PCB Experiment 2
PCB Experiment 2
Name :
Register No. :
Day / Session :
Venue : 2nd Floor, BEL307 & BEL308, Basic Engineering Lab.
Title of Experiment :
Date of Conduction :
Date of Submission :
Pre-lab questions 10
Total 50
REPORT VERIFICATION
Date :
Staff Name :
Signature :
Expt.- 2: Study and analysis of testing and measuring instruments
Aim:
• Study of various testing and measuring instruments such as Voltmeter, Ammeter,
Digital multi-meter, function generator, CRO, DSO & IC tester
• Observations using Digital multi-meter, function generator, CRO, DSO
Components Required:
Voltmeter, Ammeter, Digital multi-meter, function generator, CRO, DSO, IC tester
Theory:
Voltmeter
A voltmeter is an instrument used for measuring electric potential difference between two
points in an electric circuit. Analog voltmeters move a pointer across a scale in proportion to
the voltage of the circuit; digital voltmeters give a numerical display of voltage by use of an
analog-to-digital converter.
Voltage measurement
D.C. / A.C. Voltage Measurement
1. Connect the positive (red) test lead to the „V/mA‟ jack socket and the negative (black)
lead to the „COM‟ jack socket.
2. Set the selector switch to the desired mV D.C./D.C.V/A.C.V range.
3. Connect the test leads to the circuit to be measured.
4. Turn on the power to the circuit to be measured, the voltage value should appear on the
digital display along with the voltage polarity (if reversed only).
Current measurement
1. Connect the positive (red) test lead to the „V/mA‟ jack socket and the negative (black)
lead to the „COM‟ jack socket (for measurements up to 200mA). For measurements
between 200 mA and 10A connect the red test lead to the „10mA‟ socket.
2. Set the selector switch to the desired uA/mA/A range.
3. Open the circuit to be measured and connect the test leads in SERIES with the load in
which current is to be measured.
4. To avoid blowing an input fuse, use the 10A jack until you are sure that the current is
less than 300 mA. Turn off power to the circuit. Break the circuit. (For circuits of more
than 10 amps, use a current clamp.) Put the meter in series with the circuit and turn
power on.
Resistance measurement
1. Connect the positive (red) test lead to the “V/mA” jack socket and the negative(black)
lead to the “COM” jack socket.
2. Set the selector switch to the desired OHM (Ω).
3. If the resistance to be measured is part of a circuit, turn off the power and discharge all
capacitors before measurement.
4. Connect the test leads to the circuit to be measured.
5. The resistance value should now appear on the digital display.
Continuity test
This mode is used to check if two points are electrically connected. It is often used to verify
connectors. If continuity exists (resistance less than 210 ohms), the beeper sounds
continuously.
1. Connect the positive (red) test lead to the „V/mA‟ jack socket and the negative (black)
lead to the „COM‟ jack socket.
2. Set the selector switch to the position.
3. Connect the test leads to two points of the circuit to be tested. If the resistance is Ohms
the buzzer will sound. If the resistance to be measured is part of a circuit, turn off the
power and discharge all capacitors before measurement.
Diode test
1. Connect the positive (red) test lead to the “V/mA” jack socket and the negative (black)
lead to the “COM” jack socket.
2. Set the selector switch to the position.
3. Connect the test leads to be measured.
4. Turn on the power to the circuit to be measured and the voltage value should appear on
the digital display.
General Operation
Connection of Probes:
All multimeters come with two probes. They are to be connected to the terminals on the meter
itself. The Black probe is to be connected to the COM terminal. Red probe is to be connected
to terminal marked with:
• “V” for voltage measurement
• “mA” or “20A” for current measurement (there are two terminals, one for 2A range and
the other for 20A range)
• “Ω” for resistance measurement.
Setting of function:
The multimeter uses different circuits internally to measure different things. Therefore, you
must select the correct function before using it.
Setting of Range:
• You can change the sensitivity of the meter by selecting different range for
measurement. Set the range to the first range that is higher than the maximum value
you expect to measure. This will give a more accurate reading. If you do not know what
to expect, use the highest range first. After a reading is obtained, set the range to the
appropriate one to get a better reading.
