Appendix KOVA Counting Procedure
Appendix KOVA Counting Procedure
Material
Chamber preaparation
Fill the KOVA-Chamber by pipetting 10μl of the sample, with an accurate Eppendorf
pipette, into the small opening. Avoid air bubbles, discard incorrect filled chambers.
Let the chamber stand for one minute before counting.
Get an overview of the urine content by scanning the whole chamber (9x9 grids) at low
magnification (100x).
Count the particles in the grids and the particles touching the left and the lower border of
the grid. Do not count the particles touching the upper and the right border of the grid.
Do not count particles under the wide outer grid-lines if they do not touch the inner
boarder. (Count, do not count)
Two chambers are counted by two investigators (using a discussion microscope). Build
the mean of each chamber and afterwards build the average over these both single
chambers calculate the final result as described below.
Write the number of counted grids, the interim means and the final results in the form:
“Manual microscopy with KOVA-chamber”.
Counting of particles
Count HYA, SEC and PAT at low magnification (100x) inside the complete grid (81
small grids 0.9 μl). Multiply by 1.1 = cells/μl
Please note that single mucus strings are recorded as negative. Please find below examples
about negative and positive results.
Please note that yeast is mostly seen in chains and not that often found as single cells but
the rating is only positive or negative. There will be no counting of the yeast cells
performed. Please find below examples about negative and positive results.
1+
2+
3+
Roche Diagnostics International Ltd.
RPD Clinical Operations 19-Sept-2013
Mucus
Single mucus
string; rated as
negative
Many mucus
strings; rated as
positive