Ultrasound Service Manual DC7 e DC7T
Ultrasound Service Manual DC7 e DC7T
Ultrasound Service Manual DC7 e DC7T
Service Manual
Table of Content
Table of Content ..................................................................................................................... i
Revision History.................................................................................................................................. I
Intellectual Property Statement ......................................................................................................... II
Responsibility on the Manufacturer Party.......................................................................................... II
Warranty ........................................................................................................................................... III
Safety Precautions ............................................................................................................................V
i
3.2.7 CW Board ................................................................................................................... 3-13
3.2.8 I/O Backplane .............................................................................................................3-14
3.2.9 ECG&PCG Board .......................................................................................................3-17
3.2.10 4D&TEE Drive Board.................................................................................................. 3-18
3.2.11 Control Panels and Keyboard.....................................................................................3-19
3.2.12 Power Output Control Board.......................................................................................3-21
3.2.13 Touch Screen .............................................................................................................. 3-22
3.2.14 LCD Monitor................................................................................................................ 3-23
3.3 Power Supply Principle ......................................................................................................3-24
3.3.1 Power Supply System................................................................................................. 3-24
3.3.2 Power Boards Principle .............................................................................................. 3-27
3.4 System Software Principle..................................................................................................3-31
3.4.1 Structure......................................................................................................................3-31
3.4.2 Principle ...................................................................................................................... 3-32
ii
4.3.18 Disassembling ECG Assembly ................................................................................... 4-51
4.3.19 Disassembling Probe Board & Probe Control Board.................................................. 4-56
4.3.20 Replacing Air Inlet Screen ..........................................................................................4-64
4.3.21 Disassembling Isolation Transformer..........................................................................4-65
4.3.22 Disassembling Power Fan Assembly..........................................................................4-66
4.3.23 Replacing Casters....................................................................................................... 4-69
4.4 Installing Peripherals ..........................................................................................................4-70
4.4.1 Installing B/W Video Printer ........................................................................................ 4-70
4.4.2 Installing Color Video Printer ......................................................................................4-72
4.4.3 Installing VCR ............................................................................................................. 4-76
iii
7.2 Software Maintenance..........................................................................................................7-3
7.2.1 Setup Data Backup and Restore .................................................................................. 7-3
7.2.2 Patient Data Backup and Restore ................................................................................ 7-4
7.2.3 Software Upgrade .........................................................................................................7-5
7.2.4 System Recovery.......................................................................................................... 7-6
7.2.5 Install/Uninstall the Optional Devices Software ............................................................7-8
7.2.6 Monitor Test ..................................................................................................................7-9
7.2.7 Selftest ........................................................................................................................7-10
7.2.8 Maintenance and Processing of dump File ................................................................ 7-16
iv
Revision History
Mindray may revise this publication from time to time without written notice.
10.0 2014.2.27 Add the OrderNumber and picture of new 17-inch monitor support
arm assembly
I
©2009-2016 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights Reserved.
For this Operator’s Manual, the issue date is 2016-06.
, , , , , BeneView,
WATO, BeneHeart, are the trademarks, registered or otherwise, of Mindray in
China and other countries. All other trademarks that appear in this manual are used
only for informational or editorial purposes. They are the property of their respective
owners.
All information contained in this manual is believed to be correct. Mindray shall not be
liable for errors contained herein or for incidental or consequential damages in
connection with the furnishing, performance, or use of this manual.
Mindray is responsible for the effects on safety, reliability and performance of this
product, only if:
z all installation operations, expansions, changes, modifications and repairs of this
product are conducted by Mindray authorized personnel;
z the electrical installation of the relevant room complies with the applicable
national and local requirements; and
z the product is used in accordance with the instructions for use.
II
Upon request, Mindray may provide, with compensation, necessary circuit diagrams,
calibration illustration list and other information to help qualified technician to maintain
and repair some parts, which Mindray may define as user serviceable.
Note
Warning
It is important for the hospital or organization that employs this equipment to carry out a
reasonable service/maintenance plan. Neglect of this may result in machine breakdown
or injury of human health.
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR ANY PARTICULAR PURPOSE.
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation
or other charges or liability for direct, indirect or consequential damages or delay
resulting from the improper use or application of the product or the use of parts or
accessories not approved by Mindray or repairs by people other than Mindray
authorized personnel.
III
z Personnel unauthorized by Mindray repairs or modifies the instrument.
Return Policy
Return Procedure
In the event that it becomes necessary to return this product or part of this product to
Mindray, the following procedure should be followed:
1. Obtain return authorization: Contact the Mindray Service Department and obtain a
Customer Service Authorization (Mindray) number. The Mindray number must
appear on the outside of the shipping container. Returned shipments will not be
accepted if the Mindray number is not clearly visible. Please provide the model
number, serial number, and a brief description of the reason for return.
2. Freight policy: The customer is responsible for freight charges when this product is
shipped to Mindray for service (this includes customs charges).
3. Return address: Please send the part(s) or equipment to the address offered by
Customer Service department
Company Contact
Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Address: Mindray Building, Keji 12th Road South, Hi-tech Industrial Park,
Nanshan, ShenZhen 518057, P. R. China
Tel: +86 755 26582479 26582888
Fax: +86 755 26582934 26582500
IV
Safety Precautions
V
3. Safety Precautions
Please observe the following precautions to ensure patient’s and operator’s safety when
using this system.
VI
8. DO NOT use an aftermarket transducer other than those
specified by Mindray. The transducers may damage the
system, causing a profound failure, e.g. a fire in the worst
case.
9. DO NOT subject the transducers to knocks or drops. Use
of a defective transducer may cause an electric shock.
10. DO NOT open the covers and front panel of the system.
Short circuit or electric shock may result when the system
hardware is exposed and powered on.
11. DO NOT use this system when any digital device such as
a high-frequency electrotome, high-frequency therapeutic
device or defibrillator is applied already. Otherwise, there
is a risk of electric shock to the patient.
12. Only use the ECG leads and PCG transducer provided
with the physiology module; otherwise, electric shock
may be resulted.
13. When moving the system, you should hold the handle,
otherwise, damage may be resulted by abnormal force.
DO NOT push the system from the left/right side,
otherwise, it may be toppled over.
14. The auxiliary power output outlet in the system is used to
supply power for the recommended peripheral devices.
DO NOT connect other devices to the outlet, otherwise,
the rated output power may be exceeded and failure may
be resulted. Maximum output power of the outlet is
330VA.
