ASVCP QALS Hematology TEa Gu
ASVCP QALS Hematology TEa Gu
ASVCP QALS Hematology TEa Gu
Mary B. Nabity1, Kendal E. Harr2, Melinda S. Camus3, Bente Flatland4, Linda M. Vap5
1Department of Veterinary Pathobiology, Texas A&M University, College Station, TX; 2Urika,
LLC, Mukilteo, WA; 3Department of Pathology, University of Georgia, Athens, GA;
4Department of Biomedical and Diagnostic Sciences, College of Veterinary Medicine, University
Developed by the American Society for Veterinary Clinical Pathology (ASVCP) Quality
Assurance and Laboratory Standards (QALS) Committee
If citing this document, the following format is suggested: American Society for Veterinary
Clinical Pathology (ASVCP). ASVCP Guidelines: Allowable Total Error Hematology, Version
1. 2017. Available at http://asvcp.org/pubs/qas/index.cfm Accessed January 9, 2018.
Page 1 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
TABLE of CONTENTS
1 Introduction .................................................................................................................................. 3
2 Scope ............................................................................................................................................ 3
7 References .................................................................................................................................. 17
11 Acknowledgments.................................................................................................................... 22
12 Tables ....................................................................................................................................... 23
13 Appendices ............................................................................................................................... 27
13.1 Appendix A. Derivation of hematology TEa recommendations ...................................................................... 27
13.1.1 Selection of measurands ......................................................................................................................... 27
13.1.2 Clinician input .......................................................................................................................................... 27
13.1.3 TEa based on assessment of instrument performance ........................................................................... 29
13.1.4 Determination of recommended TEa. ..................................................................................................... 30
13.2 Appendix B. External quality assurance/proficiency testing programs.......................................................... 33
13.3 Appendix C. Allowable total error worksheet ................................................................................................. 34
Page 2 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
1 Introduction
Analytical quality requirements (a.k.a. quality specifications) are pre-determined
benchmarks used to judge analytical performance of laboratory instruments or methods.1
Quality requirements may vary by type of laboratory, measurand concentration (e.g., low vs.
high) and species tested. Quality requirements can be derived from several sources, including
governmental regulatory requirements, expert opinion, biological variation data, and current
instrument performance. Regardless of source, it is essential that quality requirements be
clinically relevant for medically important measurand concentrations and be realistic for
available laboratory technology.2
Allowable total error (TEa) defines a quantitative goal combining imprecision (random
error) and inaccuracy (systematic error, or bias) to determine acceptable variation in a single
measurement procedure without interfering with the clinical interpretation of patient data.
2 Scope
Objectives of this guideline are to provide TEa recommendations for hematology
measurands routinely assayed in veterinary practice and to provide an overview of how these
recommendations were derived. A worksheet for determining whether instrument performance
meets these recommendations is also included. Similar to biochemistry testing, TEa
recommendations for hematology testing in veterinary medicine are needed to facilitate
instrument performance evaluation, method comparison, and quality control validation. Intended
audiences include, but are not limited to, reference laboratories, in-clinic laboratories, and animal
health diagnostic companies supplying hematology instruments to the veterinary market. TEa is
proposed here for those measurands in common clinical use and likely to be evaluated with
routine quality control procedures. Furthermore, the TEa recommendations were based on
clinician input regarding dogs, cats, and horses, along with data generated from quality control
material (QCM) and dog specimens using automated methods. This guideline is thus not all-
inclusive but presents TEa recommendations considered to be suitable for the current instruments
and methods commonly used for veterinary hematology measurands. While blood from non-
mammalian species was not tested, the recommendations could serve as a baseline for
laboratories evaluating blood from non-mammalian animals.
Bias (a.k.a. inaccuracy) – Total systematic error, which includes constant and proportional bias.
Bias is the difference between the test instrument’s measured result and the true value (e.g., as
measured by a reference method or as defined by a known standard). The term bias in difference
Page 3 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
plot analysis (expressed in measurand units) equals the difference between the values of the two
methods being compared or the average of all the differences between the paired sample values.
