The document provides specifications for the Fmc2 Flow Management Computer. The Fmc2 is a flexible, user-programmable instrument that continuously monitors and controls liquid and gas flows. It includes interfaces for remote communications, temperature/pressure monitoring, loop control, reporting, and output to other equipment. Key features include a touchscreen interface, support for multiple languages and standards, automatic proving capabilities, and Modbus interfaces. It can operate as a standalone computer or integrate with supervisory systems.

Uploaded by

bahaaahmed36
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
71 views4 pages

Ss 09042

The document provides specifications for the Fmc2 Flow Management Computer. The Fmc2 is a flexible, user-programmable instrument that continuously monitors and controls liquid and gas flows. It includes interfaces for remote communications, temperature/pressure monitoring, loop control, reporting, and output to other equipment. Key features include a touchscreen interface, support for multiple languages and standards, automatic proving capabilities, and Modbus interfaces. It can operate as a standalone computer or integrate with supervisory systems.

Uploaded by

bahaaahmed36
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

SPECIFICATIONS

Fmc2™ Flow Management Computer


Bulletin SS09042 Issue/Rev. 0.2 (09/10)

Flow Management Computer


Fmc2™ Flow Management Computer is a flexible,
extremely robust, user programmable and configurable
microprocessor-based instrument designed to continuously
monitor and control flow of liquid and gas petroleum
products.

The Fmc2™ can act as a high integrity, stand-alone flow


computer or as a powerful building block for integration
into a supervisory computer control system. The instrument
includes interfaces for remote communications, tempera-
ture, pressure and density monitoring, process loop
control, report printing, and the capability of providing
analog and digital outputs to remote equipment. Main Screen Display

Touch Screen Interface

Features

• Flexible and scalable Field Input/Output Blocks (FIOBs) • Full PI (Proportion/Integral) process loop control
• Graphical and menu driven Human Machine Interface functions, with selectable secondary variable for use
(HMI) via touch screen; unprecedented in the flow in override situations (i.e. back pressure control);
computer market bumpless transfer between modes
• Multi-language, user configurable text for displays, • Standard report templates, as well as user-configu-
alarms/events and reports (English, Russian, Spanish, rable report templates. Archival of 100 records of each
Portuguese, and French) report type
• All major current and historical standards for API and • Intuitive, efficient and secure file transfer to/from
AGA (Including measurement of Ethanol per Brazilian the Fmc2 via standard USB flash drive or to a
specifications) shared network folder (Archived reports, alarms,
• Automatic proving support for displacement and master configurations, etc.)
meter provers with selectable automatic meter factor • Standard as well as user-configurable menu system
acceptance • Graphic trending of process variables
• Advanced Statistical Control of the meter factor • Onscreen viewable reports
• Multi-manufacturer frequency densitometer inputs • Graphic alarm display screen
• User-configurable Modbus Master interface – can be • Integrated Meter Run Motor Operated Valve (MOV)
used to collect data from any Modbus Slave device Control and Sequencing (Online/Offline, Onprove/
(Gas Chromatographs, Ultrasonic Meters, other flow Offprove)
computers, etc.) • Optional Remote HMI (RHMI) Software for remote
• Twelve (12) point meter factor linearization monitoring and control
• User-configurable Modbus Slave interface to host or • Flow computer Dual and Triple Redundancy support
SCADA systems, or any Modbus Master device • Condition-Based Monitoring (CBM) of all process
variables
Applications Field Input/Output Block (FIOB) (MTR-AI-FI)
Dimensions:
• Pipeline Deliveries
14.76" (375 mm) W
• Batch Deliveries 4.75" (120.65 mm) H
• Proving (Displacement/Master Meter) 3.5" (89 mm) D
Weight:
• Easily manages most fiscal-type displacement and
3.9 lb (1.77 kg)
inference meters (e.g. PD, conventional turbine, helical
turbine, ultrasonic, orifice)
• Gas and Liquid: One software version supports MTR-AI-FI
applications for gas or liquid measurement systems, as H FIOB
well as all measurement units (U.S. Customary or Metric), D
W
or any combination of liquid measurement units. A single
Fmc² can support mixed fluid types per meter run (i.e.
two meter runs measure liquid products, two meter runs Electrical/Environment Specifications
measure gas products, within the same Fmc²) Field Input/Output Block (FIOB)
Supply Voltage: 24 VDC, +20/-15%
Typical Current Consumption (meter run): 24VDC, 3A max
Operating Temperature 0°C to 55°C (32°F to 131°F)
Storage Temperature -25°C to 75°C (-13°F to 167°F)
Relative Humidity 75% Non-Condensing
Suitable for installation in ordinary locations.