15. Accessory equipment connected to the analog and digital
interfaces must comply with the relevant IEC standards
(e.g., IEC 60950 information technology equipment safety
standard and IEC 60601-1 medical equipment standard).
Furthermore, all configurations must comply with the
standard IEC 60601-1-1. It is the responsibility of the
person, who connects additional equipment to the signal
input or output ports and configures a medical system, to
verify that the system complies with the requirements of
IEC 60601-1-1. If you have any questions regarding these
requirements, consult your sales representative.
16. Prolonged and repeated use of keyboards may result in
hand or arm nerve disorders for some individuals.
Observe the local safety or health regulations concerning
the use of keyboards.
VII
CAUTION: 1. Precautions concerning clinical examination techniques:
This system must be used only by qualified medical
professionals.
This operator’s manual does not describe clinical
examination techniques. The clinician should select the
proper examination techniques based on specialized
training and clinical experience.
2. Malfunctions due to radio wave:
If a radio wave emitting device is used in the proximity of
this system, it may interfere with operations. DO NOT use
or take any devices transmitting RF signals (such as
cellular phones, transceivers and radio controlled
products) in the room placing the system.
If a person brings a device that generates radio waves
near the system, ask him / her to immediately turn OFF the
device.
3. Precautions concerning moving the system:
Please install the system on a flat plane with the four
casters locked. Otherwise, damage may result by
accidental moving.
DO NOT move the system laterally, which may result in
damage in case of toppling.
Move the system slowly on the slope by two people,
otherwise, damage may result in case of unexpected
sliding.
DO NOT sit on the system, which may result individual
falling in case of system moving.
Object placed on the monitor may fall and injure an
individual.
Fasten and fully secure any peripheral device before
moving the system. A loose peripheral device may fall and
injure an individual.
When move the system on the steps, please take care to
prevent the system from toppling.
4. If the circuit protector is tripped, it indicates that the
system or a peripheral device was improperly shut down
and the system is unstable. You can’t repair the system
under this circumstance and must call the Mindray
Customer Service Department or sales representative.
5. There is no risk of high-temperature burns during normal
ultrasound examinations. It is possible for the surface
temperature of the transducer to exceed the body
temperature of a patient due to environmental temperature
and exam type combinations. DO NOT apply the
transducer to the same region on the patient for a long
time. Apply the transducer only for a period of time
required for the purpose of diagnosis.
VIII
6. DO NOT use the system to examine a fetus for a long
period of time.
7. The system and its accessories are not disinfected or
sterilized prior to delivery. The operator is responsible for
the cleaning and disinfection of transducers and
sterilization of biopsy brackets according to the manuals,
prior to the use.
All items must be thoroughly processed to completely
remove harmful residual chemicals, which will not only
harmful to the human body, but also damage the
accessory.
8. It is necessary to press <End Exam> to end the current
scan that is in progress and clear the current Patient
Information field. Otherwise, new patient data may be
combined with the previous patient data.
9. DO NOT connect or disconnect the system’s power cord or
its accessories (e.g., a printer or a recorder) without
turning OFF the system power first. This may damage the
system and its accessories or cause electric shock.
10. If the system is powered off improperly during operation, it
may result in data damage of the system’s hard disk or
system failure.
11. DO NOT use a USB memory device (e.g., a USB flash drive,
removable hard disk) which has unsafe data. Otherwise,
system damage may result.
12. It is recommended to only use the video devices specified
in this manual.
13. DO NOT use gel, disinfectant, transducers, transducer
sheath or needle-guided brackets that are not compatible
with the system.
14. The applied contrast agents should be compliant with the
relevant local regulations.
15. Read the Acoustic Output Principle in the operation
manual carefully before operate this system on clinical
examination.
16. The cover contains natural rubber that can cause allergic
reactions in some individuals.
17. Please use the ultrasound gel compliant with the relevant
local regulations.
NOTE: 1. DO NOT use the system in the vicinity of strong electromagnetic field (such
as a transformer), which may affect the performance of the system.
2. DO NOT use the system in the vicinity of high-frequency radiation source,
which may affect the performance of the system or even lead to failure.
3. When using or placing the system, keep the system horizontal to avoid
imbalance.
IX
4. To avoid damaging the system, DO NOT use it in following environment:
z Locations exposed to direct sunlight;
z Locations subject to sudden changes in environmental temperature;
z Dusty locations;
z Locations subject to vibration;
z Locations near heat generators;
z Locations with high humidity.
5. Turn ON the system only after the power has been turned OFF for more
than 20 seconds. If the system is turned ON immediately after being turned
OFF, the system may not be rebooted properly and could malfunction.
6. Press <Freeze> key to freeze an image or turn off the power of the system
before connecting or disconnecting a transducer. This can damage the
system and/or transducer.
7. Remove the ultrasound gel from the face of the transducer when the
examination is completed. Water in the gel may enter the acoustic lens and
adversely affect the performance and safety of the transducer.
8. You should properly back up the system to a secure external storage
media, including system configuration, settings and patient data. Data
stored to the system’s hard drive may be lost due to system failure,
improper operation or accident.
9. DO NOT apply external force to the control panel. Otherwise, the system
may be damaged.
10. If the system is used in a small room, the room temperature may rise.
Please provide proper ventilation and free air exchange.
11. To dispose of the system or any part, contact Mindray Customer Service
Department or sales representative. Mindray is not responsible for any
system content or accessories that have been discarded improperly.
12. Electrical and mechanical performance may be degraded due to long usage
(such as current leakage or distortion and abrasion), the image sensitivity
and precision may become worse too. To ensure optimal system
operations, it is recommended that you maintain the system under a
Mindray service agreement.
13. The replaceable fuse is inside the chassis. Refer replacing job to Mindray
service engineers or engineers authorized by Mindray only.
14. DO NOT turn OFF the power supply of the system during printing, file
storage or invoking other system operations. An interrupted process may
not be completed, and can become lost or corrupted.
15. The iScape feature constructs a single extended image from a series of
individual image frames. The quality of the final image is user-dependent
and requires skill to efficiently apply the feature and technique. Exercise
caution when measurements are performed from an iScape image.
16. Ensure that the current exam date and time are the same as the system
date and time.
X
4. Warning Labels
The warning labels are attached to this system in order to call your attention to potential
hazards. The symbol on the warning labels indicates safety precautions.
The warning labels use the same signal words as those used in the operator’s manual.
Read operator’s manual carefully before using the system.
The name, pattern and meaning of each warning label are described as follows:
XI
1 System Overview
1-1
2 System Structure
2-1
System Structure
2-2
System Structure
<13> Color printer table Used for placing color video printer.