Bias may also be expressed as a percentage according to the formula:
𝑀𝑒𝑎𝑛𝑡𝑎𝑟𝑔𝑒𝑡 − 𝑀𝑒𝑎𝑛𝑚𝑒𝑎𝑠𝑢𝑟𝑒𝑑
𝐵𝑖𝑎𝑠% = × 100
𝑀𝑒𝑎𝑛𝑡𝑎𝑟𝑔𝑒𝑡
Bias, constant – When the degree of systematic error remains the same over the range of
measurand concentrations (i.e., results of one method are consistently above or below another
method and roughly by the same amount, regardless of measurand concentration).3
Bias, proportional – When the magnitude of systematic error changes as the measurand
concentration changes.3
Comparability Testing – A quality assurance procedure in which measurement results from two
or more instruments or methods are compared to each other for purposes of analytical
performance assessment. Comparability testing can be a component of formal EQA/PT programs
or can be carried out independently within a laboratory or network of laboratories.5 Total
allowable error is one tool that is used in comparability testing and aids in assessment of whether
results from different instruments can be used interchangeably without causing clinical error. 3
Decision Threshold – Clinical or medical decision limit (i.e., numerical value) at which
important clinical decisions regarding testing or treatment are made for a particular measurand.
Page 4 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
EQA/PT specimen, testing item, test material, or check specimen panel - A specimen
containing measurands of undisclosed concentrations or compositions sent to a participating
laboratory to assess the laboratory’s testing competency.6
Measurand - A particular quantity subject to measurement under specified conditions (e.g. the
enzymatic activity of alkaline phosphatase at 37°C).6
Peer group – Used for comparison of quality requirements and defined by the same instrument
and/or method as that used by the participating laboratory or testing site.
Precision – Closeness of agreement between independent, repeated results obtained from the
same specimen under specific conditions. These may be derived in the same day (repeatability
study) or on different days (within laboratory precision).7 Note: The definition of precision has
become more complex in recent years and is frequently being modified. Readers are referred to
other sources for further definition.8
Reference Interval – An interval that contains all the possible values between and including an
upper and lower limit. Reference limits are defined such that the reference interval contains a
specified proportion of values from a well-defined, typically clinically healthy reference
population. Reference interval is preferred over the term reference range. 10
Repeatability – Precision of analysis when repeated using the same operator, measurement
procedure, equipment, time, and laboratory.7
Page 5 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
Repeat patient testing (RPT) – Repeat testing of individual patient specimens under specified
conditions that is used as a statistical quality control method. RPT exploits the fact that
specimen deterioration under defined conditions (fixed time interval, storage conditions, etc.)
causes an expected degree of variation in results; any variation in results exceeding this threshold
may indicate a problem with the test system.11,12
QALS (Quality Assurance and Laboratory Standards Committee of the ASVCP) – The
ASVCP committee charged with encouraging and promoting the establishment of standards for
the performance of laboratory procedures on animal specimens.
QCM (quality control material) – A test material intended by its manufacturer to be used for
QC of laboratory testing. Measurement of QCM monitors the entire test system (operator,
reagents, and instrument analytical function). QCM may be used to carry out an instrument
performance study or to monitor routine analytical performance. “Assayed QCM” is QCM that
has been measured by the manufacturer, which then provides target means, ranges, standard
deviation and CV for that QCM for specific instruments or methods.13
Quality Goal Index – A numerical index that reflects whether imprecision, bias, or both are
contributing to an observed analytical error. QGI may be calculated according to the formula14
%𝐵𝑖𝑎𝑠
𝑄𝐺𝐼 =
𝐶𝑉
Standard Deviation (SD) – A measure of variability or diversity associated with random error
or imprecision. SD demonstrates the variation or dispersion from the mean (average or other
expected value) during repeated measures. A small SD indicates that data points tend to be very
close to the mean, whereas a large SD indicates that data points are spread over a wide range of
values. SD is the square root of a dataset’s variance.3
Page 6 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
TE (total error, total analytical error) – The sum of random error (imprecision) and systematic
error (bias or inaccuracy). This term may also incorporate other sources of error (e.g., pre‐
analytical variation, biologic variation, and other factors) that contribute to the variation seen in
patient results. TE may be expressed in measurand units or as a percentage.3
TEa (allowable or desirable total error) – A quality requirement that sets a limit for combined
imprecision (random error) and bias (inaccuracy, or systematic error) that are tolerable in a
single measurement or single test result to ensure clinical usefulness.3 Recommendations for
hematology TEa are found in section 5.