HMI Display Unit


Supply Voltage: 24 VDC Current Consumption: 5A max
Operating Temperature 0°C to 50°C (32°F to 122°F)
Storage Temperature -25°C to 55°C (-13°F to 131°F)
Relative Humidity 75% on average, 85% occasionally; no
condensation

Meter Stream Display Communications Connectivity


1 EIA-232 DB9 serial port, up 115k Baud 1 EIA-422/485
DB9 serial port, up 115k Baud
Technical Specifications 2 Ethernet RJ45 ports, 10/100 Mbit/s
Process Input/Output User configurable Modbus Master/Slave interface via serial
Maximum 6 dual-pulse meter runs, with a combination of or Ethernet. (Modbus ASCII/RTU/TCP)
Analog Input/Output modules and RTD input modules, with
Master Meter, Bidirectional prover, and frequency densitom-
eter inputs. Input/Output Specifications
Human Machine Interface Frequency Module
8 inch, panel mounted touch screen interface
Meter Pulse/Density Inputs
16-bit color 800 x 600 screen resolution
Type: High-speed, edge-triggered, voltage-sourcing pulse
32-bit CPU processing power transmitter input (optically-isolated), according to DIN EN
Panel Cutout Dimensions: 61131-2 Type 1.
8.69" (221mm) W, Meter pulse fidelity checking and correction per IP Level A
7.25" (184mm) H or Level B.
Weight: 24VDC Pulse Inputs
3.9 lb (1.77 kg)
Frequency Range: 0 to 10.0 kHz

2 Bulletin SS09042 Issue/Rev. 0.2 (09/10)


Output load: 500 Ohms (current, maximum), 2 kOhms (volt-
age, minimum)
Output Resolution: >14 Bit
Current Output: Short Circuit Proof
See Note 1

RTD Inputs
Type: Resistive temperature sensors input, Pt100, Ni, Cu,
10 – 3000 Ohm
Connection: 3 or 4-wire, shielded sensor cable
Accuracy: +/-0.2°F, +/-0.83 maximum (Pt100, 4-wire)
Measurement Resolution: 14 Bit plus a sign bit
Prover Display
Signal Supply Voltage: Short Circuit Proof
See Note 1

Digital Inputs
Type: Optically-isolated solid-state voltage sensor inputs
Input Voltage Range: -3VDC to + 30VDC (24 VDC typcal)
Maximum low level voltage: < 5 VDC
Minimum high level voltage: > 15 VDC
Signal Supply Voltage: Short Circuit Proof
See Note 1

FIOB (Field Input/Output Block)


Digital Outputs
Type: Electrically-isolated relay contact output
Pulse Outputs Connection: Form C relay contact output
Type: Optically-isolated, 24VDC, 6 watt solid state output Maximum Switching Voltage: 250 VDC/VAC
Pulse width/output type software configurable Output Current: 3A Maximum
See Note 1
Analog Inputs
Single-ended signal inputs for the connection of either voltage Approvals/Certifications/Listings
or current signals. • OIML R-117-1, “Measuring systems for liquids other than
Measurement ranges: 0 mA to 20 mA, 0 VDC to 10 VDC water”
Connection: Shielded pair (two conductors with shield) • EN-12405-1(/A1), “Electronic gas volume conversion
Accuracy: +/- 0.02% of range device”
Measurement resolution: > 14 Bit • GOST (Russian) Pattern Approval
Signal Supply Voltage: Short Circuit Proof • UL/CE
See Note 1
• FCC (display)
Analog Outputs
Type: Outputs to connect either current or voltage outputs
Current range: 0 mA to 20 mA
Voltage range: 0 VDC to 10 VDC
Connection: Shielded Pair (Two conductors with shield)
Accuracy: +/- 0.008% of range (voltage), +/-0.01% of range
(current)

Note 1: Analog / RTD / Digital point function programmable via software.

Bulletin SS09042 Issue/Rev. 0.2 (09/10) 3


Application Example
Typically Instrumented 3-Meter Run, Bidirectional Prover and Quality Loop System

Notes:
1: "DP" = Differential Pressure; may be a transmitter or switch for strainers, switch for 4-way seal
2: "DT" = Densitometer (frequency or analog)
3: "AE" = Analyzer Element (S&W)
4: Sample Controller may be driven by integrated pulse output on the Frequency Module

The specifications contained herein are subject to change without notice and any user of said specifications should verify from the manufacturer that the specifications are currently in
effect. Otherwise, the manufacturer assumes no responsibility for the use of specifications which may have been changed and are no longer in effect.

USA Operation
1602 Wagner Avenue
Erie, Pennsylvania 16510 USA
P:+1 814.898.5000

TechnipFMC Germany Operation


13460 Lockwood Road Smith Meter GmbH
TechnipFMC.com Building S01 Regentstrasse 1
Houston, Texas 77044 USA 25474 Ellerbek, Germany
© 2020 TechnipFMC plc. All rights reserved SS09042 Issue/Rev. 0.2 (09/10) P:+1 281.591.4000 P:+49 4101.304.0

You might also like