<14> Handle Used for pushing and moving the system.
<15> I/O Panel Input and output interface panel.
<16> Power Supply Electric interface panel.
Panel
<17> USB port Used for connecting USB devices.
<18> MIC interface Used for connecting a microphone.
<19> VCR placing room Used for placing VCR and DVD.
<20> Power switch Used for powering on/ off.
2-3
System Structure
As shown in figure above, the lever is located at the lower left of the monitor back. Only
when the lever is placed in the rightmost position can it be used for up/down adjustment
of the monitor. The monitor can be tilted 20 degrees forward or backward. See the figure
below.
2-4
System Structure
When transporting or moving the system, push the lever to the left side to place the
monitor in the horizontal position, as shown in figure below.
2-5
System Structure
Locked: The support arm will be locked without moving any further when it is in the
horizontal position.
Unlocked: The support arm of monitor is fully adjustable.
2
Ethernet port (port <3> is reserved for future use)
3
4 System reset.
2-6
System Structure
2-7
System Structure
2-8
System Structure
Controller
Type No. Name Function
Type
Start an exam of a new patient
Functional
1 End Exam while ending the last patient’s
Button
exam.
Functional Open/ exit patient information
2 Patient
Button screen.
Functional Switches Transducer and Exam
3 Probe
Button Type.
Patient
Functional Enter/ exit the image review.
operation
4 Review Button
Review the storage image of
current (or last) patient.
Functional
5 Report Open/ exit the exam reports.
Button
Enter/ exit the patient
Functional
6 iStation information management
Button
system.
2-9
System Structure
Controller
Type No. Name Function
Type
Press: enter the M mode.
16 /
Press to enter the Dual mode
from another mode;
Functional
17 Dual Or, press to switch between the
Button
two display windows in the Dual
Window mode.
Switch active window during
Functional multi-imaging mode or switch
18 Update
Button active/ fixed end of the caliper
during measurement.
Adjust depth in real-time
19 Depth Knob
imaging.
Rotate: enter pan zoom mode;
Parameter 20 Zoom Pressable knob
Press: enter spot zoom mode.
Adjusting
21 TGC Slide bar Adjust the depth gain.
Functional
22 iTouch Optimize the image.
Button
Functional
Freeze 23 Freeze Freeze/ unfreeze the image.
Button
Functional Enter/ exit the application
Measurement 24 Measure
Button measurement mode.
Comments Functional Enter/ exit the textual comment
25 Comment
(Annotations) Button status.
2-10
System Structure
Controller
Type No. Name Function
Type
Functional Enter/ exit the arrow comment
26 Arrow
Button status.
Short-press: first clear the
selected element, and then clear
all elements on screen
Functional backward.
27 Clear
Button Long-press: first clear all
elements related to current
mode, and then clear all rest
elements.
2-11
System Structure
2.6.2 Keyboard
2-12
System Structure
NOTE: “/” means the key are undefined or have no silk print. For the undefined keys, you
can customize them. For key configurations, please refer to the Operator's Manual [Basic
Volume].
Switch
Exit
Mode
B C PW Mode
Selection
NEXT
Duplex/Triplex HPRF Audio
Sweep Speed
Parameter
Med Adjusting
Doppler Freq Quick Angle
_ 2.5M + 60
2-13
System Structure
Switch
iStation Mode Exit Title Area
Activate
New Exam
Exam Main
Functions
Review Info Report Send To Area
Soft Keyboard
Widget
Area
Scroll
Knob-adjusting
Area
(1) Title area: to display the title that corresponding to the monitor-displayed dialog
box title or some functional buttons (such as the soft keyboard and exit).
(2) Main function area: click the button to go to the corresponding function (including
the page switching and main function buttons).
(3) Widget Area: widgets corresponding to operations such as selection or input on
main screen.
(4) Knob-adjusting area: the knob-controlled parameters.
2-14
System Structure
Dialog box
It refers to those interface used for selecting the system status (e.g. to confirm or to
cancel a command). The same interface is displayed in the monitor at the same time.
As shown in figure below.
Delete
Yes No
Other interface
There are interfaces provided for some other applications, such as touch screen
pages displayed in comment or body mark modes.
2-15
3 System Principle
C A B
b m e B, M
T W o p a
ra a m /I
POUT n b rd ifl C, CM O
s o ie ECHO fo
P m a r rm c
P ro is rd PW, CW D Hard-wiredo
n
r
o b s e n
P b e io Front-side FPGA r ig e
ro e in n configuration bus b
o ita Hard-wiredc /I
O
c a tin
b o te b rd lb b
e n rf Scan control C o g o
b o a o b a
tr a signal n rd Scan control signal a rd
o
a
o
lb
c
e n rd Flexible- o a
e
rd
o b c
to wired rd
a o
rd a Data bus r Front-side control bus
rd
Ethernet
Power cable
4
D PHV 1 A+5V
/T D+5V
b E A+5V A-5V
o PHV1 D 3.3V
a E
Power/control/state rd d A-100V PHV2 D+3.3V A+5V D+12V
ri A+100V A+12V D+5V
signal v
e PHV2 A-5V
D+3.3V
A-5V
A+5V D+5V A+5V
D+5V A-5V HDD Touch
D+3.3V A+12V screen
D+12V D+3.3V
A+12V D+5V
D-12V
System motherboard
Fan connecting
LED board
board Power output control board
Fan
Power motherboard
Isolation Connecting
PFC board +5V board -5V board PHV board
transformer board
External
power external Auxiliary
input auxiliary output adapter
board
output
peripheral output
3-1
System Principle
Probe socket D
Boards connecting
socket 3
socket 4
Multi-channel relays array
Probe
Probe
array
array
signal
signal
Probe array signal
Probe socket C
Relay control signal
Probe socket B
Boards connecting
socket 1
socket 2
Probe socket A
Relay control signal
3-2
System Principle
selection
signal
connecting
control board signal
signal
socket
General
probe ID
signal
CPLD
Probe connecting signal
The probe board provides probe switching and probe ID selection. The probe switching
signal controls the relays to switch between probes, and the ID reading circuit is
independent of the probe switching circuit and can read out the IDs of the probes.
The probe socket D is used to connect the special probe, differs from other sockets by
using the 260 PIN socket, thus the probe misconnection will not happen. PINs used for
different probes in the socket are independent from each other, thus the socket damage
caused by recognizing the wrong probe will not happen. But the SPI interface reserved
for TEE probe also can be used for 192-array probe, an incorrect control byte could be
sent if the probe is wrongly recognized.