TEobs (observed or calculated total error) – The sum of measured random error (imprecision)
and systematic error (bias/inaccuracy) that can be calculated from instrument performance data
according to the formula as defined in this guideline
Z score – A unitless number that is a coefficient indicating the number of standard deviations
from the mean. The z score (a.k.a. standard score, z value) is arbitrary in TEa and dependent
upon the stringency desired for the test. The ASVCP consensus-approved TEa guidelines and
CLIA documentation assign the z score of 2 for calculation of TEa in laboratory medicine. 18
Page 7 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
Page 8 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
While the above factors are all recognized to influence TEa, the recommendations proposed in
this guideline were largely based on clinician recommendations for dogs, cats, and horses as well
as instrument performance based on quality control material (QCM) and dog specimens.
Page 9 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
4.2.3 Platelets
Blood specimens from cats are known to clot quickly, and their platelets are
frequently clumped upon evaluation of a blood smear. Therefore, TEobs in cats can be
much more variable than in other species, and high TEobs for feline platelets was found
due to imprecision in one study.25 Therefore, TEa for platelets presented in this guideline
focus on dogs and horses. Platelet concentrations <50,000/µL in any species are often
associated with a higher degree of error than the recommended TEa of 20%, even when
TEobs is based only on imprecision (2*CV). Imprecision is particularly high when counts
are extremely low (e.g., <10,000/µL). However, at these extremely low concentrations,
high error means that small changes are not biologically significant despite the fact that
clinical decisions are often made based on such changes. For example, a 50% TEobs for
10,000/µL platelets means that the patient result can range from 5,000/µL to 15,000/µL.
4.2.4 Reticulocytes
Automated reticulocyte counts in cats are problematic due to the presence of both
punctate and aggregate reticulocytes, and higher CV is observed for reticulocytes in cats
than in dogs.26 Reticulocyte counts in horses are rarely clinically relevant and may be
subject to increased error, given the higher degree of error associated with reticulocyte
counts <60,000/µL (see Appendix). While some data support a relatively close
correlation between manual and automated reticulocyte counts, precision of manual
reticulocyte counts is thought to have higher error.33,34 Therefore, reticulocyte
recommendations in this document focus on reticulocyte numbers generated by
Page 10 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
For purposes of calculating the total error of a measurand using a particular method or
instrument (TEobs), Westgard originally used TE = bias(%) + 1.65CV.35 This formula is the
basis for the TEobs calculation used in this document, which is currently the most widely
accepted formula:36
𝑻𝑬𝒐𝒃𝒔 = 𝒂𝒃𝒔𝒐𝒍𝒖𝒕𝒆 𝒃𝒊𝒂𝒔% + 𝟐𝑪𝑽
If units of the measurand are used, then the equation used to calculate an instrument’s total error
(TEobs) changes to:
𝑻𝑬𝒐𝒃𝒔 = 𝒂𝒃𝒔𝒐𝒍𝒖𝒕𝒆 𝒃𝒊𝒂𝒔 (𝒊𝒏 𝒎𝒆𝒂𝒔𝒖𝒓𝒂𝒏𝒅 𝒖𝒏𝒊𝒕𝒔) + 𝟐𝑺𝑫
where SD is standard deviation.
Absolute values for bias should be used in these formulae (i.e., negative values should not be
used).15
Calculation of an instrument’s TEobs (for purposes of comparing to TEa) can be based
on routine daily QC data and/or periodic EQA/PT data, both of which are recommended to
ensure ongoing production of reliable laboratory results. The frequency of quality assurance (QA)
monitoring can be determined by the QC specialist based on the number of samples analyzed per
day, known inherent drift of the analytical method, perception of previous problems noted in the
laboratory, cost of reagents, and other factors.3
Page 11 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
measurands present within the materials, shelf-life/stability, cost, etc. In addition, both in-
clinic QC and EQA/PT require different types (concentrations, species, cell sizes, etc.) of
QCM, dependent upon the instrument, reference intervals, expected changes due to
disease, and species evaluated at the facility. Hematologic QCM must be selected based
on the methodology of the instrument; QCM for impedance-based and flow cytometric-
based instruments are not interchangeable. QCM from different lots may not have the
exact same measurand concentrations, which must be factored into the control limits used
to decide if QC data are in-control or out-of-control.37 Ideally, a minimum of two levels
(measurand concentrations) of assayed QCM should be used to determine instrument
performance.22 The concentration of measurand in the QCM should be at or near those of
decision thresholds and/or reference limits. If only one concentration of QCM is used, then
ideally it should be consistent with reference values for the species.