• When the value measured exceeds the existing range, the display will flash. When this
happens, set the multimeter to a higher range until some values are displayed.
Precautions:
1. For current measurement, the maximum input current is 2A (if the RED mA and
BLACK COM terminals are used) or 20A (if the RED 20A and BLACK COM
terminals are used). Excessive current will blow the fuse on the 2A range, which must
be replaced. The 20A range, however, is not protected by the fuse.
2. For voltage measurement (use the RED V- and BLACK COM terminals), the maximum
input voltage on:
• all DC ranges is 1200V DC or peak AC
• the 20V, 200V and 1000V AC ranges is 1000V rms continuous
• the 2V and 200mV AC ranges is 1000V rms for not more than 15 seconds.
3. To avoid electrical shock and/or instrument damage, do not connect the COM input
terminals to any source of more than 500 volts DC or peak AC above earth ground.
Resistance measurement
Procedure
1. Connect probes: black probe to COM terminal and red probe to terminal marked with
“Ω”
2. Set function to resistance measurement
3. Set to the appropriate range (refer to above)
4. Connect the two probes‟ crocodile clips to the resistor (or to the resistor circuit via
jumper wires) to make measurement
5. Note the reading, adjust range if necessary
6. Take the more accurate reading.
Voltage measurement
Procedure
1. Connect probes: black probe to COM terminal and red probe to terminal marked with
“V‟”
2. Set function to voltage measurement
3. Set to the appropriate range (refer to above)
4. Set the AC-DC selection - depends on what type of signal you want to measure
5. Touch the two points where you want to make measurement
6. Note the reading, adjust range if necessary
7. Take the more accurate reading.
Note
Reading obtained is the voltage of where the red probe touch with reference to where the black
probe touches. This may not indicate the voltage level from ground. To find the voltage level
of a point from ground, black probe should be touching a ground point and the red probe on
the point you want to measure.
Current measurement
Procedure
1. Connect probes: black probe to COM terminal and red probe to terminal marked with
“A”
2. Set function to current measurement
3. Set to the appropriate range
4. Set AC-DC selection - depends on what type of signal you want to measure
5. Off the power to the circuit
6. Break the path which we want to make measurement
7. Connect the path with the two probes so that current now flow through the multimeter
8. On the power
9. Note the reading, change range if necessary
10. Take the more accurate reading.
Note
Use the 20A range if you are not sure of the current to be measured.
Connection Methods to Measure Voltage and Current
Function generator
A function generator is electronic test equipment used to generate different types of waveforms
over a wide range of frequencies. Function generators are capable of producing a variety of
repetitive waveforms, generally from the list below
Sine wave: A function generator will normally have the capability to produce a standard sine
wave output. This is the standard waveform that oscillates between two levels with a standard
sinusoidal shape.
Square wave: A square wave is normally relatively easy for a function generator to produce.
It consists of a signal moving directly between high and low levels.
Pulse: A pulse waveform is another type that can be produced by a function generator. It is
effectively the same as a square wave, but with the mark space ratio very different to 1:1.
Triangular wave: This form of signal produced by the function generator linearly moves
between a high and low point.
Saw tooth wave: Again, this is a triangular waveform, but with the rise edge of the waveform
faster or slower than the fall, making a form of shape similar to a saw tooth.
These waveforms can be either repetitive or single-shot Function generators are used in the
development, test and repair of electronic equipment.
Types of Function Generator
Fig 6. Cathode Ray tube Oscilloscope Fig 7. Out Line Diagram of CRO
Procedure
1. Turn on the Oscilloscope
2. Adjust the intensity and the focus of the trace.
3. Use the X & Y knobs to center the trace horizontally and vertically.
4. Connect the cable from Ch1 of the CRO to Function generator.
5. A signal will appear on the screen.
6. Make sure that the inner red knobs of the Volt/Div and the Time/Div are locked
clockwise.
7. Set the frequency of the generator to 100 Hz.
8. Adjust the Volt/Div and the Time/Div knobs so that you get a suitable size signal
9. Count the number of vertical squares lying within the signal, then calculate the peak to
peak value as:
Figure 9. IC Tester
LAB Exercise