Working sequence of power on and probe connection/ disconnection:
1. The probe-online signal level will switch whenever the A, B, C or D socket accepts a
probe. The level change is detected by CPLD on probe control board, the information
then will sent to the software.
2. The software performs power on the VDD, read out IDs and other operations when
appropriate after receiving the probe-online signal.
3. Switches on the power supplies of the D probe after reorganization, e.g. VCC of 4D&
TEE, ± 100V of 192-array, then starts other operations.
3-3
System Principle
Probe socket D
Probe socket C
Relay control signal
Probe socket B
Probe socket A
Relay control signal
Comparing to probe board (single 4D socket), Probe socket (double 4D sockets) has
increased switch control of special probe signal, but principle of probe control board is
also the same as figure 3-3.
3-4
System Principle
N N
N channel transmission
The transmission board is designed for generation of high-voltage pulse and transmission
sequence. The transmission circuit consists of two parts: transmission sequence and
transmission drive. The transmission sequence circuit generates low-voltage pulse, which
passes the transmission drive circuit and is converted to high-voltage pulse. The
transmission board is provided with four programmable transmission high voltages by the
power supply, so that the board can generate different transmission waveforms
depending on the needs of the system.
The amplifier board is designed for voltage-controlled gain amplification of small signal echo.
The amplifier board consists of high-voltage isolation circuit, channel analog multiplexer,
primary voltage-controlled gain amplification circuit and secondary amplification circuit. The
control signal of the channel select logic is derived from the beamformer board, and this
signal controls the on-off of the reception channel. The primary gain amplification circuit
3-5
System Principle
DDSP
Doppler signal
Link
DDSP port
configura CDSP
tion port Switch
Flow
DDSP CDSP
Doppler signal
configura peripherial
signal tion port
CDSP
Switch control Switch
Doppler
audio signal
SSRA M2
M 2
Reserved
保留
The beamformer board consists of beamformer module, signal processing FPGA module,
and signal processing DSP module. The beamformer board is designed for AD sampling,
beamforming, radio frequency signal processing, Doppler signal processing, flow signal
processing, video signal processing, and generation of scan control signal, etc.
The channel difference analog signal, transmitted from the system motherboard, passes
the low-pass filter to avoid aliasing of subsequent AD sampling. The cut-off frequency is
15 MHz.
After the signal is filtered and passes 8 high-speed AD for sampling, the sampled data is
sent to FPGA U46 and U65 for beamforming.
The data, after beamforming, is sent to FPGA U4 for B/W signal processing. At the same
time the data is sent to FPGA U59 for color flow signal processing and partial Doppler
signal processing. Subsequently part of data is sent to FPGAU83 for remaining color flow
signal processing, and the other part of the data is sent to DSP U78 for remaining
Doppler signal processing. Then the B/W signal, color flow signal, Doppler signal pass
the system motherboard and are sent to the digital board for the subsequent processing.
3-6
System Principle
USB Video
Capture MU Decoder
PCG G Ethernet X
CPLD
CF Card
ECG Uart<->USB
Power
USB Manage
Audio
PC DVI
LPC
Module Transmitte DVI
SATA Analog VGA r
TS LCD
Analog VGA
X1 PCI Express VGA DAC LCD
SDVO to
PCIe PHY LVDS
33MHz,32bit PCI
Seria to Parallel
Cable Equalizer SDI IN
Converter
BDSP NIOS2 BCDSC
BCDSC LVDS
DDR SSRAM0
SSRAM0 DDR0
CDSP SDI
Cine
Cine
BCDSC
BCDSC FPGA
DDSP DDR
DDR
SSRAM1
SSRAM1 FPGA Config
Cine DSC Back
FPGA FPGA MDDSC FPGA
Parallel to Seria
DDR0 Video DDR1 Cable Driver SDI Out
Converter
(Interlacing)
Front Video
Encoder
End
SSRAM1
Parallel and Serial Port
FPGA Config
The upload module provides real-time uploading of the ultrasound data to the PC module.
The PC module is responsible for part of the image processing, measurement and
display. The display module provides processing and display of the video and VGA
signal.
Functions of sub-modules are described as follows:
The system cache the data according to different modes (B/C/D/M) in two storage spaces
in an alternative way, that is, one space is used for uploading data to PC while another
used for receiving real-time data. There are frame mode and line mode divided by
imaging modes.
3-7
System Principle
Frame mode (B and C mode): cache two data frame (the actual storage space varies
by line density and FOV). One frame is the data obtained from scanning, waiting for
uploading to PC; another frame is used to receiving the real-time scanning data. Two
frames work alternatively.
Line mode (D and M mode): cache several lines (configured by system software),
allocate two storage space for storage and uploading, in which the D mode includes
spectrum image and audio signal.
When in multi-imaging mode (BD/BC/BCD/BM/CM/CCM etc.):
z BD mode: upload B mode data after caching 1 frame, upload D mode data after
caching several lines.
z BC mode: upload B and C mode data together after caching the same frame of
B and C data.
z BCD mode: B/C data uploading is the same with BC mode, D data is uploaded
after caching several lines.
z BM mode: upload B mode data after caching 1 frame, upload M mode data after
caching several lines.
z CM: upload M/CM data after caching several lines.
z CCM: B/C data uploading is the same with BC mode, M/CM data uploading is
the same with CM mode.
z ECG and PCG uploading is similar with that of D mode.
To avoid two line data uploading requests generating within a B or C mode uploading
period, the system must be able to tell the line data uploading period from the current
max frame data uploading period. Thus the interval between two line data uploading
requests should be longer than that of the frame data uploading period since the line data
uploaded is in a small amount.
3-8
System Principle
Video_Input_MUX0
PC S-video
(standby)
Data wire
RGB Video
User S-Video
Decoder 245
VCR S-Video (TVP5146) Clock
Digital board connector
Video(CVBS)
USB to PC Module
EM2860
Data wire
245
Clock
Video_Input_MUX1
Line_L
EMP
Line_R 202
Video inputs are S-Video, DVBS and RGB. VCR S-Video is used for connecting VCR
recorder; user RGB input, Video (CVBS) and User S-Video are external user input video
interfaces; PC S-Video are standby.
3-9
System Principle
3-10
System Principle
Beamformer board
Amplifier board
Probe control board
Transmission board
CW board
Probe connecting board
3-11
System Principle
TO Fan SDI In
F12VFAN2FAN3 SDI Out
F12V
TO Fan
3-12
System Principle
Socket
Type Connection Purpose
number
Power signal Hard-wired 4 For all power channels
3.2.7 CW Board
The CW board diagram is shown below, which consists of four portions: transmission,
reception, power and control.