All commercially available QCM and calibration materials have a lot number and
expiration date based on proper storage of unopened vials. QCM should be labeled with
the date it is opened and the expiration date based on opening. All QCM should be
promptly discarded upon reaching either of its expiration dates (i.e., expired QCM
should never be used). Hematologic QCM may degrade more quickly upon opening
than QCM for biochemical testing, and it may show signs of degradation near the end of
(although prior to) its expected shelf life.
Manufacturer’s recommendations regarding storage and handling should be
followed and included in the laboratory’s Standard Operating Procedures.38 QCM that
are transported or stored under inappropriate conditions may lead to errors. For
example, using compromised QCM can trigger unnecessary troubleshooting and/or
calibration. Furthermore, if compromised calibration materials are used to calibrate an
instrument, it will lead to systematic error. Therefore, conditions known to alter the
stability of the QCM (temperature, light, humidity, length of storage, etc.) must be
monitored to ensure its stability. If it is suspected that the stability of the QCM is
compromised, it should be discarded and replaced. Shipment of hematology QCM and
whole blood between laboratories frequently results in disparate measurements due to
transport conditions and degradation over time. Even under appropriate conditions of
transport and storage, measurand results may vary over the lifetime of the QCM while
remaining within the expected intervals. Therefore, when comparability assessment
between laboratories is desired, assayed QCM should be analyzed at approximately the
same time (i.e., within 6 hours), on the instruments to be compared. This may require
splitting and shipping of QCM in aliquots instead of analyzing the QCM on different
days at different facilities. Timing and arrival should be planned so that weekends and
holidays are avoided. In contrast, QCM for chemistry and endocrinology can usually be
aliquoted and frozen for some time.
Page 12 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
Page 13 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
1. Measure each QCM daily at least five times.22 Five repetitions in one day is
possible but does not incorporate potential interday variation that mimics
conditions when assessing samples from hospitalized patients over time. If
Page 14 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
The mean, SD, and CV derived from these QC data are referred to as the
‘measured,’ ‘calculated,’ or ‘observed’ mean, SD, and CV.
2. Calculate the analyzer’s measured bias using the measured mean and the
QCM manufacturer’s reported mean (i.e., target mean) for the assayed control
material (using the same instrument and/or method as that used by the
analyzer) according to the formula:
𝑴𝒆𝒂𝒏𝒕𝒂𝒓𝒈𝒆𝒕 − 𝑴𝒆𝒂𝒏𝒎𝒆𝒂𝒔𝒖𝒓𝒆𝒅
𝑩𝒊𝒂𝒔% = × 𝟏𝟎𝟎
𝑴𝒆𝒂𝒏𝒕𝒂𝒓𝒈𝒆𝒕
QCM manufacturer’s reported means are commonly found in the QCM package
insert, categorized according to the instrument and method producing the assayed
values. Measured bias may be a positive or a negative number, depending upon
whether the analyzer’s results are lower or higher than the manufacturer’s
reported mean. If bias is a negative number (e.g., 5.0%), then the absolute
number (5.0%) should be used in step 3, below.
Page 15 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
inaccuracy (high bias).24 Use of special calculations, such as the Quality Goal Index may
be helpful in determining if the poor performance is due to imprecision, inaccuracy, or a
combination of both.14 If unacceptably large imprecision is suspected, then a more
rigorous precision study should be performed, including verification of manufacturer
performance claims, if this has not already been done.7 If unacceptably large bias is
suspected, then the means by which bias was determined should be re-visited to
determine if the targets are optimal. Instrument performance re-evaluation using a more
appropriate target for bias determination (i.e., a different representation of true measurand
concentration) could be considered.
