CCD Digital Radiographic Detectors
CCD Digital Radiographic Detectors
University of Alabama at
Birmingham
Outline
• Introduction
• CCD Fundamentals
• Characteristics related to imaging
• Implementations in Radiology
• Advantages and Disadvantages
• CMOS
• Conclusion
Introduction
• Each cell has a given well depth for the storage of the charge
• If the charge exceeds the well depth, electrons will spill into
adjoining cells
• Design chip to accommodate expected well depth
• In some applications overflow drains are incorporated into the
chip
Vertical Overflow Drain
• Efficiency of transmission
– ~ 0.7 to 0.8 for CsI spectrum (550 nm)
• Focal length
• Demagnification factor
• Overall collection efficiency
Collection efficiency for lenses
TL
ηL =
1+ 4 • f # • (1+ m)
2 2
• Geometric distortion
• Vignetting
• Veiling glare (light scatter)
ηTFO =
1
2
•
(n 2
2 −n 3)
2 1/ 2
• TF • (1− LR ) • FC
m n1
η = 15%
Quantum efficiency of CCD
• Absorption of light
– Light needs to pass through polysilicon layers but not so
deep that cannot be captured in potential well
– Polysilicon layer needs to be relatively transparent
• Spectrum sensitivity
– Can modify this with overlaying materials and doping to
optimize to a given wavelength
• Light collection
– microlens
Example Kodak Blue Plus CCD
• Digital Fluoroscopy
• Stereo breast biopsy
• Mammography
• General radiography
Digital Fluoroscopy
• Typical FOV is 5 x 5 cm
• Matches with CCD well
• Fischer and Lorad use 2.5 x 2.5 cm cameras
– Fischer uses tapered fiber optics
– Lorad uses lens
• Noise is a major issue
– Use 27 to 28 kVp to increase photon flux reaching CCD
– 1024 x 1024 matrix; pixel size 50 µm -- affects noise
– CsI and special scintillators used to increase light output
Stereotactic Biopsy
• Imaging characteristics
– DQE is low, unpublished values
– New cameras and scintillators have been implemented
– 512 matrix can help on contrast and noise
– but main reason for doing x-ray biopsy is to find
calcifications where resolution and noise is important
Digital Mammography
• single CCD
• Use a large high quality lens
• Have achieved high DQE, through many design features
Specialized -- Statscan (Lodox)