Waveform
Tissue generator Signal
process
Quadr
Reception RF Reception Audio A/D
ature Filter
transducer amp beamformer amp converter
demo
3.2.7.1 Transmission
The transmission signal flow of CW board is shown in figure below.
FPGA- Transmit
Passive Probe
generating signal
filter drive
transmit signal drive
3.2.7.2 Reception
The reception signal flow is shown in figure below.
The high voltage isolation circuit is used to protect the reception circuit. Although a relay
is placed in the front-end of the reception circuit to isolate from the probe, if fault occurs
for the relay control, the high voltage may be applied to the front-end of the reception
circuit. Therefore the high voltage isolation is necessary.
3-13
System Principle
The band pass filter is used to filter extremely low frequency component and sum
frequency component in the signal demodulated. It consists of two-level active high-pass
filter and four-level active low-pass filter.
3-14
System Principle
HDD
USB3 to DVD RW
TO Monitor (VGA,
AUDIO, PW OUT)
Keyboard
DVD
NOTE: other modules are boxed with broken lines
问字
3-15
System Principle
3-16
System Principle
Supports external ECG input and complies with the patient monitoring product interface
of Mindray.
z Supports ECG realized by three-cable leads I, II and III in Patient In mode; does
not support ECG auto reorganization; supports ECG lead drop check.
z Supports DC IN.
z Supports PCG.
z Supports 1 ECG channel, 1 PCG channel; does not support channel switch; the
doctor can switch channel by moving the leads.
z Supports continuous gain adjustment controlled by external gain knob.
z Supports continuous ECG and PCG wave-position adjustment controlled by
external wave-position knob.
Right leg
Lead 3
drive
3-18
System Principle
Interface of
probe connecting
Probe board
Digital board
System
Digital board
interface
motherboard
interface
Cable
3D/4D/TEE
Module interface
TEE
Progra
Signal
ADC mmable
amplifie
power
r
4D&TEE
control board 4D&TEE drive board
Power
Signal
DAC amplifi
shaper
er
Figure 3-17 Connections between 4D&TEE Drive Board and Other Boards
3-19
System Principle
3-20
System Principle
The CPU circuit consists of one FPGA and the peripheral circuit, and it is the core of
the keyboard, to process various types of information.
The DC/DC circuit is used to provide different levels required by the keyboard.
The key control circuit refers to the key processing hardware.
The trackball control circuit and encoder control circuit are used to convert the
external mechanical actions to signals which are recognizable by FPGA, and the
circuits consist of trackball, encoder, pulse shaper circuit and hardware processing
circuit.
The minor panel encoder board consists of 5 encoders with buttons that
corresponding to 5 key backlights and touch screen signal channel and I2C interface
of the display (including 8.4 inch touch screen and 17 inch monitor) contrast,
brightness, and position adjustment.
The encoder signal and key signal are generated by the encoder and sent to main
control panel to process. The key backlight is dual-color LED driven by the main
control panel. The touch screen and monitor signals are controlled by main control
panel. The minor encoder board provides connection socket for touch screen and
monitor for signal connection.
The LED control circuit is used to indicate statuses of LEDs.
The STC A/D sampling circuit consists of sliding potentiometer, amplifier and AD
sampling circuit, and it is used to convert the analog signal from the sliding
potentiometer to the digital signal which can be accepted by FPGA.
3-21
System Principle
CON1
CON2
SPI
ADT7516 VREF Max 6066
CON3
Power
CON4
output
module SPI
ADT7516_VREF
REF3040
SPI
CONNECT 232 MAX3243
TTL download
OR interface
3-22
System Principle
Model of the 8.4 inch LCD display is specified, G084SN03 by AU; the boards, cards and
wires are customized according to structure and interface specifications.
Model of the 17 inch LCD display is specified, LTM170E8-L01 by SAMSUNG; the boards,
cards and wires are customized according to structure and interface specifications.
3-23
System Principle
3-24
System Principle
programmable voltages are obtained through the BOOST converter and BUCK-BOOST
converter.
After passing the power motherboard, the outputs of the power system are connected to
the power output control board for filtering and monitoring. After the programmable
voltages are detected, they are outputted to the main unit for supplying power for relevant
hardware circuits. The outputs of the power system, after passing the fan adapter board,
are connected to the LED board. The LED board indicates the statuses of the outputs.
3-25
System Principle
isolation
隔离变压器
transformer
L LED
FL LED 指示板
indicator
N
board
FN
连接板
风扇
彩色视频
color video
Conne fan
printer
打印机 cting
Fan
board VPP 风扇连接板
connecting
board
auxiliary
辅助输出
output RY+
adapter
转接板
board
黑白视频
B/W video
FL
printer
打印机 FN
VPP
显示器
monitor RY+
PFC板
VCR
PFC SW
5B 5B
board SW LED
LED POFF
POFF ENA
ENA GND
GND SW
LED
CON(1~4)
VDD
POFF
DC+
DC-
PF
ENA
系统母板
电源母板
GND
电源输出监控板
+5V
+5V板
+5V
board +3.3V
+5V +5V
VDD
DC+
DC-
+12V
+3.3V +3.3V
-12V
-5V板
PFL
-5V -5V
+12V
PHV(1~4) PHV(1~4)
PHV板
+/-HV
PHV
board
PHV(1~4)
fan
风扇连接板
connecting
board
CON(1~4)
fan
风扇
3-26
System Principle
Isolation transformer
Voltage Voltage
selector selector
wire wire
L surge
Main unit
Breaker
Auxiliary
FG
Relay EMI filter Relay AC
output
Relay control signal (from PFC
board)
Connecting board
The mains power input passes the circuit breaker to enter the connecting board, reaches
an isolation transformer through a surge current suppression circuit, and then the
isolation transformer generates two AC outputs that are isolated with the mains power.
After passing one-level or multiple-level EMI filter, one output provides power for the PFC
board, and the other, through relay control, provides power for the peripheral device such
as monitor and printer. The high and low voltages of the mains power are selected by
tieline of the transformer. After the system is turned on, the PFC boards sends the relay
shut-off signal, and bypasses the thermistor suppressing surge current, and the AC
auxiliary output is switched on.