If these sources of error cannot be corrected or if problems occur repeatedly, the
manufacturer of the instrument and/or a board-certified clinical pathologist with expertise
in QA should be called upon for further assessment. Further assessment may include
attempts to improve performance capability by analyzer adjustments, operator training,
reagent replacement with a new reagent or a product from a different manufacturer, or,
potentially, analyzer replacement.24
Alternatively, the initial quality requirement may be relaxed. This approach is
acceptable only if diagnostic judgment deems that additional analytical error can be
tolerated. Furthermore, this option should only be used upon consultation with a board-
certified veterinary clinical pathologist or other QC specialist. Relaxation of the TEa for a
particular measurand requires education of ALL clinicians using analyzer results that this
measurand is associated with larger error than is recommended. Use of a TEa higher than
that recommended in this document should be justified and documented in a laboratory
handbook.
5 TEa recommendations
Table 1 and Table 2 summarize TEa recommendations for hematologic measurands.
These recommendations were partially based on instrument performance using QCM (see
Appendix A for the control ranges evaluated). Each instrument’s manufacturer guidelines should
be consulted to determine the range of values supported by the instrument. Most instruments
cannot accurately quantify extreme values. It is worth noting that use of quality requirements for
hematology testing focuses on automated data, and manual review of blood smears is necessary
to verify automated findings.
Page 16 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
Biological variation of hematology measurands has been studied in dogs; example quality
specifications calculated from selected data are presented in Table 3. Based on instrument
performance data gathered by the authors (Appendix A), current veterinary reference laboratory
state-of-the-art instrumentation can typically meet “minimum” BV-based TE. In fact, it can often
meet “desirable” TE for most hematology measurands. The ability to meet “desirable” and
“optimum” BV-based TE varied by measurand and institution. Unsurprisingly, “optimum” TE
(the most stringent BV-based quality requirement) was met least often and is currently not
recommended for routine assessment of hematology instrument performance. Overall, TEa
values recommended in this guideline are comparable to published minimum TE based on
biological variation data (Table 3), and either type of quality requirement can be used to evaluate
hematology instrument performance.43-45
7 References
1. Sandberg S, Fraser CG, Horvath AR, et al. Defining analytical performance specifications:
consensus statement from the 1 st strategic conference of the European Federation of Clinical
Chemistry and Laboratory Medicine. Clin Chem Lab Med. 2015; 53(6): 833-835.
2. Kjelgaard-Hansen M, Jensen AL. Subjectivity in defining quality specifications for quality
control and test validation [letter]. Vet Clin Pathol. 2010; 39: 134–135.
3. Harr KE, Flatland B, Nabity M, Freeman KP. ASVCP guidelines: allowable total error
guidelines for biochemistry. Vet Clin Pathol. 2013; 42: 424-436.
4. Freeman KP, Baral RM, Dhand NK, Saxmose Nielsen S, Jensen AL. Recommendations for
designing and conducting veterinary clinical pathology biologic variation studies. Vet Clin
Pathol. 2017; 46(2): 211-220.
5. Clinical and Laboratory Standards Institute (CLSI). User verification of performance for
precision and trueness; approved guideline. 2nd ed. (EP31-A-1R). Wayne, PA: AALA; 2005.
6. Clinical and Laboratory Standards Institute (CLSI). Using proficiency testing to improve the
nd
clinical laboratory; approved guideline, 2 ed. (GP27-A2). Wayne, PA: AALA; 2008.
7. Clinical and Laboratory Standards Institute (CLSI). Evaluation of precision performance of
nd
quantitative measurement methods; approved guideline, 2 ed. (EP15-A2). Wayne, PA:
AALA; 2006.
8. Clinical and Laboratory Standards Institute (CLSI). User verification of precision and
estimation of bias; approved guideline. 3rd ed. (EP15-A3). Wayne, PA: AALA; 2014.
9. OIE-World Organisation for Animal Health: Manual of Diagnostic Tests and Vaccines for
Terrestrial Animals. 2014. Available at:
http://www.oie.int/fileadmin/Home/eng/Health_standards/tahm/0.04_GLOSSARY.pdf.
Accessed October 25, 2016.
10. Friedrichs KR, Harr KE, Freeman KP, et al. ASVCP reference interval guidelines:
determination of de novo reference intervals in veterinary species and other related
topics. Vet Clin Pathol. 2012; 41(4): 441-453.
