This document provides installation and configuration instructions for the Electrical Control Unit Model 21-35-NM-F. Key points include: 1. Installation should follow applicable electrical codes and provide a minimum IP64 protection level. 2. Separate conduits/cable glands should be used for probe and power circuits. Unused cable glands must be sealed by the user. 3. Jumpers SW3 and SW4 determine whether the control unit responds as flow or level detection and high or low level alarm. SW1 and SW2 set delay times to prevent false alarms.

Uploaded by

felixz911
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
39 views6 pages

Configuración Tilt Switch

This document provides installation and configuration instructions for the Electrical Control Unit Model 21-35-NM-F. Key points include: 1. Installation should follow applicable electrical codes and provide a minimum IP64 protection level. 2. Separate conduits/cable glands should be used for probe and power circuits. Unused cable glands must be sealed by the user. 3. Jumpers SW3 and SW4 determine whether the control unit responds as flow or level detection and high or low level alarm. SW1 and SW2 set delay times to prevent false alarms.

Uploaded by

felixz911
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 6

Electrical CONTROL UNIT Model 21-35-NM-F

Installation 1 Installation should be in accordance with provincial or


regional Electrical Code.

2 The cable entry devices or elements for the Control Unit


must provide a degree of protection of at least IP64.

3 The Control Unit enclosure is supplied with (2) ¾”


conduit hubs (3) or M20x1.5 cable glands.
-For ATEX Models: The supplied cable glands accept
cable diameters from 8.0 mm to 10.5 mm.

4 Use separate conduits/cable glands for probe and power


circuits connection.
-For ATEX Models: It is forbidden to pass more than one
cable through the same cable gland, if necessary, use
multipoles cables.

5 The sealing of unused cable glands is the user’s


responsibility.

6 If the cable glands that are supplied are not used for the
user’s wiring, the user is responsible for using certified
cable glands with the correct ratings for the cable entries.

7 The Ground connection must be secured on the


enclosure.

8 The user is responsible for using the correct terminals for


ground connection to the frame screw.

9 The terminals will accept a 12 AWG stranded wire


(maximum) or 2 mmsq

10 The cable length of the probe shall not exceed 400 feet.

11 For electrical connections, refer to Figure 2-6.

Thermo Fisher Scientific 2-7


REC-4266 Rev E
Figure 2-6. Control Unit Model 21-35-NM-F – Terminal Wiring

Figure 2-7. Control Unit Model 21-35-NM-F – Internal Layout

2-8 Thermo Fisher Scientific


REC-4266 Rev E
Chapter 3 Operation

When the Ramsey Mercury Free Tilt Sensor for Hazardous Locations
system is used as a no-flow detector, the normal position of the
probe is tilted. When used as a level detector, the normal position
is vertical (not tilted). In both cases the tilted condition indicates
the presence of material.

FLOW APPLICATION EXAMPLES

Thermo Fisher Scientific 3-1


REC-4266 Rev E
LOW APPLICATION EXAMPLES

Typical Plugged Transfer Chute “Alarms”

3-2 Thermo Fisher Scientific


REC-4266 Rev E
System Different applications may have different probe positions that
indicate a “normal” condition (vertical vs. tilted). There are two
Configuration jumpers on the circuit board to control the behavior of the alarm
lights and contacts in regard to probe position. Use the following
Function Mode tables to determine the correct jumper settings for
your application.

Model 21­35­NM­F ­
Function Modes
Table 3-1. Model 21-35-NM-F – Function Modes Switch
Settings

SW3 SW4 Function Probe Condition Lamp Delay on


(See table (See Mode Position Alarm Transition to
3-3) table 3-4) (See table 3-2)
DEFAULT SETTING Alarm for
High Level Tilted Alarm ON X
Level (B) High (A) Detection
Vertical Normal OFF
Alarm for
Level (B) Low (B) Low Level Tilted Normal OFF
Detection
Vertical Alarm ON X
Alarm for No
Flow (A) Low (B) Flow Tilted Normal OFF
Detection
Vertical Alarm ON X
Alarm for
Flow (A) High (A) Flow Tilted Alarm ON X
Detection
Vertical Normal OFF

Thermo Fisher Scientific 3-3


REC-4266 Rev E
Model 21­35­NM­F – SW1 and SW2 Settings are used to delay the indication of alarm
conditions. This will prevent alarms due to material momentarily
Switches Setting moving the probe.

Table 3-2: SW1 and SW2 Settings

SW1 SW2 Delay in Seconds


B B 1
A B 2
B A 4
A A 6

SW3 determines whether the alarm functions of the control unit


respond like a Flow or Level application.

Table 3-3: SW3 Settings

SW3 Mode
A Flow
B Level

SW4 determines whether the alarm functions of the control unit


respond like a High or Low Level application.

Table 3-4: SW4 Settings

SW4 Mode
A High Level
B Low Level

SW5 defines the instrument electrical power supply.

Table 3-5: SW5 Selector Setting

SW5 Power Supply


Position 115VAC
Position 230VAC

3-4 Thermo Fisher Scientific


REC-4266 Rev E