The AC input power passes the EMI filter, and then passes the rectifier bridge to reach
the BOOST APFC power circuit. In the APFC control circuit, the APFC function of the
input power is completed, and a stable +390V DC current is outputted to the +5V board
and -5V board. The APFC control circuit is powered by the STANDBY power supply. The
structure of the STANDBY is single-ended FLYBACK converter. The STANDBY not only
provides +18V power VCC for the APFC control circuit, but also outputs an isolated +5V
voltage for power switch control circuit and LED control circuit.
Besides OVP, OTP and sequential control circuit, the PFC board receives the fail
protection signal PF from the +5V board, the power-off signal POFF and power-off enable
signal ENA, thus realizing the safe turn-off of the entire power system.
3-27
System Principle
The core ICs of the PFC board, PFC controller and PWM controller, are respectively
UC3854 and 38HC43 made by TI.
DC+
Rectifier 390V out
2
EMI
bridge
filter 1
+5V board/
3
-5V board
Linear VDD
regulator
APFC
control
circuit Linear
regulator
VPP Linear
regulator LED
LED
control
circuit
PWM
control
circuit ENA
Power on/ SW
off logic
POFF
VOP/ OTP circuit
circuit
3-28
System Principle
3-29
System Principle
FORWARD
transformer magamp
rectifier -12V output
adjustm
+5V filter
390Vdc Input ent 1
power filter
board overcurre
circuit PFL signal
ocp/ovp signal nt &
over-
voltage
detection
Power motherboard
magamp -5V output
VDD rectifier
adjustm
filter
ent 2
PWM
control
OVP&O
feedback VP
OTP
Photocoupler
isolation
3-30
System Principle
+HV+
flyback +HV-
12V input -HV+
converter -HV-
boost PHV1
converter
boost PHV3
control converter
voltage
buck-boost PHV2
converter
buck-boost PHV4
converter
The principle of other boards is simple and all components on the boards are almost
passive, so it isn’t necessary to describe it.
3.4.1 Structure
The software system consists of Win32 applications and Win32 support programs
running on Windows XP operating system. Also includes control programmed embedded
peripheral devices.
The software function diagram is shown in figure below.
3-31
System Principle
3.4.2 Principle
DC-7 system software is a user-interactive device control program that is even driven,
and its function can be described in several views.
It’s a device control program of the ultrasound hardware system. The software
converts user operations or commands from the control panel to signals and setting
recognized by the hardware, thus the imaging effect will be changed due to
parameter changes.
Ultrasound echo data from the hardware systems is received by software, and post
processed and converted to digital image for computer processing and display.
It provides clinic diagnosis such as comment, measurement, physiological signal
display (ECG & PCG); provides a simple patient information management system for
exam data storage and management.
It supports commonly-used peripherals (USB flash drive, recorder, VCR etc.) and
special medical & ultrasonic peripherals (footswitch, DICOM interconnection etc.).
3-32
4 Structure and Assembly/
Disassembly
4-1
Structure and Assembly/ Disassembly
4-2
4.2 Field Replaceable Unit
Assembly/
NO. Classification Description OrderNumber Photo Model Compatibility
Disassembly
Refer to
1 DSP board 801-2112-00001-00 DC-7 Includes button battery,
4.3.22
4-3
Structure and Assembly/ Disassembly
Refer to
5 Amplifier board 801-2105-00002-00 DC-7
4.3.22
Stop production;
Supports 2P2\P4-2 probes
dosen't support P7-3\P12-4
CW upgrade Refer to
6 801-2112-00021-00 DC-7 the version of doppler software
module 4.3.22
should be 02.00.00 or above.
The board whose ID is 144
can't work in CW model for
P4-2 probe
4-4
Structure and Assembly/ Disassembly
Supports 2P2\P4-2\P7-3\P12-4
CW upgrade probes, the version of doppler Refer to
7 801-2112-00021-01 DC-7
moduleECN015 software should be 02.00.00 or 4.3.22
above
Refer to
8 Transmission board 801-2112-00062-00 DC-7
4.3.22
4-5
Structure and Assembly/ Disassembly
Transducer Refer to
11 801-2112-00015-00 DC-7
connecting board 4.3.27
System
12 801-2112-00016-00 DC-7
motherboard
ECG&PCG Refer to
13 115-001362-00 DC-7 CENot include ECG cable
assembly 4.3.27
Refer to
14 Physio Module 801-2112-00069-00 DC-7 FDAincludes ECG cable
4.3.27
4-6
Structure and Assembly/ Disassembly
4-7
Structure and Assembly/ Disassembly
4-8
Structure and Assembly/ Disassembly
Refer to
23 PFC board 801-2105-00011-00 DC-7
4.3.22
Refer to
24 +5V power board 801-2105-00012-00 DC-7
4.3.22
Power related
Refer to
25 -5V power board 801-2105-00013-00 DC-7
4.3.22
Refer to
26 PHV power board 801-2112-00052-00 DC-7
4.3.22
4-9
Structure and Assembly/ Disassembly
Power connecting
28 801-2105-00015-00 DC-7
board
4-10
Structure and Assembly/ Disassembly
Refer to
30 SWITHC 250V 13A 801-2105-00030-00
4.3.29
Control panel
related&touch
screen related
Stop production;
plese both replace main panel Refer to
33 Main panel PCBA 801-2112-00008-00 DC-7
silicon keypad and control 4.3.14
panel PCBA when used
4-11
Structure and Assembly/ Disassembly
Stop production;
Main panel silicon plese both replace main panel Refer to
35 801-2112-00030-00 DC-7
keypad silicon keypad and control 4.3.11
panel PCBA when used
4-12
Structure and Assembly/ Disassembly
PC Keyboard Refer to
40 801-2112-00055-00 DC-7
assembly 4.3.17
Refer to
41 2" Trackball 801-2112-00025-00 DC-7 Includs connecting wire
4.3.13
4-13
Structure and Assembly/ Disassembly
Monitor
Display assembly Refer to
45 related 115-023467-00 DC-7 17-inch monitor assembly
(maintenance 4.3.1
spare)
4-14
Structure and Assembly/ Disassembly
Refer to
19“ LCD assembly 801-2112-00058-00 DC-7 19-inch monitor assembly
4.3.1
Back cover of
Refer to
17-inch monitor 043-003951-01 DC-7 17-inch monitor assembly
4.3.1.1
(2119)
Front cover
assembly of the
Refer to
monitor 115-023468-00 DC-7 17-inch monitor assembly
4.3.1.1
(maintenance
spare)
4-15
Structure and Assembly/ Disassembly
Damping axis
assembly Refer to
115-023473-00 DC-7 17-inch monitor assembly
(maintenance 4.3.1.3
spare)
Speaker and
Refer to
connecting cable 020-000033-00 DC-7 17-inch monitor assembly
4.3.1.2
40 hm
4-16
Structure and Assembly/ Disassembly
Fan related
4-17
Structure and Assembly/ Disassembly
4-18
4.3 Disassembly and Replacement
NOTE: Make sure the casters are locked before disassembly, replacement or whenever
the machine is not intended to move.