Page 17 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
Page 18 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
27. Camus MS, Flatland B, Freeman KP and Cruz Cardona, JA. ASVCP quality assurance
guidelines: external quality assessment and comparative testing for reference and in-clinic
laboratories. Vet Clin Pathol. 2015; 44: 477–492.
28. Clinical and Laboratory Standards Institute (CLSI). Verification of comparability of patient
results within one health care system; approved guideline. (C54-A). Wayne, PA: AALA;
2008.
29. Westgard JO. Implementing repeat patient test controls. In: Westgard JO, ed. Basic QC
Practices Training in Statistical Quality Control for Medical Laboratories, 4th ed. Madison,
WI: Westgard QC; 2016: 233-244.
30. Cook AM, Mortiz A, Freeman KP, Bauer N. Objective evaluation of analyzer performance
based on a retrospective meta-analysis of instrument validation studies: point-of-care
hematology analyzers. Vet Clin Pathol. 2017. 46(2): 248-261.
31. Vap LM, Harr KE, Arnold JE, et al. ASVCP quality assurance guidelines: control of
preanalytical and analytical factors for hematology for mammalian and nonmammalian
species, hemostasis, and crossmatching in veterinary laboratories. Vet Clin Pathol. 2012;
41(1): 8-17.
32. Kjelgaard-Hansen M, Jensen AL. Is the inherent imprecision of manual leukocyte differential
counts acceptable for quantitative purposes? [letter]. Vet Clin Pathol. 2006; 35: 268-270.
33. Lilliehöök I, Tvedten H. Validation of the Sysmex XT‐2000iV hematology system for dogs,
cats, and horses. I. Erythrocytes, platelets, and total leukocyte counts. Vet Clin Pathol.
2009; 38(2): 163-174.
34. Tvedten H, Moritz A. Reticulocyte and Heinz body staining and enumeration. In: Weiss
DJ, Wardrop KJ, eds. Schalm’s Veterinary Hematology. 6th ed. Ames, IA: Blackwell
Publishing Ltd; 2010: 1067-1073.
35. Bayat H. Westgard Web: QC-dependent risk reduction. 2015. Available at:
www.westgard.com/qc-risk-reduction.htm. Accessed September 7, 2017.
36. Clinical Laboratory Improvement Amendments (CLIA): Proficiency testing criteria for
acceptable analytical performance. Fed Reg. 1992; 57(40): 7002-7186.
37. Quam E. Selecting the right control materials. In: Westgard JO. Basic QC Practices
Training in Statistical Quality Control for Medical Laboratories, 4th ed. Madison, WI:
Westgard QC; 2016: 103-111.
38. Bellamy JEC, Olexson DW. Quality Assurance Handbook for Veterinary Laboratories.
Ames, Iowa: Iowa University Press; 2000: 26.
39. Furlanello T, Tasca S, Caldin M, et al. Artifactual changes in canine blood following
storage, detected using the ADVIA 120 hematology analyzer. Vet Clin Pathol. 2006; 35(1):
42-46.
40. Rishniw M, Pion P. Evaluation of performance of veterinary in-clinic hematology analyzers.
Vet Clin Pathol. 2016; 45(4): 604-614.
41. VetBiologicalVariation: Database tables. 2015. Available at:
http://vetbiologicalvariation.org. Accessed September 7, 2017.
Page 19 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
42. Oosterhuis WP. Gross overestimation of total allowable error based on biological variation.
Clin Chem. 2011; 57: 1334-1336.
43. Bourgès-Abella NH, Gury TD, Geffré A, et al. Reference intervals, intraindividual and
interindividual variability, and reference change values for hematologic variables in
laboratory beagles. J Am Assoc Lab Anim Sci. 2015; 54: 17-24.
44. Wiinberg B, Jensen AL, Kjelgaard-Hansen M, et al. Study on biological variation of
haemostatic parameters in clinically healthy dogs. Vet J. 2007; 174: 62-68.
45. Jensen AL, Iversen L, Petersen TK. Study on biological variability of haematological
components in dogs. Comp Haematol Int. 1998; 8: 202-204.
