformity of our products in future.
This declaration is no longer valid, once the product was modified without written consent of the
manufacturer.
Notes:
Notes