4-19
Structure and Assembly/ Disassembly
3. Unscrew 1 M4X12 screw. Unplug monitor power supply cable, signal cable and
speaker cable;
M4X12 screw
Power
supply
cable
4. Unscrew six M4X12 screws, and remove the monitor assembly upwards.
4-20
Structure and Assembly/ Disassembly
Note: The metal part is inserted into the hole of the plastic part. The damage of
the monitor’s back cover may occur if removed by force.
2. Unscrew two PT3X8 screws locking the adjustable key board, and then remove the
adjustable key board assembly;
3. Take the adjustable keys out. Be careful of the positions of those keys when
re-assembling them.
4-21
Structure and Assembly/ Disassembly
4. Unscrew eight M3X8 screws on the machine assembly and front cover, and then
remove the front cover assembly of the monitor.
2. Unplug the speaker cable. Unscrew M2.6X8 screws (four screws on each left and
right side) on the speaker, and then take out the speaker.
4-22
Structure and Assembly/ Disassembly
4-23
Structure and Assembly/ Disassembly
4. Slightly press the upper support arm, screw out the screw pins on both siedes using
a Philips screwdriver, then remove the screw pins and spring washer M5.
Upper
Screw pins & M5 t
i h
5. Upturn the upper support arm to the limit, and remove the split washer on the tail of
parallel bracket of the upper support arm using the slotted screwdriver
4-24
Structure and Assembly/ Disassembly
Split washer
6. Hold the upper support arm, and meawhile, push out the bolt on the arm from one
side.
Bolt
Note:
1. Make sure to upturn the upper support arm to the limit before pushing out
the bolt on the arm.
2. Please stabely hold the upper support arm when pushing out the bolt to
avoid the upper support arm falling, besides, the upper support arm may
be ejected out when the bolt is removed, due to the spring force.
7. Pull the upper support arm out in the direction as per the figure, and draw out the
spring from the guiding sleeve.
4-25
Structure and Assembly/ Disassembly
Spring
Guiding sleeve
8. Screw out the two damping rubber column securing nuts using the wrench, and then
remove the damping rubber column from the tail of spring guiding sleeve.
Guiding
l
Damping
bb l
Damping rubber
l i t
NOTE: after the support arm (including the monitor) is installed, please check if the
monitor can stay at any height, if can’t, then adjust the spring adjusting nut using a
wrench to adjust the spring force until the monitor can stay at any height. Then cover the
upper support arm cover.
4-26
Structure and Assembly/ Disassembly
Spring
dj ti t
2. Remove three M3x8 screws on the side of main control panel and five M4x12 screws
at the bottom (The figure below doesn’t show the main unit but shows the positions
of the screws.)
4-27
Structure and Assembly/ Disassembly
5 M4x12
screws
3 M3x8
screws
Main control
panel assembly
4. Lift the front end of the panel, and disconnect the cables, and take out the main
control panel.
4-28
Structure and Assembly/ Disassembly
Minor panel
rear cover
assembly
Captive
screw Captive
4-29
Structure and Assembly/ Disassembly
4. Remove two M4x8 screws fixing the front side of the minor panel. Gently press the
two sides of the minor panel cover, take out the minor panel assembly. Withdraw the
connectors in the back slightly (be careful not to scratch the support arm), and take
off the minor panel. Press the two sides of minor panel also when assembling the
minor panel, and be careful not to scratch the support arm.
4-30
Structure and Assembly/ Disassembly
Two M3x8
screws
Trackball
Trackball cable
Slider
Big knob Middle knob
Small knob
3. Turn over the main control panel assembly, loosen the 16 fixing screws to take off the
fixing plate, encoder control board and four encoders.
4-31
Structure and Assembly/ Disassembly
16 M3x8 screws
Fixing plate
Fastening screw
4. Remove one M3x8 screw and 21 PT3x8 self-tapping screws shown in figure below to
take off the main control panel PCBA and replace it.
1 M3x8 screw
21 PT3x8 self-tapping screws
Main control panel PCBA
Encoder knob
4-33
Structure and Assembly/ Disassembly
Minor panel
6 PT3x8 encoder board
self-tapping screws
Connector3
Connector4
Connector1 Connector2
4-34
Structure and Assembly/ Disassembly
3. Remove the seven M3x8 screws to take out the 8.4 inch screen assembly.
Monitor assembly
Seven M3x8
screws
4. Remove the eight PT3x8 screws to take out the touch screen holder, and replace the
touch screen.
Touch screen
holder
4-35
Structure and Assembly/ Disassembly
Keyboard
assembly
2. After the keyboard assembly ejects in position, as shown in the figure (bottom view),
use two hands simultaneously to toggle the two plastic rods at the first nodes of the
slide tracks on the back surface of the keyboard assembly (squat down if necessary),
and slightly pull out the keyboard assembly; draw out the keyboard with the first
nodes of the sliding tracks.
Plastic rod
Plastic rod
Moving direction
Moving direction
3. Loosen the screw (M4x12) on the keyboard connector clip, and remove the cable
connector of the keyboard to take out the keyboard.
4-36
Structure and Assembly/ Disassembly
4. Remove the fixing screws (4 M4x8 panhead screws with washers) on remaining slide
tracks and keyboard support, and install the slide tracks to the new keyboard
assembly.
5. Insert the keyboard cable connector to the new keyboard assembly, and tighten the
screw on the keyboard connector clip.
6. Hold the keyboard by hands, align the tracks, push the keyboard evenly to the end,
and push and pull repeatedly to check if the keyboard is properly installed.
Main control
panel assembly
2. Remove two screws fixing the light control PCBA (2 M4x8 screws) using a
screwdriver, remove the light control PCBA and replace it with a new one, and install
4-37
Structure and Assembly/ Disassembly
the new PCBA. (Note the light shall face right to the hole on the cast-aluminum
panel).
Cast-aluminum cover
2. Remove 13 M4x12 screws fixing the cast-aluminum cover, and remove the
cast-aluminum cover, as shown in the figure below.