46. Lilliehöök I, Tvedten HW. Errors in basophil enumeration with 3 veterinary hematology
systems and observations on occurrence of basophils in dogs. Vet Clin Pathol. 2011; 40(4):
450-458.
47. Jensen AL, Kjelgaard-Hansen M. Method comparison in the clinical laboratory. Vet Clin
Pathol. 2006; 35: 276-286.
Page 20 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
8 Web resources
Glossary of QA terms, Westgard QC glossary:
http://www.westgard.com/glossary.htm
CLIA website:
https://www.cms.gov/Regulations-and-Guidance/Legislation/CLIA/index.html
Biological variation:
http://vetbiologicalvariation.org
Page 21 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
Amelia Woolums, DVM, MVSc, PhD, DACVM, DACVIM (Large Animal Internal Medicine)
Professor
Department of Pathobiology and Population Medicine
College of Veterinary Medicine
Mississippi State University
Starkville, MS
10 Contributing Laboratories
• Department of Biomedical and Diagnostic Sciences, University of Tennessee, Knoxville, TN
• Department of Comparative Medicine, Animal Diagnostic Laboratory, Stanford University,
Stanford, CA
• Department of Microbiology, Immunology and Pathology, Colorado State University, Ft.
Collins, CO
• Department of Pathology, University of Georgia, Athens, GA
• Department of Veterinary Pathobiology, Texas A&M University, College Station, TX
• Department of Veterinary Pathobiology, University of Missouri, Columbia, MO
• Urika, LLC, Mukilteo, WA owned by Dr. Harr
11 Acknowledgments
The authors thank Drs. Roberta Moorhead and Marlyn Whitney for their contribution of
hematology data.
The authors also thank members of the ASVCP Executive Board and the following individuals
for their thorough review of the manuscript: Ms. Jill Arnold and Drs. Erica Behling-Kelly, Jean-
Pierre Braun, Jennifer Cook, Glenn Frank, Kathy Freeman, Kristen Friedrichs, Luca Giori,
Emma Hooijberg, Kate Irvine, Unity Jeffery, Ernst Leidinger, Tracy Stokol, and Harold Tvedten.
Page 22 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
12 Tables
Table 1. Allowable total error (TEa) for automated hematologic measurands, applicable for all
concentrations (low/normal/high).a See Appendix A for derivation of TEa values.
bCLIA
Measurand TEa Value
RBC 10% 6%
Hgb 10% 7%
Hct/PCV 10% 6%
MCV 7% ---
MCHC 10% ---
Reticulocytesc 20%d ---
WBC 15%e 15%
(Reference laboratory)
WBC 20% ---
(*In-Clinic laboratory)
Platelets 20%f 25%
(Reference laboratory)
Platelets 25% ---
(*In-Clinic laboratory)
human medicine in the United States and is therefore included here for comparison. 36
cThis recommendation is only applicable to canine absolute reticulocyte counts. Feline
Page 23 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
Table 2. Allowable Total Error (TEa) for Differential Cell Counts. TEobs for differential cell
counts should be based on absolute numbers and may be generated using automated techniques,
manual techniques, or a combination thereof (see Section 4.2.2). See Appendix A for derivation
of TEa values and example calculations.
Page 24 of 34
Table 3 Comparison of Selected Biological Variation Data and ASVCP TEa Recommendations
ASVCP = American Society for Veterinary Clinical Pathology; CVg = interindividual biological variation; CVi = intraindividual
biological variation; Des = desirable; HCT = hematocrit; HGB = hemoglobin; MCHC = mean cell hemoglobin concentration; MCV =
mean cell volume; Min = minimum; Neut = neutrophils; Opt = optimum; PCV = packed cell volume (spun HCT); PLT = platelets;
RBC = red blood cells; RDW = red blood cell distribution width; TEa = allowable total error; WBC = white blood cells;
aRBC, WBC, and HGB were determined using a model S560 Coulter Counter; HCT was additionally measured manually using a
Haemofuge (a microhematocrit centrifuge). BV-based quality specifications given in these tables were calculated using published
biological variation data (as cited) and formulae for optimum, desirable, and minimum thresholds from Fraser, 2001. 15
bReference laboratories
cIn-clinic laboratories using point-of-care instrumentation.
Page 26 of 34
13 Appendices
variety of specialties based on maximal allowable deviation from a given result that
would not affect their clinical decision.25 In our experience, the range of clinician
expectations was initially quite broad and included small errors that were often
unattainable with current instrumentation. However, a consensus was eventually reached
that was ultimately more aligned with instrument performance. Cook, et al, observed a
much wider decision threshold range, possibly because input was provided without
discussion between clinicians and clinical pathologists.25
As an example of the discussions, consider desired total error for neutrophil
enumeration (Example 1, below). Clinicians, particularly oncologists, often base
treatment decisions on a low neutrophil concentration of 2000 cells/µL. Clinicians were
asked how much error they could tolerate in making clinical decisions for a patient
having a “true” concentration of 2000 neutrophils/µL, and a maximum acceptable range
of error around this true value was identified as 1800 to 2200 /µL, or 200 /µL.
Expressed as a percentage, this degree of error is 10%. While this small degree of error
is not attainable with current instrumentation, it provides a goal for manufacturers.
Page 28 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
clinically important values unrelated to a reference interval. This clinician input was
considered in combination with instrument performance to determine the recommended
TEa (see below).
Therefore, the desired TEa for a moderately low platelet concentration was
calculated as 15%.
Page 29 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
Page 30 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
Table A1a. Derivation of TEa recommendations including comparison of total allowable error
(TEa) with total observed error (TEobs), clinician error limit, and Clinical Laboratory
Improvement Amendments (CLIA) for hematology measurands.
Page 31 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
Table A1b. Levels/concentrations of historical QCM used by the institutions that provided data
for this study and the interval of TEobs for each using reference laboratory instrumentation.
*Note that a recommendation for TEa for low reticulocyte concentrations was not deemed
clinically relevant.
Page 32 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
Vendors and manufacturers are listed alphabetically. This list is for informational purposes only
and does not constitute a legal contract or endorsement between ASVCP and any person or entity
unless otherwise specified. The ASVCP does not endorse any particular vendor or manufacturer.
CLIA
https://www.cms.gov/Regulations-and-
Guidance/Legislation/CLIA/Proficiency_Testing_Providers.html
Page 33 of 34
ASVCP QALS TEa Hematology Version 1.0 (final approved) December 2017
𝑺𝒕𝒂𝒏𝒅𝒂𝒓𝒅 𝒅𝒆𝒗𝒊𝒂𝒕𝒊𝒐𝒏
𝑪𝑽 (%) = × 𝟏𝟎𝟎
𝑴𝒆𝒂𝒏
𝑴𝒆𝒂𝒏𝒕𝒂𝒓𝒈𝒆𝒕 − 𝑴𝒆𝒂𝒏𝒎𝒆𝒂𝒔𝒖𝒓𝒆𝒅
𝑩𝒊𝒂𝒔(%) = × 𝟏𝟎𝟎
𝑴𝒆𝒂𝒏𝒕𝒂𝒓𝒈𝒆𝒕
3. Calculate Total Error (TEobs) using the following formula. Use the absolute bias result
(i.e., positive number only).
𝑻𝑬𝒐𝒃𝒔 = 𝒂𝒃𝒔𝒐𝒍𝒖𝒕𝒆 𝒃𝒊𝒂𝒔% + 𝟐𝑪𝑽
4. Compare TEobs calculated in the laboratory with TEa found in the ASVCP Guidelines for
TEa.
a. If TEobs ≤ TEa
i. The quality assessment passes and no further action is needed
b. If TEobs > TEa
i. Report results to Quality Assurance Personnel/Committee.
ii. Investigate pre-analytical and procedural factors (e.g., specimen quality,
instrument SOP, operator proficiency, bias determination) that may have
impacted performance and correct as needed. Reassess instrument
performance following correction.
iii. If no pre-analytical or procedural factors are identified that can be
addressed and corrected, management should report findings to the
manufacturer of the instrument so that any needed maintenance, repairs, or
replacement may be evaluated and implemented.
iv. Clinicians should be notified in writing of potentially clinically impactful
error.
5. All Total Error assessments should be catalogued in a written and/or digital archive
accessible to personnel who may operate the instrument or interpret the results.13
Page 34 of 34