4-38
Structure and Assembly/ Disassembly
13 M4x12 screws
3. After the keyboard ejects completely, remove the four (4 M3x8) screws fixing the
slide tracks to take out the keyboard assembly and slide tracks.
Slide track
Keyboard assembly
4. Disassemble the slide tracks from the keyboard assembly, replace them with new
tracks and install on the keyboard assembly.
5. Align the new tracks with the locating poles on the cast-aluminum panel, place the
four screws loosened on the tracks, repeatedly push the keyboard assembly to
confirm that the keyboard can smoothly move and eject in position, and tighten the
four screws.
4-39
Structure and Assembly/ Disassembly
6. According to the reversed order, install the cast-aluminum cover, minor panel
assembly, main panel assembly and minor panel rear cover.
Lock seat
4-40
Structure and Assembly/ Disassembly
4-41
Structure and Assembly/ Disassembly
PHV board
Digital board
-5V board Beamformer
board
5V board
Amplifier
board
PFC board
Transmiss
ion board
4-42
Structure and Assembly/ Disassembly
10 screw
caps
13 M4×8
screws
2. Remove the cables from the I/O panel, i.e., remove the connectors from the I/O
interfaces. Be careful not to damage of the cables.
4-43
Structure and Assembly/ Disassembly
I/O
cables
3. Remove the two rows of M3x8 screws fixing the I/O assembly, hold the two grips by
hands, and take out the I/O assembly. Remove the HDD data cable, HDD power
cord and network cable connecting the I/O assembly and the digital board.
4-44
Structure and Assembly/ Disassembly
Screws
Handles
I/O
assembly
4. Release the board detacher, and take out the boards one by one.
System
boards
Detacher
4-45
Structure and Assembly/ Disassembly
Screws
Power shielding
cover
Auxiliary output
3. Release the board detacher and remove the boards one by one.
Power board
Detacher
4-46
Structure and Assembly/ Disassembly
M4x8 screws
4. Remove the 10 M3x8 screws fixing the cover of mounting box, remove the nine
M3x8 screws fixing the 4D drive board to remove the 4D&TEE drive board.
4-47
Structure and Assembly/ Disassembly
M3x8 screws
4D drive board
4D module
mounting box
4-48
Structure and Assembly/ Disassembly
M3x8 screws
HDD assembly
I/O shielding
cover
2. Remove the 4 special screws fixing HDD with the support and take the HDD off.
4 screws
4-49
Structure and Assembly/ Disassembly
2 M4×8 screws
(2) Take out the DVD-R/W assembly, remove the power cord and signal cables on it.
4-50
Structure and Assembly/ Disassembly
Signal
cable DVD-R/W
assembly
Power
cord
3. Remove DVD-R/W: Remove the four M3x8 screws (fixing DVD on the DVD holder)
at the bottom of the DVD-R/W to take out the DVD-R/W.
DVD-R/W
DVD-R/W holder
4 M3x8 screws
4-51
Structure and Assembly/ Disassembly
3 M4×8 screws
Footplate
6 M4×8 screws
6 screw
caps
4-52
Structure and Assembly/ Disassembly
(3) Remove the four M4x8 screws attaching the front cover to the main unit housing.
(4) Pull the font cover down until it is detached from the buckle of the neck front
cover, and remove the front cover assembly.
4 M4×8
screws
4. Remove four M4x8 screws of the ECG assembly, pull it up until the ECG assembly is
detached from the main unit housing, and remove the ECG assembly.
4-53
Structure and Assembly/ Disassembly
Hook on
ECG
assembly
ECG
assembly
M4x8 screw
5. Remove the M3x8 screws, disconnect the cables, and remove the ECG base
assembly and ECG mounting cover.
4-54
Structure and Assembly/ Disassembly
ECG mounting
assembly
6. Remove M3x8 screws, and replace the ECG board or footswitch connecting PCBA.
M3x8 screws
ECG base
ECG board/
footswitch
connecting board
4-55
Structure and Assembly/ Disassembly
22 M3x8
screws
Grips
3. Remove the probe control board: Remove the 16 M3x8 screws fixing the probe
control board, and pull up the control board until it is detached from the probe board.
4-56
Structure and Assembly/ Disassembly
M3x8 screws
4-57
Structure and Assembly/ Disassembly
M3x8 screws
Bolt screw
Probe shielding
cover
M3x8 screws
(2) Remove the 24 M3x6 countersunk screws fixing the shielding frames and
shielding plates of the probe sockets, and then remove the three probe shielding
frames and one 4D probe socket shielding frame;
M3X6
countersunk
head screws
Aluminum shielding
frame
4D probe socket
shielding frame
4-58
Structure and Assembly/ Disassembly
(3) Remove 16 M3x6 countersunk screws fixing the probe mounting brackets and
shielding plate, and remove six probe mounting brackets and two 4D probe
mounting brackets.
M3X6
countersunk head
screws
Probe mounting
bracket
4D probe
mounting bracket
(4) Remove 8 M3x8 screws and 8 M3x7 bolt screws fixing the probe board and
shielding plates, and then remove the probe board.
4-59
Structure and Assembly/ Disassembly
M3x8
Probe board
Aluminum shielding
plate
M3×8 screws
Bolt screws
Shielding cover
of the probe
M3×8 screws
4-60
Structure and Assembly/ Disassembly
(2) Remove the 8 M3x14 panhead screws fixing the washer blocks, then you can
take out four washer blocks.
M3x14 panhead
Washer block
(3) Remove the 24 M3x6 countersunk screws fixing the shielding frames and
shielding plates of the probe sockets, and then remove the two probe shielding
frames and two 4D probe socket shielding frames;
4-61
Structure and Assembly/ Disassembly
M3x6
countersunk
screws
Probe shielding
frames
(4) Remove 16 M3x6 countersunk screws fixing the probe mounting brackets and
shielding plate, and remove four probe mounting brackets and four 4D probe
mounting brackets.
4-62
Structure and Assembly/ Disassembly
M3x6
countersunk
screws
Probe mounting
brackets
4D probe mounting
brackets
(5) Remove 8 M3x8 screws and 8 M3x7 bolt screws fixing the probe board and
shielding plates, and then remove the probe board.
M3x7 bolt
M3x8 screws
Probe board
Probe shielding
plate
4-63
Structure and Assembly/ Disassembly
2 captive screws
2. Move the nylon screen support to the side until it is detached from the hook, and take
out the nylon screen support.
4-64
Structure and Assembly/ Disassembly
3. Replace the nylon screen and the support with new ones.
4-65